Lab 1a Wind Tunnel Testing Principles & Drag Coefficients of Golf Balls
Lab 1a Wind Tunnel Testing Principles & Drag Coefficients of Golf Balls
Lab 1a
Wind Tunnel Testing Principles & Drag Coefficients of Golf balls
OBJECTIVES
- To perform air flow measurement using the wind tunnel.
- To compare measured and theoretical velocities for various area changes along the flow
path of a Venturi insert.
- To calculate and compare the drag coefficient of a dimpled and non-dimpled golf ball.
- To perform uncertainty analysis, including bias and precision errors for the pressure
measurement and uncertainty of drag coefficients.
EQUIPMENT
Name Model S/N
Flowtek Wind Tunnel
Venturi inserts
Pressure Transducers
Golf balls
NI DAQ Card PCI-6024E
Computer
The wind tunnel flow measurement workstation consists of 8 pressure tubes connected to
pressure transducers to measure pressure relative to atmospheric pressure, Flowtek wind tunnel
with control panel to control air flow, and data acquisition system (DAQ NI PCI-6024E and
LabView 2010 software). In this experiment, you will perform air flow measurement along path
of Venturi inserts, and the dimpled and non-dimpled golf ball.
This lab was designed to be completed in 2 lab sessions. Data processing may take 1 or 2 more
sessions. Data processing should be performed as data is taken, to guarantee good results.
REQUIRED READING
Read reference [1] for introduction and operation of the wind tunnel. For further reference see
[2].
1
058:080 Experimental Engineering
PROCEDURE
1. Write detailed explanation of how the different components of the experiment work. Be sure
to include information about:
- How can you control the fan of the tunnel?
- What DAQ channel does each of the pressure taps connect to?
- Describe in detailed how to mount the rake with pressure tube.
2. A LabView program (lab1a.vi) will be used to control the measurement of this experiment
using the DAQ. The LabView VI is available on the course website under the Lab 1a section
of the Laboratories page. After opening the VI, you should be able to find the device PCI-
6024E by clicking Measurement & Automation under tools in LabView and then expanding
the Devices and Interfaces list.
Figure 1 in Appendix.A shows the front panel of the program. By pressing “Ctrl + e” you can
switch between the front panel view and the block diagram view. Explore the program to
understand its functionalities before conducting the experiment.
Note: The data saved in the specified file will be overwritten if the file name is not changed
for each run of the program. Also, click the “stop” button on the front panel to stop the
program instead of “abort execution” in the menu bar.
In the front panel, 0 to 7 denotes the corresponding channel measurement (i.e. 0 is channel 0
measurement, 1 is channel 1 measurement, etc.). The “Original measurement” (V0 to V7)
provide the voltages measured by the transducers, with instant and mean values. “Pressure”
(Pressure 0 to Pressure 7) show the pressure in Paschals (Pa) relative to atmospheric
pressure. “Velocity” (Velocity 0 to Velocity 7) show the velocity in m/s. The “Motor RPM
Controller” is adjustable and controls the percent of full-scale operation (% FSO) for the
wind tunnel fan.
i. Copy the downloaded folder containing the LabView VIs a place of your preference such
that you can play around with it without modifying the original. Use the copy for the entire
experiment.
ii. Test the program. Change the user input options if necessary in the front panel, such as the
file name and path where your data is to be saved.
iii. The “Measure” button takes a single point sample of all the pressure and velocity and records
the data into the file that you specified in step 2-ii. The “Read_Counter” will tell you how
many times the data is recorded. Note that the data will be recorded at the time when you
release the mouse, i.e., click and hold the mouse will NOT take any data.
iv. Look over the wind tunnel control panel (of the wind tunnel station). The “Control Power”
key turns on the power to the control panel. The “Fan Enable” button starts the tunnel fan,
while the E-Stop button stops it. On the back of the control panel box, there is a switch that
controls the method in which the motor RPMs are adjusted. If the switch is toggled up, the
2
058:080 Experimental Engineering
RPMs are adjusted using the Fan Speed dial on the panel. If the switch is toggled down, the
RPMs are adjusted using the “Motor RPM Controller” on the front panel of LabView
software. For the whole experiment, the RPMs will be adjusted using the computer.
You can control the RPM of the fan either by hit/sliding the controller, or
increment/decrement or enter the exact percentage of the RPM (0 – 100) in the controller’s
window. The controller of the fan is connected to channel ao0 of the DAQ.
v. Look over the pressure transducers in the back of the instrumentation box on the wind tunnel.
A total of 8 pressure tubes can be traced to pressure transducers located behind the tunnel.
These eight transducers are connected to the computer and correspond to the eight pressure
and velocity measurement on the front panel of the program. One of the pressure transducers
has a tube labeled “0” which connects at the place where the incoming air flow starts, thus
will provide reference for other tap measurement for the experiment.
vi. Important: Make sure there are no loose objects within the tunnel upon start-up.
Turn on the wind tunnel by turning the key from Lock to Run. Do not press “Fan Enable”
yet. Run the LabView program. Notice the indicator “Initializing” in the bottom left corner
saying “Wait” with the red button. Wait until the button turns green and the indicator says
“Ready” before continuing (it takes one minute). During this time, the program is initializing
all the pressures to zero. This measurement initialization is very important. Be sure to
perform this at the beginning of each run of the program. Once the indicator turns green,
then press the “Fan Enable” button on the wind tunnel control panel. Notice the changes that
take place on the screen. Slowly increase the RPMs by sliding up the Motor RPM slider in
small increments and then releasing the mouse button (or by other ways to change the RPM
in step 2-iv). Note: Do not make sudden and large increases in the RPMs.
vii. For each of RPM level, wait for the air flow to reach steady state (monitor it from readings of
the front panel, such as the original, pressure, or velocity measurement.) Record several data
at the steady state of each RPM (e.g. 10%, 20%, 30% FSO) by hitting and releasing the
“Measure” button. Notice the change in the indicator “Read_Counter”.
viii. Set the RPM back to 0 (without sudden changes) and turn off the wind tunnel by pressing
the E-Stop button on the wind tunnel control panel. The E-Stop button will always shut off
the motor.
ix. Stop the LabView program. Open the saved file and check the pressures and velocities. Since
tap 0 is the only tap connected to the wind tunnel, readings from tap 0 should give you the
correct measured information.
Now you should be able to perform the air flow measurement.
3
058:080 Experimental Engineering
4
058:080 Experimental Engineering
1.256 E-02
3
1.344 E-02
4
1.593 E-02
5
1.832 E-02
6
2.062 E-02
7
12. Start LabView program and perform measurement initialization.
13. Turn on the wind tunnel by pressing “Fan Enable” once the LabView says “Ready”. Slowly
adjust the RPMs such that the velocity at Tap 0 reads Velocity 0 ~ 6 m/s. Note and record
the motor speed in % FSO from the front panel of LabView program and in RPMs from the
control panel, as well as Tap 0 velocity reading in your logbook.
14. Make 20 measurements of the velocities by clicking “Measure” 20 times.
15. Repeat steps 13 and 14 for Velocity 0 ~ 12 and 18 m/s. Remember to change your file name
to be saved to avoid overwriting the data.
16. Set the Motor RPMs to zero (Do not make sudden change), stop the program, turn off the
fan, and turn the key to the “Lock” position. This might be a good stopping point for day 1.
However, make sure you perform your data processing of the Venturi insert part and obtain
good results (See Analysis, step 27 to 30). Do not wait until the end of the experiment to do
data analysis.
Drag coefficient of non-dimpled and dimpled golf ball
17. Measure the diameters of the non-dimpled and dimpled golf ball. Take 10 measurements for
each ball so that the precision error can be found.
18. Measure the location of the pressure taps relative to the central tap. The pressure taps will be
attached to the rake to measure pressure right after the golf ball which will then be used for
deriving the drag coefficient. You may have to detach the rake with the screwdriver to
measure the distances with the calipers. These values are radii, ri (i = 2~7), with the central
tap (Tap1) being r1 = 0 mm. Note that Tap 0 is for measuring the incoming air flow and is
not used for measurement with rake. When attaching the pressure taps, be aware about their
locations. See Figure 3 below for further explanation.
5
058:080 Experimental Engineering
Figure 3 Circular cross-section of the measure region (Each dot represents pressure tap
location)
19. Reattach the rake if it was taken out.
20. Insert the non-dimpled golf ball, lining it up with the rake.
21. After making sure that everything is secure within the tunnel, close the viewing window.
Start LabView program, perform measurement initialization and wait until the program says
“Ready”.
22. Increase the Motor RPM such that Velocity 0 = 6 m/s. In your log book, record Velocity 0,
the motor RPM with %, and the actual motor RPMs (the red numbers on the tunnel control
panel).
23. Record data for 30 times.
24. Repeat steps 22 and 23 for Velocity 0 = 12 m/s and Velocity 0 = 18 m/s.
25. Reduce the Motor RPM to zero, stop LabView program and turn off the fan.
26. Repeat steps 20 through 25 for the dimpled golf ball.
Analysis
For air flow along Venturi inserts:
27. Calculate the average and the precision errors for all the pressure and velocity values. Make
comment on accuracy of the pressure transducers.
28. Knowing that A*Vel = A0*Vel0 = constant, calculate the theoretical velocities at each
pressure tap and each Velocity 0. Here, for each Velocity 0 case, use Vel0 to be the average
velocity at Tap 0 and A0 to be 2.323x10-2 m2 .
29. Plot pressure tap vs. velocity for all three Velocity 0. Discuss and compare your theoretical
and measured velocities.
6
058:080 Experimental Engineering
30. Discuss any findings of your Venturi inserts measurement, such as the distribution of
velocities and pressure along the Venturi inserts.
For golf balls measurement:
31. Calculate the Reynolds number for each Velocity 0 and each ball:
𝑉𝑉0 × 𝑑𝑑 (2)
𝑅𝑅𝑅𝑅 =
𝜐𝜐
Where d is the ball diameter and ν is the kinematic viscosity of air (1.557×10-5 m2/s at 25
°C).
32. Calculate the drag force, Fdrag , using equation 3:
Where Pi is the pressure reading of Tap i (i = 2 ~ 7), P1 is the pressure reading at Tap 1, Ai is
area of annulus for the Tap i. See Figure 3 for reference. For instance, with the radii of ri (i =
1 ~ 7), 𝐴𝐴2 = 𝜋𝜋(𝑟𝑟22 − 𝑟𝑟12 ); 𝐴𝐴3 = 𝜋𝜋(𝑟𝑟32 − 𝑟𝑟22 ), etc. That is, Ai can be calculated with the
difference of circular area of tap i and tap (i-1). Make sure you understand why equation 3
can be used to approximate drag force.
33. Calculate the drag coefficient, Cd , for each Velocity 0 and each ball:
𝐹𝐹𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑 (4)
𝐶𝐶𝑑𝑑 =
1 2 𝜋𝜋𝑑𝑑 2
2 × 𝜌𝜌 × 𝑉𝑉𝑉𝑉𝑉𝑉0 × ( 4 )
Where ρ = 1.1845 kg/m3 , is air density and d is the ball diameter in [m]
34. For the non-dimpled and dimpled balls, plot the Cd vs. Re with Re on the x-axis.
35. Compare and discuss these plots with the theoretical curve for a sphere seen in Figure 4 of
Appendix.A. Explain the physical phenomena that your experimental results show.
36. Perform detailed uncertainty analysis on drag coefficients of non-dimpled and dimpled golf
balls. Estimate and discuss the uncertainties of the drag coefficients. Discuss all your
elemental errors and identify the largest sources of error.
REFERENCES
[1] GDJ Inc, “FLOTEK 360 Wind Tunnel Instructions and Operations Manual”,
http://www.engineering.uiowa.edu/~expeng/equipment/Flotek360_Wind_Tunnel_Instruction
.pdf (1998).
[2] Frank M. White, Fluid Mechanics, McGraw Hill (2010)
7
058:080 Experimental Engineering
APPENDIX.A
8
058:080 Experimental Engineering