Manual ATN-1100 Automatic Kjeldahl Nitrogen Determination Apparatus
Manual ATN-1100 Automatic Kjeldahl Nitrogen Determination Apparatus
Table of contents
Safety instructions ……………………………………………..………Page 1
Packing list …………………………………………………………….Page 3
Product overview and characteristics ………………………………Page 4-9
A. Product Overview…………………………….…..Page 4-8
B. Working Principle…………………………….…..Page 8
C. Technical Standard………………………….…..Page 9
Instrument structure and installation …………………………… Page10-11
Instrument initiate adjustment …………………………….……... Page 12-17
A. Reagent pump flow calibration ……………… Page12
B. System settings……………………………….…..Page15-17
C. Formula managemen……………………….…..Page17
Data inquiry……………………………………………………………… Page 18
Operation method ……………………………………………………Page19-23
Titrating solution calibration ………………………………….….Page 24
Safety system ………………………………………………………… Page25
Attention and maintenance ……………………………………….. Page26
Common problems and solution ………………………………… Page26-29
Reagent preparation method ………………………………………Page30-33
Commonly-used consumables …………………………………… Page31-34
Version: 2018/v01
Dear user:
Thank you for using our products. Please read the operation manual of the product very carefully
prior to using our products to ensure that the instrument is able to run effectively.
Please don’t hesitate to contact us for further assistances and / or services; we are willing to provide
We tried to make our manual as “detailed”, “professional” and “simple” as possible; nevertheless, we
are sorry for the misunderstanding and inconveniences of reading due to the inevitable errors.
We always looking forward to your valuable comment and / or idea which about the operation of the
Safety Instructions!
1、 Operator shall use the instrument according to the operation requirements, or else, property losses or
3、 Please use independent outlet, and make sure that outlet ground wire is in good condition (it is strictly
4、 Please do not plug or unplug the instrument without turn off power switch.
5、 Please do not repair without permission, and professional guidance is needed if the repair is
6、 Please make sure the equipment stay away from the electromagnetic interference sources in order to
7、 Please use in working environment with temperature of 10℃--45℃ and relative humidity less than
80%.
8、 Please use in the environment without dust and combustible and explosive gas
9、 Please do not install instrument at position with direct sunlight to ensure data accuracy and prevent
10、 For reagent and other solutions is corrosive! Please wear protection gear and operate according to
user manual.
full-automatic Kjeldahl azotometer packing list
One set of DW-ATN-1100 One 25L waste liquid barrel (green label)
mainframe One 1L dropping bottle One 10L distilled water barrel (blue label)
One 5L boric acid barrel (yellow label)
One 5L alkaline solution barrel (red label)
One Cooling water inlet pipe One Cooling water outlet pipe One Distilled water delivering pipe (blue label 3 m)
(including adapter) 5 meter 1 One Alkaline liquor delivering pipe (red label 3m)
5 meter One Boracic acid delivering pipe (yellow label 3m)
One Liquid waste delivering pipe (green label 3m)
One bottle catalyzer 250g One Cooling water inlet One Sample room water
Five rolls of printing paper pipe seal ring receiver
One
One Digestion tube (it has been installed in instrument
One wrench
Power cord (110V/10A) sample chamber)
Φ42*300mm)
Product qualification certificate Product serial number sticker 2 pieces
Product Manual And
2 copies Product warranty card
Chapter I: Overview and technical standard of DW-ATN-1100 series
azotometer
Kjeldah method is the classical method for determining nitrogen contents. At present, kjeldah
method is widely used for determining nitrogen-containing compound in soil, food, agricultural products,
fodder and others. Three procedures including sample digestion, distillation separation and titrimetric
analysis shall be conducted successively in case of using above method
DW-ATN-1100 series full-automatic kjeldahl azotometer is a kind of analysis system used for
determining titration endpoint with full-automatic distillation ——color method and developed according
to classic kjeldah method. Products of the series include base-type DW-ATN-1100, professional type
DW-ATN-1100A, computer type DW-ATN-1100B products, and their operating methods are basically same.
The manual gives guidance and description based on DW-ATN-1100A
I. DW-ATN-1100 series products’ characteristics
1、 DW-ATN-1100 characteristics
The system uses 7.5-inch color touch screen+ Siemens PLC
control module and an interactive touch screen in order to be
operated simply and conveniently and easily to learn. (see
Figure 1-1)
Feed liquid pump—— Anti-corrosion self-suction diaphragm
pump is used for feeding, abandoned air-filled feeding Figure1-1
methods which is used by many other manufactures. Use of
self-suction pump for feeding can generate many advantages
such as enable liquid feeding barrel to be away from main
frame or placed under operating floor and so on.
Pipeline system—— Uses imported (made in USA) modified
silicone pipe that cannot be deformed and corroded under
strong acid and alkali environment for a long time. Figure 1-2
Use of American CPC quick-connect pipe fitting (see
Figure1-2) enable instrument to be used conveniently also
prevent liquid and gas leakage in the system.
Titration system —— Enable user easily to check the accuracy
and reliability of the result at any time during titration
processes at any time. (Figure 1-3)
Visible titration cup—— Enable user to observe the change
during titration processes, improve the convenience for
cleaning and changing the cup. Figure 1-3
Pre-heating function—— Maintain steam generator
temperature at pre-set temperature range to maintain the starting-up or standby status, which is
beneficial to generating steam quickly in case of analysis and shortening analysis time
Automatic alkali solution adding mode for preventing overreaction between alkali and acid
Automatic alkali solution adding——In case of insufficient alkali solution added during operation,
adding procedure can be conducted repeatedly to avoid analysis error or sample scrapping
caused by insufficient quantity of alkali solution. (see Figure1-4)
Automatic titration cup cleaning——Automatic spraying of titration
cup cleaning can prevent residual liquid in the cup influencing
low-content sample or blank sample and improve system accuracy
Automatically deducting blank value in sample——blank value of
sample can be deducted automatically.
Titration curve——Enable user to observation the whole processes to
located the error appeared during titration processes. (see Figure 1-4
Figure1-6)`leakage and electric shock due to various factors and
protect operator.
Digestion tube absence protection——If digestion tube is not
provided, reagent adding and heating shall not be conducted to
prevent personal injury caused by reagent and steam.
Safety door protection——Transparency safety door with thickness of
Figure1-5
5mm can be operated conveniently and safely to ensure personal
safety.
Water shortage protection for steam system——System can shut down automatically and send
alarm in case of water shortage in the system to prevent accidents
happening.
Steam kettle over-temperature protection system——System can
shut down automatically and send alarm in case of over-temperature
in steam kettle caused by distillation system fault.
Safety alkali solution adding mode——Use of this mode can prevent
reduce strong neutralization reaction during adding alkali and
Figure1-6
prevent pipe bursting.
2、 DW-ATN-1100A newly-increased functions (in addition to possession of above
characters)
Formula management——User can add 10 preset parameters into instrument, increase the
convinces of the operation.
Analysis mode —— Two modes including “automatic analysis” and “manual analysis” provide
convenience for operator.
Automatic discharge for waste liquor in digestion tube——After completing titration, waste
liquid in digestion tube will discharge automatically and be collected, improvement of degree of
automation reduces labor strength of staff and enable concentrated collection of waste liquid
after analysis
Alkali liquid pipeline cleaning—— implementation of alkali liquid cleaning by pressing “alkali
pipeline cleaning” button regularly can prevent blocking in the pipeline and improve feeding
accuracy (see Figure 1-6)
Titrating solution concentration (mol) auto-calibrating system——the calibration method can be
used to eliminate difference caused by “titration by instrument” after using “manual calibration”
and make system error smaller and analysis accuracy higher.
Simultaneous titration and calibrating —— Increase the efficiency and save the energy and
resource.
Steam flow adjustment——Two levels include 100% “high steam content” and 75% “middle
steam content” are provided for users to choose.
Cooling water flow display——user can observe condition of cooling water flow at any time to
prevent insufficient cooling due to too small water volume or wasting water due to too large
water volume
Steam overpressure protection —— If pipeline blocking causes abnormal high pressure in steam
system, the system can shut down automatically and send alarm to prevent accidents becoming
more severe and protect operators and property safety.
Temperature monitoring system of distilling samples——Prevent distilled samples data error
caused by insufficient cooling water or too high temperature, monitoring system will shut down
automatically and send alarm.
Cooling water volume statistic (XX3) —— Through the data, user can know cooling water
consumption of cooling water (in statistics) to calculate the expenses.
Multi-function interface:
A. Reagent barrel level display alarm interface ——in case of insufficient reagent in reagent barrel
and distilled water barrel, the system will send alarm and shut down to prevent instrument operating
under condition of no reagent and distilled water. (the system is “optional”)
B. Recirculation cooler interface——after azotometer is connected to “intelligent cooling water
system”, “cooling system” doesn't need to be set or opened or closed again. When azotometer is
opened, “cooling system” automatically starts up and execute inner cycle program, when there is
steam generated in azotometer, above program in “cooling system” is automatically changed into
outer cycle program to lower temperature of azotometer (see Figure1-7 and Figure 2-2) (note: the
system shall be used together with “intelligent cooling water system” produced by the
manufacturer)
USB interface——Mainly used for connecting wireless mouse and analyzing data and exporting
data. (see Figure 1—7,Figure 2--2)
3、 DW-ATN-1100A newly-increased functions (in addition to possession of
above performances)
Computer operating software, computer interface ( 485 ) ——Enable
computer control and management (see Figure 1—7,Figure2--2)
Internet interface ——Enable remote control and instant data transfer (see
Figure1—7,Figure2--2)
Group user’s management center——control center consists of 2 sets-50
sets of “full-auto Kjeldahl azotometer” and can realize dynamic
monitoring, instantaneous transmission of data and control center data
storage for any instrument
4、 Working principles Figure 1--7
Distillation and titration analysis can be conducted after sample digestion is completed and
sample solution is cooled to indoor temperature. Process principles are as follows:
(NH4)2SO4+2NaOH= Na2SO4+2H2O+2NH3↑
That is, under alkaline condition, after separating ammonia with high-temperature steam,
collecting ammonia and steam mixture in titration cup containing boric acid absorption liquid
(including mixed indicator), use Sulphur acid titration liquid to conduct titration analysis in
photometric titration system (see Figure 1—8)
Figure 1—8
Automatic
Operating mode Automatic analysis/manual analysis
analysis
Safety alkali adding mode None Yes: any setting within range of 0—9999s
Automatic cleaning of alkali solution None Yes (automatically wash when starting up)
Front structure (see Figure 2—1) Back structure (see Figure 2—2)
Users selecting reagent barrel level display and alarm system insert reagent barrel level
sensor into “multifunction interface” (Figure 1—9, Figure2-2) (Optional)
9、Add reagent and distilled water into reagent barrel (see Chapter IX Reagent Preparation Method)
10、Put water receiver into instrument sample chamber and (see
Figure 2—11) close safety door
11、Insert power wire into “power socket” on back of instrument (see
Figure 2—12), and then insert another end into outlet.
12、Check(Figure 2-1-C) “power switch” shall be at “turn off ”state,
turn on “leakage protector” switch upwards (see Figure 2-13),
13、Put titration bottle which contain titrating solution into instrument
and insert right end of pump pipe into hole at upper end of the
bottle (see Figure 2—14) Figure 2—10
14、Close the instrument’s door, installation complete.
Figure 2—11
Figure 2—14
Chapter III, Instrument initiate adjustment
I. Startup:
1. Close safety door after confirming the digestion tube in sample chamber is at designated position
(see Figure 3—5)
2. Open power switch (see Figure3-6)
a) After turned on, enter self-check program (spend about 3 minutes)
Working process after DW-ATN-1100A/B starts up——discharge of residual liquid in
digestion tube—→ discharge of distilled water in steam kettle—→discharge of residual
liquid in titration cup—→ water adding in steam kettle—→heating—→ standby
Working process after DW-ATN-1100 starts up——discharge of distilled water in steam
kettle—→ discharge of residual liquid in titration cup—→ water adding in steam
kettle—→heating—→ standby
b) Startup page display after completing self-check shows below. (see Figure3-1)
3. Click (Figure3-1) center of the clock, an dialog window will appears on page (see Figure3-2)
5. Pump change solution twice (if change from high concentration to low concentration, it is
recommended to go through procedure twice)
6. After solution changed. Insert burette back into (Figure 3-7-c)
IV. System setting:
"System setting" is necessary to guarantee the normal operation of equipment, before
conducting sample analysis, confirmation shall be conducted
for the setting parameter to avoid analysis data error caused
by wrong parameter setting.
1. DW-ATN-1100 system setting(See Figure 3—15)
A. Absorption liquid—— Refers to setting of added volumes of
absorption liquid (boric acid) in titration cup automatic
Figure 3—15
analysis.
B. Alkali liquid——Refers to setting of added volumes of alkali liquid in digestion tube during
automatic analysis
C. Diluent——Refers to setting of added volumes of diluent in digestion tube during automatic
analysis. Note: before adding alkali liquid, add “diluents” (distilled water) properly to reduce
sharp reaction between alkali and acid
D. Distillation waiting time——It means that during automatic analysis, and after adding alkali
solution, “heating and distillation” shall not be conducted immediately until “the setting time”
ends. It is mainly used for analysis for partial samples containing nitrate nitrogen (It takes certain
reaction time after adding Nitrogen fixing alloy)
Hint! If instrument cannot heat at normal working state, such phenomena are related to the
setting.
E. Stand-by time——the instrument can heat automatically when it starts up, and the temperature
will keep “constant” within set “stand-by time”. If no persons operate under condition of
exceeding set time, heating system will not keep “constant” and will enter dormant state for
energy conservation. In case of operating system again, the system automatically enters heating
and constant temperature state
F. Standby temperature——Refers to temperature maintained “in distillatory kettle” within above
standby time and displayed at top right corner,
Hint! Set value of “standby temperature” shall not be too high, and
maximum temperature shall not be more than 70℃.
G. Safety alkali solution adding mode time——Refers to a kind of
relatively safe alkali solution adding mode. In order to prevent
sharp and strong reaction between acid and alkali, first, conduct
Figure 3—16
heating for sample containing acid properly, and then add alkali
solution after temperature of above sample rises to reduce
reaction degree properly. "Safety mode time" refers to suitable
"heating time" for sample.
Hint! “Heating time” shall not be too long, and shall be suitable
for making sample be at “almost boiling” state to prevent acid mist
generating after boiling.
Figure 3—17
H. Titrating solution concentration input value——Titrating solution
H+ concentration value input.
I. Titrating cup cleaning time——Refers to setting of times during
Figure 3—18
the process of using distilled water to wash away residual liquid in titrating cup.
2、 ATN-1100A/B system setting (see Figure 3—16)
Other settings are the same as those of DW-ATN-1100 (see Figure 3—15), newly-added
setting is simply described as follows
A. Cooling water flow switch——Turn on switch and monitor, the system displays cooling water
flow state, and if flow is less than 0.6L/min, the system will send alarm and shut down
Attention! If cooling water flow is not stable, turn off flow switch to prevent system sending
alarm and shutting down due to unstable flow.
B. Cumulative cooling water flow——Accumulative statistics of total cooling water consumption
(cubic meter).
C. Digestion tube waste liquid discharging time——Refers to setting of time of digestion tube
waste liquid discharging after completing analysis.
D. Steam flow selection:
High flow—— Steam flow reaches to maximum value during distillation and quantity of
steam counts for 100%
Middle flow—— Steam flow is reduced during distillation, and quantity of steam accounts
for 75%
E. Automatic alkali solution adding—— When function on, before distillation processes end, add
3ml alkali solution again to help blow out the sample attached on the tube.
F. Simultaneously distillation and titration—— When function on, when distillation sample around
70ml, titration processes automatic start and shutdown when titration finish.
Attention: Before use this mode, please make sure it match your analysis methods.
3. Formula management—— User can set 10 groups of different formulas by different analysis con
dition with their own preference, to provide convinces for future operation.
Setting and storage method——Click “formula management” button (Figure 3—4) and
enter (Figure 3—17.
Click “formula number” (Figure 3—17) and dialog box pop out. (Figure 3-18)
Select number (see Figure 3-8) in the dialog box, for example: “2”.
Input conditions as needed successively (see Figure 3—18), such as:
Absorption liquid——25ml
Alkali liquid——50ml
Diluent——10ml
Distillation waiting time——0min
Digestion tube waste liquid discharging time——15s
Alkali adding mode——common
Steam flow——75%
Click “return” button and complete setting and storage
Overview: After completing installation and setting according to above procedures, the
instrument has obtained basic conditions of sample analysis, in
consideration of relatively large difference among sample treatment
and analysis methods of different industries, so the operating method
is only used as interpretation for procedures concerning way of the
instrument to complete sample analysis
I. Automatic analysis operating button and window
Figure 5—1
interpretation:
1、 DW-ATN-1100 operating system
A. Enter "automatic analysis" page
a) Click “automatic analysis” on “system home page” in Figure
5-1
b) Enter “alkali adding mode selection” page (see Figure 5—2)
c) Select “safety” mode
Figure 5—2
(Before adding alkali solution into samples, first heat
samples, and the heating time needs to be set in “safety
mode time” on “system setting” page. So, confirm that
“safety mode time” has been set before selecting “safety
mode”. Note! "Safety mode time" shall be set to be suitable
for making samples be at almost boiling state to prevent
acid mist pollution generating after boiling) Figure 5-3
d) Or select “common” mode
e) Enter "automatic analysis” page (see Figure 5—3)
B. Automatic analysis page window, button function
interpretation (see Figure 5—3)
Samples “analysis results” display area is at left side of the
page.The system state display and operating
button are at right side of the page Figure 5—4
a) Distilled water level——Indicates water level state in steam kettle
b) Steam temperature——Indicates temperature in steam kettle, and alarm will be sent and
shutdown will be caused in case of over temperature
c) Alkali solution adding mode—— Indicates alkali solution adding mode selected
d) Mode time—— Indicates set time of safety mode
e) Absorption liquid—— Indicates added volume of absorption liquid
f) Diluent—— Indicates added quantity of digestion solutions
g) Alkali solution—— Indicates added volume of alkali solution
h) Titrating solution concentration—— Indicates H+ concentration of titrating solution used
i) Blank value input window—— Sample blank value input window
Click and dialog box (see Figure 5—4) pop up, input data and confirm
j) Exit button—— exit from the page after clicking
k) Sample data button——Click it and enter sample data input page (see Figure 5—5)
Sample number window——Users can number within 3 digits according to needs, or
else, the system will take current number as default value, but automatic analysis
results will not be affected (see Figure 5—6)
l) Sample weight window——The final
display results (nitrogen content) on “automatic
analysis” page will be affected without the data input,
but titration volume will not be affected. If users have
forgotten to fill in, results can be obtained through
manual calculation according to titration volume and
sample weight recorded (see Figure 5—6)
m) Curve observation button——Click and enter page Figure 5—5
displaying sample titration process real-time curve in( Fi
gure 5—7) used for analyzing sample working state
Too steep (vertical) curve indicates that titrating
solution concentration value is too high. Results: big
error
Straight (skew line) curve indicates that titrating
solution concentration value is too low. Results: no Figure 5—6
end point
n) Titration cup cleaning button——After clicking, spraying
and cleaning are conducted for titration cup, and after
that, waste liquid discharges automatically
o) Additional alkali adding button——If quantity of alkali
solution added is not enough during system operation,
such adding can be conducted many times by pressing
the button Figure 5—7
Dialog pop ups after clicking (see Figure 5—8), and
then input added quantity of alkali solution and then
click “confirm” to begin adding
p) Operation button/stop button——Analysis system
startup/stop button
Dialog box pops up after clicking (see Figure 5—9),
Figure 5—8
and analysis system operation begins after clicking
“confirm”
System will stop operating after user confirm the
“stop” dialog. \
2、 DW-ATN-1100A/B operating system
Basic operating method of DW-ATN-1100A/B is the same as of
DW-ATN-1100, this part only provides description of
Figure 5—9
newly-added contents and functions (see Figure 5—10)
Enter “automatic analysis” page
a) Click “automatic analysis” on “system home page” in Figure 5-10
b) Enter “digestion tube waste liquid discharging” page (see Figure 5—11)
Hint! Confirm that sample particle size is more than or eq
c) ual to 60 mesh before selecting “automatic discharging”
to prevent pipe blocking caused by too large particle
c) Select “automatic discharging” and enter “alkali solution
adding mode selection” page (see Figure 5—12)
d) Select “common” mode and enter automatic analysis
page (see Figure 5—13)
Automatic analysis page window and button function Figure 5—10
j) Titration curve observation —— Curve observation and printing during titration process.
II. "Automatic analysis" operating method:
Now “automatic analysis” operation is expainl according to DW-ATN-1100A full-auto
Kjeldahl azotometer
1. Reagent, distilled water barrel, reagent of titration bottle
2. Water tap is at open position
3. See relevant parameters on “system setting page” according to sample conditions,
4. For example: Chose pre-set formula # “2”
5. Place sample at designated position and close safety door
6. Enter “system home page”
7. Select “automatic analysis”
8. Select “automatic discharging” in “digestion tube waste liquid discharging mode selection”
9. Select “common” in “alkali solution adding mode”
10. Enter "automatic analysis” page (see Figure 5—13)
A. Click “formula selecting” pull-down menu, select # “2” formula (see Figure 5—14)
B. Input “0” in “blank input”
C. Click “sample data” and enter “sample data input” page
a) Input sample number, for example:101
b) Input sample weight, for example:0.1234g
D. Click “return” button to return “automatic analysis” page
E. Click “operation” button and “Yes” to confirm
a) Add boric acid in titration cup
b) Motor operates to stir acid in the cup
c) Add diluent in digestion tube
d) Add alkali solution in digestion tube
e) Distillation system works and temperature of steam kettle begins to rise
f) Cooling water valve automatically opens after rising to 65℃, tap water flows in for
cooling
g) After rising to 100℃, a large amount of steam containing samples is changed into
distilled water containing samples flowing into titration cup
h) If volume of titration cup reaches to value as required (about 110ml),distillation system
stops working
i)Cooling water valve automatically closes
j)Distillation system automatically begins working (photometric lamp lights)
k) After reaching to titration end point (liquid in the cup becomes red), distillation system
automatically stops working
l)Waste liquid in digestion tube discharges
m) Waste liquid in titration cup discharges
n) Digestion tube and titration cup cleaning begins
o) Waste liquid (after washing) in digestion tube discharges
p) Waste liquid (after washing) in titration cup discharges
hours, its weight is constant, take it out and put into dry
container for cooling to indoor temperature for use
3. Titrating solution preparation—— it is conducted according to
concentration about 0.1000 mol/L, see Chapter IX Reagent
Preparation
After completing titration solution preparation, put such Figure 6-2
solution into Titration bottle and then put the bottle into
instrument (see Figure 2—14)
4. Weighting three “ammonium dihydrogen phosphate” samples
accurately (e.g.:0.1010g,0.1020g ,0.1030g), put them into
three clean digestion tubes respectively and mark these tubes
with 1#, 2#, 3# for use Figure 6—3
II. Instrument pre-heating operation
1. Take small amount of “ammonium dihydrogen phosphate”
sample (about 0.1 g ) and put it into digestion tube provided
for instrument
2. Click “flow calibration” (see Figure 6-1) and enter “flow
calibration page” (see Figure 6—2)
Figure 6—4
3. Click “next page” (Figure 6—2) and enter “confirmation of
effective components of reagent” page (see Figure 6—3)
4. After clicking “coefficient” window, data input box pops up
(see Figure 6—4)
Coefficient value :0.0000——1.000
5. Input effective components of reagent, such as 0.995, and
confirm (see Figure 6—4) Figure 6—5
6. Click “confirm” (Figure6—4) and enter “concentration
calibration” page (Figure 6-5)
“Titration concentration calibration” page window contents
interpretation
Sample weight input window—— accurately input
sample weight (3 values in total)
Figure 6—6
Titrating solution volume display window—— Display
volume consumed in this titration (3 values in total)
7. After clicking “operate”, “calibration confirmation” dialog pops up (see Figure 6—6)
8. Click “confirm”
9. Instrument preheating begins
10. Processes automatically stopping (Note: pre-heating operation of this time is not involved in
calculation, so “titration flow volume” is not displayed)
III. Liquid concentration calibration
1. Remove digestion tube for pre-heating
2. Put 1# digestion tube into instrument and close safety door
3. After clicking “operate” (Figure 6—6), “calibration
confirmation” dialog pops up(see Figure 6-6)
4. After clicking “confirm”, calibration begins
Figure 6—7
5. 1# digestion tube operates continuously before automatically
stopping
“Titration volume display” will display titrating solution volume consumed by sample in 1#
digestion tube (for reference)
6. Click “the first time” “sample weight input” window (Figure 6—7)
7. Digital input dialog pops up (see Figure 6—7)
8. Input weight of 0.1010g of sample in 1# digestion tube and confirm
9. 1# tube calibration is completed
10. Calibration method of 2# and 3# digestion tubes is the same as 1# tube
11. Repeat calibration for 2# and 3# digestion tubes
12. After completing calibration for these three tubes, system calculate and stores average value of
three “calibration value”, and display in “calibration value” window.
Hint! “calibration value" in the window and “titrating solution concentration input value” on
"system setting" page are connected, so "value" update in the window and "value" update on
"system setting" page happen at the same time. Similarly, "value" update on "system setting"
page and "value" update in this window happen at the same time
13. Calibration is completed
Φ42*300mm
A000123 Digestion tube
(340mL)
Φ50*300mm
A000123 Digestion tube large volume
(500mL)
A000454 Kjeldahl high 750g
efficiency
A000123 catalytic pill 250g
Reagent barrel
A000246 5L
(made in US)