User Manual Osiloskop

This document describes a 100MHz dual-channel oscilloscope, the GOS-6103/6103C. It has features like a 100MHz bandwidth, cursor readout for voltage, time and frequency measurements, 10 sets of memory for saving front panel settings, and auto-ranging of the time base. The vertical system has 2 input channels with 11 sensitivity settings from 2mV to 5V per division and a bandwidth of DC to 100MHz. The horizontal system provides sweep times from 0.5s to 50ns per division.

Uploaded by

neddylalola
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
437 views

User Manual Osiloskop

This document describes a 100MHz dual-channel oscilloscope, the GOS-6103/6103C. It has features like a 100MHz bandwidth, cursor readout for voltage, time and frequency measurements, 10 sets of memory for saving front panel settings, and auto-ranging of the time base. The vertical system has 2 input channels with 11 sensitivity settings from 2mV to 5V per division and a bandwidth of DC to 100MHz. The horizontal system provides sweep times from 0.5s to 50ns per division.

Uploaded by

neddylalola
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

100MHz CURSOR READOUT ANALOG OSCILLOSCOPE

OSCILLOSCOPES

OSCILLOSCOPES
The GOS-6100 Series is a 100MHz, two-channel, dual-sweep, portable oscilloscope for general purpose
use. A microprocessor-based operating system controls most of the functions of the instrument, in
cluding cursor readout and digitized panel setting.

On-screen alphanumeric readout and cursor function for voltage, time, frequency and phase
measurement provide extraordinary operational convenience. The advance Time Base Auto-rang fun
ction conveniently acquires waveforms at the push of button (GOS-6103/ 6103C). Ten different user
defined instrument settings (GOS-6103/ 6103C) can be saved and recalled without restriction.

The vertical deflection system has two input channels. Each channel has 11 basic deflection factors
from 2mV to 5V per division. The horizontal deflection system provides single, dual or delayed sweeps
GOS-6112/6103/6103C (100 MHz) from 0.5s to 50ns per division (delayed sweep, 50ms to 50ns per division). The trigger system provides
stable triggering over the full bandwidth of the vertical deflection system.
GOS-6112 GOS-6103/6103C
GOS-6103C Without Approval

FEATURES SPECIFICATIONS SPECIFICATIONS


CRT CURSOR READOUT FUNCTION
* 100MHz Bandwidth, Dual Channel, Delayed Type 6-inch rectangular type with internal graticule; 0%, 10%, 90% and 100% markers Cursor Measurement Function V, V%, VdB, T, 1/ T, T%,
Sweep 8 x 10 div (1 div = 1 cm) Cursor Resolution 1/100 div
Accelerating Potential 16 kV approx. (GOS-6103/GOS-6103C), 12kV approx. (GOS-6112) Effective Cursor Range Vertical: + _ 4 div
_ 3div ; Horizontal: +
* Built-In 6 digits Universal Counter (GOS-6103C) Illumination Continuously adjustable (GOS-6103/GOS-6103C) Panel Setting Display Vertical: V/div(CH1,CH2),UNCAL,ALT/CHOP/ADD,INV, probe factor,AC/DC/GND
* 10 Sets Memory for Front Panel Setting Save Z-axis input Coupling : DC Horizontal: s/div(MTB, DTB), UNCAL, x 10MAG, delay time , Hold-off
& Recall (GOS-6103/GOS-6103C) Sensitivity: 5V or more Trigger: source, coupling, slope, level, TV-V, TV-H
Maximum input voltage : 30V (DC + AC peak) at 1kHz or less Others: X-Y, lock, save/recall MEM 0-9 (GOS-6103/GOS-6103C)
* Time Base Auto-range (GOS-6103/GOS-6103C) Bandwidth : DC ~ 5 MHz Voltage Measurement
AUTO MEASUREMENT FUNCTION (GOS-6103C)
* Cursor Readout with 7 Measurements VERTICAL SYSTEM _ WIDTH, +
Parameter Function FREQ, PERIOD, + _ DUTY (+ or - polarity selected by trigger slope)
* Panel Setup Lock of Digital-Control Functions Sensitivity 2mV~5V/div, 11 step in 1-2-5 sequence Display Digits Max. 6-digits, decimal
* Buzzer Alarm Sensitivity Accuracy _ 3% (5div at the center of display)
< Frequency Range 50Hz ~ 100MHz
Vernier Vertical Sensitivity Continuously variable to 1/2.5 or less of panel-indicate value Accuracy 1kHz ~ 100MHz : + _ 0.05%
_ 0.01% ; 50Hz ~ 1kHz : +
* LED Indicators Bandwidth(-3dB) DC~100MHz(2mV/div:DC~20MHz) Measuring Sensitivity > 2 div (Measuring source selected from CH1 and CH2 as synchronous signal sources)
* TV Synchronization Rise Time 3.5ns (2mV/div:17.5ns) SPECIAL FUNCTION
Signal Delay Leading edge can be monitored
* Trigger Signal Output TIME/DIV Auto Range Provided (GOS-6103/GOS-6103C)
Max. Input Voltage 400V(DC+AC peak) at 1kHz or less
Panel Setting Save & Recall 10 sets (GOS-6103/GOS-6103C)
* Z-Axis Modulation Input Input Coupling AC, DC, GND
_2% // approx. 25pF Panel Setups Lock Provided Voltage Percentage Measurement
Input Impedance 1M +
* SMD Technology, High Stability and Reliability Vertical Mode CH1,CH2,DUAL(CHOP/ALT), ADD, CH2 INV. POWER SOURCE
Bandwidth Limited 20MHz _10% , 50/60Hz
AC 100V/120V/230V +
Common-Mode Rejection Ratio 50:1 or better at 50kHz
DIMENSIONS & WEIGHT
CURSOR MEASUREMENT Dynamic Range 8 div at 60MHz; 5div at 100MHz (GOS-6112)
8 div at 100MHz (GOS-6103/GOS-6103C) 310(W) x 150(H) x 455(D) mm ; Approx. 9kg
FUNCTIONS HORIZONTAL SYSTEM
Horizontal Modes MAIN(A), ALT, DELAY(B) ORDERING INFORMATION
A(main) Sweep Time 50ns~0.5s/div, continuously variable (UNCAL) GOS-6112 100MHz, 2-channel, Analog Oscilloscope
B(delay) Sweep Time 50ns~50ms/div GOS-6103 100MHz, 2-channel, Analog Oscilloscope
Accuracy +
_ 3% ( +_ 5% at x 10 MAG) GOS-6103C 100MHz, 2-channel, Analog Oscilloscope with 100MHz Frequency Counter Frequency Measurement
Sweep Magnification x 10 (maximum sweep time 5nS/div)
Hold Off Time ACCESSORIES :
Variable
Delay Time 1 s~5s User manual x 1, Power cord x 1
The unique, easy-to-use cursor and numerical GTP-100A-2 : 100MHz ( 10 : 1/1 :1 ) Switchable Passive Probe (one per channel)
Delay Jitter Better than 1:20000
readouts make waveform observation and Alternate Separation Variable OPTIONAL ACCESSORIES
measurement easier, faster and more accurate. TRIGGER GTC-001 Instrument Cart, 450(W) x 430(D) mm (120V Input Socket)
The on-screen cursors provide seven Trigger Modes AUTO, NORM,TV GTC-002 Instrument Cart, 330(W) x 430(D) mm (120V Input Socket)
measurement functions ( V, V%, VdB, T, Trigger Source CH1,CH2,LINE,EXT GTL-110 Test Lead, BNC-BNC Heads
Trigger Coupling AC,DC,HFR,LFR
1/ T,T%, ) Trigger Slope "+" or "-" polarity or TVsync polarity Time Percentage Measurement
Trigger Sensitivity Mode Frequency INT EXT
10 Hz ~ 20 MHz 0.35 div 50 mV
AUTO
20 MHz ~ 100 MHz 1.5 div 150 mV
DC ~ 20 MHz 0.35 div 50 mV
NORM
20 MHz ~ 100 MHz 1.5 div 150 mV
Rear Panel sync signal 1 div 200 mVpp
TV

TV sync TV-V, TV-H


Max. External Input Voltage 400V(DC+AC peak ) at 1kHz Time Measurement
External Input Impedance 1M W +_5% // approx.25pF
X-Y OPERATION
Mode X-axis: selectable CH1, CH2, EXT ; Y-axis: selectable CH1, CH2, CH1 and CH2
Sensitivity Accuracy 2mV~5V/div+ _ 3% ; EXT : 0.1V/div +
_ 5%
X-axis Bandwidth DC~500kHz(-3dB)
Phase Error 3 or less from DC~50kHz
OUTPUT SIGNAL
GOS-6112 Trigger Signal Output Voltage: approx. 25mV/div into 50 ; Frequency response : DC ~ 10MHz
Calibrator Output 1kHz Square wave, 2Vpp +_ 2%
Phase Measurement

A29 Made to Measure. Good Will Instrument Co., Ltd. A30


100MHz CURSOR READOUT ANALOG OSCILLOSCOPE
OSCILLOSCOPES

OSCILLOSCOPES
The GOS-6100 Series is a 100MHz, two-channel, dual-sweep, portable oscilloscope for general purpose
use. A microprocessor-based operating system controls most of the functions of the instrument, in
cluding cursor readout and digitized panel setting.

On-screen alphanumeric readout and cursor function for voltage, time, frequency and phase
measurement provide extraordinary operational convenience. The advance Time Base Auto-rang fun
ction conveniently acquires waveforms at the push of button (GOS-6103/ 6103C). Ten different user
defined instrument settings (GOS-6103/ 6103C) can be saved and recalled without restriction.

The vertical deflection system has two input channels. Each channel has 11 basic deflection factors
from 2mV to 5V per division. The horizontal deflection system provides single, dual or delayed sweeps
GOS-6112/6103/6103C (100 MHz) from 0.5s to 50ns per division (delayed sweep, 50ms to 50ns per division). The trigger system provides
stable triggering over the full bandwidth of the vertical deflection system.
GOS-6112 GOS-6103/6103C
GOS-6103C Without Approval

FEATURES SPECIFICATIONS SPECIFICATIONS


CRT CURSOR READOUT FUNCTION
* 100MHz Bandwidth, Dual Channel, Delayed Type 6-inch rectangular type with internal graticule; 0%, 10%, 90% and 100% markers Cursor Measurement Function V, V%, VdB, T, 1/ T, T%,
Sweep 8 x 10 div (1 div = 1 cm) Cursor Resolution 1/100 div
Accelerating Potential 16 kV approx. (GOS-6103/GOS-6103C), 12kV approx. (GOS-6112) Effective Cursor Range Vertical: + _ 4 div
_ 3div ; Horizontal: +
* Built-In 6 digits Universal Counter (GOS-6103C) Illumination Continuously adjustable (GOS-6103/GOS-6103C) Panel Setting Display Vertical: V/div(CH1,CH2),UNCAL,ALT/CHOP/ADD,INV, probe factor,AC/DC/GND
* 10 Sets Memory for Front Panel Setting Save Z-axis input Coupling : DC Horizontal: s/div(MTB, DTB), UNCAL, x 10MAG, delay time , Hold-off
& Recall (GOS-6103/GOS-6103C) Sensitivity: 5V or more Trigger: source, coupling, slope, level, TV-V, TV-H
Maximum input voltage : 30V (DC + AC peak) at 1kHz or less Others: X-Y, lock, save/recall MEM 0-9 (GOS-6103/GOS-6103C)
* Time Base Auto-range (GOS-6103/GOS-6103C) Bandwidth : DC ~ 5 MHz Voltage Measurement
AUTO MEASUREMENT FUNCTION (GOS-6103C)
* Cursor Readout with 7 Measurements VERTICAL SYSTEM _ WIDTH, +
Parameter Function FREQ, PERIOD, + _ DUTY (+ or - polarity selected by trigger slope)
* Panel Setup Lock of Digital-Control Functions Sensitivity 2mV~5V/div, 11 step in 1-2-5 sequence Display Digits Max. 6-digits, decimal
* Buzzer Alarm Sensitivity Accuracy _ 3% (5div at the center of display)
< Frequency Range 50Hz ~ 100MHz
Vernier Vertical Sensitivity Continuously variable to 1/2.5 or less of panel-indicate value Accuracy 1kHz ~ 100MHz : + _ 0.05%
_ 0.01% ; 50Hz ~ 1kHz : +
* LED Indicators Bandwidth(-3dB) DC~100MHz(2mV/div:DC~20MHz) Measuring Sensitivity > 2 div (Measuring source selected from CH1 and CH2 as synchronous signal sources)
* TV Synchronization Rise Time 3.5ns (2mV/div:17.5ns) SPECIAL FUNCTION
Signal Delay Leading edge can be monitored
* Trigger Signal Output TIME/DIV Auto Range Provided (GOS-6103/GOS-6103C)
Max. Input Voltage 400V(DC+AC peak) at 1kHz or less
Panel Setting Save & Recall 10 sets (GOS-6103/GOS-6103C)
* Z-Axis Modulation Input Input Coupling AC, DC, GND
_2% // approx. 25pF Panel Setups Lock Provided Voltage Percentage Measurement
Input Impedance 1M +
* SMD Technology, High Stability and Reliability Vertical Mode CH1,CH2,DUAL(CHOP/ALT), ADD, CH2 INV. POWER SOURCE
Bandwidth Limited 20MHz _10% , 50/60Hz
AC 100V/120V/230V +
Common-Mode Rejection Ratio 50:1 or better at 50kHz
DIMENSIONS & WEIGHT
CURSOR MEASUREMENT Dynamic Range 8 div at 60MHz; 5div at 100MHz (GOS-6112)
8 div at 100MHz (GOS-6103/GOS-6103C) 310(W) x 150(H) x 455(D) mm ; Approx. 9kg
FUNCTIONS HORIZONTAL SYSTEM
Horizontal Modes MAIN(A), ALT, DELAY(B) ORDERING INFORMATION
A(main) Sweep Time 50ns~0.5s/div, continuously variable (UNCAL) GOS-6112 100MHz, 2-channel, Analog Oscilloscope
B(delay) Sweep Time 50ns~50ms/div GOS-6103 100MHz, 2-channel, Analog Oscilloscope
Accuracy +
_ 3% ( +_ 5% at x 10 MAG) GOS-6103C 100MHz, 2-channel, Analog Oscilloscope with 100MHz Frequency Counter Frequency Measurement
Sweep Magnification x 10 (maximum sweep time 5nS/div)
Hold Off Time ACCESSORIES :
Variable
Delay Time 1 s~5s User manual x 1, Power cord x 1
The unique, easy-to-use cursor and numerical GTP-100A-2 : 100MHz ( 10 : 1/1 :1 ) Switchable Passive Probe (one per channel)
Delay Jitter Better than 1:20000
readouts make waveform observation and Alternate Separation Variable OPTIONAL ACCESSORIES
measurement easier, faster and more accurate. TRIGGER GTC-001 Instrument Cart, 450(W) x 430(D) mm (120V Input Socket)
The on-screen cursors provide seven Trigger Modes AUTO, NORM,TV GTC-002 Instrument Cart, 330(W) x 430(D) mm (120V Input Socket)
measurement functions ( V, V%, VdB, T, Trigger Source CH1,CH2,LINE,EXT GTL-110 Test Lead, BNC-BNC Heads
Trigger Coupling AC,DC,HFR,LFR
1/ T,T%, ) Trigger Slope "+" or "-" polarity or TVsync polarity Time Percentage Measurement
Trigger Sensitivity Mode Frequency INT EXT
10 Hz ~ 20 MHz 0.35 div 50 mV
AUTO
20 MHz ~ 100 MHz 1.5 div 150 mV
DC ~ 20 MHz 0.35 div 50 mV
NORM
20 MHz ~ 100 MHz 1.5 div 150 mV
Rear Panel sync signal 1 div 200 mVpp
TV

TV sync TV-V, TV-H


Max. External Input Voltage 400V(DC+AC peak ) at 1kHz Time Measurement
External Input Impedance 1M W +_5% // approx.25pF
X-Y OPERATION
Mode X-axis: selectable CH1, CH2, EXT ; Y-axis: selectable CH1, CH2, CH1 and CH2
Sensitivity Accuracy 2mV~5V/div+ _ 3% ; EXT : 0.1V/div +
_ 5%
X-axis Bandwidth DC~500kHz(-3dB)
Phase Error 3 or less from DC~50kHz
OUTPUT SIGNAL
GOS-6112 Trigger Signal Output Voltage: approx. 25mV/div into 50 ; Frequency response : DC ~ 10MHz
Calibrator Output 1kHz Square wave, 2Vpp +_ 2%
Phase Measurement

A29 Made to Measure. Good Will Instrument Co., Ltd. A30