Nitrogen, Ammonia: Test Preparation
Nitrogen, Ammonia: Test Preparation
Test Preparation
IF
LL
IL
EN
1. Select the test. 2. Insert the Multi-cell 3. Prepared Sample: 4. Blank Preparation:
Adapter with the 1-inch Fill a square sample cell to Fill a second square
square cell holder facing the 10-mL mark with sample cell to the 10-mL
the user. sample. mark with deionized water.
5. Add the contents of 6. Press TIMER>OK. A 7. When the timer 8. Press TIMER>OK. A
one Ammonia Salicylate three-minute reaction expires, add the contents 15-minute reaction period
Powder Pillow to each cell. period will begin. of one Ammonia Cyanurate will begin.
Reagent Powder Pillow to
Stopper and shake to A green color will develop if
each cell.
dissolve. ammonia-nitrogen
Stopper and shake to is present.
dissolve.
9. When the timer 10. Press ZERO. The 11. Wipe the sample and
expires, insert the blank display will show: insert it into the cell holder
into the cell holder with the 0.00 mg/L NH3–N with the fill line facing the
fill line facing the user. user. Results are in mg/L
NH3–N.
Nitrogen, Ammonia
Page 2 of 6 NitrogenAmm_8155_PP_SAL.fm
Nitrogen, Ammonia (0.01 to 0.50 mg/L NH3–N)
Interferences
Table 1 Interfering Substances and Levels
Interfering Substance Interference Levels and Treatments
Calcium Greater than 1000 mg/L as CaCO3
All levels. Correct for iron interference as follows:
1. Determine the amount of iron present in the sample by following one of the Iron, Total,
Iron procedures.
2. Add the same iron concentration to the ammonia-free water in step 4. The interference will
be successfully blanked out.
Magnesium Greater than 6000 mg/L as CaCO3
Monochloramine present in chloraminated drinking water interferes directly at all levels, giving
Monochloramine high results. Use Method 10200, Free Ammonia and Monochloramine, to determine free
ammonia in these sample matrices.
Nitrate Greater than 100 mg/L as NO3––N
Nitrite Greater than 12 mg/L as NO2––N
Phosphate Greater than 100 mg/L as PO43––P
Sulfate Greater than 300 mg/L as SO42–
Sulfide will intensify the color. Eliminate sulfide interference as follows:
1. Measure about 350 mL of sample in a 500-mL Erlenmeyer flask1.
Sulfide 2. Add the contents of one Sulfide Inhibitor Reagent1 Powder Pillow. Swirl to mix.
3. Filter the sample through a Folded Filter Paper1 and Filter Funnel1.
4. Use the filtered solution in step 3.
Less common interferences such as hydrazine and glycine will cause intensified colors in the
prepared sample. Turbidity and color will give erroneous high values. Samples with severe
Other Substances
interferences require distillation. Use the distillation procedure with the General Purpose
Distillation Set.
Adjust the pH to 2 or less with concentrated (about 2 mL per liter) Sulfuric Acid. Store samples
at 4 °C or less. Samples preserved in this manner can be stored up to 28 days. Just before
testing the stored sample, warm to room temperature and neutralize with 5.0 N Sodium
Hydroxide Standard Solution. Correct the test result for volume additions.
Accuracy Check
1. After reading test results, leave the sample cell (unspiked sample) in the instrument.
3. Press OK to accept the default values for standard concentration, sample volume, and
spike volumes. Press EDIT to change these values. After values are accepted, the
unspiked sample reading will appear in the top row. See the user manual for more
information.
Nitrogen, Ammonia
NitrogenAmm_8155_PP_SAL.fm Page 3 of 6
Nitrogen, Ammonia (0.01 to 0.50 mg/L NH3–N)
5. Prepare three sample spikes. Fill three mixing cylinders with 25 mL of sample.Use the
TenSette® Pipet to add 0.2 mL, 0.4 mL, and 0.6 mL of standard, respectively to the
cylinders and mix each thoroughly.
6. Analyze each sample spike as described in the procedure above, starting with the 0.2 mL
sample spike. Accept each standard additions reading by pressing READ. Each addition
should reflect approximately 100% recovery.
7. After completing the sequence, press GRAPH to view the best-fit line through the standard
additions data points, accounting for matrix interferences. Press IDEAL LINE to view
relationships between the sample spikes and the “Ideal Line” of 100% recovery.
1. Diluting 4.00 mL of the Ammonia Nitrogen Standard Solution, 10-mg/L, to 100 mL with
deionized water. Or, use the TenSette® Pipet to prepare a 0.40 mg/L ammonia nitrogen
standard by diluting 0.8 mL of an Ammonia Nitrogen Voluette® Standard Solution, 50-mg/L
as NH3–N, to 100 mL with deionized water.
2. To adjust the calibration curve using the reading obtained with the standard solution, press
OPTIONS>MORE on the current program menu. Press STANDARD ADJUST.
Method Performance
Precision
Standard: 0.40 mg/L NH3–N
Sensitivity
Summary of Method
Ammonia compounds combine with chlorine to form monochloramine. Monochloramine reacts
with salicylate to form 5-aminosalicylate. The 5-aminosalicylate is oxidized in the presence of a
sodium nitroprusside catalyst to form a blue-colored compound. The blue color is masked by
the yellow color from the excess reagent present to give a final green-colored solution. Test
results are measured at 655 nm.
Nitrogen, Ammonia
Page 4 of 6 NitrogenAmm_8155_PP_SAL.fm
Nitrogen, Ammonia (0.01 to 0.50 mg/L NH3–N)
Required Apparatus
Description Quantity/Test Unit Cat. No.
Sample Cells, 1-inch square, 10-mL, matched pair 2 2/pkg 24954-02
Nitrogen, Ammonia
NitrogenAmm_8155_PP_SAL.fm Page 5 of 6
FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING: HACH COMPANY
In the U.S.A. – Call toll-free 800-227-4224 WORLD HEADQUARTERS
Outside the U.S.A. – Contact the HACH office or distributor serving you. Telephone: (970) 669-3050
On the Worldwide Web – www.hach.com; E-mail – [email protected] FAX: (970) 669-2932
© Hach Company, 2005. All rights reserved. Printed in Germany. Updated April 05 Edition 1