C80-Specification V1.5 2021.11.04

This document provides specifications for the COMEN C80 patient monitor. It can monitor ECG, respiration, temperature, SpO2, NIBP, and heart rate as standard. Optional configurations include IBP, EtCO2, AG, ICG, C.O., BIS, and multi-parameter monitoring. The monitor has a 12.1" color touch screen, stores data for future recall, and can connect to a network for central monitoring. It provides accurate measurements and alarms for key vital signs.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
164 views6 pages

C80-Specification V1.5 2021.11.04

This document provides specifications for the COMEN C80 patient monitor. It can monitor ECG, respiration, temperature, SpO2, NIBP, and heart rate as standard. Optional configurations include IBP, EtCO2, AG, ICG, C.O., BIS, and multi-parameter monitoring. The monitor has a 12.1" color touch screen, stores data for future recall, and can connect to a network for central monitoring. It provides accurate measurements and alarms for key vital signs.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 6

Specification: C80

SHENZHEN COMEN MEDICAL INSTRUMENTS CO.,LTD


No.2 of FIYATA Timepiece Building, Nanhuan Avenue, Gongming
sub-district, Guangming New District, Shenzhen, P.R. China
Tel: +86-755-26408879
Fax:+86-755-26431232
Email: [email protected]
Web: www.comen.com

2021 V 1.5
Patient Monitor
C80

Standard Configuration: Power indicator


5-lead ECG, RESP, Temp (Single Channel), COMEN SpO2, Battery indicator
NIBP, HR QRS beep and alarm sound
Optional Configuration: Operating key sound
Dual-IBP, EtCO2, AG, ICG, C.O., BIS, Nellcor/Masimo SpO2, Interface:
3/12-lead ECG, Thermal Recorder, Dual-Temp, Suntech Parameter cable interface
NIBP, Trolley, Wall mount, Ground wire AC power input socket
Safety Standards: Two USB port
IEC 60601-1 IEC 60601-1-8 IEC 60601-2-27 EN 1060-3 IEC RJ45 port
80601-2-30 IEC60601-2-34 IEC60601-2-49 ISO 80601-2-56 Optional Multi-functional interface
ISO 80601-2-61 VGA output
Physical Characteristics: Data storage
Product Size: 347mm*291mm*165mm Alarm Event Recall: 200 groups
Weight: 3.9kg Wave Recall: 6 hours (8 waves)
Display: 12.1” color TFT touch screen NIBP Recall: 2000 groups
Resolution: 800*600 Trend Graph: 160 hours
Trace: 10 waveforms Trend Table: 160 hours
IP grade IPX1 Power-off Yes
storage:
Operation Environment: Alarm: User-adjustable High and Low 3-level
Working Limits;
Temperature 5-40°C Prioritized audible and visual alarm
Humidity: ≤93% Network: Connected to Central Monitoring
Power Supply 100-240V~, 50/60Hz±1Hz System by hardwire/wireless
Battery Type: Rechargeable Lithium-ion battery Recorder:
Battery Capacity: 2200mAh (optional: 4400mAh) Type: Built-in; Thermal array
Battery Recharging Channel: 3 channel waveforms
Time: Maximum 5.5 hours for charging; Speed: 25mm/s,50mm/s
Battery backup: 2 hours for continuous working Record width: 50mm
Real-time record
Indicator: time 8s, 16s, 32s or continual
One alarm indicator Alarm record Yes
Respiration: ST SEGMENT
Method: RA-LL Impedance Method detection: -2.0mV~+2.0mV (Automatic)
RR measurement Arrhythmia
Range Adult: 0-120 rpm Analysis: 26 types
Pediatric/Neonate:0-150 rpm Pacemaker
Accuracy: 7~150rpm: ±2rpm or 2%, whichever detection: Detectable
is greater Alarm: Yes, audible and visual alarm, alarm
0-6rpm: unspecified events reviewable
Resolution: 1 rpm 12 lead ECG
RESP Apnea 10s-60s (Adu); 10s-40s (Ped/Neo) Analysis: Yes
Alarm: Audible and visual alarm; alarm NIBP:
events reviewable Method Automatic oscillation
Sweep Speed: 6.25,12.5,25mm/s Work mode: Manual / Automatic/Continual (5min,
Gain Selection: X0.25, X0.5, X1, X2, X4 not applicable to neonates)
ECG: Measurement time: Adjustable(1-480min)
Lead Type: CardioTecTM 5-leads ECG Analysis, 12- Maximum
Lead and 3-leads electable measurement time Adu/Ped: 120s; Neo: 85s
Lead selection 12-Lead I; II; III; aVR; aVL; aVF; V1-V6. Measurement Unit: mmHg / kPa selectable
5-lead: I; II; III; aVR; aVL; aVF; V Measurement
3-lead: I; II; III types: Systolic, Diastolic, Mean
Waveform 5-lead: 2 -channel Range of systolic
3-lead: 1 -channel pressure: Adult Mode: 40-270mmHg
Gain Selection: X0.125, X0.25, X0.5, X1, X2, X4, auto Pediatric Mode: 40-200mmHg
error<±5% Neonate Mode: 40-135mmHg
Sweep Speed: 6.25,12.5, 25, 50mm/s, error≤±10% Range of diastolic
Resp, lead AC waveform: Current :<0.1µA; pressure: Adult Mode: 10-215mmHg
disconnection Frequency: 64kHz, ±10% Pediatric Mode: 10-150mmHg
detection and Neonate Mode: 10-100mmHg
active noise Range of mean
control: pressure: Adult Mode:20-235mmHg
CMRR ≥105 dB Pediatric Mode:20-165mmHg
Heart Rate Neonate Mode 20-110mmHg
measurement Static pressure
Range: Adult: 15~300bpm range and accuracy:
Pediatric/Neonate:15~350bpm 0~300mmHg(0kPa~40.0kPa)
Accuracy: ±1% or ±1bpm (whichever is greater) ±3mmHg(±0.4kPa)
Protection: Withstand 4000VAC/50Hz voltage in Over-pressure
isolation, Again electrosurgical protection: Adult Mode: 297mmHg
interference and defibrillation Pediatric Mode: 240mmHg
Accuracy: ±1% or ±1bpm (whichever is greater) Neonate Mode: 147mmHg
Band width: Monitoring Mode: 0.5-40Hz Accuracy: ±3mmHg
Diagnosis mode: 0.05-150Hz Initial pressure
Surgery mode:1-20Hz range(mmHg): Adult: 80~240;Pediatric: 80~200;
ST mode: 0.05-40Hz Neonate:60~120
Alarm: Systolic, Diastolic, Mean Resolution: 1%
PR from NIBP: Measurement & alarm range: 40- Accuracy: ±2% (70~100%, Adu/Ped, non-
240bpm motion)
Resolution: 1bpm ±3% (70-100%, Neo, non-motion)
Accuracy: ±3bpm or ±3%, whichever Unspecified (1-69%)
is greater Data averaging and
Nellcor SpO2: other signal
Measurement processing time: 2s
Range 0-100% Data refresh rate: 8s
Alarm Range 20-100% PR Measurement
Resolution: 1% Range: 20--254bpm
Accuracy: ±2% (70-100%, Adu/Ped, non- Resolution: 1bpm
motion) Accuracy: ±2bpm
±3% (70-100%, Neo, non-motion) Alarm range: 20~254bpm
1-69% unspecified Perfusion index: 0.05%~20%
Alarm Range: 20-100% Resolution: 0.01% (within
PR Measurement 0.05%~9.99% range) or 0.1% (within
Range: 20-300bpm 10.0%~20.0% range)
Resolution: 1bpm Temperature (Dual Channel)
Accuracy: ±3bpm (20-250bpm); Range: 0-50°C
Unspecified (251-300bpm) TEMP sensor: Skin/rectal TEMP sensor
Alarm Range: 20~300bpm Resolution: 0.1°C
Masimo SpO2: Accuracy: ±0.1°C (exclusive of error of sensor)
Measurement & Channel: T1, T2, TD (Temperature Difference)
alarm Range 1~100% EtCO2
Resolution: 1% Unit: mmHg, kPa
Accuracy: ±2% (70-100%, Adu/Ped, non- Measurement
motion) range: 0mmHg~150mmHg
±3% (70-100%, Neo, non-motion) Resolution 1mmHg or 0.1kPa or 0.1%
1-69% unspecified Accuracy 0mmHg ~40mmHg should
Alarm range 1~100% be±2mmHg;
PR Measurement 41mmHg ~70mmHg should
Range 25~240bpm be±5%×reading;
Resolution: 1bpm 71mmHg ~100mmHg should
Accuracy: ±3bpm (stationary) be±8%×reading;
±5bpm (motional); 101mmHg~150mmHg should
Alarm range: 25~240bpm be±10%×reading
Perfusion index: 0.02~20% Oxygen
Resolution: 0.01% (within compensation 0~100 mmHg
0.02%~9.99% range) or 0.1% (within Equilibrium gas Helium, room air, nitrous oxide
10.0%~20.0% range)
COMEN SpO2: IBP
Measurement Channel: 2 Channels
& alarm Range 0~100%
Measured Pressure: ART, PA, CVP, RAP, LAP, ICP, LV, AO, HAL, ISO, ENF: 0~8%: ±
UAP, BAP, FAP, UVP, IAP, P1, P2, P3, (0.15%+reading×5%); 8~25vol%:
P4 unspecified
Measurement Unit: mmHg/ kPa selectable SEV: 0~10%: ± (0.15%+reading×5%);
Measurement 10~25vol %: unspecified
Range: ART: 0~300mmHg DES: 0~22%: ± (0.15%+reading×5%);
PA: -6~120 mmHg 22~25%: unspecified
CVP: -10~40mmHg O2: 0-100%: ± (1%+reading×2%)
RAP: -10~40mmHg Data output: Fi and Et values
LAP: -10~40mmHg AG resolution: CO2: 1mmHg
ICP: -10~40mmHg awRR: 1rpm
LV: 0~300mmHg Accuracy: For all measured values complies
AO: 0~300mmHg with EN ISO 21647:2004 and EN
UAP: 0~300mmHg 864:1996
BAP: 0~300mmHg Alarm: EtCO2: 0mmHg~190mmHg
FAP: 0~300mmHg Fi CO2: 0mmHg~190mmHg
UVP: -10~ 40mmHg AwRR: 2mmHg~150mmHg
IAP: -10~40mmHg EtO2: 18% ~ 100%
P1, P2, P3, P4: -50~300mmHg FiO2: 18% ~ 100%
Accuracy: ±2% or ±1mmHg, whichever is EtN2O: 0% ~ 100%
Resolution: greater FiN2O: 0% ~ 82%
0.1kPa or 1mmHg EtHal/EtEnf/EtIso/EtSev/EtDes: 0% ~
(-50mmHg~300mmHg) 25%
Alarm Range: -50mmHg~300mmHg FiHal/FiEnf/FiIso/FiSev/FiDes: 0% ~
Pressure sensor: sensitivity: 5 V/V/mmHg 25%
Impedance range: 300~3000Ω Others: Up to 4 waveforms displayed
PR from IBP: Measurement & alarm range: MAC value displayed
20bpm~350bpm ISATM (AG) Sidestream Gas Analyzer
Resolution: 1bpm Method: Infrared gas measurement
Accuracy: ±1bpm or ±1%, whichever No Breaths
is greater Timeout range Adult: 10s, 15s, 20s, 25s, 30s, 35s,
40s ,45s, 50s, 55s or 1min;
AG Pediatric and neonate: 20s, 25s, 30s,
AG (complies with ISO 80601-2-55) 35s or 40s
Method: Infrared Radiation Absorption Accuracy: ±5s
Characteristics No Breath Alm
AG preheating time Delay: 10s, 15s, 20s, 25s, 30s, 35s, 40s, 45s,
<20s 50s, 55s, 1min or Off
Gas sorts: CO2, N2O, DES, ISO, ENF, SEV, HAL, Working
O2 (optional paramagnetic sensor) conditions: ISA AX+: 0~50°C (32~122°F); ISA OR+:
Measurement 5~50°C (41~122°F)
range: CO2: 0~15%: ±(0.2kPa+reading×2%), Storage conditions -40~70°C (-40~158°F)
15~25%: unspecified RH <4kPa H2O (non-condensing) 95%
N2O: 0~100 %: ±(2kPa+reading×2%) RH, 30°C
Barometric BT, IT: ±0.1°C (no sensor)
pressure 52.5~120kPa (4572m) Alarm Range: BT Hi limit: (LO limit +0.4)-43°C
Water treatment Sampling tube: patented dehydration BT Lo limit:25.0~(Hi limit-0.4) °C
tube Step: 0.1°C
Data output: Fi and Et values Impedance Cardiography (ICG)
Waveform: Display up to 4 gas concentration
waveforms at a time Method: Indirect measurement by the
Diagnostic impedance cardiogram
parameter: Barometric pressure Measurement
ISA sensor: 2~9-channel NDIR gas analyzer range: SV: 5~250 ml/ beat
(measurement range: 4~10µm) HR: 40~250 bpm
Compensation: CO2 broadening effect C.O.: 1.4~15 L/min
Calibration No calibration is required. The Accuracy: HR: ±2bpm
Monitor will auto perform zeroing C.O.: Unspecified
when powered on and perform auto SV: unspecified
zeroing every 24h (ISACO2) or 8h (ISA Alarm range: C.I.:0~15.0L/min/m2
AX+/OR+) subsequently. TFC:10~150KΩ
Preheating time ISA CO2: <10s; ISA OR+/AX+: < 20s
Alarm Deviation C.I.: ±0.1L/min/m2
Rise time CO2: ≤250ms; N2O: ≤ 350ms;
TFC: ±1k Ω
anesthetic gases: ≤ 350ms; O2: ≤
450ms BIS
Overall system Measurement
response time <3s (2m sampling tube) range: BIS: 0~100; Accuracy: 1%
Respiration SQI: 0~100%; Accuracy: 1%
detection Self-adaptive threshold (minimum EMG: 0~100dB; Accuracy: 1%
CO2 concentration change: 1 vol%) ESR: 0~100%; Accuracy: 1%
RR 0~150 breaths/min Resolution: 1
Anesthetic gas Alarm range: BIS: 0~99
threshold Threshold of main anesthetic gases Wall Mounting
(ISA OR+/AX+): 0.15 vol%. The Net weight: 3.1kg
concentration of any identified Bracket: 275*150*165mm
anesthetic gas will be reported, even Length 330mm
if it is lower than 0.15 vol% (Support Arm)
Cardiac Output (C.O.) Max. Slip 140mm
Method: Thermodilution Rotation 180°
Range: C.O.: 0.1~20L/min
BT: 25~43°C
IT: 0~25°C *Notice: Specifications subject to changes
Resolution: C.O.: 0.01L/min without prior notice. All rights reserved by
BT, IT: 0.1°C COMEN
Accuracy: C.O.: ±5% or ±0.1 L/min,whichever
is greater