Optical Density (Radiographic Density) : T o T o o T
Optical Density (Radiographic Density) : T o T o o T
Relationship of the Optical Density of Radiographic Film to Light Transmission Through the Film
Percent of Light Transmitted(It/Io x 100) Fraction of Light Transmitted(It/Io) Optical Density(log Io/It)
100 1 0
50 1/2 0.3
32 8/25 0.5
25 1/4 0.6
12.5 1/8 0.9
10 1/10 1
5 1/20 1.3
3.2 4/25 1.5
2.5 1/30 1.6
1.25 1/80 1.9
1 1/100 2
0.5 1/200 2.3
0.32 2/625 2.5
0.125 1/800 2.9
0.1 1/1000 3
0.05 1/2000 3.3
0.032 1/3125 3.5
0.01 1/10000 4
*Sensitometry is the study of the relationship between the intensity of exposure of the film and the blackness after processing.
*The two principal measurements involved in sensitometry are the exposure to the film and the percentage of light transmitted
through the processed film. Such measurements are used to describe the relationship between OD and radiation exposure. This relationship is called
characteristic curve.
*Characteristic curve or sometimes the H & D curve after Hurter and Driffield, who first described this relationship.
*Two pieces of apparatus are needed to construct a characteristic curve: an optical step wedge, sometimes called a
sensitometer, and a densitometer, a device that measures OD.
▪ Black is numerically equivalent to an OD of 3 or greater, clear is less than 0.2.At and OD of 2,only 1% of viewbox light
passes through the film.
▪ Is determined by dividing the incident light intensity by the amount of light transmitted through the process film.
*Most unexposed and processed radiographic film has an OD in the range of 0.1 to 0.3, corresponding to 79% and 50%transmission, respectively.
These ODs of unexposed film are attributable to base density and fog density.
*Base density is the OD that is inherent in the base of the film. It is attributable to the composition of the base and the tint added to the base to make the
radiograph more pleasing to the eye.Base density has a value of approximately 0.1.
*Fog density is the development of silver grains that contain no useful information.Fog density results from inadvertent exposure of film during
storage, undesirable chemical contamination, improper processing, and a number of other influences. Fog density on a processed radiograph should
not exceed 0.1.
*Base and fog densities reduce radiographic image contrast and should be as low as possible.
*The useful range of OD is approximately 0.25 to 2.5. Most radiographs, however, show image patterns in the range of 0.5 to 1.25 OD.
*A(overexposed),B(underexposed).Both chest radiographs are unacceptable because no detail to the lung field is apparent.
screens.
*The reciprocity law fails for screen-film exposures at exposure times less than approximately 10 ms or longer than approximately 2s.
*The reciprocity law is important for special procedures that require very short or very long exposure times, such as angiointerventional
radiography and mammography, respectively. For these few situations, increasing the mAs setting may be required if automatic exposure control
2. Distance
Source-to-image receptor distance (SID)
▪ Distance affects exposure of the image receptor according to the inverse square law.
▪ Refers to the distance from the x-ray source to the image receptor(IR).
▪ Affects the number of x-ray photons reaching the IR when as it changes.
*SID and density are inversely proportional: As SID increases, density will decrease increase.
▪ Affects radiation intensity: As SID increases, radiation intensity decreases; As SID decreases, radiation intensity increases.
This is known as the inverse square law.
*The inverse square law states that the intensity of radiation is inversely proportional to the square of the distance from the x-
ray source, or
* However, the most common situation in radiography is a need to maintain an acceptable density while changing the distance.To maintain density,
mAs for the density change. The density maintenance formula is used for this purpose. This formula is based on the inverse square law but it is reversed
to a dricet square law because the mAs must increase when distance increases, and vice versa, in order to maintain density.
The distance between the image receptor and the object being imaged.
An increase in OID decreases density by creating space for scatter
radiation to escape before reaching the image receptor. If less scatter
reaches the image receptor, less density is created on the image.
An air-gap technique increases object-to-film distance, and this permits
scatter radiation to avoid the image receptor. This scatter radiation would
normally contribute radiation fog to the density.
3. Anatomical Part
▪ The patient is the prime attenuator of the beam, the anatomical part being examined has a great deal of influence on the
density of the film. The amount of attenuation is dependent on the thickness and type of the tissue being radiographed. The
tissue type is affected by the average atomic number and the density(quantity of matter per volume) of the tissue. The use of
contrast media will alter the average atomic number of the tissue and can affect density. Pathology can alter tissue thickness
and/or type.
*There is an inverse relationship between tissue thickness/type and radiographic density. In other words, as tissue thickness, average atomic number of
the tissue, and/or tissue density increases, radiographic density decreases.
Classifying Pathology
4. Film Processing
▪ The condition of the film processing solutions can dramatically alter density. Density will increase when the developer
solution temperature increases, immersion time increases, or replenishment rates increase. Density will decrease when the
above factors decrease as well as when contamination decreases solution strength.
Developer solution Temperature: Maintained at 35˚C (95˚F)
Immersion Time :Manual-5 min, Automatic-22s
Replenishment Rate : Approximately 60 to 70 mL of developer and 100 to 110 mL of fixer for every 35 cm (14 in)
of film.
▪ The development time recommended by the manufacturer is the time that will result in maximum contrast. When
development time extends far beyond the recommended period, the image receptor contrast decreases.
6. Collimation
▪ Collimation is the use of any beam restricting device to decrease the field of view of the image and exposure area of the
patient.
▪ Beam restrictors reduce the amount of scatter that is produced.
▪ Every technologist should follow the principle of ALARA (As Low As Reasonably Achievable) which includes collimating
only to the necessary area of interest.
▪ Collimation reduces patient dose and decreases scatter radiation.
▪ Beam restriction, or collimation, changes the size of the x-ray field and, therefore, the volume of tissue being exposed to
radiation. As collimation increases, the size of the light field and the volume of tissue irradiated are decreased. Increasing
collimation results in decrease scatter production and, as a result, decreased optical density.
*Increasing collimation deceases optical density.
*Decreasing collimation increases optical density.
7. Grids
▪ The radiographic grid was invented in 1913 by Gustave Bucky and continues to be the most effective means for limiting the
amount of scatter radiation that reaches the IR."
▪ Grids absorb scatter which would otherwise add density to the film. The more efficient the grid, the less the density. Grid
with high ratios, low frequency, and dense interspace material; moving grids and improperly used grids all reduce density.
▪ Compensation for varying grid ratios is generally accomplished by increasing mAs. The amount of mAs needed can be
calculated using the grid conversion factors. Changes between grids, which is the most common clinical problem, are
accomplished by using the following formula:
mAs1 GCF 1
=
mAs2 GCF 2
where: mAs1=old mAs
mAs2 = new mAs
GCF1 = original grid conversion factor
GCF2 = new grid conversion factor
Bucky Factor at
No grid 1 1 1
5:1 2 2.5 3
8. Filtration 8:1 3 3.5 4
▪ Three types of x-ray filtration are used: inherent, added,and
compensating. All x-ray beams 12:1 3.5 4 5 are affected by the inherent filtration
properties of the glass or metal envelope of the x-ray tube. For general-
16:1 4 5 6
purpose tubes, the value of inherent filtration is approximately 0.5
mm Al equivalent.
▪ The variable-aperture light-localizing collimator usually provides an additional 1.0 mm Al equivalent. Most of this is
attributable to the reflective surface of the mirror of the collimator. To meet the required total filtration of 2.5 mm Al, an
additional 1-mm Al filter is inserted between the x-ray tube housing and the collimator. The radiologic technologist has no
control over these sources of filtration but may control stages of added filtration.
▪ Some x-ray imaging systems have selectable added filtration, as shown below.
▪ Usually, the imaging system is placed into service with the lowest allowable added filtration. Radiographic technique charts
usually are formulated at the lowest filtration position. If a higher filter position is used, a radiographic technique chart must
be developed at that position.
▪ Under normal conditions, it is unnecessary to change the filtration. Some facilities may be set for higher filtration during
examinations of tissue with high subject contrast, such as the extremities, joints, and chest. When
properly used, higher filtration for these examinations results in lower patient radiation dose. When added filtration is
changed, be sure to return it to its normal position before beginning the next examination.
▪ As added filtration is increased, the result is increased x-ray beam quality and penetrability. The result on the
image is the same as that for increased kVp, that is, more scatter radiation and reduced image contrast.
▪ Filtration and its ability to alter beam intensity affect density. All types of filtration alter density. Density decreases when
filtration is increased.
▪ When added filtration is changed, which is rare, a half-value layer calculation should be made to permit adjustment of mAs
or other factors to maintain density.