MAN0010407 QuantStudio 3 5 InstallUseMaint UG
MAN0010407 QuantStudio 3 5 InstallUseMaint UG
Systems
INSTALLATION, USE, AND MAINTENANCE
Firmware v1.4
Publication Number MAN0010407
Revision D.0
C.0 December 2015 Updates – describes new features in Firmware v1.2.x, including:
• Updated workflow to create an instrument profile
• Streamlined workflow for connectivity to Thermo Fisher™ Connect
• Improved display of VeriFlex™ Zones
• Plate insert reminder before starting run
• Support for 96-well Fast (0.1 mL) plates
(continued)
Important Licensing Information: These products may be covered by one or more Limited Use Label Licenses. By use of these
products, you accept the terms and conditions of all applicable Limited Use Label Licenses.
Trademarks: All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified. TaqMan is a
registered trademark of Roche Molecular Systems, Inc., used under permission and license. Cy is a registered trademark of GLOBAL
LIFE SCIENCES SOLUTIONS GERMANY GMBH. Microsoft and Windows are trademarks of Microsoft Corporation. Pentium is a
trademark of Intel Corporation. Google and Chrome are trademarks of Google, Inc.
©2021 Thermo Fisher Scientific Inc. All rights reserved.
Contents
4 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Contents
Workflow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Options for running an experiment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Create and run an experiment from a template . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Run an experiment from a saved file . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Repeat the last instrument run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Edit an experiment before starting a run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Enter or edit template properties . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Edit the run method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Define, assign, and view well details . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Load and unload the plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Load and unload a plate in the instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Load and unload a plate in the instrument with a compression pad . . . . . . . . . . . . . . 38
View, pause, or stop a run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
View real-time data and plots on the instrument touchscreen . . . . . . . . . . . . . . . . . . . . 40
Pause or stop an instrument run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Adjust the display of real-time plots on the instrument touchscreen . . . . . . . . . . . . . . 41
Lock the touchscreen during a run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Transfer, view, or manage files and results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Transfer EDS files from the instrument home screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
View run history and delete or transfer files from the instrument . . . . . . . . . . . . . . . . . 42
Manage templates (EDT files) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 5
Contents
Initial start‑up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Manage profiles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Create an administrator instrument profile during initial start-up . . . . . . . . . . . . . . . . . 76
Manage all instrument profiles (Administrator only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Edit an instrument profile . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
™
Link an instrument profile to Thermo Fisher Connect . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Recommended order to set up profiles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
™
Overview of local instrument profiles and Thermo Fisher Connect profiles . . . . . . . 78
™
Thermo Fisher Connect instrument profile roles and functions . . . . . . . . . . . . . . . . . . 79
™
Test the connection to the Thermo Fisher Connect platform . . . . . . . . . . . . . . . . . . . . 79
™
Link the instrument to your Thermo Fisher Connect account . . . . . . . . . . . . . . . . . . . 80
™
Link a local profile to a Thermo Fisher Connect profile . . . . . . . . . . . . . . . . . . . . . . . . . 82
™
Unlink a Thermo Fisher Connect account . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
If you link when you are signed in to the instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
™
Change the PIN for a Thermo Fisher Connect profile . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Enable SAE mode (Administrator only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Require instrument profile sign-in (Administrator only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
6 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Contents
■ APPENDIX B Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 7
Contents
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
8 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
About this guide
Purpose
This guide provides information about installing, using, and maintaining the QuantStudio™ 3 and 5
Real‑Time PCR Systems.
Note: For information and instructions on performing experiments on these systems, see QuantStudio™
Design and Analysis Desktop Software User Guide (Pub. No. MAN0010408).
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 9
1 Product information
Password security
Thermo Fisher Scientific strongly recommends that you maintain unique passwords for all accounts in
use on this product. All passwords should be reset upon first sign in to the product. Change passwords
according to your organization's password policy.
It is the sole responsibility of your IT personnel to develop and enforce secure use of passwords.
10 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 1 Product information
Update the instrument firmware 1
4. If backing up to a USB drive, insert a USB drive into the front-panel USB port.
7. Touch Backup.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 11
Chapter 1 Product information
1 Instrument hardware description
The instrument can be run directly from the touchscreen to create and start experiments. To design
experiments or to analyze data, the instrument can be integrated with the QuantStudio™ Design and
Analysis Software.
An optional barcode scanner and optional wireless adapter can be purchased separately.
FAM™ dye,
x1-m1 Blue 470 ± 15 520 ± 15 SYT09
SYBR™ Green
JOE™ dye,
x2-m2 Green 520 ± 10 558 ± 12 VIC™ dye
HEX™ dye, TET™ dye[2]
x4-m4 Orange 580 ± 10 623 ± 14 ROX™ dye, JUN™ dye Texas Red™ dye
MUSTANG PURPLE™
x5-m5 Red 640 ± 10 682 ± 14 LIZ™ dye
dye, Cy®5
12 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 1 Product information
Instrument hardware description 1
Wavelength
(nm)
Emission
Spectra
Custom dyes
The instrument can run assays designed with custom dyes.
Custom dyes include:
• Dyes that are not supplied by Thermo Fisher Scientific.
• Dyes or formulations of dyes that are not system dyes for the instrument.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 13
Chapter 1 Product information
1 Instrument hardware description
14 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 1 Product information
Software description 1
These independent zones are ideal for real-time PCR optimization or the ability to run multiple
experiments in the same run. Unlike standard gradients which give a sigmoidal temperature curve
across the columns, blocks with VeriFlex™ Zones help deliver accurate temperatures across every zone.
The instrument includes three additional USB ports on the back of the instrument.
Note: The instrument recognizes only one external storage device at a time for data transfer.
Software description
Instrument, desktop, and cloud software features
The instrument software and the QuantStudio™ Design and Analysis Software (desktop and Thermo
Fisher™ Connect) include the features described below.
Thermo Fisher™
Feature Instrument Desktop
Connect
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 15
Chapter 1 Product information
1 Software description
(continued)
Thermo Fisher™
Feature Instrument Desktop
Connect
Start run ✓ ✓ —
Analyze results — ✓ ✓
16 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 1 Product information
Software description 1
Folders in Load Location in which you can store templates (EDT files) on the Instrument
Experiment instrument:
• My Instrument – Displayed if you are signed in
• Public – Location in which all experiments run by guest
users are stored
• USB – USB for manual transfer to and from a computer
• Post Read – If you run an endpoint experiment, the pre-
read experiment is automatically saved in this folder for
post-read analysis
Experiment Template Default settings for an instrument run, can be modified before Instrument
File (EDT) instrument run.
Desktop
Two types:
Thermo Fisher™
• Factory-provided, accessed from Open Template Connect
(instrument) or Create Experiment (desktop or Thermo
Fisher™ Connect)
• User-created, accessed from Load Experiment
Experiment Run File Settings and data for a completed instrument run. Instrument
(EDS)
Desktop
Thermo Fisher™
Connect
Template and
experiment Instrument Desktop Thermo Fisher™ Connect
components
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 17
Chapter 1 Product information
1 Network connection options
(continued)
Template and
experiment Instrument Desktop Thermo Fisher™ Connect
components
Plate Sample names Define and assign samples, targets or SNP assays, and
tasks in the Quick Setup and Advanced Setup panes of the
You cannot edit targets/SNP
Plate tab.
assays or tasks on the instrument.
Run Start and monitor a run in progress Start and monitor a run in Monitor a run in progress
progress (link to Thermo Fisher™
View: time remaining and
Connect Instrument Details)
temperature, method, plots View: time remaining and
temperature, method, plots View: time remaining and
Pause, resume, stop a run
temperature, method, plots
Third-party software
Before installing third-party software on the computer running QuantStudio™ Design and Analysis
Software, confirm that the third-party software will not do the following:
• Restrict Ethernet communication.
• Interfere with instrument or computer operation.
You can connect an instrument to a network or computer in the configurations listed below. For specific
information on networking, see “Networking” on page 95.
LAN
LAN
18 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 1 Product information
Types of runs 1
Types of runs
Purpose Description
Determines absolute 1. The software measures amplification of the target in a standard dilution series and in
target quantity in test samples.
samples. 2. The software generates a standard curve using data from the standard dilution series.
3. The software uses the standard curve to interpolate the absolute quantity of target in
the test samples.
Determines relative 1. The software measures amplification of the target of interest and of an endogenous
target quantity in control target in a standard dilution series, in a reference (calibrator) sample, and in test
samples. samples.
The endogenous control is a target that is expressed equally in all samples;
examples of endogenous controls are β‐actin, GAPDH, and 18S ribosomal RNA. The
software can algorithmically incorporate multiple endogenous control targets in relative
quantification calculations.
The reference sample is used as the basis for relative quantification results (or
1× sample). For example, in a study of drug effects on gene expression, an untreated
control is an appropriate reference sample.
2. The software generates standard curves for the target of interest and the endogenous
control using data from the corresponding standard dilution series.
3. The software uses the standard curves to interpolate the quantities of the target of
interest and the endogenous control in each sample. The target quantity in each
sample is then normalized to the sample's endogenous control quantity.
4. To determine the relative quantity of the target in test samples, the software divides
the normalized target quantity in the sample by the normalized target quantity in the
reference sample.
Determines relative 1. The software measures amplification of the target of interest and of an endogenous
target quantity in control target in a reference (calibrator) sample and in test samples.
samples. The endogenous control is a target that is expressed equally in all samples;
examples of endogenous controls are β‐actin, GAPDH, and 18S ribosomal RNA. The
software can algorithmically incorporate multiple endogenous control targets in relative
quantification calculations.
The reference sample is used as the basis for relative quantification results (or
1× sample). For example, in a study of drug effects on gene expression, an untreated
control is an appropriate reference sample.
2. The measurements for the target of interest are normalized to the endogenous control.
3. To determine the relative quantity of the target in test samples, the software compares
the normalized ΔCq (ΔCt or ΔCrt) for the sample to the normalized ΔCq (ΔCt or ΔCrt) for
the reference sample.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 19
Chapter 1 Product information
1 Types of runs
Purpose Description
Genotyping run
Detects single Genotyping runs use preformulated TaqMan™ SNP Genotyping Assays that include the
nucleotide following components:
polymorphism (SNP)
variants of a • Two sequence-specific primers for amplification of sequences containing the SNP of
target nucleic acid interest
sequence. • Two allele-specific TaqMan™ probes for Allele 1 and Allele 2
1. The software normalizes the fluorescence of the reporter dyes to the fluorescence of
the passive reference dye in each well.
2. The software plots the normalized reporter dye signal of each sample well on an Allelic
Discrimination Plot, which contrasts the reporter dye intensities of the allele-specific
probes.
3. The software algorithmically clusters the sample data, and assigns a genotype call to
the samples of each cluster according to its position on the plot.
Presence/absence run
Determines the The software calls the target present or absent based on an algorithmically determined call
presence or absence threshold. (The call threshold is different from the Ct threshold; the Ct threshold is not used
of a target nucleic
to make calls.)
acid sequence in a
sample.
20 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
2 Start, sign on, and configure the
instrument
Note:
· Instruments are factory calibrated, so no calibration is necessary at installation. However, we
recommend that you verify instrument performance before using the instrument.
· Regular calibration and verification should be performed according to the calibration and
verification schedule (see “Calibration and verification schedule” on page 43).
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 21
Chapter 2 Start, sign on, and configure the instrument
2 Precautions for use
CAUTION! FIRE HAZARD. For continued protection against the risk of fire, replace fuses only with
listed and certified fuses of the same type and rating as those currently in the instrument.
CAUTION! PHYSICAL INJURY HAZARD. During instrument operation, the sample block
temperature can reach 100°C. Allow it to cool to room temperature before handling.
CAUTION! Before using a cleaning or decontamination method other than those recommended by
Thermo Fisher Scientific, confirm with Thermo Fisher Scientific that the proposed method will not
damage the instrument.
CAUTION! Use flat caps for tubes. Rounded caps can damage the heated cover.
22 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 2 Start, sign on, and configure the instrument
Parts of the home screen 2
1 2
9 6
8 7
Element of the
Function For more information, see...
home screen
Avatar and Set by the administrator to uniquely identify instrument. “Manage the instrument name
Instrument name (Administrator only)” on page 88.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 23
Chapter 2 Start, sign on, and configure the instrument
2 Parts of the home screen
Element of the
Function For more information, see...
home screen
Current user Displays the user name of the current signed-in user and —
name and block the instrument block type.
type
Note: If no user is signed-in, the instrument defaults to
the Guest profile.
Settings button Touch Settings to configure, calibrate, or learn about “Configure instrument settings” on
the instrument. page 26.
Buttons for Load Experiment – Touch to open a user-created .edt “Run an experiment from a saved
accessing file from a Thermo Fisher™ Connect account, USB, file” on page 30.
experiment and instrument folder, or network drive.
template files
Open Template – Touch to open a system .edt template “Create and run an experiment from
file. a template” on page 30.
Run Last – Touch to open the last .edt file run on the “Repeat the last instrument run” on
instrument. page 31.
• If you are signed-in, the file opened will be the last
file you ran when signed-in.
• If you are not signed-in, the file opened will be the
last file the guest profile ran.
Sign In button • Touch Sign In to sign into an instrument profile “Sign in” on page 25.
and link to a Thermo Fisher™ Connect account.
(My Profile button
when a user is • Touch My Profile to change instrument profile
signed in) settings, link to a Thermo Fisher™ Connect account,
or lock the instrument during a run.
24 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 2 Start, sign on, and configure the instrument
Use the instrument without signing in 2
3. Touch PIN Code, enter a four-digit numerical password, then touch Enter.
Note: Touch the Show PIN checkbox to switch PIN display on or off.
Note: To enable access to Thermo Fisher™ Connect, see “Link an instrument profile to Thermo Fisher™
Connect” on page 77.
Sign in
Create an instrument profile before signing into the instrument. See “Create a new instrument profile”
on page 25.
Note: An instrument profile is a user account specifically for the instrument. It is not related to any
other user account for the system or software.
Note: To enable access to Thermo Fisher™ Connect, see “Link an instrument profile to Thermo Fisher™
Connect” on page 77.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 25
Chapter 2 Start, sign on, and configure the instrument
2 Sign out
Sign out
1. In the home screen, touch My Profile.
Options Description
Instrument Settings
Instrument Name Instrument name. If connected to a network, the instrument name must be unique.
(Administrator only) (Optional) Add an avatar image for the instrument (.jpg, .png, or .gif) from a USB.
Sleep Mode Enable the instrument to enter a standby mode after a set length of inactivity.
Heated Cover Temperature Set the temperature of the heated cover for instrument operation and standby mode.
Enable heated cover to automatically turn off during standby mode.
Network Drive Specify a default network location for the signed-in user.
Insert Plate Reminder Enable a reminder to insert a plate before starting a run from the instrument.
OEM Connection Only Required for API access to the instrument. When enabled, the desktop cannot
connect to the instrument.
(Administrator only)
API access to the instrument is exclusive to authorized OEM partners.
Cloud Region Specify the regional server location to access Thermo Fisher™ Connect.
(Administrator only) Note: Once you set a region for the cloud platform, restore factory defaults to change
the region.
26 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 2 Start, sign on, and configure the instrument
Configure instrument settings 2
Options Description
Network Connection Set wired or wireless network connection for the instrument.
IMPORTANT! Back up the instrument before restoring factory defaults (see “Backup
or restore the instrument” on page 67).
(QuantStudio™ 5 Instrument • Runs must be started from the desktop software. The instrument locks out
only) access to load experiments, open templates, repeat the last run, start a run,
change the instrument name, and change the date/time.
• Only an administrator can perform calibrations.
• A user can monitor a run and view the method from the instrument touchscreen.
About Instrument
About Instrument Displays the instrument Model Name, Block Serial Number, and Firmware Version.
License Agreement Displays the End User Software License Agreement and the Limited Product Warranty.
You can export the License Agreement to a USB drive.
Notifications
— Enable home screen notifications of instrument errors and software updates. This
function is not related to the Notifications function in the cloud software.
The number of new, unviewed notifications displays over Settings in the home screen.
Monitoring Enable:
(Administrator only) • Remote Monitoring Service to automatically send critical device statistics to
Thermo Fisher Scientific (does not monitor or send data).
• Thermo Fisher Scientific Cloud Monitor to allow real-time monitoring of
instrument runs from a Cloud account.
Instrument Statistics Displays instrument usage information: Block Cycle Count and LED Life.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 27
Chapter 2 Start, sign on, and configure the instrument
2 Configure instrument settings
Options Description
Ship Prep Mode Place the instrument in a safe state for shipping, moving, or long-term storage.
Run History
— Displays the experiments run on the instrument and whether the data was transferred.
Note: Experiments run while a user is signed-in can only be viewed or transferred by
that user or an administrator.
Touch an experiment to view experiment run details and to transfer or delete .eds file.
Manage Users
Sign In Required Enable the restriction of instrument use to only signed-in users (disables Guest profile
use).
(Administrator only)
Sign Out Timer Set the duration of inactivity before a user is automatically signed out.
(Administrator only)
Displays the profile information for the instrument and the associated Thermo Fisher™
Connect cloud-based account.
28 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
3 Create and run experiments on the
instrument
■ Workflow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
■ Options for running an experiment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
■ Edit an experiment before starting a run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
■ Load and unload the plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
■ View, pause, or stop a run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
■ Transfer, view, or manage files and results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Workflow
Start the instrument
Transfer results (EDS file) to the desktop software or Thermo Fisher™ Connect, then analyze
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 29
Chapter 3 Create and run experiments on the instrument
3 Options for running an experiment
4. (Optional) Enter or edit template properties, including Run File Name (EDS file name), Plate
Barcode, Reagent Information, and Data Destination (see “Enter or edit template properties”).
5. (Optional) Edit the run method (see “Edit the run method” on page 33).
• Add, remove, or edit a step, stage, melt curve, or data collection point.
• Adjust the heated cover temperature, sample volume, or number of cycles.
• Configure VeriFlex™ Zones, ramp rate, and pause settings.
6. (Optional) Define plate wells with sample names and view Well ID, Targets, or Dyes (see “Define,
assign, and view well details”).
7. Load a plate into the instrument (see “Load and unload the plate” on page 37).
Note: To disable this reminder, select Do not show again or select Settings4Insert Plate
Reminder in the home screen.
5. (Optional) Enter or edit template properties, including Run File Name (EDS file name), Plate
Barcode, Reagent Information, and Data Destination (see “Enter or edit template properties”).
30 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 3 Create and run experiments on the instrument
Options for running an experiment 3
6. (Optional) Edit the run method (see “Edit the run method” on page 33).
• Add, remove, or edit a step, stage, melt curve, or data collection point.
• Adjust the heated cover temperature, sample volume, or number of cycles.
• Configure VeriFlex™ Zones, ramp rate, and pause settings.
7. (Optional) Define plate wells with sample names and view Well ID, Targets, or Dyes (see “Define,
assign, and view well details”).
8. Load a plate into the instrument (see “Load and unload the plate” on page 37).
Note: To disable this reminder, select Do not show again or select Settings4Insert Plate
Reminder in the home screen.
2. (Optional) Enter or edit template properties, including Run File Name (EDS file name), Plate
Barcode, Reagent Information, and Data Destination (see “Enter or edit template properties”).
3. (Optional) Edit the run method (see “Edit the run method” on page 33).
• Add, remove, or edit a step, stage, melt curve, or data collection point.
• Adjust the heated cover temperature, sample volume, or number of cycles.
• Configure VeriFlex™ Zones, ramp rate, and pause settings.
4. (Optional) Define plate wells with sample names and view Well ID, Targets, or Dyes (see “Define,
assign, and view well details”).
5. Load a plate into the instrument (see “Load and unload the plate” on page 37).
Note: To disable this reminder, select Do not show again or select Settings4Insert Plate
Reminder in the home screen.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 31
Chapter 3 Create and run experiments on the instrument
3 Edit an experiment before starting a run
• Edit the file name for the EDS file for the run.
a. Touch the Run File Name field.
b. Enter the name for the EDS file, then touch Done.
c. Touch the Name, Type, Lot #, Reagent Barcode, Part #, or Expiration Date field to enter
individual reagent information.
d. Touch Done.
• Automatically transfer run data (EDS file) when an instrument run ends.
a. Touch Data Destination.
d. Touch Done.
• Enter a comment.
a. Touch Comments.
32 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 3 Create and run experiments on the instrument
Edit an experiment before starting a run 3
2. Hold the scanner 20–30 cm away from a plate or container label and aim at the center of the
barcode, then press the trigger.
3. Slowly move the scanning beam across the barcode until the scanner emits a high-pitched tone.
When the scanner scans a barcode, it automatically transmits the following information:
• Transmits the alphanumeric equivalent of the barcode to the barcode field.
• Transmits other reagent information (Lot #, Part #, Expiration Date, etc.)
For more information about the hand-held barcode scanner, see the user documentation provided with
the barcode scanner.
For an overview of the method as it is graphically represented on the touchscreen, see “Method
elements” on page 34.
In the Method screen:
1. Touch Edit.
3. Touch Manage Steps to add or remove a step, stage, melt curve, or data collection point.
Note: The Manage Steps screen also provides access to configuring VeriFlex™ Zones, Ramp
Rates, and Add Pause settings (see “Configure VeriFlex™ Zones, ramp rates, and pause settings”
on page 35).
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 33
Chapter 3 Create and run experiments on the instrument
3 Edit an experiment before starting a run
Method elements
2 3
Option Procedure
1. Touch Add/Remove steps4Add steps.
2. Touch on the left or right border of a step to add a step before or
Add a step after the step, respectively.
3. Enter parameters for the new step, then touch Enter.
4. Touch Done.
34 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 3 Create and run experiments on the instrument
Edit an experiment before starting a run 3
(continued)
Option Procedure
1. Touch Add/Remove stages4Add stages.
2. Touch on the left or right border of a stage to add a stage before or
after the stage, respectively.
Add a stage
3. Edit parameters of the new stage in the Method screen. See “Edit the
run method” on page 33.
4. Touch Done.
e. Touch Done.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 35
Chapter 3 Create and run experiments on the instrument
3 Edit an experiment before starting a run
c. Touch Enter.
b. Touch in a stage.
c. Enter the pause temperature, and the cycle after which the pause will occur.
d. Touch Enter.
The pause is represented by a P in the corner of the stage.
e. Touch Done.
Option Procedure
Define and assign samples in the 1. Touch .
plate layout view
2. Touch Manage or touch an individual well.
3. Touch the Samples subtab.
4. Select one or more wells, then touch Edit.
5. Enter sample names for the selected wells, then touch Done.
36 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 3 Create and run experiments on the instrument
Load and unload the plate 3
CAUTION! Use flat caps for 0.2-mL tubes and 0.1-mL tubes. Rounded caps can damage the heated
cover.
b. Load the plate onto the plate adaptor so that the following criteria are met.
• Position well A1 of the plate in the top-left corner of the plate adapter.
• Ensure that the barcode faces the front of the instrument.
IMPORTANT! The instrument should be used by trained operators who have been warned of
the moving parts hazard.
Note: (For 96-well 0.2-mL blocks only) Do not remove the black plate adapter before loading
a plate or tube strips. If used, tube strips can fit loosely in the adapter, but the heated cover
will apply the appropriate pressure to seat the tube strips securely in the adapter.
Note: The 384-well and 96-well Fast (0.1-mL) block configurations do not require a plate
adapter.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 37
Chapter 3 Create and run experiments on the instrument
3 Load and unload the plate
CAUTION! PHYSICAL INJURY HAZARD. During instrument operation, the plate temperature
can reach 100°C. Allow it to cool to room temperature before handling.
Note: If the instrument does not eject the plate, contact Support.
b. Load the plate onto the plate adaptor so that the following criteria are met.
• Position well A1 of the plate in the top-left corner of the plate adapter.
• Ensure that the barcode faces the front of the instrument.
IMPORTANT! The instrument should be used by trained operators who have been warned of
the moving parts hazard.
Note: (For 96-well 0.2-mL blocks only) Do not remove the black plate adapter before loading
a plate or tube strips. If used, tube strips can fit loosely in the adapter, but the heated cover
will apply the appropriate pressure to seat the tube strips securely in the adapter.
38 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 3 Create and run experiments on the instrument
Load and unload the plate 3
c. Before loading the plate in the drawer, place the MicroAmp™ Optical Film Compression Pad
with the gray side down on the surface of the plate to ensure a proper seal between the
instrument and the adhesive film.
IMPORTANT! Be careful to place the compression pad with the brown side up and the
gray side down, centered on top of the plate. Ensure that the compression pad is free from
wrinkles and signs of deterioration prior to use.
d. Remove the MicroAmp™ Optical Film Compression Pad from the plate and store it inside the
pack.
IMPORTANT!
· If the compression pad becomes stuck inside the instrument, call service to clean the
heated cover.
· Between each use, place the pad back in the pouch so that it does not dry out.
· Each compression pad may be used up to 20 times before discarding. Do not use more
than 20 times.
CAUTION! PHYSICAL INJURY HAZARD. During instrument operation, the plate temperature
can reach 100°C. Allow it to cool to room temperature before handling.
Note: If the instrument does not eject the plate, contact Support.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 39
Chapter 3 Create and run experiments on the instrument
3 View, pause, or stop a run
CAUTION! PHYSICAL INJURY HAZARD. During instrument operation, the plate temperature
can reach 100°C. If you want to access the plate during a run pause, enter room temperature as
the pause temperature and allow the plate to cool to room temperature before handling.
3. (Optional) After pausing a run, touch Edit to change the number of cycles.
40 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 3 Create and run experiments on the instrument
Transfer, view, or manage files and results 3
2. Touch Zoom.
4. Touch the arrows to pan left, right, up, or down on the graph.
4. Touch OK.
5. Touch Transfer.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 41
Chapter 3 Create and run experiments on the instrument
3 Transfer, view, or manage files and results
View run history and delete or transfer files from the instrument
In the home screen, touch Settings4Run History.
• Touch an individual run record to view its details, then complete one of the following actions:
– Touch Delete to delete the run record.
– Touch Transfer to export the run data.
• Touch Manage to select multiple run records for simultaneous viewing, deletion, or transfer.
Note:
· Guests (users not signed-in) can only view guest run records.
· Users signed into their instrument profiles can also view their own run records.
· Administrators can view all run records.
Note: If the connection between the instrument and the desktop software is interrupted during the run,
the instrument still completes the run. However, the run data (EDS file) must be transferred from the
instrument to the desktop software using a USB drive or a network drive.
4. (Optional) Navigate through the My Instrument, Public, USB, and Post Read folders to access an
EDT file. Folder availability depends on your sign-in status.
7. Touch Done.
42 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
4 Calibrate and verify instrument
performance
Note: After instrument installation, we recommend performing instrument verification using the
provided RNase P plate.
IMPORTANT! Perform calibrations and instrument runs under the environmental conditions that
are specified in “Environmental requirements” on page 108. Exposure to extreme temperatures can
adversely affect the instrument performance and shorten the life span of the instrument components.
To set the calibration frequency for the instrument, touch Settings4Maintenance and
Service4Calibrations4History and Reminders4Edit4Exp interval field.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 43
Chapter 4 Calibrate and verify instrument performance
4 Calibration descriptions
(continued)
Note: Performing a Dye calibration for a given dye plate does not invalidate
any other calibration.
Note: To prepare custom dye plates and to perform custom calibrations, see “Calibrate custom dyes”
on page 59.
Calibration descriptions
Calibration description and purpose Pass Criteria
ROI/Uniformity
• The software captures images for each optical filter. The image for each filter
• The software uses calibration data to map the increase in fluorescence distinguishes all wells of the plate.
to the plate wells during subsequent runs and to evaluate well-to-well Each well in the image is distinct.
consistency of the signals.
Background
• The software captures background images for each optical filter in the The plate images for all filters are free
absence of sample and reagent, and it checks that the fluorescence of abnormal fluorescence.
from each well is below a fluorescence threshold.
• The software uses calibration data to remove background fluorescence
during a run.
Note: You can also run this calibration to determine if contamination is
related to the sample block or the plate.
44 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 4 Calibrate and verify instrument performance
View the calibration status and set reminders 4
(continued)
Calibration description and purpose Pass Criteria
Dye
• The software extracts a spectral profile for each dye standard, then Dye spectra peak within the same
produces a set of spectral profiles plotted as fluorescence vs filter. filter as their group.
• The software uses calibration data to characterize and distinguish the
individual contribution of each dye in the total fluorescence signals
collected by the instrument.
2. In the Calibration Reminders screen, view the status of each calibration type.
3. (Optional) Touch a calibration row to view the history of that specific calibration type, then touch
Done.
4. Touch Edit to set the calibration reminder settings. For each calibration type:
a. Slide the control On to enable the calibration reminder.
c. Touch Save.
5. (Optional) To transfer the calibration report, touch Export then follow the directions on the screen.
6. Touch Done.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 45
Chapter 4 Calibrate and verify instrument performance
4 Perform ROI/uniformity, background, and dye calibrations
Note: More than one email address can receive the calibration reminders.
(Optional ) Perform custom dye calibrations (“Calibrate custom dyes” on page 59)
Perform any time that ROI and background calibrations are current.
46 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 4 Calibrate and verify instrument performance
Perform ROI/uniformity, background, and dye calibrations 4
Note: We recommend calibrating with all Spectral Dye Calibrations Plates available for your
block configuration even if you are not using all the dyes in the plates.
Note: Do not discard the packaging for the calibration plates. Each calibration plate can be used
up to 3 times if the following conditions are met:
· The plate is stored in its packing sleeve at –15 to –25°C.
· The plate is used within 6 months after opening.
· The plate is used before the plate expiry date
IMPORTANT! Do not remove the plate from its packaging until you are ready to use it. The
fluorescent dyes in the wells of calibration plates are photosensitive. Prolonged exposure to light
can diminish the fluorescence of the dyes.
2. While wearing powder‐free gloves, remove the calibration plate from its packaging and retain the
packaging. Do not remove the optical film.
3. Vortex the plate for 5 seconds, then centrifuge at 750–1,000 × g for 2 minutes.
4. Confirm that the liquid in each well is at the bottom of the well and free of bubbles. If it is not,
centrifuge the plate again.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 47
Chapter 4 Calibrate and verify instrument performance
4 Perform ROI/uniformity, background, and dye calibrations
IMPORTANT! Keep the bottom of the plate clean. Fluids and other contaminants on the bottom
of the plate can contaminate the sample block and cause an abnormally high background signal.
Perform calibrations
1. In the instrument home screen:
Calibration Touch
ROI/Uniformity[1] Settings4Maintenance and Service4Calibrations4ROI and Uniformity
Background[2] Settings4Maintenance and
Service4Calibrations4Custom4Background
Dye Settings4Maintenance and Service4Calibrations4Dye
[1] Automatically followed by Background calibration.
[2] Initiate via this route if performing Background calibration only.
Note: Dye calibration only: Select the Dye Plate to run, then touch Next.
48 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 4 Calibrate and verify instrument performance
Perform ROI/uniformity, background, and dye calibrations 4
b. Load the plate onto the plate adaptor so that the following criteria are met.
• Position well A1 of the plate in the top-left corner of the plate adapter.
• Ensure that the barcode faces the front of the instrument.
IMPORTANT! The instrument should be used by trained operators who have been warned of
the moving parts hazard.
Note: (For 96-well 0.2-mL blocks only) Do not remove the black plate adapter before loading
a plate or tube strips. If used, tube strips can fit loosely in the adapter, but the heated cover
will apply the appropriate pressure to seat the tube strips securely in the adapter.
Note: The 384-well and 96-well Fast (0.1-mL) block configurations do not require a plate
adapter.
4. Touch Start.
5. When the run is complete and the screen displays Calibration Complete, touch View Results to
check the calibration status.
Note: You can view the calibration images only after the ROI/Uniformity and Background
calibrations pass.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 49
Chapter 4 Calibrate and verify instrument performance
4 Perform ROI/uniformity, background, and dye calibrations
CAUTION! PHYSICAL INJURY HAZARD. During instrument operation, the plate temperature
can reach 100°C. Allow it to cool to room temperature before handling.
Note: If the instrument does not eject the plate, contact Support.
Note: Each calibration plate can be used up to 3 times if the following conditions are met:
· The plate is stored in its packing sleeve at –15 to –25°C.
· The plate is used within 6 months after opening.
· The plate is used before the plate expiry date
2. In the Details screen, touch a calibration type to view its images and plots.
50 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 4 Calibrate and verify instrument performance
Perform ROI/uniformity, background, and dye calibrations 4
(continued)
Calibration Example results indicating successful calibration
Background Few, if any, signals with abnormally high fluorescence.
Dye Signals from each well following a uniform trend, and each dye
peaks at the correct filter.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 51
Chapter 4 Calibrate and verify instrument performance
4 Perform ROI/uniformity, background, and dye calibrations
Identify contamination
Signals that exceed the limit of normal fluorescence may indicate fluorescent contaminants on the
calibration plate or the sample block. Common contaminants include ink residue from permanent pens,
powder from disposable gloves, and dust.
1. View the calibration data and note the wells that failed the calibration.
2. Remove the plate from the instrument, rotate the plate 180°, then perform the calibration again.
4. If the calibration fails after you decontaminate the sample block and replace the plate, contact
Support.
52 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 4 Calibrate and verify instrument performance
Perform instrument verification using RNase P plates 4
1. Remove a reaction plate from its box and place it on a clean, dry surface.
3. Seal the plate using an optical adhesive cover or optical flat caps.
The instrument requires valid ROI/uniformity, background, and dye calibrations to perform instrument
verification.
Confirms the Quantifies the number of copies of The instrument passes performance specifications
performance of the the human RNase P gene in samples if the following inequality is true and the
instrument. with known concentrations of the instrument successfully distinguishes between
corresponding DNA template. unknown populations A and B.
[(CtA) – 3(σCtA)] > [(CtB) + 3(σCtB)]
where:
• CtA = Average Ct of unknown population A
• σCtA = Standard deviation of unknown
population A
• CtB = Average Ct of unknown population B
• σCtB = Standard deviation of unknown
population B
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 53
Chapter 4 Calibrate and verify instrument performance
4 Perform instrument verification using RNase P plates
2 3 4
5 6 7
1 3 5
7 8
2 4 6
54 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 4 Calibrate and verify instrument performance
Perform instrument verification using RNase P plates 4
The software automatically adjusts the threshold and omits a defined number of wells from the
unknown populations to meet the performance specifications. To view any omitted wells, open the
EDS file for the verification in the desktop software.
1. Remove the RNase P plate from the freezer, then thaw the plate in its packaging.
• Thaw the plate for approximately 5 minutes.
• Use the plate within 30 minutes of thawing.
2. Confirm that the bench, vortex, and centrifuge are clean. Before use, wipe the vortex and
centrifuge using a lint‑free tissue.
3. While wearing powder‐free gloves, remove the plate from its packaging.
4. Vortex the plate for 5 seconds, then centrifuge at 750–1,000 × g for 2 minutes.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 55
Chapter 4 Calibrate and verify instrument performance
4 Perform instrument verification using RNase P plates
5. Confirm that the liquid in each well is at the bottom of the well and free of bubbles. If it is not,
centrifuge the plate again.
IMPORTANT! Keep the bottom of the plate clean. Fluids and other contaminants on the bottom
of the plate can contaminate the sample block and cause an abnormally high background signal.
b. Load the plate onto the plate adaptor so that the following criteria are met.
• Position well A1 of the plate in the top-left corner of the plate adapter.
• Ensure that the barcode faces the front of the instrument.
IMPORTANT! The instrument should be used by trained operators who have been warned of
the moving parts hazard.
Note: (For 96-well 0.2-mL blocks only) Do not remove the black plate adapter before loading
a plate or tube strips. If used, tube strips can fit loosely in the adapter, but the heated cover
will apply the appropriate pressure to seat the tube strips securely in the adapter.
Note: The 384-well and 96-well Fast (0.1-mL) block configurations do not require a plate
adapter.
56 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 4 Calibrate and verify instrument performance
Perform instrument verification using RNase P plates 4
3. Touch Start.
4. When the run is complete and the screen displays Verification Complete, touch View Results to
confirm the status of the run.
CAUTION! PHYSICAL INJURY HAZARD. During instrument operation, the plate temperature
can reach 100°C. Allow it to cool to room temperature before handling.
Note: If the instrument does not eject the plate, contact Support.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 57
Chapter 4 Calibrate and verify instrument performance
4 Perform instrument verification using RNase P plates
58 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 4 Calibrate and verify instrument performance
Calibrate custom dyes 4
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 59
Chapter 4 Calibrate and verify instrument performance
4 Calibrate custom dyes
2. Dispense aliquots of each dilution into the center of a reaction plate, then seal the plate.
A full plate is not needed. See the following figures for suggested replicates.
100 nM
200 nM
400 nM
800 nM
1600 nM
100 nM
200 nM
400 nM
800 nM
1600 nM
100 nM
200 nM
400 nM
800 nM
1600 nM
3. Vortex the plate for 5 seconds, then centrifuge at 750–1,000 × g for 2 minutes.
4. Confirm that the liquid in each well is at the bottom of the well and free of bubbles. If it is not,
centrifuge the plate again.
IMPORTANT! Keep the bottom of the plate clean. Fluids and other contaminants on the bottom
of the plate can contaminate the sample block and cause an abnormally high background signal.
60 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 4 Calibrate and verify instrument performance
Calibrate custom dyes 4
CAUTION! The instrument should be used by trained operators who have been warned of the
moving parts hazard.
CAUTION! PHYSICAL INJURY HAZARD. During instrument operation, the plate temperature
can reach 100°C. Allow it to cool to room temperature before handling.
Note: If the instrument does not eject the plate, contact Support.
1. In the Results tab of the desktop or cloud software, select Raw Data Plot.
This plot displays the raw fluorescence signal of each optical filter, for individual wells.
Note: The Raw Data Plot cannot be viewed on the instrument touchscreen.
2. For each replicate population of dilutions, select the wells in the Plate Layout to view in the plot.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 61
Chapter 4 Calibrate and verify instrument performance
4 Calibrate custom dyes
3. Examine the raw data to identify the wells yielding signals according to the ranges shown in the
following table.
4. (Optional) Export the raw data, then calculate the average fluorescence value for each
concentration.
5. Select the lowest (optimal) dye concentration that falls within the acceptable signal range.
Create a full plate of the custom dye diluted to the optimal concentration:
2. Pipet the appropriate volume of the diluted custom dye to the plate wells.
4. Vortex the plate for 5 seconds, then centrifuge at 750–1,000 × g for 2 minutes.
5. Confirm that the liquid in each well is at the bottom of the well and free of bubbles. If it is not,
centrifuge the plate again.
IMPORTANT! Keep the bottom of the plate clean. Fluids and other contaminants on the bottom
of the plate can contaminate the sample block and cause an abnormally high background signal.
62 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 4 Calibrate and verify instrument performance
Calibrate custom dyes 4
Field/option Action
Custom Dye Name Enter a name for the custom dye.
IMPORTANT!
· Do not use a system dye name for a custom dye name.
· Dye names are spacing sensitive and cannot contain special characters.
Type Select:
• Reporter—The dye works in conjunction with a quencher dye to report an
increase of PCR product.
• Quencher—The dye suppresses the fluorescence of a reporter dye until
amplification of PCR product.
• Both—The dye can be used as a reporter or quencher dye.
4. Touch Save.
IMPORTANT! You must also add the custom dye to the desktop software dye library before
creating, running, or analyzing experiments that use the custom dye. For detailed information, see
the QuantStudio™ Design and Analysis Desktop Software User Guide.
CAUTION! The instrument should be used by trained operators who have been warned of the
moving parts hazard.
4. Review the custom dye information, (optional) make changes, then touch Update.
7. Touch Start.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 63
Chapter 4 Calibrate and verify instrument performance
4 Calibrate custom dyes
8. When the run is complete and the screen displays Calibration Complete, touch View
Results4Details.
9. Review the plot. Passing calibration results show uniform signals with peaks that are aligned with
the dye wavelength.
10. Select an action depending on whether the custom dye calibration passed or failed.
Failed • Perform the calibration again using a new custom dye plate.
• See “Troubleshoot calibration failure” on page 51.
CAUTION! PHYSICAL INJURY HAZARD. During instrument operation, the plate temperature
can reach 100°C. Allow it to cool to room temperature before handling.
Note: If the instrument does not eject the plate, contact Support.
64 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 4 Calibrate and verify instrument performance
Calibrate for a custom melt curve experiment 4
2. Touch PCR + Melt or Melt only as appropriate for the kit you are using.
4. Select or add a dye, then select a filter set appropriate for your dye's wavelength (see filter-
wavelength table below).
Note: Refer to your reagent kit documentation for dye name and wavelength information.
IMPORTANT! If the selected filter set does not match your reagent kit documentation, then the
incorrect wavelength may be collected during a run.
6. Touch Start.
7. When the run is complete and the screen displays Calibration Complete, touch View
Results4Details.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 65
Chapter 4 Calibrate and verify instrument performance
4 Calibrate for a custom melt curve experiment
8. Review the plot. Passing calibration results show uniform signals with peaks that are aligned with
the dye wavelength.
9. Select an action depending on whether the custom dye calibration passed or failed.
Failed • Perform the calibration again using a new custom dye plate.
• See “Troubleshoot calibration failure” on page 51.
Note: You must also add the custom dye to the desktop software dye library before creating, running,
or analyzing experiments that use the custom dye. For detailed information, see the QuantStudio™
Design and Analysis Desktop Software User Guide.
Note: To perform a custom melt experiment, you can either create a Standard Curve or a Custom
experiment with melt, then specify the data points per degree in the method.
66 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
5 Maintain the instrument
IMPORTANT! This chapter contains user maintenance procedures for the instrument. Procedures
other than those described in this document must be performed by a qualified Thermo Fisher Scientific
service representative.
To Action
Backup 1. Touch Backup Instrument.
2. Select a location to store the backup settings for your instrument.
3. If backing up to a USB drive, insert a USB drive into the front-panel USB
port.
4. Enter a backup file name, then touch Done.
5. Select the elements to backup, or leave them all selected.
6. Touch Backup.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 67
Chapter 5 Maintain the instrument
5 Decontaminate the sample block
CAUTION! PHYSICAL INJURY HAZARD. Do not remove the instrument cover. There are no
components inside the instrument that you can safely service yourself. If you suspect a problem,
contact Support.
CAUTION! PHYSICAL INJURY HAZARD. During instrument operation, the sample block
temperature can reach 100°C. Allow it to cool to room temperature before handling.
CAUTION! Before using a cleaning or decontamination method other than those recommended by
Thermo Fisher Scientific, confirm with Thermo Fisher Scientific that the proposed method will not
damage the equipment.
Materials required
• Safety glasses
• Powder-free gloves
• Tissue, lint-free
• Cotton or nylon swabs and lint-free cloths
• Pipette (100-µL) with pipette tips
• Deionized water
• Ethanol, 95% solution
• Na-hypochlorite, (0.1% v/v) solution
CAUTION! PHYSICAL INJURY HAZARD. During instrument operation, the sample block
temperature can reach 100°C. Allow it to cool to room temperature before handling.
IMPORTANT! Always use deionized water to rinse wells after cleaning with bleach or ethanol solution.
1. Identify the contaminated wells of the sample block (see “Identify contamination” on page 52).
68 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 5 Maintain the instrument
Decontaminate the sample block 5
3. Rinse the contaminated wells with deionized water (see “Detailed procedures for cleaning the
sample block” on page 70).
4. Close the drawer and test the sample block for contamination:
a. Push the instrument drawer back in to the instrument.
c. Perform a background calibration to confirm that you have eliminated the contamination.
5. If the contamination remains, clean the contaminated wells using a 95% ethanol solution:
a. Repeat step 2 – step 3.
b. Clean the contaminated wells using a 95% ethanol solution (see “Detailed procedures for
cleaning the sample block” on page 70).
c. Repeat step 3 – step 4 to rinse the sample block with deionized water and to confirm that you
have eliminated the contamination.
IMPORTANT! Always use deionized water to rinse wells after cleaning with bleach or ethanol
solution.
6. If the contamination still remains, clean the contaminated wells using a Na-hypochlorite (0.1% v/v)
solution:
a. Repeat step 2 – step 3.
b. Clean the contaminated wells using a Na-hypochlorite (0.1% v/v) solution (see “Detailed
procedures for cleaning the sample block” on page 70).
c. Repeat step 3 – step 4 to rinse the sample block with deionized water and to confirm that you
have eliminated the contamination.
IMPORTANT! Always use deionized water to rinse wells after cleaning with Na-hypochlorite
or ethanol solution.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 69
Chapter 5 Maintain the instrument
5 Decontaminate the sample block
b. In each well, pipet the water up and down several times to rinse the well.
b. In each well, pipet the solution up and down several times to rinse the well.
IMPORTANT! Always use deionized water to rinse wells after cleaning with bleach or ethanol
solution.
b. In each well, pipet the solution up and down several times to rinse the well.
IMPORTANT! Always use deionized water to rinse wells after cleaning with bleach or ethanol
solution.
70 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 5 Maintain the instrument
Replace the instrument fuses 5
Materials required
• Fuses (2) – 10A, Time-Lag T, 250VAC, 5 × 20mm
• Safety glasses
• Powder-free gloves
• Screwdriver, flathead
3. Remove each fuse from its fuse holder and inspect it for damage. Carbon typically coats the inside
of failed fuses.
Good Failed
Note: The voltage and amperage ratings are on the fuse holder.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 71
Chapter 5 Maintain the instrument
5 Power on or off, store, and move
Note: Fuse failure can result from fluctuations in the supplied power to the system. To prevent further
failures, consider installing an electrical protective device, such as a UPS or a surge protector. If issues
with the fuse persist, contact Support.
3. Touch Edit Time, then enter the time length of inactivity before the instrument enters sleep mode.
Note: To power off the instrument for >1 week, see “Prepare the instrument to store, move, or ship” on
page 73.
1. Power off the instrument using the power switch on the back of the instrument.
72 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 5 Maintain the instrument
Power on or off, store, and move 5
3. Load the packing plate or an empty plate, then touch to close the drawer.
5. Power off the instrument using the power switch on the back of the instrument.
CAUTION! PHYSICAL INJURY HAZARD. Do not attempt to lift the instrument or any other heavy
objects unless you have received related training. Incorrect lifting can cause painful and sometimes
permanent back injury. Use proper lifting techniques when lifting or moving the instrument. At least
two people are required to lift it.
IMPORTANT! Moving your instrument can create subtle changes in the alignment of the instrument
optics. Recalibrate the instrument if necessary.
• Ensure that the surface on which you place the instrument can support at least 35 kg (77 lbs).
• Ensure that the path to transport the instrument is clear of obstructions.
• At least two people are needed to lift and carry the instrument.
• Keep your spine in a good neutral position.
• Bend at the knees and lift with your legs.
• Do not lift an object and twist your torso at the same time.
• Coordinate your intentions with your assistant before lifting and carrying.
IMPORTANT! After moving the instrument, perform an RNase P instrument verification run. If the run
fails, perform ROI/ uniformity, background, and dye calibrations.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 73
Chapter 5 Maintain the instrument
5 Power on or off, store, and move
1. Contact your local customer care center or technical support group to obtain a copy of the
Certificate of Instrument Decontamination, a service notification, a service call number, and
packaging materials (if required).
5. Pack the instrument in the provided packaging and follow the instructions in the table below.
6. Attach the postage provided with the Certificate of Instrument Decontamination to the box, then
ship the instrument to the designated facility.
74 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
6 Configure the instrument and
manage instrument profiles
■ Initial start‑up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
■ Manage profiles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
■ Link an instrument profile to Thermo Fisher™ Connect . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
■ Enable SAE mode (Administrator only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
■ Require instrument profile sign-in (Administrator only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
■ Enable remote instrument monitoring (Administrator only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
■ Update instrument software (Administrator only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
■ Restore factory defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
■ Configure the network . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
■ Select a region for Thermo Fisher™ Connect (Administrator only) . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
■ Manage the instrument name (Administrator only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
■ Set the date and time (Administrator only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
■ Manage the Sign Out Timer (Administrator only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Initial start‑up
Perform these tasks during initial instrument start‑up or after you restore factory defaults (see “Restore
factory defaults” on page 87).
2. (Optional) Enter an instrument name (see “Manage the instrument name (Administrator only)” on
page 88).
3. Set the date and time (see “Set the date and time (Administrator only)” on page 89).
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 75
Chapter 6 Configure the instrument and manage instrument profiles
6 Manage profiles
5. Create an administrator profile (see “Create an administrator instrument profile during initial start-
up” on page 76).
Manage profiles
To See
Create a profile • “Create an administrator instrument profile during initial start-up” on page 76
• “Create a new instrument profile” on page 25
View or manage all • “Manage all instrument profiles (Administrator only)” on page 76
profiles
Note: Touch the Show PIN checkbox to switch PIN display on or off.
A list of users, the date the profile was created, and the user type displays.
76 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 6 Configure the instrument and manage instrument profiles
Link an instrument profile to Thermo Fisher™ Connect 6
4. Touch Done.
2. Touch Edit.
4. Touch Done.
To link an instrument profile to Thermo Fisher™ Connect, see the following sections:
• “Recommended order to set up profiles” on page 78
• “Overview of local instrument profiles and Thermo Fisher™ Connect profiles” on page 78
• “Thermo Fisher™ Connect instrument profile roles and functions” on page 79
• “Link the instrument to your Thermo Fisher™ Connect account” on page 80
• “Link a local profile to a Thermo Fisher™ Connect profile” on page 82
• “Unlink a Thermo Fisher™ Connect account” on page 83
• “If you link when you are signed in to the instrument” on page 84
• “Change the PIN for a Thermo Fisher™ Connect profile” on page 85
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 77
Chapter 6 Configure the instrument and manage instrument profiles
6 Link an instrument profile to Thermo Fisher™ Connect
78 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 6 Configure the instrument and manage instrument profiles
Link an instrument profile to Thermo Fisher™ Connect 6
Standard Thermo • Create, save, open, import, and run template files
Fisher™ • Create and modify run settings
Connect
• View and export data files
Administrator Thermo All the permissions of a local administrator profile, plus the following
Fisher™ functions performed in Thermo Fisher™ Connect:
Connect • See a list of all the Thermo Fisher™ Connect profiles that are linked
to the instrument
• Assign Thermo Fisher™ Connect administrator roles to one or more
users
• Remove a user from an instrument
• Disconnect the instrument from Thermo Fisher™ Connect
• Change the instrument name
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 79
Chapter 6 Configure the instrument and manage instrument profiles
6 Link an instrument profile to Thermo Fisher™ Connect
3. Tap Close.
80 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 6 Configure the instrument and manage instrument profiles
Link an instrument profile to Thermo Fisher™ Connect 6
After the PIN is set up on the instrument, it must be changed with your Thermo Fisher™ Connect
account on a browser, see “Create an administrator instrument profile during initial start-up” on
page 76.
The first time the instrument is used with a Thermo Fisher™ Connect account, a region must be
selected.
1. In the Sign In screen, tap Get started4Create a local instrument file4Connect to your
Thermo Fisher™ Connect account.
On the instrument:
1. Tap Mobile devices.
2. Hold the camera on your mobile device over the QR code that is displayed on the
touchscreen.
3. Tap Close.
4. (Optional) In the Enter PIN screen, tap the PIN (4 digits required) field, enter a four-digit numerical
PIN, then tap Enter.
Tap the Show PIN checkbox to show or hide the PIN.
5. (Optional) Tap the Confirm PIN field, enter the four-digit numerical PIN again, then tap Enter.
6. Tap Done.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 81
Chapter 6 Configure the instrument and manage instrument profiles
6 Link an instrument profile to Thermo Fisher™ Connect
82 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 6 Configure the instrument and manage instrument profiles
Link an instrument profile to Thermo Fisher™ Connect 6
On the instrument:
1. Tap Mobile devices.
2. Hold the camera on your mobile device over the QR code that is displayed on the
touchscreen.
3. Tap Close.
5. (Optional) In the Enter PIN screen, tap the PIN (4 digits required) field, enter a four-digit numerical
PIN, then tap Enter.
Tap the Show PIN checkbox to show or hide the PIN.
6. (Optional) Tap the Confirm PIN field, enter the four-digit numerical PIN again, then tap Enter.
7. Tap Done.
4. Tap Confirm.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 83
Chapter 6 Configure the instrument and manage instrument profiles
6 Link an instrument profile to Thermo Fisher™ Connect
Before you link: • You enter your local instrument profile name in the Sign In screen.
• Your local instrument profile (UserABC) is displayed in the home screen of the
instrument.
• All plates and results that you create are accessible only when you are signed in
with your local instrument profile.
When you link: • You link your local instrument profile (see “Link a local profile to a Thermo Fisher™
Connect profile” on page 82).
• If this is the first time you link, a Thermo Fisher™ Connect instrument
profile is created using the FirstNameLastInitial of the user name from your
thermofisher.com account.
Example: [email protected] First name is User, Last name is Gray.
Thermo Fisher™ Connect account username is User G.
• Your local instrument profile (UserABC) is linked to your Thermo Fisher™ Connect
account ([email protected]).
• Your Thermo Fisher™ Connect instrument profile (User G.) replaces your local
instrument profile.
After you link: • Your Thermo Fisher™ Connect instrument profile (User G.) and is displayed in
the home screen of the instrument.
• Plate files and data files from your local instrument profile can be transferred to
Thermo Fisher™ Connect.
• New plate files and data files are saved under your Thermo Fisher™ Connect
instrument profile.
If your Thermo Fisher™ • Your local instrument profile (UserABC) is displayed in the home screen of the
Connect account is instrument.
unlinked: • Plate files and data files that were saved under your Thermo Fisher™ Connect
instrument profile are accessible under your local instrument profile.
• Plate files and data files are saved under local instrument profile and can be
copied to Thermo Fisher™ Connect (see “Manage templates (EDT files)” on
page 42 and “Transfer EDS files from the instrument home screen” on page 41).
• Your local instrument profile name (UserABC) is available for selection in the Sign
In screen.
84 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 6 Configure the instrument and manage instrument profiles
Enable SAE mode (Administrator only) 6
3. In the Update PIN number dialog box, enter a new PIN, then enter the new PIN a second time to
confirm it.
4. Click Send.
The 4-digit PIN to sign in to instruments with your Thermo Fisher™ Connect profile is updated. This
change applies to all of the instruments that you access with your Thermo Fisher™ Connect profile.
3. Touch OK.
Note: All instrument runs must be started from desktop software that has SAE functions enabled.
• Edit an instrument name.
• Edit an instrument date or time.
• Restore to factory defaults.
• Perform a software update.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 85
Chapter 6 Configure the instrument and manage instrument profiles
6 Require instrument profile sign-in (Administrator only)
3. Touch Done.
3. Touch OK.
86 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 6 Configure the instrument and manage instrument profiles
Restore factory defaults 6
IMPORTANT! Back up the instrument before restoring factory defaults (see “Backup or restore the
instrument” on page 67).
After restoring factory defaults, perform initial start‑up tasks. See “Initial start‑up” on page 75.
2. In the Network Configuration screen, touch a radio button to connect to a network either by
DHCP or by a Static IP address.
Option Action
Automatic Touch DHCP.
Manual 1. Touch Static IP.
2. Enter the IP address for the instrument, the Subnet Mask, the Default Gateway, and
(optional) the Primary DNS Server and the Secondary DNS Server.
Addresses are in the form of X.X.X.X, where each X is a number from 0 to 255.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 87
Chapter 6 Configure the instrument and manage instrument profiles
6 Select a region for Thermo Fisher™ Connect (Administrator only)
Note: Ask your system administrator if the IP address is assigned statically or dynamically. For
static addresses, you need to know the IP address for the instrument, the Subnet Mask, and the
Default Gateway.
3. Touch OK.
2. Select a network or touch Join other network, enter the network password, then touch Enter.
If you select Join other network, enter the network name and security type.
5. Touch OK.
6. In the Network Configuration screen, edit the Network details or touch Done.
2. Touch a region.
The server used to store your data is located in this region.
3. Touch OK.
2. Touch the Instrument Name field, enter an instrument name, then touch Done.
3. Touch OK.
88 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Chapter 6 Configure the instrument and manage instrument profiles
Set the date and time (Administrator only) 6
b. Touch Next, touch the date field, then enter the date.
c. Touch Next, touch the time field, then enter the time.
5. Touch Done.
2. Touch the Edit Time field, then enter the desired duration of inactivity before automatic user sign
out.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 89
A Install and connect the instrument to
a network
90 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Appendix A Install and connect the instrument to a network
Before you begin A
2. Follow the pre-printed instructions on the instrument box to unpack the instrument, accessories,
and reference documentation. Save the packing material for future use or recycle it.
The instrument box contains:
• Reference documentation: Welcome note, unpacking and set up instructions card, system
documentation insert
• One instrument
• Accessories: power cable, Ethernet cable, USB drive, reaction tube retainer
• Shipping plate
Note: Save the shipping plate but do not use it to operate the instrument.
4. Plug the power cable into the power port on the back panel of the instrument, then plug the cable
into an electrical receptacle.
5. Connect your instrument as required by your network configuration (see “Supported options for
instrument and computer connections” on page 95 and “Instrument and computer connections”
on page 93).
• Connect an Ethernet cable to the Ethernet port on the back panel of the instrument, then
connect the cable to a computer or a networking port.
or
• Connect wirelessly via the wireless adapter.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 91
Appendix A Install and connect the instrument to a network
A Power on and follow the startup wizard
Note: Do not connect the High Power USB WiFi Module (Cat. No. A26774) to the instrument if
it is connected to a network by an Ethernet cable. Configuring the instrument for both wired and
wireless connection can interfere with instrument operation.
Note: You can perform any of the steps above at a later time if you do not have the information
needed to complete the startup screens. See “Configure instrument settings” on page 26.
IMPORTANT! Before using the instrument for the first time, we recommend that you perform
instrument verification using RNase P plates (see “Perform instrument verification using RNase P
plates” on page 53).
3. Open the Control Panel, then open the Network and Sharing Center.
4. In the left pane of the Network and Sharing Center window, click Change Adapter Settings.
7. Click Properties.
92 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Appendix A Install and connect the instrument to a network
Instrument and computer connections A
8. In the Internet Protocol Version 4 (TCP/IPv4) Properties dialog box, select one of the following
options.
• Obtain an IP address automatically
• Use the following IP address
10. Close all dialog boxes by clicking OK, then re-start the computer.
5
6
7
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 93
Appendix A Install and connect the instrument to a network
A Download and install the desktop software
3 2
1 1
1 Detachable power supply cord compatible with local power supply receptacle.
2 Connection between the computer and the instrument.
3 Connection between the computer and the monitor, keyboard, and mouse.
4 Connection between the computer and the (optional) handheld barcode scanner.
94 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Appendix A Install and connect the instrument to a network
Networking A
2. Go to thermofisher.com/quantstudio3-5softwaredownloads.
3. Double-click setup.exe.
7. If you are installing the software on a computer provided by Thermo Fisher Scientific, install
the software on the D:\ drive. If you are installing on a customer-provided computer, install the
software in your preferred location.
8. Click Finish.
Networking
IMPORTANT! This section provides general networking information. It does not provide adequate
detail to integrate the instrument into all possible network architectures. Because a network may
contain advanced features (such as a firewall or network domains), we recommend that you consult
a network administrator before connecting the instrument to your laboratory network.
IMPORTANT! Do not connect both an Ethernet cable and the Wi‑Fi module (Cat. No. A26774) to
the instrument. Configuring the instrument for both wired and wireless connection can interfere with
instrument operation.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 95
Appendix A Install and connect the instrument to a network
A Networking
Note: For detailed information about networking your instrument, see the QuantStudio™ 3 and 5
Real‑Time PCR Systems IT Checklist (Pub. No. MAN0013738).
• A computer provided by Thermo Fisher Scientific with the QuantStudio™ Design and Analysis Software
• Computer‑to‑instrument connection:
– Direct, wired connection between the computer and the instrument
using an Ethernet cable
• A computer provided by Thermo Fisher Scientific with the QuantStudio™ Design and Analysis Software
• The computer and instrument must have the same subnet mask within the same network.
• Internet access and a computer with the Chrome™ web browser to access Thermo Fisher™ Connect
96 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Appendix A Install and connect the instrument to a network
Networking A
Note: Because mDNS is limited to direct network connections, an instrument configured for
mDNS may not be visible to other nodes that are separated by a router, hub, or another network
device.
• IPv4 linknlocal (IPV4LL) in the RFC (also known as Automatic Private IP Addressing [APIPA] or
Internet Protocol Automatic Configuration [IPAC]).
Note: When an instrument is set for DHCP, APIPA is automatically enabled, and the instrument
provides an IP address when no address is supplied by the DHCP server.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 97
Appendix A Install and connect the instrument to a network
A Networking
98 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
B Troubleshooting
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 99
Appendix B Troubleshooting
B Networking
100 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
C Parts and materials
10 plates N8010560
MicroAmp™ Optical 96-Well Reaction Plate
500 plates 4316813
20 plates 4306737
MicroAmp™ Optical 96-Well Reaction Plate with Barcode
500 plates 4326659
MicroAmp™ EnduraPlate™ Optical 96-Well Clear Reaction Plates with 20 plates 4483354
Barcode 500 plates 4483352
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 101
Appendix C Parts and materials
C 384-well consumables
(continued)
Consumable Amount Cat. No.
Region of Interest (ROI) and Background Plates, 96‑Well 0.2‑mL (2 plates) 4432364
384-well consumables
Consumable Contents Cat. no.
102 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Appendix C Parts and materials
Consumables (96-well, 0.1-mL format) C
MicroAmp™ Optical Fast 8-Tube Strip (0.1 mL) 125 strips 4358293
MicroAmp™ Optical Fast Tube with Cap (0.1 mL) 1000 tubes 4358297
MicroAmp™ Optical 96-Well Fast Reaction Plate (0.1 mL) 10 plates 4346907
MicroAmp™ EnduraPlate™ Optical 96-Well Fast Reaction Plate with 20 plates 4481194
Barcode (clear) (0.1 mL)
500 plates 4483494
96-Well Fast Region of Interest (ROI) and Background Plates (2 plates) 4432426
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 103
Appendix C Parts and materials
C Accessories
Accessories
Item Amount Cat. No.
Material/Consumable Source
104 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
D Instrument specification and layout
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 105
Appendix D Instrument specification and layout
D Configured system dimensions
15 cm (6 in.)
50 cm (20 in.)
27 cm (10.5 in.)
30 cm (12 in.)
40 cm (16 in.)
106 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Appendix D Instrument specification and layout
Electrical requirements D
Instrument clearances
During instrument installation and maintenance, it is necessary to access the back of the instrument. If
the back of the instrument faces a wall, ensure that there is sufficient clearance on the bench to rotate
the instrument for access.
IMPORTANT! For safety, the power outlet for the instrument must be accessible.
Electrical requirements
WARNING! For safety, the power outlet used for powering the instrument must be accessible at all
times. See “Instrument clearances” on page 107 for information about the space needed between the
wall and the instrument. In case of emergency, you must be able to immediately disconnect the main
power supply to all the equipment. Allow adequate space between the wall and the equipment so
that the power cords can be disconnected in case of emergency.
WARNING! Use of accessories, transducers, and cables other than those specified or provided by
the manufacturer of the equipment could result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and result in improper operation.
Circuit Rated
Device Rated voltage Rated power
required frequency
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 107
Appendix D Instrument specification and layout
D Environmental requirements
Environmental requirements
Table 2 Environmental requirements
Electromagnetic interference Do not use this device in close proximity to sources of strong electromagnetic
radiation (for example, unshielded intentional RF sources). Strong electromagnetic
radiation may interfere with the proper operation of the device.
Altitude Between sea level and 2000 m (6500 ft) above sea level
Thermal output During operation, the net thermal output, based on the actual current draw of the
instrument, is expected to be approximately 960 W (3275 Btu/h).
Vibration Ensure that the instrument is not adjacent to strong vibration sources, such as
a centrifuge, pump, or compressor. Excessive vibration will affect instrument
performance.
Pollution degree The instrument has a Pollution Degree rating of II. The instrument may only
be installed in an environment that has nonconductive pollutants such as dust
particles or wood chips. Typical environments with a Pollution Degree II rating are
laboratories and sales and commercial areas.
The noise output of the instrument is £ 60 dB when running.
Other conditions Ensure the instrument is located away from any vents that could expel particulate
material onto the instrument components.
Avoid placing the instrument and computer adjacent to heaters, cooling ducts, or
in direct sunlight.
Network requirements
The instrument:
• Is factory-configured for IPv4 TCP/IP communication and includes an Ethernet adapter
(100/1,000 Mbps) with an RJ45-type connector for integrating the device into a local area network
(LAN).
• Can alternatively be configured for wireless networking (High Power USB WiFi Module required,
sold separately as an optional accessory).
108 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Appendix D Instrument specification and layout
Network requirements D
The instrument can be configured for either wired or wireless networking, not both.
If a Thermo Fisher Scientific service representative is to install the instrument:
• If the instrument will be connected to a LAN, an active, tested network jack must be in place before
the scheduled installation date.
• A representative from your information technologies department must be available during the
installation to help connect the instrument to your network.
–or–
• Wireless—High Power USB WiFi Module (Cat. No. A26774, sold separately)
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 109
E Safety
WARNING! GENERAL SAFETY. Using this product in a manner not specified in the user
documentation may result in personal injury or damage to the instrument or device. Ensure that
anyone using this product has received instructions in general safety practices for laboratories and
the safety information provided in this document.
· Before using an instrument or device, read and understand the safety information provided in the
user documentation provided by the manufacturer of the instrument or device.
· Before handling chemicals, read and understand all applicable Safety Data Sheets (SDSs) and use
appropriate personal protective equipment (gloves, gowns, eye protection, and so on). To obtain
SDSs, see the “Documentation and Support” section in this document.
110 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Appendix E Safety
Symbols on this instrument E
(continued)
On On (marche)
Standby En attente
Terminal that can receive or supply Borne pouvant recevoir ou envoyer une
alternating current or voltage tension ou un courant de type alternatif
Terminal that can receive or supply Borne pouvant recevoir ou envoyer une
alternating or direct current or voltage tension ou un courant continu ou alternatif
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 111
Appendix E Safety
E Symbols on this instrument
(continued)
Do not dispose of this product in unsorted Ne pas éliminer ce produit avec les déchets
municipal waste usuels non soumis au tri sélectif.
Conformity symbols
Conformity mark Description
112 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Appendix E Safety
Safety alerts on this instrument E
English Français
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 113
Appendix E Safety
E Safety information for instruments not manufactured by Thermo Fisher Scientific
Instrument safety
General
CAUTION! Do not remove instrument protective covers. If you remove the protective instrument
panels or disable interlock devices, you may be exposed to serious hazards including, but not limited
to, severe electrical shock, laser exposure, crushing, or chemical exposure.
CAUTION! Solvents and Pressurized fluids. Wear eye protection when working with any
pressurized fluids. Use caution when working with any polymeric tubing that is under pressure:
114 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Appendix E Safety
Instrument safety E
Physical injury
CAUTION! Moving and Lifting Injury. The instrument is to be moved and positioned only by the
personnel or vendor specified in the applicable site preparation guide. Improper lifting can cause
painful and permanent back injury.
Things to consider before lifting or moving the instrument or accessories:
· Depending on the weight, moving or lifting may require two or more persons.
· If you decide to lift or move the instrument after it has been installed, do not attempt to do so
without the assistance of others, the use of appropriate moving equipment, and proper lifting
techniques.
· Ensure you have a secure, comfortable grip on the instrument or accessory.
· Make sure that the path from where the object is to where it is being moved is clear of
obstructions.
· Do not lift an object and twist your torso at the same time. Keep your spine in a good neutral
position while lifting with your legs.
· Participants should coordinate lift and move intentions with each other before lifting and carrying.
· For smaller packages, rather than lifting the object from the packing box, carefully tilt the box on its
side and hold it stationary while someone else slides the contents out of the box.
CAUTION! Moving Parts. Moving parts can crush, pinch and cut. Keep hands clear of moving parts
while operating the instrument. Disconnect power before servicing.
WARNING! Do not attempt to lift or move the instrument without the assistance of others. Use
appropriate moving equipment and proper lifting technique, improper lifting may result in serious
injury.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 115
Appendix E Safety
E Safety and electromagnetic compatibility (EMC) standards
Electrical safety
WARNING! Fuse Installation. Before installing the instrument, verify that the fuses are properly
installed and the fuse voltage matches the supply voltage. Replace fuses only with the type and
rating specified for the unit. Improper fuses can damage the instrument wiring system and cause a
fire.
WARNING! Ensure appropriate electrical supply. For safe operation of the instrument:
· Plug the system into a properly grounded receptacle with adequate current capacity.
· Ensure the electrical supply is of suitable voltage.
· Never operate the instrument with the ground disconnected. Grounding continuity is required for
safe operation of the instrument.
WARNING! Power Supply Line Cords. Use properly configured and approved line cords for the
power supply in your facility.
WARNING! Disconnecting Power. To fully disconnect power either detach or unplug the power
cord, positioning the instrument such that the power cord is accessible.
CAUTION! Cleaning and Decontamination. Use only the cleaning and decontamination methods
specified in the manufacturer's user documentation. It is the responsibility of the operator (or other
responsible person) to ensure the following requirements are met:
· No decontamination or cleaning agents are used that could cause a HAZARD as a result of a
reaction with parts of the equipment or with material contained in the equipment.
· The instrument is properly decontaminated a) if hazardous material is spilled onto or into the
equipment, and/or b) prior to having the instrument serviced at your facility or sending the
instrument for repair, maintenance, trade-in, disposal, or termination of a loan (decontamination
forms may be requested from customer service).
· Before using any cleaning or decontamination methods (except those recommended by the
manufacturer), users should confirm with the manufacturer that the proposed method will not
damage the equipment.
116 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Appendix E Safety
Safety and electromagnetic compatibility (EMC) standards E
Safety compliance
Reference Description
IEC 61010-1 Safety requirements for electrical equipment for measurement, control, and
laboratory use – Part 1: General requirements
EN 61010-1
UL 61010-1
CSA C22.2 No. 61010-1
IEC 61010-2-010 Safety requirements for electrical equipment for measurement, control and
laboratory use – Part 2-010: Particular requirements for laboratory equipment
EN 61010-2-010
for the heating of materials
IEC 61010-2-081 Safety requirements for electrical equipment for measurement, control and
laboratory use – Part 2-081: Particular requirements for automatic and semi-
EN 61010-2-081
automatic laboratory equipment for analysis and other purposes
EMC
Reference Description
EN 61326-1/ IEC 61326-1 Electrical Equipment for Measurement, Control and Laboratory Use –
EMC Requirements – Part 1: General Requirements
ICES-001, Issue 4 Industrial, Scientific and Medical (ISM) Radio Frequency Generators
FCC Part 15 Subpart B (47 CFR) U.S. Standard Radio Frequency Devices
This equipment has been tested and found to comply with the limits for a Class A digital device,
pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection
against harmful interference when the equipment is operated in a commercial environment. This
equipment generates, uses, and can radiate radio frequency energy and, if not installed and used
in accordance with the instruction manual, may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful interference in which case the
user will be required to correct the interference at his own expense.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 117
Appendix E Safety
E Safety and electromagnetic compatibility (EMC) standards
Environmental design
Reference Description
Directive 2012/19/EU European Union “WEEE Directive” – Waste electrical and electronic equipment
PCBA X O O O O O
118 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Appendix E Safety
Chemical safety E
Chemical safety
WARNING! GENERAL CHEMICAL HANDLING. To minimize hazards, ensure laboratory personnel
read and practice the general safety guidelines for chemical usage, storage, and waste provided
below. Consult the relevant SDS for specific precautions and instructions:
· Read and understand the Safety Data Sheets (SDSs) provided by the chemical manufacturer
before you store, handle, or work with any chemicals or hazardous materials. To obtain SDSs, see
the "Documentation and Support" section in this document.
· Minimize contact with chemicals. Wear appropriate personal protective equipment when handling
chemicals (for example, safety glasses, gloves, or protective clothing).
· Minimize the inhalation of chemicals. Do not leave chemical containers open. Use only with
sufficient ventilation (for example, fume hood).
· Check regularly for chemical leaks or spills. If a leak or spill occurs, follow the manufacturer
cleanup procedures as recommended in the SDS.
· Handle chemical wastes in a fume hood.
· Ensure use of primary and secondary waste containers. (A primary waste container holds the
immediate waste. A secondary container contains spills or leaks from the primary container.
Both containers must be compatible with the waste material and meet federal, state, and local
requirements for container storage.)
· After emptying a waste container, seal it with the cap provided.
· Characterize (by analysis if needed) the waste generated by the particular applications, reagents,
and substrates used in your laboratory.
· Ensure that the waste is stored, transferred, transported, and disposed of according to all local,
state/provincial, and/or national regulations.
· IMPORTANT! Radioactive or biohazardous materials may require special handling, and disposal
limitations may apply.
WARNING! 4L Reagent and Waste Bottle Safety. Four-liter reagent and waste bottles can crack
and leak. Each 4-liter bottle should be secured in a low-density polyethylene safety container with the
cover fastened and the handles locked in the upright position.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 119
Appendix E Safety
E Biological hazard safety
WARNING! BIOHAZARD. Biological samples such as tissues, body fluids, infectious agents,
and blood of humans and other animals have the potential to transmit infectious diseases.
Conduct all work in properly equipped facilities with the appropriate safety equipment (for example,
physical containment devices). Safety equipment can also include items for personal protection,
such as gloves, coats, gowns, shoe covers, boots, respirators, face shields, safety glasses, or
goggles. Individuals should be trained according to applicable regulatory and company/ institution
requirements before working with potentially biohazardous materials. Follow all applicable local,
state/provincial, and/or national regulations. The following references provide general guidelines when
handling biological samples in laboratory environment.
· U.S. Department of Health and Human Services, Biosafety in Microbiological and Biomedical
Laboratories (BMBL), 6th Edition, HHS Publication No. (CDC) 300859, Revised June 2020
https://www.cdc.gov/labs/pdf/CDC-BiosafetymicrobiologicalBiomedicalLaboratories-2020-
P.pdf
· Laboratory biosafety manual, fourth edition. Geneva: World Health Organization; 2020 (Laboratory
biosafety manual, fourth edition and associated monographs)
www.who.int/publications/i/item/9789240011311
120 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
F Documentation and support
Related documentation
Publication
Document Description
number
QuantStudio™ Real‑Time PCR System Help MAN0010422 Describes the instrument touchscreen and
provides procedures for configuration,
calibration, and performing a run.
QuantStudio™ Design and Analysis MAN0010409 Describes how to use the command-line
Desktop Software Command-Line software application and provides the
Application Guide procedure to automate the creation of
new experiment files and export data from
existing files.
QuantStudio™ Design and Analysis MAN0010408 Describes how to design and perform
Desktop Software User Guide experiments using the software.
QuantStudio™ Design and Analysis MAN0010415 Describes the software and provides
Desktop Software Help procedures for common tasks.
QuantStudio™ 5 Real‑Time PCR System MAN0010417 Describes the SAE Administrator Console
SAE Admin Console Help and provides procedures for common
tasks.
QuantStudio™ 5 Real‑Time PCR System MAN0010410 Describes how to use the SAE
SAE Admin Console User Guide Administrator Console.
QuantStudio™ Design and Analysis Cloud MAN0010414 Describes the Cloud version of the
Software Help software and provides procedures for
common tasks.
QuantStudio™ 3 and 5 Real‑Time PCR MAN0010405 Explains how to prepare your site to
Systems Site Preparation Guide receive and install the system. Intended
for personnel who schedule, manage, and
perform the tasks required to prepare the
site for installation.
Note: For additional documentation, see “Customer and technical support” on page 122.
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 121
Appendix F Documentation and support
F Obtain information from the Help system
Note: For SDSs for reagents and chemicals from other manufacturers, contact the
manufacturer.
122 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Index
A set up 92
Connect
accessories 104
Link to instrument 80
add to method 35
Roles
administrator instrument profile 76
Administrator 79
Administrator profile 79
Standard 79
Unlink from instrument 83
B Connect profile, Overview 78
connect to network, workflow 90
background calibration
connection, not recognized
create plate 52
consumables
description 44
384-well 102
plate, prepare 47
96-well 0.1-mL 103
Background calibration
96-well 0.2-mL 101
run 48
contamination, identify 52
when to perform 43
custom dye
backup instrument 11, 67
add to the instrument 63
barcode scan 33
calibration plate, create 62
biohazard safety 120
calibration, perform 63
description 13
C dilution guidelines 59
optimal concentration 61
calibration
wavelength requirements 59
reminders 45
custom dye calibration, workflow 59
Background 48
custom dye dilution plate, prepare 60
custom dye. See custom dye
custom melt calibration 65
custom melt 65
Dye 48
plate, prepare 47 D
reminders 45
data collection 14
ROI/Uniformity 48
data collection points, add to method 34
schedule 43
date and time 89
status 45
delete files 42
transfer results 50
desktop software
view images 50
download 95
workflow 46
download and install 94
calibration failure, troubleshoot 51
install 95
Change password 85
disk space message
Change PIN 85
documentation, related 121
Cloud server, select location 88
download desktop software 95
comparative Ct runs 19
dye calibration
computer
description 44
connection to instrument or LAN 95
plate, prepare 47
connections 93
Dye calibration
requirements for desktop software 94
run 48
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 123
Index
N
H network
Help system, access 122 configurations supported 95
home screen, parts of 23 guidelines and best practices 98
wired connection set up 87
wireless connection set up 88
I network configuration and security 10
install, desktop software 95 networking 95
installation
before you begin 91
desktop software 94 O
workflow 90 online Help. See Help system
instrument Order to set up profiles 78
profile, create 25
connections 93
enable for monitoring 86 P
move 73 Password, Change 85
name 88 password security 10
overview 11 pause run 40
parts of 15 performance verification. See RNase P verification
power off 72 PIN, Change 85
power on 22, 72 PIN reset 76
prepare to store or move or ship 73 presence/absence runs 19
124 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
Index
Profile S
Connect
SAE mode (Security, Audit, and e‑Signature) 85
Link 80
safety, biohazard 120
Unlink 83
sample block
Edit 85
clean 68
profiles
decontaminate 68
administrator 76
settings 26
delete 76
sign in, require 86
edit 77
sign out
guest, disable 86
profile 26
instrument 25
timer, set 89
require sign-in 86
sleep mode 72
reset PIN 76
software features 15
sign in 25
software updates 86
Profiles
stages, add to method 34
Connect
standard curve runs 19
Administrator 79
Standard profile 79
Standard 79
standby mode 72
Order to set up 78
startup wizard 92
properties, enter or edit 32
steps, add to method 34
protocol. See method
stop run 40
support, customer and technical 122
R supported dyes 12
real-time plots symbols, safety 110
view on instrument touchscreen 40 system dyes 12
zoom 41
related documentation 121 T
relative standard curve runs 19
template properties. See properties
Replace with index term
terms and conditions 122
require sign in 86
thermal protocol. See method
restore instrument 11, 67
Thermo Fisher Connect profile, Edit 85
RNase P verification
third-party software 18
pass criteria 55
transfer files 42
plate description 53
troubleshooting 99
run 56
RNase P verification, plate, prepare 55
ROI/uniformity calibration U
description 44
uniformity calibration
plate, prepare 47
description 44
ROI/Uniformity calibration
plate, prepare 47
run 48
unload plate 37, 38
when to perform 43
unpack instrument 91
run
use the instrument, workflow 29
pause or stop 40
post-PCR 14
real-time 14 V
run data, transfer 41
verification failure, troubleshoot 57
run history 42
run last experiment 31
run protocol. See method W
runs, types 19
warranty 122
well details
QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide 125
Index
126 QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide
QS 35 Instrument UG_RUO_MAN0010407EN no translations-v12-GUID-88D256DF-7F65-4A04-
B662-0DCAA6DC61A6-2021/07/21 16:29:30 en
16:15:10.886+01:00
thermofisher.com/support | thermofisher.com/askaquestion
thermofisher.com
22 July 2021