All the necessary precautions have been taken to ensure that the information set
out in this manual is correct at the time of its publication. Nonetheless Biosystems,
S.A. reserves the right to make any changes that may be necessary without notice,
as an inseparable part of the product's ongoing development.
Any change made to the instrument by the client will render the warranty void
and without effect.
http://www.biosystems.es
Biosystems
4
Table of contents
Staff for whom this manual is intended.............................................11
Notices and warnings........................................................................11
Software usage licence.......................................................................17
1. Intended use...............................................................................18
2. Contents.....................................................................................18
2.1. Content of the accessory box................................................................18
3. Identification of the main components.......................................21
4. Installation..................................................................................23
4.1. Location...............................................................................................23
4.2. Front door opening..............................................................................24
4.3. Installing the waste containers and washing solution...........................25
4.3.1. Preparing the washing solution..................................................................... 25
4.3.2. Emptying the high contamination waste container........................................ 26
4.4. Purified water connection....................................................................26
4.4.1. Positioning the tubes..................................................................................... 27
4.5. Low contamination waste connection..................................................28
4.6. Installing the sample and reagent rotor................................................28
4.7. Specifications of the barcode labels......................................................29
4.8. Affixing the identification labels...........................................................30
4.9. Installing the reaction rotor.................................................................31
4.10. Connection to the mains and start-up..................................................31
4.11. Connection to the computer................................................................33
4.12. Installing the user programme in the computer...................................33
4.12.1. Energy options configuration........................................................................ 36
4.12.2. Configure programmes in second plane........................................................ 37
4.12.2.1. Flash updates........................................................................................ 37
4.12.2.2. Java updates.......................................................................................... 38
4.12.2.3. Operating system services configuration................................................... 38
4.12.3. Cyberseurity configuration............................................................................ 39
4.12.3.1. Execution rights.................................................................................... 39
4.12.3.2. Configure the application user rights....................................................... 39
4.12.3.3. Activate and configure the operating system antivirus............................... 39
4.12.3.4. Run the antivirus before installing the application................................... 41
4.12.3.5. Do not install any extra program............................................................ 41
4.12.3.6. Update of windows versions................................................................... 42
5
User manual
Biosystems
6
10.2.4. Arrangement of tests for report..................................................................... 74
10.2.5. Barcode......................................................................................................... 75
10.2.6. LIS configuration.......................................................................................... 76
10.2.6.1. Work session setup................................................................................. 76
10.2.6.2. Communication settings......................................................................... 77
10.2.6.3. Protocol................................................................................................ 78
10.2.7. LIS mapping................................................................................................. 79
10.2.8. Users............................................................................................................. 80
10.2.9. Change user.................................................................................................. 82
10.3. Programming.......................................................................................82
10.3.1. Tests.............................................................................................................. 82
10.3.1.1. Ordering of tests.................................................................................... 83
10.3.1.2. Test parameters: general......................................................................... 84
10.3.1.3. Test parameters: procedure...................................................................... 85
10.3.1.4. Test parameters: calibration and blank.................................................... 87
10.3.1.5. Test parameters: quality control.............................................................. 88
10.3.1.6. Test parameters: options......................................................................... 89
10.3.2. Calculated tests............................................................................................. 91
10.3.3. Contaminations............................................................................................ 93
10.3.4. Profiles.......................................................................................................... 94
10.3.5. Calibrators.................................................................................................... 95
10.3.6. Controls........................................................................................................ 97
10.3.7. Patient data................................................................................................... 99
10.3.8. ISE test....................................................................................................... 100
10.3.9. External test................................................................................................ 102
10.4. Initialising and switching off the analyser..........................................103
10.5. Work session......................................................................................104
10.5.1. Sample request............................................................................................ 104
10.5.2. Additional entering of patient information................................................. 109
10.5.3. Rotor positioning........................................................................................ 111
10.5.4. Creating the worklist from the sample positioning screen............................ 114
10.5.5. Worklist setup............................................................................................. 115
10.5.6. Save session................................................................................................. 116
10.5.7. Load session................................................................................................ 116
10.5.8. Delete session.............................................................................................. 117
10.5.9. Delete virtual rotors.................................................................................... 117
10.6. Current state monitor........................................................................117
10.6.1. Main........................................................................................................... 117
10.6.2. Worksession................................................................................................ 118
10.6.3. Sample and reagent..................................................................................... 120
10.6.4. Reactions.................................................................................................... 121
10.6.5. ISE module................................................................................................. 122
10.6.6. Alarms........................................................................................................ 123
10.7. Worksession results............................................................................124
10.7.1. Results by patient........................................................................................ 124
10.7.2. Results by test............................................................................................. 126
10.7.3. Reaction graphs.......................................................................................... 129
10.7.4. Result repetitions........................................................................................ 130
7
User manual
Biosystems
8
14.3. Maintenance.......................................................................................172
14.3.1. Changing the reaction rotor........................................................................ 172
14.3.2. ISE module maintenance............................................................................ 172
14.3.2.1. Changing the electrodes........................................................................ 172
14.3.2.2. Changing the reagent kit...................................................................... 175
14.3.2.3. Changing the peristaltic pump tubes..................................................... 176
14.3.2.4. Turning off the ISE module for a long period of time ............................ 178
14.3.2.5. ISE module reactivation...................................................................... 179
14.3.3. Maintenance frequency............................................................................... 180
14.3.4. End of the analyser's useful life................................................................... 180
15. Technical characteristics............................................................181
15.1. General characteristics.......................................................................181
15.2. Sample and reagent management.......................................................181
15.3. Reaction rotor....................................................................................182
15.4. Cuvette washing system......................................................................182
15.5. Optical system....................................................................................182
15.6. ISE module (optional)........................................................................182
15.7. Environmental requirements..............................................................183
15.8. Dimensions and weight......................................................................183
15.9. Electrical requirements.......................................................................183
15.10. Fluid requirements.............................................................................183
15.11. Minimum computer requirements.....................................................183
16. Measuring and calculation procedures......................................185
16.1. Operating sequence. Preparation and reading cycles..........................185
16.2. Calculation of the absorbances...........................................................186
16.2.1. Endpoint monoreagent............................................................................... 187
16.2.2. Endpoint bi-reagent.................................................................................... 188
16.2.3. Differential switch....................................................................................... 189
16.2.4. Fixed time monoreagent.............................................................................. 189
16.2.5. Fixed time bi-reagent.................................................................................. 190
16.2.6. Kinetic monoreagent................................................................................... 191
16.2.7. Kinetic bireagent......................................................................................... 192
16.3. Concentration calculation..................................................................193
16.4. Repetition criteria..............................................................................195
16.5. Calculation of the ISE ion concentration...........................................196
16.6. Internal Quality Control....................................................................198
16.6.1. Basis............................................................................................................ 198
16.6.2. Admissible value interval............................................................................. 198
16.6.3. Selection of control rules............................................................................. 199
17. Summary of workflow scenarios with the LIS...........................201
9
User manual
Biosystems
10
Staff for whom this manual is intended
This manual is intended for the use of clinical laboratory professionals who will
use the BA200 analyser to determine analyte concentrations.
This manual describes the characteristics and general operating concepts of the
BA200 analyser. The installation, programming, execution and maintenance
procedures are described in detail.
Symbol Description
The symbol warns of operating risks that could cause personal injury.
WARNING
The symbol warns of a potential biohazard.
BIOHAZARD
The symbol warns of potential damage to the system or unreliable results.
CAUTION
The symbol warns that the information requires your attention.
NOTE
Risk of electric shock
11
User manual
Explanations of the symbols used on the analyser labels and in the manual
Symbol Description
This product is compliant with directive 98/79/EC on medical devices for In
Vitro Diagnostics.
Serial number
Expiry date
Lot code
Catalogue number
Temperature limit
Manufacturer
Irritant
Date of installation
Distilled water
Biosystems
12
Symbol Description
Caution, Mechanical parts in motion, keep clear
Safety precautions
Symbol Description
Preventing electric shock
To reduce the risk of electric discharges, do not remove the analyser cover.
There are no parts inside that can be repaired by the user, for which reason it is
necessary to contact the technical assistance service.
Preventing biological risks in handling the samples
Inappropriate handling of samples, controls and calibrators could cause
biological infection. Do not touch the samples, mixtures or waste with your
BIOHAZARD hands. Wear gloves and protective clothing when necessary.
In the event that the samples come into contact with the skin, wash
immediately with abundant water and seek medical advice. It is advisable to
follow good laboratory practice.
13
User manual
Symbol Description
Prevention in handling reagents
Handle reagents and washing solutions with care, they contain substances that
could be corrosive.
WARNING In the event that the reagents or washing solutions come into contact with the
skin, wash immediately with abundant water and seek medical advice.
Consult the reagent or washing solution adaptation sheet and follow the safety
instructions.
It is advisable to follow good laboratory practice.
Preventing biological risks in handling liquid waste
Handle the high contamination waste container with care. Wear gloves and
protective clothing when handling the container.
BIOHAZARD Dispose of the waste in accordance with national or local legislation for
disposing of dangerous biological waste, and consult the reagent manufacturer
or distributor for more details.
Preventing biological risks in handling solid waste
Take care in handling parts of the analyser that are converted to waste such as
the reaction rotor, sample tubes and reagent bottles.
BIOHAZARD Wear gloves and protective clothing when handling such waste.
Dispose of the waste in accordance with national or local legislation for
disposing of dangerous biological waste, and consult the reagent manufacturer
or distributor for more details.
Prevention of electro magnetic interferences
The analyser complies with the requirements with respect to emissions
and immunity set forth in the standard UNE -EN 61326-2-6:2006. This
NOTE equipment has been designed and tested for class B of standard UNE-EN
55022:2000. In a household environment, it may cause radio interference, in
which case the necessary measures must be taken to mitigate such interference.
Do not use the analyser near strong electro magnetic radiation sources (such
as centrifuge appliances, radio transmitters, mobile telephones), as they could
interfere with its correct operation.
Preventing laser light emission risks
The analyser has a bar code reader that emits laser light. The reader only works
when the analyser is in the execution mode and its rotor covers are in place. In
the event of a failure or during adjustment by technical maintenance staff, the
light beam could be activated without the cover in place; in such cases, do not
look directly at the laser beam.
Prevention at the end of the analyser's useful life
At the end of the useful life of the analyser, disposal of the product must be
carried out in accordance with the environmental legislation in force in each
country. In EU member states, the terms of the WEEE directive on electrical
and electronic appliances will apply. In other words, when the appliance's
useful life has ended, it is converted into waste and must be separated from
household waste for correct recycling. For this purpose, contact the distributor
for the product to be properly recycled.
Biosystems
14
Screenshots The screenshots shown in this manual have a merely illustrative function. They
do not necessarily reflect valid data.
Abbreviation Definition
∅ Diameter
ASTM American Society for Testing and Materials (www.
astm.org)
EC European Community
EMC Electromagnetic compatibility
CTRL Control key on the computer keyboard
EN European norm
F Fast (fuse type)
FUS Fuse
HL7 Health Level Seven (www.hl7.org)
IHE Integrating the Healthcare Enterprise (www.ihe.net)
ISE Ion-selective electrode
IVD In Vitro Diagnostics
LED Light-emitting diode
LIS Laboratory information system
prep Preparation
WEEE Waste Electrical and Electronic Equipment
REF Reference solution for the ISE unit
UPS Uninterruptible power source
TAS Technical assistance service
SD Standard deviation
ES Electrical safety
USB Universal Serial Bus
UV Ultraviolet
Units Definition
" Inch
˚C Degrees centigrade
A Ampere / Absorbance
GB Gigabyte
h Time
Hz Hertz
Kg Kilogram
L Litre
MB Megabyte
15
User manual
Units Definition
m Metre
min Minute
mL Millilitre
mm Millimetre
mmol Millimol
mv Millivolt
nm Nanometre
s Second
VA Volt-ampere
V Volt
W Watt
µL Microlitre
µm Micrometre
Biosystems
16
Software usage licence
BioSystems, S.A., the exclusive owner of all the rights over this computer applica-
tion, grants one non-transferable, non-exclusive licence governing the use of the
computer application to the user, who accepts it solely and exclusively for executing
this computer application in a single central processing unit (CPU) of a computer.
This licence does not permit the execution, use, access, reproduction, transforma-
tion, translation, lease, sale, distribution, commercial exploitation or provision to
third parties in any way, and above all in a computer network or through remote
access technologies, of all or part of the content included in this DVD.
BioSystems, S.A. will in no case be liable or assume payment of any compensation
whatsoever:
with respect to any infringement of the intellectual and/or industrial property
rights of others caused by copying images, audio and/or text as part of the content
of this DVD;
due to the exhaustiveness or accuracy of the data incorporated through incorrect
use of the computer application in the DVD;
for damages, losses or indirect, special, incidental or consequential damages to
people or property due to the incorrect use or putting into practice of any of the
methods, theories, products, instructions, ideas or recommendations included in
this DVD or which make reference to its content.
The computer application in this DVD is delivered with no guarantee of the
results obtained following incorrect use or adaptation for a specific purpose. The
user fully assumes all risks with regard to the results obtained by incorrect use of
the computer application.
None of the contents of this usage licence grants the user any intellectual or indus-
trial property rights over the confidential information of BioSystems S.A., and/or
of persons with rights over the content of this DVD.
The licence granted and constituted pursuant to these terms and conditions will
be interpreted in accordance with and governed by Spanish law, with the jurisdic-
tion being the courts of the city of Barcelona, Spain. The user waives any other
applicable legislation and/or jurisdiction that is competent, if any.
The user of this licence knows and agrees that the user licence grants no rights
over the use of computer programmes and/or applications belonging to others
that may be used or which may be necessary for using or operating this computer
application, for which the user will obtain the respective legitimisation of use.
17
User manual
1. Intended use
The BA200 analyser is used to determine analyte concentrations by in vitro bio-
chemical, turbidimetric and electrolyte measurements of human samples of serum,
urine, plasma, cerebrospinal fluid or total blood.
The BA200 analyser is optimised to work with the BioSystems biochemical, turbi-
dimetric and electrolyte reagents line. Reagents not included in the BA200 analyser
validation performed at BioSystems SA require a full and exhaustive validation by
the user or the laboratory.
We recommend a validation of the overall operation of the analyser and of the
reagents by the laboratory, taking into account the preanalytical phase and any
other relevant aspect.
The analyser is exclusively for professional use, i.e., for users who have the ap-
propriate training and expertise to use it. In addition to how to install the instru-
ment, users are instructed on the operation of the analyser and the software that
goes with it.
The environmental conditions for the functioning of the analyser are normal
clinical analysis laboratory conditions. These conditions are described in the
specifications chapter.
2. Contents
The elements that the user will find on unpacking the analyser are listed below.
Make a visual check to ensure that none of the elements has suffered any apparent
damage during transport.
1. Analyser.
2. Unpacking instructions sheet.
3. Analysis certificate sheet (Instrument Release Certificate).
4. Accessory box (supplied separately in a different box from the analyser).
Biosystems
18
Accessory Description
Labels for identifying the empty bottles.
Sample wells (1000)
Paediatric adapter (50)
Tube adapter (50)
19
User manual
Accessory Description
Mains connection cable, American plug
USB cable.
Fuses (2).
Biosystems
20
3. Identification of the main components
The different component parts of the analyser are marked and numbered in the
following figures and their associated lists:
3 5 6 4 7 2 1 8
Figure 1 Main components
21
User manual
4
6 5
Figure 2 Rear view components
4 5
Figure 3 Internal components
1– LED status lamp 5 – Cleaning solution bottle
2– ISE module
3– Reagent kit for ISE module
4– High contamination waste bottle
Biosystems
22
4. Installation
4.1. Location
Location Install the analyser in a large space. It occupies a minimum space of
107 cm x 75 cm.
Leave a space of at least 10 cm behind the analyser for the air to circulate from
the fan outlet.
Leave a space of least 70 cm above the analyser to allow the main cover to be opened.
Leave a space of least 40 cm at the front to allow the doors to be opened for ac-
cessing the ISE module.
Leave a space of at least 40 cm on the left side to allow room for accessing the
partial and general switches.
Environmental conditions Install the analyser in a dry, non-corrosive environment. The relative humid-
ity should not exceed 85%, with no condensation. It is advisable for the room
temperature to be less than 35 °C or 30 °C in the event of using the analyser
ISE module. Do not install the analyser in areas that are exposed to draughts.
Lighting Do not place the analyser below potent light sources. Keep the lighting as stable
as possible and ensure that no flashing light falls directly on the analyser. Direct
sunlight should also be avoided.
Electromagnetic radiation Make sure the analyser is not near any electromagnetic radiation sources (such
as motors, centrifuging appliances, mobiles telephones) or heat sources.
Anchoring Move the analyser to its definitive location by pushing it gently.
Once in the final position, anchor it. Unscrew the four adjustable legs (1) until
they touch the table. (See Figure 4).
Level the analyser by lengthening or shortening the legs, as necessary. Use a span-
ner to turn the nut (2) (See Figure 5).
When it is properly levelled, secure the nuts by turning the counter nut (3) to
the upper limit.
Do not turn the nut too much (3) to prevent the leg from being separated from
the structure.
23
User manual
1 1 2 1 1
Figure 4 Adjustable legs and front cover opening
3
2
Biosystems
24
4.3. Installing the waste containers and washing solution.
If you open the front door, you will see two bottles inside. The one on the left
with red cap is the high contamination waste bottle (1) and the one on the right
with green cap contains the washing solution (2). See Figure 6.
To handle the bottles in safety, leave them on the front door after opening it, to
enable the tubes to be disconnected and connected. The door is designed to sup-
port a weight of no more than 6 kg.
1 2 3
Figure 6 Location of the waste bottles and washing solution
25
User manual
Dispose of the waste in accordance with the applicable national or local govern-
ment legislation governing the disposal of dangerous biological waste.
Handle the high contamination waste container with care. Wear gloves and protec-
BIOHAZARD tive clothing when handling the container.
Biosystems
26
1
27
User manual
Connection Insert the red tube of the accessory box into the central connector of the analys-
er (2). See Figure 7. It is marked “LOW CONCENTRATION OUTLET”. Place
the other end of the tube directly in the drain if the legislation of the country
allows this. If not, install an external tank and connect the tube inside it.
Biosystems
28
1
Replacing the rotors To remove the rotor in order to easily install the sample tubes or reagent bottles,
proceed as follows:
1. Remove the rotor cover.
2. Press the central button on the handle to release the rotor.
3. Remove the rotor from its housing. Be careful, as if the rotor is full, it may
weigh up to 5 kg.
4. When reinserting the rotor into its housing, press the release button and let
NOTE
the rotor descend as far as it will go. Turn it until the positioning tab coincides
at the base and is correctly seated.
5. When inserting a full rotor into the housing, ensure it descends slowly without
falling, to prevent it from knocking against the base and the reagent bottles
from splashing.
6. Place the cover of the rotor on its housing. Ensure that it is properly seated
in the housing, it has only one position.
NOTE
29
User manual
Affix each label to the reagent bottles supplied or to the tubes in accordance with
the above chart. When the barcode reader scans the reagent rotor and detects an
auxiliary bottle, the programme will ask you to associate that bottle to a reagent
in the list.
Biosystems
30
User manual
Biosystems
31
Model continuous UPS (on-line)
Power 1.5 KW
Battery capacity Over 15 min
32
2. Connect the mains cable firstly to the appliance and secondly to the mains.
3. Put the general switch (1) to the connected position (I).
4. There are three separate switches, one for the analyser, another for the refrig-
erator and a third for the ISE module.
5. To turn on the analyser, put the switch (4) to the connected position (I).
6. To turn on the refrigerator open the protective cover and put the switch (3)
to the connected position (I).
7. To turn on the ISE module open the protective cover and put the switch (2)
to the connected position (I).
WARNING
33
User manual
Ensure that no Microsoft SQL server version has been previously installed in the
computer. To verify this, open the following programme in Home:
Control panel\All Control panel elements\Programmes and characteristics
and check there is no input with the name: Microsoft SQL server
Install the program by proceeding as follows:
1. Insert the disk into the DVD drive of the computer.
2. Press Start, select Execution and write:
3. D:\setup\setup.exe, or the name of the DVD unit
4. Follow the steps indicated by the installer programme.
5. The installer program automatically installs the application programme, the da-
tabase manager and USB controller driver without the user having to intervene.
6. The database manager installation process may take some time; wait until the
installation process has ended.
7. Configure the display of operating system with the following characteristics:
• Select All settings
• Select System
• Select Display
• Select Resolution 1024 x 768
• Select Scale and distribution: 100%
))See Figure 13
8. Disable the screen saver
• Select All settings
• Select System
• Select Power&sleep
• Select Never, in all the options
))See Figure 14
Biosystems
34
Figure 13 Screen of text configuration
35
User manual
Biosystems
36
Figure 17 Change the USB power options
37
User manual
39
User manual
Biosystems
40
4.12.3.4. Run the antivirus before installing the application
Before installing the BA400 application run the antivirus program to make sure
that the system is clean of viruses.
Follow the following steps:
• Select All settings
• Select Update and security
• Select Windows defender
• Select Open windows defender security center
• Select Advanced Exam
))See Figure 22
• Select Complete Exam
41
User manual
Biosystems
42
Figure 24 Restart options
Windows and antivirus updates can cause a lack of compatibility that affects the
operation of the BA400 software: User Application, Service application or Com-
munications Libraries with LIS.
To quickly recover the operativeness, we recommend that you program the au-
tomatic and periodic realization of points of restoration of the operating system.
If you detect a compatibility problem after an update, immediately report the
problem to your Technical Service, attaching a SATReport and the date on which
you observed the problem. While Biosystems analyzes the problem and designs a
solution to guarantee compatibility, restore the operating system to the last date
with stable operation, this will allow you to continue working.
Follow the steps to configure the restore points:
• Select All settings
• In the search bar type: Create a restore point
• A window opens to configure the restore points
• Select OS C Unit: (system) and press Configure
))See Figure 25
• Enable system protection and configure the maximum disk space that will be
used for system protection. Select how at least 10 Gbytes
43
User manual
Biosystems
44
4.12.3.8. Backup of the data
Each time a session is reset, the database is backed up.
You can locate this backup in the following folder:
c: \ Users \ [User name] \ documents \ BA400 \ RestorePoints \
Where [User name] is the username to enter the system.
The last 10 backup copies are saved. The oldest copies are erased.
This file is password protected. It can only be opened with the application. Through
the sat-report reading option.
Configure a method to make copies of these files out of the system. So that in case
of system failure you can recover the data without problems.
45
User manual
Biosystems
46
4.12.3.11. Use a VPN system
For those systems that require the privacy of the data during communications
with the host, use the VPN (Virtual Private Network) in the windows system to
perform encrypted communications with the host.
Follow the following steps:
• Select All settings
• Select Network and internet
• Select VPN
• Select Add a VPN connection and configure the configuration options according
to the data of the system administrator or the VPN provider.
Electrode installation After opening the front cover the ISE module can be directly accessed. See Fig-
ure 23.
1. Turn off the ISE module power supply.
2. Unscrew both screws by hand (1) and remove the cover (2) to access the
compartment for positioning the electrodes.
3. Unpack each electrode. Ensure that the sealing ring (O-ring) is in position.
Carefully dry any traces of liquid.
4. First put the reference electrode in position. Pull out the identification wire
with a label that is connected inside the electrode circulation channel. Ensure
47
User manual
there are no traces of salt in the channel. Keep the wire with the label in case
you need to uninstall the electrode.
5. To insert the reference electrode press the yellow tab downward (3) and insert
it as far as it will go. Then release the tab.
6. Insert the other electrodes in the positions shown in Figure 23. Check that
their sealing rings (O-rings) are correctly positioned. Carefully dry any traces
of liquid.
7. Each electrode has a single position to prevent errors in putting them in place.
8. In the event of not having a Li+ electrode, insert an empty electrode in its place
(it is marked by a line of dots), to close the channel pathway.
9. Release the yellow button to supply pressure to all the electrodes and ensure
good fluid communication.
10. To ensure that the electrodes are properly placed, press them at the front until
you hear a click or they have been correctly seated.
Tube installation Insert a tube into each peristaltic pump. To insert the tube into the head of the
peristaltic pump release the pressure on the head by pulling to the side of the
clamp, see Figure 24.
Each tube has two labels. The labels help guide the tube correctly in the peristaltic
pump. The number on the label of each tube must coincide with the number on
the pump label.
• The tubes marked W must be installed in the pump (6). The order for putting
them in place starting from the bottom is W1 and W2.
• The tubes marked B must be installed in the pump (5). The order for putting
them in place starting from the bottom is B2 and B1.
• The tubes marked A must be installed in the pump (4). The order for putting
them in place starting from the bottom is A2 and A1.
Biosystems
48
Take care when connecting the tubes of the waste pump (6) as they are connected
in reverse order to the tubes of the pump for calibrators A (4) and B (5).
Installing the reagent kit Unpack the kit, remove the three red protective caps from the connections and
the red warning label. Keep the caps in case you need to uninstall the reagent
kit. Position the connector in the correct direction and press lightly until you
hear a “click”. Write the installation date on the side of the kit.
Do not press the sides of the box too strongly or put the reagent kit face downwards
without the caps, as the reagent could be spilled. It is advisable to wear gloves when
performing this operation.
NOTE
Place the kit in its housing.
With the user programme, execute the actions in the number and order indicated
in the Installation/Activation optionof the ISE functions section.
))See chapter 10.9.3
49
User manual
Biosystems
50
10. Select the Communications configuration tab in the General submenu of the
Configuration menu.
11. Select the automatic option.
12. In the same menu select the Analyser tab.
13. Selection one of the two water inlet selection options, depending on the water
inlet installation.
))See chapter 10.2.1
14. Press the initialisation button on the analyser.
15. Perform 5 conditionings to ensure that the internal water tank fills up and that
the fluidic system is correctly primed.
))See chapter 10.9.2
16. In the event of having an ISE module, install the electrodes and reagent kit.
))See chapter 10.9.3
17. Fill in the calibrator concentration fields and controls of the tests to be used.
))See chapters 10.3.5 and 10.3.6
18. Make a list of blanks, calibrators and controls.
))See chapter 10.5.3 and 10.5.4
ing the Pause button. In the event that the main cover opens unexpect-
NOTE edly, the analyser will stop any action it is performing, and the preparations
already made in which the sample has not yet been dispensed will be lost.
• Ensure that the sample, reagent and reaction rotor cover are on while the
appliance is operating. The analyser will not perform any operation if any of
these cover is missing.
• Keep the analyser work surface free from obstacles that could collide with the
preparation or stirring arms.
• Make sure the barcode labels on the sample tubes are correctly affixed and
properly centred. They must be aligned properly on the tube. If the label has
a barcode with only a few digits affix it lengthwise and centered, without plac-
ing it on the top part of the tube. Position the sample tube with the barcode
label facing outwards.
51
User manual
• Do not duplicate any sample tube identification code on the barcode labels
during the same session. If several sample tubes have the same barcode identi-
fier while the equipment is enabled for working with LIS communications,
NOTE the analyser will not automatically assign any test to those tubes and it will
display a message indicating this on the screen. In the case of manual opera-
tion (without LIS communications), first the analyser will pipette the tube
that is in the lowest position in the sample rotor.
• The positions of the sample tubes and reagent bottles must be maintained when
the Automatic Barcode Verification is not used before starting / continuing
the work session.
NOTE
• When using sample tubes of 15 mm in Ø or with a capacity of 15 mL, do not
fill them to the brim with the sample. Fill them with no more than 14 mL in
order to correctly detect the sample.
• Use the positive identification of sample tubes whenever it is possible. That
is, use the bar code to identify sample tubes. This will prevent possible errors
in the placement of tubes in the rotor of samples.
NOTE
Biosystems
52
To obtain precise results, the samples must be free from clots, fibrin, etc., which
could block the sample tip or ISE module reader channel. Also check that the
serum has no air bubbles.
NOTE If using a tube with serum separator gel, check it has sufficient serum volume in
order to avoid inserting the sample tip into the gel layer. This could block the tip.
Never insert samples that are extremely lipaemic and/or haemolysed.
Primary plasma tubes Plasma tubes can be used only for those tests already validated (see IFU). Pro-
ceed with the sample preanalytical phase of the plasma tubes as follows:
1. Obtain the sample by venous puncture in a blood collection tube with an
anticoagulant. The anticoagulant must be compatible with the determinations
that have to be made.
If this sample is to be used for measuring ISE determinations, heparin sodium
must be used as an anticoagulant. The heparin level should not exceed 15
UI/ml of the tube volume. Do not use heparin ammonium, heparin lithium,
EDTA or NaF tubes.
NOTE
2. Mix the sample by inverting the tube several times. Do not shake it.
3. Centrifuge the sample for 10-15 min within one hour of collecting or follow
the manufacturer’s instructions. Carefully remove the plasma layer at the top
for analysis. Use a Pasteur pipette or a syringe fitted with a blunt-tipped needle
for this procedure. Make sure the cell phase is properly separately from the
aqueous phase.
Dilution of urine for ISE If wanting to make ISE determinations in urine, the urine must be diluted. Per-
form the dilution manually outside the analyser with a dilution factor of 1/10
and a diluent code of 5412 or 5408.
• The analyser uses 200 µL to make an ISE determination in urine. Prepare a
larger quantity of diluted urine.
• Take one part of the urine and pipette it into a primary tube or paediatric vial.
• Take nine parts of the urine diluent (it is in the ISE module accessory box)
and pour it into the same primary urine tube.
• Mix it and place it in the sample rotor.
ISE washing solution Every day in which ISE determinations are made you should wash the module
to remove proteins from the fluidic canal. It is advisable to perform this washing
operation at the end of the day.
The ISE module accessory box contains the ISE module washing solution kit. It
has 6 bottles with washing powder (peptin) and a diluent.
• Add the diluent until the peptin bottle (12 mL) is full, shake it well and take
note of the preparation date
• When it is not in use, keep it in the refrigerator.
• Discard it 4 weeks after preparing it.
Washing solutions The washing solutions supplied with the equipment are of two types, the wash-
ing solution (WS) used for internal washing of the tips and reaction rotor and
the acid washing solution (WS1) used to prevent contamination.
53
User manual
The washing solution (WS) is prepared in the 2.4 L bottle located on the left side
of the front of the equipment.
The acid washing solution (AC17201)(WS1) is installed in the reagent rotor. It
is supplied in 20 mL bottles.
Handle these solutions with care and wear the appropriate clothing and gloves.
Never mix both washing solutions as they could give off dangerous gases.
Biosystems
54
7. Operating principle
The analyser has various operating states: Warm-up, standby, running and pause.
Sleeping state During this state the analyser is stopped and all its electronic systems are
dormant, to save energy. The analyser will not perform any operation until the
warm-up state is activated.
Warm up state During this state the analyser initialisation process is started. The initialisation
process consists of the conditioning of the fluidic system and thermostating
the reaction rotor so that it reaches the correct temperature and ends with the
rotor baseline execution. The baseline consists of filling all the reaction rotor
wells with distilled water, measuring the absorbance in each well with all the
wavelengths and checking that the measurements are within the ranges. If the
measurements are correct the rotor is emptied and dried. These baseline read-
ings serve to correct the preparation reading absorbances. The whole baseline
reading process takes about 20 minutes.
Stand-by state In that state the analyser waits for the operating state to start. While in this
state, the user can perform maintenance tasks and/or execute the analyser
functions.
Running state During this state the analyser performs several repetitive cycles to prepare the
reactions and take the measurements. The dispensing arm performs two differ-
ent movements, firstly suctioning of R1 and sample and secondly, suctioning
and dosing of R2. A reagent is prepared in the following way:
1. Suctioning of reagent 1 and sample, dispensation in the reaction rotor.
2. Stirring of the reagent 1 and sample mixture.
3. Start the reading period.
4. Suctioning of reagent 2 and dispensation into the reaction cuvette 5 min.
after dispensing R1.
5. Stirring of the mixture with the second reagent.
6. Finalise the reading processes.
7. Washing of the cuvettes.
The reading process is based on the optical absorption spectrophotometer prin-
ciple. The concentration is determined by comparing the luminous intensity of
a certain wavelength that passes through the cuvette when there is a reaction and
when there is no reaction. In some cases the concentration is a direct function of
the absorbance and in others, it is a function of the change of the absorbance over
time, depending on the analysis mode.
Pause state During this state the analyser stops the sample and reagent dispensing process,
allowing the user to access the sample and reagent rotors and to add new sam-
ples or replenish reagents. During this state, the analyser continues to perform
the reaction rotor reading process.
55
User manual
2
3
Figure 31 Covers
Biosystems
56
1 – Main cover 5 – Washing Solution and HCW Bottles
2 – Reaction rotor cover and ISE module front door access
3 – Sample and reagent rotor cover
4 – LED status lamp
Main cover This covers the surface of the analyser. Open this cover to access the reagent,
sample or reaction rotors. To ensure the safe operation of the analyser, this cover
must be kept closed. It has an open or closed casing detector. The analyser will
stop executing the worklist if the cover is opened while it is operating.
Sample and reagent This gives access to the sample and reagent rotor. The patient samples, calibrators,
rotor cover controls and reagents are placed in this rotor. The rotor unit is refrigerated. The
cover has a detector that enables the program to detect the presence of the cover.
Reaction rotor cover This allows you to access the reaction rotor. This rotor is where the reactions are
made and the photometric readings are taken. This rotor is thermostatted at
37 SDgrC. The cover has a detector to allow the programme to detect the presence
of the cover.
Front door It provides access to the washing solution and high contamination waste bottles,
and also to the ISE module (optional unit).
Led status lamp This lamp indicates the analyser state. Possible states:
Beeper states The analyser has a beeper that warns the user in the event that an alarm is
triggered.
When the analyser is first switched on (supplied with power), it will carry out a
series of internal checks. Once these checks have been completed the instrument
will generate a short beep, indicating that it is ready to establish connection with
the User Software/Service.
While in the list execution state, if an alarm is triggered, for instance, exhaustion of
reagent, samples, etc., the analyser will indicate this through a beeper that sounds
until the user turns it off manually.
57
User manual
Positions There are 88 positions in all, laid out in two rings. The bottle and sample bar-
codes can be read in both rings.
Sample tubes To position the tubes, an adapter must first be inserted (see chapter 4.6 explain-
ing how to position the adapter) Primary tube dimensions:
• Minimum diameter: ∅12 mm
• Maximum diameter: ∅16 mm
• Minimum height: 70 mm
• Maximum height: 100 mm
Sample wells To insert the sample wells in their positions, an accessory is supplied with the
analyser for adapting the diameters. See chapter 4.6 explaining how to position
the adapter.
Bottles 2 types of bottle can be inserted. The volumes of the bottles are as follows:
• 60 mL, only for positioning in the inner ring.
• 20 mL, for positioning in the inner and the outer ring.
Refrigeration The refrigeration system power supply is separated from that of the analyser,
which means the analyser can be switched off and the refrigeration system can
be left on.
Biosystems
58
8.3. Reaction rotor
The reaction rotor consists of a thermostatted channel containing an optical-quality
plastic rotor that permits the transmission of UV light.
Positions There are 120 positions. The reagent and sample are dispensed into each cu-
vette. While the mixture is reacting the optical reading is taken, to obtain the
absorbance.
Volume The reaction volume is between 180 µL and 440 µL.
Temperature The rotor is maintained at a stable temperature of 37 °C by a Peltier-based ther-
mostating system.
Dispensing cycles for each of the arms:
• Cycle 1: Dispensing Reagent 1 and sample
• Cycle 2: Stirring Reagent 1 and sample
• Cycle 3: Initiation of photometric readings
• Cycle 18: Dispensing of Reagent 2
• Cycle 19: Stirring of Reagent 2
• Cycle 35: End of the reading processes
• Cycles 36 - 46: Washing of cuvettes in the wash station
59
User manual
The optical system is physically located in the reaction rotor, below the wash station.
Wavelengths 340 nm, 405 nm, 505 nm, 535 nm, 560 nm, 600 nm, 635 nm and 670 nm
Measuring range From -0.2 A to 3.5 A
Resolution 0.0001 A
The system automatically performs a cuvette blank of the whole rotor when the
analyser is initialised. This cuvette blank absorbance serves to correct the reaction
absorbance measurements. When the well enters the wash station, it is read again
to check that it is in optimal conditions. If this value exceeds a pre-established
limit, the well is rejected.
Biosystems
60
Figure 34 Stirring arm
61
User manual
Biosystems
62
8.9. ISE module (optional)
The ISE reader module is optional and serves to determine the concentration of
the Na+, K+, Cl- and Li+ ions in serum, plasma and urine samples.
63
User manual
In addition, for each determination the module measures one of the two liquids
from the reagent kit: A for determinations in serum and plasma, and B for deter-
minations in urine.
Both liquid A and liquid B are supplied with the reagent kit. That kit is connected
directly to the ISE module.
The kit is supplied as an accessory and its housing is accessible through the front
door of the analyser.
Biosystems
64
9. Description of the software
65
User manual
Biosystems
66
Icon Description of the icon
Access to the quality control screen.
67
User manual
Biosystems
68
10. Working procedure
The first time the programme is started, the login and password to be entered are:
Parameter Value
User name Admin
Password BA200
Table 6 Initial login and password
Click on the icon to change the password. You can only change the user pass-
word that was entered in the home screen. Once you entered for the first time,
the program ask to you to change the password.
Figure 39 shows the screen for changing the password. Enter the different values
required to change the password.
69
User manual
10.2. Configuration
In this menu you can access the different setup options:
• General: General programme configuration.
• Languages: Selection of the programme language.
• Reports: Setup of the report headers and page footers.
• Order Printing of Tests: Selection of the test order for the patient reports.
• Barcode: Barcode setup.
• LIS: LIS communication system setup.
• Mapping for LIS
• Users: Generation of users for accessing the programme.
• Change User: Change in user.
Biosystems
70
Figure 40 Work session setup
Default sample tube type Select the tube that will normally be shown when creating the list of patients. It
may be: tube or sample well.
Automatic checking barcodes Check this option is you want the analyser to automatically check the position
of the reagent bottles and sample tubes with the barcode before starting the
session.
Download only patient tubes Check this option if you only want to delete the sample rotor tubes when reset-
from rotor when resetting ting the session. The information and position of the sample wells (calibrators
worksession and controls) will be saved for the next session.
Automatic processing of Check this option if you want the repetitions to be made automatically. If not,
repetitions they can must be made manually.
Automatic printing of patient Check this option to automatically print out the results for a finalised pa-
results tient. When this option is selected the report and frequency type options are
activated.
Type Select the type of report to be used for printing out the patient results.
• Compact - Report with no patient header and with the results of all the patients
in a continuous list with no page separations.
• Individual - Individual report by patient. Each report is printed out on separate
pages with a patient header.
Frequency Select the frequency to be used for printing the results.
• On restarting the work session
• After completing each session
• After completing each patient
71
User manual
Water inlet selection Select the water inlet mode for the analyser.
The water may enter through different channels which are mutually exclusive:
• Water tank
• Water mains
))See chapter 4.4 for the purified water installation.
Alarm sound disabled Check this option if you do not want the beeper sound whenever an alarm is
indicated.
Automatic Select this option and the programme will automatically search for the compu-
ter output port for communicating with the analyser.
Manual Select this option for the port to be selected manually.
Port:
• Select the available COM port, the USB uses a virtual COM port.
Biosystems
72
If no value appears in the port pop-up menu, wait a few minutes and try the
connection again. The USB driver registration process may take a few minutes.
NOTE
Protected mode Select this mode when you do not want the supervisor and operator user to be
able to modify and display the main parameters of the test. This option will not
be visible to the supervisor or operator users.
10.2.2. Language
This allows you to select the application language.
10.2.3. Reports
This permits the configuration of the patient report format. It allows you to change
the header, footer and add logos.
73
User manual
Biosystems
74
Figure 46 Screen for arranging the tests
Select a test or test group and press one of the buttons until the test is placed in
the desired position. Normal arrows orders the selected test by one step posi-
tion, while arrows aiming to an horizontal bar orders it at the beginning (aim-
ing up)/end (aiming down) of the list
Press this button to display the tests in alphabetical order again.
10.2.5. Barcode
Screen for configuring the barcode reader options. Figure 40 shows the screen with
different options that allows the configuration can be made.
Disable Barcode Select this option to disable the barcode reader. Disabling the barcode reader is
incompatible with either LIS connectivity or sample ID by labelling.
Sample barcode type Select the barcode type for performing the sample barcode readings. More than
one barcode type can be selected. The barcode printed on the labels of the pri-
mary tubes must coincide with the code selected in the configuration.
Barcode fields activation If this field is not activated the barcode reader identifies the whole barcode as
the sample identifier and the reader can read any code with length of between
1 and 30 characters. Codes with different lengths can be combined in the same
session.
75
User manual
When this field is activated, information on the sample type is provided in addi-
tion to the sample identifier. The total size of the barcode is still flexible, between
1 and 30. The following fields are enabled too.
Start and end Select the start and end positions of the sample identifier in the barcode. The
sample type will always be shorter in length than the total barcode length.
Sample type Specific characters which define sample type within the barcode. The number of
characters must match with total length for sample type identifier. Every code
identifies one sample type.
Biosystems
76
and the packet size of each query message. It is configured in the option Host
Query Packages size.
Upload patients results When this option is activated the patient results created manually through the
requested from the analyzer analyser are sent.
Upload QC results When this option is activated the results of controls requested manually
requested from analyzer through the analyser are sent.
Upload results during When this option is activated all the results of the session following a reset are
Reset worksession sent. All the results requested by the LIS will be sent and when the above pa-
rameters are active, the results requested manually from the BA200 will also be
sent.
Automatic results upload When this option is activated you can select the frequency with which the
results are automatically sent to the LIS.
77
User manual
LIS communications enabled This allows communication with a LIS application to be activated or
deactivated.
Data transmission type The transmission may be:
• ASTM: TCPIP-Client, TCPIP-Server
• HL7: TCPIP-Client, TCPIP-Server, TCPIP-transitory connection
Host name This field is only completed when the data transmission type option has been
selected: TCPIP-Client. Enter the IP of the computer IP when the LIS for mak-
ing the connection is executed.
TCP port Number of the TCP-IP port through which the LIS connection is made.
When the TCPIP-Transitory HL7 Connection is selected, 2 different ports must
be configured: client port and server port.
Client TCP port Number of client port in a TCP connection.
Server TCP port Number of server port in a TCP connection.
10.2.6.3. Protocol
Screen for configuring the necessary parameters for the low level LIS communica-
tion protocols.
Protocol name Select the type of protocol to be used in the communications: HL7 or ASTM.
Transmissions code page Select the type of coding for messages to be sent between the analyser and the
LIS. This is used in transmitting and receiving messages. You should configure
the page code used by your LIS system.
Host ID Identifier used by the LIS application.
Host provider Name of the LIS application supplier.
Instrument ID Name used to identify the instrument; that field is transited in each message.
Biosystems
78
Analyzer provider Name of the instrument supplier.
IHE compliant Select this option when the message transmission strictly follows the IHE com-
munication standard.
Host Query packages size Number of specimens sent in one Query message by specimen when using the
ASTM protocol.
Max time to retry sending Configuration of the maximum time during which another attempt is made to
messages to LIS send a message to LIS when no response is received.
Max time to consider LIS Configuration of the maximum wait time for receiving an acceptance or con-
offline firmation message from the LIS. After this time the LIS state is modified (Led
lamp on red) indicating that there are problems in communication which must
be solved.
Delimiters Enter the delimiters to be used in transmitting and receiving messages.
79
User manual
• The second column shows the name of the element as it appears in the analyser.
• The third column shows the name of the element used in communications
with the LIS (messages received and sent). These names must be edited in
order to adapt them to each LIS. When installed, the same names as those
used in the analyser will appear.
LIS mapping types Selection box for filtering the elements shown by each of the types.
You are able to edit any LIS Name by pressing the edition button, and saving it.
10.2.8. Users
This allows you to create, edit and delete the names of the users accessing the
application.
There are three user levels. Administrator, supervisor and operator.
Level Description
Administrator Has complete access to the application. This user is allowed
to create the supervisor and/or operator user. There is only
one administrator user.
From this level of permissions you access the configuration
screen of the test visualization. The supervisor and operator
users do not have access to this screen.
Biosystems
80
Level Description
Supervisor Has limited access. This user is allowed to create users with
operator permits. He is permitted to change the calibrators
and control values and create a limited number of tests
established by the Administrator. Only one supervisor
can be created. Supervisor is able to create up to 5 new
additional tests.
Operator This is the most restrictive access level. This user can only
execute lists, view and print out results and consult the test
parameters. As many users as required can be created.
Table 7 User levels
81
User manual
10.3. Programming
In this menu you can access the different options for programming the necessary
parameters to measure with the analyser. The different programming options are:
test, calculated test, contaminations, profiles, calibrators, controls, patient data,
ISE tests and external tests.
10.3.1. Tests
Check the test parameter programme is according to the instructions for use
(IFU) for each reagent.
CAUTION This option in the programme allows you to create, change, delete and list the
tests and their parameters.
The screen is divided into two parts. The left-hand part shows a list of all the
tests and the right-hand part shows the different parameters and their values. The
parameters are grouped by different sections: general, procedure, calibration and
blank, quality control and options.
Press on the name of the section to access each group of parameters.
Press the icon to create a new test. To indicate that the fields can be edited, the
background colour will change to white. Some parameters will appear with
default values.
Biosystems
82
To edit the parameters of a test already entered, first select the name of the test
to be changed from the test list and then press the icon. You can also edit the
test parameters by double-clicking on the name of the test in the list of tests.
Select the test name and press the icon. The programme will ask for confirma-
tion before deleting it. Only the tests created by the user can be deleted (the test
icon is in yellow). The original tests (with the blue icon) cannot be deleted.
Press the icon to open up an auxiliary screen which allows you to clear and
order tests. The ordering and/or clearing of tests is applied in the test selection
screen.
))See chapter 10.3.1.1
Press this icon to make a copy of a test. The name of the new test must be
changed.
You can delete or print out several tests at once, and make multiple selections
from the list of tests.
Non-consecutive multiple selection Select a test and keep the CONTROL key pressed while you select the other
tests.
Consecutive multiple selection To make a consecutive multiple selection of several tests, select the initial test,
press the UPPER CASE key and selec the last test. All the tests between the ini-
tial and last test will be selected.
Ordering of tests Press the header on the test list to show the tests in ascending order. If you press
a second time, they will be shown in descending order.
This icon will appear when a compulsory parameter needs to be entered, or if
there is an error in entering the value.
83
User manual
Select a test or test group and press one of the buttons until the test is placed in
the desired position.
Press this button to display the tests in alphabetical order again.
Check the box whenever a test should be available in the Selection Test screen.
Uncheck it if you are not going to work with.
Analysis modes
Endpoint monoreagent
Endpoint bi-reagent
Differential be-reagent
Fixed time monoreagent
Fixed time bi-reagent
Kinetic monoreagent
Kinetic bireagent
Biosystems
84
))See how to make the absorbance calculations depending on the analysis mode in chapter 16.
Unit Select from the list the unit that will use the test. This value will be displayed
along with the concentration results.
Decimals Number of decimal points for displaying the concentration values.
Number of replicates Number of replicates performed by the analyser for each sample.
Type of reaction Select the type of reaction: increasing or decreasing.
Report name Name of the test which will appear in the patient report. If there is no name
entered in this box, the test name will be shown in the patient report.
Reading mode Select one of the two following options: monochromatic or bichromatic. The
absorbance calculation depends on the reading mode.
))See how to perform the absorbance calculations in chapter 16.
Main filter Select the main filter value to be used for taking the readings.
Reference filter Select the reference filter value. This box will only be activated if the bichro-
matic reading mode has been selected.
Sample volume Enter the sample volume for making the preparation. The sample volume range
is from 2 mL to 40 mL. The volume can be entered in decimal fractions in mL.
Volume of reagent 1 Enter the reagent 1 volume for making the preparation. The volume range is
from 90 mL to 300 mL.
Volume of reagent 2 Enter the reagent 2 volume for making the preparation. The volume range is
from 10 mL to 100 mL. This box will only be activated if the bi-reagent option
is selected in the analysis mode.
85
User manual
Reading time 1 Enter the reading time for calculating the absorbance. This can be entered as
seconds or cycles. The time range is from cycle 2 to 35 when the reaction is in
monoreagent mode or from cycle 2 to 18 when the reaction is in bireagent.
Reading time 2 Enter the time for making the last reading. This box will be activated for kinet-
ic, fixed-time or bireagent differential calculation methods. The time ranges are
from cycle 3 to 35. Reading time 2 must always be greater than reading time 1.
Diluent Select the diluent for preparing the dilution. It may be distilled water, saline
solution or a generic diluent. This bottle will be used for pre-dilutions and for
performing repetitions and it must be installed on the reagent rotor.
Pre-dilution factor Activate this option if the sample requires a pre-dilution. The pre-dilution can
be made automatically by the analyser or manually, which it requires the user’s
intervention. Its final result will be calculated automatically taking into account
the applied dilution ratio. This function affects all the samples. The parameters
required are:
Automatic repetition Activate this option if you want automatic repetitions to be made whenever a
result is out of any established limit in Options screen tab. This function affects
only individual test and it is in dependency of the ‘Automatic processing of
repetitions’ function. See General Setup configuration 10.2.1
Post-dilution factor Enter the reduced and increased factor. When performing automatic repeti-
tions, make the preparations using these post-dilution ratios. Manual repetitions
show these values by default, but they also allow the user to enter a different
dilution factor before sending them.
Biosystems
86
Repetition factor Description
Increased factor This will be used when making a preparation
with a higher sample proportion (to
solve problems of samples with very low
concentrations that are below the detection
limit)
Enter the factor for the repetition
concentration to be increased and exceed
the detection limit. The analyser changes
the sample/reagent volume ratio with
the programmed factor. The result of the
preparation will automatically be divided by
the factor.
87
User manual
Blank replicates Number of replicates for making the blank. The range is from 1 to 3. The mean
of the replicates is used to calculate the concentration.
Factor If the test is not calibrated enter the multiplicative factor value for calculating
the concentration.
Experimental calibrator Enter the calibrator data and its concentration. Press the icon again to open up
the calibration screen and enter the calibrator parameters.
))See how to enter the calibrator parameters in chapter 10.3.5
Calibrator replicates Number of replicates for making the calibrator. The range is from 1 to 3. The
mean of the replicates is used to calculate the factor.
Use alternative calibrator Calibration reference it will be used for another sample type (i.e. you can use
calibration data obtained from SER type to be applied on URI type)
Calibration and curves values Shows the calibrator values assigned to the test. They are only shown for in-
formative purposes. To create new calibrators and/or change them, edit them in
the calibrator screen.
Activate quality control Check this option to activate the quality control for this test. If this option has
not checked, the selection of the test in the Controls Programming screen will
not available.
Control replicates Number of replicates for measuring the controls. The range is between 1 and 3.
Rejection criteria It establishes the limit considered as an alarm. This value is calculated in stand-
ard deviations (SD). The margin is between 0.1 and 4.1.
Biosystems
88
Calculation mode The calculation mode may be manual or statistical. Indicate how to calculate the
ranges in order to draw the Levy-Jennings graph and activate the Westgard rules
alarms.
Calculation Description
mode
Manual Use the theoretical ranges of the serum control seating values
entered when registering a control. They remain unaltered,
unless new cumulative values are to be assigned. Lot Average
and SD values have been obtained experimentally from
different instruments.
))
See chapter 10.8.3
Statistical Use the ranges calculated from the mean and SD of the above
series.
The minimum number of series indicates how many values
will be used to calculate the expected average and SD. During
these first series, the manual mode is used internally. The
minimum number of series to be programmed is 5. After the
internal series, the obtained average and SD results will be
compared against the expected.
Different quality regulations in the laboratory make it
advisable to assign 20 minimum series when starting to use a
specific control lot.
Rules to apply Select which Westgard rules you want to apply to the quality controls for this
test.
It serves to register the controls with their batch and concentration values.
))See how to register a control in chapter 10.3.6.
Control selection This table shows the different controls registered for the test. In the box, activate
the controls to be used, as you may have created various controls. It can activate
up to 3 controls at one time.
89
User manual
Blank absorbance limit Limit value established for comparing against the result of the blank absorb-
ance. It is used for checking the state of the reagent.
Kinetic blank limit Enter the correct kinetic blank value limit. It is only applied to tests with the
kinetic calculation mode.
Linearity limit Enter the value for which the reagent is no longer linear. If the concentration
value is higher than this value, the programme will issue a warning message and
if it is activated in the programme a repetition will automatically be launched.
Only linear calibrations have pre programmed values for this. In case of multi
point calibrations, which they may depend on lot experimental values, this cor-
responds to the highest concentration value.
Detection limit Enter the value for which the reagent does not detect a value. If the concentra-
tion value is lower than this value, the programme will issue a warning mes-
sage and if it is activated in the programme a repetition will automatically be
launched.
Factor limits Upper and lower range for verifying that the calibration factor is correct.
Repetition range The analyser repeats the sample automatically if both the concentration value is
within the range and the automatic repetition is enabled. This range serves to
confirm the result automatically.
Prozone effect The prozone effect may occur in tests based on the principle of the formation
of antigen-antibody complexes (agglutination). This effect is usually detected in
samples with a high antigen content. Excess antigen inverts the reaction direc-
tion and may cause incorrect sample measurements. To detect that effect you
should activate the prozone effect option and enter the 3 parameters: Time 1,
Time 2 and ratio in (%).
The program will calculate the absorbance increases in times 1 and 2, obtain the
increase quotient and compare the result against the ratio. If the quotient exceed
Biosystems
90
the ratio, an alarm is triggered indicating that the sample could have the prozone
effect, in which case the user must make a manual repetition with a dilution factor,
to finally determine the exact value of the sample.
The prozone effect verification parameters must be determined by the end user.
In the case of doubt in detecting the prozone, the user must make a sample
dilution, to ensure the result is correct.
CAUTION
Slope function Enter parameters a and b of the formula Y=aX+b. These parameters change
the value of the result concentration in a linear manner. This option serves
to proportionally correct the concentration of a sample for the same absorb-
ance. Where X will be replaced by the concentration value and Y will be the
changed concentration value. a and b are the slope and the intercept with Y axis
respectively.
Substrate depletion Enter the value in absorbances. When a test in kinetic analyse mode has a point
below this limit, this means it has consumed the reagent substrate, and the re-
sult is thus not correct. When this alarm is activated the programme automati-
cally launches a repetition.
Reference ranges Indicate the normal reference values for the population. If values are entered in
the fields, the results are shown on the screen along with the patient report and
the concentration result.
Panic ranges Enter the values for which a result is pathological. The values entered must meet
the following conditions:
Minimum panic < Minimum normality < Maximum normality < Maximum panic
91
User manual
On the left-hand side of the screen is a list of calculated tests and on the right-hand
side are the parameters to be entered for each calculated test.
))See how to create, edit, delete and print icons in chapter 10.3.1.
Name Name of the calculated test.
Short name Abbreviated name of the calculated test.
Sample type Indicates the type of sample to be used for selecting the standard tests.
Unit Unit in which the results of the calculated tests will be shown. This unit may be
different from that of the standard tests.
Decimals Number of decimal points for displaying the calculated test concentration val-
ues. This number of decimal places may be different from that of the standard
tests.
Print experimental tests Check this option if the patient report also has to show the results of the stand-
ard tests in addition to the result of the calculated test.
Formula definition Formula relating the calculated test to the standard tests. To enter the formula,
select the standard tests, other calculated tests, numbers and operators. The
programme verifies whether the formula entered is correct and indicates this
through one of the following icons:
Biosystems
92
This icon indicates that the formula has been correctly entered with no errors.
This icon indicates that there are errors in the formula. Change the formula
until the icon disappears.
Delete the last character entered.
Delete the whole formula entered.
A reference range screen tab is also available to enter the expected values in Generic
or Detailed manner.
))See how to create, edit, delete and print icons in chapter 10.3.1.
10.3.3. Contaminations
When contamination exists between programmed reagents, analyzer throughput
drops.
CAUTION This screen is used to program contaminations between reagents and cuvette
contaminations.
To eliminate the contamination, first of all the programme orders the tests for a
patient to avoid dispensing them in consecutive order. If it is not possible to elimi-
nate the contamination by ordering them, an extra wash cycle is added between the
contaminant test and the contaminated test, to clean the tip. If nothing is indicated
in the programme, the wash cycle is performed with purified water, otherwise the
cycle will be executed with the programmed washing solution.
It exists up to 3 different washing solutions. washing wolution 1 corresponds to the
acid washing solution (AC17201), washing solution 2 corresponds to the alkaline
washing solution (AC17205). You can define one more washing solution, WS3.
93
User manual
In the left-hand column a list of all the potential contaminant tests is displayed.
The tests for which contamination has already been programmed are marked in red.
Select a test and then press any of the following icons:
Press this icon to add the contaminated tests to the selected test. The contami-
nated test table will be activated for you to select the contaminated tests. The
washing solution to be used by the analyser to prevent contamination can be
indicated for each contaminated test.
Contaminated well Select this box if the contaminant test contaminates the reaction rotor cuvette.
Step 1 Enter the washing solution to be dispensed in the reaction cuvette if the con-
taminant is reagent 1.
Step 2 Enter the washing solution to be dispensed in the reaction cuvette if the con-
taminant is reagent 2.
In the cuvette section, the contaminant test is shown, with the washing solutions
to be used in step 1 and step 2.
10.3.4. Profiles
Profiles are names given to a set of tests with a diagnostic significance. Profiles are
sets of tests that are usually requested at the same time.
Biosystems
94
Figure 62 Screen for entering profile information
))See how to create icons, edition, deletion and printing of the test screen in chapter 10.3.1.
Parameters to be programmed in the profiles:
Name Profile name.
Sample type Select the type of sample the profile will have.
Test type Select the type of test, which may be: standard, calculated, ISE or external tests.
It serves to filter the number of tests to be displayed in the selection column.
Select the different tests you want to form part of the profile. Use the CTRL and
BLOCK CAPITALS keys to make a multiple selection.
Add the selected tests to the profile.
Eliminate a test from the profile.
Add all the test at one time to the profile.
Eliminate all the tests at one time from the profile.
10.3.5. Calibrators
Check the calibrator concentration values entered against those of the instruc-
tions for use (IFU) values of the calibration serum. Make sure you enter the
values with the test programme units.
CAUTION
Screen for programming the different calibrator parameters: name, batch, expiry
date, concentration, number of calibrators.
95
User manual
In the first tab it is entered the general information on the calibrator. Calibrators
for all the pre programmed tests are already programmed but you can create new
ones, edit or print with the corresponding icons.
The information to be entered by the user is the following:
Calibrator name Enter a name for the calibrator. If you don’t change its name, you can leave the
default ID
Biosystems
96
Lot Enter the calibrator batch. When the lot is changed, you must reprogramme the
concentrations of all the tests that use this calibrator. The programme issues a
warning, showing the tests affected.
Expiry Enter the days the calibrator will last once reconstituted.
No. Enter the number of calibrators.
Dilution mode Visible only when the standard number is greater than one. Select the dilution
mode: manual or automatic. In automatic multipoint calibrations, it is the
analyzer that automatically prepares dilutions for each point from the highest
concentration standard. In the manual option there are as many calibrators
prepared as points are defined in the curve.
In the second tab the calibrator is assigned to the test and the concentration value
is entered.
First select the test to which you want to assign the calibrator and then press the
edit button.
Calibrators Select the name of the calibrator you want to assign to the test.
Enter the calibrator values for the test.
If the calibrator is for a specific value, you only need to enter the concentration
value. In multipoint calibrators, you must enter the following parameters:
Increasing / decreasing This indicates whether the calibration curve will be ascending or descending.
Creating the curve Enter the method for creating the calibration curve. It may be one of the fol-
lowing methods: polygonal, linear regression, parabolic regression or spline.
Also select the axes on which you want to show the calibration curve: linear axes
or logarithmic axes.
Concentration values Enter the concentration values for each standard in descending order.
Diluent Choose the diluent to perform the automatic curve.
Max. concentration Enter the highest standard concentration.
Factor Enter each dilution factors for each point in decreasing order. It is important
to introduce the number one at the first point to indicate which is the highest
standard.
Ask for the test calibration programming in the test programming screen. You
can only change the calibration value in this screen.
10.3.6. Controls
In this screen the controls to be used are registered. You can create new controls
and edit, delete and print them.
You can also edit the minimum and maximum values of each test for each control
level.
Control name Enter the control name.
Sample type Enter the type of sample that will use this control.
Level Enter the control level (1, 2 or 3). This is used to facilitate the selection of an
entire level in the sample request screen.
97
User manual
Biosystems
98
Figure 66 Screen for changing the lot of a control
At the top of the screen is a list with all the patients already entered.
))See how to create, edit, delete and print icons in chapter 10.3.1.
Patient ID Enter a patient identifier in order to associate the patient data to the results.
Gievn name Enter the name of the patient.
Last name Enter the surname of the patient.
99
User manual
Complete one or more fields to enter the search criteria. For the date of birth and
age fields you must enter a range of dates and ages respectively.
Press this icon to make a search after entering the criteria.
Press this icon to delete all the search criteria. It is activated when data is entered
in any field.
Biosystems
100
Name Test name. This name is used to identify the test in the programme. The maxi-
mum length is 16 characters. Do not change this field content.
Short name Abbreviated name with up to 8 characters for use in certain screens in the
application.
Sample type Select the type of sample to be used. Tests which the blue symbol (plus) appears
they have more than one programming available in sample type dependency.
Available ISE test Select this option to view the lithium test on the sample selection screen (option
only available for Li+).
Report name Name of the test which will appear in the patient report. If there is no name
entered in this box, the test name will be shown in the patient report.
101
User manual
Biosystems
102
Figure 71 Screen for programming external tests
103
User manual
is launched, the state LED lamp turns orange, indicating that the analyser is ready
to be connected with the user programme.
The user programme is initiated and the communication connections are estab-
lished automatically.
If there is no communication with the analyser, check the communications setup
and press the connect button.
))See chapter 10.2.1
Once connection is established the analyser is connected. In this state the user can
access any part of the programme, except the parts that operate with the analyser
(start worklist, change rotor, etc). Press the initialisation button to operate with it.
The initialisation process is automatic and includes the following steps:
• Process for previously checking the state of the analyser and priming the
fluidic system.
• Wait 20 minutes for the reading rotor temperature to stabilise. Since the
baseline is totally dependant to the rotor to reach the 37ºC temperature, It is
not advisable to terminate the thermo time by pressing the Terminate button.
• Initialisation of the reading rotor baseline. This process consists of filling the
120 rotor wells with distilled water, making the photometric readings with
all the wavelengths in each well and emptying and drying the reading rotor.
The photometric readings will serve to compensate the samples values. The
photometric readings will serve to compensate the readings of the determina-
tions. If any reading is outside pre-established margins, this well is discarded.
If there are more than 10 discarded wells, the program suggests changing the
readings rotor. This whole process takes about 15 minutes.
The switching off process is performed by pressing this button. During the switch
off, the program will ask you to positionate the ISE clenaing solution whenever
the last sessions total number of ISE tests run exceed 50.
))See chapter 14.2.9
Once the washing operation has been completed the analyser will automatically
go off.
Biosystems
104
Figure 72 Screen for entering new samples
The screen has two parts. The top part contains the fields for entering the patients
and tests. The patients entered are shown in list format, in the bottom part. Each
individual patient and test can be edited and deleted until the list is positioned on
the rotor. Once a sample is being positioned, you are not able to delete it.
Sample class This field is used to select the type of sample to be entered. The types may be:
patient, blank, calibrator or control. It serves to make only lists of blanks and
calibrators or only lists of controls.
Urgent This is used to indicate that the sample is urgent. Only available for patient type
samples. Urgent patients are the first ones to be executed. If a work session is
interrupted and urgent patients are added, they will be processed as soon as the
instrument finishes the preparations in progress.
Patient/sample Field in which the sample code is entered. This code may be alphanumerical. If
no code is entered, it is generated automatically. The automatic code starts with
the character #, followed by the date in numerical format and a consecutive
number. The selection of a preprogrammed patient implies acquiring the iden-
tification assigned to that patient. Identification of the patient entered manually
corresponds to the same identification as the sample.
Number To enter several patients with the same test profile, enter the number of patients
to be created. If the previous field has data entered, this field will be deactivated.
The patient code is automatically generated and starts with #, to distinguish it
from those entered manually.
Sample type Select the type of sample before going to the test selection screen. Patients with
several sample types will have different tubes, one for each type. (For example:
for a patient with a serum sample and a urine sample, it will place both tubes in
the rotor. Each tube can only be assigned the tests of the sample type selected).
In the event of determining ions in urine, dilute it manually and place the dilu-
tion in a different tube.
105
User manual
Tests Button for accessing the list of tests and assigning them to the patient. See Fig-
ure 72.
This screen shows all the tests related to a sample type. The tests are divided into
functional groups: standard tests, calculated tests, ISE tests, external tests and
profiles.
Colour code for the test selection status.
Colour
Name Description
code
Selected Test selected
New tests can be added to a work session that has samples and reagents in position.
To do this select the patient for which the tests are to be added and press the tests
button. The programme will ask the user about whether new tests are to be added,
and if the answer is yes, the screen in Figure 72 appears. Samples or tests can also
be added to work sessions that are already initiated or finished.
Biosystems
106
After entering the tests for a patient, the information is shown in the list in Figure
71.
The information is separated into two tabs: patient information and information
on blanks/calibrators/controls.
The following information is shown in the patient tab:
107
User manual
This button allows you to eliminate tests and samples from the work session.
First select the row to be eliminated and then press the button. The blanks and
Biosystems
108
calibrators are automatically eliminated if there is no other patient with that
test. The controls can also be eliminated.
It allows you to save a session and retrieve it later. The programme will ask you
to enter a name for the session.
It allows you to load a previously-saved session. The programme will let you
select the name from a list of saved sessions.
Press this button to send the samples and reagents to be positioned in the ro-
tors. The programme will automatically change screen. Once the samples have
been sent to the positioning screen they are marked in grey.
))Chapter 10.5.2 describes the procedure for positioning on the rotor.
Press the button to open up the screen for entering the results for all patients
with external tests.
))See Figure 74
109
User manual
The samples sent via LIS are shown in the list, marked with an identification label
and in grey, to facilitate their identification. You can consult their data but they
cannot be changed.
The right-hand side of the window shows the patient data fields for the user to
enter the information. All the fields are optional.
The information can be entered quickly using ENTER or TAB which takes you
to the next field. When you reach the last field and press ENTER or TAB you are
taken to the next sample on the list.
Given name Patient name. On entering the letters of the name, a search is made and a list is
shown with the matching names stored. The user can select a name from the list
or enter a new one. In the latter case, the name will be stored for future use.
Last name Patient surname. A pop-up window is also generated with suggested surnames
to facilitate data entry. The suggested surnames are only those with the name
entered in the previous box.
Gender Male/Female/No value. Press the initial to quickly select the option.
Date of birth Patient’s date of birth. If the date is not entered correctly, it will be shown in
red and not memorised, but the field will not be blocked. It allows you to enter
the date in any operating system regional setting format (short or long format).
For patients aged less than 1 year, it is not necessary to enter the year. The age is
calculated automatically.
Performed by Open field for entering the name of the person who requested the tests of each
patient. This field will appear in the patient report.
Comments Open field with several lines for entering comments. This field will appear
in the patient report. You can enter comments with a line break by using the
CTRL + ENTER key combination.
Deletes all the data of the current patient from the pop-up screen.
Press this icon to undo the changes made for the current patient.
Biosystems
110
Press this icon to go to the last patient.
Press this icon to go to the first patient.
Press this icon to go to the previous patient with no data entered.
Press this icon to close the pop-up window. By default, all the changes are stored
without having to press any other icon to save the data. You can also close this
screen with the ESC key.
111
User manual
The samples placed in sample wells will not be read by the barcode reader and the
user must position them by dragging them manually or using the auto-positioning
button.
In the event of automatic barcode reading, when the session start button is pressed
(with Host Query activated), all the rotor positions will be read and the codes will
be matched with the orders pending in LIS.
When more than one bottle with the same reagent has been positioned, the pipet-
ting order will be the same as the order of each bottle on the rotor.
If the samples have barcodes that do not correspond to the work session samples,
the user can enter the necessary additional information: type of sample and test
to be performed on each sample.
If the programme detects an erroneous barcode, this is indicated by an icon over
the rotor position. Erroneous barcodes can be corrected manually.
Press this button if you want the analyser to automatically identify the session
samples positioned on the rotor. These samples will be shown in yellow. The
read barcodes that have not been matched with any of the requests on the list
will be shown in grey, and the user can assign the tests using the screen shown
in Figure 77. Elements in sample wells such as calibrators and controls will have
to be positioned manually or with the auto-positioning button.
Button for warnings related to elements required for the work session. When
this button is pressed a message will appear informing you of all the elements
that have still to be positioned. The same message will appear if an element has
not yet been positioned when the positioning screen is closed.
The programme will block all tests/samples in the event that a component required
for their execution is missing. Below are examples of cases in which blocking occurs:
Biosystems
112
Component not Rotor Blocking
in position
Distilled water Samples The test will be blocked when it is
performed for the first time and there are
no blank result values.
Calibration Samples The test will be blocked when it is
performed for the first time and there are
no factor result values.
Diluent bottle Reagents This is used to perform automatic or
manual repetitions with a sample dilution.
The test will be shown as blocked when
no diluent is detected on activating the
repetition (either because the bottle is not
in position or the volume is insufficient).
The test programmed with the pre-dilution
will be blocked.
This is used to pre-dilute samples (for
URINE or other tests). If it is not
positioned on the rotor, the test will be
shown as blocked.
Press this button to print out a report of the positions of all the elements in the
current session.
Press this button to memorise the elements positioned in the visible rotor. The
program will request a name for identifying the memorised rotor. When per-
forming a session reset, the positioned reagents are maintained, but the samples
are emptied.
Press this button to load the positions of the elements of a rotor that was previ-
ously memorised.
Press this button to delete the reagents or samples positioned on the rotor.
When the button is pressed a selection menu appears for the user to select the
sample or the reagent rotor.
Press this button whenever you run out of volume. Perform this action when
the volume termination alarm is triggered. For reagents with barcodes you only
need to place a new bottle on the rotor and press the barcode reading button
and the information on the volume of the new bottle will automatically be
updated.
Press this button to release the samples and reagents shown in red due to detect-
ing a clot or a tip collision. Before pressing it, the user must carefully check
the tubes of elements in red and decide on the appropriate course of action:
eliminate the clot, correct the tube size, position the tube correctly, remove the
reagent bottle cap, etc.
On pressing the button, a pop-up window emerges in which the user can enter
additional patient information from the work list.
))Chapter 10.5.2 explains how to enter additional patient information.
113
User manual
Figure 77 Screen for creating the work list from positioning
Biosystems
114
Column name Description of the fields
Status Information only appears when making a Host Query.
It indicates the status of requests to the LIS. It has the
following states:
• ASKING: Request sent to the LIS
• PENDING: Request already sent and awaiting
receipt of the worklist for the sample.
• REJECTED: Request rejected by the LIS.
• NO INFO: The LIS has no information about this
sample.
Select all It allows the selection of all the specimens in order to request the LIS for the
work order or to manually create the worklist.
Button allowing a Query All to be made directly to the LIS from this auxiliary
screen.
The bottom of the screen allows you to create the worklist manually after the
specimen barcodes have been read. To do this, select the specimen or group of
specimens and assign the sample type, if it is urgent, and assign tests using the
TEST button. Once you have done this, press save and continue with the next
specimen.
If the sample type is not coded in the barcode, the type must be assigned manually
to all the specimens. To do this, select all the specimens or a group of specimens
and assign the sample type from the pop-up box at the bottom of the screen.
115
User manual
Press this button to stop or abort the work session. Once the button has been
pressed, the screen shown in Figure 78 will appear.
Select one of the following options:
• Stop: This action stops the work session, ends the preparations in progress
until a concentration result is given and does not prepare any more. The next
action to be taken is to reset the work session.
• Abort: This action aborts the work session. The execution of the list stops im-
mediately and the preparations in progress in the analyser are lost. The next
action to be taken is to reset the work session.
• Cancel: Closes the window and continues executing the current list.
Biosystems
116
10.5.8. Delete session
Menu option at Work session that allows you to delete a previously memorised
session. When this option is pressed a pop-window opens. Select the name and
press the delete icon.
10.6.1. Main
Screen which informs about the status of the analyser: the analyser elements that
are switched on (refrigeration system, ISE module), main sensors (covers, tem-
peratures), work session times, graphic information on alarms and information
on processes being executed by the analyser.
117
User manual
10.6.2. Worksession
This tab displays information about the work session being performed by the
analyser, the state of the samples and of the tests.
The information is organised in a table with all the samples and tests of the session
and a graph summarising the state of the preparations.
The work session is ordered so that emergencies are always executed first. Before
the patient samples the blanks, calibrators and controls of the tests assigned to
the patients are executed.
Biosystems
118
Figure 80 Work session screen
The table shows the state of the samples and tests through a colour code and ad-
ditional information using icons.
119
User manual
Test status This graph informs you about the total number of preparations requested and
their status: completed, in progress, pending, blocked and completed with
alarms.
Biosystems
120
Colour code Name Description of the bottle identification
Insufficient Reagent bottle in which the analyser has
vacuum / detected insufficient volume for making
Blocked due to the preparation. The programme will block
collision all the subsequent preparations using that
reagent. To release the bottles due to insuf-
ficient volume the user must insert a new
bottle and read the barcode.
To release the bottles due to a collision the
user must check the positions of the bottles
affected by the collision and press the release
button to be able to reuse that bottle.
Insufficient Warning low volume.
volume
Not used Reagent positioned but not used in the work
session.
Error in code Error in barcode reading.
reading
Unknown Bottle positioned but not identified.
Selected Bottle selected.
This screen is for information purpose and cannot be used for making changes in
the sample positions or resolving volume alarms. To do this go to the sample and
reagent rotor position screen.
))See chapter 10.5.3
10.6.4. Reactions
Screen informing the status of each of the reaction cuvettes. It corresponds to the
reactions tab of the monitor screen.
The user can press any cuvette to see detailed information about its content or the
preparation it contains. It also allows the reaction curve to be accessed when the
cuvette contains a preparation.
There is a colour code for identifying the status of each cuvette.
121
User manual
Biosystems
122
Figure 81 ISE module monitor display
10.6.6. Alarms
Screen showing a list of all the alarms that appear while the analyser is operating.
It corresponds to the alarms tab of the monitor screen.
Each alarm has the following information:
• type, indicating the severity. Serious alarms can interrupt the analyser opera-
tion.
Icons Description
Warning icon. Indicates that an alarm has gone off which requires
the intervention of the user. This type of alarm does not interrupt
the operation of the analyser.
Icon indicating that the alarm has been resolved.
Icon indicating a serious alarm. Indicates that a serious alarm has
gone off and operation is interrupted. Depending on the type of
alarm, for instance, detection of a collision in one of the tips, the
user will have to press the analyser recovery button to resolve the
alarm.
• Date
• Time
• Alarm name
• Alarm description
• Possible solution
123
User manual
The alarms are ordered by date and time of arrival, but they can be ordered using
any other criterion. Press the column header in which you want to order them.
If pressed once, they will be ordered in ascending order, and if pressed twice, in
descending order.
On the left is a list of patients and tests performed in the session (separated into
two tabs). It allows you to see all the results for each patient or all the results for
each test. Select an element from the list to see the results in the tables on the right.
Biosystems
124
Patient tab Field Description
Buttons allowing you to show or hide the replicates
with a result.
If you press the header icon, all the patient tests will
be shown or hidden.
Only one of the two icons appears. Every time you
press the icon, one will be changed to the other
alternatively.
OK Indicates that the result has been accepted. It will
be sent to the historical log and to LIS. When
repetitions are executed the last result is always
accepted.
This indicates whether it was sent to LIS
automatically or manually.
New This allows you to repeat a preparation.
This option is deactivated if the repetition
option has only been selected for LIS in the LIS
configuration screen.
))
See the different repetition options in chapter 10.7.4
Test Test name
Type Sample type
This allows you to access the kinetic reaction graph.
Press this icon to print out a report with all the patients in a compact way.
Press this icon to see a summary table of the results. A table appears with the
results of all the patients and all the tests of the session in progress.
125
User manual
Press this icon if you want to print out a list of the patient results including the
absorbance values obtained.
Press this icon to manually send the selected results to a LIS laboratory informa-
tion system.
Press this icon to enter the external test results. When the button is pressed an
auxiliary screen appears in which you can enter the external test values (the test
must be selected previously for the button to be activated).
))See chapter 10.3.9
Press this icon to send the selected samples for repetition.(I.e., tests with the
New field activated).
This option is deactivated if the repetition option has only been selected for LIS
in the LIS configuration screen.
))See the different repetition options in chapter 10.7.4
Biosystems
126
Field Description
Absorbance Reagent This shows the R2 absorbance value. Only shown in
tests with differential programming.
Kinetic blanc limit This value is programmed in the test. It is used to
check the slope of the kinetic reaction exceeds that
limit.
Initial abs. This shows the first value of kinetic absorbance.
Only shown in tests with kinetic and fixed time
programming.
Absorbance main filter This shows the main filter blank absorbance
value. Only shown in tests with bichromatic
programming.
Blank absorbance limit Blank limit value. This value is programmed in
the test. It is used to check that the reagent is in
good condition. Either in incresing reactions, if
the absorbance exceeds the established limit, or
in decreasing reactions, if the absorbance is below
the established limit, the programme will issue a
warning message in remarks.
Remarks This displays the alarms that could appear in the
results.
))
See the possible alarms in chapter 13
Date Date on which the result was delivered.
Repetition mode This shows the mode in which the sample was
repeated.
127
User manual
Field Description
Number Indicates the replicate number, if there is more than
one replicate.
Absorbance Calibrator absorbance value that will intervene in
calculating the factor.
Theoretic concentration Calibrator concentration value. This value comes
from the test programming.
Units This shows the units in which the test has been
programmed.
Factor Value calculated based on the calibrator absorbance
that will intervene in calculating the concentration.
Factor limits Factor limit entered in programming the test. If
the factor value is off limits, a warning message will
appear in the remarks field.
Remarks This displays the alarms that could appear in the
results.
))
See the possible alarms in chapter 13
Date Date on which the result was delivered.
Repetition mode This shows the mode in which the sample was
repeated.
Biosystems
128
Field Description
Number Indicates the replicate number, if there is more than
one replicate.
Absorbance Absorbance value of the control obtained.
Concentration Calculated concentration value of the control.
Units This shows the units programmed in the test.
Concentration limits This shows the maximum and minimum limits for
the controls entered in programming the test.
Remarks This displays the alarms that could appear in the
results.
))
See the possible alarms in chapter 13
Date Date on which the result was delivered.
Repetition mode This shows the mode in which the sample was
repeated.
When the patient tab is selected the information viewed is detailed in chap-
ter 10.7.1.
129
User manual
The reaction graph button can be accessed in the results, work session and reac-
tions screens.
You can request the repetition of tests manually by selecting the New field in the
results table.
To do this press the New field and a window like the one shown in Figure 84
will appear. For automatic repetitions, the increase/decrease factor will be that
programmed by default in Procedure screen of Tests Programming.
You can select the repetition criterion at that time for the patient samples. The
criteria are:
• Repeat with an increased sample. The programmed test value is usually shown.
If the value is changed, the programme warns the operator that the test limit
established has been exceeded. This repetition changes the ratio of volumes
between sample and reagent for increasing the sample absorbance. The re-
sulting concentration is divided by the increase factor. This repetition is used
to increase the sensitivity of samples below detection level limit. Its limit is
according the default test programming and the limit of pippeting for both
reagents and samples. The lower the sample pippeting volume is, the more
times you will be able to concentrate.
• Repeat with a reduced sample. Enter the dilution factor of the sample. The
programmed test value is usually shown. The user can enter a value between 2
and 200. The repetition makes a prior preparation for diluting the sample and
then it prepares the measurement based on the diluted sample. The resulting
concentration is multiplied by the dilution factor. This repetition is used for
samples that are outside the linearity limits, that have a substrate depletion.
alarm or for samples that are outside the calibration curve.
• Repeat under the same conditions programmed in the test.
Biosystems
130
Figure 84 Screen for selecting repetitions
Recover the recommended dilution factors (the same ones programmed in the
test).
The blanks, calibrators and controls have these options deactivated, as they are
always repeated with the same initial conditions.
Each row of results has an informative icon indicating the type of repetition crite-
rion applied (for both automatic and manual repetitions). There is an identifying
icon for the original result and another for the repeated result
131
User manual
After making the selection, press the icon to execute the search and view the
results at the bottom of the screen.
Date range Enter the start and end date for selecting the results by a date range.
Patient Enter the patient code, name or surname to select the results by patient. All the
results for a patient starting with the value entered are displayed.
Barcode Enter a sample barcode to select the results.
Urgent The available options are: All, urgent or routine.
Test type The available options are: All, standard, calculated, ISE, external
Sample type The available options are: All, SER, URI, PLM, WBL, CSF, SEM, LIQ.
Test name Enter the name of the test to make the selection.
The results are shown in a table, ordered by date. On pressing the heading of a
column in the table, the results of that column will be rearranged.
Press this icon to display the results of the next page.
Press this icon to display the last results.
Press this icon to display the results of the previous page.
Press this icon to display the first results.
Press this icon to print out the results that were previously selected. If you want
to select all the results, press the heading selection box.
Select this icon to print out the results with a compact report, i.e., with no
patient header and all the results shown continuously.
Press this icon to send the selected results to a LIS laboratory information sys-
tem. This is a manual export.
Biosystems
132
Press this icon to eliminate the selected results. Once eliminated, they cannot be
recovered.
On pressing the button, a pop-up window emerges in which the user can
complete and change additional patient information from the historical results
selected.
))Chapter 10.5.2 explains how to enter additional patient information.
At the top of the screen there are several fields that permit the entry of selection
criteria for restricting the viewing of the results. More than one selection criterion
may be selected at one time.
After making the selection, press the icon to execute the search and view the
results at the bottom of the screen.
Date range Enter the start and end date for selecting the results by a date range.
Test name Enter the name of the test to make the selection.
The results will be shown in two tables, ordered by date. The first table shows
the results of the blanks and the second one shows the results of the calibration
standards. On pressing the date or test heading in the table, the results of that
column will be rearranged.
The blank and calibrator fields displayed are the same as the fields shown on the
current session results screen, selected by test.
))See chapter 10.7.2 for a description of each field in the blank and calibrator screens.
133
User manual
Date range Range of dates of the results to be viewed. The date range reported by default is
that which allows all the results pending review to be displayed.
Biosystems
134
formula24-ENG
Rejection criterion Number of standard deviations for determining the admissible value interval
limits for the results:
formula18-ENG
Range=Mean± Rejection criterion⋅SD
∑ Xi
Mean= i=1
n
n
∑ xi−Mean2
i=1
SD= criterion cdot SD )
Range = Mean +- (Rejection
n−1
SD
CV = ⋅100
Mean
The target values for each control are shown in the columns on the left: means,
standard deviation (SD), coefficient of variation (CV) and admissible range of
values, calculated in accordance with the selected calculation mode and rejection
criteria.
The columns on the right (Results area) show the statistical values for each control,
calculated based on the available results:
∑ Xi
Mean= i=1
n
SD Standard deviation of the results:
n
∑ xi−Mean2
i=1
SD=
n−1
If the statistical calculation mode is selected, the results used for calculating the
target values are not used in the calculation.
• Details of individual results. This shows the individual results for all the
controls selected within the specified date range, validated pursuant to the
indicated calculation criteria. The information displayed for each result is the
following:
Parameter Description
n Number of series executed.
Name Name of the control to which the result belongs
Date Result date/time.
Mean = {sum from i=1 to n {X_i }} over {n}
Result Result value.
Manual result The result changed icon is shown if:
indicator • the result value has been changed manually
• the result has been entered manually
Unit Result measuring unit
SD = sqrt {sum from i=1 to n (x_i - Mean)^2 over {n-1}}
Absolute error Difference between the result value and the target mean.
Relative error Absolute percentage of error divided by the target mean.
Biosystems
136
Parameter Description
Alarms Alarms generated while the result is being validated.
These include:
• result outside admissible value range
• infringing of rules applied
Results with alarms are shown with red lettering.
In addition, if the calculation mode is statistical and the results used to calculate
the target values are included in the results group in the displayed set of results,
the symbol Xm is shown to the left of the number of the series executed.
This opens up the auxiliary screen for entering a new series manually, giving the
date, time and value of the result for one or more of the available controls.
The added results are shown with the icon changed in the listed Individual results
table
))Figure 88 shows the screen for entering new results.
Opens the auxiliary screen for changing the value of the selected result (only the
value, the date cannot be changed) or temporarily exclude it from the calcula-
tion and validation processes. An excluded result can be included later.
The Listed individual results table shows the changed results with the respective
icon and the excluded results are shown deleted and on a grey background.
))Figure 89 shows the screen for editing results.
It allows you to permanently delete the results selected.
Opens the auxiliary screen that shows the results in graphic format. The display
type can be selected: Levey-Jennings or Youden. The controls to be graphed can
also be selected: between 1 and 3 for Levey-Jennings and between 1 and 2 for
Youden.
In the Levey-Jenning graph the Y axis values will depend on the number of con-
trols graphed:
• If only one control is graphed, it will display concentration values and the
scale of the standard deviation in multiples.
• If several controls are graphed, it will show the values of the scale, in standard
deviation multiples.
))Figure 90 and 84 show the screens of the Levi-Jenning and Youden graphs respectively.
137
User manual
Figure 90 Screen with the Levy-Jennings graph
Biosystems
138
Figure 91 Screen with the Youden graph
The buttons located at the bottom of the screen are always available:
Opens the screen that allows the current quality control results to be accumu-
lated by control and test.
))See chapter 10.8.4.1
139
User manual
formula24-ENG
On the left is a list of controls with results for accumulation. In selecting a control
from the list, the number of the active lot will be shown on the right of the screen,
along with the list of tests with results that must be accumulated for the control.
Load the information to be accumulated for each test for the selected control.
Double-click on a control in the list for this button to function.
The information of the list of tests with results to be accumulated for the control
is set out in two tables:
• Last accumulated values: this shows the mean, standard deviation (SD)
and range of values admitted for the last accumulated item for each test, if
accumulated items exist previously for the selected control; otherwise the
respective boxes are shown as empty. Use the tick to select/unselect the test
for accumulation (the table is loaded/downloaded from Data to accumulate).
• Data to accumulate: only for the tests selected in the above table; it shows
the calculation of the values that will be accumulated:
∑ Xi
Mean= i=1
n
Biosystems
140
Parameter Calculation method
SD Standard deviation of the results:
n
∑ xi−Mean2
i=1
SD=
n−1
If the mode for calculating the test is statistical, the results used for calculating the
target values are not included in the set of values to be accumulated.
Double click on a test in this table, open the Quality control results screen which
displays a list of the set of results to be accumulated.
))See chapter 10.8.3
Icon indicating that the set of results to be accumulated includes one or more
series with validation alarms. This icon is shown to the right of the mean.
Execute the accumulation process for the control results of the selected tests. If
all the tests are accumulated for the selected control, the control is downloaded
from the control's list.
141
User manual
On the left is a list of tests with accumulated quality control results. In selecting
a test from the list the information of all the controls with accumulated results
for the test appears on the right of the screen. There are two clearly defined zones
in this list area:
• Results accumulated by control/batch: this allows you to specify the date
range of the accumulated results to be consulted. In changing the date range,
the content of both list areas is emptied and it is necessary to select the search
Biosystems
142
formula19-ENG
formula25-ENG
button to reload them. The information on controls with accumulated series
for the test within the specified date range specified is shown in the table, with
the following structure:
Parameters Description
N Number of accumulated series
Mean Weighted mean of the accumulated results.
Calculation method:
n
∑ Meani⋅ni
Mean= i=1 n
∑ ni
i=1
The tick is used to activate/deactivate the accumulated series for the control/lot in
the list. A maximum of 3 controls can be selected.
Date range Range of dates of the accumulated results to be displayed. The date range that is
normally given displays all the accumulated results for the selected test.
Execute the search and validation of results applying the selected calculation
Mean = {sum
criteria.from i=1 to n {Mean_i cdot n_i }} over {sum from i=1 to n {n_i}}
Execute the updating of the target values defined for the test and control/batch
selected, assigning the latest accumulated statistical values to them. Functional-
ity only available for an active control lot and for Supervisor level users.
• List of accumulated series: only for the controls selected in the previous
table; it shows the accumulated series list within the date range selected, in
tabled and graphic format. In the tabular display, the information shown for
each control is the following:
Parameter Description
CV = SD over Mean cdot 100
Number of accumulated series
Control Control name
Parameter Description
Lot Number Control lot number
Date Date and time when the accumulated series was created
n Number of individual accumulated results in the series
Mean Statistical mean of the accumulated results in the series
Unit Results measuring unit
SD Standard deviation of the accumulated results in the series
CV Coefficient of variation of the accumulated results in the
series
Range Range of admissible values for the accumulated results in
the series
Biosystems
144
Figure 95 Screen showing ISE electrodes calibration historical log
145
User manual
At the top of the screen there are several fields that permit the entry of selection
criteria for restricting the viewing of the results. More than one selection criterion
may be selected at one time.
After making the selection, press the icon to execute the search and view the
results at the bottom of the screen.
Date range Enter the start and end date for selecting the results by a date range.
Type The available options are: All, error and warnings.
The results are shown in a table, ordered by date. On pressing the heading of a
column in the table, the results of that column will be rearranged.
Press this icon to display the results of the next page.
Press this icon to display the last results.
Press this icon to display the results of the previous page.
Press this icon to display the first results.
Press this icon to eliminate the selected results. Once eliminated, they cannot be
recovered.
10.9. Utilities
Biosystems
146
Figure 99 Rotor change
147
User manual
Action Description
Maintenance Empties the tubes. Only activates the waste pump. In
the repetitions parameter indicate how many times the
action must be executed.
Purge A It performs a priming cycle with calibrator A, using a
volume of 100 µL. In the repetitions parameter indicate
how many times the action must be executed.
Purge B It performs a priming cycle with calibrator B, using a
volume of 100 µL. In the repetitions parameter indicate
how many times the action must be executed.
Priming A It performs a priming cycle with calibrator A, using a
volume of 300 µL. In the repetitions parameter indicate
how many times the action must be executed.
Priming B It performs a priming cycle with calibrator B, using a
volume of 300 µL. In the repetitions parameter indicate
how many times the action must be executed.
Wash It performs a wash cycle with the ISE washing solution. In
the sample rotor pos. parameter indicate the position of the
tube with the washing solution. In the volume parameter
indicate the volume to be dispensed for washing.
Biosystems
148
Action Description
Activate reagent Execute this action to activate and memorise the reagent
pack kit in the programme. It is also used to memorise the
installation date and record the consumption of the
calibrators. The program issues a warning when the
calibrators are no longer usable.
Activation of Execute this action to activate and memorise the electrodes
electrodes in the programme. It is used to record the consumption
of the electrodes and warn the user when they are no
longer usable.
Activation of ISE Use this action to tell the programme you have installed
preparations an ISE module.
149
User manual
by sending a cancelled order message; in this way the LIS can record the user and
date on which this action was performed.
The intensive use of this tool is not recommended. Formally, it must always be the
LIS system that distributes orders among the laboratory instruments and reports
the cancellations to each instrument.
NOTE
The reasons why the LIS decides to cancel orders in the BA200 may be the fol-
lowing: the sample tubes do not reach the instrument, instrument alarms exist
which prevent the work from being done and the orders may be sent to another
instrument.
This action is only available if the analyser is in the STAND-BY mode and the LIS
communications are activated and free from error.
Delete internal storage queues Delete the queue of messages to be uploaded to the central LIS system. If there
is a communication problem with the LIS, the messages may accumulate in
the queue, waiting for communications to be restored. If the user is no longer
interested in sending these messages, they can be deleted using this option.
This action is only available if the analyser is in the STAND-BY mode.
Selection of Tracing Level for Allows the trace level to be configured for capturing information in the LOG
LIS communication log about the operation of communications between the LIS and the BA200. This
option is used to diagnose potential communication problems during the inte-
gration of LIS with the BA200.
Biosystems
150
Figure 102 Screen for generating a report for the technical service.
151
User manual
Biosystems
152
To be able to perform an analysis of the data:
• Select the text to be analyzed.
• Copy it
• Paste it on a text editor (for example, Notepad ++). From the editor you can
perform searches in an easier way.
10.10. Help
10.10.3. About
This option displays the installed software version.
10.11. Exit
To exit the programme go to the exit menu and select one of the two options:
Exit with shut-down This option closes the programme and tells the analyser to switch itself off and
complete the closing process.
Exit without shut-down This option will only close the programme and leave the analyser on and on
standby.
153
User manual
Accessories list
Code Representation Description
AC17274 DVD User and service programme
Biosystems
154
Accessories list
Code Representation Description
AC16 362 60 mL reagent bottles (20)
155
User manual
Accessories list
Code Representation Description
AC16749 High contamination waste bottle cap
Biosystems
156
List of ISE module accessories (optional)
Code Representation Description
5625 Set of ISE module tubes
5202 K+ electrode
157
User manual
Biosystems
158
12. Support and warranty
The BA200 analyser is designed to perform biochemical and turbidimetric analyses.
Its operation is optimised for the BioSystems Reagents line. For information about
all the available measuring procedures please contact your habitual distributor.
159
User manual
Biosystems
160
Type of Alarm/Error Cause of the problem Solution proposed
alarm
The purified water The purified water tank has not Check that the water inlet
tank has been been filled for a long time. This configuration is correct. If there is
empty for too action blocks the work list. an external tank, check that it is full.
long Resolve the problem of the absence
of water and press the bottle change
confirmation button.
Warning that the The reaction rotor cover has Close the reaction rotor cover.
reaction rotor opened during the stopping phase.
cover is open
Warning that You have started a worklist without Insert a new reaction rotor with the
there is no a reaction rotor. rotor change utility.
reaction rotor
Error due to the The wash station has collided. This Check the correct positioning of the
reaction rotor action stops the work list. reaction rotor. Check that the wash
stopping station suspension is not blocked.
Press the recovery button. If the
alarm continues notify the technical
service.
Warning that The analyser has detected a Centrifuge the sample again.
a clot has been blockage in the sample tip.
detected
The analyser has a fluidic problem Check the connections and the
water inlet configuration. Check
that there is sufficient water in the
external tank.
Warning The sample or calibrator volume Press the pause button. When the
that there is is insufficient. The programme program tells you to, access the
insufficient sample blocks the next tests for the current sample rotor and replenish the
volume patient. sample physically and confirm it by
pressing the refilling button. Press
the continue button.
Warning of There is insufficient volume in the Press the pause button. Check the
insufficient rotor cuvette where the sample sample or diluent volume. Press the
volume in the dilution is prepared. The program continue button.
diluted sample blocks the diluted sample in
progress.
Error in adjusting Baseline adjustment values off Change the reaction rotor. Check
the baseline limits. This action is done with the that the wash station is operating
rotor change. correctly. If the alarm continues
notify the technical service.
Warning message Too many reaction rotor cuvettes Change the reaction rotor.
to change the rejected. This warning does not
reaction rotor block the execution of the work
list.
161
User manual
Biosystems
162
Type of Alarm/Error Cause of the problem Solution proposed
alarm
Reagent kit connector not correctly Check the reagent kit connector.
positioned
Warning that the The reagent kit has expired Change the reagent kit
reagent kit has
expired
Warning that Calibrator A or B has been used up. Change the reagent kit
calibrator A or B
of the reagent kit
has been used up
Warning that One of the electrodes has expired Change the expired electrode.
an electrode has
expired
Warning that the The number of uses necessary Change the electrode.
number of uses for for the correct operation of an
an electrode has electrode has been exceeded
been exceeded
Error in Insufficient sample in the ISE Check the sample volume and repeat
dispensing the module reader or air bubbles the sample.
sample detected
Slope value under Non-alignment of the electrodes Remove the electrodes. Inspect the
the established toric joint (O-ring)
limit Reinstall the electrodes.
The calibration solutions have been Replace the reagent kit
used up
End of the electrodes' useful life Replace the electrodes
Air bubbles in the reference Remove the electrode. Tap it several
electrode times to eliminate the air bubbles.
Reinstall the electrode. Recalibrate
Shunt from an This may occur if the electrode is Purge calibrator A and recalibrate
electrode new or calibrator A has just been
installed.
If the electrode is new, it may shunt
initially while rehydrating for 15
minutes.
End of the electrode's useful life Replace the electrode
Air in the sample Insufficient sample volume. Check that there is sufficient sample
and/or calibrator volume.
Check that the tip is not partly
blocked.
Loss of fluid Determine the leak.
Call the technical service.
163
User manual
Biosystems
164
Type of Alarm/Error Cause of the problem Solution proposed
alarm
Principal Abs < This message will appear for Check the reagent status, it may be
Blank Abs limit tests programmed as descending damaged: this may be due to the
endpoint tests. reagent expiry date being exceeded
This result indicates the status of or to poor conservation of the
the reagent. reagent.
Working Reagent This message will appear for Check the reagent status, it may be
Abs < Blank Abs tests programmed as descending damaged: this may be due to the
Limit differential tests. reagent expiry date being exceeded
This result indicates the status of or to poor conservation of the
the reagent. reagent.
Initial Blank Abs This message will appear for Check the reagent status, it may be
< Blank Abs Limit tests programmed as kinetic or damaged: this may be due to the
descending fixed time tests. reagent expiry date being exceeded
The initial blank Abs value is not or to poor conservation of the
used to calculate the concentration. reagent.
This result indicates the status of
the reagent.
Kinetic blank > This message will appear for tests Repeat the blank.
Kinetic blank programmed as kinetic and fixed
limit time tests.
For descending reactions
calculation of the kinetic blank
will be converted into positive, to
correctly compare it with the limit.
It is used to check that the blank is
correctly executed.
Incorrect curve For an ascending calibration Repeat the different calibration
curve: all the absorbance points curve points.
must be in ascending order as the
concentration increases
For a descending calibration curve:
all the absorbance points must
be in descending order as the
concentration increases
Factor calculated The factor value is outside the Repeat the calibration.
off limits limits entered in programming the
test
Calibration factor The calibrator absorbance is lower Repeat the calibration.
NOT calculated than the blank absorbance.
It was not possible to calculate the
calibrator absorbance.
It was not possible to calculate the
blank absorbance.
The calibrator absorbance has
exceeded the photometric limit
>3.5
165
User manual
Biosystems
166
14. Maintenance and cleaning
After performing any maintenance operation, make sure that the instrument is
in the correct installation conditions for its operation.
NOTE
167
User manual
Biosystems
168
14.2.5. Cleaning the barcode reader window
If the programme reports a high number of errors in reading the barcodes, check
the status of the barcode reading window.
1. Turn off the analyser.
2. Remove the reagent and sample rotor cover.
3. Take out the rotor.
4. Clean the interior window with a damp cloth.
169
User manual
Biosystems
170
1
Cleaning the ISE To prevent blockages of the ISE module waste tube due to poorly coagulating
module tubes samples, proceed as follows to clean the waste tube:
1. Remove the ISE module waste tube from the peristaltic pump housing.
2. Separate the yellow L-shaped part from the bottom part of the module.
3. Use commercial bleach. Check that it has a sodium hypochlorite concentra-
tion of between 5 and 6%.
4. Fill a syringe with 5 mL of bleach and connect it to the L-shaped connector
that was previously removed.
171
User manual
5. Slowly inject the syringe to fill the waste tube with the bleach. Let the bleach
act inside the tube for 6 minutes.
6. Remove the syringe from the L-shaped connector and fill it with distilled water.
7. Reinsert the syringe into the L-shaped connector and rinse out the waste circuit.
8. Remove the syringe from the L-shaped connector and plug the L-shaped
connector into the module.
9. Replace the waste tube in the peristaltic pump.
10. Execute a couple of priming operations with CAL A.
14.3. Maintenance
Biosystems
172
Unpack the reference electrode. Remove the wire that has a yellow label (keep the
wise in case you need to turn off the module and store the electrode for a long time).
Ensure that there are no accumulated salts at the ends of the measuring channel.
Other electrodes Unpack the new electrode. Remove the adhesive tape that protects the fluid
transportation channel. Check that the rubber seal in the opening is in place. If
there is no rubber seal, put it back in its position. Each box of spares has pair of
seals, in case one is lost.
Proceed as follows to replace the electrode (both the reference and the other
electrodes)
1. With the user programme in the ISE utilitiessection, perform 1 Maintenance
cycle, to empty the ISE module channel.
2. Turn off the ISE module power supply.
3. Open the doors and remove the front cover of the ISE module.
4. Press the yellow button down to release the pressure in the electrodes.
5. Remove all the electrodes.
6. Discard the electrode that must be changed.
7. To put the electrodes back in position, press the yellow button down and
first insert the reference electrode and then the other electrodes, in the order
shown in Figure 108.
8. If there is no Li+electrode, put an empty electrode in its place, to ensure con-
tinuity in the channel through which the sample flows.
9. Release the yellow button to supply pressure to all the electrodes and ensure
good fluid communication.
10. To ensure that the electrodes are properly placed, press them at the front until
you hear a click or they have been correctly seated.
11. Turn on the ISE module power supply.
12. Put the front cover back in position and close the doors.
13. With the user programme, execute the actions in the number and order in-
dicated in the ISE functions, change electrodes section.
173
User manual
Perform the last 4 actions 3 times. If the calibration is not satisfactory, wait 5
minutes and repeat the last 4 actions.
LI+
Na+
K+
Cl−
REF
WASTE
Biosystems
174
14.3.2.2. Changing the reagent kit
Open the front door and you will see the ISE module reagent kit on the left.
Remove it from its housing and take the connector out of the packet. Press the
yellow button to separate the kit from the connector.
Dispose of the waste in accordance with the applicable national or local govern-
ment legislation governing the disposal of dangerous biological waste. Do not
perforate or open the reagent kit.
BIOHAZARD
Check that the new kit is from a zone with room temperature similar to that of
the analyser.
Unpack the kit and remove the three red protective caps. Do not press the side of
the kit after removing the caps as the solutions it contains could flow out. Have
some paper at hand to dry the surface where the connector is coupled in case any
liquid flows out.
Position the connector in the correct direction and press lightly until you hear a
click. Write the installation date on the side of the kit.
Place the kit in its housing.
With the user programme, execute the actions in the number and order indicated
in the ISE utilities, change reagent kit section.
))See chapter 10.9.3
175
User manual
Biosystems
176
• The tubes marked B must be installed in the pump (3). The order for putting
them in place starting from the bottom is B2 and B1.
• The tubes marked W must be installed in the pump (4). The order for putting
them in place starting from the bottom is W1 and W2.
1
2
3
Take care when connecting the tubes of the waste pump (4) as they are connected
in reverse order to the tubes of the pump for calibrators A (2) and B (3).
With the user programme, execute the actions in the number and order indicated
in the ISE utilities, change pump tube section.
))See chapter 10.9.3
177
User manual
14.3.2.4. Turning off the ISE module for a long period of time
If the analyser is not to be used for a long period of time, for instance, during the
holiday period, proceed as follows to conserve each electrode, tube and reagent kit.
To leave the module inactive, uninstall the electrodes and the reagent kit and clean
the tubes to prevent salts or traces of serum from blocking the circuit.
Execute the actions in the number and order indicated in the user programme,
in the ISE utilities section.
))See chapter 10.9.3
Step Action Repetitions Description
1 Filling 3 This action dispenses 300 µl of solution
with Cal A into the module cup. Use the syringe
A and long tip supplied with the accessories
box to suction the liquid and deposit it in
a sample well or in any other container.
Repeat this action 3 times. This solution
will be used to fill the electrode channel in
the storage procedure.
Biosystems
178
Na+ and Cl-electrodes Place each electrode separately in a sealed bag.
Reference electrode Reinsert the wire with the yellow label it into the reference electrode opening
and then place it in an individual sealed bag.
K+ and Li+electrodes K+ and Li+ electrodes Suction a small quantity of the calibrator A dispensed
in a sample well during the first step.
Inject the appropriate quantity of calibrator A into the channel of the K+ and
Li+ electrodes until the liquid fills the channel.
Cover the two ends of the channel (both sides of the K+ and Li+electrodes) with
adhesive tape to hold calibrator A in place.
Place the K+ and Li+ electrodes in a sealed bag.
Reagent kit Remove the reagent kit from the analyser and dispose of it.
Peristaltic pump tubes Remove all the tubes from the fluids and rinse them with purified water. Use
the syringe with the medium tip.
Thin tubes Rinse the thin tubes with purified water. Use the syringe with the small sized
tip.
179
User manual
Biosystems
180
15. Technical characteristics
181
User manual
Biosystems
182
Type of measurement Direct ISE
Sample volume Serum: 100 mL
Urine: 200 mL
183
User manual
Biosystems
184
16. Measuring and calculation procedures
This chapter describes the different analysis modes of the analyser and the calcula-
tions made to obtain the analytical results, i.e., the concentration values of the dif-
ferent sample analytes. The different formulae used are indicated in each case. The
controls are treated in the same way as the patient samples in all the calculations.
Symbol Description
ABS Absorbance value read in one instant of the reaction.
A Absorbance value calculated based on the chosen analysis
mode.
[...]λprincipal Absorbance value at the main wavelength.
[...]λreference Absorbance value at the reference wavelength.
[...]L1 Absorbance value in time L1
[...]L2 Absorbance value in time L2
DABS Increase in absorbance
VM Sample volume
VR1 Volume of reagent 1
VR2 Volume of reagent 2
C Analyte concentration
F Factor
ABlank Blank absorbance
ACalibrator Calibrator absorbance
Asample Sample absorbance
Ccalibrator Known calibrator concentration
185
User manual
Analysis mode
Endpoint monoreagent
Endpoint bi-reagent
Differential switch
Fixed time monoreagent
Fixed time bi-reagent
Kinetic monoreagent
Kinetic bireagent
Each of the analysis modes executed by the analyser is shown below in detail, with
a graphic interpretation of the dispensing and reading points and the calculation
made to obtain the absorbance.
Each of the above analysis modes may be ascending or descending.
If the test is ascending, the evolution of the absorbance increases depending on
the time. It has an ascending form.
If the test is descending, the evolution of the absorbance decreases depending on
the time. It has a descending form. To obtain positive absorbance values using
these calculation methods, the result is multiplied by -1.
Biosystems
186
16.2.1. Endpoint monoreagent
In endpoint reactions, once initiated the reaction lasts until it is balanced and then
the absorbance value remains stable. The absorbance reading is programmed at
this point. See Figure 112.
First reagent A is dispensed and the sample is dispensed in cycle 1, it is stirred and
the reaction commences. Once it is balanced, the reading is taken, L1. The change
in absorbance is directly proportional to the analyte concentration.
!"
f
ormul
a1-
ENG
formula2-ENG
A=ABSλL1main
(1)
If the reaction is bichromatic two readings are made in time L1. Each of the read-
ings is made with a different wavelength. The absorbance is the difference between
the two wavelengths.
187
User manual
formula2-ENG
A=ABSλL1main (3)
If the reaction is bichromatic two readings are made in time L1. Each of the read-
ings is made at a different wavelength. The absorbance is the difference between
the two wavelengths.
Biosystems
188
16.2.3. Differential switch
Differential tests make two readings, the first one before dispensing reagent B
and the second one after the end of the reaction. These tests are used to eliminate
potential turbidity effects in the sample, or eliminate the potential absorbance
levels of reagent A.
First reagent A is dispensed and the sample in cycle 1, in the next cycle it is stirred
and the reaction commences. Before dispensing reagent B, the L1 reading is taken.
Reagent B is dispensed in cycle 18, and stirred in the next cycle, and the second
part of the reaction commences. When the second reaction is balanced, reading
L2 is taken.
f
ormul
a5-
ENG
V M +V R1 (5)
A=ABSλL2main−(ABSλL1main∗ )
V M +V R1+V R2
189
User manual
First reagent A is dispensed and the sample is dispensed in cycle 1, it is stirred and
the reaction commences. Reading L1 is taken and after a few cycles reading L2 is
taken. The absorbance is the difference between the readings.
!
formula6
formula21-ENG
"
"
If the reaction is bichromatic, two readings are made at time L1 and two readings
at time L2. The absorbance is the difference between the two wavelengths at reach
reading time.
A= ABS L2main − ABS L2reference − ABS L1main− ABS L1reference (7)
Biosystems
190
The absorbance calculation may be monochromatic or bichromatic.
If the reaction is monochromatic, it is only measured at one wavelength and the
absorbance calculation is performed with the following formula:
If the reaction is bichromatic, two readings are made at time L1 and two readings
at time L2. The absorbance is the difference between the two wavelengths at reach
reading time.
A= ABS L2main − ABS L2reference − ABS L1main− ABS L1reference (9)
A={%ABS}_L2-{%ABS}_L1
# !"
191
User manual
If the correlation coefficient r < 0.9 then the programme says that the result of
the kinetic reaction is non-linear.
First reagent A is dispensed and the sample is dispensed in cycle 1, it is stirred
and the reaction commences. The analyser starts to take the readings from time
L1 to time L2.
The absorbance calculation is as follows:
main
(7)
A= [
ABS
min ]
!
!
Biosystems
192
formula9-ENG
!
main
A= [
ABS
min ] (7)
193
User manual
Calibrator The calibrator is a sample with the known concentration of the analyte to be de-
termined. It is a standard or reference material. To correctly measure the calibra-
tor blank absorbance, the same analysis mode must be used as the one used with
formula10-ENG
the samples.
If the ratio between the analyte absorbance and its concentration is linear, then
the calibration function is a line. So it will only be necessary to measure the blank
and a calibrator.
ABS
Acalibrator
Asample
Ablank
For the linear standard functions the blank absorbance ordinates is taken as the
source and the inverse of the factor as the slope.
The factor is calculated as follows:
C standard
F=
A standard − Ablank (8)
For calibration functions that are non-linear several known concentration calibra-
tors are used, approximating the curve with regression functions.
The following regression functions can be programmed:
Biosystems
194
ABS C4
Asample
C3
C2
C1
Ablank
Csample
CONC
195
User manual
formula23
RT
E x =E S ⋅log C (12)
nF
Biosystems
196
formula22
formula14
Symbol Description
Ex Electrode potential in the sample solution
ES Potential developed in standard conditions
RT/nF Constant, depending on the temperature
a Ion coefficient of activity measured in the solution
C Ion concentration measured in the solution
A comparative method is used for measuring. First, the ISE module measures the
potentials developed by the sample when it is positioned in the electrodes. Then
calibrator A for serum samples or calibrator B for urine samples is positioned in the
electrodes. The difference between both measurements for each ion is proportional
to the quotient logarithm between the ion concentration in the sample and in
the calibration solution. The proportionality factor (S) is calculated in a previous
calibration operation described below. Since the differences in potential of the ion
concentrations in the calibrating solution are known, the concentration of the ions
in a sample solution can be calculated, using the Nernst equation, and rewriting:
formula15-ENG
E −E =S⋅log
x
C
S
Cx
s
(13)
E x −E S
C x =C s · 10 S (14)
Symbol Description
Ex ISE voltage in the sample solution
ES ISE voltage in the calibrator solution
S Electrode slope calculated during the calibration process
Cx Ion concentration of the sample
CS Ion concentration of the calibrator solution
“S”, slope, is determined during the calibration using calibrators A and B, in which
the sodium, potassium and lithium levels are known.
When a two-point calibration is started, the slope is calculated based on the
difference between the reading of calibrator A and the reading of calibrator B.
Excessive shunt or noisy readings will be indicated and an error message will be
sent to the system.
The slope is defined as:
E B −E A
Slope=
E_x-E_S = S cdot log left(CC_x C_s right)
over (15)
log B
CA
C_x= C_s· 10^{{E_x-E_S}over S}
197
User manual
Symbol Description
CA Concentration of calibrator A in mmol/L
CB Concentration of calibrator B in mmol/L
EA Voltage measured in the ISE of calibrator A in mV
EB Voltage measured in the ISE of calibrator B in mV
The calibration slope value is affected by the temperature and by the ageing of the
electrodes. It is verified that the slope value is within certain limits.
16.6.1. Basis
The result obtained for a control material is compared with an admissible value
interval and a decision is taken:
• The result is within the interval: It is considered that the measuring procedure
maintains its accuracy within certain limits (it is stable) and the results of the
series are accepted.
• The result is outside the interval: It is considered that the measuring procedure
returns an error that is above tolerable levels and the results of the series are
rejected.
Biosystems
198
The dispersion of the results obtained is due to the imprecision of the measuring
procedure between series. This dispersion must have a normal distribution char-
acterised by the mean values and the standard deviation.
Therefore it is possible to establish a value interval with a known probability of
the result being included in that interval.
As it is required that the probability be high, it is common to select intervals be-
tween Xm ± 2s and Xm ± 3s. The selected criterion for establishing the admissible
value interval is a decision-based criterion or a control rule.
Internal control is based on the idea that it is not very likely that a result outside
the established limits will be obtained.
Control rules based on Gaussian statistics are usually represented by the expres-
sion Ans, where “A” is the number of control results and “ns” is the admissible
limit selected.
Different control results belonging to one control material or more than one may
also be used. Likewise the control results may have been obtained in one series or
in various consecutive series.
Rules that are more complex may be entered with various control results.
The ones most often used are the following:
22s Series rejected when 2 results are obtained that exceed 2s of the same type (posi-
tive or negative).
41s Series rejected if 4 results are obtained that exceed 1s of the same type.
10X Series rejected if 10 results are obtained on the same side of the mean.
R4s Series rejected if one result exceeds the +2s limit and the other exceeds the –2s
limit.
The rules for several control results may also give a clue as to the possible cause
of the increase in error. Rules 22s, 41s and 10X are particularly sensitive to systemic
error, whereas rule R4s is better at detecting increases in imprecisions.
Another interesting option is the combination of several rules in a logical or
algorithmic sequence. The best known combination is known as the Westgard
algorithm or rules for two control results.
In some cases it is not possible to make a statistical estimate of the dispersion of
results and apply control rules, because there are no accessible control materials
or because the measuring procedure is not used very often. In such situations it is
common to use a control material furnished by the supplier of the reagents or meas-
uring system, for which an admissible value interval is indicated (Manual mode).
199
User manual
• High likelihood of detecting important increases in error. The lower the value
interval of the control rule, the greater probability there will be of detecting
increases in error.
The idea is to have the lowest possible number of false alarms and guide error-
detection to increases that are considered important, based on the understanding
that smaller errors may occur (tolerable errors) without being detected.
Biosystems
200
17. Summary of workflow scenarios with
the LIS
This chapter describes the different scenarios defining interaction between the
BA200 and the information management software of a laboratory (LIS-Labora-
tory Information System). It describes the exchange of information between the
BA200 analyser and the LIS, such as for example in receiving work orders from
the laboratory for creating the worklist in the analyser and sending results from
the analyser to the LIS.
The BA200 implements two types of message or protocol message flows:
• The HL7 (Health Level 7) applied pursuant to the IHE (Integrating the
Healthcare Enterprise) recommendation
• The ASTM (American Society for Testing and Materials)
In this context the terminology used to describe data-transmission from the LIS
to the analyser is called Download and the data-transmission from the analyser
to the LIS is calledUpload.
The specimen is the content of each patient or control tube and may be one of the
types admitted (serum, urine, whole blood, etc). A patient may have two different
specimens, a serum one and a urine one. The tests indicated through a worklist
are performed on the specimen.
Transmission between the analyser and the LIS system is done through TCP/IP
connections.
• The TCP/IP for ASTM and HL7 is established when the system is initiated
and it must be permanently maintained provided the analyser is on. The
communication supports two setup modes: establishing the analyser as client
or as server.
• The HL7 also permits the transitory connection mode, which establishes two
connections at one time: When the BA200 starts a conversation a network
connection is established (a socket with an IP address and a port is opened) and
all the messages related to the conversation are sent and received by the socket.
When the LIS wants to initiate a conversation, another network connection
is initiated (another socket is opened with an IP address and a port) and all
the messages related to this conversation are sent and received through this
other socket.
201
User manual
Biosystems
202
The chain of actions is as follows:
1. Press the Query All button on the bar at the top. The analyser makes a generic
request for the worklist to the LIS.
2. The LIS sends all the work orders it has for the analyser. The programme
processes the orders and creates the worklist.
3. The user positions the sample specimens on the rotor and reads the barcodes.
4. The programme assigns to each specimen the information of the work order
programmed for the list.
5. The user can start the work session.
6. Once the list has been completed and the results obtained for each specimen,
the analyser sends them to the LIS. The sending of the results in real time is
done with the frequency configured by the user. (End of Test, End of Patient,
End of Work Session).
))See chapter 10.2.6 for the LIS operation configuration.
7. The LIS must send a cancellation of all orders not executed.
203
User manual
These LIS orders that have not yet been processed in the analyser are automatically
added to the next work LIS session by pressing the button Add orders to LIS. After
adding them to a new session the memorised session is automatically eliminated.
To eliminate Pending LIS orders, the LIS must send the respective Cancellation
messages.
In the LIS utilities screen requests from the LIS that have not been processed can
be eliminated.
17.4. Repetitions
The parties who can request test repetitions are established in the LIS configuration
screen. The options are the following:
• LIS: Patient tests can only be repeated from the LIS. Requests for repeti-
tions are launched during the clinical validation from the LIS manager. The
manual option of patient sample repetitions in the current results screen will
be blocked. Blanks, calibrators and controls may be repeated via the analyser.
• Analyser: Patient tests can only be repeated from the analyser. Requests are
launched for repetitions during the technical validation of the results. Repeti-
tion orders received from the LIS will be rejected.
• Both: All the repetition requests coming from the LIS or from the analyser
are accepted.
Description Cause
The user deletes requests The required specimen has not been received
accepted from the LIS Lack of reagent
pending execution by the
Other reasons
BA200
Description Cause
Lack of knowledge about the The cancelled test or sample type does not
type of patient sample or test exist in the BA200.
The test or sample type to be Execution of the test or sample type to be
cancelled has ended cancelled has already ended (the results have
already been obtained)
Biosystems
204
Description Cause
Request from the LIS for a The repetition mode in the analyser has only
non-permitted repetition been selected in the BA200.
Request from LIS to repeat It is not permitted to request repetitions of
an internal QC control internal controls from the LIS.
Description Cause
Unknown test or sample type The test and sample type identifier fields
are known by the BA200, but the test
was not programmed for the sample
type.
Internal control request for an An internal control was requested for a
erroneous test or sample type. calculated test or for an external test.
An internal control was requested for
a normal or ISE test and the quality
control parameters are not programmed
in the BA200.
Calculated test requires more than The calculated test is formed by tests
one sample type with different sample types.
Duplicate specimen The same specimen identifier was sent
for different patients
Duplicated request The same specimen identifier/test
identifier /sample type identifier was
already requested for the same patient
and their result has not yet been sent.
205
User manual
Biosystems
206
Menu Submenu Description
Programming Tests Programming of test
parameters
Calculated Tests Programming of calculated test
parameters
Contaminations Contamination programming
Profiles Profile programming
Calibrators Programming of calibrators
Controls Programming of controls
Patient data Patient data entry
ISE Tests ISE test programming
External tests External test programming
Programming export Export the parameters of the
text to an external file
Programming import Import the parameters of the
test trom an external file
207
User manual
Biosystems
208
Menu Submenu Description
Exit With Shutdown Option of exiting and switching
off the analyser
Without Shutdown Option of exiting the analyser
without switching it off
209