H30 Removal Instruction v1.0

The document provides instructions for disassembling and replacing various modules and components of an instrument. It describes removing the sampling module probe and o-rings, as well as the rinsing head. Instructions are also given for replacing the x-axis and y-axis sensors. Finally, steps are outlined for removing the entire sampling module from the instrument by disconnecting connectors, tubing, and removing screws and covers.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
86 views45 pages

H30 Removal Instruction v1.0

The document provides instructions for disassembling and replacing various modules and components of an instrument. It describes removing the sampling module probe and o-rings, as well as the rinsing head. Instructions are also given for replacing the x-axis and y-axis sensors. Finally, steps are outlined for removing the entire sampling module from the instrument by disconnecting connectors, tubing, and removing screws and covers.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 45

Disassembling Instruction V01

Elements Replacement

1. Modules locations

SAMPLING MODULE
SYRINGE VALVE MODULE

REAGENT SYRINGE MODULE WASTE SYRINGE MODULE

COUNTING MODULE

COUNTING VALVE MODULE

1
Disassembling Instruction V01

2. Sampling Module

3. Elements replacement
a. Sample probe/Sample probe O-ring

Removal
Action Steps Reference

Prerequisite 1. Drain the instrument from any liquids.

2. Switch off the instrument and disconnect the


main power cord.

3. Open the fluidic door

Sampling 1. Pull on the top of the probe to remove it from


Probe the probe carriage.

2
Disassembling Instruction V01

2. Pull on the rinsing head to remove it from the


sampling module.

1
3. Pull up the probe to remove it from the rinsing
head.

4. To replace the probe disconnect it from the


tubing # 9 (1) and remove it from the
instrument.

5. Replace the probe by following the reverse


instruction.

O-ring 6. To replace the O-ring of the rinsing head,


unscrew the probe guide and remove the
assembly probe/guide/O-ring from the rinsing
head.

Note: Be careful to not lose the plastic washer (2)


located under the O-ring inside the rinsing head.

7. Remove the O-ring from the probe and replace


it.

8. To replace the O-ring, insert first the probe


guide, then the O-ring on the probe, then the
plastic washer.

9. Install this assembly on the rinsing head and 2


screw back the probe guide.

10. Reinstall the complete assembly on the sampling


module

3
Disassembling Instruction V01

b. Rinsing head

Removal
Action Steps Reference

Prerequisite 1. Drain the instrument from any liquids.

2. Switch off the instrument and disconnect the


main power cord.

3. Open the fluidic door

Rinsing 4. Pull on the top of the probe to remove it from


Head the probe carriage.

5. Pull on the rinsing head to remove it from the


sampling module.

4
Disassembling Instruction V01

6. Pull up the probe to remove it from the rinsing


head. 1
7. Disconnect tubings #10 (1) and #21 from the
rinsing head and replace it.

8. Put back the assembly following the reverse


instruction
2

2
1

5
Disassembling Instruction V01

c. Y axis sensor

Removal
Action Steps Reference

Prerequisite 1. Switch off the instrument and disconnect the


main power cord.

2. Open the fluidic door

Y Axis 3. Remove the sensor fixation screw (1).


Sensor

4. Disconnect the sensor from its connector (2).


2

5. Remove all the cable ties (3) which maintain the


cable.

6. Remove the sensor and replace it following the 3


reverse instructions.

6
Disassembling Instruction V01

d. X axis sensor

Removal
Action Steps Reference

Prerequisite 7. Switch off the instrument and disconnect the


main power cord.

8. Open the fluidic door

X Axis 9. Remove the sensor fixation screw (1).


Sensor

10. Disconnect the sensor from its connector (2).

11. Remove the sensor and replace it following the


reverse instructions. 2

7
Disassembling Instruction V01

e. Module replacement

Removal
Action Steps Reference

Prerequisite 1. Switch off the instrument and disconnect the


main power cord.

2. Remove the fluidic door.

3. Remove the top cover.

4. Remove the front cover.

5. Remove the reagent plate.

6. Remove the right plate of the reagent


compartment.

7. Remove the assembly probe/rinsing head from


the module.

8. Remove the Y sensor from the module.

Complete 9. Disconnect the elbow connector from tub#10.


Module

10. Remove the tubing tie (1) and remove tub #10
and #21 from the module.

8
Disassembling Instruction V01

11. Disconnect the Y motor connector (1).

12. Disconnect the X motor connector (2).

13. Remove the three screws from Left bearing 3


support (3).

14. Remove the two screws (4) which maintain the X


motor protection cover and remove it from the 4
motor.
4

9
Disassembling Instruction V01

15. With a torque tool, turn the X axis screw to


place the Y axis approximately at the middle of
the X axis screw.

16. Remove the two screws from right bearing


support (5).

17. Remove the two screws (6) which maintain the X


axis guide and remove it.

6 6

18. Remove the two screws (7) which maintain the


plastic square and remove it.

7 7

10
Disassembling Instruction V01

19. Rotate the complete module to pass the X axis


screw and the shaft on the fluidic side.

20. The X axis screw and the shaft must pass


through the “window”.

21. Once the screw and shaft have passed through


the “window”, continue to rotate the complete
module until to be able to pass the X axis motor
on the fluidic side.

Note: Top view.

11
Disassembling Instruction V01

22. Once the complete module is perpendicular to


the fluidic plate pull on it and remove it.

23. Replace the complete module respecting the


opposite instruction.

f. RBC Electrode

Removal
Action Steps Reference

Prerequisite 1. Drain the instrument from any liquids.

2. Switch off the instrument and disconnect the


main power cord.

3. Open the fluidic door

Electrode 4. Turn the electrode plastic holder from a quarter


RBC turn anticlockwise.
5. Pull on it and remove the RBC electrode
assembly from the RBC counting bath.

6. Remove the ground cable fixation screw (1) of


1
the RBC electrode.

7. Disconnect the RBC coaxial connector (2) from


the counting manifold and then remove the RBC
electrode from the instrument. 2

8. Replace the RBC electrode respecting the


reverse instruction.

12
Disassembling Instruction V01

g. RBC Counting Bath

Removal
Action Steps Reference

Prerequisite 1. Drain the instrument from any liquids.

2. Switch off the instrument and disconnect the


main power cord.

3. Open the fluidic door.

4. Remove the RBC electrode.

RBC 5. Release the maintaining screw (1) of the


Bath protection cover and remove it.

6. Remove the three remaining fixation screws (2)


of the RBC bath. 2 2

7. Pull on the RBC bath and remove it.

8. Replace the RBC bath following the reverse


instruction.

Note: Verify the presence of the O-ring (3) on the


red fitting before to perform the reassembling.
3

13
Disassembling Instruction V01

Note: Verify the presence of the three O-rings (4) on


the bath before to perform the reassembling.

4 4

Important note: Screw back the four


fixation screws respecting a torque 50cNm
to 50cNm

14
Disassembling Instruction V01

h. RBC Counting Head

Removal
Action Steps Reference

Prerequisite 1. Drain the instrument from any liquids.

2. Switch off the instrument and disconnect the


main power cord.

3. Open the fluidic door.

4. Remove the RBC electrode.

5. Remove the RBC counting bath.

RBC 6. Remove the two fixations screws (1) of the RBC


Counting counting head.
Head

7. Remove the RBC counting head using pliers to


pull it if necessary.

15
Disassembling Instruction V01

8. Replace the RBC counting head following the


reverse instruction.

Note: Verify the presence of the O-ring (2) before to


perform to the reassembling.

40cNm

Important note: Screw back the two


fixation screws, respecting a torque
to 40cNm.

16
Disassembling Instruction V01

i. WBC Electrode

Removal
Action Steps Reference

Prerequisite 9. Drain the instrument from any liquids.

10. Switch off the instrument and disconnect the


main power cord.

11. Open the fluidic door.

Electrode 12. Turn the electrode plastic holder from a quarter


WBC turn anticlockwise.
13. Pull on it and remove the RBC electrode
assembly from the RBC counting bath.

14. Remove the ground cable fixation screw (1) of


the RBC electrode. 1

15. Disconnect the RBC coaxial connector (2) from


the counting manifold and then remove the RBC
electrode from the instrument. 2

16. Replace the RBC electrode respecting the


reverse instruction.

17
Disassembling Instruction V01

j. WBC Counting Bath

Removal
Action Steps Reference

Prerequisite 9. Drain the instrument from any liquids.

10. Switch off the instrument and disconnect the


main power cord.

11. Open the fluidic door.

12. Remove the WBC electrode.

13. Remove the protection cover.

WBC 14. Disconnect the HGB cable connector (1).


Bath

15. Remove the three remaining screws (2) of the


WBC bath.
2 2
16. Pull on the WBC bath and remove it.

17. Replace the WBC bath following the reverse


instruction.

Note: Verify the presence of the O-ring (3) on the


red fitting before to perform the reassembling.

18
Disassembling Instruction V01

Note: Verify the presence of the three O-rings (4) on


the bath before to perform the reassembling.

4 4

Important note: Screw back the four


fixation screws respecting a torque 50cNm
to 50cNm

19
Disassembling Instruction V01

k. WBC Counting Head

Removal
Action Steps Reference

Prerequisite 1. Drain the instrument from any liquids.

2. Switch off the instrument and disconnect the


main power cord.

3. Open the fluidic door.

4. Remove the WBC electrode.

5. Remove the WBC counting bath.

WBC 6. Remove the two fixations screws (1) of the WBC


Counting counting head.
Head

7. Remove the WBC counting head using pliers to


pull it if necessary.

20
Disassembling Instruction V01

8. Replace the WBC counting head following the


reverse instruction.

Note: Verify the presence of the O-ring (2) before to


perform to the reassembling.

40cNm

Important note: Screw back the two


fixation screws, respecting a torque
to 40cNm.

21
Disassembling Instruction V01

l. Counting Manifold

Removal
Action Steps Reference

Prerequisite 1. Drain the instrument from any liquids.

2. Switch off the instrument and disconnect the


main power cord.

3. Open the fluidic door.

4. Remove the complete equipped RBC bath.

5. Remove the complete equipped WBC counting


bath.

Counting 6. Remove the two nuts (1) which maintain the


Manifold RBC and WBC coaxial connectors on the
counting manifold. 1

7. Once the two nuts removed, push on the coaxial


connectors to remove them from the counting
manifold.

8. Remove the ground terminal fixation screw (2).

22
Disassembling Instruction V01

9. Disconnect tubings #1; 2A; 7; 17 and 26B from


the counting manifold.

10. Remove the three fixation screws (3) of the


counting manifold then, remove it. 3
3
11. Equip the new counting manifold with all
elements and reassemble it following the
reverse instruction.

23
Disassembling Instruction V01

m. Counting Valves

Removal
Action Steps Reference

Prerequisite 1. Drain the instrument from any liquids.

2. Switch off the instrument and disconnect the


main power cord.

3. Open the fluidic door.

Counting 4. Disconnect the valve connector.


Valves

5. Remove the two fixation screws (1) of the valve


solenoid.
1
6. Remove the solenoid from the valve body.
1

7. Remove the two fixation screws (2) of the valve


body. 2

8. Remove the valve body. 2

9. Install the new valve following the reverse


instruction.

Important notes:

Be careful to the mounting direction


of the valve body.
The “NC” inscription must be
oriented through the bottom.

24
Disassembling Instruction V01

Tighten the two fixation screws of


the valve body, respecting a torque
to 40cNm. 40cNm

Be careful to the mounting direction


of the valve solenoid.
The solenoid connector must be
oriented through the right.

Tighten the two fixation screws of


the valve solenoid, respecting a
torque to 25cNm. 25cNm

25
Disassembling Instruction V01

n. Counting Valves Module

Removal
Action Steps Reference

Prerequisite 1. Drain the instrument from any liquids.

2. Switch off the instrument and disconnect the


main power cord.

3. Open the fluidic door.

4. Remove the reagent metallic plate.

Counting 5. Disconnect the tubings from valves #1 and #2.


Valves
Module

6. Disconnect the two valves connectors.

7. Remove the two fixation screws (1) of the


counting valves module and remove it.
1
8. Reassemble the counting valves module
following the reverse instruction

26
Disassembling Instruction V01

o. Syringe Valves

Removal
Action Steps Reference

Prerequisite 1. Drain the instrument from any liquids.

2. Switch off the instrument and disconnect the


main power cord.

3. Open the fluidic door.

Counting 4. Disconnect the valve connector of the valve you


Valves want to remove

5. Remove the two fixation screws (1) of the valve


solenoid.
1
6. Remove the solenoid from the valve body.

7. Remove the two fixation screws (2) of the valve


body.
2
8. Remove the valve body.

9. Install the new valve following the reverse


instruction.
2

Important notes:

Be careful to the mounting direction


of the valve body.
The “NC” inscription must be
oriented through the back.

27
Disassembling Instruction V01

Tighten the two fixation screws of


the valve body, respecting a torque
to 40cNm.
40cNm

Be careful to the mounting direction


of the valve solenoid.

Tighten the two fixation screws of


the valve solenoid, respecting a
torque to 25cNm.

25cNm

28
Disassembling Instruction V01

p. Syringe Valves Module

Removal
Action Steps Reference

Prerequisite 1. Drain the instrument from any liquids.

2. Switch off the instrument and disconnect the


main power cord.

3. Open the fluidic door.

4. Remove the reagent metallic plate.

Syringe 5. Disconnect all the tubings (excepted tub#27and


Valves #16) from the syringe valves module.
Module

6. Disconnect the nine valves connectors.

1 1
7. Remove the four fixation screws (1) of the
syringe valves module and remove it.

8. Reassemble the counting valves module 1 1


following the reverse instructions.

29
Disassembling Instruction V01

q. Syringe body equipped

Removal
Action Steps Reference

Prerequisite 1. Drain the instrument from any liquids.

2. Switch off the instrument and disconnect the


main power cord.

3. Open the fluidic door.

Syringe 4. Remove the four fixation screws (1) maintaining


Body the syringe body. 1 1
Equipped
5. Pull the assembly out and remove it from the
syringe module.

1 1

6. Replace the part you want change.

Sampling piston (2).


Lyse piston (3).
Diluent piston (4).
Syringe body (5).
5

3 4

30
Disassembling Instruction V01

7. Reassemble the module following the reverse


instruction. 6 6

Note: Before to reassemble, verify the presence of


the four O-rings (6). 6

IMPPORTANT NOTE: Tighten the


four fixation screws of the syringe
body, respecting a torque to 60cNm. 60cNm

31
Disassembling Instruction V01

r. Syringe motor holder.

Removal
Action Steps Reference

Prerequisite 1. Drain the instrument from any liquids.

2. Switch off the instrument and disconnect the


main power cord.

3. Open the fluidic door.

4. Remove the reagent metallic plate.

5. Remove the syringe body equipped.

Syringe 6. Disconnect the tubings # 6; 9; 12 and 15 from


Motor the module.
Holder

7. Push down the piston carrier in order to access


to the upper module fixation screws.

8. Remove the four fixation screws (1) maintaining


the syringe motor holder.

9. Pull the assembly and remove it.

1 1

1 1

32
Disassembling Instruction V01

10. Remove the ground cable fixation screw (2).

11. Reassemble the module following the reverse


instruction.

s. Syringe Module Sensor.

Removal
Action Steps Reference

Prerequisite 1. Switch off the instrument and disconnect the


main power cord.

2. Remove the reagent metallic plate.

3. Disconnect the sensor connector (1).


Waste
Syringe 4. Remove the sensor fixation screw (2).
2
Sensor
5. Pull the sensor and remove it.

6. Replace the sensor following the reverse


instruction.

33
Disassembling Instruction V01

t. Waste Syringe body equipped.

Removal
Action Steps Reference

Prerequisite 8. Drain the instrument from any liquids.

9. Switch off the instrument and disconnect the


main power cord.

10. Open the fluidic door.

Waste 11. Remove the four fixation screws (1) maintaining


Syringe the waste syringe body.
Body 1 1
12. Pull the assembly and remove it from the waste
Equipped
syringe module.

1 1

13. Replace the part you want change.

Waste piston (2)


Syringe body (3).

34
Disassembling Instruction V01

14. Reassemble the module following the reverse


instruction.

Note: Before to reassemble the waste syringe


body, verify the presence of the twelve O-rings (4). 4

IMPPORTANT NOTE: Tighten the


four fixation screws of the waste
syringe body, respecting a torque to 60cNm
60cNm.

35
Disassembling Instruction V01

u. Waste Syringe Motor Holder.

Removal
Action Steps Reference

Prerequisite 12. Drain the instrument from any liquids.

13. Switch off the instrument and disconnect the


main power cord.

14. Open the fluidic door.

15. Remove the reagent metallic plate.

16. Remove the syringe body equipped.

Waste 17. Disconnect the tubings # 5B; 5C; 14A; 18; 19;
Syringe 24; 29 from the module.
Motor
Holder

18. Push down the piston carrier in order to access


to the upper module fixation screws.

19. Remove the four fixation screws (1) maintaining


the waste syringe motor holder.

20. Pull the assembly and remove it.

1 1

1 1

36
Disassembling Instruction V01

21. Remove the ground cable fixation screw (2).

22. Reassemble the module following the reverse


instruction.

v. Waste Syringe Module Sensor.

Removal
Action Steps Reference

Prerequisite 1. Switch off the instrument and disconnect the


main power cord.

2. Remove the reagent metallic plate.

Waste 3. Disconnect the sensor connector (1).


Syringe
Module 4. Remove the sensor fixation screw (2).
Sensor
5. Pull the sensor and remove it.

6. Replace the sensor following the reverse 2


instruction.

37
Disassembling Instruction V01

w. Connection Board.

Removal
Action Steps Reference

Prerequisite 1. Switch off the instrument and disconnect the


main power cord.

2. Remove the reagent metallic plate.

Connection 3. Remove the three fixation screws (1) of the


Board connection board.
1

38
Disassembling Instruction V01

4. Disconnect the two connectors (2) from the


connection board.

5. Replace the board following the reverse


instruction.

39
Disassembling Instruction V01

x. Q7 Micro Processor Board.

Removal
Action Steps Reference

Prerequisite 1. Switch off the instrument and disconnect the


main power cord.

2. Remove the front cover.

Q7 3. Remove the two fixation screws (1) of the


Micro board.
Processor 1
Board
1

4. Lift the right side of the board first and pull it to


the right.

5. Replace the board following the reverse


instruction.

40
Disassembling Instruction V01

y. CPU Board.

Removal
Action Steps Reference

Prerequisite 1. Switch off the instrument and disconnect the


main power cord.

2. Remove the front cover.

Q7 3. Remove the seven fixation screws (1) of the


Micro board support.
Processor
4. Remove the shielding metallic plate (2), then
Board
the CPU/LCD assembly.
1 1

5. Disconnect the six connectors (3) from the CPU


board.
3
3 3

3 3

Note: Concerning the touch screen connector, push


the clip of the connector onto the left before to pull
the ribbon cable.

6. Remove the five fixation screws (4) of the CPU,


then remove the board from its support.
4 4 4
7. Replace the CPU board following the reverse
instruction.
4 4

41
Disassembling Instruction V01

z. LCD screen

Removal
Action Steps Reference

Prerequisite 1. Switch off the instrument and disconnect the


main power cord.

2. Remove the front cover.

Q7 3. Disconnect the three LCD connectors (1) from


Micro the CPU board.
Processor
Board

Note: Concerning the touch screen connector,


push the clip of the connector onto the left
before to pull the ribbon cable.

4. Release the two LCD’s fixation screws (2)


located on the right. 2
3

5. Remove the two LCD’s fixation (3) screws


located on the left.

6. Remove the LCD screen and replace it following 3


2
the reverse instruction.

42
Disassembling Instruction V01

aa. Front USB Board.

Removal
Action Steps Reference

Prerequisite 1. Switch off the instrument and disconnect the


main power cord.

2. Remove the front cover.

Front 3. Disconnect the cable (1) from the USB board.


USB 1
Board 4. Remove the two fixation screws (2) of the USB
board.

5. Replace the board following the reverse


instruction.

2 2

bb. ON/OFF Switch Board.

Removal
Action Steps Reference

Prerequisite 1. Switch off the instrument and disconnect the


main power cord.

2. Remove the front cover.

3. Remove the LCD screen/CPU assembly.

ON/OFF 4. Disconnect the cable (1) from the board.


2 2
Switch
Board 5. Push the plastic retainer (2) towards one
another and pull the board to remove it.

6. Replace the board following the reverse


instruction.
1

43
Disassembling Instruction V01

cc. Start Key Switch.

Removal
Action Steps Reference

Prerequisite 1. Switch off the instrument and disconnect the


main power cord.

2. Remove the front cover.

ON/OFF 3. Push manually on the start key (between the


Switch two clips) to unclip it.
Board
4. Remove the start key.

5. Push on the start key switch body to remove it.

Note: You can use a torx T10 screw driver to


remove the switch.

6. Replace the start key switch following the


reverse instruction.

44
Disassembling Instruction V01

dd. Fluidic door switch

Removal
Action Steps Reference

Prerequisite 1. Switch off the instrument and disconnect the


main power cord.

2. Remove the front cover.

3. Remove the reagent metallic plate.

Fluidic 4. Disconnect the fluidic door switch connector


Door (1). 1
Switch

5. Release the two fixation screws (2) of the fluidic


door switch and remove it.

6. Replace the fluidic door switch following the 2


reverse instruction.

IMPORTANTE NPOTE: Re-place the new switch in


the same position than the previous one and verify
the operating in “Check sensor” screen opening and
closing the fluidic door.

45