Mirror 5 Exp User's Manual (English)
Mirror 5 Exp User's Manual (English)
CE Declaration
This product is provided with a CE marking in accordance with the regulations stated in
Council Directive 93/ 42/ EEC of June 14, 1993 concerning Medical Devices. Shenzhen
Landwind Industry Co., Ltd., is certified by notified body TUV Rheinland Product Safety
GmbH. with notified No.0197 to the EC Commission.
Authorized EC Representative:
Company Name:Shanghai International Holding Corp .GmbH(Europe)
Company Address:Eiffestrasse 80,20537 Hamburg,Germany
Tel:+49-40-2513175
Fax:+49-40255726
Dimdi No.:DE/0000040627
Web:www.bfarm.de www.dimdi.de
E-mail:[email protected]
Manufactured by
Shenzhen Landwind Industry Co., Ltd.
Address: 4F, Block E, Bijing Bldg. 81, Jingtian Road, Futian District, Shenzhen China. Tel:
+86-755-27353247
Fax: +86-755-27353240
Email: [email protected].
User ’s Manual 1
All rights reserved.
This Manual and the information contained herein are proprietary information of Shenzhen
Landwind Industry Co., Ltd. No information or any part hereof can be copied, reproduced,
compiled, modified, distributed or transmitted, in any form or by any means, without the prior
written consent of Shenzhen Landwind Industry Co., Ltd. This Manual is intended for users
who are authorized to use such Manual as a part of the product purchased from Landwind.
Use of this Manual by unauthorized personnel is strictly forbidden.
Shenzhen Landwind Industry Co., Ltd. makes no representations or warranties of any kind
concerning the document, express or implied, including, without limitation, warranties of
merchantability and fitness for a particular purpose. Although every effort has been exerted
to ensure the accuracy of the information contained herein, Shenzhen Landwind Industry Co.,
Ltd. assumes no liabilities for any errors or omissions and reserves the right to modify the
product to improve its reliability, function or design without further notice. Shenzhen
Landwind Industry Co., Ltd. may modify or improve the product or program described herein
at any time.
The product may contain recycled parts whose performance is equivalent to brand-new parts
or rarely used parts.
Warning: Indicates a danger. Warnings are intended to alert you to the importance of
following the correct operating procedures where risk of injury to the patient or
system user exists. Do not proceed to any further operation until the specified
conditions are fully understood and satisfied.
Caution: Indicates a danger. Cautions are intended to alert you to the importance of
following correct operating procedures to prevent the risk of damage to the system.
Do not proceed to any further operation until the specified conditions are fully
understood and satisfied.
Note: Notes contain special information concerning the proper use of the system
and/or additional instructions.
2 User ’s Manual
Safety Classification
1. Type of protection against electrical shock: Class I equipment
2. Degree of protection against electrical shock: Type BFequipment
3. Degree of protection against harmful ingress of water: Ordinary equipment
4. Degree of safety of application in the presence of a flammable anesthetic
mixture with air or with oxygen or nitrous oxide:
Do not use this equipment in an environment in the presence of a flammable
anesthetic mixture with air or with oxygen or nitrous oxide.
5. Operation mode
Continuous operation
Applicable Scope
This instrument is applicable to the abdomen, urinary tract, and the small organ
(mammary gland,testicle, thyroid, etc.), blood vessels, heart, brain, gynecology,
obstetrics and orthopaedic department system inspection, is a general machine.
Warranty
Please do not disassemble this instrument on your own. The warranty does not
include to follow items, even during the warranty period:
1. Disassemble the instrument on your own.
2. Damage or loss resulting from disoperation or improper use (any application
beyond the application scope of this instrument).
3. Damage or loss due to the failure to follow regulations on power supply, installation and
operating conditions.
4. Damage or loss caused by the installation, modification and checkup or repair by an
unauthorized service engineer.
5. Damage or loss caused by natural disasters, such as fires, earthquakes, lightning, and
So on.
User ’s Manual 3
Environmental
Landwind designs the ultrasound diagnostic system according to the requirements
of safety and environmental protection. If user does not disassemble the instrument
or perform an improper use, the system will not cause any harm to anyone and
environment. If must use the materials which will harm the environment, please
dispose it properly under the regulations permitted.
Warning:
— Do not dispose the waste of ultrasound diagnostic system together with industrial waste
and household garbage.
— According to the local Environment Protection Law, must dispose the waste generated
by ultrasound diagnostic system properly, and process the obsolete equipment with
correct approach when it reaches its useful life.
— Recycling the recyclable materials by qualified Waste Management, thereby reducing
environment pollution.
For relevant information, please contact the Landwind service department, or dispose the
waste properly according to local refuse collection method.
4 User ’s Manual
Contents
Table of Contents
Table of Contents ...................................................................................................................................... 5
Chapter 1 Safety Precautions .................................................................................................................. 9
1.1 Safety Information ...................................................................................................................... 9
1.1.1 Safety Classification ........................................................................................................ 9
1.1.2 Electrical Safety ............................................................................................................... 9
1.1.3 Ultrasound Safety Information General ......................................................................... 13
1.1.4 Mechanical Safety ......................................................................................................... 18
1.1.5 Acoustic Output Limits ................................................................................................... 19
1.1.6 Probe Surface Temperature ........................................................................................... 20
1.1.7 Electromagnetic Compatibility ....................................................................................... 20
Chapter 2 Introduction ........................................................................................................................... 24
2.1 User’s Manual........................................................................................................................... 24
2.2 Working Theories ..................................................................................................................... 24
2.3 System Applications ................................................................................................................. 24
2.4 Imaging Modes ......................................................................................................................... 25
2.5 General View ............................................................................................................................ 25
2.6 Connectors ............................................................................................................................... 27
2.7 Probe ........................................................................................................................................ 29
2.7.1 Probe Types and Scanning Principles ........................................................................... 29
2.7.2 Probe Connection .......................................................................................................... 30
2.7.3 Selecting Active Probe ................................................................................................... 30
2.7.4 Probe Applications ......................................................................................................... 30
2.8 Peripherals ............................................................................................................................... 31
2.8.1 List of Peripherals .......................................................................................................... 31
2.8.2 Connections of Peripheral Devices ............................................................................... 31
2.9 Control Panel ............................................................................................................................ 32
2.9.1 Control Panel Brightness ............................................................................................... 32
2.9.2 Introduction of Control Panel Buttons ............................................................................ 32
2.10 System symbols ..................................................................................................................... 43
Chapter 3 Starting Examinations ........................................................................................................... 44
3.1 System Start-up and Running Sequence Inspection................................................................ 44
3.1.1 System Start-up ............................................................................................................. 45
3.1.2 System Operation .......................................................................................................... 45
3.1.3 System Continuous Operation ....................................................................................... 45
3.1.4 System Shutdown .......................................................................................................... 45
3.2 Information on the Imaging Screen .......................................................................................... 46
3.3 Patient Information ................................................................................................................... 47
3.3.1 New Patient information................................................................................................. 47
3.3.2 Historical Patients .......................................................................................................... 47
3.4 Probe and Exam Items ............................................................................................................. 48
3.5 End Exam ................................................................................................................................. 48
Chapter 4 Optimizing the Image............................................................................................................. 49
4.1 Imaging Mode ........................................................................................................................... 49
4.2 B Mode ..................................................................................................................................... 49
4.2.1 B mode Exam Steps ...................................................................................................... 50
4.2.2 B mode Parameter Settings........................................................................................... 50
User ’s Manual 5
Contents
4.3 Dual-Mode ................................................................................................................................ 54
4.4 M Mode ..................................................................................................................................... 55
4.4.1 M mode Operating Methods .......................................................................................... 55
4.4.2 M mode Parameter Settings .......................................................................................... 55
4.5 Color Mode ............................................................................................................................... 56
4.5.1 Color mode Operating Methods..................................................................................... 56
4.5.2 Color mode Parameter Settings .................................................................................... 56
4.6 PDI Mode (Power Doppler Imaging Mode) .............................................................................. 60
4.6.1 PDI mode Operating Steps ............................................................................................ 60
4.6.2 PDI mode Parameter Settings ....................................................................................... 61
4.7 Pulse-wave Doppler (PW) ........................................................................................................ 61
4.7.1 PW mode Operating Steps ............................................................................................ 62
4.7.2 PW mode Parameter Settings ....................................................................................... 63
4.8 3D Mode ................................................................................................................................... 67
4.8.1 3D Mode Operating Steps ............................................................................................. 67
4.8.2 3D Mode Parameter Settings ........................................................................................ 67
4.9 4D Mode ................................................................................................................................... 69
4.9.1 4D Mode ........................................................................................................................ 69
4.9.2 3D Static Mode............................................................................................................... 71
4.10 CW imaging model(optional) .................................................................................................. 72
4.10.1 CW Mode Operating Steps .......................................................................................... 72
4.11 ImagingPreset ........................................................................................................................ 73
4.12 LanScape (optional) ............................................................................................................... 73
4.13 Tissue Doppler Imaging (TDI) (optional) ................................................................................ 75
Chapter 5 Scan/Display Function .......................................................................................................... 77
5.1 Image Zoom ............................................................................................................................. 77
5.2 Image Freeze ........................................................................................................................... 77
5.3 CINE Mode ............................................................................................................................... 78
5.3.1 Cine loop ........................................................................................................................ 79
5.3.2 Cine Editing.................................................................................................................... 79
5.3.3 Cine Storage .................................................................................................................. 80
5.3.4 Thumbnails .................................................................................................................... 80
5.3.5 Cine Review ................................................................................................................... 81
5.4 Image Annotation ..................................................................................................................... 82
5.4.1 Mark Annotation ............................................................................................................. 83
5.4.2 Text Annotation .............................................................................................................. 83
5.4.3 Arrow Annotation ........................................................................................................... 83
5.5 Biopsy ....................................................................................................................................... 84
5.5.1 Introduction .................................................................................................................... 84
5.5.2 Adjustment of Biopsy Guideline ..................................................................................... 84
5.5.3 Biopsy-supporting Modes .............................................................................................. 85
5.5.4 Disposable Protective Case........................................................................................... 85
5.5.5 Patient Preparations ...................................................................................................... 86
5.5.6 Puncture......................................................................................................................... 86
5.6 Viewing the User’s Manual ....................................................................................................... 87
Chapter 6 Measurements and Calculations ........................................................................................... 88
6.1 Measurement Overview ........................................................................................................... 88
6.1.1 Measurement Interface .................................................................................................. 88
6.1.2 Caliper Type................................................................................................................... 89
6.1.3 Default Measurement Methods and Units ..................................................................... 90
6 User ’s Manual
Contents
6.1.4 Basic Measurement Procedures.................................................................................... 90
6.2 General Measurement .............................................................................................................. 91
6.2.1 General 2D-mode Measurements and Calculations ..................................................... 91
6.2.2 General M Mode Measurements and Calculations ....................................................... 95
6.2.3 General Doppler mode Measurements and Calculations .............................................. 96
6.2.4 Measurement Errors ...................................................................................................... 98
6.3 Special Measurement ............................................................................................................. 100
6.3.1 Abdominal Measurement ............................................................................................. 100
6.3.2 Gynecological Measurement ....................................................................................... 103
6.3.3 Obstetric Measurement ............................................................................................... 106
6.3.4 Urological Measurement .............................................................................................. 114
6.3.5 Vascular Measurement ................................................................................................ 116
6.3.6 Small Parts Measurement ........................................................................................... 120
6.3.7 Orthopedic Measurement ............................................................................................ 122
6.3.8 Cardiology Measurement............................................................................................. 124
6.3.9 Transcranial Measurement .......................................................................................... 135
6.3.10 Emergency Medicine Measurement .......................................................................... 137
Chapter 7 Record of Images and Reports ........................................................................................... 141
7.1 Worksheet and Report............................................................................................................ 141
7.1.1 Report Viewing ............................................................................................................ 141
7.1.2 Worksheet Viewing ...................................................................................................... 145
7.2 Storage of Image and Report ................................................................................................. 145
7.2.1 Reviewing Image and Report ...................................................................................... 145
7.2.2 Export Image and Report............................................................................................. 146
7.3 Data Transmission ................................................................................................................. 148
7.4 Add a Printer........................................................................................................................... 149
7.4.1 Connected to an Analog Signal Printer ........................................................................ 149
7.4.2 Connected to a Digital Signal Printer ........................................................................... 149
7.5 Report Printing........................................................................................................................ 150
7.5.1 Print Preview ................................................................................................................ 150
7.5.2 Report Printing ............................................................................................................. 150
7.6 DICOM Connection ................................................................................................................ 150
7.6.1 DICOM Storage ........................................................................................................... 151
Chapter 8 Preset .................................................................................................................................. 152
8.1 Preset ..................................................................................................................................... 152
8.1.1 Operation methods of preset........................................................................................ 152
8.1.2 System Preset.............................................................................................................. 152
8.1.3 Exam Preset ................................................................................................................ 156
8.1.4 Imaging Preset ............................................................................................................. 158
8.1.5 Measure Preset ........................................................................................................... 158
8.1.6 Text Preset................................................................................................................... 164
8.1.7 Body Mark Preset ........................................................................................................ 165
8.1.8 Network Preset ............................................................................................................ 167
8.1.9 Peripheral Preset ......................................................................................................... 168
8.1.10 Service Preset ........................................................................................................... 171
8.1.11 Acquire Settings ......................................................................................................... 172
Chapter 9 Maintenance ........................................................................................................................ 174
9.1 Caring for the System ............................................................................................................. 174
9.1.1 Daily Checklist ............................................................................................................. 174
9.1.2 Maintenance ................................................................................................................ 174
User ’s Manual 7
Contents
9.2 Cleaning and Disinfecting System Parts ................................................................................ 175
9.3 Caring for and Using Probes .................................................................................................. 177
9.3.1 Using Probes ............................................................................................................... 177
9.3.2 Cleaning and Disinfection ............................................................................................ 179
9.3.3 Storage ........................................................................................................................ 180
9.3.4 Protective Case ........................................................................................................... 180
9.3.5 Helpful Hints for Endocavity Probes ............................................................................ 180
9.3.6 Protect shell ................................................................................................................. 180
9.3.7 Caring Coupling Agent................................................................................................. 180
Appendix A Specifications ............................................................................................................ 181
A.1 Working, Storage and Transport Evironment and Packaging ........................................ 181
A.2 Probe Surface Temperature ........................................................................................... 182
A.3 Monitor ............................................................................................................................ 182
A.4 DVD-Drives..................................................................................................................... 182
A.5 Power Supply ................................................................................................................. 182
A.6 Imaging Modes ............................................................................................................... 183
A.7 Regulatory Compliance .................................................................................................. 184
Appendix B Acoustic output reporting table according to IEC 61157 ............................... 185
Appendix C Acoustic output reporting table according to IEC60601-1 ........................... 189
8 User ’s Manual
Chapter 1 Safety Precautions
Note: To avoid electrical shock, never modify the ultrasound system’s AC power connector
plug as doing so may overload your equipment’s power circuits. To ensure grounding reliability,
connect the system only to an equivalent outlet.
Note: Using an extension cord or multi-socket outlet setup to provide power to the
imaging system or to the system’s peripheral devices, may compromise the system
grounding and cause your system to exceed leakage current limits.
Note: Operators of the instrument shall be qualified service-trained professionals or shall use
the instrument under the supervision of qualified professionals.
Warning: Connect the system only to a sound wall outlet provided with a protective ground
lead and therefore use only a proper power cable. You are not allowed to remove or
disconnect the ground lead, no matter when.
User ’s Manual 9
Chapter 1 Safety Precautions
state of both wall outlet and wiring in the operating room should be inspected by qualified
personnel.
Note: Install the system only in rooms intended for medical use. The system complies with the
IEC60601-1 Electrical Safety Requirements ( IEC60601-1).
Note: Install the system only in rooms intended for medical use. The system complies with the
IEC60601-1 Electrical Safety Requirements ( IEC60601-1).
Caution: The power characteristics should comply with those of industrial enterprises and/or
health care institutions.
Caution: The system should be provided with an uninterrupted power supply (UPS) in places
where the main power supply is likely to suspend temporarily. Otherwise, the service warranty
clause will become invalid automatically.
10 User ’s Manual
Chapter 1 Safety Precautions
Caution: Do not operate the system near the sources of heat, strong electrical and magnetic
fields such as transformers as well as close to equipment generating high- frequency signals,
for example, electrosurgical units (ESUS)). Failure to comply with this requirement will have a
negative impact on ultrasound image quality and the subsequent diagnosis.
Caution: Do not remove the system hood due to the presence of dangerous voltage. When
the system is being used, the control panel shall be in place. All internal adjustments and
replacements should only be conducted by qualified service-trained Landwind customer
service professionals.
Caution: Equipment connected to the ultrasound system and located in the patient zone must
be powered from a medically-isolated power source or must be a medically- isolated device.
Equipment powered from a non-isolated source can result in chassis leakage currents
exceeding safe levels. Chassis leakage current created by an accessory or device connected
to a non-isolated outlet may add to the chassis leakage current of the ultrasound system.
Caution: To make electrical or mechanical connection between this instrument and the
equipment of other manufactures, confirm that it complied with the IEC60601-1-1:1995
standard, and recheck the leakage current and other safety indices of the entire set of system,
to avoid possible hazard arising from leakage current overlapping.
Caution: To avoid electrical shock and damage to the ultrasound system, shut down and
disconnect the equipment from the AC power source before cleaning and disinfection.
Caution: Do not use the system in an environment with combustible gases or anesthetic, to
avoid explosions.
Caution: Do not use the system in damp or wet environments. Keep it away from
inflammables and explosives.
Caution: Do not pour any fluid onto the system surfaces, as fluid seepage into the electrical
circuitry may cause excessive leakage current or system failure.
Caution: To avoid the possibility of static shock and damage to the system, avoid using
aerosol spray cleaners on the monitor screens.
Caution: Use the probe with care. Do not use damaged probes. Do not use probes with
cracks or surface crimps, since such probes might have their insulation damaged, which easily
causes electric shock to patients.
User ’s Manual 11
Chapter 1 Safety Precautions
Caution: When the system is displaced to a room with a higher temperature from one with a
lower temperature, it is highly recommended to turn on the system several hours later when
the temperature inside the system has been stabilized.
Caution: Do not attempt to modify the instrument. If it is really necessary to do so, please
notify us or our agent.
Caution: To maintain the safety and functionality of the ultrasound system, maintenance must
be performed every 12 months. Electric safety test must be performed at periodic intervals at
regular intervals as specified by local safety regulations, or as needed.
Caution: Upon the expiration of the service life of the system, please notify your dealer or
Landwind to handle properly.
Caution: Although the instrument is produced according to the existing EMI/EMC requirement,
its use in electromagnetic environments may still cause instantaneous attenuation of the
ultrasound images. If such a phenomenon often occurs, Landwind recommends you to check
the system operating environment to locate the possible radiation sources. Such radiation may
arise from other electric devices in the same or neighboring room. Such communication
devices as phones and pagers may generate such radiation, and such devices as radio
receivers, TVs or microwave transmitting devices in proximity may also produce such radiation.
If electromagnetic interferences exist, please move the instrument to another place.
EMC Note: Operating the ultrasound imaging system in close proximity to sources of strong
electromagnetic fields, such as radio transmitter stations or similar installations may lead to
interference visible on the monitor screen. However, the device has been designed and tested
to withstand such interference and will not be permanently damaged.
12 User ’s Manual
Chapter 1 Safety Precautions
Caution: For neonatal head imaging, Landwind recommends that you exercise special care
during neonatal cephalic scanning to avoid possible damage to the posterior region of the eye.
The ultrasound energy emitted by the probe easily penetrates the fontanels of the infant.
Caution: To avoid possible damage, without doctor’s guidance, it prohibits from using the
working probe scan the eye directly.
Caution: Landwind makes every effort to manufacture safe and effective probes. You must
take all necessary precautions to eliminate the possibility of exposing patients, operators, or
third parties to hazardous or infectious materials. These precautions should be considered in
the use of any application that may indicate the need for such care,
And during end cavity scanning; during biopsy or Biopsy procedures; or scanning patients with
open wounds.
Caution: Do not use the system if irregular and unusual repeated video refreshes occur.
Discontinuous scanning frequency is a hardware fault, please remove it before use.
Caution: Use the diagnostic ultrasound only when there is a good medical reason, minimize
exposure times, and select the lowest scanning power value required to create images that
are suitable for clinical exam.
Caution: When making ultrasonic scanning, the instrument shall be used with ultrasound
coupling agent, and no biologic compatibility problem is found. However, to ensure safety, only
qualified ultrasound coupling agent in compliance with related standard shall be used.
User ’s Manual 13
Chapter 1 Safety Precautions
Caution: If error information occurs to the video display, it indicates that a danger happens to
the system; please suspense the use of the system immediately. Record the error information,
turn off the system, and notify your supplier.
Warning: The system is not intended for transorbital or other ophthalmologic exams!
Warning: To ensure the safety of examined patients, do not scan the same part of an
examined patient consecutively for a long period of time.
Warning: The values of the mechanical index (MI) and technical index (TI) are not
applicable for ophthalmologic ultrasound. Take special cautions when scanning the head of a
fetus to avoid damage to the posterior region of the eye.
Warning: To avoid any patient health hazard, use probes to examine only those exam zones,
where a given probe is intended for.
Warning: Take Caution when contrast substances are used to scan heart, abdominal. Cavity
and other organs. Be careful that the lower the MI value is, the less possible ultrasonic
cavitation is.
Warning: Landwind recommends adhering to the principles of ALARA (as low as reasonably
achievable). In accordance with these principles, a physician shall select the lowest ultrasound
exposure (the lowest ultrasound intensity and exposure time) that produces diagnostically
acceptable information and, in doing so, minimize harmful effect risk.
Warning: Ultrasound power is better transmitted through water than tissues. The actual MI
and TI values may be higher than those displayed on the imaging screen when buffer
materials such as gel pad are used.
14 User ’s Manual
Chapter 1 Safety Precautions
temperature to rise. TI is the ratio of the acoustic signal power to the power causing a
temperature rise in 1°С (value based on the temperature model). There are three thermal
indices which are based on specific temperature model respectively to evaluate the
temperature rise near the surface of and inside soft tissue when the ultrasound beam passes
through soft tissue and focuses on the immediate vicinity of bone:
l Soft tissue thermal index (TIS): indicates the potential temperature rise of soft
tissue when ultrasound beam passes through and focuses on them.
l Cranial thermal index (TIC): indicates the potential temperature rise of bones near
the beam’s entrance into the body when ultrasound beam passes through the body.
l Bone tissue thermal index (TIB): indicates the potential temperature rise where the
ultrasound beam passes through soft tissue and a focal region is in the immediate
vicinity of bone.
l Both MI and TI are relative indicators which reflect potential temperature rise. The
higher the MI or TI value is, the higher the temperature rise is. They provide a
relative value for applying the concept of ALARA.
Please change the ultrasound transmission power, mode set, probe type, focus number,
Focus position depth, scanning depth and/or other parameters to change TI value.
User ’s Manual 15
Chapter 1 Safety Precautions
rise for all possible conditions and tissue types. It only means that, for a given particular
ultrasound beam shape, TI = 2 (the value that corresponds to a temperature rise in a model
specimen of 2°C) reflects a greater temperature rise in tissue than TI = 1. The temperature
rise in the actual tissue, however, is not 2°C. Refer to the following for a detail description of
the limitation regarding the use of MI and TI.
It should also be kept in mind that the margin of safety for MI and TI has not been
established. Researches on margin of safety based on bioeffects, up to now, are rather
fragmentary and still at the development stage. However, it has been found that the
temperature rise in embryonic or fetal tissue of not below 41°C (i.e., 5°C above the normal
temperature) within 5 minutes must be considered potentially hazardous. Potential hazards
are also considered present in cases where ultrasound exams with a sound pressure
amplitude near the pulmonary tissue of a neonate not below 1MPa at MI=0.7 (Pinter mode)
or MI=0.5 (Resold mode) as well as any ultrasound exams at a frequency of above4 MHz
with MI=0.5.Moreover, in the obstetric practice, ultrasound scanning modes in which TI value
is higher.
than 1.0 should be preferably avoided to comply with the reasonable margin of safety
recommended by WFUMB. If an acceptable diagnostic result is unable to be achieved under
such conditions, a short-time output increase is permitted. But, any additional thermal load
on the fetus is impermissible if the body temperature of the mother has gone up.
The models created to forecast TI take into account of the cooling effect of tissue with blood
flowing through it. Therefore, TI value should be maintained at a lower level if the hem
perfusion of the tissue under exam is poor. However, if the organs scanned are known to be
well washed with blood, for example, liver and heart, TI value can be maintained at a higher
level.
In some cases, when the soft tissues to be examined as well as bones behind them are
present in the scanning region, TIB (the thermal index for bone) other than TIS (the soft
tissue thermal index) should be used for a better gauge of temperature rise. Such a situation
occurs in examining the female breast, as well as in vascular exams if a vessel lies on a
bone surface. MI becomes of special importance in examining heart when lungs (the ‘gas-
soft tissue’ boundary) may be in the scanning region. The likelihood of cavitations is
especially high in ultrasound exams with the application of bubble materials. In such exams,
the limitation of the MI should be paid special attention. In summing up, the following should
be emphasized once more:
16 User ’s Manual
Chapter 1 Safety Precautions
Note: In a scanning mode, a temperature rise in the surface of soft tissue is always higher
than that in bone deep. But this is not the case sometimes, and Caution should be therefore
taken in interpreting TI values in both B mode and Doppler mode for second and third
trimesters of pregnancy.
Note: In cavities filled with liquid, ultrasound power could not be absorbed to such substantial
extent as expected from model calculations. So the values of MI and TI can often be
underestimated when ultrasonic scans through a full urinary bladder or amniotic fluid.
Note: TI is not intended for use in ophthalmologic practice and cannot be used in assessing
thermal effects in ophthalmologic exams.
Note: In a scanning mode, TI values give an idea about tissue heating near the probe surface
which is only based on an absorbed ultrasound beam power but take no account of a
temperature rise in tissue due to contact with the surface heated up by the probe , though
this temperature rise can be rather significant.
Note: The displayed TI values are averaged values and should not, therefore, be interpreted
as the actual temperature rise (unit: °C). TI and MI calculation models are based on many
simplifications of an actual situation. Therefore, it should be born in mind, in some especially
unfavorable situations; the actual temperature rise can be three times as much as the
displayed TI value.
Note: Unlike X-Ray beams, an unfavorable bioeffect of ultrasound on tissues only manifests in
situations where it exceeds certain threshold value. For instance, temperature of a scanned
tissue portion that goes up within a range of 37°C to 41°C is acceptable for a rather prolonged
period while a temperature elevation of up to 45°C is impermissible: no cavitations effect and
therefore, no bioeffect would occur below the threshold level.
User ’s Manual 17
Chapter 1 Safety Precautions
Note: A preferable exam approach, in terms of safety, would appear as follows: first, set the
lowest index and then increase it until a satisfactory image or a Doppler signal is produced.
Exposure time is another important parameter. Landwind recommends that you select the
lowest possible exposure time that produces diagnostically acceptable information.
Warning: Be aware of wheels and control panel, especially when you moved the machine.
Depending on its configuration and accessories, the weight of the system could be over 88
kilograms, it might cause bodily injury when the machine roll over a foot or press against tibia.
The control panel can be turned right or left, Land wind recommends pushing but not pulling
when moving the machine. Beware especially when moving on a slope.
Caution: Ensure all probe cables are safe. Avail of the cable management system for ensuring
the probe cables are free from damage. When moving the diagnostic equipment or attempting
to go over obstacles, don’t use the probe cable hook to lift up or pull the system.
Note: The wheels of the system are installed with brakes. The wheels can be locked with
stepping down on brakes, and freed when the brakes are lifted. The brakes should be
released when moving the system.
Note: If the system does not function properly after moving, please contact your supplier
immediately. The installation of each component within the system can be relied upon and can
suffer relative great force, but excessive force could cause malfunction to the system.
Note: In the process of machine use, pay attention to the load bearing of each part. Being
overweight may damage the machine parts. The main load-bearing components: keyboard
panel 5kg, probe hook 1kg, probe placed rack 2kg, 2kg storage basket.
18 User ’s Manual
Chapter 1 Safety Precautions
Transmission Power
The system is provided with the function of displaying ultrasound power. The mechanical and
thermal indices enabling ultrasound power transmitted to the patient to be monitored and
limited.
The range of power transmission (scanning signal amplitude) is -18dB~0dB. The selection
Of 0dB (or 100%), in conjunction with other controls or functions of the system, sets the
maximum sound intensity and the mechanical index of each probe where:
Warning: The use is responsible for selecting the signal level used.
To adjust transmission power applied to the patient through the probe, use the respective
keys on the system control panel. Please refer to the introduction to keys in Chapter 2
System Controls for the application of available keys. Rotate [Select] and move the cursor to
select [A. Power], and click to adjust the transmission power. As the fundamental function of
the system, it determines the transmission power of all probes and modes in the period of
real-time scanning. However, it is not the only factor to influence MI and TI.
The range and, in particular, the maximum value of MI and TI vary with probes. In addition, MI
and TI preset values are provided for each type of diagnostic exam.
Note: In addition to rotate [Select] and move the cursor to select [A. Power], and click to adjust
the transmission power, the acoustic output may be affected by such functions or system
controls as follows:
l Power ON/OFF;
l Pulse repetition frequency (PRF) of transmitter emission in PW-mode;
l PRF in Color-mode;
l Exam zone (Zone);
User ’s Manual 19
Chapter 1 Safety Precautions
l Focusing, the number of focuses and focus position depth;
l Freeze Mode;
l Imaging depth;
l Color beam density;
l Color sensitivity;
l Color window size in color mapping modes;
l System presets (See Chapter 8 Preset);
l Probe type;
l using a gel pad;
l Operating mode.
Caution: Be extremely careful! Never to operate a probe whilst emitting without contacting the
patient’s body.
The surface temperature of a probe which is being operated but makes no contact with the
patient’s body may rise until the emission function is cut off automatically (the Freeze Mode
will be automatically activated in 5 minutes).
Caution:A description of the means used to limit the surface heating of INVASIVE
TRANSDUCER ASSEMBLIES to no more than 43 °C in the event of a SINGLE FAULT
CONDITION
RF emissions
Class B
CISPR 11
Harmonic
[SYSTEM or EQUIPMENT] is applicable to all
Emissions Class A
facilities that are not directly connected to the
IEC 61000-3-2
public low voltage supply network of residential
and residential housing.
Voltage
Fluctuations/flicker
Complies
emissions
IEC 61000-3-3
User ’s Manual 21
Chapter 1 Safety Precautions
IEC 60601 test Compliance Electromagnetic environment —
Immunity test
level level guidance
IEC 61000-4-5 mode tIa l
±2kV common mode
mode ± 2 k V comon
mode
<5% UT
Mains power quality
(>95% dip in UT)
Voltage dips, Should be that of a typical
For 0.5 cycle <5% UT for 0.5
short commercial or hospital environment.
40% UT cycle
interruptions If the user
(60% dip in UT) 40% UT for 5
and voltage of the [EQUIPMENT or SYSTEM]
for 5 cycles cycles
variations on requires continued operation during
70% UT 70% UT for 25
power supply power mains interruptions, it is
(30% dip in UT) cycles
input lines recommended that the
for 25 cycles <5% UT for 5
IEC 61000-4- [EQUIPMENT or SYSTEM]
<5% UT sec
11 Be powered from an uninterruptible
(>95% dip in UT)
power supply or a battery.
for 5 sec
Power Power frequency magnetic
frequency Fields should be at levels
(50/60 Hz) 3 A/m 3 A/m characteristic of a typical location in
Magnetic field a typical commercial or hospital
IEC 61000-4-8 environment.
Portable and mobile RF
communications equipment should
be used no closer to any part of the
[EQUIPMENT
Or SYSTEM], including cables, than
the recommended separation
distance calculated from the
equation applicable to the frequency
of the transmitter. Recommended
separation
Distance
d = 1.16 P
3Vrms d = 1.16 P 80MHz to 800MHz
Conducted RF
150 kHz to d = 2.33 P 800MHz to 2.5GHz
IEC 61000-4-6 80MHz 3V Where P is the maximum output
Radiated RF 3V/m power rating of The transmitter in
3V/m
IEC 80MHz to watts (W) according to the
61000-4-3 transmitter manufacturer and d is
2.5GHz the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey, should
be less than the compliance level in
each frequency range.
Interference may occur in the
vicinity of equipment marked with the
following
Symbol:
22 User ’s Manual
Chapter 1 Safety Precautions
Notice: The production and usage of this equipment may emit some radio frequency energy.
Equipment may cause radio frequency interference for other medical and non- medical
equipment and radio communication. To prevent to produce such interference, this product
has passed the tests that confirmed its compliance with electromagnetic compatibility (EMC)
requirements of IEC 60601-1-2, which is from European standard (EN60601-1-2). the IEC
standard used for European norms in Europe. But the company does not guarantee would
never have interference in individual installation environment.
Notice: If found the equipment emerge interference which can be determined through open
and close the equipment. Users (or qualified maintenance personnel) should take the following
one or more measures to solve interference problems:
l Adjusting the affected equipment direction or location.
l Add the distance between this equipment and the affected equipment.
l Using other power supply (not affected equipment used) for this equipment.
l Contact the after-sales maintenance personnel of the Landwind company in time.
Notice: For the following caused circumstances of EMC interference, the manufacturer shall
not be responsible for: Using the other Internet cables which not recommended. Apart or refit
the equipment without permission.
Notice: All of the enclosure should be closed when operate system, ensure the intactness of
the machine before system operating, it may affect the system EMC performance in the
enclosure open operating system.
User ’s Manual 23
Chapter 2 Introdution
Chapter 2 Introduction
2.1 User’s Manual
User’s Manual including this manual and online help, Introduces the system of basic
performance, safety precautions, imaging mode, Measurements and Calculations, and
maintenance of probe and system.
This manual contains all the necessary and sufficient information for securely operating the
system, and the after-sales service personnel of Landwind will provide advanced training
for operators within the valid period of agreement. Before you try to use this Mirror 8
system, please make sure you have read and understood all of the information contained
in this manual.
Note: Please always keep this manual being placed next to the device, and periodically review
relevant operation procedures and security considerations.
24 User ’s Manual
Chapter 2 Introdution
User ’s Manual 25
Chapter 2 Introdution
Back view of the system
26 User ’s Manual
Chapter 2 Introdution
2.6 Connectors
I/O Panel:
1 USB Port
2 VGA Output
3 HDMI Output
4 S-video outputt
6 Audio outputt
User ’s Manual 27
Chapter 2 Introdution
Monitor Keys:
Keys in the rear of the monitor can directly adjust the settings of monitor brightness and
contrast. You can use the menu options to adjust the other settings of the monitor.
Keys Description
Menu Menu
Up
Down
Exit
Note: Language settings of the monitor menu are separate from language settings of the
ultrasonic system.
Note: The manufacturer defaults the settings of monitor brightness, contrast and backlight
control as imaging presets defined by the manufacturer. Adjusting monitor brightness, contrast
and backlight control may have an impact on the image optimization effects arising from preset
values defined by the manufacturer.
In order to guarantee consistency in the process of image reproduction, you should adjust
brightness and contrast of the image monitor before adjusting print quality of the installed
recording equipment.
The monitor has the functions of automatic adjustment, contrast adjustment, brightness
adjustment, backlight control, reset and exit, as shown in the figure below, through which you
can adjust these controls to match your preferences, and make it in line with the lighting
conditions of the room.
The back wall of the system is respectively attached with a nameplate (top) for full-digital
color ultrasonic diagnosis system and a label for system warning (bottom), on which
relevant information about the system is noted.
28 User ’s Manual
Chapter 2 Introdution
Caution: Risk of
electric shock, Do not
Danger: Risk of
explosion if used in the
presence of
Flammable
Note: See operating
instruc- tions.
2.7 Probe
User ’s Manual 29
Chapter 2 Introdution
Note: Before connecting or disconnecting the probe, make sure that "Freeze" mode is
enabled!
Note: Please make sure that the ultrasonic probe is always applied in accordance with the
intended uses that are prescribed for the related exam items in the following table.
Probe
Exam Subjects C4-
C2-5p L5-10p L4-13p C4-9ERp P2-3p EC4-9p
9MCp
Abdomen + + + +
Gynecology + + + +
Obstetrics + + +
Urology +
Carotid artery + +
Peripheral arterial + +
Intravenous +
small organs + +
Heart +
pediatric
musculoskeletal + +
30 User ’s Manual
Chapter 2 Introdution
2.8 Peripherals
Note: Please choose the printers recommended bythe personnel working in the Landwind.
Customer Service Center, and connect the printer with the system.
Note: when you connect additional peripherals, please observe the following requirements:
when you need to connect this system to the equipment produced by any other
manufacturers for electrical connection or mechanical connection, make sure they conform
to IEC60601-1 standards, and re-examine leakage current and other safety performance
indicators of the whole system, so as to avoid potential dangers caused by leakage current
overlay.
You can connect this system with LAN of the medical institutions with a category-5 twisted
pair wire (not included in the kit of the system).
Caution: If the system is connected with LNA, the warranty and after-warranty services
shall be implemented in accordance with the independent service agreement signed
between the operating organization and Landwind.
Caution: Please be careful when you connect this system with LAN (i.e. Ethernet). To do
this kind of connection, consult Landwind Customer Service Center. Any person who
connects this medical unit with other devices shall take responsibilities for such connection
safety.
Caution: To connect this system with LAN, the network administrator shall be responsible
for the correct settings and antivirus security for the network connection. If necessary,
consult Landwind Customer Service Center.
User ’s Manual 31
Chapter 2 Introdution
Power button
32 User ’s Manual
Chapter 2 Introdution
Located at the top right side of the system control panel and the right of alphanumeric
keyboard, TGC/LGC slide controls are used to control gain by ultrasound scanning depth.
Patient
Enter the new patient interface, that you can input a new patient information; also you can
view or edit the current and historical patients information.
Probe/Exam
Enter the probe and exam selection interface which displays a list of probes and exams,
including available system-defined exam types and user-defined settings.
Review
Enter the interface to review, select and save the patient's basic information and image
information.
Report
User ’s Manual 33
Chapter 2 Introdution
Enter the Patient Report interface to display the reports about corresponding exam items of
the patient, and execute the functions of print preview, print, export, and page turning.
End Exam
End the current patient exam, and start a new patient exam.
Toggle
Adjust the corresponding parameters on the soft menu area, and the functions of these
toggles may be different according to current imaging mode selected.
Page key
Highlight a line of soft menu options which could be activated or adjusted only after being
highlighted.
A function is activated when pressing the knob; and the other function is activated when
rotating the knob.
Press [B] knob to enter B mode for grayscale imaging. If a combination mode or other
imaging mode is activated, pressing B knob will allow the system to exit the current mode or
function and return to B mode.
Rotating [B] knob can change the total gain of B mode. Clockwise rotating can increase gain;
and counter-clockwise rotating can decrease gain.
Gain value will be displayed in the imaging parameter area on the top left corner of the
screen.
34 User ’s Manual
Chapter 2 Introdution
Dual
Dual activate the "double" mode, can display two separate collection of images side by side.
Pressing [M] knob can display M-mode cursor in the image of B mode. After that, you can roll
the trackball to place the cursor on the desired area. Press [M] knob again to display B/M
mode.
Rotating [M] knob can change the total gain of M mode. Clockwise rotating [M] knob can
increase gain; and counter-clockwise rotating [M] knob can decrease gain.
Gain value will be displayed in the imaging parameter area on the top left corner of the
screen.
CW
Press the activation of continuous wave doppler imaging.Is suitable for the purpose of heart
ultrasound (the CW mode is only available under the phased array probe) .
Color
Note: in the Power mode, rotating [Color] knob can change the total gain of power mode.
Gain value will be displayed in the imaging parameter area on the top left corner of the screen.
User ’s Manual 35
Chapter 2 Introdution
PDI
PW
Pressing [PW] knob can display a Doppler Sampling line and a Sampling gate in the B-mode
image. According to the system presets, pressing [PW] knob for the first time can also
enable the display of Doppler spectrum. For pulsed Doppler, pressing PW knob again can
enable the system to display B/PW modes and give out Doppler sound signals. Rotating [PW]
knob can change the total gain of PW mode. Clockwise rotating [PW] knob can increase
gain; and counter-clockwise rotating [PW] knob can decrease gain. The gain
Value will be displayed in the imaging parameter area on the top left corner of the screen.
Steer/Angle
In the PW mode (when the Angle function is activated), rotating Steer/Angle can adjust the
Doppler angle.
Angle value is displayed in the imaging parameter area on the top left corner of the screen.
The system can also display the angle values greater than 1° on the image screen. When the
angle is greater than or equal to 60°, the system will highlight angle values in red.
When using a linear array probe, you can press [Steer/Angle] to switch functions of angle
and steering in the process of Doppler imaging. The system will highlight cine and image in
green under the control of the trackball, so as to tell users which function is activated.
36 User ’s Manual
Chapter 2 Introdution
Depth/Zoom
Rotating the knob will change imaging depth; Press the knob will activate a zoom window,
and start amplification; We can Zoom the image no matter in real-time state or freeze
state ,Press again can exit zoom mode.
Depth value will be displayed on the right lower corner of the image.
Focus
Press this button to change the number of focuses in the image, and rotate this button to
position the focus.
Trackball
The trackball can position the image body mark, measuring mark and text. You can also use
trackball in combination with the Set key on the control panel.
The System will tell users which functions are controlled by the trackball, for which the
method is: the system highlights the selected image body mark, measuring mark and text in
green, and/or lists their active functions on the bottom left of the screen.
You can use the system presets to adjust the moving speed of the trackball.
Update
In a mixed-mode imaging process, you can switch between frozen display and real-time
display.
For example, if a B-mode image is in a frozen state, while Doppler spectrum is in a real- time
state, after you press [Update], the system will display real-time 2D images, and freeze the
Doppler spectrum.
User ’s Manual 37
Chapter 2 Introdution
When performing a measurement, pressing [Update] can circularly view the measuring mark
for editing. After the measurement is completed, pressing [Update] can circularly view the
measurement for editing.
Escape
Exit the current display mode, function or page, and reactivate the previous mode, function or
page.
Measure
Set
Used to confirm selection of a specific function or command. For example, you can use it to
lock the caliper, select a menu or body mark, including measuring values in the patient report
or delete a measurement from the worksheet.
When it is used along with the trackball, the Confirm function is similar to the mouse click
based on Windows technology.
Cursor
Activate the cursor, the cursor can move in the entire screen
38 User ’s Manual
Chapter 2 Introdution
Clear
Remove all of the comments on the screen, arrow, label and measurement.
Freeze
Freeze the image, scan or spectrum on the screen. If images or scans have been frozen,
pressing [Freeze] can restore to real-time Imaging.
System presets can change the response of the Freeze key, when you press [Freeze], other
functions will be activated.
Mark
A mark/body mark is an anatomic image displayed on the screen. It displays the direction of
the probe and the anatomic structure according to current exam type. Pressing the [Mark]
key can display a list of the available body marks of the selected exam type.
You can press the [toggle] button to select a proper body mark.
Press the [Escape] key to clear the body mark images displayed on the screen.
Somebody mark images will contain a sensor mark. Using [Select] can rotate the probe mark.
Text
Activate the annotation function and place the text cursor on the image screen. You can enter
the texts through the keyboard or by selecting the terms in the annotation list.
To reposition the text cursor, please roll the trackball. To clear the text displayed on the
screen, press the [Escape] key on the control panel or press the [Clear] button on the
Keyboard.
To display the annotation list of current exam, please press the [toggle] key, and then the
selecting library will open.
User ’s Manual 39
Chapter 2 Introdution
Note: in order to access to the other page of annotation, please roll the trackball to the
position of Page 1/2 or Page 2/2 on the bottom of the list, then press [Set].
When the annotation function is activated, you can use the system presets to display the
annotation list and customized text library, and select a library for the initial display.
Menu
To activate the main menu, and select certain item on the menu.
User Customize
F1 for the system optional 3D function button, The button does not work if there is no
matching 3D function; F2, F3 for the user-defined function keys, you can use the default
system for users to define a new custom buttons function, you can choose no, puncture
guide line, snap Settings, etc.
7.Storage Controls:
Cine Save
Image Save
Store images, the images displayed in the thumbnail area. You can use system presets to set
the area for image storage.
40 User ’s Manual
Chapter 2 Introdution
Save or print the displayed image, or send it to the target location configured in the system
presets.
2.9.2.2 Keyboard
You can use the keyboard to Input patient data, Select exam, Annotate clinical images and
Configure system preset.
The keyboard layout is similar to standard computer keyboard. Functions of keys and other
special keys as shown below:
Help
Acquire Set
Access to the screen for saving, deleting, and overwriting quick settings (used for imaging
settings of certain types of probes and exams).
Preset
Display the first page of the preset screen. Use the system presets to modify and customize
the system, including General Settings, Images, Annotations and Measurement Settings.
Display the screen of DICOM job queue to view the sending status of DICOM images.
puncture
On the video screen, activate the live puncture adjustment function for a specific probe.
User ’s Manual 41
Chapter 2 Introdution
Arrow
Placed an arrow on the screen. Rolling the trackball, can be to reposition the arrow. Press
[ confirm ] key to fixed position of the arrow. Change the direction of the arrow, please rotate
+ or
Fine tuning function of moving the biopsy guide line towards left and right when correcting a
biopsy guide line; and turning pages of the user's manual.
Backspace
To delete characters.
Caps Lock
Volum
Adjusting Doppler sound volume ; fine tuning function of moving the biopsy guide line up
and down ,and turning page up/down of the user's manual.
42 User ’s Manual
Chapter 2 Introdution
Symbol Description
AC
Dangerous voltage
BF-type equipment
Equipotential Connection
Footswitch Connector
Contrast control
Probe port
USB port
Bar code
Grounding mark
User ’s Manual 43
Chapter 3 Starting Examinations
Warning: This system will generate, receive and radiate RF electromagnetic waves. If it is not
installed or used strictly in accordance with the system installation and operating instructions, it
may have a detrimental impact on other devices, and may also interfere with itself.
We encourage users to take the following measures to exclude interferences (by switching
"ON/OFF" the system to identify interferences):
l To changes the direction or position of the device;
l To improve the isolation between the system and the equipment;
l The socket connected to this system is different from those to other devices;
l If you need help, please contact Landwind’ Customer Service Center.
Install the system monitor and avoid its screen being exposed to local lighting and direct
sunlight as well as other light sources.
This system shall be installed, assembled, customized an adjusted by the personnel from
Landwind Customer Service Center. Landwind Customer Service Center can also provide
services upon the expiry of warranty in accordance with the independent service agreements
signed between the operation institutions and Landwind. After transport or storage, please
contact Landwind Customer Service Center for any questions about the system's basic
information about re-storage, installation and assembly.
Before being installed at the scheduled location, the system shall be unpacked, and stored in
the environment at the temperatures below 5ºC. After being placed for at least two hours at
the room temperature, turn on/off the system and ultrasonic probe.
Caution: Before moving the system from one place to another place, first open the lock on the
caster brake, and lock the caster near the step or ramp.
Caution: This system shall be always installed on a level surface; therefore, the casters must
be locked. Any failure in following these requirements may cause overturning or rolling of the
system.
Caution: In the process of using the machine, rotate the monitor slowly, not too large. When the
rotation Angle reaches the limit, stop rotating in the same direction.
44 User ’s Manual
Chapter 3 Starting Examinations
Caution: Before starting this system with mains power supply for the first time, please make
sure that voltage and frequency are in line with the technical specifications described on the
nameplate on the back panel of the system.
Start the system, and perform the following steps:
l Make sure that the system has been connected with the feeder line of the monitor, and
has been properly connected to the three-hole socket with protective grounding in 220V
50Hz;
l After being connected to the mains power supply, adjust the electric switch "POWER"
l On the back panel to "I" for starting, and press on the top right corner of the control
panel;
l Turn on the printer and load paper into it.
Note: If the system loading fails, an error message will prompt. In this case, please contact the
personnel in Landwind Customer Service Center.
Warning: Users are not allowed to install other software without authorization.
Note: Due to the compatibility issues arising out of the English operating system, the system
will come to occasionally stop responding, which can be resolved by restarting the system, and
the system needs to put on a patch so as to completely solve the problem.
and press on the right corner of the control panel. The prompt information of "Do you
want to shut down?" pops up and the system will be automatically off in several seconds after
confirmation.
Caution: When the machine is downloading, you cannot shut down the system; otherwise, this
could cause the failure of software downloading. If you need to shut down the system, please
wait until the operating system has finished the downloading.
User ’s Manual 45
Chapter 3 Starting Examinations
1. Hospital name;
2. Current patient name and ID number;
3. Mechanical index (MI) and thermal index for soft-tissue ( TIS);
4. System time and date (the time does not go when being frozen);
5. Thumbnails area;
6. Gray scale (right); color scale will occur on the left of gray scale in Color Doppler
Blood Flow Imaging (color mode) or Power Doppler Blood Flow Imaging (PDI mode).
7. Scale and focus marking;
8. Option buttons from left to right: Page-down, Page-down, save to a USB Disk or CD,
Sent to the report and Delete images;
9. From left to right followed by network connection, CD, USB flash drive icon, System
time, Capital/Minuscule switching, System language;
10. Depth and frame rate
11. values; Soft menu;
12. Image display area;
13. Active probe name, active exam type and imaging parameters;
14. Probe direction signs
15. main menu
46 User ’s Manual
Chapter 3 Starting Examinations
User ’s Manual 47
Chapter 3 Starting Examinations
with [﹡] are the default exam items of this probe. When the port is not
Connected with a probe, the corresponding probe area will be displayed in blank.
Click the probe icon, select the desired probe and return to the main interface; the user can
also click the [Exam Item] of the corresponding probe, and the main interface will carry out
the measurement to the exam item of the current probe.
Note: If the port is connected with the probe, and the corresponding area is still displayed in
blank, please contact the appropriate service personnel for inspection.
48 User ’s Manual
Chapter 4 Optimizing the Image
Note: When operating in M mode, Color mode, PDI mode or PW mode, you can exit from these
mode interfaces by pressing the B button, B mode will be displayed.
4.2 B Mode
B mode is the most commonly used imaging mode in two dimensional imaging modes, which
has the measurement functions to display two-dimensional images and provide related soft-
tissue anatomy structure; in B mode, the images are displayed in gray scale.
To enter B mode, press [B] on the control panel. When you switch on the system, you can
directly access to this mode.
User ’s Manual 49
Chapter 4 Optimizing the Image
In the image above, when you change the parameters of the B mode menu bar, the
parameter area on the top left of main menu will display the corresponding indicator
changes (as shown below):
In B mode, to optimize the specific image scanning parameters above, you can make the
following adjustments:
Gn (Gain)
Operation: Make adjustments by rotating the [B] knob on the control panel.
Function: With this knob, you can set the brightness of a two-dimensional image. This knob
decides the coefficients of the increasing amplitudes of echo signals. All incoming echo
signals make use of the sample gain for amplification non-relevant with the depth.
You can also use TGC slider knobs to make compensations for gain through depth
(Sensitivity time control), so as to select gain for different images in different depths by
meams of separate methods. In this way, you can determine the accurate attenuation
compensation of echo signal according to the entering time (depth of emission non-
uniformity). The standard position of the slider knob is located in the central position. This
position corresponds to the propagation time compensation of echo signal.
50 User ’s Manual
Chapter 4 Optimizing the Image
F (Frequency)
frequency range.
Function: For special types of patients, you can change the frequency parameters so as to
achieve the best possible optimization effects.
DR (Dynamic Range)
Operation: view the soft menu [DR], you can accordingly press the functional key [2] or ▼
Persist
Operation: View the soft menu [Persist], and press ▼ or ▲ on the toggle [3] .
Operation: View the soft menu [R/S] and press ▼ or ▲ on the toggle [4], and change the
Colorize
Operation: View the soft menu [Colorize] and press ▼ or ▲ on the toggle [5] .
Scanning Area
Operation: View the soft menu [Fov] and press ▼ or ▲ on the toggle [1] to change the size
Rejection
Operation: Press [Page] on the control panel to turn to the next menu, then press ▼ or ▲
User ’s Manual 51
Chapter 4 Optimizing the Image
Smooth
Operation:View the soft menu [Smooth] and press ▼ or ▲ on the toggle [3] to change the
smoothness.
Function: This function is to reject image noise and to make images smooth.
LanSRI
Operation: View the soft menu [LanSRI] and press ▼ or ▲ on the toggle [4] to change the
image edge.
Function: Optimize the clarity of image edge.
Operation: View the soft menu [Map] and press ▼ or ▲ on the toggle [5] .
Function: This function applies the gray correction to obtain the optimum images, but it would
be easy to affect the information in B mode.
Note: Before making other adjustments, please choose the grayscale map, because there is
mutual dependence among grayscale map, gain and dynamic range. If you change the
grayscale map, you have to reset gain and dynamic range.
Focus
Operation: Press [Focus] key on the control panel.
Function: Selection of focal area has determined the depth range for the purpose of
having the best shape in this ultrasonic beam. The position of focal area can be
represented with the arrow on the right side of the image.
Scanning Depth
Operation: Press [Depth / Zoom] on the control panel;
Function: rotate this key, and you can set the depth range of the echo image in a certain
desired area. Only real-time depth adjustments are allowable. Changing echo image depth
will cause the changes of depth proportion, MI/TIS/TIB/TIC and other indexes as well as
frame rate.
You can rotate the menu button to change B-mode image display, Functions of specific
parameters and methods of operation are as follows:
52 User ’s Manual
Chapter 4 Optimizing the Image
Acoustical Power
Operation:Roll the [select] knob on the lower left corner of the control panel, and move he
cursor to select [A Power], press and roll the knob, of which rotating to the left is to educe
the parameter of acoustical power; otherwise rotating to the right means increase, hen press.
Trapezoidal Imaging
Operation: when the probe is the linear array probe, this area will be in the optional state,
while the remaining probes are gray and in non-optional states. Rotate the [select] knob and
move the cursor to select [Trapezoidal], click and the trapezoidal imaging will be [On], then
press to roll it [Off].
Split mode
Operation:Rotate the [select] knob to move the cursor to select [Spilt], and press the button
to turn the 2B real-time [On]; the two images on the left and right of the screen will be
displayed; press the button again, and the 2B real-time will be closed and displayed as
[Off].
LanTissue
Operation:Rotate the [select] key to move the cursor to select [LanTissue], press the button
to continuously rotate the key, and there will be four options, namely General, Liquid, Fat and
muscle, then press [select].
Anatomical M Mode
Reserved setting
User ’s Manual 53
Chapter 4 Optimizing the Image
4B Mode
Operation: Rotate the [select] knob and move the cursor to select [4B], consecutively press
the button, and the image display area will be followed by four images, among which when
the image icon gets gray, the image will be in a frozen state; the image in yellow is in the
real-time state, and only one image is real-time of these four images, and the others are in
the frozen state.
4.3 Dual-Mode
To enter the Dual-mode, press the [B] button on the control panel, and press [L] , and a B-
mode image will occur on the left of the image, then press [R] to display another B-mode
image on the right of the screen. In this case, the left image is in the frozen state, while the
right image is in the real-time state, which can be switched with each other by pressing the
left and right keys. The image parameters can be adjusted just similar to B-mode
adjustment.
54 User ’s Manual
Chapter 4 Optimizing the Image
4.4 M Mode
M model is a movement mode used to identify objects under the ultrasonic beam, and the
most common mode used to inspect movement of the heart.
Sweep
Operation: canning spepress ▼ or ▲ in the toggle [3] to increase or decrease the scanning
speed. Function: Scan Speed is used to change the sed of the time line in M mode.
User ’s Manual 55
Chapter 4 Optimizing the Image
Edge
OperationView the soft menu [Edge] and press ▼ or ▲ on the toggle [4] .
Function: Optimizes the clarity of image edge.
In the main menu of M mode(as shown below), you can also set acoustical power, screen
layout and others through [Select] .
You can set the parameters in the following figure to optimize images through the menu
above.
56 User ’s Manual
Chapter 4 Optimizing the Image
In color mode, the methods for setting the images shown in the figure above:
Gn (Gain)
Operation: You can change the gain by rotating [Color] on the control panel, in which left/right
adjustment is to decrease/increase gain.
Function: Gain can amplify the overall echo intensity processed in the color-mode window or
spectrum Doppler timeline, and its value will vary depending on the probes and applications.
Gain has no effects on power output. But increasing gain will lower the power output level to
so as to produce images in equal quality.
Scale
Function: Scale will affect the output energy, frame rate and wall filter.
LanFlow
Operation: Press ▼ or ▲ in the [3] of the toggle [LanFlow] to change the color blood flow.
Function: This function is used for image optimization by providing adjustment on parameters,
including scales and filters.
Operation: Press ▼ or ▲ of [4] in the toggle [R/S] to adjust the linear density.
Function: Line density helps to optimize color blood flow or spatial resolution in order to get
the best color images, which will be greatly helpful for high-frame-rate fetal heartbeats, adult
heart applications and clinical radiology applications. Changes of line density will also
change the volume density and frame rate, and at the same time, it willChange thermal index
and/or mechanical index, or may also affect the output display.
User ’s Manual 57
Chapter 4 Optimizing the Image
Baseline
Function: In the Color Mode, you can move the baseline so as to avoid aliasing in the
direction of blood flow (similar to the baseline displacement in pulse Doppler mode). By
moving the baseline, you can expand the speed range in one of the directions. The zero line
of color scale will also make corresponding displacement. The maximum and minimum blood
speed will be respectively displayed on the top and bottom of the color scale.
Operation: Press [Page] and then Press ▼ or ▲ of [1] in the toggle [Persist] .
Function: Frame correlation can determine the persistent time of color data in the sampling
box, and the persistent color data will remain on the screen until the persistent time has
expired or the system has detected a higher rate of blood flow.
Operation: Press [Page] and then press ▼ or ▲ in the toggle [2] to adjust the size of the
Priority
Function: Color priority can be used to choose whether to display arbitrary pixel color or
threshold value of B mode data, and color priority increase will lead to displaying more color
pixel information on the B mode image.
Map
Operation: Click on the touch screen menu [Map] and Rotary switch key menu [Map] in order
to change the color of the maps.
Function: Color-scale map refers to the color range that is allocated to the desired color or
power area, and the color-scale map is based on the flow rate of color flow imaging and the
flow intensity of power mode.
For color imaging, color changes mean the changes of flow rate. A dark color means a lower
rate while a light color means a higher flow rate, and these maps have the red and blue
colors to indicate the flow direction, while the color flow rate and deviated maps use green
shadows to indicate the flow disorders
58 User ’s Manual
Chapter 4 Optimizing the Image
A. Power
Please see the methods of B mode.
Color Flip
Operation:Rotate [select] and move the cursor to select [Invert],press this button to start
the color flip, through which the color scale on the right side of the image will make up-
and-down flip, and the color in the collection box will also flip, in which the original color
on the top of the color scale will become the color at the bottom of the color scale; on
the contrary, the color on the bottom will also be turned into the color on the top. Press
[Select] again to return to the color flip.
Split Mode
Operation:Rotate [Select] and move the cursor to select [Split], and press this button, thus
two real-time uniform images will be displayed in the image area, in which the image
sampling box is displayed in the right area, while the non-color B-mode image is displayed in
the left area, allowing to easily compare and view images.
Sensitivity
Operation:Rotate [Select] and move the cursor to select [Sensitivity]; press and rotate this
button in turn, then the letters of L, M and H will occur on the right of [Sensitivity], of which H
has the highest resolution and L has the lowest. What you need to do is just to press [Select]
again.
Function:By using this function, you can change the scanning operation frequency of color
Doppler imaging to adjust the sensitivity. Sensitivity increase will lead to decrease of frame
rate, while sensitivity reduction will lead to increase of frame rates (See frame display area
on the bottom right of the image for more details).
Peak
Operation:Rotate [Select] and move the cursor to select [Peak]; press this button and rotate
the key in turn, then the characters of 1sec, 2sec, 3sec and OFF will occur on the right of
[Peak]. What you need to do is just to press [Select] again.
Function:When the ultrasonic system continues accumulation, and displays the peak color
User ’s Manual 59
Chapter 4 Optimizing the Image
rate associated with flow rate in the selected time, there will be a phenomenon of persistent
peak in order to designate the time period for collecting the color rate of flow peak; when
you select 1sec, the system will maintain the peak for 1 second.
4B
Operation:Move the cursor to the 4B mode and press [select], and four images will be
consecutively displayed, in which the image will be in the frozen state when the image
icon becomes gray; and when the image gets yellow, it is in the real-time state, and
the sampling box is always in the real-time image.
.
60 User ’s Manual
Chapter 4 Optimizing the Image
These parameters will be displayed on the left side of the screen, as shown below:
For the relevant adjustment functions in PDI mode, the user can refer to the operating
methods of Color mode, which will not be repeated here.
User ’s Manual 61
Chapter 4 Optimizing the Image
1. Sampling line
62 User ’s Manual
Chapter 4 Optimizing the Image
2. Blood flow angle Indicator
3. Sampling gate
Sampling Line
Function: When activating Doppler in the B mode, [Sampling line] will be displayed in the B-
mode image and this cursor is the graphical indication of the acoustical line.
Sampling Gate
Function: For the B-mode images, Doppler-mode sampling line moves the sampling gate,
and the gate is located on a specific position. Press the function keys to change the size of
the gate. Changing sampling gate will change heat index and/or mechanical index.
Please note that it may affect output display.
User ’s Manual 63
Chapter 4 Optimizing the Image
Gn (Gain)
Operation: Use the rotary knob [PW] to change the amplitude of Doppler signal and their
display brightness on the imaging screen.
Function: The selected gain shall be matched with the signal amplitude and the used
grayscale. In this case, make sure that spectrum diagram does not appear on a lot of noise.
When rotating the control in a clockwise direction, the entire spectrum diagram becomes
brighter. When rotating the control in a counter-clockwise direction, the entire spectrum
diagram becomes darker.
DR(Dynamic Range)
Operation : Press the toggle [DR] for adjustments.
Function: Affect the amount of the displayed Doppler amplitude data.
Filter
Operation: Press the corresponding function keys on the soft menu [Filter].
Function: Remove the noises that are caused by cardiovascular or heart wall movements. It
will be at the price of lowering the sensitivity of flow.
Baseline
Operation: you shall use the toggle [Baseline]. Moving the baseline on the spectrogram can
extend the speed range in one direction, while the speed range displayed in another
direction is reduced.
64 User ’s Manual
Chapter 4 Optimizing the Image
Function: As the leveling line corresponding to the zero velocity and zero Doppler frequency
on the spectrum diagram.
Spectrum Invert
Operation: Invert the spectrum curve in the vertical direction without affecting the baseline
position, press the toggle[Invert].
Function: The function key [Invert] will be displayed on the top left of the screen, and [+] and
[-] on the speed scale will be accordingly inverted. By doing this, you can also change the
speed (frequency) calibration. In the initial state, the area above the baseline means the
blood-flow movement towards the direction of the probe, while the area below the baseline
means the blood-flow movement away from the probe. If the inverting function is enabled,
the area above baseline means the blood-flow movement away from the probe, while the
area below baseline means the blood-flow movement towards the probe.
Rejection
Function: Noise and interference arising from excessive Doppler signal spectrum can be
properly controlled by pressing [select] on the keyboard.
Sweep
Operation: Rotate [Select] and move the cursor to select [Sweep], and press this button.
Function:The scan speed can control the updating frequency of the spectrum, and the
scale will be displayed on the bottom of the screen.
Full Screen
Operation: Rotate [Select] and move the cursor to select [Full], and press this button,then
User ’s Manual 65
Chapter 4 Optimizing the Image
you can open the full-screen display of PW mode and observe the images in a meticulous
and clear manner.
Auto Tracing
Operation: Rotate [Select] and move the cursor to select [Auto trace], and press this button,
followed by Off, Up, Down and Both; for Up, the red and green lines appear at the tracing
above the baseline; for Down, the green and red lines appear at the tracing below the
baseline; for Both, the red and green lines appear above and below the baseline at the
same time, among which the green is the largest tracing and the red is the average.
Duplex
Operation: Move the cursor to the [Triplex/Duplex] option, and press [select] to enable the
dual-synchronous function; in this point, B mode and PW mode are both in the real-time
state; when the dual-synchronous function is disabled, B mode is frozen while the PW time
line remains enabled.
Triplex
Operation: Press [PW] twice after opening Color (PDI) mode, then select [Triplex/Duplex] in
left PW menu; press [select] to enter the triplex mode in combination with B, Color and PW.
In this mode, image optimization adjustment is broadly the same as the B mode, Color
mode(PDI mode) and PW mode. To switch among B mode, Color mode (PDI mode) and PW
mode, align the cursor with the top mode option bar of the main menu, and press [select].
Triplex activated mode: B + Color + PW, B + PDI + PW.
Update
This feature has not been achieved yet.
Format
Operation: Rotate [Select] and move the cursor to select [Format], and press this button,
followed by V1:1, V1:2, V2:1 and H2:3; the proportion between the two images is different,
thus you can use it according to your own preferential habits.
Note: In inverting the Doppler, the acoustical signals will also be inverted.
66 User ’s Manual
Chapter 4 Optimizing the Image
4.8 3D Mode
The free arm 3D image (FreeHand 3D) is the operator's uniform and slow moving probe for
scanning. The ultrasonic instrument records the storage scanning data, and after the scan is
completed, the 3D image is obtained.
Reconstruction mode
Operation: press the soft menu [rebuild mode] to switch the reconstruction mode
Function: define the mode of reconstruction, with surface, composite, maximum, minimum
XRay, cubeview six modes. Under the surface mode, the threshold can be adjusted and the
brightness and contrast can be adjusted in any mode. In the composite mode, transparency
can be adjusted.
User ’s Manual 67
Chapter 4 Optimizing the Image
smoothness
Operation: press the soft menu [smooth] to switch the key to smooth and change the
smoothness.
Function: reduce noise and smooth the image
Gray-scale figure
Operation: press the soft menu (fake color) to switch the fake color and change the fake color.
Function: change image color.
rotation
Operation: press the soft menu [rotation] for the switch key.
Function: to rotate the 3D image 90 degrees. You can also adjust the X, Y, Z axis of the
image through the Angle, pw,color key.
Threshold
Operation: Rotate [Select] and move the cursor to select[Threshold], and press this button.
Function: Gray value threshold definition for the reconstruction of gray value and regarded as
noise. Opacity define a threshold for distinguishing the stringency Degrees, low opacity value
makes the surface look stiff, high opacity value makes the image reconstruction with a
transparent appearance.
transparency
Operation: press the soft menu [transparency] to switch keys.
Function: to set the transparency of image, the higher the value, the higher the transparency
of grey level information.
Iuminance
Operation: press the soft menu [brightness] for the switch key.
Function: set the image brightness, the higher the value, the higher the gray level information
brightness.
contrast
Operation: press the soft menu [contrast] to switch keys.
Function: set the image contrast, the higher the value, the higher the gray level information
contrast
Format
Operation: Rotate [Select] and move the cursor to select[Format], and press this button.
Function: When the screen can be divided into 1, 2, 4 window, switch to a lower number of
windows Will keep the image from left to right.
68 User ’s Manual
Chapter 4 Optimizing the Image
4.9 4D Mode
4.9.1 4D Mode
4D mode is a slow-moving probe uniform operator scans, ultrasound equipment records storage
Scan data, after scanning, reconstruction, get a four-dimensional image. 4D provides two views for
display and use of images Sectional view and reconstruction view, cross-sectional view showing
each section.
User ’s Manual 69
Chapter 4 Optimizing the Image
4.9.1.2 4D Mode Parameter Settings
You can use the soft menu to set a specific parameter, and its value will be displayed on
the left of the screen, as shown below:
In color mode, the methods for setting the images shown in the figure above:
Threshold
Operation: Rotate [Select] and move the cursor to select[Threshold], and press this
button.function: Gray value threshold definition for the reconstruction of gray value and
regarded as noise. Opacity define a threshold for distinguishing the stringency Degrees,
low opacity value makes the surface look stiff, high opacity value makes the image
reconstruction with a transparent appearance.
Format
Operation: Rotate [Select] and move the cursor to select[Format], and press this button.
Function: When the screen can be divided into 1, 2, 4 window, switch to a lower number of
windows Will keep the image from left to right.
Quality
Operation: Rotate [Select] and move the cursor to select[Quality], and press this
button.Function: Balance of speed and line density. Hi2 combination of highest density and
lowest speed; Low combines the lowest density and maximum speed Degrees.
Reset Rotate
Operation: Rotate [Select] and move the cursor to select[Quality], and press this
button.Function: Select all the parameters are reset back to the original value or the selected
preset.
Cutting
Operation: Rotate [Select] and move the cursor to select [Cut], and press this
button.Function: Cutting 4D image section is divided into polygon cutting and tracings cutting,
cutting polygon, click the SET button to confirm the cut Point, double-click the SET after
cutting is completed, when the tracings cutting, click the SET key cutting.
Cutting revocation
70 User ’s Manual
Chapter 4 Optimizing the Image
Operation: Rotate [Select] and move the cursor to select [Cutting revocation],and press this
button.Function: Only undo the last cut.
Clear cutting
Operation: Rotate [Select] and move the cursor to select [Clear cutting].Function:
Consumers since entering the Scalpel (scalpel) all cut operation mode.
Sweep
Operation: Rotate [Select] and move the cursor to select [Sweep].Function: Probe scanning
angle.Other parameters please see the methods of B mode.
User ’s Manual 71
Chapter 4 Optimizing the Image
1. Press [CW] for the first time, show a green sampling line
2. Press [CW] for the second time, and two images will occur respectively on the top and
bottom, in which on the top is the B-mode image, and on the bottom is the CW
spectrum ;
3. Press [freeze] to image;
4. Press [Measure] to perform imaging measurements;
5. Press [CW] for the third time to exit the CW mode,enter B real-time imaging status
For more information about controls, refer to the "PW doppler" optimization control
specification
72 User ’s Manual
Chapter 4 Optimizing the Image
4.11 ImagingPreset
For the imaging modes described above, the relevant parameter settings on the main
interface have been introduced; you can reset the default setting in the Preset, if necessary;
the imaging preset operating methods are as follows:
1. Click on menu [Setup] to pop up the Preset Interface;
2. Click the module [Imaging Preset] on the left side of the screen, respectively followed by
the sub-menus of B, THI, M, Color, PDI and PW;
3. In the sub-menus above the screen, click the imaging preset menu of the corresponding
mode, through which the user can default the settings of exam item, probe and
parameters and so on.
User ’s Manual 73
Chapter 4 Optimizing the Image
LanScape acquisition
Configure the programmable key F2 as LanScape before the scan: press the [Setup] key to
enter preset menu, select [system preset], then select [User defined], then configure [User2]
as LanScape.
74 User ’s Manual
Chapter 4 Optimizing the Image
l The system accommodates a reasonable range of motion velocity. DO NOT make
abrupt changes in speed of motion. Deeper scans generally require reduced speed.
PW Tissue Doppler(PTD)
User ’s Manual 75
Chapter 4 Optimizing the Image
76 User ’s Manual
Chapter 5 Scan/Display Function
Click on menu [Setup] → Click [Exam Preset] on the left of the screen, and then you can
Click on menu [Setup] → Click [Peripheral Preset] on the left of the screen → Click
User ’s Manual 77
Chapter 5 Scan/Display Function
Pedal Switch] of the sub-menu above the screen to carry out relevant settings for left-foot
pedal switch and right-foot pedal switch.
You can enter the cine mode by selecting the [Freeze Response] of the [Exam Preset]
module to Cine. In Cine mode, "ring Indicator" will be displayed on the bottom right of the
screen (as shown below):
78 User ’s Manual
Chapter 5 Scan/Display Function
No. Name Explanation
Frame-by-frame review:
1. Press [Freeze] to move the trackball for viewing the cine loop frame by frame.
2. Press [Freeze] again to exit the CINE mode, remove memory area and start the real-
time imaging.
User ’s Manual 79
Chapter 5 Scan/Display Function
5.3.4 Thumbnails
In order to distinguish between images and cine in the thumbnail, there will be
displayed between cine and images. By using the icon below the thumbnail, we can flip up
and down, store, transmit and delete the images.
Specific operations:
1. Click on menu [Image Save] or [Cine Save] on the control panel, and the image or cine
can be seen in the thumbnail area in order.
2. Press the [Pointer] on the keypad, and use the trackball to move the cursor. Place the
cursor on the selected thumbnail, press [Set], and the selected image will occur in a
yellow box.
3. Use the trackball and [Set] to select multiple images, and click the icon to store, transmit
and delete images.
4. Select the thumbnail and double-press [Set] to view the big image, and Click [Freeze] to
exit the larger image;
5. Press [Set] again in the selected thumbnail to cancel the selection.
Note: Cineloop will occupy a great portion of the system memory. Therefore, in order to avoid
system memory overflows, it is recommended to output the recorded cine into CD or U Disk,
and then delete the recorded contents from the database in accordance with the sequences.
Note: In order to store images in the thumbnail into the preview area, you must establish a
new patient; otherwise, the images cannot be viewed in the Review.
80 User ’s Manual
Chapter 5 Scan/Display Function
l The images that are saved in the thumbnail will be stored in the D disk of the system by
default. The green area below is the currently used capacity and prompt of D disk so as
to avoid image(s) from failing to be saved due to memory overflow.
l When there are a larger number of patients, we can find a patient through [Search Key];
l Click the menu below the screen to carry out the following actions such as, "select all',
[Unselect all] [send], [cancel], [delete Exam],[Hide] and so on.
l Click the menu [Image Screen] on the right to enter the interface [Image Review];
When clicking [Image Screen], you will enter the interface [Image Review], as shown
below:
User ’s Manual 81
Chapter 5 Scan/Display Function
l On the upper left is the system's image default saving path, folder and storage time.
l You can view the image size in line with the operator's habits and page layout, and
double-press [Set] to view the single-frame image, click on the "Freeze" key to exit
display;
l Click the button below the image to transmit the selected image(s) to a U disk and/or
CD, and carry out other actions such as "select all', "delete", "send to report", "send to
DICOM", Page up and Page down, etc..
l Click the exam screen to return to the [Review] interface;
82 User ’s Manual
Chapter 5 Scan/Display Function
Press [Setup] → Click [Text Preset] on the left of the screen → [Text Properties] on the sub-
User ’s Manual 83
Chapter 5 Scan/Display Function
5.5 Biopsy
5.5.1 Introduction
The function of Biopsy can help the doctors to accurately locate the transducer and the
ancillary Biopsy needle in the process of Biopsy; this is because the Biopsy guideline
displayed on the imaging screen is just pointed to the preset path of the Biopsy needle.
Note: Biopsy is of high risk to the patients and the doctors, thus the user must comply with the
correct needling sequences when using the Biopsy stand and its accessories so as to avoid
unnecessary risks to the patients.
Note: The Biopsy guideline displayed on the screen is used for reference only. The user
must verify the exact puncturing position in the operations of biopsy, tapping and aspiration,
etc.
Press [Setup] on the keyboard → Click [System Preset] on the left of the screen → Click
[Biopsy] on the sub-menu of the screen, then you can carry out the default settings to the
relevant Biopsy functions.
Note: The Biopsy line shall be demonstrated by the representative or user from Customer
Service Center, and if the probe or biopsy bracket is replaced, the Biopsy program must be
repeated.
Note: Before Biopsy, make sure the track that the Biopsy needle enters is in line with the guide
line displayed on the imaging screen, which can be pre-tested in a water tank filled with 47℃
water. For an individual user, each type of probe needs to make only one debugging.
84 User ’s Manual
Chapter 5 Scan/Display Function
Note: Before Biopsy, make sure that the track of pointer is consistent with the guideline on the
screen.
2. Press Ctrl+ on the keyboard can left move the biopsy guide line; Ctrl+ can
right move the guide line; Ctrl+ can up move the guide line; and Ctrl+ down
move the guide line.
3. Roll the [Steer/Angle] knob to change the biopsy guide line angle.
4. Press [F12] again to exit the fine tuning function .
Note: All medical institutions must provide a disposable case for the probe so as to prevent
infection. Currently, it is recommended to use formal and standard disposable cases without
containing any pyrogen or other pathogenic agents.
Note: Each of the probes is provided with an independent case, which shall be consisted of
protective shell and cable as well as fixed ring. The sterile probes shall be placed in the
biopsy-dedicated probe device; in addition to the above, the case shall be equipped with
relevant biopsy-dedicated accessories.
Note: The latex contained in the product is likely to cause a latex-allergic patient to suffer from
a strong allergic reaction.
User ’s Manual 85
Chapter 5 Scan/Display Function
Note: Do not use condoms containing lubricants to protect the case, as this may damage
the probe. Condom lubricants are not compatible with the probe materials.
Note: Do not use any expired probe case; before use, please be sure of the service life of the
case.
Note: Only the doctors experiencing with full testing are qualified to carry out a centesis.
Appropriate preventive measures must be taken and a sterile environment must be
provided.
Note: The Biopsy function cannot be used for IVF and chorionic tissue test and umbilical cord
blood sampling and so on.
5.5.6 Puncture
Note: Before centesis, make sure that the puncturing path of the needle-holder is
consistent with the guideline on the screen. It can be tested by putting it into a water tank
filled with 470C water.
Note: Do not use the guide needle that has been tested whether the puncturing path is
consistent with the guide line on the screen, or use the guide needle that is used for
positioning the puncturing path. Make sure that a new guide needle is used for every Biopsy.
86 User ’s Manual
Chapter 5 Scan/Display Function
Note: There are several guide lines that are commonly used for Biopsy and do not make any
Biopsy before the guide line is completely determined.
Note: When setting the guide line, we do not take into account the bending or deformation of
the Biopsy needle.
Note: The intersection of Biopsy guide line on the screen and the offset of the probe scanning
central axis (depth) are able to be calculated, which can be tested in the water tank filled with
470C water.
User ’s Manual 87
Chapter 6 Measurements and Calculations
Note: In order to guarantee the desired area from any miscalculation due to the
measurement value in the measuring process, and therefore we should adjust the
parameters so as get the best image quality for measurement.
Note: In order to obtain accurate Doppler blood flow measurements, we should keep the
sampling angle at 60 degrees, as Doppler angle lower or higher than 60 degrees is prone to
misdiagnosis arising from reading errors.
Note: If the image is not frozen or the measurement mode is changed in the measuring mode,
the measurements and calculations data will disappear from the screen, and the General
measurement data will not be saved (Special measurement data will be saved in the
worksheet and the report)
Note: If the system is suddenly shut down in the measuring process, the measurement data
will not be saved.
In general, the user can carry out the functions of measurements and calculations in the real-
time, zoomed or frozen image, but in M mode and PW mode, the measurement function
cannot be implemented in the real-time image, for which the image shall be frozen first.
88 User ’s Manual
Chapter 6 Measurements and Calculations
1. Special Measurement main menu: from top to bottom: measurement item, page turning
display of measurement item ( this option will occur in the case there is a single page),
special measuring Item, worksheet , report and cursor shows state.
2. Image measuring area: Use the trackball and [Set] to activate the measurement display
of the caliper.
3. General measurement menu: Click the soft menu to select the measurements of the
corresponding general items.
4. Measurement result display area: Move the cursor to the top column and press
[Set], and rotate the trackball to press [Set] to move the measurement results to your
preferential area. This area can support 8 groups of data display, and the data after the
eighth group will automatically overwrite the previous measurement data.
User ’s Manual 89
Chapter 6 Measurements and Calculations
Caliper Type Description Mode
B mode: distance, tracing length,
++ To indicate the starting point
area, angle, volume, stenosis rate,
×× and end point of each pair of marks
A/B ratio
To draw the starting position for a
M mode: time, HR
vertical dotted line, capable of
PW mode: time, HR, PI (automatic),
marking the horizontal position or
PI (manual), flow
spacing
Measurement methods:
1. Press [Measure] to enable the measurement function;
2. Press menu [Trace Length];
3. Roll the trackball to position the first measurement mark, and press [Set], then the
system will fix this position.
4. Roll the trackball again to position the second measurement mark, and press [Set] to
complete the measurement, then the system will record the measuring results to the
result box on the screen.
User ’s Manual 91
Chapter 6 Measurements and Calculations
Measurement methods:
1. Activating measurement function;
2. Press menu [Area] and select [Ellipse];
3. Roll the trackball to position the first measurement mark, and press [Set]; roll the
trackball again, and the system will display an ellipse, press [Set] to position an axis of
the ellipse.
4. Roll the trackball again, the system will adjust the shape of the ellipse along its second
axis; press [Set], the system will display the measuring data to the result box.
5. Roll the trackball again, and press [Set] to position the ellipse on the image. If you are
not satisfied in the process of measurement, you can press [Update] to return to the
previous step and revise the measurements.
Ellipse+Distance method
Measurement methods:
1. Activating measurement function;
2. Press menu [Volume], and press the toggle to select [Ellipse+Distance].
3. Roll the trackball to position the first measurement mark, and press [Set], then the
system will fix the mark; roll the trackball again, and the system will display an ellipse,
press [Set] to position an axis of the ellipse.
4. Roll the trackball again, and the system will adjust the shape of the ellipse along its
second axis; press [Set].
5. Roll the trackball again to position the ellipse on the image and press [Set] to fix the
position. If you are not satisfied in the process of measurement, you can press [Update]
to return to the previous step and revise the measurements.
6. Roll the trackball again and move the caliper to position the starting point of the line
segment, then press [Set];
7. Roll the trackball again. Move the caliper to position the end point of the line segment,
then press [Set] to complete the measurement; the measuring results will be displayed
in the result box.
Formula:
π ( D1×D2xD3)
V=
6
Elliptic long axis D1, Elliptic short axis D2, Segment line D3. Formula unit: mm.
User ’s Manual 93
Chapter 6 Measurements and Calculations
1-Distance Method
Function: The system uses a single distance measurement to calculate the volume.
1. Activating measurement function;
2. Press menu [Volume] and then press [1-Distance];
3. Roll the trackball to position the first measurement mark, and press [Set], then the
system will fix this position, and display the second mark;
4. Roll the trackball to position the second measurement mark, and press [Set]. The
system will fix this mark, and the measuring results will be displayed in the measuring
result box.
Formula:
(π × D3 )
V =6
Segment length D, Formula unit :mm
2-Distance Method
Function: The system uses two distance measurements to calculate the volume. Use the
caliper to respectively measure the length and the depth. The measuring results are
successively displayed in the result box of the imaging screen. After the measurement is
completed, the measuring results will be displayed in the result box .
Formula:
(π × D1× D22 )
V =
6
Segment length D1,D2 ,Formula unit :mm
3-Distance Method
Function:Use the caliper to measure the three distances of D1 length, D2 width and D3
thickness of elliptical-similar object. The results of these linear measurements will be
accordingly displayed in the result box of the imaging screen. After the thickness is measured,
the volume will be displayed in the result box .
Formula:
(π ×D1×D2 ×D3)
V=
6
Segment length D1,D2 ,D3,Formula unit :mm
1-Ellipse Method
Function: Use the ellipse method to calculate the volume, and please refer to the Ellipse Area
Method for specific measurement methods. After the measurement is completed, the volume
will be displayed in the result box.
94 User ’s Manual
Chapter 6 Measurements and Calculations
Formula:
(8 × A2 )
V=
3π × D
Segment length D ,ellipse area A ,Formula unit :mm.
6.2.2.2 HR Measurement
Function: To measure the HR between two vertical dotted lines on the M image.
1. Click menu [HR] and a dotted line will occur; rotate the trackball to position the dotted
line and press [Set];
2. Move the trackball and another dotted line will occur, press [Set] to position the dotted
3. line. The HR measurements between two points will be displayed in the result box.
Measurement methods:
Click menu [Slope], and refer to Distance Measurement for the methods of slope
measurement, which will not be introduced in details here.
User ’s Manual 95
Chapter 6 Measurements and Calculations
6.2.3.1 Velocity
Function: This measurement is used to calculate speed.
1. Enter PW mode , and press [Freeze ] , then press [Measure] to enter the PW
measurement mode;
2. Click on menu [Velocity] to switch to the flow velocity measurements;
3. A measuring cursor arising from the intersection of vertical line and horizontal line
occurs, and roll the trackball to adjust the position of focus, press [Set] to position the
focus, and the velocity measurement will be displayed in the result box.
6.2.3.2 PI (Auto)
1. After entering PW measurement mode, Click the touch screen menu [PI Auto] to select
automatic tracing measurement;
2. A tracing caliper occurs on the screen, press [Set] to position the caliper and fix the
starting point;
3. Move the trackball to position the end point, and press [Set] for positioning;
4. 4The system will automatically fix these two calipers and find out the maximum value
between the two points. In addition, a caliper will automatically occur, and rotate the
trackball and press [Set] to confirm the second caliper; after the measurement is
completed, the measuring results will be displayed in the result box.
6.2.3.3 PI (Manual)
1. After entering PW measurement mode, Click the touch screen menu [PI Manual] to
select the manual tracing measurement;
2. A tracing caliper occurs on the screen, press [Set] to position the caliper and fix the
starting point;
3. Move the trackball for the maximum spectrum value required by manual tracing, and
96 User ’s Manual
Chapter 6 Measurements and Calculations
press [Set] to position the end point.
4. A caliper will occur, and rotate the trackball to set the second caliper; after the
measurement is completed, the measuring results will be displayed in the result box.
Measurement method:
1. Click on menu [Acceleration] and switch to the corresponding measurement.
2. Roll the trackball to place the caliper on the starting point, and press [Set];
3. Roll the trackball to place the caliper on the end point, and press [Set];
4. The measuring results will be displayed in the result box.
6.2.3.6 HR Measurement
Function: To depict a cardiac cycle through the measurement mark displayed in the form of
vertical line, based on which the HR can be determined.
1. Measurement methods:
2. Click on menu [HR].
3. Roll the trackball to place the caliper on the starting point of the cardiac cycle on the
spectrum, and press [Set];
4. Roll the trackball to place the caliper on the end point of the cardiac cycle, and press
[Set];
5. The measuring results will be displayed in the result box.
Measurement methods:
1. Click on menu [Flow Volume], in which there are [D-Flow] and [A-Flow], then select [D-
Flow];
2. Place the cursor in the starting point of PW mode image, and press [Set];
3. Roll the trackball to place the cursor on the end point, and press [Set];
4. Roll the trackball and move the caliper to the B mode image, then select the starting
point and press [Set];
5. Roll the trackball to place the caliper on the end point, and press [Set];
6. The measuring results will be displayed in the result box .
Measurement methods:
l Click on menu [Flow], in which there are [D-Flow] and [A-Flow], then select [A-Flow];
l Place the cursor in the starting point of PW mode image, and press [Set];
l Roll the trackball to place the cursor on the end point, and press [Set];
l Roll the trackball and move the caliper to the B mode image for area measurement
(Refer to Area Measurement in 2D-mode General measurement for details).
l The measuring results will be displayed in the result box.
98 User ’s Manual
Chapter 6 Measurements and Calculations
l Circumference measuring error: ± 3% or less;
l Area measuring error: ± 6% or less;
l Volume measuring error: ± 9% or less;
l When the ultrasonic beam axis is aligned with the blood flow, the measuring error of flow
velocity shall be within ±5%.
The measuring errors are acquired from assessment to the testing targets in the laboratory
environment.
In the process of diagnosis, the measurement data of the ultrasonic systems must be used in
combination with other clinical data, and no diagnosis shall be made only in terms of
measurement data. When the data are acquired by using the ultrasonic system, other factors
shall also be taken into account. Analysis of these factors has shown that
the accuracy of measurement and further calculation depend on the quality image to a large
extent. In turn, the image quality depends on system design and skills of operators as well as
their understandings of system features.
The ultrasound imaging algorithms are based on the sound velocity in the 1540 m/s tissues.
However, in different tissues, sound velocity may be different. In soft tissue, the error is
usually by 2%, and sometimes even reaches 5%, particularly when a fat tissue occurs in the
measured area.
In order to ensure the best measurement precision, please align the probe carefully and keep
the sound beam aligned with the direction of flow.
User ’s Manual 99
Chapter 6 Measurements and Calculations
ultrasonic diagnostic system.
Measurement methods:
1. Press [B] to enter the B mode measurement;
2. Press [Measure] and the abdominal measuring items will be displayed in the main menu
on the left; rotate the trackball to select the corresponding item and press [Set], then
move the trackball to the image for distance measurement (the measurement method is
displayed in the soft menu);
3. The measurement result is displayed in the result box; click the main menu or the
4. [Report] to view the report or worksheet.
3-Distance π
Renal volume (long diameter, thick diameter and Vol = × D1 × D2 × D3
wide diameter) 6
The reference author and corresponding estimating formula can be seen from the
following table:
Measurement method from fetal heart aortic arch diameter to right ventricle outflow tract
diameter:
1. Press [Exam/Item] to select the fetal heart measurement Item in the obstetric menu or
directly click [OB]; and click [Measure] to select [AoD Arch] in the main menu (taking
AoD Arch for example);
2. Place the third caliper on the starting point of D1 segment;
3. Press [Set];
4. Roll the trackball to place the second caliper on the end point of D1 segment;
5. Press [Set];
6. The measurements will be displayed in the result box of the screen, and saved to the
worksheet and report. If you want to continue with the following measurements, please
repeat steps 2-5.
Measurement Desired
Description Unit
Item Measurement
left ventricular end-systole volume and
left ventricular end-diastole volume are
estimated by measuring the distance from
left ventricular internal diameter (LVID ).
Teichholz Volume method:
EDV = 7(LVIDd3/1000)/(2.4 + LVIDd/10)
ESV = 7(LVIDs3/1000)/(2.4 + LVIDs/10)
LVIDd
EDV &ESV Cube: mL
LVIDs
EDV = LVIDd³/1000
ESV = LVIDs³/1000
Gibson:
EDV = 0.52(0.98*LVIDd/10 + 5.90)*
LVIDd2/100
ESV = 0.52(1.14*LVIDs/10 + 4.18)*
LVIDd2/100
Ejection fraction refers to the ratio
EF between stroke volume and end- diastolic EDV ESV %
volume: EF=SV/EDV*100
Preload Index RI
Time interval between beginning and
Tel Index a Time
end of mitral valve filling
Time interval between beginning and
Tel Index b Time
end of aortic flow
Left ventricular outflow tract maximum
LVOT Max Velocity
velocity
LVET Left ventricular ejection time Time
l From the graph, you can view the corresponding obstetric measurement items and the
selected growth curve of the reference authors, and these reference authors can be set
User ’s Manual 111
Chapter 6 Measurements and Calculations
in the Preset;Preset of reference author: press [Setup] on the t menu → Click [Measure
Preset] on the left of the screen → Click the sub-menu [Measure] on the screen, and
then click [Graph] for setting.
l There are three curves in the obstetric graph, respectively showing the maximum value,
normal value and minimum value of the measurement item of the fetus. Red point
represents the position of the current measurement value. The fetal growth can be
judged by viewing the growth curve.
l Click the [single]or [Quad] below the graph to carry out image switching and selection of
multiple births, as well as other functions such as graph print previewing, printing, page
up and page down, switching to worksheet or report, etc.
Vol = 0.7 × D1 × D2 × D3 or
Bladder-V 3-Distance π
Vol = × D1 × D2 × D3
6
3-Distance
π
Prostate-V
(long diameter, Vol = × D1 × D2 × D3
wide diameter and
6
thick diameter)
%Steno = ( A1 − A2 ) ×100
A1
A1, larger area; A2, smaller area.
Operating steps of area fractional shortening to measure arteries:
1. Press [Measure] to enter the measurement interface;
2. Measurement items will pop up in the main menu, and select the measurement item of
Carotid (taking Carotid measurement for example);
3. Click [A-%stenosis] in the soft menu:
4. Rotate the trackball and press [Set] to position the starting point of the ellipse;
5. Rotate the trackball and press[ Set] again to position an axis of the ellipse;
6. Then rotate the trackball and press [Set] to position the other axis of the ellipse;
7. And finally rotate the trackball to select the position of ellipse and press [Set] for
positioning;
8. After the measurement of an ellipse is completed, repeat steps 4-7 to complete the
measurement of another ellipse;
9. The measurements will be displayed in the result box of the screen, and saved to the
worksheet and report. If you want to continue with the following measurements,
10. please repeat steps 3-8.
Thyroid Measurement:
Users can conduct the following measurements on thyroid volume:
1. To measure thyroid right lobe size, click [Right] in the measurement items of the main
menu, and use the distance method to measure the length, thickness and width of the
right lobe, then the system will automatically calculate the right lobe volume in
accordance with the formula listed in the table below.
2. In order to measure thyroid left lobe size, please refer to the measurement method of
thyroid right lobe size.
3. After measuring the sizes of two lobes of the thyroid, the total sum is the size of the
whole thyroid. The results will be saved to the worksheet and report.
1. Click the thyroid measuring Item in the Small Organ menu of [Probe/Exam] or directly
click the Small Organ item; after clicking [Measure], select “Thyroid“ in the main menu;
to measure the left thyroid, click [Left] on the top of the menu;
2. Place the first caliper on the starting point of line segment;
3. Press [Set];
4. Roll the trackball to place the second caliper on the end point of line segment;
5. Press [Set];
6. Roll the trackball to place the third caliper on the starting point of line segment;
7. Press [Set];
8. Roll the trackball to place the fourth caliper on the end point of line segment;
9. Press [Set];
10. Roll the trackball to place the fifth caliper on the starting point of line segment;
11. Press [Set];
12. Roll the trackball to place the sixth caliper on the end point of line segment;
13. Press [Set];
14. The measurements will be displayed in the result box of the screen, and saved to the
worksheet and report.
Breast Measurements:
Click the Breast Measuring Item under the Small Part menu of [Probe/Exam] or directly click
the Small Organ item, and select “Breast” to measure breast thickness, Mass 1, Mass 2 and
Mass 3 through the distance method; refer to “Thyroid Measurements” for its operating
methods, which will not be repeated here.
Testical Measurements:
Click the Testicle Measuring Item under the Small Part menu of [Probe/Exam] or directly click
User ’s Manual 121
Chapter 6 Measurements and Calculations
the Small Organ item, and select “Testicle” to measure long diameter, thick diameter and
wide diameter through the distance method, then the system will automatically calculate the
testicular size; refer to “Thyroid Measurements” for its operating methods, which will not be
repeated here.
Note: The black font refers to the measuring Items of the following measurement items.
As shown in the table above, the measurement Item mainly includes Distance Method and
Trace/Ellipse Method, and operations of Distance Method are:
1. Press [Probe/Exam] to select [Cardiac], and enter B mode;
2. Press [Measure] and select MV measurement Item in the main menu (taking mitral valve
for example), and then click [MV] to display the measurement items of mitral valve;
3. Place the first caliper on the starting point of line segment;
4. Press [Set];
5. Roll the trackball to place the second caliper on the end point of line segment;
6. Press [Set];
7. Roll the trackball to place the third caliper on the starting point of line segment;
8. The measurements will be displayed in the result box, and saved to the worksheet and
report.
Calculation Desired
Description Unit
Item Measurement
Cardiac index refers to the cardiac output
per square meter of body surface area
(BSA):
CI = CO÷BSA.
If the height is expressed in cm, the weight will
CI (cardiac be expressed in kg: HR, BSA L/min/m
index) BSA=0.007184 x (weight0.425) x EDV ESV 2
(height0.725)
If the height in expressed in feet/inches, the
weight will be expressed in pounds:
BSA=0.007184 x (weight x 0.454)0.425 x (height
x 2.54)0.725
Cardiac output refers to the effective left
CO (Cardiac EDV
ventricular ejection volume per unit time: L/min
Output) ESV HR
CO=[(EDV – ESV) ÷ 1000] (HR).
Cubed Volume Method:
EDV = LVIDd3
ESV = LVIDs3
Teichholz Volume Method:
EDV = 7(LVIDd3 / 1000) / (2.4+LVIDd/10) ESV =
LVIDd
EDV, ESV 7(LVIDs3 / 1000) / (2.4+LVIDs/10) Gibson ml
LVIDs
Volume Method:
EDV =π/ 6*(0.98*LVIDd/10+5.90) * LVIDd/
10*LVIDd/10
ESV =π/ 6*(1.14*LVIDs/10+4.18) * LVIDs/
10*LVIDs/10
CA amp
CA/CE CA/CE=CA amp (mm) / CE amp (mm)
CE amp
Ejection fraction refers to the ratio between
EF (ejection
stroke volume and end-diastolic volume: EF = EDV ESV %
fraction)
100[(EDV-ESV)÷EDV]
LA
AO/LA Aorta / Left Atrium Ratio
AO
MVA
CO Cardiac output
HR Heart rate HR
AVA(VTI)
HR Heart rate HR
TV Tricuspid valve
PV Pulmonary valve
CO Cardiac output
HR Heart rate HR
When the measurements are completed, the system will perform the PW mode
calculation, and the calculation results will be displayed on the screen and saved to the
cardiac worksheet and report, of which the calculation and description of the measurement
items will be listed in the table below for reference.
Desired
Calculation Item Description Unit
Measurement
A/E= mitral valve A point
Mitral valve A point Mitral valve A
(m/s)/mitral valve N/A
and E point ratio (A/E) point (Distance)
E point (m/s)
Mitral valve E point and A E/A=MV E pt(m/s) / MV A Mitral valve E
N/A
point ratio (E/A) pt (m/s) point (Distance)
Mitral valve maximum MV PGmax(mmHg)=(MV Mitral valve maximum
pressure gradient (MV PG mmHg
max) Vmax(m/s))2*4 velocity
Left ventricular
LVIMP=(MV C-Odur(ms) – MV C.O duration LV
intramyocardial pressure --
LVET(ms)) / LVET(ms) ejection time (ET)
(LVIMP)
Right ventricular
RVIMP=(TV C-Odur(ms) – TV C.O duration LV
intramyocardial pressure --
RVET(ms)) / RVET(ms) ejection time (ET)
(RVIMP)
The measurement
Heart rate (HR) HR = 60/R-R interval. bpm
covers a cardiac cycle
Left ventricular outflow
MVA(VTI)( cm2) = tract diameter Left
Mitral valve area (Velocity
time integral) π*(LVOT diam(mm)/2/10) ventricular outflow tract
2
cm2
*(LVOT VTI (cm) / MV velocity time integral
[MVA (VTI)]
VTI (cm)) Mitral valve velocity
time integral
Left ventricular
AVA(VTI)(cm2) = π*(LVOT
outflow tract diameter
Aortic valve area (velocity diam(mm)/2/10)2 Left ventricular outflow
time integral) * tract velocity time cm2
[AVA (VTI)] (LVOT VTI(cm) / AV integral Aortic valve
VTI(cm)) velocity time integral
Left ventricular ((LVOT
LVOT Diam
outflow tract cardiac index Diam)2*0.785*(LVOT LVOT VTI HR
(LVOT CI) VTI)*(HR)/1000)/ BSA
Ratio between left
ventricular pre- ejection LVPEP/
LVSTI=LVPEP/ET LVPEP LVET
period and left ventricular ET
ejection time
3.14/4*(LVOT
Mitral valve area (velocity LVOT Diam
time integral)[MVA (VTI)] Diam)2*(LVOT VTI)/(MV LVOT VTI MV VTI
VTI)
Mitral valve area
MVA(P
(Pressure half-time) [MVA 220/(MV PHT(ms) MV PHT
HT)
(PHT)]
Ratio between right
ventricular pre- ejection RVPEP/
RVSTI=RVPEP/ET RVPEP ET
period and right ventricular ET
ejection time
Please refer to Appendix D for professional contents of cardiac measurement.
The following figure is the worksheet and report of cardiac measurement, in which you can
view the measurement results. If you are not satisfied with the result, you can enter the
worksheet to revise and save the data, which will be saved into the report.
6.3.10.3 EM Report
l The system will set the Report Database as the database of the current measurement
item by default;
l The basic information filled by users in the patient-Information interface will be
displayed in [Patient Information];
l For measurement data, you can click [Worksheet] to enter the worksheet for editing; in
case there are a large number of data, you can move the roll bar on the right of the
screen for viewing.
l Users can perform the functions such as image print, print preview, select images,
export, access to worksheet and page turning.
l You can click [Select Images] and in the thumbnail of main interface to select
images in the report. However, a maximum of four images can be simultaneously
displayed in the report.
l Users can enter relevant comments below each image.
l Click [Select Image] to pop up the interface below, and the left image is the one saved
in the current measurement item by clicking [Image Save], while the cine images in Cine
Save cannot be displayed in this interface:
Available Images The left column shows the images saved in the hard disk;
The right column shows the images that are selected into the
Selected Images
report;
Select all images in the left column to add to the report, which
are displayed in the [Selected Images]; if there are more than
four images uploaded to the left column, the report will only
show the first four images;
Section Three is shown in the following figure, in which users can enter the
If you want to check out the complete report, please click [Print Preview].
The worksheet can be entered in the same way as report, and you can change or delete
data in the worksheet and repeat measurements and save the final measurement result to
the report; as shown below,press [Report] to enter the report interface:
When the measurement is completed, patient reports and graphs can also be exported to an
optical disk or USB disk;
If you want to view the patient's images and reports, you can perform the following
operations:
1. First press [Patient] to enter the interface Patient Information, and search the desired
patient in the patient information list, and press [Set] to exit;
2. This patient's images will be displayed in the thumbnail of the main interface, or press
the [Review] button to enter the Reviewing interface; search the patient in the patient
information list, and lick [Image Screen] on the interface to view the images.
3. Press [Report] and then click [Worksheet] to view patient report and worksheet.
User ’s Manual 145
Chapter 7 Record of Images and Reports
When you insert a U disk, the icon will appear on the bottom right of the screen,
then you can store images into the U disk, and export the reports and images; click [Export]
or the thumbnail to pop up the export dialog box (as shown below):
l The default display of U disk is the letter following the name of the last disk of hard
drive, for example, if the system is named to D disk, the default optical drive is E disk
and displayed as CD/DVD E, then the first inserted U disk will be automatically
identified as USB F disk, and the second will be USB G disk, likewise; click [▼] on the
right of hard drive to select the location of data storage.
l When a data is stored, the system will automatically name the file and displayed in the
[File Name], and users can change the name by clicking.
l The list below the file name is the files in the U disk; double-click the folder to enter and
l Users can save images and reports by selecting different formats, for reports, there are
PDF, HTML and RTF; for images, there are JPEG, BMP and TIFF.
l Press [Set], and the data will be uploaded to the U disk and a prompt for successfully
sending data to the removable storage device will be displayed.
Note: U disk is suggested to use the 512M Netac U265 or 1 G Netac U208 matched by
Landwind; The system supports FAT32 file format, and before the U disk is used, it shall be
formatted into FAT32 on the computer. The system update can only be used by the service
staff of our company; all the consequences arising from the user's unauthorized operation
shall be undertaken by the user.
When you insert a CD/DVD, the icon will appear on the bottom right of the screen,
through which you can burn the reports and images; click [Export] in the menu or the
thumbnail to pop up the export dialog box, then you can select CD/DVD E from the
drop-down drive list, and modify the file names and data formats; after the changes are
finished, press [Set] to pop up the following dialog box:
l The system will automatically create a folder, click [Burn] to save the desired files into
this folder.
l You can view the burning progress through [Progress], and click [Close] when the
burning is completed.
l Close Media: When clicking [Close Media], the system will not perform a second burning
after the optical disk finishes its burning.
l Eject When Finished: When clicking [Eject When Finished], the CD/DVD burning is
finished and will be automatically ejected.
Export: to export the patient information on the Local HD to an optical disk (CD/DVD)
or USB disk.
1. Click [Export], and the patient information list on the first row will pop up the patient data
stored in the system;
2. On the bottom left corner occurs from [Local HD] to [], and click the drop-down list of []
to select the export device;
3. Click [Select All]or select single patient information as required;
4. Then click [Data Transfer] to transmit the patient data from the system to the
removable device.
Import: to import patient information from an optical disk (CD/DVD) or USB device to
the Local HD.
1. Click [Import], and the patient information list on the second row will pop up the patient
data stored in the system;
2. On the bottom left corner occurs from [] to [Local HD], and click the drop-down list of []
to select the import device;
3. Patient data in the import device will be displayed in the patient information list in the
first row;
4. Click [Select All] or select single patient information as required;
5. Then click [Data Transfer] to transmit the patient data from the removable device to
the system.
In the DICOM Job Spooler, you can carry out the operations to the uploaded images, such
as refresh, delete, resend, select an uploading location, hide and select all, etc..
Chapter 8 Preset
8.1 Preset
Press [Setup] on the keyboard to enter the preset interface, through which users can preset
the system parameters according to their own needs for faster and accurate data mea-
surement. The preset menu includes System Preset, Exam Preset, Imaging Preset, Measure
Preset, Text Preset, BodyMark Preset, Peripheral Preset and Service. We can change the
settings according to our needs.
Parameter Description
Enter the hospital name in the keyboard, and press Ctrl + space
Hospital Name
key to switch language input.
Select the appropriate language from the drop-down list:
Language
Chinese, English;
Time Format Time format selection: 12 h, 24 h
Date Format Date format selection: YY/MM/DD, MM/DD/YY, DD/MM/YY;
System Date Input system date and select calendar;
System Time Input system time for and select time;
Weight & Height Unit Select measurement units: Metric System, American System
Format (Metric system available only)
Surface Area Formula Set the surface area formula: East, West (East available only)
Note: After the language is switched, the system will restart to enable the new language;
please save the data before switching the language.
Parameter Description
Select: when saving images, and if there is no patient information,
the system will automatically create a patient ID, which will be
Save the images when
displayed in the report and on the main screen;
without patient info
Unselect: When saving images, and if there is no patient
information, it will prompt to create a new patient.
Select: When a new patient is created, it will automatically
generate a patient ID (date + current patient number, for example,
Auto generated ID 20101122), which can also be edited;
Unselect: To create a new patient, you can manually enter the
patient ID.
Available ports when starting the system: Port1, Port 2 and Port
3;
Active probe port
If the preset port is not installed with the probe, the system will
automatically identify the next port in sequence;
Trackball travel speed Select trackball movement speed: high, medium and low.
Doppler Audio Volume Regulate Doppler audio volume: 1% to 100 %.
Select the image area to be stored: image area 640 ×480
Image Size
Full screen 1024 x 768
Edit the stored cine time, and support backward storage.
Cine Store Time
Storage Time: 0 ~ 120 s, Default Value: 60 s.
Select: When there is no system operation exceeding the setting
Enable Screen save time, the system will enter the screensaver status. Unselect: The
screen saver function is not enabled;
Screen saver Type Select the screen-saver type: black screen, text3D ;
Screen saver Waiting Select the screensaver waiting time from the drop-down menu
(min): 5, 10, 15, 15, 20 and 25.
Parameter Description
Probe Type N/A
Bracket Type N/A
Default Angle N/A
Angle 1, Angle 2 and
N/A
Angle 3
Initial Position N/A
Select: activate Biopsy depth marking every 0.5 cm.
Enable 0.5cm Markers
Unselect: no depth marking.
Biopsy Guide Line size Select the size of Biopsy guide wire: small, medium and large
There are three customized function keys on the right of the control panel, and you can
carry out the following shortcut settings in this interface:
l None
l Acquire Set
l Biopsy guide
Parameter Description
Caliper Size Caliper size setting: large, medium, small;
Caliper Pattern "+" and "X " display option;
Measurement Result Measurement result background display:
Background Display: transparent; Non-display: non-transparent
Result Font Size Measurement result font size: small, medium, large
Result Font Color Final measurement result font color: white, yellow, blue
Distance Units of general distance measurements and distance-type
measurement results: mm, cm
Circumference Units of general circumference measurements and
circumference-type measurement results: mm, cm
Area Units of general area measurements and area-type measurement
results: mm2, cm2, m2
Volume Units of general volume measurements and volume-type
measurement results: mm3, cm3, L
Velocity Units of general velocity measurements and velocity-type
measurement results: mm/s, cm/s, m/s
Time Units of general time measurements and time-type measurement
results: ms, s
Slope Units of general slope measurements and slope-type
measurement results: mm/s, cm/s, m/s
Acceleration Units of general acceleration measurements and acceleration-
type measurement results: mm/s, cm/s2, m/s2
Parameter Description
Default B mode measurements Value: distance, Trace length,
B mode
area, angle,volume, stenosis rate, A/B ratio
Default M mode measurements Value: distance, HR, slope, time,
M mode
A/B ratio;
Default PW mode measurements Value: velocity, PI (Auto),
PW mode
acceleration,velocity rate, HR, RI, PI (Manual), flow, time;
Area Methods of setting default area measurement: ellipse, trace;
Methods of setting default volume measurement: 1-distance, 1-
Volume
ellipse, 2-distance, 3-distance, ellipse+distance;
Methods of setting default stenosis rate measurement: D-
%stenosis
%stenosis, A- %stenosis;
Methods of setting default B-mode ratio measurement: D ratio,
B mode Ratio
area ratio and volume ratio;
Methods of setting default M-mode ratio measurement: D ratio,
M mode Ratio
time ratio;
Flow volume Methods of setting default flow measurement: A-flow, D-flow.
4. [Select item], then press on the right to delete it; to add a new measurement item,
you can click , and the item will be selected to [Select item];
5. Click to move the selected item to the upper location; click to move the
selected item to the lower location;
6. Click [OK] to save the measurement item settings and return to main interface; click
[Apply] to save settings; click [Cancel] to return to the main interface without saving
settings.
click to select the annotation library on the left; when clicking , all the items in
the annotation library of the current item will be added.
3. You can manually enter the annotation on the upper-right corner, and click [Add] to
add the annotation into the annotation library; select the annotation and click [Modify]
to revise the annotation in the annotation library;
4. Click to move the selected item to the upper location; click to move the
5. selected item to the lower location;
6. Click [OK] to save the measurement item settings and return to main interface; click
[Apply] to save settings; click [Cancel] to return to the main interface without saving
settings.
Parameter Description
Text Font Size Annotation text font size: 10, 12, 14, 18, 20; unit: Pound(s)
Set annotation text font and arrow color, and values are: yellow,
Text Color
red, blue, orange, white and purple;
When selected, it will automatically wrap when the text length
Word Wrapping exceeds the image boundary; otherwise it needs to wrap
manually;
Arrow Size Set the arrow size: small, medium, large
erase Text Image is Select the text, and the annotation will be erased when the image
Unfrozen is frozen; otherwise the erasing function will not be performed.
Retain while Entering When selected, upon entering and exiting the timeline mode, the
or Leaving Time Line system will reserve the annotation on the monitor display;
Mode otherwise it will not reserve the annotation;
When the text annotation is activated, set whether the default text
Default Text Library
annotation based is displayed: on, off.
Parameter Description
Host Name Host name of the ultrasonic system
IP Address IP address of the ultrasonic system host
Subnet Mask Subnet mask of the ultrasonic system host
Gateway Default gateway of the ultrasonic system host
AE Title Application entity title of local DICOM service
Port Store service port number (local port number). 104 by default
Largest PDU packet size and range: 16k to 64k. If the settings
PDU
are beyond this range, the system will set 32k by default.
Printing device selection: In accordance with differences of peripheral signal ports, the
printers can support:
l Numeric Print: to print numerically (including picture and text print and digital video
print):
l Analog Print: to print analog videos.
l Disk Storage and DICOM Storage:When Choose this and return the interface, press
[print ] , the Image will be stored in Disk and send to DICOM .
Add Printer: click the button to enable "Add Printer Wizard". You can add the printer driver
from the system or optical disk according to your needs; after the completion of adding, the
list will show the added printer model and status information (see Chapter 6 Add Printer).
Remove Printer: select a printer from the printer list, and click [Delete] to remove the printer.
Parameter Description
You can set two types of video signal output sizes:
Output size
Full-screen (1024 * 768); image area (640 * 480).
Output Format Select the standard for video printer output. PAL, NTSC
After video reserve is started, and if the video output is full-
screen, the image area remains unchanged, and non-image
Video Invert video signal will be reverse; if [image area] is selected, the video
inverse will be automatically restricted to close, that is, the video
reserve would only be enabled in full- screen output.
This system can connect a USB footswitch to replace some control panel keys.Before
connect you need to setup the footswitch as below:
Connect the footswitch to a computer,and insert the footswitch driver to define the key.
Define CTRL+SHIFT+Left key as left pedal,CTRL+SHIFT+right key as right pedal.
l Connect the footswitch
1. Connect the footswitch to the USB port of the system.
2. Preset the footswitch function as shown above the figure.
3. Step on the footswitch in the main surface to perform the function.
Note: Before importing preset, please save the measurement data; otherwise the system
restart will lose data.
Chapter 9 Maintenance
9.1.2 Maintenance
Note: To maintain the safety and functionality of the ultrasound system, maintenance must
be performed every 12 months. Electric safety test must be performed at periodic intervals
according to the local safety regulations or as needed.
Repair:
For the repair or replacement of system equipment parts, please contact Landwind or your
dealer.
Note: To avoid electrical shock and damage to the ultrasound system, always power off and
disconnect the equipment from the AC power source before cleaning and disinfection.
Note: Do not pour any fluid onto the system surfaces, as fluid seepage into the electrical
circuitry may cause excessive leakage current or system failure.
Note: remove gel residuals out of the probe after each use!
Note: To avoid the possibility of static shock and damage to the system, avoid the use of
aerosol spray cleaners on the monitor screens.
To remove and clean the holders for probes and coupling agent (gel):
Use neutral cleaning agent to clean the surfaces of the holders for probes and coupling
agent, and then use a piece of lint-free cloth to dry them.
Note: Do not drop or place foreign objects inside the trackball assembly. This may affect the
trackball’s operation and damage the system.
Cleaning
l Clean the bezel and trackball with a cotton swab or lint-free pad moistened with mild
detergent solution.
l Clean the inside of the trackball assembly, using a cotton swab moistened with mild
detergent solution.
l Allow the trackball components to completely dry before reassembly.
Caution: Ensure surfaces of the probe are always disinfected during procedures requiring
disinfection.
Caution: To minimize the risk of cross contamination and infectious diseases, endocavity
and intraoperative probes must be cleaned and high-level disinfected after each use.
Caution: Probes are sensitive devices——irreparable damage may occur if they are dropped,
knocked against other objects, cut, or punctured. Do not attempt to repair or replace any part
of a probe.
Caution: To avoid cable damage, do not roll the system over probe cables.
Caution: Take extreme care when handling or storing probes. They must not be dropped,
jarred, or knocked against other objects. Do not allow probes to come into contact with any
sharp-edged or pointed object.
Caution: Maintenance and repair should only be carried out by the duly authorized personnel.
Do not attempt opening the probe or connector. This could void your warranty!
Caution: All probes, no matter what their design is, are sensitive to strikes. And therefore,
handle them with care. Pay attention to any cracks, through which a current-conducting liquid
can enter the probes!
Caution: Prevent twisting, kinking, and looping of the probe cables, as well as protect them
from exposure to mechanical factors, in particular, from getting to under equipment wheels or
feet!
Caution: Damage to the probe due the above causes could void your warranty!
Caution: The system and probes should be inspected regularly for damages by the duly
authorized personnel!
Warning: Damages to the probe or probe cable can affect safe operation of the system
and therefore, repair them immediately!
You should inspect periodically the probes on the following signs of damage(pins shouldn’ t
be bent or damaged).When you are not using the probe, place it in the probe holder located
on the side on your system to assure safe, convenient storage.
To improve acoustic contact, use only a special ultrasound transmission gel. Never use the
following substances to improve acoustic contact:
1. Mineral oils;
2. Mineral oil based gels;
3. Gels containing lotions or any softeners;
Cleaning
Remove gel residuals out of the probe after each use. Where required, wipe the
probe and cable with a water-moistened gauze pad. Wipe the probe by any of the
recommended disinfectant. Remove disinfectant solutions residuals out of the probe with a
water-moistened gauze pad. Do not allow the disinfectant solution drying up on the probe.
Do not wipe the working surface of the probe with hard material, do not use a stiff brush to
clean the probe, and do not scrape gel residuals with sharp tools.
Disinfection
You should disinfect the probe after the probe lens contacts affected tissues, mucous
membranes and blood. The essential disinfection level of the ultrasound probe is defined by
tissue type, probe contacted with.
l Unplug the probe.
l Sterilize the probe with isopropanol in terms of requirements.
To prevent probe damage, follow the immersion depth guidelines for each probe type.
Caution: To avoid electrical shock and damage to the system, disconnect the probe prior to
cleaning or disinfecting.
Caution: Do not immerse the probe cable or the connector housing in liquid and prevent
them from soaking!
Caution: Do not immerse the probe in a disinfectant for more than 1 hour! Wipe the residual of
disinfector solution, if any, from the probe with a clean cloth which dips bolder water. Then
cool the probe to room temperature. Forbid to air dry the disinfector solution on the probe.
After disinfection, wipe the probe with a dry clean cloth. When cleaning, avoid any the
Caution: Never sterilize the probe with sterilization techniques such as gas or heat
sterilization techniques. It can cause the serious probe damages!
Caution: Do not use abrasive cleaning agents, organic solvents such as benzene,
isopropyl alcohol, or phenol-based substances, or cleaning agents containing organic solvents
to clean or disinfect probes. These substances can damage the probes.
Caution: Do not use a spray cleaner on a probe, as this may force cleaning fluid inside the
housing and damage the probe.
Caution: All probes should be cleaned and disinfected prior to their use on each patient.
Endocavity and intraoperative probes require high-level disinfection prior to use.
9.3.3 Storage
Store the probes in a clean and dry environment. Extreme temperature or humidity may
damage the probes.
Caution: Do not immerse the probe in a disinfectant for more than 1 hour!
Caution: Do not use abrasive cleaners, substances based on acetone, phenol, ethyl alcohol,
dechlorinating agents and organic solvents to clean and sterilize the probe!
Caution: Before use, examine the coupling agent for any material flaws. Thinning,
bulging, or brittleness of the material indicates damage. Any product showing flaws should not
be used.
Do not store coupling agent below 5°C nor above 57°C. Coupling agent has a limited shelf
life. Before use, examine these products for any material flaws. Some packaging may list an
expiration date. Any product showing flaws, or whose expiration date has passed, should
not be used.
Appendix A Specifications
Protective Measures:
l Liquid Protection: This system does not have the function of liquid protection;
l Heat Protection: There are two fans installed on the back of the system host for heat
protection.
Note:
1. Far away from equipment with strong electric and magnetic field and high voltage;
2. Avoiding the exposure of the screen in direct sunlight; light shielding indoors for easy
observation; good ventilation and anti-dust measures;
3. Operate system in the environment GB/T14710-93 specifying for.
A.3 Monitor
Monitor: High-resolution 15”TFT LCD, noninterlaced scan.
M34
Resolution:
1024x768 pixels.
Synchronization: Horizontal:30 Hz to 70 Hz; Vertical:50 Hz to 85 Hz.
Controlled tilt/rotation: Tilt angle: Up/down 90°; Rotation angle: Left/right 90°.
Safety classification: Class 1 in accordance with IEC 60601-1/EN60601-1.
A.4 DVD-Drives
Reading speed: 48x (7.2 Mb/s); Caching: 2M;
CD/DVD-R and Writing speed: CD-R: 48x (7.2 Mb/s); CD-RW: 32x (4,8 Mb/s);
CD/DVD-RW drive : DVD+/-R 20X DVD+RW 8X
Access time:100 ms.
A.6.1 B Mode
Imaging Mode B, real-time 2B, non-real-time 2B,4B
Frequency 2.5MHz~5 MHz
DR 30~120
Persist 0~4
Maximum Scan Depth 31cm
R/S 0~3
Colorize 0~15
Fov 19.2~36.1fps
Rejection 0~5
Smooth 0~3
LanSRI 0~4
Map A~P
Flip U/D,L/R
LanTissue Normal,Liquid,Fat,Muscle
Frame counter 1055
A.6.2 M Mode
Imaging Mode B/M, Full-M
Frequency 2.5MHz~5 MHz
DR 30~120
Sweep 1~7
LanSRI 0~3
Colorize 0~15
Map A~P
Rejection 0~5
A.6.5 PW Mode
Imaging Mode B/PW, Full-PW
Frequency 2.2MHz~2.6 MHz
PRF 15.87KHz~732.49Hz
Gate 0.5~20
Colorize 0~11
DR 30~60
Filter 0~4
Angle -60~60
Invert Available
Map A~H
Rejection 0~5
Sweep 1~7
Auto Trace Off, Up, Down
TriPlex/Duplex On,Off
Format V1:1,V1:2,V2;1,H2:3
Ip (MM)/Ip (mm) 5 65 5 65 75 5 65
ssr(Hz) - - 24 24 - 14 14
Dimensions of output
1×13 16×13 1×13 16×13 16×13 1×13 16×13
beam (mm)
Fawf (MHz) 3.2 3.2 3.2 3.2 2.5 2.8 2.8
APF (%) 0 0 0 0 0 0 0
Power-up mode B B B B B B B
AIF (%) 0 0 0 0 0 0 0
Initialization mode B B B B B B B
Acoustic output freeze Yes Yes Yes Yes Yes Yes Yes
Parameter M B PW Color
Ip (MM)/Ip (mm) 14 14 20 14
ssr(Hz) - 55 - 14
Dimensions of output
4×4 4×4 4×4 4×4
beam (mm)
Fawf (MHz) 6.6 6.6 5.1 5.2
APF (%) 0 0 0 0
Power-up mode B B B B
AIF (%) 0 0 0
Initialization mode B B B B
Parameter M B PW Color
Ip (MM)/Ip (mm) 12 12 45 12
ssr(Hz) - 55 - 24
Dimensions of output
3.69×9 3.69×9 6.56×9 3.69×9
beam (mm)
Fawf (MHz) 5.81 5.81 4.57 4.71
APF (%) 0 0 0 0
Power-up mode B B B B
AIF (%) 0 0 0 0
Initialization mode B B B B
Freq Frequency
MI Mechanical index
Mid Middle
P Output power
td Pulse duration
TI Thermal index
b
p