Instrumentation Lab: Malla Reddy Engineering College (Autonomous) Department of Mechanical Engineering
Instrumentation Lab: Malla Reddy Engineering College (Autonomous) Department of Mechanical Engineering
(Autonomous)
Department of Mechanical Engineering
Instrumentation Lab
OBSERVATION MANUAL
FOR
ROLL NO.:
SECTION:
1
MALLA REDDY ENGINEERING COLLEGE
(Autonomous)
Department of Mechanical Engineering
Instrumentation Lab
List of Experiments
APPARATUS:
1. Standard pressure gauge
2. Pressure cell
3. Digital pressure indicator
4. Foot air pump
THEORY:
Pressure is defined as force per unit area and is measured in Newton per square meter (Pascal)
or in terms of an equivalent head of some standard liquid (mm of mercury). A typical pressure gauge
will measure the difference in pressure between two pressures. Thus a pressure gauge is connected to
the hydraulic line and the gauge itself stands in atmospheric pressure. The gauge reading will be the
difference between the air pressure and the atmospheric pressure and is called gauge pressure. The
absolute pressure (the actual pressure within the air line) is the sum of the gauge pressure and
atmospheric pressure.
The material most commonly used for manufacture of diaphragms are steel, phosphor bronze,
nickel silver and beryllium copper. The deflection generally follows a linear variation with the
diaphragm thickness.
SPECIFICATIONS:
Sensor : Strain gauges bonded on steel diaphragm for pressure measurement
Type : Diaphragm
Range : 10 Kg / cm2
Excitation :10V D.C
Accuracy and linearity : 1%
Display : 3.5 Digit LED display of 200 mV FSD
Power supply :230 V +/- 10% 50Hz
DIAGRAM:
Pressure C shaped
3 Transducer Display
Foot air pump Bourdon
Vessel
tube
Fig: Block Diagram
PROCEDURE:
1. Check connection made and switch ON the instrument by rocker switch at the front panel.
2. Display glows to indicate the instrument is ON.
3. Allow the instrument in ON position for 10 min for initial warm up.
4. Select the mode of display in MM through FUNCTION SELECT switch provided on the front
panel
5. Adjust the ZERO potentiometer on the front panel till the display reads “000”.
6. Close the release valves fixed to the pressure vessel and apply the pressure on the sensor using
foot air pump.
7. The instrument reads the pressure coming on the sensor and displays through LED. Readings can
be tabulated and the % error of the instrument, linearity can be calculated.
OBSERVATION TABLE:
4
S.No Actual pressure Indicator Error % Error
(Kg/cm2) reading
(Kg/cm2)
GRAPHS:
APPLICATIONS:
RESULT:
5
CALIBRATION OF MCLEOD GAUGE
AIM:
To calibrate the given vacuum cell. (Low Pressure Cell)
APPARATUS:
Vacuum pump
Vacuum chamber
McLeod gauge setup
THEORY:
In everyday usage, vacuum is a volume of space that is essentially empty of matter, such
that its gaseous pressure is much less than atmospheric pressure. The word comes from
the Latin for “empty”. A perfect vacuum would be one with no particles in it at all, which
is impossible to achieve in practice. Physicists often discuss ideal test results that would
occur in a perfect vacuum, which they simply call “vacuum” or “free space”, and use the
term partial vacuum to refer to real vacuum. The Latin term in vacuum is also used to
describe an object as being in what would otherwise be a vacuum.
Vacuum is useful in a variety of processes and devices. Its first widespread use
was in the incandescent light bulb to protect the filament from chemical degradation. The
chemical inertness produced by a vacuum is also useful for electron beam welding, cold
welding, vacuum packing and vacuum frying. Ultra-high vacuum is used in the study of
atomically clean substrates, as only a very good vacuum preserves atomic-scale clean
surfaces for a reasonably long time (on the order of minutes to days). High to ultra-high
vacuum removes the obstruction of air, allowing particle beams to deposit or remove
materials without contamination. This is the principle behind chemical vapor deposition,
physical vapor deposition, and dry etching which are essential to the fabrication of semi
conductors and optical coatings, and to surface science. The reduction of convection
provides the thermal insulation of thermos bottles. Deep vacuum lowers the boiling point
of liquids and promotes low temperature out gassing which is used in freeze drying,
adhesive preparation, distillation, metallurgy, and process purging. The electrical
properties of vacuum make electron microscopes and vacuum tubes possible, including
cathode ray tubes. The elimination of air friction is useful for flywheel energy storage
and ultracentrifuges.
Boyle’s Law: For a fixed mass of gas at constant temperature, the volume of the gas is
inversely proportional to the pressure. This is only strictly true foe ideal gases (it can be
derive from the ideal gas equation), but it remains a close approximation for real gases. It
6
is one of the earliest scientific laws, and remains useful to this day.
PROCEDURE:
1. Rotate the Mcleod gauge horizontally so that the mercury is collected completed
inside the reservoir.
2. Connect the vacuum pump to the hose nipple provided at the back of the gauge.
3. Start the vacuum pump and it starts absorbing all the air inside the gauge.
4. Rotate the gauge slowly so that the mercury fills the two manometers. You can
notices that one arm of the manometer is closed (reading end) and other is opened to
the vacuum chamber (reference end). Tilt the gauge so that the mercury level touches
exactly the ZERO mark in the reference end of the manometer.
7
5. The mercury level in the reading end will show the vacuum built inside the vacuum
chamber. Vacuum can be measured from 0.01 to 10mm Hg
RESULT:
8
CALIBRATION OF RESISTANCE TEMPERATURE DETECTOR FOR
TEMPERATURE MEASUREMENT
AIM:
To calibrate the given RTD by using Thermometer
APPARATUS:
Temperature sensor (RTD), electrically heated kettle with water , thermometer signal
conditioning and amplifying and analog to digital converter in singal unit power supply
THEORY:
Resistance Temperature Detector (RTD)
Resistance thermometers, also called resistance temperature detectors or resistive
thermal devices (RTD), are temperature sensors that exploit the predictable change in
electrical resistance of some materials with changing temperature. As they are almost
invariably made of platinum, they are often called platinum resistance thermometers
(PTR). They are slowly replacing the use of thermocouples in many industrial
applications below 600oC, due to higher accuracy and repeatability. There are many
categories like carbon resistors, film and wire wound types are the most widely used.
Carbon resistors are widely available and are very inexpensive. They have very
reproducible results at low temperatures. They are the most reliable from at extremely
low temperatures. They generally do not suffer from significant hysteresis or strain gauge
effects.
Film thermometer have a layer of platinum on a substrate, the layer may be extremely
thin, perhaps one micrometer. Advantages of this type are relatively low cost and fast
response. Such devices have improved performance although the different expansion
rates of the substrate and platinum give “strain gauge” effects and stability problems.
Wire – wound thermometers can have greater accuracy, especially for wide
temperature ranges.
Coil elements have largely replaced wire wound elements in industry. This design has a
wire coil which can expand freely over temperature, held in place by some mechanical
support which lets the coil keep its shape.
POWER: 3 pin main cable is provided with the instrument .connect the 3 pins socket to
the instrument at the rear panel and to the AC main 230 V supply .
Connect the kettle to 230V supply with cable supplied .
NOTE: Before connecting ensure the voltage is 230V and power switch is in OFF
position .
PROCEDURE:
1. Check connections made and switch on the instrument by rocker switch at the front panel.
The display glows to indicate instrument is ON
2. Allow the instrument in ON position for 10 minutes for initial warm-up
3. Pore around ¾ full of water to the kettle and place sensor and thermometer inside the kettle
4. Note down the initial water temperature from the thermometer.
5. Select the sensor on which the experiment to be conducted through selection switch on the
front panel
6. Adjust the initial set potentiometer in the front panel till the display reads initial water
temperature
7. Switch on the kettle and wait till the water boils note down the reading of thermometer ans
set final potentiometer till display reads boiling temperature
8. Remove the sensor from the boiling water immerse it in cold water .set the cold water
temperature using initial set potentiometer.
9. Repeat the process till the display reads exact boiling water and cold water temperature.
10. Change the water in the kettle and re-heat the water .Now the display starts showing exact
temperature raise in the kettle
11. Experiment can be repeated for all three sensors .Temperature in the thermometer and the
indicator readings in steps of 10 degrees can be tabulated.
10
GRAPHS:
11
CALIBRATION OF THERMISTOR FOR TEMPERATURE MEASUREMENT
AIM:
APPARATUS:
Temperature sensor (thermistor), Heating coil to heat water in water bath, Digital
temperature Indicator & Thermometer.
THEORY:
Thermisor (semiconductor resistance sensor) when a semiconductor material is subjected
to increase in temperature, its resistance decrease. A resistance thermometer which uses
semiconductor as sensing element is called thermistor.
Thermisor have negative temperature coefficient of resistance in contrast to the positive
temperature coefficient of resistance exhibited by RTD. The variation of resistance with
temperature is exponential and is given by .
POWER: 3 pin main cable is provided with the instrument .connect the 3 pins socket to
the instrument at the rear panel and to the AC main 230 V supply .
Connect the kettle to 230V supply with cable supplied .
NOTE: before connecting ensure the voltage is 230V and power switch is in OFF
position .
PROCEDURE:
1. Check connections made and switch on the instrument by rocker switch at the front panel.
The display glows to indicate instrument is ON
2. Allow the instrument in ON position for 10 minutes for initial warm-up
3. Pore around ¾ full of water to the kettle and place sensor and thermometer inside the kettle
4. Note down the initial water temperature from the thermometer.
5. Select the sensor on which the experiment to be conducted through selection switch on the
front panel
6. Adjust the initial set potentiometer in the front panel till the display reads initial water
temperature
7. Switch on the kettle and wait till the water boils note down the reading of thermometer ans
set final potentiometer till display reads boiling temperature
8. Remove the sensor from the boiling water immerse it in cold water .set the cold water
temperature using initial set potentiometer.
9. Repeat the process till the display reads exact boiling water and cold water temperature.
10. Change the water in the kettle and re-heat the water .Now the display starts showing exact
temperature raise in the kettle
11. Experiment can be repeated for all three sensors .Temperature in the thermometer and the
indicator readings in steps of 10 degrees can be tabulated.
13
OBSERVATION AND TABULAR COLUMN:
S.NO Thermometer Thermistor Indicator Error(3-2) % Error
Reading Reading
14
CALIBRATION OF THERMOCOUPLE FOR TEMPERATURE MEASUREMENT
AIM:
To calibrate the given Thermocouple by using Thermometer
APPARATUS:
Temperature sensor (Thermocouple), Heating coil to heat water in water bath, Digital
temperature Indicator & Thermometer.
THEORY:
A thermocouple is a sensor used to measure temperature in a number of processes.
Thermocouples consist of two wire legs made from dissimilar metals which are fixed together at
one end, creating a junction. When this junction experiences a change in temperature a voltage
is created, this voltage can then be measured and referenced back to the temperature.
Thermocouples are available in different combinations of metals, each with its own unique
characteristics in terms of temperature range, durability, vibration resistance, chemical
resistance, and application compatibility.
• Base metal thermocouples or type J, K, T, & E are relatively low cost and therefore the most
popular thermocouples, they are commonly used in a broad range of low to medium temperature
applications. (-185 to 1100°C)
• Noble metal thermocouples or type R, S, and B have greater resistance than base metal
thermocouples, however they have platinum conductors, making them far more expensive. They
are commonly used in high temperature applications. (Up to 1600°C)
SPECIFICATIONS :
POWER : 3 pin main cable is provided with the instrument .connect the 3 pins socket to
the instrument at the rear panel and to the AC main 230 V supply .
15
Connect the kettle to 230V supply with cable supplied .
SENSORS : connect RTD , Thermistor And Thermo Couple to the connector on the rear
panel .
NOTE : before connecting ensure the voltage is 230V and power switch is in OFF
position .
PROCEDURE:
1. Check connections made and switch on the instrument by rocker switch at the front panel.
The display glows to indicate instrument is ON
2. Allow the instrument in ON position for 10 minutes for initial warm-up
3. Pore around ¾ full of water to the kettle and place sensor and thermometer inside the kettle
4. Note down the initial water temperature from the thermometer.
5. Select the sensor on which the experiment to be conducted through selection switch on the
front panel
6. Adjust the initial set potentiometer in the front panel till the display reads initial water
temperature
7. Switch on the kettle and wait till the water boils note down the reading of thermometer ans
set final potentiometer till display reads boiling temperature
8. Remove the sensor from the boiling water immerse it in cold water .set the cold water
temperature using initial set potentiometer.
9. Repeat the process till the display reads exact boiling water and cold water temperature.
10. Change the water in the kettle and re-heat the water .Now the display starts showing exact
temperature raise in the kettle
11. Experiment can be repeated for all three sensors .Temperature in the thermometer and the
indicator readings in steps of 10 degrees can be tabulated.
16
CALIBRATION OF LVDT
AIM: - Study and calibration of LVDT (linear variable differential transformer) transducer for
Displacement measurement.
THEORY: -
Differential transformer which work on a variable inductance principle or also used to measure
displacement the most popular variable inductance transformer for linear displacement is the
LVDT.
17
SPECIFICATIONS:
RANGE: 10 mm
LINEARITY: 1%
DISPLAY: 3-DIGIT seven segment red LED display to read upto 1999 counts
.200mv for full scale deflection
SENSTIVITY : 0.1 mm
REPATIBILITY: 99%
OPERATION PROCEDURE:-
3 PIN power chord is provided with an instrument connect the 3 –pin plug to 230V 50HZ
power supply
Before connecting the cable ensure that the power switch is in OFF position
6 Core shielded Cable is connected to the LVDT with male connector of different colors
fixed to each wire. Connect the male pins sockets matching the colors
After checking the entire connections switch on instruments by passing down the Toggle
switch. The display glows to instrument is ON.
Allow the instrument is ON position for 10 min that initial warm up rotate the micro
meter to read 200mm
Adjust the potentiometer at the Front pond so the the display reads 10.0
18
Bring the micrometer back to read20.0 once again the adjust the potentiometer to read the
display 10.0 .
Move the core in by operating the micro meter in step 1 or 2mm tabulate the reading
micrometer reading will be actual reading and display will give indicator reading plot the
graph between actual & indicator reading.
OBSE5RVATON:-
19
CALIBRATION OF CAPACITIVE TRANSDUCER FOR ANGULAR DISPLACEMENT
APPARATUS:
1. Sensor: Parallel plate capacitor/Angular plate capacitor.
2. Capacitate transducer trainer.
THEORY:
Principle of operation of capacitate transducer is based upon familiar equation for
Capacitance of parallel plate capacitor
A
C=ε
d
Where,
C = Capacitance
ε r = Relative permittivity
These changes are caused by physical variables such as displacement, pressure and force
in most of the cases. The change in capacitance may be caused by change in dielectric
constant as in case of measurement of liquid or gas levels. The principle of change in
capacitance with change in over lapping area can be employed for measurement of
Angular displacement.
20
Fig: Angular displacement capacitance
SPECIFICATIONS:
ACCURACY :5-10 %
DISPLAY :3.5 DIGIT LED DISPLAY TO READ +/- 1999 COUNTS FOR
+/- 200 MV fsd
PROCEDURE:
Allow the instrument in ON position for 10 min for initial warm up.
Move the plate in step of 5 to 10mm and note down the reading in the tabular column till 50mm.
RESULT:
AIM: -To find the actual strain using strain Gauge setup
THEORY: -
Transducer Converts energy from one form to another form. Strain gauge in a kind of transducer.
Its used to sense strain due to load pressure vibrations etc. property of strain gauge is used to
measure the strain. Gauge factor (G.F) refer to sensitivity of strain gauge. Its defined as the ratio
of unique change in its resistance to unit change in its length.
G.F=(AR/R)/(ΔL/L).
Our setup consists of a cantilever beam with a provision to load up to 1 Kg signal and a
digital strain indicator. Strain indicator conditions and amplified strain gauge signal strain gauge
is bounded on the cantilever beam and is connected in form of Wheatstone bridge. The strain
indicator is provided with zero balancing facility through adjustable potentiometer.
Strain gauge converts mechanical quantity into electrical quantity. Wheatstone bridge
converts change in resistance into change in voltage amplifier with again 10 to 1000 are used to
increase their signals to levels where they are compatible with voltage measuring devices.
RECORDER: -
They are voltage measuring devices they are used to display the measurement in a form that can
be read and interpreted digital/analog voltmeter are used to measure static voltages.
DIAGRAM:-
22
SPECIFICATIOINS: -
Cantilever beam: -
Thickness: 2.5mm
23
OBSERVATION:-
OPERATING PROCEDURE:
Connect the 3-pin power supply to 230V power supply and to the instrument at the rear
panel
Connect the strain gauge the terminals at the panel as mentioned below
Check the connections made switch ON the instruments by toggle switch at the back of
panel box. The display glows to indicate the instrument is ON
Select the bridge configuration through rotary switches on the panel
Allow the instrument ON position for about 10min for initial warm up
Adjust the zero potentiometer on the panel till display read ‘000’
Apply the load using loading arrangement in steps of 100 gm to 11Kg
Strain will be displayed in micro strain for full bridge total Strain will be displayed and
for half bridge half of the strain will be displayed .for half bridge double the treading to
get strain in micro strain
Note down the reading in tabular form calculate the theoretical strain and find the error
Draw the graph taking actual on x- axis, indicated on y-axis
SPECIMEN CALCULATONS:
M/I = F/Yy
F = (M×Y)/I
e= 3.77x100
RESULT:
25
SPEED MEASUREMENT CALIBRATION OF MAGNETIC PICK UP AND
AIM:- Study and calibration of speed Measurement by using photo reflective and magnetic
pickup.
APPARATUS:- Speed measurement Trainer consisting regulation to select the speed photo and
magneto Pickup sensor LED Display digital RPM trainer Contacts/Non contact tachometers
system. Works at 6VDC derived from 230V AC 50HZ
This consists of
THEORY:-
The shaft of the tachometer is connected to the speed source. The disc starts
rotating. As the disc rotates, the light passes through the hole for some time and the rest is
obstructed by the opaque disc.
Hence, only when a hole is present, the light falls on the light sensor. Thus the
intermittent light falling on the photocell produces voltage pulses. The frequence of this
voltage pulses is a measure of the speed of the disc and hence, the speed of the light
source.The output is obtained on a frequency measuring unit.
26
OBSERVATION:-
PROCEDURE:-
Before Switching on the Experiment Instruments Ensure that the connections are made
properly.
Switch on the Instruments by pushing down the toggle Switch provided at the rear side of
the box LED display glow to indicate the Instrument is ON.
Allow the Instrument for 10 min in ON position for Initial Warmup
Select the sensor with the help of the toggle Switch
Switch on the electronic regulation the fan rotates which will in turn rotates the tone
wheels. The display will start indicating Exact RPM of the motor.
Reading can be tabulated and Sensor Comparison can be made.
Experiment can conduct by tabulating the speed indicated for both sensors at various
speed. The speed can varied by using electronic regulator at the particular speed note
down the display readings for both the sensors selecting through the toggle switch.
Tabulating at least through toggle switch calculating the at least 5 to 10
reading taking the photo reflective as per the RPM at the motor. Accuracy of the
magnetic pickup can be calculated.
RESULT: -
27
ROTA METER FOR FLOW MEASUREMENT
APPARATUS REQUIRED:
Potentiometer
Auto manual switch
Electro servo amplifier
Spin connector to control unit
6. hand valve
7. Reservoir tank
8. Pump
9. Wheel control
THEORY:
The purpose of control system is to make the process flow changes in stem point as
close as possible. This operation is called “ Servo operation “. The disturbance
doesn’t change when the stem point undergoes a change . The feedback control acts
in such a way as to keep the output & process close to stem point. Servo operation is
typically of electrical & mechanical system & has received the most attention in
tests on automatic control.
28
Fig: Rotameter
Specifications:
Timer :3 Digital timer with set point and relay control time set in Sec
Power : 230 V 50 Hz
PROCEDURE:
Fill the collecting tank with water till ¾ of collecting tank and make sure that pump is
immersed in water fully.
Switch on the instrument ON switch. The timer will switch on and indicates the previous
set value. If needed alter the set value using incremental and decrement keys.
Switch on the pump ON switch and press the start button set the flow rate for 30 LPH by
using control knob provided on the rota meter.
Press the start button the panel and the timer start counting and when it reaches zero the
pump be off.
29
Note down the initial water level in cms and final water level in cms for each flow rate
and tabulated in the tabular column.
𝜋
Area of measuring tank : a= 4 (D) 2
RESULT:
30
MEASUREMENT AND STUDY OF SPEED BY USING STROBOSCOPE.
AIM:
Calibration of stroboscope for speed measurement.
a) Speed source contains a wheel with reflective mark. Rpm range; 500-2750.
THEORY
31
3. First find out the speed of wheel with tachometer non-contact type.
4. On the stroboscope, select the range say 150-4500, or if the rpm is more than
4500. Select the button range 4500-18000.
5. Then focus on the rotating wheel and adjust the coarse and fine knobs , till the
rotating object gives a stand still appearance and observe the digital read-out
which displays exact rpm.
PRECAUTIONS:
1. The stroboscope should not be used for more than 30 min continuously, and a
break of 5 minute is essential between two readings.
OBSERVATIONS:
S.NO Voltage Stroboscope Speed = r1rn(n-
reading in rpm 1)/rn-r1
RESULT:
32
MEASUREMENT OF VIBRATION AMPLITUDE OF AN ENGINE BED BY
USING SEISMIC PICKUP AT VARIOUS LOADS.
AIM: To study and use of seismic pickups for the measurement of vibration
amplitude of an engine bed at various loads
APPARATUS:
Shaker/vibrator
Vibrator meter
Power supply
THEORY:
The acceleration may be attached to the spindle through M.S stud and signal
output from accelerometer is connected.
PROCEDURE:
1. Connect the sensor to the instrument through the BNC socket provided in
the back panel mentioned sensor
2. Connect vibration generator to the instrument through cable provided at
real panel of instrument make exciter.
3. Connect instrument to 230 v , 50 Hz supply through cable provided at rear panel.
4. Keep the frequency pot and the volt pot in the minimum position
5. Switch on the instrument, the display glows to indicate the power is ON in
this position pressure tare button to make readings zero.
6. Turn the volt pot to maximum position.
7. Now turn frequency pot in steps of 100Hz and note down the reading of
acceleration, velocity and displacement.
8. Tabulate the reading in the tabular column experiment can be repeated for
different voltage levels settable though pot knob provided.
33
VARIATION AMPLITUDE FORMULAE:
OBSERVATION TABLE
S.NO Frequency Calculated Indicated
RESULT
34