OCTAVIUS 4D White Paper Note en 91320006 00
OCTAVIUS 4D White Paper Note en 91320006 00
06/00
PTW D913.200.06/00 2
3.3 Dose distribution in the OCTAVIUS 4D reconstructed this way. The dose reconstruction time
phantom needed by OCTAVIUS 4D is approximately 25 s.
OCTAVIUS 4D and its associated software VeriSoft are
designed to measure and reconstruct a 3D dose grid without 3.4 Dose distribution in the patient
using dose information from the treatment planning system. In order to reconstruct the 3D dose distribution in the
Therefore, the measuring results of the QA test tool are patient, the CT data containing the densities of the
completely independent of the treatment plan. The basics of structures must be imported into VeriSoft. The patient dose
the 3D dose reconstruction algorithm are easy to reconstruction follows the steps described in chapter 3.3,
understand: however with a different procedure in step g). In contrast to
the phantom, the patient is not homogeneous along the ray
a) Convert the PDDs measured in water upon line but features structures with different densities. Also, the
commissioning to PDDs in the OCTAVIUS 4D source-to-surface-distance differs because of the irregular
phantom, using the known relation of the electron shape of the patient contour. This situation is depicted in
densities of water and phantom material Fig. 2.
b) At the current gantry angle (time) consider one
detector of the detector panel (“current detector”)
c) Measure the dose DDet in Gy at this position
d) Construct a ray line through the current detector to the
focus of the beam
e) Determine the current field size from the irradiated
detectors
f) Apply corrections for non-central-axis TPRs according
to [14] in case of beams with flattening filter, or
according to [15] in case of flattening filter free (FFF)
beams
g) Inside the phantom, using the PDD appropriate for the
current field size, reconstruct dose values D(r) in Gy at
the distance r from the current detector along the ray
line according to equation (1)
h) Do this for all detectors of the detector panel
i) Do this for all gantry angles
j) Sort all of the dose values obtained this way into
voxels of 2.5 x 2.5 x 2.5 mm³ (can be changed by the
user) by linear interpolation
k) For the Detector 729, remove a layer of 3 cm in
thickness from the outer shell of the phantom to obtain
a dose grid in a cylinder with 26 cm in diameter and
Fig. 2 OCTAVIUS 4D phantom with ray line (3) through focus and
26 cm in length. For the Detector 1000 SRS the
current detector (2), not to scale. The contour (4) is the patient
reconstructed cylinder is 11 cm in diameter and 11 cm surface from the CT image. The current detector measures the
in length. dose DDet at the water-equivalent depth zPhantom. The algorithm
reconstructs the dose DCT for the current voxel (1) at the water-
The dose D(r) along a ray line is obtained from the dose equivalent depth zCT in the CT image. aDet and aCT are geometrical
DDet(0), measured by the current detector, using the distances from the focus to the current detector and to the current
following relationship voxel, respectively.
PDD(r )
D(r ) DDet (0) (1) To determine the patient dose at a point on the ray line, a
PDD(0) relationship between the dose measured in the phantom by
the current detector, DDet, and the dose at the point on the
where PDD(r) and PDD(0) are the percentage depth dose ray line in the CT image, DCT, is established. The steps of
values at the distance r from the current detector and at the the algorithm are:
position of the current detector, respectively.
g.1) Convert the PDDs to tissue-phantom-ratios (TPRs)
The outer shell of the phantom is removed from the g.2) Select a point (“current voxel”) on the ray line
reconstructed data as in this region no detectors exist. constructed in d)
Scattered radiation is accounted for by the detectors g.3) Along the ray line, convert the Hounsfield units from
surrounding the current detector. For a typical VMAT plan the CT to electron densities [16]
approximately 1.1 million voxels (dose points) are
PTW D913.200.06/00 3
g.4) Determine the water-equivalent depth of the current
detector in the phantom, zDet, by multiplication of the
geometric depth with the ratio of the electron densities
of the phantom material and water
g.5) Determine the water-equivalent depth of the current
voxel in the patient, zCT, according to equation (3)
g.6) Determine the geometrical distances between focus
and current voxel, aCT, and between focus and current
detector, aDet
g.7) Calculate the dose in the patient DCT at the current
voxel by application of equation (2)
g.8) Do this for all voxels along the ray line inside the
patient’s contour.
Fig. 3 Dose volume histograms in the patient’s geometry
determined by OCTAVIUS 4D (dashed lines) and by the TPS (solid
The reconstruction of the dose in the patient (or in the CT
lines)
image) along a ray line through the current detector and the
focus is based on equation (2). The meaning of the symbols
is explained above.
Mostly, dose distributions in the patient are visualized by
means of dose volume histograms (DVHs). Usually, DVHs
2
TPR ( zCT ) a Det are presented not for the complete patient but for different
DCT DDet (2)
TPR ( z Det ) aCT
structures such as organs at risk or target volumes. In order
to determine DVHs the VeriSoft software must import the
geometry of the patient structures from the TPS. The DVH
The first term in equation (2) accounts for different algorithm uses the reconstructed 3D dose distribution in the
thicknesses of overlaying material in front of the current patient and determines the histograms from the dose values
detector in the phantom and in front of the current voxel in inside the structures. VeriSoft only considers structures that
the patient (or CT). The water-equivalent depth zCT of the are completely inside the OCTAVIUS 4D measuring volume.
current voxel in the patient is obtained from the geometrical If the TPS allows DVH curves to be exported, they can be
geom
depth z CT and the electron densities of the materials compared with the DVH curves determined by
OCTAVIUS 4D as depicted in Fig. 3.
involved:
n
CT , i
4. Reconstruction accuracy
zCT z geom
i 1
(3)
n Water
CT
This algorithm is not able to account for the change of The accuracy of the DVHs was also tested employing two
scattered radiation as the position on the ray line passes different methods. First, the reconstructed maximum dose in
structures with different density. the patient was compared with the TPS dose calculation.
Second, a full-volume gamma analysis on the reconstructed
The time needed by OCTAVIUS 4D to reconstruct the 3D dose volume was performed by comparison with the
dose distribution in the patient is typically in the order of two treatment plan (chapter 4.4).
minutes.
PTW D913.200.06/00 4
Before the measurements and the data analysis the
OCTAVIUS 4D system was correctly commissioned as
described in chapter 3.2. Special attention was given to the
correct adjustment of the phantom’s electron density in the
treatment planning system.
PTW D913.200.06/00 5
4.3 Comparison with a stationary detector panel A similar comparison between OCTAVIUS II and
and with TPS results OCTAVIUS 4D was made by the delivery of a RapidArc lung
A more realistic test is the comparison of OCTAVIUS 4D plan with nodes. Here the pass rates were 100% for
with a system using a stationary detector panel by applying OCTAVIUS II and 99.6% for OCTAVIUS 4D.
an IMRT plan or a rotational plan. For this purpose, an
OCTAVIUS II system, consisting of a stationary octagonal Finally, a clinical prostate plan was measured with
phantom and a Detector 729, was irradiated with an IMRT OCTAVIUS 4D using detector panels with different spatial
Head & Neck plan consisting of 11 fields. The coronal plane resolution. Fig. 8 shows the results for the Detector 729 with
defined by the detector panel was evaluated with VeriSoft. 10 mm resolution and for the Detector 1000 SRS with 2.5 mm
Then the same irradiation plan was applied to an resolution. At 3% / 3 mm both pass rates were 100%,
OCTAVIUS 4D system, and the same coronal plane was therefore the criterion was changed to 2% / 2 mm. The pass
extracted from the 3D dose grid. Fig. 7 shows the rates were then 95% and 98.5%, respectively.
comparison of both measurements with the TPS. The
OCTAVIUS II achieved a pass rate of 99.8%, while the pass
rate of the OCTAVIUS 4D was 99.2%.
PTW D913.200.06/00 6
of clinical treatment plans the maximum doses were additional data consist of PDDs at different field sizes and
determined from the reconstructed plan and from the TPS. the phantom’s electron density. For the dose distribution in
The differences between reconstruction and TPS were the patient, geometric information on the patient’s structures
noted and their mean values calculated. The uncertainty of and their densities are needed in addition. No dose
the maximum DVH dose measured with OCTAVIUS 4D information from the TPS is used to establish the measured
was found to be ± 2% for relatively homogenous plans, and dose distributions, making the OCTAVIUS 4D system a real
± 6% for lung plans. These uncertainties are approximately independent QA tool.
twice as high as the uncertainty of the DVHs calculated by
the TPS. The biggest discrepancies between OCTAVIUS 4D and
TPS data is found for cross profiles measured with the
For a deeper insight into the accuracy of the reconstructed Detector 729 in static fields (Fig. 4). This is the result of the
3D dose grid in the patient, the commercial version of relatively coarse spatial resolution of the detector panel. The
VeriSoft was modified to allow the comparison of individual detectors are 0.5 x 0.5 x 0.5 cm³ in size and
reconstructed dose planes in the patient with those from the separated by 10 mm center to center. The fact that the
TPS. Fig. 9 shows the result of such a gamma analysis in a phantom normally rotates with the beam smears out this
patient’s plane. resolution effect to a big degree, increasing the pass rate in
the evaluated plane from 83.3% to 99.8%. Also, the use of a
detector panel with a higher spatial resolution improves the
results (Fig. 5).
5 Discussion The data shown in Fig. 8 clearly demonstrate the fact that
the pass rate for clinical treatment plans raises as the
The algorithms used by OCTAVIUS 4D are straightforward, spatial resolution of the detector panel is increased.
using only the gantry-resolved measured dose matrices, the
associated gantry angles, and a very limited set of additional Although the problem is much reduced by the rotation of the
data provided by the user upon commissioning of the phantom, the spatial resolution of 10 mm for the
system. For the dose distribution in the phantom the Detector 729 and the associated linear interpolation
PTW D913.200.06/00 7
algorithm result in small, but detectable deficiencies in the [3] Ann Van Esch et al., “On-line quality assurance of rotational
radiotherapy treatment delivery by means of a 2D ion chamber
gamma analysis. In this context, however, it is worth noting
array and the Octavius phantom”, Med. Phys. 34 (10), October
that comparable devices are having similar restrictions. The 2007, 3825 - 3837
Delta-4, for instance, also uses a linear interpolation
[4] Paul A Jursinic et al., “MapCHECK used for rotational IMRT
algorithm [8], and the ArcCHECK requires agreement
measurements: Step-and-shot, Tomotherapy, RapidArc”, Med.
between measurements and treatment plan within Phys. 37, (6), June 2010, 2837 - 2846
approximately 10% in order to be able to reliably correct the
[5] M Stasi et al., “D-IMRT verification with a 2D pixel ionization
calculated patient dose distribution [10]. For the COMPASS chamber: dosimetric and clinical results in head and neck cancer”,
system dose errors of up to 17% have been reported [9]. Phys. Med. Biol. 50 (2005) 4681 - 4694
of dose volume histograms in the patient’s geometry has [17] “Central Axis Depth Dose Data for Use in Radiotherapy”,
been estimated to vary from ± 2% to ± 6% depending on British Journal of Radiology, Supplement 11, London 1972
the magnitude of the density variations in the patient. The
[18] Tom Depuydt et al., ”A quantitative evaluation of IMRT dose
accuracy of the system can be improved by the use of distributions: refinement and clinical assessment of the gamma
detector panels with higher spatial resolution such as the evaluation”, Radiotherapy and Oncology 62 (2002) 309 - 319
SRS
Detector 1000 . [19] I. Ali et al., “Quantitative assessment of the accuracy of dose
calculation using pencil beam and Monte Carlo algorithms and
requirements for clinical quality assurance”, Medical Dosimetry, in
press, 2013
PTW D913.200.06/00 8