6-Gas Chromatography-Mass Spectrometry (GC-MS) - Working Principle and Applications.
6-Gas Chromatography-Mass Spectrometry (GC-MS) - Working Principle and Applications.
1. Introduction
Gas chromatography-mass spectrometry (GC-MS) is a hyphenated technique
consisting of two analytical procedures coupled in series that is a Gas Chromatography
(GC) for separation of volatile compounds and Mass Spectrometry (MS) for
identification and quantification of molecular weight of each separated component. The
GC helps in separation of multiple components mixture so that they can reach the mass
spectrometry one at a time. Thus, the molecular mass and other fragments mass values
describe only for a single component at a time.
1
Fig. 1. GC-chromatogram
Electron-capture detector (ECD): ECD is sensitive like FID, but has a limited
dynamic range. Mostly used in the analysis of electronegative atoms containing organic
molecules such as halogens, phosphorous and nitro groups.
Amobile Astationary
Where, [CM] is molar concentration of the analyte in the SP and [CM] is molar
S M
concentration in the MP. If K is constant over a wide range of analyte concentrations,
then [CM] is directly proportional to [CM] and chromatographic peak is symmetrical,
S M
Gaussian distribution and retention time are independent of the amount of analyte
injected.
Retention Times of an analyte is defined as the time it takes after sample injection for
the analyte to reach the detector and elute. The time for non-interacted species with SP to
reach the detector is defined as the dead time (tD), if taken in volume called dead volume
(VD). Thus.the rate of migration of a non-interacted species is the same as the rate of
motion of the mobile phase expressed as,
Rate = L / t
M D ………………….(i)
Where, L is the column length and t is the retention time of the mobile phase/
D
non-interacted species.
Similarly, the linear rate of a solute molecule is calculated by dividing the column
length by the retention time t ,
R
The linear rate of a solute molecule can be expressed as a function of the rate of
migration of the unretained species,
Where, S (t) is the fraction of time the solute spends in the mobile phase.
M
HETP = L / N …….………….(iv)
Where, W is the peak width measured in the same units as t and W is the peak
R 1/2
width measured at half the peak height.
The resolution of a column is defined as its ability to separate a mixture of
compounds. For example, compounds Ψ and Θ were separated by the GC column as
shown in Fig. 3.
Fig. 3. GC-chromatogram
Determining the column resolution: After some mathematical manipulation, the
required parameters to compute the column resolution is as follow:
R = (Θt
- Ψt ) / (W + W ) = Δt / (W + W ) …………(vi)
S R R Θ Ψ R Θ Ψ
The chromatogram generally plotted in detector signal vs. time, gave Gaussian-
shaped (or bell shaped) peaks. In reality, non-Gaussian peak shapes will often occur. The
front side of a peak might be drawn out while the tail on the right is steep called fronting,
might be due to large amount of a sample is introduced into the column or wrong
combination of packing material. The right side of the peak is drawn out while the tail on
the left is steep called tailing and usually occurs when the solute has a concentration
dependent distribution coefficient. Fronting and tailing will result in less accurate
quantitative analysis and peaks broadening.
Equipment calibration: The GC-MS instrument needed to be tuned before each
analysis. The calibration consists of injecting a known volume of a standard and
measuring the time between injection and elution. The retention time of a compound
remained the same for a given set of variables (temperature, flow rate and column
length). If the quantitative analysis is required, it might need to compare peak heights or
areas of the peak with that of one of the standards. The most ancient method for
quantitative analysis is the internal standard calibration method (ISTD) and involves
the preparation of a series of standards that approximate the composition of unknown. In
this procedure, a measured quantity of an internal standard is introduced into each
standard and sample. Chromatograms for the standards are generated and peak areas are
plotted as a function of analyte amount or concentration. Such plot should yield a straight
line passing through the origin and gives qualitative and quantitative analysis of different
components present in the mixture (Fig. 4).
Fig. 4. An ISTD calibration curve (2.5 area counts = 1.0 units of compound amount)
Instrumentation: All mass spectrometers have three major components (Fig. 6):
(i) Ionizer (ion source), (ii) Ion/mass Analyzer and (iii) Detector
Fig. 6. MS instrument
0 -3
(i) Ionizer: A sample vaporized at >400 C and 10 Pa pressure is introduced in the mass
spectrometer where molecule get ionized using different ionization methods that include
electron impact (EI), In-beam (desorption) electron impact (DEI), Chemical ionization
(CI), Alternate CI-EI (ACE), Rapid heating (desorption) Chemical ionization (DCI),
Atmospheric pressure ionization (API), Atmospheric pressure chemical ionization
(APCI), Electrospray ionization (ESI), Thermospray ionization (TSP), Laser ionization
(Laser desorption), Matrix assisted Laser ionization (MALDI), Field ionization (FI) and
Field desorption (FD), Fission fragment ionization or Plasma desorption mass
spectrometry (PDMS), Secondary ion mass spectrometry (SIMS), Liquid SIMS, Fast
atom bombardment (FAB), Photon and multiphoton ionization (MPI) and Resonance
Ionization, etc. Mostly, MS ionizers work with positive ions (negative is rare). For
example, in EI-MS technique, 70 eV electron energy is used for molecule ionization
where atom knocked out one or more electrons off to give a positive ion (molecular ion)
and lower the electron energy. The high energy radical cation (mol. ion) undergoes
further fragmentation to minimize its energy (Scheme 1).
General reaction:
70 eV
- .+ -
A-B-C (vapor) + e ------------Æ A-B-C + 2e
(radical cation)
In the ionization chamber, the vaporized molecules are bombarded with a stream
of electrons emitted from electrically heated metal coil. Some of the collisions are
energetic enough to knock out one or more electrons to make positive ions. Mostly, the
positive ions formed carry single charge because it becomes much more difficult to
remove further electrons from a positive ion. The free electron(s) are attracted towards
the electron trap (+vely charged plate called electron collector). The ions produced in the
ionization chamber travel forward direction only due to repeller ((+vely charged plate)
placed from other side. In ion source and mass analyzer sections, a high vacuum is
required to avoid high energy electrons to strike with air molecules (Fig. 7).
The positive ions are repelled away from the very positive ionization chamber and
pass through three slits, having potential difference in decreasing order from + 10,000 to
0 volts. Therefore, all the ions are accelerated and have the same kinetic energy at the end
of 0 volts (Fig. 8).
In the mass analyzer, the accelerated ions are then entering into mass separation
unit, depending upon their mass/charge (m/z or m/e) ratio. For example, magnetic sector
analyzer (MSA) deflects ions by an electromagnetic field according to their masses (Fig.
9). Lighter ions are deflected more compared to heavier. The amount of deflection also
depends on the number of positive charges on ion. The more the ion is charged, the more
it gets deflected.
In Fig. 9, ion stream A is most deflected because ions have minimum mass/charge
ratio. Ion stream C is the least deflected because ions have maximum mass/charge ratio,
assuming all are +1 ions. Thus, stream A has the lightest ions, stream B the next lightest
and stream C the heaviest. Lighter ions are going to be more deflected than heavy ones.
Only stream B ions will reach to the detector in this case. For stream A ions to be
recorded, we need less electromagnetic field to be applied. Conversely, for stream C ions
to be recorded, we need stronger electromagnetic field to be applied which can be done
by changing the magnetic field.
(iii) Detector: Only stream B ions reach to the ion detector. The other ions collide with
the walls where they will pick up electrons and be neutralized (uncharged). Finally, they
get removed from the mass spectrometer by the vacuum pump. Ions detection is based on
either charge or momentum value. In modern instruments, for large signals a Faraday cup
is used to collect ions and measure the current. In old instruments, photographic plates
are used to measure the ion abundance at each mass/charge ratio. Most detectors used
amplifying the ion signal using a collector similar to a photomultiplier tube. These
amplifying detectors include: electron multipliers, channeltrons and multichannel plates.
The gain is controlled by changing the high voltage applied to the detector. A detector is
selected for its speed, dynamic range, gain and geometry. Some detectors are sensitive
enough to detect single ions.
When an ion hits the metal surface, its charge is neutralized by an electron
jumping from the metal on the ion (right hand diagram). That leaves a space amongst the
electrons in the metal and the electrons in the wire shuffle along to fill it. A flow of
electrons in the wire is detected as an electric current which can be amplified and
recorded through computer software. Larger the ions knocking, the greater will be the
current which indicate the intensity of the peak.
The output from the chart recorder is usually simplified into a "stick diagram".
This shows the relative current produced by ions of varying mass/charge ratio.
(a) (b)
1.3 Applications
GC-MS is used as the tool of choice for tracking organic pollutants in the
environment. There are some compounds for which GC-MS is not sufficiently sensitive,
including certain pesticides and herbicides, but for most organic analysis of
environmental samples, including many major classes of pesticides, it is very sensitive
and effective.
GC-MS can analyze the particles from a human body in order to help link a
criminal to a crime. The analysis of fire debris using GC-MS is well established. There is
even an established American Society for Testing Materials (ASTM) standard for fire
debris analysis. GCMS/MS is especially useful here as samples often contain very
complex matrices and results, used in court, need to be highly accurate.
GC-MS is used for detection of illegal narcotics and supplant drug-sniffing dogs.
It is also commonly used in forensic toxicology to find drugs and/or poisons in biological
specimens of suspects, victims, or the deceased.
GC-MS is the main tool used in sports anti-doping laboratories to test athletes'
urine samples for prohibited performance-enhancing drugs, for example anabolic
steroids.
1.3.7 Astrochemistry
Several GC-MS have left earth. Two were brought to Mars by the Viking
program. Venera 11 and 12 and Pioneer Venus analyzed the atmosphere of Venus with
GC-MS. The Huygens probe of the Cassini-Huygens mission landed one GC-MS on
Saturn's largest moon, Titan. The material in the comet 67P/Churyumov-Gerasimenko
will be analyzed by the Rosetta mission with a chiral GC-MS in 2014.
1.3.8 Medicine
References
-----