Aestiva/5 Anesthesia Machine: GE Healthcare
Aestiva/5 Anesthesia Machine: GE Healthcare
Aestiva/5
anesthesia
machine Aestiva®/5
Two vaporizer configuration
More than superior ventilation
Features
Superior ventilation: 7900 SmartVent™
• Volume Mode, pressure
• Volume Mode, Pressure Control Mode, Pressure Support
(PSVPro®), Synchronized Intermittent Mandatory
Ventilation (SIMV), electronic PEEP
• Tidal volume compensation
• One motion from mechanical to manual mode
• Two key presses to total standby: end case
• Cardiac bypass case mode
Aestiva/5
Open systems architecture Three vaporizer configuration
Innovative patient breathing system Improved low flow/reduced life cycle costs
• Eight machine hoses/cables integrated • Fresh gas flow compensation—automatically
• “No tools” disassembly of components • Smooth, faster acting fresh gas flow control
• Autoclavable and latex-free • Minimum O2 flow of 50 mL
• Responsive location of common gas outlet • Dual air flow tube for low flow
• Two scheduled maintenance checks per year
Physical Specifications
Dimensions Top drawer (1 standard)—locking (internal dimensions)
Tec 6 Plus
Pneumatic specifications Flowmeters
Internal common gas outlet O2 ranges: Two tubes: 0.05 to 0.95
L/min and 1 to 15 L/min
Connector: ISO 22 mm OD and 15 mm ID
Minimum O2 flow:
Auxiliary common gas outlet (optional) 50 mL/min ±25 mL
N2O ranges: Two tubes: 0 to 0.95
Connector: ISO 22 mm OD and 15 mm ID
L/min and 1 to 10 L/min
Gas supply Air range: One tube option:
1 to 15 L/min
Pipeline
input range: 240 kPa to 600 kPa/ Two tube option: 0 to
35 psig to 88 psig 0.95 and 1 to 15 L/min
Pipeline (low flow tube optional)
connections: DISS-male
CO2 (optional): One tube: 0 to 0.5 L/min
All fittings available for O2, N2O,
Heliox range
and Air, and contain pipeline filter
(optional): One tube: 0 to 15 L/min
and check valve.
Calibration: Percent of Accuracy
Cylinder input: Pin indexed in accordance with
full scale flow (% of flowrate)
CGA-V-1; contains input filter and
check valve 100 ±2.5%
90 ±2.5%
Note: Maximum 5 cylinders total; one
oxygen required. 80 ±2.6%
Electromagnetic compatibility
Immunity: Complies with all requirements
of EN 60601-1-2
Emissions: CISPR 11 group 1 class B
Approvals: UL 2601-1,
CSA C22.2 #601.1
IEC 601-1
EN 60601-1
© 2005 General Electric Company – All rights reserved.
GE and GE Monogram are trademarks of
General Electric Company.
GE Healthcare
P.O. Box 7550
Madison, WI 53707-7550
USA
www.gehealthcare.com
AN2594-K 7/05