0% found this document useful (0 votes)
129 views

MV STD Ox CS

Uploaded by

vipmama
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
129 views

MV STD Ox CS

Uploaded by

vipmama
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

DAILY CONTROL TESTS

Motor Vessel (<32 Bar LPSG) Control and Dosage Chart Standard Boiler Water Treatment with DREWPLEX® OX Corrosion Inhibitor
Read MSDS before running these tests. Before testing, samples must be cooled to 25 OC (77 OF) by collect- colored or turbid, filter before testing. If the sample remains cloudy after the first filtering, the sample should
ing through a sample cooler for safety and to prevent flashing which concentrates the sample. If the sample is be refiltered through the same filter paper since the filter becomes more retentive on the second filtration.
DEHA DAILY TEST PROCEDURES
1. Fill the sample cup to the 25-ml mark with feedwater sample. (Fig. 1).
2. Add 2 drops of Activator Solution. Stir gently with tip of an ampoule from the DREWPLEX OX 1 2 3 4 5 6
corrosion inhibitor Ampoule Test Kit to mix the contents of the sample cup (Fig. 2).
ml

4
25
20
3. Immediately snap the tip of the ampoule by pressing the ampoule against the side of the
sample cup. The sample will fill the ampoule and begin to mix with the reagent, leaving a
15
10
ml
25
20
10 : 0 0
5
small bubble to facilitate the mixing (Fig. 3).
15
10
5

4. Remove the ampoule from the cup. Mix the contents of the ampoule by inverting it several
times, allowing the small bubble to travel from end to end each time (Fig. 4). MEANS OF ADJUSTMENT DOSAGE INFORMATION
5. Wipe all liquid from the exterior of the ampoule and WAIT EXACTLY 10 MINUTES for full • DREWPLEX® OX corrosion inhibitor LESS THAN 0.4 PPM – INCREASE BY 25%
color development (Fig. 5). • Continuous Dosage 0.4 - 0.8 ppm – SATISFACTORY - NO CHANGE
6. Place the DREWPLEX OX corrosion inhibitor ampoule, flat end downward into center tube GREATER THAN 0.8 PPM – DECREASE BY 25%
of comparator. Direct the comparator toward a source of bright white light while viewing from
Initial dosage* is 100-300 ml/day. Increase dosage by 50-100ml/day until the DEHA control range is reached.
bottom. Hold comparator in a nearly horizontal position and rotate it until the color standard
below the DREWPLEX OX corrosion inhibitor ampoule shows the closest match (Fig. 6). Feed continuously with DREWPLEX AT boiler water treatment. (*When starting up a new system or one that
has been down for some reason).
7. Record the result and adjust dosage as necessary.
PHOSPHATE – Filter sample before running test DAILY TEST PROCEDURES
1. Fill the sample cup to the 25-ml mark with sample (Fig. 1).
1 2 3 4 5
2. Place the Boiler Phosphate ampoule’s tapered tip into one of the four depressions in the ml

bottom of the sample cup. Snap the tip by squeezing the ampoule toward the side of the cup.
4
25
20

The sample will fill the ampoule and begin to mix with reagent (Fig. 2). 15
10 5:0 0
3. Remove the Boiler Phosphate ampoule from the cup. Mix the contents of the ampoule by 5

inverting it several times allowing the bubble to travel from end to end each time (Fig. 3).
4. Wipe all liquid from the exterior of the ampoule and WAIT 5 MINUTES for full color MEANS OF ADJUSTMENT DOSAGE INFORMATION
development (Fig. 4).
• ADJUNCT® B PHOSPHATE BOILER WATER 0 - 10 ppm 30 grams per ton of boiler water in the system
5. When using the comparator, be sure it is illuminated by a white light directly above it. The
filled Boiler Phosphate ampoule should be placed between the color standards for viewing. TREATMENT 10 - 20 ppm 15 grams per ton of boiler water in the system
It is important that ampoule be compared by placing it on both sides of the standard tube 20 - 40 ppm SATISFACTORY - NO CHANGE
before concluding that it is darker, lighter or equal to the standard (Fig. 5). Record the result. 40+ ppm HIGH, BLOWDOWN REDUCE DOSAGE
INITIAL DOSAGE: 30 grams per ton of boiler water in the system

P&T ALKALINITY DAILY TEST PROCEDURES


PHENOLPHTHALEIN (“P”) ALKALINITY
1&2 3 5&6 7 8&9
1. Collect 50-ml cooled boiler water sample (Fig. 1 & 2).
2. Pour into evaporating dish (Fig. 1 & 2).
3. Add 4 drops phenolphthalein (Fig. 3).
4. Pink color - Alkaline. No Color - “P” alkalinity zero.
5. Add sulfuric acid N/10 drop by drop until pink color disappears (Fig. 5 & 6).
6. Note level of acid in buret. Convert ml to ppm using chart at right and record result in ppm as MEANS OF ADJUSTMENT DOSAGE INFORMATION
“P” alkalinity. Keep sample for “T” alkalinity test. Record result & adjust dosage as necessary. • GCTM concentrated alkaline liquid ML SULFURIC ACID N/10 PPM DOSAGE ADJUSTMENT
0 - 0.3 0 - 30 0.15 liter per ton of boiler water in the system
TOTAL (“T”) ALKALINITY 0.4 - 0.7 40 - 70 0.10 liter per ton of boiler water in the system
7. Add 3 drops Total Alkalinity Indicator-GP (Fig. 7). 0.8 - 0.9 80 - 90 0.05 liter per ton of boiler water in the system
8. Add sulfuric acid N/10 until pink color develops (Fig. 8 & 9). 1.0 - 1.5 100 - 150 SATISFACTORY
1.5+ 150+ BLOWDOWN — REDUCE DOSAGE
9. Note the level of acid in buret. Convert ml to ppm using chart at right and record result in ppm
as “Total Alkalinity”. Record the result & adjust dosage as necessary. TOTAL ALKALINITY TEST RESULT MUST BE IF NOT, DOSE 1 LITER OF GC REGARDLESS
LESS THAN TWICE THE “P” ALKALINITY. OF THE “P” ALKALINITY. TEST AGAIN IN 2 HRS.
©2012 Drew Marine • ®Registered trademark, TMTrademark, Drew Marine • 1Registered trademark of CHEMetrics Inc. – 1 – MV_STD_OX_CS
CHLORIDE TEST DAILY TEST PROCEDURES
1. Measure 2-ml cooled sample into glass tube (Fig. 1). SATISFACTORY NO BLOWDOWN BLOWDOWN
2. Add 3 drops phenolphthalein. 1 3 N/10
Silver Nitrate Drops N/10 1 2 3 4 5 6 Over 6
Sulfuric
Acid
3. Add N/10 Sulfuric Acid dropwise until pink color disappears. Add 1 more drop (Fig. 3).
4. Add 6 drops Potassium Chromate. Sample will be yellow. PPM Chloride 50 100 150 200 250 300 Over 300
5. Add N/10 Silver Nitrate dropwise until the sample turns orange.
6. Drops N/10 Silver Nitrate X 50 = ppm chloride.
7. Record the result and refer to chart at right for blowdown adjustment. MEANS OF ADJUSTMENT
• BLOWDOWN

NEUTRALIZED CONDUCTIVITY DAILY TEST PROCEDURES


1. Connect the Conductivity Electrode-ATC to the meter. Press the power button to power up 2 3
the meter.
2. Add 2 drops of phenolphthalein indicator to the cooled boiler water sample and stir. If
the sample turns pink, add Gallic Acid while stirring until the pink color disappears. This
neutralizes the sample (Fig. 2).
3. Place the electrode into the sample and take care to ensure that the liquid level is above
its upper steel band. Stir the electrode gently in the sample to create a homogenous MEANS OF ADJUSTMENT DOSAGE INFORMATION
sample (Fig. 3). • BLOWDOWN Up to 700 µS/cm - Satisfactory Above 700 µS/cm
4. Allow time for the reading to stabilize. Note the reading on the display. NORMAL MAINTENANCE BLOWDOWN INCREASE BLOWDOWN FREQUENCY
5. Record the result and refer to chart at right for blowdown adjustment.
Blowdown should be carried out as per boiler manufacturer’s instructions. Blowdown is necessary to
reduce dissolved and suspended solids. If suspended solids are noticed in boiler water samples, blowdown
regardless of conductivity test results. Weekly flash bottom blowdown for 10-20 seconds duration is advisable
to remove normal sludge and to ensure clear operational lines.

CONDENSATE pH TEST - Test immediately after drawing sample. DAILY TEST PROCEDURES
1. Collect 50-ml cooled condensate sample and pour into evaporating dish (Fig. 1).
1 2 3
2. Add 3 drops phenolphthalein. Sample should turn pink (Fig. 2).
3. Add sulfuric acid N/10 drop by drop until pink color disappears (Fig. 3).
4. Record the result and adjust dosage as necessary.

MEANS OF ADJUSTMENT DOSAGE INFORMATION


• SLCC-ATM condensate corrosion NO PINK WITH PHENOLPHTHALEIN – INCREASE BY 25%
inhibitor PINK COLOR – 1-2 DROPS N/10 ACID TO CLARIFY– SATISFACTORY NO CHANGE
• Continuous Dosage PINK COLOR – 3 OR MORE – DROPS N/10 ACID TO CLARIFY– DECREASE BY 25%
INITIAL DOSAGE: 0.15 liter per ton of boiler water in the system
LIQUID COAGULANTTM Boiler Sludge Conditioner - No Test Required (For Oil Contamination Only)
Daily dosage 28-ml (1.0 oz.) per day for each ton of water capacity to condition suspended solids (oil and sludge) for removal by boiler sludge conditioner. Dose for two weeks or until oil removal is complete. If severe contamination persists, the possibility of
bottom blowdown. Flash bottom blow and scum blow several times at 20 minute intervals after dosing the LIQUID COAGULANT continuing oil leakage should be investigated and eliminated.
APPARATUS
Storage: Water Treatment Test Cabinet: PCN 1AA0016 (220 Volt) or 1AA2148 (110 Volt) Gallic Acid, Brass Spoon, Plastic Beaker, 2 x 4 AAA Batteries and Manual
DEHA: DREWPLEX OX AmpouleTest Kit: PCN 0387010 includes: Sample Cup, Comparator, Phosphate: Boiler Phosphate Ampoule Test Kit: PCN 1AA0003 includes: Sample cup, Comparator, 30 Ampoules
30 Ampoules, 1 x 10 ml Activator Solution. The Activator Solution is also supplied separately under Boiler Phosphate Ampoule Refill (PCN 1AA0004)
PCN 0387036 (6 x 7 ml)
Filter Paper: PCN 0225-01-2, Box of 100
Alkalinity/Chloride: Alkalinity/Chloride Reagent Set PCN 0301-02-8 : 1 bottle Phenolphthalein, 1 bottle Total Alkalinity Indicator GP,
Funnel: PCN 0221-01-0
2 bottles Sulfuric Acid N/10, 2 Chloride LMP Test Kits Alkalinity/Glassware Set PCN 0201-02-0 includes: LP
Alkalinity Titration Assembly, Evaporating Dish (140ml), Graduated Cylinder (50ml, plastic), 30ml Dropper Bottle
for Total Alkalinity Indicator GP, 30ml Dropper Bottle for Phenolphthalein, 3 Plastic Stirring Rods (150mm)
Conductivity: Conductivity Meter Kit Complete: PCN 0173062 includes Meter and Electrode, Case, Stand, Rubber Armor,
©2012 Drew Marine • ®Registered trademark, TMTrademark, Drew Marine • 1Registered trademark of CHEMetrics Inc. – 2 – MV_STD_OX_CS

You might also like