GP-100 Service Manual. Vers1.4.3.edoc
GP-100 Service Manual. Vers1.4.3.edoc
This symbol means that the labeled item is hot while the
instrument is in use. Don’t touch the labeled item as you could be
scalded.
The symbol is labeled on the lamp support of optic system.
This means that the labeled item could lead to personal injury
and/or damage to the analyzer.
The symbol is labeled beside the power outlet and some external
interface.
General
Before you start installing and working with the analyzer, you should read the safety
precautions and regulations shown in this chapter.
Operator Qualification
Please note that the operation with Chemistry analyser should be carried out only by the
doctor or clinical inspector who has undergone necessary training provided by the sales
agent.
To install, maintain and repair the instrument, a service technician has to be trained on the
instrument by the manufacturer or their representative. A service technician is also expected
to be familiar with the normal operation of the instrument as described in the User’s manual
and the special operations as described in the service manual.
Electrical
To prevent the risk of electrical shock and/or damage to the instrument Operator should not
open the white cover of the instrument. Only authorized personnel, for example, service
technicians, may open the instrument to perform maintenance or repair.
Touching the main board when the power is on may cause severe injury or death. Any
problem, please ask for helps from your supplier.
Mechanical
There is no risk presented by the mechanical parts of the instrument when the instrument is
closed. If the covers of the instruments are removed, mechanical parts could cause personal
injury or the instrument may be damaged if the following advice is not being considered: DO
NOT wear loose garments or jewellery that could catch in mechanisms. DO NOT put your
fingers/hands into the pathway or any part while the analyzer is in operation. DO NOT attempt
mechanical repair unless the instrument is not in operation or OFF.
Lamp
The source lamp becomes extremely hot during operation; never touch the lamp when it is on!
Look directly into the light path of the lamp may cause eyes damage, too.
Chemical
The operator is responsible for taking all necessary precautions against hazards associated
with the use of clinical laboratory chemicals. Specific recommendations for each reagent
used with the analyzer are normally found on the manufacturer’s package inserts or the on
the product information sheets for each chemical. Wipe up any reagent spillage on the
instrument immediately.
Biohazardous Materials
As with all vitro diagnostic equipment, patient samples and serum-based quality control (QC)
products that are assayed on this system, as well as all waste from the waste container,
should be treated as the potentially biohasardous. All materials and mechanical components
associated with the sampling and waste system should be handled according to your facility’s
biohazard procedure. Use the personal protective equipment recommended by your facility
when handling any of these components. Detailed recommendations:
- Samples
Treat all samples as potentially biological pollution and infectious. If any sample is spit on the
instrument, utilize the correct personal protective equipment (PPE- gloves, lab coat, etc...),
wipe it up immediately and clean the contaminated with a disinfectant.
Avoid direct contact with waste solution and/or solid waste. Both should be handled as
potentially biohazardous.
Dispose of waste solution and/or solid waste according to the relevant governmental
regulations.
Consult the reagent manufacturer for information on the concentrations of heavy metals and
other toxic constituents in each reagent.
- Biohazardous parts
Avoid direct contact with the sipper tubing and all parts of the sample flow path. Treat these
areas as potentially biohazardous and /or infectious
Avoid direct body-contact with reagents. Direct body-contact may result in irritation or
damage to your skin. Refer to the manufacturer’s reagent kit box and package inserts, or
product information sheets for specific instructions.
Avoid direct body-contact with cleaning solution. Direct body-contact amy result in skin
irritation or damage. Refer to the manufacturer’s kit box and package inserts, or product
information sheets for specific instructions.
ATTENTION! Cautions
Additional Precautions
- Flammables
For proper use of the instrument, measure control samples and monitor the instrument during
the operation.
Treat all reagents according to the manufacturer’s recommendations. Refer to the reagent kit
box and package inserts, or product information sheets for specific instructions.
Make sure that the sample/reagent mixture does not contain any blood clots, dust or other
insoluble contaminants. If insoluble contaminants are contained in the sample, correct
measuring values may not be obtained.
- Application
The instrument is designed for clinical chemistry test analysis using water-soluble samples
and reagents.
Please note that other types of analysis may not be applicable to the instrument.
During operation and maintenance of the instrument, proceed according to the instructors
and do not touch any parts of the instrument other than those specified.
Never leave a Reagents/sample mixture in the flow-cell for longer than necessary. Always
clean the flow-cell after a batch of measurement and keep the flow-cell filled with distilled
water when not in use.
Avoid touching the mechanism, such as the sipper mechanism inside the instrument, while
the instrument is operating. This may cause operation stop or damage the instrument.
1.1 Foreword
⚫ 47 of the most commonly assays are stored in the flashing RAM of the GP-100,
which enables user can start working right away with the instrument. 130 of
assays can be programmed.
⚫ All original programs setting of tested items can be revised by user.
⚫ It has double coloring system: One is flow-cell system and the other is cuvette
system.
⚫ 6 Analytical Modes:
Endpoint
Two-point
Kinetics
Dichromatic
Absorbency
Multi-Calibration
⚫ It can analyze samples of blood serum, blood plasma, whole blood and urine
etc.
⚫ It has no restriction of reagent.
⚫ The user could connect the unit to the PC of clinic lab by RS232 interface.
⚫ Either built-in printer or external printer can be selected.
⚫ The external standard (XT/AT) keyboard, mouse.
⚫ Power-off protection, real-time data saving.
Precision≤±0.5℃,amplitude of fluctuation≤±0.4℃
Dimension 460mm(L)×330mm(W)×190mm(H)
Power 80VA
Fuse T3.15AL250V,Φ5×20
Unpack the instrument and remove the materials used for delivery. Keep the packing
boxes and materials for repacking the instrument in the future.
1) Take the instrument out from the boxes.
2) Take away the packing materials; take out the instrument from the plastic
packing bags.
3) Check the articles in the packing box and confirm that the following articles are
contained::
⚫ GP-100 mainframe
⚫ User’s manual
⚫ Packing list
⚫ Repair guarantee certificate of the agent
⚫ Power cable
⚫ Product approved certificate
The instrument should not be directly exposed to the sun.. The working table selected shall
be smooth in surface and has enough space to mount GP-100. The front edge of instrument
shall be close to the edge of working table. The working table is prevented to have great
vibration (e.g. centrifuge is placed on the table).
To ensure the normal operation of the instrument, it is forbidden to mount the instrument
in the following places:
⚫ a.c.220V±22 V
⚫ 50Hz±1Hz
⚫ 80VA
1) Insert one end of power source line to the power source socket of the
instrument.
2) Insert the other end of power source line to the alternating current power source
socket.
As indicated in the following drawing, move the LCD display screen to an appropriate
position deemed by you.(beware to impose proper strength).
Figure 2.1.1
As indicated in the following drawing, take away the cover of flow-cell (beware proper
strength is used)
Figure 2.1.2
Figure 2.2.1
Figure 2.2.2
1. Power Switch
2. Fuse
3. AC power connector
4. Fan
5. Pump
6. Brightness
2.2.3 Keyboard
Figure 2.2.3
2) Rinse On, Off: Pressing this control will switch on the sipper pump
in a continuous mode, press the key again will stop
the pumping action at the end of a complete cycle.
3) Feed Paper: Control the printer paper for the embedded printer.
7) Right: This key will advance the cursor to the next position.
Note: When opening the cover of flow-cell or inner cover of light lamp,
press the two sides of cover so that it deform innerward, and then
open it according to the direction of the following drawing.
Figure 2.2.4
Incubation
Colorimetric system
Figure 2.2.5
Figure 3.1.1
Figure 3.2.1
Since the test sample is different and have different density, it can offer different
3.3.1 Function
When you press aspirate button in the test, system drives the sipper pump to rotate
and enable the tested liquid in pipeline extrude to liquid drainage port. So the tested
liquid is absorbed into the flow-cell. When sample volume set by measuring item is
absorbed, the system stops the sipper pump motor. Liquid to be tested is in the
3.3.2 Structure
Figure 3.3.1
3.4.1 Function
The optical structure of GP-100 is, as the below diagram, spectrums structure after
going through filter. After lightening tungsten halogen lamp, radiate light with filament
as center is given out. The radiate light is converted to parallel ray after going through
convex lens. The protection glass plays the role of protecting convex lens. After the
parallel ray pass through the flow-cell with sample liquid to be tested, a part of light
energy is absorbed by liquid. Light that shoot up is called shoot-up light. The shoot-up
light is changed to single color light after going through single color filter. The single
color ray shoots on the photoelectric receiver. The photoelectric receiver transfer the
light signal into electric signal and input it to front board for processing. The color of
filter and color of tested liquid shall be additive complementary colors.
3.4.2 Structure
①.Lamp 6V/10W
②.Protection glass
③.Convex lens
④.Flow cell
⑤.Filter
⑥.Photometer Detector
If the user meets some problems and needs to open the instrument, you can do as
follow:
1) Turn over the instrument.
2) From bottom of the instrument you can see seven plastic pillars with screw hole.
Uses a screwdriver unscrew these seven screws. (Figure 8.1)
3) Turn the instrument back to normal position.
4) Set the LCD screen to vertical position. (Figure 8.2)
5) Lift the top cover of instrument gradually, and open the front side about 3 inches
first. Do not use too much force, because there are cables between top and foot
cover.
6) After open the machine, you can see the inside distribute of top and foot cover like
figuer8.3
7) Find the LCD connector on the main board, and open it carefully.
8) Then turn over the top cover, disconnect 5 other cables
Figure 4.1.1
Top cover
Foot cover
Front side
Figure 4.1.3
1. Turn off the machine, unscrew the bolts on base of instrument; take out the upper
cover of instrument.
2. Pull out plugs①、②、③.
3. Press firmly the flip plate④ on both end of switch, push outwards the entire switch
seat and remove the switch seat.
4. Pull out tie-in of power source line⑤、⑥ with switch. Press firmly the flip plate and
push outwards the switch.
5. Take a fine quality switch.
6. Connect the tie-in of power source line⑤、⑥and switch.
7. Press the switch socket outwards into the base of instrument, the power source
line socket shall face the fan.
8. Connected power source line①、②、③.
9. Place the upper cover, firmly screw bolts.
Note:
Don’t connect line wrong. Various connection lines do not press on the bolt port to
prevent damage of connection line.
Adjust Screw
Clip Button
Pump Tube
Motor
Figure 4.3.1
Clip Bar
Figure 4.3.2
Clip Button
Pump Tube
Figure 4.3.3
3. Take out the pump tube, tear the belt, then the pump tube can be removed
from the white cuff.
Cable Tie
White Connector
Figure 4.3.4
4. Replace a new pump tube, tie the belt, and operate reservedly so that the
replacement can be completed.
1. Open the cover of the flow-cell and lamp cover; you will see the flow-cell located in the
incubator.
Flow-cell
Figure 4.4.1
Figure 4.4.2
3. Replace with a new flow-cell, following the reverse procedure. Be careful not to fold the
sampling tube.
NOTE:
The flow-cell should be installed in right direction. The side in the following left
picture should be facing the sampling tube; the other side should be facing the drain
tube.
Figure 4.4.3
Turn off the instrument, open the cover of lamp, pull out power supply wire of lamp,
unscrew the screw ③ and ② then unscrews ①,take out the lamp from the holder.
The structure of lamp is showing in the following figure:
②
A
①
B
③
S
Figure 4.5.1
1. Replace a new lamp and install it to the holder, screw the screw①,but do not screw it
too firm, connect the lamp wire to power supply wire.
2. Screw the screws ②、③ ,but do not screw it too firm
3. Turn on the instrument and enter the main menu, select“ABS”; enter setting window,
set Wavelength as 340nm and Reading as 200 second,interval time 1 second. Then
aspirate distilled water to test.
Figure 4.5.2
4. Move the lamp forwards and backwards to maximize the tested AD value. The value is
in the small bracket (show as 2500…), which is displayed on the screen.
5. Use a screwdriver to screw the screws ① (not too firm).
6. Adjust screws ②、③,which enable the tested AD value is about 3500(2500—3900
is also permissible), then to fasten screw ①.
Figure 4.5.3
1. Open the upper cover of GP-100,beware of the connection lines between upper cover
and main board.
Cassette
Fix Screw
Figure 4.6.1
2. Unscrew 2 fixed screws of filter cassette, and then take out the cassette (Figure 4.6.2).
At the same time, unfasten the 3 screws in the filter wheel (Figure 4.6.3).
Figure 4.6.2
3. Screw out 2 screws and dismantle filter wheel motor (Figure 4.6.4).
4. Take out the filter wheel as showed in Figure 4.6.5. Filter is installed in the filter wheel
follow certain sequence (Figure 4.6.6).
340nm
546nm (A gap beside)
5. Select replacement needed filter, take out the fixed ring and filter carefully (Figure
4.6.7).
Filter
Fix Ring
Figure 4.6.7
6. Replace a new filter with same wavelength, fix it with cooper ring.
7. Reverse the above dismantling procedure, and the filter replacement is completed.
The function of photometer detector is to change the variable light signal into the
electric signal, and then calculated and process. It is the most important component in
the light path system.
Photometer Detector
Figure 4.7.1
2. Unscrew 3 screws with red gasket and draw outwards the photometer detector
board (Figure 4.7.2).
Figure 4.7.2
3. Replace a new photometer detector, connect power source line, screw the screws,
add cassette and equip with cassette.
Figure 4.7.3
4. Connect the other end of power source line in J5 of front board. The replacement
is finished. The signal receiver can work normally.
1. Open the upper cover of the instrument; you can see two boards (big and small) at
the bottom. The small board is the front board (Figure 4.8.1).
Figure 4.8.1
2. Unscrew four screws in the front board, pull out all plugs on the board carefully and
connect them to new board, operate reversely to finish the replacement.
Note:
There is a layer of glue in the plug, scrape the glue first when pull out
the plug; if you are not familiar with positions of plug, write down the
position first, in case of plug wrong.
1. Open the upper cover of the instrument; you can see two boards (big and small) at
the bottom. The big board is the main board (Figure 4.9.1).
Figure 4.9.1
2. Unscrew four screws in the front board, pull out all plugs on the board carefully and
connect them to new board, operate reversely to finish the replacement.
1. Open the upper cover of the instrument; you can see the power supply assembly
as follow.
Screw
Screw
Figure 4.10.1
2. Pull out all lines connected in the front board and main board; unscrew two fixed
screws in figure 4.10.1, and replace with a new board and operate reversely to
finish the replacement.
Figure 4.11.1
Figure 4.11.2
Data Cable
Power Lines
Figure 4.11.3
5. Cut the 2 power lines showed in Figure 4.11.3, pull out data cable and unscrew 4
fixed screws. The LCD can be taken out.
6. Replace with a same model LCD; operate reversely to finish the replacement.
Figure 4.12.1
1. Open the upper cover and the optical assembly can be seen as follow (Figure
4.13.1).
Incubator
Peltier
Thermal discharge
Slices and Fan
Figure 4.13.1
2. Changing the peltier, you need to dismantle the whole optical assembly. Turn over
the instrument and unscrew 4 screws as showed in Figure 4.13.2 to dismantle
optical assembly.
Figure 4.13.2
3. After you take out the optical assembly, unscrew 2 screws showed in Figure 4.13.3,
you will find the peltier (Figure 4.13.4).
4. Replace a new peltier with same model and operate reversely to finish the
replacement.
NOTE:
While changing the peltier, pay attention to the direction of two lines
(black and red) and do not connect wrong. The peltier should be lain
on a layer of silicone evenly.
1. Open upper cover of GP-100 and unscrew 4 screws of optical assembly at the
bottom of instrument and take out the optical assembly.
2. Unscrew 1 screw showed in Figure 4.14.1 to take out the incubator. The
temperature sensor is fixed on it.
Figure 4.14.1
3. Cut the cable tie of temperature sensor and pull it out (Figure 4.14.2).
Temperature
Sensor
Figure 4.14.2
4. Replace a new temperature sensor with same model and operate reversely to
finish the replacement.
NOTE:
Before replace the temperature sensor back, make sure there is
silicone in the hole for good heat conduction.
1. Open upper cover of GP-100 and unscrew 4 screws of optical assembly at the
bottom of instrument and take it out.
2. Unscrew 2 screws at the bottom of optical assembly (Figure 4.15.1).
Figure 4.15.1
3. Replace a new peltier fan with same model and operate reversely to finish the
replacement.
1. Open upper cover and turn it over, you can see the printer driver board (Figure
4.16.1).
2. Pull the connected lines out carefully and unscrew the fixed screws to take out the
board.
3. Replace a new printer driver board with same model and operate reversely to
finish the replacement.
1. Open upper cover, you can see the internal printer (Figure 4.17.1).
Internal Printer
2. Pull the connected lines out carefully and unscrew the fixed screws to take out the
internal printer.
3. Replace a new printer with same model and operate reversely to finish the
replacement.
NOTE:
There are two kinds of internal printer. Pinter in Figure 4.17.1 has two
fixed screws, the other only has 1 fixed screw.
The cleanness of flow-cell must be kept to ensure the accuracy and reliability of test
results.
1. Outer cleaning
a、Place flow-cell according to requirements.
b、If the outer flow-cell is polluted, use soft cloth with alcohol to clean it softly.
2. Inside cleaning
a) Put distilled water under the aspirate tube, press RINSE key to start continuous
cleaning function. Press RINSE key again to stop cleaning. Generally, the cleaning
takes half a minute.
b) Detergent, dilute HCL (0.1N), dilute Tween 20 liquid (2-3drop/L) can be used to
clean flow-cell. Push RINSE key to aspirate detergent, press RINSE key again to
stop. Keep the detergent stay in flow-cell for 5 minutes, and then clean the
flow-cell with distilled water for 1 minute. If it cannot clean well once, detergent
should be used to clean twice.
The flow-cell shall be cleaned in the following situation:
⚫ The water blank value’s deviation is too large when power on the
instrument.
⚫ Change test item.
⚫ Power off the instrument.
Warning:Do not let reagent and other pollutant in the flow-cell for a long time.
Malfunction Solution
Time pop up randomly, date can not 1. Battery power is used out, replace another battery.
be saved after change.
ABS value is negative 1. Flow-cell is dirty or has bubble inside, clean it.
2. Air is aspirated when first turned on, turn it off and
power on again to aspirate.
1. The flow-cell has not been inserted to bottom.
“Absorbency deviation is too big, 2. Liquid has been aspirated or air is aspirated,
retry?” pop up after power on the aspirate distilled water again.
instrument. 3. Light intensity of lamp is low, change the lamp.
4. Flow-cell is dirty and easy to get bubble or flow-cell
is damaged, clean or replaces flow-cell.
5. Photometer detector is damaged, replace it.
Note:
This manual may contain other ownership declaration and copyright
information. Its articles shall be observed and abided by. The
information in the manual may have inaccurate parts in technology
and printing error.
The manual can be revised or upgraded without additional notice
Greiner also can improve/revise the product/program described in the
material at any time without additional notice.
_______________________________________________________________________________