Measuring systems for the continuous and dynamic
measurement of quantities of liquids other than water
Published by:
WELMEC Secretariat
E-mail: [email protected]
Website: www.welmec.org
Page 2 of 32
CT-005, 2020, Measuring systems for the continuous and dynamic measurement of quantities of liquids
other than water, OIML R 117: 2019 – MID Annex I and VII (MI-005)
Foreword
1. The column “Comments” indicates when necessary the relevant text of OIML R 117 and
related explanations concerning the compliance with relevant MID requirement.
2. The column “Conclusion” gives the conclusion on the compliance between MID and OIML
R 117 for the relevant requirement.
The indication “Not Covered” means that the MID requirement is either not compatible with the
relevant OIML R 117 requirement or not included in OIML R 117.
The indication “Not relevant” means that the MID Annex I requirement is not applicable for
measuring systems for the continuous and dynamic measurement of quantities of liquids other than water.
The text in italic is an extract from the relevant clause of the OIML Recommendation.
Page 3 of 32
CT-005, 2020, Measuring systems for the continuous and dynamic measurement of quantities of liquids other than water, OIML R 117: 2019 – MID Annex I and VII
(MI-005)
ANNEX I
Page 4 of 32
CT-005, 2020, Measuring systems for the continuous and dynamic measurement of quantities of liquids other than water, OIML R 117: 2019 – MID Annex I and VII
(MI-005)
Page 5 of 32
CT-005, 2020, Measuring systems for the continuous and dynamic measurement of quantities of liquids other than water, OIML R 117: 2019 – MID Annex I and VII
(MI-005)
(b) The following influence quantities shall No mechanical shock test is specified considering the conditions of
be considered in relation with mechanical installation and use of such measuring systems. This test is seen as
environments: not relevant for these kind of measuring instruments.
- Vibration
- Mechanical shock
Page 6 of 32
CT-005, 2020, Measuring systems for the continuous and dynamic measurement of quantities of liquids other than water, OIML R 117: 2019 – MID Annex I and VII
(MI-005)
Page 7 of 32
CT-005, 2020, Measuring systems for the continuous and dynamic measurement of quantities of liquids other than water, OIML R 117: 2019 – MID Annex I and VII
(MI-005)
Page 8 of 32
CT-005, 2020, Measuring systems for the continuous and dynamic measurement of quantities of liquids other than water, OIML R 117: 2019 – MID Annex I and VII
(MI-005)
Page 9 of 32
CT-005, 2020, Measuring systems for the continuous and dynamic measurement of quantities of liquids other than water, OIML R 117: 2019 – MID Annex I and VII
(MI-005)
3 Repeatability 3.1.2.2 In OIML R 117 the requirement for repeatability is only stated for the Partially covered
meter and not for the complete measuring system (MI-005). This
The application of the same measurand
article is therefore partially covered by OIML R 117.
under the same conditions of
measurement shall result in the close
agreement of successive measurements.
The difference between the measurement
results shall be small when compared with
the MPE.
2.3.2
4 Discrimination and Sensitivity Covered
2.5.3
A measuring instrument shall be 3.1.2.4
sufficiently sensitive and the 3.1.3
discrimination threshold shall be 3.1.5.4
sufficiently low for the intended 3.1.6.1
measurement task. 3.1.7.1
3.1.8.1
3.1.9.1
3.2.1.4
3.3.3
3.4.1
3.4.8
3.6.8
3.9.1.3
5.1.8
5.5.1
Page 10 of 32
CT-005, 2020, Measuring systems for the continuous and dynamic measurement of quantities of liquids other than water, OIML R 117: 2019 – MID Annex I and VII
(MI-005)
5.9.1
5.11.3
5 Durability 3.1.2.3 Two different aspects have been considered: mechanical durability Covered
and electronic durability.
A measuring instrument shall be designed R 117-2: 5.4
to maintain an adequate stability of its “Endurance
metrological characteristics over a period
Capability of the measuring system to keep its performance
of time estimated by the manufacturer’s 4.1.3
characteristics over a period of use.”
instruction when in the environmental
A.1.5
conditions for which it is intended. “Durability for electronic devices
Capability of the electronic devices of a measuring system to keep
their performance characteristics over a period of use.”
“The requirements in 4.1.1 shall be met durably. For this purpose,
electronic measuring systems shall be provided with the checking
facilities specified in 4.3.”
6 Reliability 4 “Electronic measuring systems shall be designed and manufactured Partially covered
such that their metrological functions are safeguarded and their errors
A measuring instrument shall be designed A.1.5
do not exceed the maximum permissible errors as defined in 2.5
to reduce as far as possible the effect of a
under rated operating conditions.” (4.1.1)
defect that would lead to an inaccurate
measurement result, unless the presence Not covered for non-electronic devices (mechanical devices). The MID
of such a defect is obvious. requires in that case that the manufacturer covers this in his risk
analysis.
7 Suitability
7.1 A measuring instrument shall have no 2.9 – 2.18 No requirement is stated regarding the suitable timing of installation of Partially covered
feature likely to facilitate fraudulent use, downloaded software.
2.20
whereas possibilities for unintentional
Section 5 gives additional requirements for certain types of measuring
misuse shall be minimal. 5
systems and therefore takes into account relevant practical working
Annex A conditions.
Page 11 of 32
CT-005, 2020, Measuring systems for the continuous and dynamic measurement of quantities of liquids other than water, OIML R 117: 2019 – MID Annex I and VII
(MI-005)
7.2 A measuring instrument shall be suitable 2 Section 5 gives additional requirements for certain types of measuring Covered
for its intended use taking account of the systems and therefore takes into account relevant practical working
3
practical working conditions and shall not conditions.
require unreasonable demands of the 5
user in order to obtain a correct
Annex A
measurement result.
7.3 The errors of a utility measuring These measuring systems are not utility measuring instruments. Not relevant
instrument at flows or currents outside the
controlled range shall not be unduly
biased.
Page 12 of 32
CT-005, 2020, Measuring systems for the continuous and dynamic measurement of quantities of liquids other than water, OIML R 117: 2019 – MID Annex I and VII
(MI-005)
3.1.4 R 117-1 does not require that test procedures be included in the
7.6 A measuring instrument shall be designed Partially covered
3.1.10.1 operating manual.
so as to allow the control of the measuring
3.7.4 R 117-1 does not require that special equipment or software needed
tasks after the instrument has been
3.7.5 for control is part of the instrument.
placed on the market and put into use. If
3.7.6
necessary, special equipment or software
4.3.4
for this control shall be part of the
4.3.5
instrument. The test procedure shall be
5.4.3
described in the operation manual.
5.4.4
5.5.3
5.5.9
5.9.5
6.2.2.3
6.1.2.1
When a measuring instrument has Covered
6.1.3
associated software which provides other
A.1.1
functions besides the measuring function,
A.1.3
the software that is critical for the
A.1.4
metrological characteristics shall be
A.2.1
identifiable and shall not be inadmissibly
A.2.2
influenced by the associated software.
Page 13 of 32
CT-005, 2020, Measuring systems for the continuous and dynamic measurement of quantities of liquids other than water, OIML R 117: 2019 – MID Annex I and VII
(MI-005)
2.20.1
8.2 A hardware component that is critical for Only covered if protected by hardware seals. Partially covered
3.1.3
metrological characteristics shall be
designed so that it can be secured.
Security measures foreseen shall provide
for evidence of an intervention.
6.1.3
8.3 Software that is critical for metrological Evidence of an intervention is covered only for parameter and not Partially covered
2.20.2.1.5
characteristics shall be identified as such covered for other items such as software.
A.1.1
and shall be secured.
A.1.3
Software identification shall be easily
A.1.4
provided by the measuring instrument.
A.3
Evidence of an intervention shall be
available for a reasonable period of time.
2.20.1
8.4 Measurement data, software that is critical The protection against accidental or intentional corruption for Partially covered
3.5
for measurement characteristics and measurement data is only covered for the primary indication.
4.3.2
metrological important parameters stored
4.3.3.1 The protection against accidental or intentional corruption for
or transmitted shall be adequately
A.1.3 parameters is only covered if protected by hardware seals.
protected against accidental or intentional
A.1.4
corruption.
A.2.3
A.2.4
A.3
8.5 For utility measuring instruments the These measuring systems are not utility measuring instruments. Not relevant
display of the total quantity supplied or the
displays from which the total quantity
supplied can be derived, whole or partial
reference to which is the basis for
payment, shall not be able to be reset
during use.
Page 14 of 32
CT-005, 2020, Measuring systems for the continuous and dynamic measurement of quantities of liquids other than water, OIML R 117: 2019 – MID Annex I and VII
(MI-005)
9.1 A measuring instrument shall bear the 2.2.3 In the Directive 2014/32/EU article 8.6 and 10.3 shall be taken into Covered
following inscriptions: account.
2.19
(a) manufacturer’s name, registered trade
name or registered trade mark
(b) information in respect of its accuracy,
and, where applicable
(c) information in respect of the conditions
of use
(d) measuring capacity
(e) measuring range
(f) identity marking
(g) number of EU-type examination
certificate or the EU design examination
certificate
(h) information whether or not additional
devices providing metrological results
comply with the provisions of this
Directive on legal metrological control.
Page 15 of 32
CT-005, 2020, Measuring systems for the continuous and dynamic measurement of quantities of liquids other than water, OIML R 117: 2019 – MID Annex I and VII
(MI-005)
makes this unnecessary. Information shall conditions for compatibility with interfaces, sub-assemblies or
be easily understandable and shall measuring instruments.
include where relevant:
(a) rated operating conditions
(b) mechanical and electromagnetic
environment classes
(c) the upper and lower temperature limit,
whether condensation is possible or not,
open or closed location
(d) instructions for installation,
maintenance, repairs, permissible
adjustments
(e) instructions for correct operation and
any special conditions use
(f) conditions for compatibility with
interfaces, sub-assemblies or measuring
instruments.
9.4 Groups of identical measuring instruments OIML R 117:2019 does not require that an instruction manual shall be Not covered
used in the same location or used for on location of the measuring instrument.
utility measurements do not necessarily
require individual instruction manuals.
3.2.1.2 “The scale interval shall be in the form 1 × 10n, 2 × 10n or 5 × 10n
9.5 Unless specified otherwise in an Covered
3.2.2.1 authorized units of quantity, where n is a positive or negative whole
instrument-specific annex, the scale
3.3.3 number, or zero.”
interval for a measured value shall be in
the form 1x10n, 2x10n, or 5x10n, where n 2.9.1 “The name of the unit or its symbol shall appear in the immediate
is any integer or zero. The unit of vicinity of the indication. For mass, according to the case, the name of
measurement or its symbol shall be the unit or its symbol shall be accompanied by the term “mass” (actual
shown close to the numerical value. mass) or “conventional mass” (comparison to weights).”
Page 16 of 32
CT-005, 2020, Measuring systems for the continuous and dynamic measurement of quantities of liquids other than water, OIML R 117: 2019 – MID Annex I and VII
(MI-005)
9.6 A material measure shall be marked with Measuring systems are not material measure. Not relevant
a nominal value or a scale, accompanied
by the unit of measurement used.
9.7 The units of measurement used and their 2.9.1 Note: The unit "cubic decimetre" is not allowed according to MI-005, Covered
symbols shall be in accordance with the art. 8.
provisions of Community legislation on
units of measurement and their symbols.
9.8 All marks and inscriptions required under 2.19.1 Only covered for non-transferable of marks and inscriptions. Partially covered
any requirement shall be clear, non-
erasable, unambiguous and non-
transferable.
10 Indication of result
T.m.2
10.1 Indication of the result shall be by means Note: Indication of result by means of only a hard copy is not allowed Covered
T.m.3
of a display or a hard copy. according to definitions of a meter in Annex VII (MI-005) and OIML
2.1
R 117-1 paragraph 2.1 and definitions T.m.2 (Measuring system) and
2.9.2
T.m.3 (Meter).
3.2
3.4
3.5
Page 17 of 32
CT-005, 2020, Measuring systems for the continuous and dynamic measurement of quantities of liquids other than water, OIML R 117: 2019 – MID Annex I and VII
(MI-005)
2.2.3 Covered
10.2 The indication of any result shall be clear
2.9
and unambiguous and accompanied by
2.19.4
such marks and inscriptions necessary to
3.2.1.1
inform the user of the significance of the
3.2.1.3
result. Easy reading of the presented
3.3
result shall be permitted under normal
3.4
conditions of use. Additional indications
5.1.6
may be shown provided they cannot be
5.1.12
confused with the metrologically
5.5.1
controlled indications.
5.9.1
5.11.3
A 2.2
3.4.2
10.3 In the case of hard copy the print or A non-erasable printout is not covered. Partially covered
3.4.3
record shall also be easily legible and
3.5.2
non-erasable.
3.5.3
3.5.4
3.5.5
A.2.3
A.2.4
A.2.5
A.2.6
2.2.3 Partially covered
10.4 A measuring instrument for direct sales The presentation to both parties involved is only covered for
3.9
trading transactions shall be designed to measuring systems with a self-service device.
2.21
present the measurement result to both
parties in the transaction when installed
as intended. When critical in case of direct
sales, any ticket provided to the consumer
by an ancillary device not complying with
the appropriate requirements of this
Directive shall bear appropriate restrictive
information.
Page 18 of 32
CT-005, 2020, Measuring systems for the continuous and dynamic measurement of quantities of liquids other than water, OIML R 117: 2019 – MID Annex I and VII
(MI-005)
Not relevant
10.5 Whether or not a measuring instrument
intended for utility measurement purposes
can be remotely read it shall in any case
be fitted with a metrologically controlled
display accessible without tools to the
customer. The reading of this display is
the measurement result that serves as the
basis for the price to pay.
Page 19 of 32
CT-005, 2020, Measuring systems for the continuous and dynamic measurement of quantities of liquids other than water, OIML R 117: 2019 – MID Annex I and VII
(MI-005)
6.1
12 Conformity evaluation Covered
6.2
A measuring instrument shall be designed A.1
so as to allow ready evaluation of its A.3
conformity with the appropriate A.4
requirements of this Directive.
Page 20 of 32
CT-005, 2020, Measuring systems for the continuous and dynamic measurement of quantities of liquids other than water, OIML R 117: 2019 – MID Annex I and VII
(MI-005)
1.2
1.2 The properties of the liquid to be Covered
2.3.1
measured by the instrument by specifying
the name or type of the liquid or its 3.1.1.1
relevant characteristics, for example:
— Temperature range;
— Pressure range;
— Density range;
— Viscosity range.
Page 21 of 32
CT-005, 2020, Measuring systems for the continuous and dynamic measurement of quantities of liquids other than water, OIML R 117: 2019 – MID Annex I and VII
(MI-005)
2.1 For quantities equal to or greater than 2 2.5.1 OIML R 117-1 does not specify an accuracy class 2,5. Partially covered
litres the MPE on indications is: 6.1.1 LNG is considered as a cryogenic fluid and therefore the accuracy
class for cryogenic fluids given in the MID could also be valid for LNG
(accuracy class 2,5). Where OIML R 117 does not recognize accuracy
class 2,5 and only accuracy class 1,5 is stated for LNG.
2.2 For quantities less than two litres the MPE 2.5.2 Covered
on indications is:
Page 22 of 32
CT-005, 2020, Measuring systems for the continuous and dynamic measurement of quantities of liquids other than water, OIML R 117: 2019 – MID Annex I and VII
(MI-005)
Page 23 of 32
CT-005, 2020, Measuring systems for the continuous and dynamic measurement of quantities of liquids other than water, OIML R 117: 2019 – MID Annex I and VII
(MI-005)
(a) Calculator
MPEs on quantities of liquid indications
applicable to calculation, positive or
negative, are equal to one-tenth of the
MPEs as defined in line A of Table 2.
Page 24 of 32
CT-005, 2020, Measuring systems for the continuous and dynamic measurement of quantities of liquids other than water, OIML R 117: 2019 – MID Annex I and VII
(MI-005)
2.7.2 Values defined in OIML R 117-1 Table 5.1 are greater than those
(c) Accuracy for calculating function Partially covered
specified in the MID since a quadratic sum has been considered to
The MPE for the calculation of each define the relevant MPEs.
characteristic quantity of the liquid,
positive or negative, is equal to two fifths OIML R 117-1 does not specify an accuracy class 2,5.
of the value fixed in (b).
Page 25 of 32
CT-005, 2020, Measuring systems for the continuous and dynamic measurement of quantities of liquids other than water, OIML R 117: 2019 – MID Annex I and VII
(MI-005)
Page 26 of 32
CT-005, 2020, Measuring systems for the continuous and dynamic measurement of quantities of liquids other than water, OIML R 117: 2019 – MID Annex I and VII
(MI-005)
5 Suitability
Page 27 of 32
CT-005, 2020, Measuring systems for the continuous and dynamic measurement of quantities of liquids other than water, OIML R 117: 2019 – MID Annex I and VII
(MI-005)
Page 28 of 32
CT-005, 2020, Measuring systems for the continuous and dynamic measurement of quantities of liquids other than water, OIML R 117: 2019 – MID Annex I and VII
(MI-005)
2.21
5.4.2 The display of the quantity on which the The presentation to both parties involved is only covered for Partially covered
3.9.1.7
transaction is based shall be permanent measuring systems with a self-service device.
3.9.2.1.1
until all parties in the transaction have
3.9.2.1.3
accepted the measurement result.
3.9.3.1.1
4.1.2
5.4.3 Measuring systems for direct sales shall Covered
be interruptible. 5.1.10
5.5.1
5.9.1
5.11.3
Page 29 of 32
CT-005, 2020, Measuring systems for the continuous and dynamic measurement of quantities of liquids other than water, OIML R 117: 2019 – MID Annex I and VII
(MI-005)
Page 30 of 32
CT-005, 2020, Measuring systems for the continuous and dynamic measurement of quantities of liquids other than water, OIML R 117: 2019 – MID Annex I and VII
(MI-005)
Page 31 of 32
CT-005, 2020, Measuring systems for the continuous and dynamic measurement of quantities of liquids other than water, OIML R 117: 2019 – MID Annex I and VII
(MI-005)
7 Putting into use 2.4 OIML R 117-1 does not specify an accuracy class 2,5. Partially covered
6.2 LNG is considered as a cryogenic fluid and therefore the accuracy
class for cryogenic fluids given in the MID could also be valid for LNG
(accuracy class 2,5). Where OIML R 117 does not recognize accuracy
class 2,5 and only accuracy class 1,5 is stated for LNG.
8 Units of measurement 2.9.1 Note: The unit "cubic decimetre" is also allowed in OIML R 117-1. Covered
The metered quantity shall be displayed in
millilitres, cubic centimetres, litres, cubic
metres, grams, kilograms or tonnes.