Specification Sheet: Tamson Low-Temperature Circulator
Specification Sheet: Tamson Low-Temperature Circulator
Specification sheet
TLC15-5
Tamson Low-Temperature Circulator
General Safety
The TLC15-5 is a very compact general purpose The bath conforms to CE regulation. It further is
circulator and this apparatus can be used in various equipped with a mechanical adjustable safety
applications where cooling is needed. Experiments, thermostat. The TLC15-5 is also equipped with a float
refractometers, rotary evaporators, distillation units, to warn the user when the bath fluid is too low.
water baths, flash point testers all are applications of
where the TLC15-5 can be used for. Because of its RS232
very small size, the TLC15-5 can easily be fitted into The bath comes standard with RS232 communication
your laboratory, either on or under the workbench. and when using the free software tool Tamcom, it
provides temperature logging or a predefined
The apparatus holds a bath containing 5 litres and the temperature set point curve.
filling opening measures 85 x 150 mm. Due to its
special design the system is extremely quiet, compact Control mechanism
and saves costly laboratory space. Readout is in The TLC15-5 combines a wide temperature range and
degrees °C or on in °F. maintenance free -long life- cooling system. Accurate
temperature regulation is provided by a
Cooling medium microprocessor controlled PID system of which the PID
The cooling system inside the TLC15-5 is ozone parameters can be set freely. The cooling system
friendly and doesn’t contain any CFK/HCFK gas. provides rapid cooling down of the 5 litres bath
content.
Span
The working temperature ranges from:
-15..+60C / -5..+140F
Optional feature:
-15..+120°C / -5..+284°F
Accuracy
The set point can be set in steps of 0.01C. The overall
system accuracy is better than 0.02C. Temperature
readout in 0.01C.
Pump
A pressure pump provides circulation in the bath or via
an external circuit. The pump pressure is 0.3 Bar and 7
litres per minute (no counter pressure). Optional is a
performance pump, P/N 04T0225 (pump pressure 0.7
Bar, flow max 10 L/min) .
20
Temperature
10
[°C]
0
TLC15-5
-10 Time
[hrs:min]
-15
-20
10
0
0.2 0.4 0.6 0.8
Pressure [Bar]