Oprators Manual CMS9000V3.0 Software Monitor
Oprators Manual CMS9000V3.0 Software Monitor
Catalogue
1 Preface................................................................................................................................................. 1
1 .1 Caution.....................................................................................................................................1
1 .2 Manipulator Obligation.......................................................................................................... 1
2 General Introduction to Software..................................................................................................... 2
2 .1 General Introduction to Aim..................................................................................................2
2 .2 Function................................................................................................................................... 2
3 Main Technical Index..........................................................................................................................3
3 .1 Dimension and Weight.......................................................................................................... 3
3 .2 Working Conditions................................................................................................................3
3 .3 Performance Index.................................................................................................................3
3 .4 Network Performance Index.................................................................................................4
3 .5 Printer Performance Index....................................................................................................4
4 System Configuration Demand........................................................................................................ 5
5 Function Instructions.......................................................................................................................... 6
5 .1 Main Monitoring Interface..................................................................................................... 6
5 .2 Key Observation – Patient Information Interface............................................................11
5 .3 Single Bedside Monitor Instruction................................................................................... 12
5 .4 Operation Introduction to Waveform Review.................................................................. 17
5 .5 Trend Review Instruction.................................................................................................... 21
5 .6 Alarm Review Instruction.................................................................................................... 26
5 .7 CO Review............................................................................................................................ 29
5 .8 NIBP Review.........................................................................................................................31
5 .9 Display setting instructions.................................................................................................32
5 .10 History Review - History Patient List and Inquiry.....................................................34
5 .11 History Review- Medical History Amendatory Interface...........................................36
5 .12 History Review - Other Reviews.................................................................................37
5 .13 Instructions to System Setting......................................................................................... 37
5 .14 Help Operation Instructions............................................................................................. 44
5 .15 Printing Function Introduction..........................................................................................45
5 .16 Alarm Management........................................................................................................... 45
5 .17 Central Monitor System Conglomerate (Interior Crunode Data Transmission)...... 48
6 Whole Set of Complete Accessories............................................................................................. 50
7 System Network Configuration.......................................................................................................51
8 Appendix.............................................................................................................................................52
8 .1 Waveform and Parameter Monitored by this System.................................................... 52
8 .2 TCP/IP Transfer Control Protocol......................................................................................53
8 .3 Dual-screen Display Configuration................................................................................... 54
I
1 Preface
You are welcome to use Central Monitor System Software ("this system" or "this product" is called for
short hereinafter). the latest edition, works well in Microsoft Windows XP, of which the free interface
and ways of operation will bring you much convenience during your work.
This specification includes all the functions and operation flow, some usual questions and detailed answers
offered.
Please read the specification carefully before using this product to smooth your usage .
Thanks for using products and look forward to your valuable suggestions and advice.
1 .1 Caution
Before usage, please be aware of safety and validity stated below:
The system should be used under doctors' guide.
The reliability of the product depends on whether the operation and maintenance are properly done.
Warning: system errors may arise if non-proper accessories are replaced; therefore, any maintenance
men who haven't been trained in our company or normal maintenance organization should not maintain
this system.
1 .2 Manipulator Obligation
1.Our company will resolve any software and hardware problems of the product in one year after your
purchase with the precondition of proper usage of it.
2.Responsible for users' training.
3.Responsible for free update and life maintenance of the software.
4.Responsible for free maintenance of hardware after your purchase.
5.Not responsible for damage caused by improper usage or non-fixed accessories usage.
1
2 General Introduction to Software
The system, the center of the monitors network, can be networked to many kinds of monitors. It can collect,
deal with ,analyze and output the information from the multi-monitors to realize concentrative monitor
with much force of lightening the huge burden of doctors and improving the monitory work.
2 .2 Function
Central Monitor System Software , the monitor assistant, has some functions below:
a. Up to 64 bedside monitors connectable
b. Display of holographic waveforms of the view bed of single bedside monitor; display of
ECG,RESP,CO2,PLETH,IBP waveforms and so on.
alarm manage: parameter alarm events for each bedside monitor; S-T segment alarm events; arrhythmia
alarm events
c. Alarm prompt modes :sound and light wave.
d. Storage and review of 24-hour and 64-lead holographic waveforms
e. Storage and review of up to 240-hour trend data for each bedside monitor
f. Storage and review of u to 720-item alarm events for each bedside monitor
g. Review of five parameters and 6 minutes short trend for each bedside monitor
h. All the review information will be stored under off-state
i. Display of continuous multi-lead ECG and single-lead arrhythmia analysis for each bedside monitor.
j. Trend records and alarm records can be output according to certain format.
k. Patient information management: receive, cancel, input and educe
l. 10,000 history patient information management: inquire, scan, maintain.
m. Central Monitor System gather function: relying on the data transmission ,the system has the
capability that in different position could simultaneously monitor the bedside monitors in the courtyard
n. All-round system help information including operation guide of this system and bedside monitor.
o. Multi-language display interface
p. Support screen (dual- screen display),laser printer, and sound output equipment
q. Support keyboard and mouse input
2
3 Main Technical Index
3 .2 Working Conditions
Temperature
Working 0 to 40 ℃
Transmit and store -20 to 60 ℃
Humidity
Working <= 80 %
Transmit and store <= 90 %
Altitude
Working -500 to 4,600m ( -1,600 to 15,000 ft)
Transmit and store -500 to 13,100m( -1,600 to 43,000ft)
Electric specification
AC 100-127/200-240 V 7A/4A
Applicability
Non patient area
Warning
Proper input pressure must fit the pressure indication of power supply, or else the machine will be
seriously damaged.
3 .3 Performance Index
Display: High-resolution 1024 * 768, 65536 color, optional LCD display
Language: Chinese and English
Indicator:
Waveform vision:2 path waveforms each bed , key bed observed extend to 12 path
Alarm indicator light
Mainframe indicator light
Interface:
Power supply input receptacle
Two serial interfaces
One parallel interface
Two USB interfaces
One RJ-45 network interface
3
One keyboard interface
One PS/2 mouse interface
Sound output interface
Mic input interface
Trend time
240 hours trend display for each bed
6 minutes short trend dynamic display for each bed
Alarm
Parameter alarm
S-T segment alarm
Arrhythmia alarm
Technical alarm
Audible and visual alarm with alarm range setting
Waveform storage
Storage and review of 24-hour, 64-path holographic waveforms
4
4 System Configuration Demand
5
5 Function Instructions
After the first startup of the program ,enter the main interface Double click the left key of the mouse in a
certain bed area, then enter the key observation function for the corresponding bed. several operation
interfaces(modules) as follows:
1.patient information 2.single bed observation 3.waveform review 4.trend review 5.alarm review 6.CO
review 7.NIBP review 8.display setting
Users can review history , set the system and examine help information via the menu of the main interface.
Detailed function specification of each module as follows.
6
System Prompt Information.
Part 1, 2,3, in District A of Picture 5.1 shows this product name ,startup time and current time respectively,
which can't be operated by users.
Control Icon
Four parts on the right hand of district A in Chart 5.1 are control icon parts, which show black and white.
When user puts the mouse on the icon, the corresponding ones will change into colors Users can obtain
different functions by mouse click.
Functions as follows:
Icon Function
Section B in Chart 5.1 consists of 8 patient information section.(part 5 shows one patient information
section ).As follow chart:
7
In the blue section (color blue means alarm signal appears and black no alarm. The background color can
be set up in system setting. ), section D on the left hand shows waveforms, section E on the right hand
shows parameters. the meaning of other parts as follows:
8: Red character is the serial number of this very section in 64 sections and link state; unlink bed is
This menu will be showed when right clicking waveform area in D section. Functions listed in below
chart:
in top left corner, the first two one are for moving
time sect, the third one is for canceling freeze, the 4th one is for
print.
review Enter review interface
Clear background Erase alarm back color
Amplify wave Amplifying waveform swing, make a waveform more clear; wave
could be amplified many times.
Shrink wave Dwindling waveform swing, canceling amplifying waveform.
8
13: Pop-up menu by right clicking;
This menu will be showed by right clicking a parameter area in Section E. It is used to set the alarm on-off,
when alarm is off, icon 13-1 will be displayed in relative parameter area.
13-1: Icon , when alarm of a parameter is off, this icon will be displayed in the parameter area.
64-bed status information:
doctor could manage No. 9 to No. 16 at the same time. There , By clicking button or on the top
can achieve page up and page down for viewing 8-bed status display. For example, number 1-8 are
displaying on the screen at present, when clicking , beds number 9-16 will be displayed after page
down. And 17-24 will be displayed when click again and vice versa ; another way is achieving page
up and down by bed indicating button, take 8-bed status chose and bed number 1-8 displayed for example,
when clicking any bed indicating button from number 17-24, 17-24 bed machine will be displayed, when
clicking any bed indicating button from number 49-56, 49-56 bed machine will be displayed on the screen.
Central Monitoring Systems adopts this kind of method to transfer all the information collected from the
64 bed machines to display them group by group on the screen.
Notice 1: under multi-bed observation mode, when left-clicking the mouse on one group, the 8 beds
machine of this group will be displayed on the screen.
Under single (key) bed observation mode, left-clicking one bed indicating button to change the bed
Number to the indicating button.
(Double display and Multi-bed and single-bed mode chose at the same time, left-clicking a bed to
change single bed display, the bed number will be displayed accordingly, if multi-bed display mode
want to be chose, or should be used; under double display and only multi-bed observation
mode, left-click one bed to change the whole group.
Notice 2: when conducting 2 , 4, 8 or 16 bed observation, the bed machines will be divided to 32, 16,
8 or 4 groups according to the screen display.
9
System Menu
The right part of the section c in Diagram 5.1 is System Menu. It contains five option:
Menu Function
System Setup
History review
Supplement
Users can choose bed from No. 1-64 by using combo box (blue part), and then click "Admit Patient"
button to admit patient (add new patient record) or use "Dismiss Patient" button to dismiss patient (e.g.
patient leaves hospital) . For each patient of the relative bed, relevant bed number and button should be
chose. Click the close button to close the dialog box when operation completed.
This System can store 1000 records at most.
Notice: a, Dismiss patient after admit him or her for a period of time, this patient can become
history patient.
b, Patient data can not be record if the patient not admit, so for this patient, data review and
freeze waveform can not be achieved.
c. If there are also some patients that not dismissed when the system closed, the data of the
10
patients will no be recorded the restart the system next time. Doctor or Nurse should re-click
2.Left-click Main Review Menu , a submenu will be popped up, that will be 2,4,8,16 bed
observation from the top, click the submenu to enter 2,4,8,16 bed observation interface. Double click
, or after clicking but not click submenu, instead to click other area on the
screen, last observation interface will be resumed on the main screen, if 2-bed observation displayed last
time, it will be resumed.
3.
Notice: breviary chart of the main screen display will be showed in the left part of the submenu; bed
number correspondingly displayed with the submenu will be 2 times as before under double display
mode.
11
Users can check names and information of the on line patients and also can change case number.
Users can input doctor name and remarks on the right part of the interface, click 『 』
button can store the new data to the database, click『 』 button to cancel the modification.
Notice: the interface can not be edit if no on line patient available on the bed machine.
12
Graph 5.3-2 The operation interface with multi-leads of ECG and none-trendline for single-bed
observation
Under default state, it protracts relevant waveform in display area according to configuration files, if this
waveform is not exist, it's area will be set off according to existing waveform number, every waveform
have got its name on top left corner of this waveform.
Because of the importance of ECG parameter, this system provide comprehensive browse mode for ECG
parameter, which can assist doctor in diagnosis. This system provide waveform of seven leads: I, II, III ,
aVR, aVL , aVF, V, when you click switch button on left top corner, it will be switch from general
waveform to multi-lead, then this display area includes seven waveforms as figure: 5.3-2, click this button
again, it will be switch from multi-lead to general waveform as figure: 5.3-1.
The waveform in this system is based upon information from monitors, if the monitor does not provide
signal of seven lead, this system can not calculate relevant parameter. Multi-lead inspection requires the
monitors provide ECG data from 2 channels of different leads, and this data should be identical.
This six minutes brief treadline is in top right corner of screen, which displays trend of relevant parameter
in 6 minutes, these parameters are consistent with parameters on its right side. User can click button on top
13
right corner of interface, which can switch the trendline between conceal and display.
Brief trend can memory and display average of parameters every 2 seconds, which is more accurate, more
timely reflect change of relevant signals. If the divided monitor does not provide relevant parameters, the
systems can not display relevant brief trend of parameters, and adjust number and position of brief trend
automatically.
This area includes two parts, the first five among all the parameters which connected currently are shown
at the right top corner in the interface, and these five parameter have five relevant area for display the
alarm higher and lower limit, user can select whether display alarm information by click right key or
14
Graph 5.3-4: Parameter Display Area: Alarm on/Alarm off
If the quantity of parameter is more than five, the redundant parameters will be displayed in the down right
corner of the screen, if you choose brief trend review, 20 parameters can be displayed at the same time, if
not, only 10 parameters can be displayed. If any parameter causes alarm, you can find alarm notice at
relevant parameters entry.
15
Graph 5.3-5: Two kinds of display of other parameters
If this system connected with a patient monitor, which starts alarm, but user did not shield this alarm, a
information as following will be displayed in the upper center of screen. Its color represents alarm level.
16
5 .4 Operation Introduction to Waveform Review
In the waveform review, every screen can display at most 5 waveforms of lately 24 hours. It also includes
parameter information at appointed time. If you need detailed information of any waveform, please choose
the waveform by double click on appointed waveform area. It includes waveform detail and parameters of
12 seconds, alarm information, in this column you can measure the space between waveform with
electronic caliper.
Click button "Wave Review" you can enter wave review main interface. Which displays waveform
information according to page number, every page includes 12 seconds of waveform, there are 7200 pages
(24 hours) totally in memory. The main interface includes following area under default state:
The waveform be displayed according to user's selection. Every screen include at best five waveforms of
12 seconds, waveform name be displayed at top left corner of screen. In waveform display area includes
several triangles for orientation of appointed waveform. The appointed information be displayed at the
bottom of waveform display area.
You can select appointed waveform from 12 seconds of waveform by click left key of mouse, if the
parameter be displayed now, you can review the parameter information according to data file.
Click left key of mouse twice in waveform display area, a relevant waveform frame will be displayed
17
according to you're the position you click. The detailed operation waveform frame of is as following:
You can choose appointed parameter by "ShowPara" button, which located in central area of the buttons.
This area includes an orientation bar, you can go to one page from page1 to page7200 by click the bar.
Under the bar it displays start time middle time and conclude time of retrospection, their sequence is from
left to right. It also displays page number under the time information.
This area includes buttons as following: Print, Record, Save As, ReviewSetup, Refresh, ShowPara,
AutoPlay and six direction keys. The direction keys includes: First page, last page, page up, page down.
18
Waveform Review Setting
Click the button "ReviewSetup", then you can set up the waveform review. This setting includes following
items:
Please select the waveform group, which is needed to be reviewed in this field. There are 4 groups of
selectable waveforms with the first 3 groups fixed and the 4th one user-defined. Then please select any
waveform which is needed to be displayed in 14 waveforms, but when more than 5 are selected there will
be only the first 5 selected waveforms displayed in the procedure.
19
(3) Option
User please set time internal of automatic page turning and page turning internal when turning pages.
Please press the button "Default" to recover the option of the dialog box to establish for recognizing the
setting, then press "OK" to save the setting in the configuration files. Please press "Cancel" to give up.
Please double-click the left mouse button in the waveform draws district of the waveform review interface,
then the process will establish banding box, which corresponds the waveform, in the waveform distinct
that the mouse double-clicked. Including the follows:
20
The functions of several buttons:
Trend is used to observe the change trend of the patient's certain physiology signal in a period. It is mainly
used for assistant analysis of the condition of an illness. The trend usually takes 1 minute as unit of account,
that is, that each parameter brings an average value in one minute denotes the trend value of this parameter.
It usually can be reviewed with trend draw and trend diagram.
Trend group/category
Trend group means to organize the associated parameter of physiology properly and provide the user more
integrated information of a certain physiologic index aspect.
The definition of trend group:
1. Vital signals: including the parameters of HR, SP02, NIBP, RR, TEMP.
2. ST changes: including HR, ST-Ⅰ/Ⅱ, RR, SP02, PVCS.
3. Hemodynamics: IBP1, IBP2, IBP3, IBP4, HR.
4. GAS: AA, AwRR, N20, 02, CO2Et.
Please clickbutton『Trend Review』in single bed or history mode, then enter the trend review interface. The
recognizing state is trend gram operation:
21
Diagram5.5-1 Trend Review Operation Interface (trend Graph)
Instructions as follows:
Trend gram is to turn the trend statistics sketch to embody the trend of a certain parameter numerical value
more. It will show trend waveform which is needed to review according to the user's choice, and every
screen can show trend waveforms of 5 parameters at most at the same time. There is a moveable dynamic
glide in the waveform display district. Drag it to the named position, and then the corresponding trend
numerical value will be shown at the parameter display district.
1. Continuous
Monodromy continuity, a colorized ploy-line, such as HR.
2. Non-periodic
Discontinuous parameter, such as NIBP.
3. Many parameters
To display with separated poly-lines, every poly-line shows color with corresponding parameter, to
display the name of parameter and corresponding numerical value with same color, such as ST1, ST2.
22
Trend Graph Instruction
1. Please review the specified bed's trend gram's information through choosing bed number.
2. The trend gram can be printed and output. Please click "Print" button to print the current screen with a
length of one page.
3. Please choose a trend group that is needed to be observed in the "Select trend group", and then the
corresponding trend group parameter will be shown in the parameter diagram on the left side of the screen.
4. Please choose the parameter which is needed to be shown in this parameter group in the parameter
diagram on the left side of the screen with 『 』as a checked state, and then the corresponding trend
group parameter will be shown in the parameter diagram on the left side of the screen.
5. Please choose the time length of reviewing in "select review time". There are 8 choices, including 12
hours, 24 hours, 48 hours, 72 hours, 96 hours, 120 hours, 168 hours, 240 hours from the current time.
6. Please choose resolution ratio of the waveform in "select resolution ratio". The time segment which is
shown by the waveform data in every page can be changed through choosing resolution ratio.
8. Please click the button『 』to get the trend data before the current clicking time.
23
9. Please click the button "Trend Table" 『 』 to observe the specified data value of the
current trend gram shown on the screen.
10. Please click and drag the triangle glide piece to the named position at the glide area, and then the
corresponding trend data will be shown at the parameter area of the trend gram.
11. Please click 『 』down at the right side of the screen, and then you can turn over
page forward or backward and go to the first or the last directly as well.
Trend Table
Please click the button 『 』down at the screen in the trend gram interface to go to the trend
table operating interface. Shown as follow:
Trend table is to organize data to do trend review by way of tables, through which you can check the
specified parameter data value of every minute. The trend table can be printed as well.
What the trend table shows is the corresponding parameter value of the trend gram review; every line from
left to right is: serial number, each parameter value and the corresponding measurement time. You can
24
change the data unit in the system setting. Number serial is only used to count and has nothing to do with
data value.
1. Please click the button『 』to reset the review of starting time which is shown at "Query
Time" down in the left side of the screen. You can also choose the time length and resolution ratio of the
review through two buttons "select review time" and "select resolution ratio" to adjust the data shown in
the trend table;
2. The trend table also has the function of turning over pages. There are 240/12 = 20 pages in total under
the recognizing situation. The specified time segment in the current page will be shown above the list.
3. Please print the data according to your need. The trend table can be printed and output while users can
print the current screen with a length of 1 page by clicking the "print" button.
6. Please check the page number of the current trend table page through 『 』down at the
right side of the trend table. And you can also input the page number directly and then press "enter" to
jump to the named page.
25
Enact the displayed parameter group.
In the trend table, the page number and total pages are shown. User can
input page number directly and press enter to jump to the named page.
26
Bar Table for Alarm
The alarm band list is up at the left side area of the alarm review interface. It shows all the alarm band
information of alarm level which accords with that is needed to query within the time range enacted by the
user according to the time order of the alarm, including the alarm time, types, level, code, and information
of the band. When entering the alarm review interface which is under the recognizing state, the first piece
of record in the list box is set to be picked state. User can click the corresponding entry in the list to choose
alarm band which is needed to be checked as needed.
There is an orientation glide under the list box, through which user can know what kind of alarm event
happens at what time roughly. User can click any area in the orientation glide by mouse, and then the list
will roll to the nearest alarm band record automatically in order to be easy for the user to choose.
The right side of the alarm band list is the alarm review setting and operation area, including the set of
alarm band level and time segment that are needed to be reviewed, starting time of review (the alarm event
in a certain time segment before the time). User can also save, delete, record the picked alarm band, refresh
the alarm review interface to get the latest alarm events, and do the operation of printing the band list and
alarm band.
alarm information
27
Special Alarm Bar Specification
Under the alarm review is the special alarm band specific description area, which can offer 2 waveforms
data of 8 seconds before and 4 seconds after the alarm, trend data in the latest 960 seconds of 1
corresponding parameter and numerical value information of some other important parameter at alarm time.
It can also offer calipers to measure the space between of the waveform. Information like names of two
waveforms and the alarm code, position of the band and speed shown by the band waveform is displayed
at the bottom of this area.
Operation Instructions
The specific ways of operation after entering the alarm review interface:
1. Please enter the alarm review interface of the certain bed through choosing bed number.
2. Alarm band and alarm list can both be printed and output. Please click 『 』button to
print the alarm information list on the current review interface and click 『 』button to
print the current picked alarm band, including 2 waveforms and numerical value information of other
important parameter of alarm time.
3. Alarm band can be recorded and output. Please click "record" to print, and the length is 12 seconds.
4.Uses can select the alarm categories needed observation in "select alarm type"
5.Users can select time length in "select review period". There are 8 choice: 24, 48, 72, 96, 120, 168 or
240 hours from present time.
28
6.Users can set alarm review time by clicking " " button.
7.Users can acquire present new alarm information by clicking " " button. The review
will stop when Refresh is clicked.
8.The result by users inputing alarm categories and review time will show on the bottom of the display
together with present alarm in the result.
9.The waveform name and 960-second trend description will show on the bottom of the display.
10.All parameters in alarm period will show on alarm strip.
11. Users can measure the waveform in alarm strip with calipers. Click the right key of the mouse and
daggle from the beginning to the end of the measurement, snd the system will show the measurement time
of waveform.
5 .7 CO Review
General Introduction
This system has separate review function for CO, which can review 240-hour CO record of appointed
patient (24 hours for patients checked before ) in sinfle-bed mode.
Users can enter CO review interface by clicking " " under single-bed or history mode:
29
Graph 5.7 COReview
User operation
The list shows CO data within 240 hours acquiescently; from left to right of the line: serial number, CO
value, TB (temperature of blood) value and corresponding measurement time. Data unit can be changed in
system setting. Serial number is used for taking count and has nothing to do with data value.
Users can select time beginning and length; Users should click " " after correct if just
want to change the length of the time period.
Each page (list) only shows data within 12 hours actually regarding data ammount; page up/down
function is supplied at the same time. Time period could be advanced or traced back 12 hours, in all
240/12 = 20pages acquiescently. The specific time period is displayed on top of the list.
Users can print the 12-hour data in present page as needed; patient information will also be printed.
Operation Instructions
The corresponding function of all buttons and options is the same as NIBP review:
30
To the first page
Display the page number and total pages. Users can input the
page number and click enter to the right page
5 .8 NIBP Review
General Introduction
This system has separate review function for NIBP, which can review 240-hour NIBP record of appointed
patient (24 hours for patients checked before ) in single-bed mode.
Users can enter CO review interface by clicking " " under single-bed or history mode:
31
Operation Instructions
The list shows NIBP data within 240 hours acquiescently; from left to right of the line: serial number,
contraction pressure, diastole pressure and corresponding measurement time. BP unit can be changed in
system setting. Serial number on left is used for taking count and has nothing to do with blood pressure
value.
Users can select time beginning and length; Users should click " " after correct if just
want to change the length of the time period.
Each page (list) only shows NIBP data within 12 hours actually regarding data ammount; page
up/down function is supplied at the same time. Time period could be advanced or traced back 12
hours, in all 240/12 = 20pages acquiescently. The specific time period is displayed on top of the list.
Users can print the 12-hour data in present page as needed; patient information will also be printed.
Display the page number and total pages. Users can input
the page number and click enter to the right page
32
Graph 5.9-1 Display setting interface
Single-bedside observation setting lists 14 waveforms for users to select. If the waveform is selected and
this bed transfers the waveform data, the waveform will displayed when observing this bed. Users can set
the wave speed and plus effect after selection.
Multi-bedside observation setting groups parameter display. Users can set which group of parameters to be
displayed. Each bed could display 4 waveforms at most. Users can select waveforms to display and set
corresponding wave protracting speed and plus effect.
33
Graph 5.9-3 Multi-bedside observation display setting
Button
Default setting
General Introduction
History data mean information of patient who has used this system before will be reserved if "patient
release" is select when the patient leaves the hospital. The system could reserve history data for 10,000
patients; the data include parameter information one day before the patient's leave and recent 50 alarm
issues.
History data review allow users to inquire according to patient name, medical history number, principal
physician and diagnosing content. Medical history number is distributed by hospital for patient when he
comes for the treatment. It is unique. For example, the patient named "Jack" may be not only one, but the
patient having the medical history number "20060616001" is only one. This system adopts fuzzy inquiry
for convenience, that is, users can check the patient information in smaller extension after listing the
records which include the demanding qualification.
34
Graph 5.10 History review interface
1.Users can set patient name, medical history number, physician name or diagnosing content as inquiry
terms. If several terms are input, system will pick out the patients who meet all demands
.
2.Users click "Query" button to inquiry all records that meet the demands and will be listed in "Inquiry
result" which lists separately patient name, medical history number, time to leave hospital, principal
physician name and diagnosing result.
4.Click "View" or dbclick chosen information in list. When interface changes to "Medical history rework
interface", the red font on upper left of the interface shows it is history review presently and the patient
name is XXX; all data are corresponding to the chosed history patient.
Please refer to History patient information for means to input patient data.
Attention: when turning to interface of history patient list, patient reviewed presently doesn't change and
needs rechoose to change.
35
Instructions to button in the interface are as follows:
Inquire according to patient
information
Select patient information
in list
Update present list of
history patient
Delete chosen history
patient records in list
Print present history patient
list
Page up/dowm to desplay
more history patients
Since medical history information is reserved to database and unchangeable, it couldn't be amended
directly but add new instructions to the original one. The operation of this interface is similar with the
interface of patient information under concernful observation.
There are also enter and educe functions of history patient in the interface.
36
Educing Operation
Click " " to educe history data of present history patient (chosen from History patient list)
to some file, including PatientInfo.xml, HisAlarm and HisTrend. Users can shift the files to other central
stand to make entering operation by moving storage facility.
Entering Operation
Click " " to choose the file including patient data and enter. The red font on upper left of
the interface shows entering data and patient name is XXX; the data can be operated the same as history
patient data.
Graph 5.11-2 File for choosing data of Educing & Entering operation
General Introduction
There are three interfaces to carry out monitoring, accessory and maintenance settings.
Monitoring Setup
It is mainly for general monitoring setting, such as alarm level of arrhythmia, choosing monitoring
parameter unit, setting means of alarm reference, etc.
Accessory Setup
37
It is mainly for setting system facility, such as recorder, audio card, modem, etc. Users can do internet
setting in senior version.
Maintain Setup
It is mainly for senior users and maintanence personnel. Password is needed to enter the interface in which
the following functions can be carried out: scan information, change languages, change password, change
system time, set hospital, etc.
User can enter setting window by clicking " " on lower right of the display. Monitoring
setting window is preset..
38
Instruction
Producer
Item Purpose Choosing scope
setting
Alarm silence Set time length of system alarm Manual 3 minutes, 10 minutes,
Duration Manual
Alarm Suspend Set time length of alarm 2 minutes 2 minutes, 3 minutes, 10
Duration hanging time minutes, permanent
Alarm Color Set time of setting reference 1 minute 1 minute, 3 minutes, 6
Screen hours
Arrhythmia Alarm Record when arrhythmia alarms Yes Yes, No
Record
39
PVC Set PVC alarm level Medium High, Medium
RonT Set RonT alarm level Medium High, Medium
TPT Set TPT alarm level Medium High, Medium
COUPLET Set COUPLET alarm level Medium High, Medium
VT>2 Set VT>2 alarm level Medium High, Medium
BIGEMINY Set BIGEMINY alarm level Medium High, Medium
TRIGEMINY Set TRIGEMINY alarm level Medium High, Medium
PNP Set PNP alarm level Medium High, Medium
PNC Set PNC alarm level Medium High, Medium
PNF Set PNF alarm level Medium High, Medium
MTI Set MTI alarm level Medium High, Medium
BEAT MISS Set BEAT MISS alarm level Medium High, Medium
SPB Set SPB alarm level Medium High, Medium
IPB Set IPB alarm level Medium High, Medium
APB Set APB alarm level Medium High, Medium
1.Users can set alarm mute, length of alarm hang and reference time of alarm background in all options of
2.Users can set the recording function of all parameters in " " and could do multiple
choice.
3.Users can set alarm level of all arrhythmia parameters in " "
4.Following is the dialogue box which appears when users click『 』area on the lower right of the
window to make sure what color is needed.
40
Graph5.13-2 Dialogue box for choosing alarm background color
7.The change could be effective only when click " " button after setting. If not, the setting
will be ineffective next time.
8.All setting will turn back to manufacture's setting when click " " button.
Click " " button on top of system setting to enter the interface:
41
Graph 5.13-3 Accessories setting interface
1.The above options can not be set in present version and part functions can be used in advanced version.
2 . IP, Subnet mask and gateway, which is used to correct or set internet IP when connecting, can be
3. All setting will turn back to manufacture's setting when click " " button. Internet IP
shows as follows in present version:
4.The change could be effective only when click " " button after setting.
42
Maintain Setup Interface and Operation
Click " " button on top of system setting to enter the interface:
1. System will show "Identity Confirmation" interface automatically before entering maintenance setting
window as the following graph. Input the password. Click "OK" if it is right, or "Cancel" if wrong, then
the maintenance setting window will not be able to use.(initial password:123456)
43
3.Users can set " " and input the target IP address in IP column
Users can choose ", so that it can be logged off only with password.
Users can choose " ", so that keyboard will be locked and part keys are forbidden to use.
5.Users can click " " button to change present systemic time.
8.Setting in this part (except " ") will be effective after clicking " " button; other
setting will be changed successfully after setting.
!Attention !
Avoid intended/unintended false manipulation in order to ensure system's normal operation.
Chaning password termly is adviced.
!Attention!
Please contact the maintenance for all-purpose one when password is forgotten and reset a new one
asap.
Notice
It will affect stored system data if time is adjusted forward, so it is forbidden to set time ahead of
present time. Therefore, changing the system time requests much caution.
Click the link going to be checked with left key of the mouse to replace the website with the corresponding
oneclicked.
Click the blank of the assisting website with right key of the mouse; the menu, which is different due to
different installed software but all including "Backwards" & "Forwards", will appear. Choose "Backfords"
for page up, "Forwards" for page down.
Or, when scanning website, click the top right button of the interface
44
or
or
in order to carry out the function of "Backwards" and "Forwards" for website scanning as foregoing menu.
Attention:
1、 The menu excludes backfords and forwards options when click pictures with right key. The right
operation menu will come out only when click the blank website.
2、 When assisting website is available, or and or can only operate the assisting
website. The button can be effective for observing beds when quit the scanning operatiin or switch into
another interface.
General Introduction
This system has two kinds of alarms: physiological information alarm and technical alarm
■ Physiological information alarm includes parameter alarm, ST alarm and arrhythmiaalarm
Physiological information alarm reminds doctor via visual and aural suggesting idea which meets
international standard. Visual aspect mainly includes lightening and glitter of color light; aural aspect
mainly includes suggestive sounds in different level.
Physiogical alarm actualizes via "alarm limit", which defines certain physiogical parameter is within
normal range. When parameter is beyond upper limit or lower limit of this range , the alarm will be
triggered.
!Notice!
Parameter alarm limit is from bedside through network; this system cannot revise it.
■ Technical alarm
Technical alarm mainly reflects fault reference of facility or hardware and remides doctor via two means of
vision and hearing.
45
Alarm Levels
Alarm level reflects the severity degree of alarm. There are three level:
Alarm level 1――― critical and highest-level alarm, also named first-degree alarm
Alarm level 2――― severe alarm, medium alarm, also named second-degree alarm
Alarm level 3――― common alarm, the lowest-level alarm, third-degree alarm
The alarm limit of parameters is transferred from bedside through network; it can't be revised.
This system judges the existent alarm by alarm information showing different background color through
single & multiple bed observation or by the changes of indicator lights of 64-bed situation under main
interface:
Red light on--highest-level alarm
Orange light on--medium-level alarm
Yellow light on--lowest-level alarm
Alarm Mute
Alarm mute means system will prompt by vision but sound when there is alarm.
Time limit of alarm mute can be set as 3 minutes, 10 minutes and permanence.
When 3 minutes or 10 minutes is set, the system will turn to non-mute alarm automaticly when time
reaches the setting; and it will be in alarm mute excepting users' intervention when "Permanence" is set
except users' intervention.
Users can unchain alarm mute at any time when mute is set.
Please refer to "Setup instructions" for time limit set of alarm mute
Inuring and unchaining of alarm mute are via icons in main interface. Please refer to "Main monitoring
interface--Controlling icons"
Alarm Hang
Alarm Hang means the system doesn't judge alarm in a period of time, ignoring psysiological information
alarm.
persisting time of alarm hang can be set as 2 minutes, 3 minutes, 5 minutes.
System will turn back to the former condition when time for alarm hang reaches the setting.
Users can unchain alarm hang at any time when it is set..
Please refer to "Setup instructions" for time limit set of alarm mute.
Inuring and unchaining of alarm mute are via icons in main interface. Please refer to "Main monitoring
interface--Controlling icons"
Alarm informationcan be prompted by visual and aural methods. This system supplies many means to
suggest users about the alarm:
■ Grounding Change
46
When psysiological information alarm occurs, system will prompt by updating the grounding in patient
section of corresponding bedside.
persisting time of grounding change can be set as 1 minute, 3 minutes or 6 hours. Please refer to "Setup
instructions" for detail.
Users can set specific updated grounding color which can also be found in Setup Instruction.
Users can click on right top of the procedureto adjust the volume.
!Notice !
The system alarms by choosing the highest alarm level when different bedsides alarm or different
parameters of one bedside alarm, accoring to corresponding alarm volume and tune.
47
5 .17 Central Monitor System Conglomerate (Interior Crunode Data
Transmission)
General Introduction
One central system can receive patient data information from not only the bedside but also another central
system.
Central monitor system can receive patient data information from bedside. At the same time, it can
transmit the information to other central systems completely or partially, which enhances real-time
observation and management in different places but not in one computer. As shown below:
1.Enter "Maintain setup" in "System Setup" and input password. Function interface transferred will show
as follows:
2.Pitch on " Host Destination " and input target IP on the right IP address column when transmitting is
needed. Tick off 『 』if not needed, and here the column does not allowed to input any IP.
48
3.Remember to click " " button after setting which will be effective when central monitor
system is restarted.
49
6 Whole Set of Complete Accessories
CD software: 1
USB softdog: 1
PC mainframe: 1
Display: 1 or 2, optional
Printer: 1, optional
50
7 System Network Configuration
A monitoring network system can be made up of bedside monitor the ( such as plug-in monitor, portable
monitoring device. Multi-Parameter Monitor) and this system which can be aslo extended to connect
defibrillator monitor. connecting the hospital network that connects to the Internet and other medical
information system and the formation of large-scale networks. Monitoring network system using client /
server (/) structure, Client for bedside monitoring equipment; Server refers to receive bedside monitoring
devices, data devices, in-system. Monitoring System was carefully observed in the physical layer using
RJ45 and linked by five kinds of non-shielded cable way . Data linking layer uses IEEE 802.3 Ethernet
standard, the network layer and transport layer use TCP/IP protocol.
Monitor netework system which can be connected by 64 units patient monitor works under transmission
speed of 10M or 100M . Followed is illustration of monitoring net:
51
8 Appendix
Monitored waveform:
Color
Name: Description:
RGB symbolize Red, Green and Blue
ECG1 RGB(0,255,0)
ECG2 RGB(0,255,0)
ECG3 RGB(0,255,0)
RESP RGB(0,255,255)
ART, #ART RGB(255,0,0) 0-300
PA, #PA RGB(255,0,0) -6 – 300
CVP, #CVP RGB(255,0,0) -10 – 50
RAP #RAP RGB(255,0,0) -10 – 50
LAP #LAP RGB(255,0,0) -10 – 50
ICP #ICP RGB(255,0,0) -10 – 50
AUX1,#AUX1 RGB(255,0,0) -10 - 300
AUX2,#AUX2 RGB(255,0,0) -10 - 300
HOCUS RGB(255,128,0)
CO2 RGB(255,255,0)
SPO2 RGB(255,255,0)
Waveform of restore factory settings: ECG1, ECG2, SPO2, RESP.
Monitored parameter:
52
NIBP RGB(255,255,255) Effective 1- 300
Ineffective: 0
TEMP RGB(255,255,255) Effective 0 – 50.0
Ineffective: -100
CO2 RGB(255,255,0) Effective 1-100
Ineffective:: 0
AWRR RGB(255,255,0) Effective 1-255
Ineffective: 0
PVCs RGB(255,255,0) Effective 1-255
Ineffective: 0
RR RGB(0,255,255) Effective 11-300
Ineffective: 10
CO RGB(255,255,255) Effective -50 - 50
HOCUS RGB(255,128,0) Effective 0 -270
The method of selecting parameter setting by factory is parameter group 1 ( Followed list)
53
of the agreements relating to the TCP network books.
2. Choose "set"attributes page from attributes dialoge box in the systems, Followed is the setup page:
54
3. Click "advanced"button from setup page, then shows the advanced setup dialog box as follow:
then you can set up by graphic card and driver of your own computer.
55
2.782.051.01ESS/1.0
1.4.15.05.006
2021.12