EE450: High Voltage Engineering: January 25, 2017
EE450: High Voltage Engineering: January 25, 2017
Lecture 21
Page 2
Measurement of High Voltages
• Different types of high voltages are generated in the laboratory to determine the
breakdown voltage of insulating materials.
• The high voltage equipment have large stray capacitances with respect to the
grounded structures and hence, large voltage gradients are set up.
• A person handling these equipments and the measuring devices must be protected
against these over voltages.
• The intensity of stray electric fields can be controlled by :
˗ Shaping the conductors to reduce stress concentrations
˗ Proper clearance between the equipment and the grounded structures to avoid
flashover
˗ Insertion of insulation of higher dielectric strength at high stress points
Page 3
Measurement of High Voltages
Page 4
Sphere Gaps
Page 5
Sphere Gaps
• The voltage at which the breakdown of the gap occurs is called breakdown voltage
and depends on the electric field Eb and the gap spacing d.
Sphere Gaps in
High Voltage
Laboratory
Page 6
Sphere Gaps
• There are some practical difficulties when very high or low voltages are measured
with the help of sphere gap.
• For the measurement of high voltages, the gap distance d should be increased.
Consequently, the uniform field between the spheres becomes distorted, and
accuracy falls.
• For maintaining the uniformity of the electric field, the radius of the electrode has to
be increased.
• The limits of accuracy are dependant on the ratio of the spacing d to the sphere
diameter D, as follows.
˗ d < 0.5 D, accuracy = ±3%
˗ 0.75 D > d > 0.5 D, accuracy = ±5%
Page 7
Sphere Gaps
• For the measurement of voltages less than 50 kV, or places where the availability of
ultraviolet radiation is low, irradiation of the gap by radioactive or other ionizing
media should be used.
• The irradiation of the sphere gap will increase the availability of initiating electrons in
the gap, thus reducing the statistical time lag.
• The development of spark process consists of two components:
˗ Statistical time-lag caused by the need of an electron to appear in the gap
during the application of the voltage.
˗ Formative time-lag which is the time required for the breakdown to develop
once initiated.
˗ Typically, the total time lag is of the order of few micro-seconds.
Page 8
Sphere Gaps
• The sphere gap can be used for measurement of impulse voltage of either polarity
provided:
˗ Wave-front and wave-tail times should be at least 1 μs and 5 μs respectively
˗ Gap length between the sphere should not exceed a sphere radius.
Page 9
Sphere Gaps
Page 10
Sphere Gaps
Page 11
Sphere Gaps
Page 12
Sphere Gaps
• It has been observed that the metal of which the spheres are made does not affect
the accuracy of measurements, so the spheres may be made of brass, bronze, steel,
copper, aluminium, or light alloys and that their surfaces should be:
˗ Surface of the sphere should be smooth
˗ Curvatures should be as uniform as possible
• The surface of the sphere should be cleaned, and free from dust, grease, moisture
and any other coating.
• The curvature should be measured by a spherometer at various positions over an
area enclosed by a circle of radius 0.3 D about the sparking point where D is the
diameter of the sphere and sparking points on the two spheres are those which are
at minimum distances from each other.
Page 13
Sphere Gaps
Measuring Circuit
• Continuous voltage measurement is not possible with sphere gaps, since the voltage
source is short circuited at the instant of measurement.
• A series resistance is connected between the source and the spheres to limit the
current that flows after the breakdown of the gap. The breakdown current may also
cause pitting on the surface of the spheres.
• The protective resistance should have a value of 1 Ω/V.
• For higher frequencies, the voltage drop would increase and it is necessary to have
a smaller value of the resistance.
• For impulse voltage, the protective resistors are not required.
Page 14
Sphere Gaps
Procedure
• For the measurement of a.c. or d.c. voltage, a reduced voltage is applied to begin
with so that the switching transient does not flash over the sphere gap and then the
voltage is increased gradually till the gap breaks down.
• Alternatively, the voltage is applied across a relatively large gap and the spacing is
then gradually decreased till the gap breaks down. Corresponding to this gap the
value of peak voltage can be read out from the calibration tables.
Page 15
Sphere Gaps
• For the measurement of 50% impulse disruptive discharge voltages, the spacing
of the sphere gap or the charging voltage of the impulse generator is adjusted in
steps of 3% of the expected disruptive voltage.
• Six applications of the impulse should be made at each step and the interval
between two applications is 5 seconds. The value giving 50% probability to disruptive
discharge is preferably obtained by interpolation between at least two gap or voltage
settings, one resulting in two disruptive discharges or less out of six applications and
the other in four disruptive discharges or more out of again six applications.
• Another method, simple though less accurate, is to adjust the settings such that four
to six disruptive discharges are obtained in a series of ten successive applications.
Page 16
Sphere Gaps
• If these conditions are satisfied and the specifications regarding the shape,
mounting, clearances of the spheres are met, the spheres will spark at a peak
voltage which will be close to the nominal value shown in standard calibration
tables.
• These calibration values relate to standard temperature and pressure (STP)
conditions:
˗ Temperature of 20℃
˗ Pressure of 760 mm Hg
Page 17
Sphere Gaps
Sphere Gaps
Sphere Gaps
• It should be noted:
˗ For a.c. and impulse voltages, the tables are considered to be accurate within
±3% for gap lengths upto 0.5 D.
˗ If the gap length is greater than 0.5 D, the results are less accurate and are
shown in brackets.
˗ The tables are not valid for gap lengths less than 0.05 D and impulse voltages
less than 10 kV.
Page 20
Sphere Gaps
Page 21
Sphere Gaps
Atmospheric Conditions
• The value of disruptive voltages as given in calibration tables correspond to 760 mm
Hg pressure and 20 ℃ .
• The breakdown voltage of a sphere gap increases with increase in pressure and
decreases with increase in temperature.
• The relative air density δ is given by,
Page 22
Sphere Gaps
Page 23
Sphere Gaps
Page 24
Sphere Gaps
Page 25
Sphere Gaps
Page 26
Sphere Gaps
Dust Particles
• When a dust particle is floating between the gap this results into erratic breakdown in
homogeneous or slightly inhomogenous electrode configurations.
• When the dust particle comes in contact with one electrode under the application of
d.c. voltage, it gets charged to the polarity of the electrode and gets attracted by the
opposite electrode due to the field forces and the breakdown is triggered shortly
before arrival.
Page 27
Sphere Gaps
Page 28
Uniform Field Spark Gaps
• Although the sphere gaps are widely used for voltage measurements, they have
only limited range with uniform electric field. Hence, it is not possible to ensure that
the sparking always takes place along the uniform field region.
• The sparkover voltage in a uniform field gap is given by,
Page 29
Uniform Field Spark Gaps
Page 30
Uniform Field Spark Gaps
• Bruce made uniform field electrodes with a sine curve in the end region. For the
Bruce profile,
˗ A = 24.22 and B = 6.08 under STP conditions
• According to Bruce, the electrodes with diameters of 4.5, 9.0, and 15.0 inches can
be used for maximum voltages of 140, 280, and 420 kV respectively.
• Rogowski presented a design for uniform field electrodes for sparkover voltages up
to 600 kV. Under STP conditions, that is, at temperature T = 25℃ and pressure =
760 torr,
˗ A = 24.4 and B = 7.50
˗ Constants A and B differ for a.c., d.c., and impulse voltages.
Page 31
Uniform Field Spark Gaps
• Later, it was found that with humidity the sparkover voltage increases, and the
relationship for sparkover voltage was modified as,
Page 32
Uniform Field Spark Gaps
Page 33
Uniform Field Spark Gaps
• Advantages:
˗ The sparkover voltage of uniform field electrode gaps can also be found from
calculations. However, no such calculation is available for sphere gaps.
˗ No influence of nearby earthed objects.
˗ No polarity effect.
• Disadvantages:
˗ Very accurate mechanical finish of the electrodes and careful alignment of the
electrodes are difficult to obtain.
˗ Presence of dust particles may cause erratic breakdown of the gap.
Page 34
Rod Gaps
Page 35
Rod Gaps
• Table shows the power frequency breakdown voltage for 1.27 cm square rods in
air at 27℃ and at a pressure of 760 torr with the vapour pressure of water of 15.5
torr.
Page 36
Rod Gaps
• Except for low voltages (less than 120 kV), where the accuracy is low, the
Page 37
Rod Gaps
• Like sphere gap, the breakdown voltage of rod gap depends upon:
˗ Atmospheric pressure
˗ Atmospheric temperature
˗ Humidity
• Because of the large variation in breakdown voltage for the same spacing and the
uncertainties associated with the influence of humidity, rod gaps are no longer used
for measurement of a.c or impulse voltages.
Page 38
Rod Gaps
• Electrostatic Voltmeters
• Ammeters in series with high ohmic resistors
• Generating voltmeters
Page 39
THANK YOU FOR YOUR ATTENTION