GC Detector Design Troubleshooting Flame Ionization Fid Theory Basics Gas Flows July212020
GC Detector Design Troubleshooting Flame Ionization Fid Theory Basics Gas Flows July212020
Agilent Technologies
YES! This means you can have up to 4 detectors on your 8890 GC or 3 detectors on your 8860 GC!
D1 Position
• FID
• TCD
• uECD
• FPD+
• NPD
• XCD D2 Position
• FID
• TCD
• uECD
TCD
FID
µECD
sim MSD scan
NPD (N)
NPD (P)
FPD (S)
FPD (P)
SCD
NCD
10-15 g 10-12 g 10-9 g 10-6 g 10-3 g
fg pg ng µg mg
1 ppt 1 ppb 1 ppm 0.1% %
• H2 / Air = Flame
• Sample is burned producing
positively charged ions
• Positive Ions are attracted to
the negative voltage on the
collector
• Collector current is converted
to FID output in the
Electrometer
Ignitor
Insulators Collector
Air
Jet
H2 and Make-Up
gas
Column nut
• 6890/7890 GC –
Capillary optimized FID base
OR
Packed FID base with
capillary adapter included
• Intuvo GC –
Capillary optimized FID base only
-no packed column capabilities on Capillary columns Adapters are available
Intuvo only for capillary columns or
1/8” Packed Columns
• 8860/8890 GC –
Capillary optimized FID base only ¼” Packed columns
A packed column adapter is can be installed
included without an adapter
m0 10 20 30 40 50 60 70
4
8
m
m
6
8
m
m 1
m
m
General Rule:
Push column up to inside the tip
of the jet and then pull back 1 mm
O
p
t
i
mu
mh
y
d
r
og
e
nf
l
ow
ra
t
e
ve
rs
u
s
ca
r
r
ie
r
f
lo
wr
a
t
e
5
0 Gas Flow Guidelines:
Hydrogen Flow (ml/min or kPa)
(
1
56
)
Hydrogen 30-35 ml/minute
4
0
(
1
32
)
Carrier + Makeup 30-35 ml/minute
Air – 400 ml/minute
3
0
(
1
07
) Nitrogen is the recommended makeup gas
2
0
H2-to-Air ratio should be between 8-12%
(
7
5)
2
0 3
0 4
0 5
0 6
0
Total Inert Gas should have ~1:1 ratio to H2
C
a
r
r
i
er
Fl
o
w (
N
2)
,m
l
/
mi
n
• It is best to operate the FID at 300oC or hotter – But always 20oC hotter than the highest
GC Oven Temp
Gas Flow Guidelines:
• Optimizing Flows 1:1 ratio of Carrier + Makeup to H2 flow Hydrogen 30-35 ml/minute
• Nitrogen is the recommended makeup gas Carrier + Makeup 30-35 ml/minute (≈1:1 ratio to H2)
Air – 400 ml/minute
• Make sure you check the base of the FID for any chunks of graphite when changing the column or
jet to prevent those chunks from getting into the jet and causing a partial blockage
• Check your Lit Offset. Default is 2pA and sometimes with very clean gases the baseline will drop
below 2pA causing your GC to think the flame is not it.
Pass Fail
• Set H2 Flow to 75 ml/minute
• Turn the Makeup Gas “OFF”
• Monitor Makeup Gas “Actual Flow”
➢ Low flow reading/minimal change – Jet is clear!
➢ High Flow reading/big change - indicates Jet Plugging…
Makeup and H2
Flows
An increase in the Makeup Gas flow reading indicates pressure backing up from the jet
Now the GC has the ability to run the Jet Restriction test from the front panel and generate a pass or fail.
Possible Causes:
• Contamination of the jet, collector, and/or the base by particulates, stationary phase, graphite
- Clean and Sonicate or Replace the Jet
- Thoroughly clean the collector and chimney assembly
- Thoroughly clean the base of the FID – especially if ferrule chunks are visible
New!
Smart
Gas Clean
Sensor
Possible Causes:
• Contaminated/Saturated Gas Supply Traps
• Buildup of high MW sample contamination in the GC Column
• Column Bleed
• Contamination in the Base of the Detector
• Gas Leaks
• Column Fitting Leaks
Remove/replace the
Remove the collector and silicone rubber gasket
Upper/lower PTFE Insulators
26 August 11, 2020 DE.6789930556
FID Maintenance Procedures
Jet Removal, Collector Removed
Shine a flashlight up
From the oven to check Look for graphite chunks
For impacted graphite Clean out the Detector Base – Top down falling into the oven
A syringe plunger works well
FID Jet
Interconnect
Spring
These on-board
procedures take you
through common
maintenance operations,
guiding you step-by-step,
cooling the instrument
zones that are necessary,
and even updating logs
and EMF counters when
finished!
Videos for Simple Maintenance and Theory can be found on the Agilent YouTube page:
https://www.youtube.com/user/agilent/
Checkout Agilent Communities where you can get answers and share insights:
https://community.agilent.com/welcome
SI-7693-1100s Agilent 7693A ALS Basic and Advanced Operation and Maintenance
GC-9000-2101cV3 Agilent 7890/9000 GC with OpenLAB 2.3 Essential and Advanced Operation – 4 Days
GC-9000-2103c Agilent Intuvo 9000 GC with OpenLAB CDS ChemStation Operation – 4 Days
GC-8890-2100c Agilent 8890 GC Operation with OpenLAB CDS ChemStation Edition – 4 Days