Prospeed Error 117 XG Fail
Prospeed Error 117 XG Fail
Modality: CT YMS
Model Type: ,
Problem Description:
When trying to scan, system kicks out an MA accuracy error, XG 117 Code 531. FE started the call as a DOA
report and found out the the board he was filling the report out on was good. Filamnet cable, filament board,
and CPU were replaced, but system still had the same problem intermittenlty, Tube has over 100k slices.
Solution Description:
Tube was the solution too. Just to mention, when scanning using small filament, off-line scans, no errors.
When switching back to large filament, errors right away.
Modality: CT YMS
Model Type: ,
Problem Description:
FE was trouble shooting system for intermittent XG failures ER-117(Code=541) anode overcurrent.
Bleeder was put in place of tube to run KV check seeing if any spits were occuring from
HV tank.
Tyraps had to be clipped on HV cables comming from HV tanks so they could reach the bleeder.
Upon hooking back up tube ER-117(Code=531) mA accuarcy error began.
Problem was that anode and cathode HV cables had be swapped.
Solution Description:
Uncrossed HV cables - Anode cable went to anode side of tube, cathode cable went to cathode tube well.
Tube fans are not running due to fuse blown under CTVRC cover on Prospeed CT
CT Resolution: Replace blown fuses on small board inside the CTVRC cover of the Prospeed Gantry
See Long Description for Details.
Replace blown fuses on small board inside the CTVRC cover of the Prospeed Gantry. This small board contains AC power and fuses
for the Tube Fans. Use spares from Prospeed Fuse Kit 2122788
CT Software Version: 6.04 for for HiSpeed series and CT/e, CT ProSpeed A1/F1 series, DXi,LXi,FXI, ZXi (18811)
6.04 for for HiSpeed series and CT/e, CT ProSpeed A1/F1 series, DXi,LXi,FXI, ZXi (18811)
6.04 for for HiSpeed series and CT/e, CT ProSpeed A1/F1 series, DXi,LXi,FXI, ZXi
(18811)
Also came as FMI26521 for Lemage, Lemage SX, Lemage Supreme, Lemage SX/E, ProSpeed S, ProSpeed SX, ProSpeed, ProSpeed S
Fast, ProSpeed SX Power, ProSpeed SX Advantage, ProSpeed with SSD option, ProSpeed SX Power with SSD option
Scanner Hardware Stopped Scan WITH Error: xg117 00000072 Modification Parameter is not
allowed:ProSpeed Qj and Zj (Lemage, Qj, Qj-Base, Qj-Hlt):Replacing Filament Board eliminates XG117
Code72 Errors
CT Issue: Scanner Hardware Stopped Scan WITH Error: xg117 00000072 Modification Parameter is not allowed
Scanner Hardware Stopped Scan WITH Error: xg117 00000072 Modification Parameter is not allowed
Scanner Hardware Stopped Scan WITH Error: xg117 00000072 Modification Parameter is not allowed
Follow procedures in YMS service documentation 2124012 Component Replacement. See directions for replacement of Filament
Supply
CT Component: 2119708-2 Filament Control Board for YMS Prospeed Systems old part # - 2119708
See LD for additional details
FE should check/replace the C17 on filament board, 470uf 63volts before replacement of this board - also check C40!
This is the primary failure part on this board. C17 has been replaced
with 470uf 80 volts purchased locally and this fixed the filament errors. Also check C40.
Part #: 2119708-2
Part Type: GP CONSUMABLES
Part Weight: 0.60 LBR
Repairable: No
Returnable: No
Hazardous Material: N
Description: FILAMENT HEATER BOARD
FRU: Yes
Status: Orderable
Parts Offered as:
Current: 2119708-2
Harvested: Not Available
Comments:
FLS - FILAMENT CONTROL BRD U/O CT QJ & CT MINI REPLACES 2119708 PER PCN P 10/99 02 ON CONSUMABLE
COLLECTION PER G TOURNE 1/16/04 MS5277 DO THE CHANGE UP ONCE OLD MATERIAL PULLED BACK DO NOT
DISCOUNT - PER SHAKUN CHECK CAPACITORS: C17 AND C70
GELS - 17-09-2003 20PCES WILL SEND TO BUC FOR TEST AND 30 PCES IN FW 39 FLORIANE 16/09/03,PCB CONT FIL
CT Software Version: 6.04 for for HiSpeed series and CT/e, CT ProSpeed A1/F1 series, DXi,LXi,FXI, ZXi (18811)
6.04 for for HiSpeed series and CT/e, CT ProSpeed A1/F1 series, DXi,LXi,FXI, ZXi (18811)
6.04 for for HiSpeed series and CT/e, CT ProSpeed A1/F1 series, DXi,LXi,FXI, ZXi
(18811)
Also came as FMI26521 for Lemage, Lemage SX, Lemage Supreme, Lemage SX/E, ProSpeed S, ProSpeed SX, ProSpeed, ProSpeed S
Fast, ProSpeed SX Power, ProSpeed SX Advantage, ProSpeed with SSD option, ProSpeed SX Power with SSD option
Modality: CT YMS
Problem Description:
System: Prospeed Qj and Zj CT
Subsystem: X rays Tube (D3112T)
Software: V/R 4.X, 5.x, 6.X
Date: 10/2003
Solution Description:
The stator impedance should measure as follows:
Cables Black to White (common) 23 ohms.
Cables Green to White (common) 23 ohms
and cables Black to green 46 ohms.
All wires to ground should be open. This is a good place to check for error code XG Error 117
codes:0000072, 0000216.
Low current errors
ID: 16895 : ProSpeed Qj: Scanner Hardware Stopped Scan Message During Scan WI
xray tube
CT Issue: Scanner Hardware Stopped Scan WITH Error: xg117 00000072 Modification Parameter is not allowed
Scanner Hardware Stopped Scan WITH Error: xg117 00000072 Modification Parameter is not allowed
Scanner Hardware Stopped Scan WITH Error: xg117 00000072 Modification Parameter is not allowed
CT Cause: A faulty XRAY TUBE CAUSED Error: xg117 00000072 Modification Parameter is not allowed
See LD for Troubleshooting
A faulty XRAY TUBE CAUSED Error: xg117 00000072 Modification Parameter is not allowed
Troubleshooting
a.“Most common” Check C17 on Filament Board. If board smells like urine replace Filament board.
b. Check that Filament Cable is secure(E-Rack XJ06---Cathode HV tank J2).
c. Check that the Rotor Board is seated
d. Check the L.V. Power Supply on E-Rack
e. Perform KV test.
f. Check Filament resistance.
g. Check HV Cables for an open or short.
h. Replace E-Rack Mother Board
See page 337 of the component replacement manual for tube replacement procedure
http://olcweb.olc.med.ge.com:8020/servlet/ClientServletProp/2326233_6.pdf?
DOCCLASS=A&REQ=RAC&DIRECTION=2326233&FILENAME=2326233_6.pdf&FILEREV=6&DOCREV_ORG=6&SUBMIT=+ACCEP
NOTE: This may also be a good time to replace the collimator primary filter - see component if you need the part number.
Part #: P9184MV
Part Type: GP CONSUMABLES
Part Weight: 3.00 LBR
Repairable: No
Returnable: No
Hazardous Material: N
Description: FILTER
FRU: Yes
Status: Orderable
Comments:
Not Available
Part #: D3112T
Part Type: GP TUBES
Part Weight: 0.00 LBR
Repairable: No
Returnable: Yes
Hazardous Material: N
Description: X-RAY TUBE
FRU: Yes
Status: Orderable
Comments:
FLS - RELOADED CT ZJ/QJ TUBE UNIT FOR PROSPEED S & SX, SYTEC- 6000/PROSPEED-VX, SYTEC-8000/ PROSPEED .........
ORDERING TUBE............... MAXIRAY EQUIVALENT IS D3119T AM SOURCE - GEMS-MEXICO BLACK MOUNT. PLATE= YMSP91
CT Software Version: 6.02 for HiSpeed series and CT/e, CT ProSpeed AI/FI series, DXi,FXi, LXi, ZXi
6.02 for HiSpeed series and CT/e, CT ProSpeed AI/FI series, DXi,FXi, LXi, ZXi This software Rev is Obsoleted by FMI No.25870C V/R6.03 P
6.02 for HiSpeed series and CT/e, CT ProSpeed AI/FI series, DXi,FXi, LXi, ZXi
http://3.36.10.5/../ctdoc/FMI/12/25869_22706.pdf
This software Rev is Obsoleted by FMI No.25870C CT HiSpeed, CT/e, CT ProSpeed AI/FI Series
RELEASE
Xi FMI 25869_22706 6.02 Sftwr.pdf
COMMENTS: Some KV Bds received "may need" a slight TWEEK to I/F with
your system.
Modality: CT YMS
Problem Description:
System is having ER-117 XG error code 72 (Modification of Parameter is not allowed)
Normally this error refers to the filament.
1.- Look at the LEDs on the FILAMENT-bd, in stand by we should have 5 green LEDs and 1 yellow LED
In this case, FIL_SUPPLY_OK and FIL_OVER were ON (both LEDs are red)
2.- With the System off, measure the Filament Impedance in the x ray tube as follows . . .
Large-Common and Small-Common --> both readings should be less than one ohm.
If one of them is more than 1 ohm, we have an open filament.
(Note: In Sytec SRi and Synergy systems We use one filament only, which is the large one,
but the short filament is shorted out, the short circuit between small-common should be
present always.
3.- To make sure that we only have an open filament, with the system off, take out the HV Cathode
cable and short out the 3 pins between themselves with a piece of solder.
Bring the power back on, DO NOT MAKE ANY EXPOSURE, look at the LEDs at the FILAMENT-BD
and the 2 RED LEDs, FIL_OVER and FIL_SUPPLY_OK should be OFF, the other 5 GREEN LEDs and the
YELLOW one should be ON only
Solution Description:
System: Prospeed Qj
Software: V/R 6.03
Problem:
-------
FE having ER-117 XG Errors ( Code=72 Modification of Parameters is not Allowed )