This document provides specifications for an instrument housing type XD-I, including: 1. Dimensions and material details for the body, cover, and various threaded connections. 2. Service temperature ranges depending on the seal material. 3. Coating and seal material options. 4. Protection concept details including type of protection, ATEX marking, permitted use, approval body and certificate number.
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
68 views

Data Sheet - IECEx

This document provides specifications for an instrument housing type XD-I, including: 1. Dimensions and material details for the body, cover, and various threaded connections. 2. Service temperature ranges depending on the seal material. 3. Coating and seal material options. 4. Protection concept details including type of protection, ATEX marking, permitted use, approval body and certificate number.
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 3

31.03.

2016

Limatherm S.A.

INSTRUMENT HOUSING – type XD-I


125.5
Ø110
48*
M100X2 6H/6g
Ø90 D3

A-A

l3
LOCKING EARTH
SCREW TERMINAL
PROTECTION
80 max

* SCREW

D1
~108

127

145
115
48

58
Ø2x4
A M4
D1

A (4x)
l1
36

26

l2
6

40 D2
* - other dimensions possible

HT W HEN C Second caution inscription


TIG IR
ER CU - in french language standard
IT
OV - german, italian, spanish on request
C

- customer logo on request


AL
EP

IV E
KE

NE
O

I PA
NS SO
TE UVR
IR SOUS

25
8

customer logo

FOR ALL VERSIONS of XD-I


Material of body and cover: aluminium pressure die-casting
Service temperature: Tserv = -40 ÷ 100°C (85°C with window) – oil proof rubber
Tserv = -40 ÷ 100°C (85°C with window) – silicone rubber
Tserv = -20 ÷ 200°C (85°C with window) – FKM rubber
Coating: yellow chromating and chemically resistant paint (outside only)
SEAL
THREAD D3 l3 [mm] THREAD D2 l2 [mm] THREAD D1 l1 [mm] COATING
MATERIAL
M20×1.5 SE
M20×1.5 M20×1.5
M24×1.5 3
M24×1.5 20 M24×1.5 20 20 Cal
M25×1.5
M25×1.5 M25×1.5
M27×2 4 SPU
½ NPT mod ½ NPT mod ½ NPT mod PPU
16 16 16
¾ NPT mod ¾ NPT mod ¾ NPT mod 5
P
PD PD PD

PROTECTION CONCEPT
Type of protection
Protection Protection
IECEx Permitted use Approval Body Certificate No.
method ATEX marking Principe
marking
Flameproof
EN 60079-0
EN 60079-1 1026 II2G Ex db IIC Gb Ex db IIC Gb FTZU 03 ATEX 0207U Contain the
Zone 1, Zone 2 FTZU
EN 60079-31 explosion and
Zone 21, Zone 22 Czech Republic IECEx FTZU 12.0017U
IEC 60079-0 1026 II2D Ex tb IIIC Db Ex tb IIIC Db quench the flame
IEC 60079-1
IEC 60079-31

34-600 LIMANOWA, ul. Tarnowska 1; tel. +48 18 337 98 20 fax +48 18 337 98 25

www.limatherm.com
31.03.2016

Limatherm S.A.

ORDERING

XD-I... – . - .. / .. - .. / .. - .. / .. - . ..

Type of instrument housing

Version of instrument housing


= standard
H = high cover
win = cover with window
Lwin = low cover with window
Hwin = high cover with window
C = without lugs
CH = without lugs, high cover
Cwin = without lugs, cover with window
CLwin = without lugs, low cover with window
CHwin = without lugs, high cover with window

Material of seal
3 = TPE rubber 4 = VMQ rubber 5 = FKM rubber

Symbol of conduit thread D3 - D2


Standard: On request:
M2 = M20×1.5 M24 = M24×1.5
N2 = ½ NPTmod – modified acc. to OIT-17/03 M25 = M25×1.5
N3 = ¾ NPTmod – modified acc. to OIT-17/03 PD = plugged
Notice:
Type: “size of thread / PD” if thread is plugged, e.g. M2/PD, etc.

Symbol of process thread D1


Standard: On request:
M25 = M25×1.5 M2 = M20×1.5
N3 = ¾ NPTmod – modified acc. to OIT-17/03 M24 = M24×1.5
M27 = M27×2
N2 = ½ NPTmod – modified acc. to OIT-17/03
PD = plugged
Notice:
Type: “size of thread / PD” if thread is plugged ex. M2/PD, etc.

Symbol of paint type


SE – spray epoxy – standard; SPU – spray polyurethane (on se layer); C – creodur; PPU – pulyurethane powder; P – polyester powder

Symbol of paint colour


sh = RAL 9006 shine – standard; yc = RAL 1007 yellow chromic; ys = RAL 1003 yellow signaling; bs = RAL 5005 blue signaling;
sb = RAL 5015 sky-blue; sg = RAL 7032 silicon-grey; gr = RAL 7035 grey; ag = RAL 9002 ashen grey-white; gs = RAL 9007 grey-silver;
cw = RAL 9010 cream-white; al = alu natural colour (for C paint type only)

STANDARD PRODUCTS: PROTECTION DEGREE


IP D3 D2 D1
XD-I4-D3-D2-D1-SEsh plugged – IP 68 plugged – IP 68 sensor connection
54 – thread without
cable gland IP66 cable gland IP66 sealing – IP 54
Sample:
closed – IP 68 closed – IP 68 sensor connection
XD-I4-M2-M2-M25-SEsh 68 – sealed thread –
cable gland IP68 cable gland IP68
XD-Iwin-4-N3/PD-N3-N3-SEsh IP 68
For more details see technical information

34-600 LIMANOWA, ul. Tarnowska 1; tel. +48 18 337 98 20 fax +48 18 337 98 25

www.limatherm.com
31.03.2016

Limatherm S.A.

I standard body IC version body


XD-I XD-IC

Standard cover

XD-IH XD-ICH

H version cover

XD-Iwin XD-ICwin

win version cover

XD-ILwin XD-ICLwin

Lwin version cover

XD-IHwin XD-ICHwin

Hwin version cover

34-600 LIMANOWA, ul. Tarnowska 1; tel. +48 18 337 98 20 fax +48 18 337 98 25