Installation Manual of AJ15 US Dental Chair: P/N:8021151 Version A Date 2017-08-01
Installation Manual of AJ15 US Dental Chair: P/N:8021151 Version A Date 2017-08-01
com/
Installation Manual
of AJ15 US Dental Chair
P/N:8021151 Version:A Date:2017-08-01
Installation manual of AJ15 US dental chair
Copyrights
CONTENTS
1 Introduction……………………………………………………………1
1.1 Products information…………………………………………1
1.2 Symbols explanation……….…………………………………1
2 Composition……………………………………..………………………3
3 Parameters………………..........………………………………………4
3.1 Specifications……………………………………………………4
3.2 Si ze . …………………………………………………………4
4 Preparations………..……………………………………………………5
4.1 Attentions...…...……………………………………………………5
4.2 Tools recommended………………..…………….…………….6
4.3 Installation requirements………………………………………..7
5 Installation…………………………………………………………7
5.1 Place the chair ...………………………………………………7
5.2 Fix the chair………………………………………………………8
5.3 Install the backrest………………………………………………9
5.4 Install the headrest ……………………………………………10
5.5 Install backrest upholstery……...………………………………11
5.6 Install seat upholstery………………………………………12
6 Operations……………………………………………………………13
6.1 Power………….………………………………………………13
6.2 Chair positioning…………………………………………14
6.2.1 Operate by foot control………………………………………14
6.2.2 Control panel and function set …………………………………14
6.2.3 System state ……………………………………………………15
6.2.4 Initial settings of the chair……..................................……15
6.2.5 Initial setting description………………...………...…………15
6.2.6 Chair positions description……………………………………18
6.3 Dual articulating headrest……………………………………20
6.4 Armrest …………………………………………………………21
Installation manual of AJ15 US dental chair
1 Introduction
• Model:AJ15US
• Working conditions:
① Ambient temperature:5 ℃~40 ℃;
② Relative humidity:≤ 80%;
③ Air pressure:86 kPa~106 kPa;
④ Nominal voltage:AC110 V±22 V;
⑤ Nominal frequency:50 Hz±1 Hz;
Symbol Explanation
Manufacturer EU representative
1 P/N:8021151
Installation manual of AJ15 US dental chair
Symbol Explanation
0
Original position Chair Down
P Chair Up
Set
2 P/N:8021151
Installation manual of AJ15 US dental chair
2 Composition
AJ15US dental chair (figure 1)combined with new design concepts ,
provide a good balance for usability ,simplicity and comfortability.
1 2 3 4
5 6 7 8 9
5 Foot control
3 P/N:8021151
Installation manual of AJ15 US dental chair
3 Parameters
3.1 Specifications
Load bearing
Patient bearing:Max 300 pounds(135kg)
4 P/N:8021151
Installation manual of AJ15 US dental chair
4 Preparations
4.1 Attentions
Before install the unit, you should:
• Clear all the debris in the room ,and clean the floor thoroughly.
• Apply two manually operated valves for the air and water source.
• Exhaust the dirty water and air before connecting.
• May have no necessity to use all the component described in this document when
installing. Components should be evaluated before installation.
Leather maintenance:
• No organic solvent or alkaline detergent (such as: gasoline、alcohol ,etc.), or the
leather will crack, be smelly or fade;
• No exposure, keep the room in good ventilation(exposure can cause the leather
crack,bad ventilation lead to go mouldy);
• Mention the patient to remove edged tools(like keys)before operation,to avoid
scratch the leather.
5 P/N:8021151
Installation manual of AJ15 US dental chair
Hammer Scissors
Leveling Flat-bladed
instrument Screw driver
Hexagon
Spanner socket
wrench
6 P/N:8021151
Installation manual of AJ15 US dental chair
5 Installation
5.1 Place the chair
7 P/N:8021151
Installation manual of AJ15 US dental chair
4、Remove the cover and mark the installing position , then move the unit away.
Drill on the marked position, press the explosive bolt, then return the unit to its
8 P/N:8021151
Installation manual of AJ15 US dental chair
Figure 2
2
3
1 Metal backrest
5 Push rod
Pass the M6*20 cuntersunk screws through the M6 lock washer , then
fix the push rod head on the backrest by the M6 cuntersunk screws.
9 P/N:8021151
Installation manual of AJ15 US dental chair
Figure 3
10 P/N:8021151
Installation manual of AJ15 US dental chair
Figure 4
1 2 3
1 Metal backrest
2 Fixing shaft
3 Backrest upholstery
11 P/N:8021151
Installation manual of AJ15 US dental chair
A Fixing bracket
B Chair frame
12 P/N:8021151
Installation manual of AJ15 US dental chair
6 Operations
6.1 Power
The power switch of the power box controls power supplying of the
Figure 6
13 P/N:8021151
Installation manual of AJ15 US dental chair
Figure 7
0 Original position
Program Chair
1 Position 1
Program Chair
2 Position 2
P Set
LP Spitting position
Backrest down
Backrest up
Chair up
Chair down
14 P/N:8021151
Installation manual of AJ15 US dental chair
Program Chair
Position 1
Program Chair
Position 2
Program Chair
Position 3
P Set
LP Spitting position
Backrest down
Backrest up
Chair up
Chair down
15 P/N:8021151
Installation manual of AJ15 US dental chair
Press the “O” button on the main control panel of dentist unit to descend the
chair motor till the lower limit switch is triggered, raise the backrest motor till
the higher limit switch is triggered. System initialize automatically ,it can be
normally performed after initialization finished .
1、Press “SET” button about 12 second until the buzzer sound 2 short
beeps(the indicator flashes twice and then illuminates),enter original setup
status; press P1,the buzzer sound 1short beep( P light off),escape setup
status. This steep eliminates the original stroke parameters.
2、Press “SET” button about 12 second until the buzzer sound 2 short beeps,
(the indicator flashes twice and then illuminates),enter initial setup status ,
press the four moving chair button to make backrest up and press “R”, and
then press the four moving chair button again to make chair up to the highest
and the backrest is on the lowest. At this time, press “LP” and then finish the
setting.
16 P/N:8021151
Installation manual of AJ15 US dental chair
Attention:
When setting up the system,do not extend the maximum range of the
cylinder t prevent the hydraulic pump from not working properly.
Process P or or P
• Move the chair and backrest to the working position, and adjust dental light
to the desired condition ( On or Off), press the “P” button(3s),the
indicator light is on (beep 1 time),( loose button so as not to enter the
original setting state),enters the setup procedure.
• Press the “P1” button ,and then press the “P “button one time
again,(beep1-2times),setup finished and escape the setup procedure.
• P2, P3 position setup are same step as the P1.
17 P/N:8021151
Installation manual of AJ15 US dental chair
NOTE: If you press the button continuously, when the chair backrest
reach the limit position, it will automatically stop the movement.
b. Backrest Down
Press this button continuously and the backrest will move downpwards. When the
backrest reach the desired position, stop pressing the button and the backrest will
stop immediately.
NOTE: If you press the button continuously, when the chair backrest
reach the limit position, it will automatically stop the movement.
c. Chair up
Press this button continuously and the chair will move upwards. When the backrest
reach the desired position, stop pressing the button and the backrest will stop
immediately.
NOTE: If you press the button continuously, when the chair backrest
reach the limit position, it will automatically stop the movement.
d. Chair down
Press this button continuously and the chair will move downwards. When the backrest
reach the desired position, stop pressing the button and the backrest will stop
immediately.
NOTE: If you press the button continuously, when the chair backrest
reach the limit position, it will automatically stop the movement.
18 P/N:8021151
Installation manual of AJ15 US dental chair
e. Original position 0
Pr e ss t h e b ut t o n, t h e c h a i r a n d ba c kr e st wi l l m o v e t o t h e o r i g i n a l p o s i t i o n
automatically(Original position : backrest highest , the chair lowest) and light off . Press
this button again , the chair returns to former position and light on. The chair shall stop at
a new position by halfway stop.
f. Spitting button
Press the button, the chair and backrest will move to spitting position
automatically(Spitting position : backrest highest , the chair is still) and light off . Press
this button again , the chair returns to former position and light on. The chair shall stop at
a new position by halfway stop.
19 P/N:8021151
Installation manual of AJ15 US dental chair
A Headrest
B Quick-release button
C Glidebar
NOTE:
Please fix the headrest after adjustment to avoid patient’s head not
being settled.
20 P/N:8021151
Installation manual of AJ15 US dental chair
6.4 Armrest
The armrest is easy to operate ,which provides convenience to users .
Press the button you can rotate the armrest forward or backward .
Figure11 Armrest
positioning
Button
21 P/N:8021151
Installation manual of AJ15 US dental chair
6.5 Chair rotation
Tighten
Loosen
22 P/N:8021151
Installation manual of AJ15 US dental chair
The special design of upholstery is easy to remove and replace .They are
installed on backrest, seat and headrest of the chair .
Backrest upholstery
To take off the upholstery from backrest ,you should hold the bottom
edge of gasket firmly and lift it up (1 inch or 25 mm) ,then lift the
upholstery from the backrest of the chair ( Figure15).
To reinstall the upholstery,you can put the keyhole of gasket on nail
head, then push till the gasket is fixed.
23 P/N:8021151
Installation manual of AJ15 US dental chair
Seat upholstery
Loose the screw at the front end of the upholstery, push the upholstery
forward,then remove the seat upholstery.
24 P/N:8021151
Installation manual of AJ15 US dental chair
8 Reserved lines
Swing Type
A B C
1
A 21
32
43
54
65 1. 1/8" white , reserved for dentist unit
76
87 2. 1/8" brown,reserved for dentist unit
8 3. 1/8" red, reserved for dentist unit
9
10 4. 1/8" orange, reserved for dentist unit
11 5. 1/8" black, reserved for dentist unit
6. 1/8" blue,reserved for dentist unit
12
7. 1/8" yellow,reserved for dentist unit
13 8. 1/8" green,reserved for dentist unit
14
9. 1/4" black,reserved for dentist unit
15
10. 1/4" blue,reserved for dentist unit
11. 1/4" red,reserved for dentist unit
C 12. Control wire for delivery head
16 13. 24V, 7 core power cord for delivery head
14. 24V power cord for dental light
17 15. Ground wire for delivery head
16. 1/4" green, reservd for assistant's control
18
17. 1/4" yellow, reservd for assistant's control
18. Control wire for assistant control
B. Drainage pipe
25 P/N:8021442
Installation manual of AJ16US dental chair
Lines
Lines offorOTP
OTP version
version
9 Connecting diagrams
Connecting diagram(Electric system)
27 P/N:8021151
Installation manual of AJ15 US dental chair
28 P/N:8021151
We reserve the right to make any alterations which may be due to
technical improvements.