PartLow MRC7700 Operqating Manual
PartLow MRC7700 Operqating Manual
MRC 7000
Installation, Wiring, Operation Manual
Brand
I nformation in this installation, wiring, and operation
manual is subject to change without notice. One
manual is provided with each instrument at the time of
shipment. Extra copies are available at the price published
on the front cover.
2
TABLE OF CONTENTS
SECTION 3 - CONFIGURATION
3.1 Configuration (Set Up) 20
3.2 Configuration/Jumper Positioning 21
3.3 Operation Summary 21
3.4 Start Up Procedures 21
3.5 Front Panel Operation 22
SECTION 4 - OPERATION
4.1 Off Control Mode 37
4.2 Alarm Operation 41
4.3 Tune Mode Operation 42
SECTION 5 - SERVICE
5.1 Service 45
5.2 Changing Charts 45
5.3 Changing Pens 45
5.4 Calibration 46
5.5 Test Mode Procedures 51
5.6 Troubleshooting and Diagnostics (Error Code Definitions) 55
APPENDICES
A - Board Layouts and Jumper Positioning
A-1 - Processor Board 62
A-2 SPST Relay/SSR Driver Output Board 63
A-3 SPDT Relay/SSR Driver Output Baord 64
A-4 Current Output Board 65
B - Glossary 66
C- Order Matrix 68
D- Product Specifications 69
E- Software Record/Reference Sheet 72
Warranty Inside back page
3
FIGURES & TABLES
Figure 1-1 Recorder Description 5
Figure 1-2 Recorder Display 7
Figure 2-1 Installation Panel Dimensions Conduit Opening Locations 9
Figure 2-2 Noise Suppression 11
Figure 2-3 Noise Suppression 11
Figure 2-4 Board and Terminal Locations 14
Figure 2-5 AC Power Input 15
Figure 2-6 Thermocouple Inputs 15
Figure 2-7 RTD Inputs 15
Figure 2-8 Milliamp, Volt and Millivolt Inputs 16
Figure 2-9 Remote Setpoint Input VDC, mADC 16
Figure 2-10 Digital Communications 17
Figure 2-11A SPST Relay Output 17
Figure 2-11B SPDT Relay Output 18
Figure 2-12 SSR Output 18
Figure 2-13 Current Output 18
Figure 2-14 24 VDC Power Supply Option 19
Figure 2-15 Position Proportioning Control Output 19
Figure 3-1 Keypad Features 24
Figure 5-1 Changing Pens 45
Table 3-1 Program Mode Configuration Procedure 28
Table 3-2 Tune Mode Configuration Procedure 34
Table 3-3 Enable Mode Configuration Procedure 36
Table 5-1 Calibration Procedure 47
Table 5-2 Test Procedures and Description 52
FLOW CHARTS
Flow - Calibration 46
Flow - Enable Mode 36
Flow - Program Mode 25
Flow - Setpoint Select 38
Flow - Standby 41
Flow - Test 50
Flow - Tune Mode 33
4
Product Description 1.1
1.1.1 GENERAL
The instrument is a microprocessor based circular chart Recorder Controller capable of
measuring, displaying, recording and controlling from a variety of inputs. Applications include
temperature, level, pressure, flow and others. The instruments can be specified as either a
single or as a dual pen model.
Recording, control functions, alarm settings and other parameters are easily entered via the
keypad. All user's data can be protected from unauthorized changes by the Enable mode
security system, and is protected against memory loss, as a result of AC power outage, by
battery back-up.
The process sensor input for each terminal block is user configurable to directly connect to
either thermocouple, RTD, mVDC, VDC, or mADC inputs. Changes in input type can easily be
made by the user. Thermocouple and RTD linearization, as well as thermocouple cold
junction compensation, are performed automatically. The instrument process variable inputs
are isolated. An isolated 24 VDC regulated transmitter power supply can be provided in the
instrument for use with up to two 4 to 20 mADC process sensor transducers.
The instrument can be ordered to operate on either 115 VAC or 230 VAC power at 50/60 Hz.
The 230 VAC option includes a switch for selecting either 230 VAC or 115 VAC operation.
The instrument is housed in a structural foam enclosure suitable for panel or surface
mounting.
FIGURE 1-1
Pen 1 Display
Up Key
Down Key
Pen 2 Auto/Man Key
Pen 2 Display
The number of keys and LED's will depend upon the configuration of individual unit.
5
1.1.2 RECORDING
The instrument records the selected process variable on a 10-inch circular chart. One box of
standard charts is provided with each recorder. Charts are available in a wide selection of
ranges. Chart rotation speed is programmable from 0.1 to 999.9 hours per revolution in 0.1
hour increments. The instrument can be ordered with one or two pens. Pen 1 is red and Pen 2
is green. Pens are the disposable fiber-tip type.
1.1.3 DISPLAYS
Each instrument is provided with a digital display and status indicator for each pen provided
(See Figure 1-1, page 5). The display may be configured to display the Process Value,
Process Value and Setpoint, Deviation from Setpoint only, Deviation and Setpoint, or Setpoint
only. During configuration the display(s) is/are used to show the enabled modes of operation
and the parameter codes.
The display in the upper right corner is for Pen 1, the display in the lower right corner is for
Pen 2 (if provided). The display includes status indicators for Manual mode operation, Output
1, Output 2 , Alarm, degrees C, degrees F, engineering units, setpoint and minus sign. See
Figure 1-2 ( page 7).
Display resolution is programmable for 0.1 or 1 degree for thermocouple and RTD inputs, and
none, one, two or three decimal places for other input types.
1.1.4 CONTROL
The instrument can be provided with relay, solid state relay driver and milliamp DC outputs.
Instruments can be programmed for on-off, time proportioning, current proportioning or
position proportioning control depending upon the hardware present. Switching between the
Control mode and the Manual mode of operation is easily accomplished with a dedicated key
on the keypad. Switching is bumpless from the Control to the Manual mode, and while in
manual, manipulation of proportional outputs is possible. Each pen of a dual pen recording
controller is provided with its own AUTO/MANUAL key . Other standard control features
include proportional control output limits, setpoint limits, anti-reset windup and a unique
Automatic Transfer function. If configured, the Automatic Transfer function allows manual
control of the proportional output until the process reaches the setpoint at which time the
instrument will go into the Control (automatic) mode of operation.
1.1.5 ALARM
An Alarm indicator is standard for each pen. Two alarm functions are provided for each pen
and the alarm indicator will light if either alarm for that pen is on. Alarm settings are program-
mable. Alarm type may be selected as process direct or reverse (high or low), deviation from
setpoint direct or reverse, and deviation band open or closed within the band. Alarm outputs
can be provided by assigning any relay(s) Single Pole/Single Throw (SPST)or Solid State
Relay (SSR) driver to the respective alarm.
6
1.1.7 DIGITAL COMMUNICATIONS
The instrument can be ordered with a Digital Communications option that provides the
capability of bi-directional communications with a supervisory computer. A dual pen
instrument can have an individual address selected for each pen. Refer to the Communica-
tions Protocol Manual (Form 2878) for more details regarding the communications option.
This manual is included with the unit when the communications option is specified.
FIGURE 1-2
SP
Setpoint
C
F
Minus Sign U
Digital Display
7
Installation and Wiring 2.1
Read these instructions carefully before proceeding with installation and operation. Electrical
code requirements and safety standards should be observed. Installation should be performed
by qualified personnel.
CAUTION: The Instrument AC power input is specified in the model number and on the
wiring label affixed to the the top center of the platen. Verify the AC power input required
by the instrument prior to proceeding with installation.
Unpacking 2.2
Remove the instrument from the carton and inspect for any damage due to shipment. If any
damage is noticed due to transit, report and file a claim with the carrier. Write the model
number and serial number of the instrument on the inside of the front cover of this Operation
Manual for future reference.
Location 2.3
Locate the instrument away from excessive moisture, oil, dust, and vibration. Do not subject
the instrument to operating temperatures outside of the 32°F to 131°F (0°C to 55°C) range.
Mounting 2.4
Figure 2-1 (page 9) shows an installation view and physical dimensions for a panel mounted
instrument. The panel where the instrument will be mounted must provide rigid support for
the approximately 20 pound instrument. Adjacent instruments may be mounted within a
minimum of 2 inches horizontally and 3 inches vertically, providing that proper panel support
is supplied.
PANEL MOUNTING HARDWARE REQUIRED: (not provided with instrument)
(4) 1/4"-20 x 2" flat head bolts w/nuts
(4) appropriate lock washers
PANEL MOUNTING:
1) Cut panel opening to the dimensions illustrated in Figure 2-1 (page 9).
2) Insert the instrument in the panel opening. Firmly fasten the instrument to the panel using
the nuts, bolts and lock washers.
SURFACE MOUNTING:
1) Install the mounting brackets,ordered separately,, on the vertical sides of
instrument housing. Use the brackets to fasten the instrument to the surface.
8
FIGURE 2-1
7/8" Dia hole for
wiring - 3 locations,
EC1, EC2, and EC3
EC1 EC3
13 3
16
(335 mm)
4
PLACES
EC2
21 (64 mm) 3 43
2
15 1 (384.2 mm)
8 11
WIDTH OF COVER 4
16
(119.1 mm)
2 19 (65.9 mm)
32
Mounting Bracket (2)
9
32 DIA.(7.1mm) Holes to mount
bracket to surface
5
12 8 71
2
(320.7 mm)
(190.5 mm) Screw that mounts to case
1 7
13 2 (342.5 mm) 32 (5.5 mm) Note: Surface Mount
13 15 ( 354 mm) Dimensions
16
Top edge of case
Panel cut-out for flush mounting
7 3/4"
196.9mm
2.5"
63.5mm
14 11/16"
373mm
Holes should be sized to accommodate screw type
needed to fasten instrument. This is up to the installer.
Before using any instrument near the devices listed, the instructions below should be
followed:
1. If the instrument is to be mounted in the same panel as any of the listed devices,
separate them by the largest distance possible. For maximum electrical noise
reduction, the noise generating devices should be mounted in a separate
enclosure.
2. If possible, eliminate mechanical contact relay(s) and replace with solid state
relays. If a mechanical relay being powered by an instrument output device
cannot be replaced, a solid state relay can be interposed to isolate the instrument. 9
(Continued on next page)
3. A separate isolation transformer to feed only instrumentation should be
considered. The transformer can isolate the instrument from noise found on the
AC power input.
4. If the instrument is being installed on existing equipment, the wiring in the area
should be checked to insure that good wiring practices have been followed.
For those devices that do not have surge suppressors supplied, RC (resistance-capacitance)
networks and/or MOV (metal oxide varistors) may be added.
Inductive Coils - MOV's are recommended for transient suppression in inductive soils con-
nected in parallel and as close as possible to the coil. See Figure 2-2. Additional protection
may be provided by adding an RC network across the MOV.
Contacts - Arcing may occur across contacts when the contact opens and closes. This results
in electrical noise as well as damage to the contacts. Connecting a RC network properly
sized can eliminate this arc.
For circuits up to 3 amps, a combination of a 47 ohm resistor and 0.1 microfarad capacitor
(1000 volts) is recommended. For circuits from 3 to 5 amps, connect 2 of these in parallel.
See Figure 2-3.
10
FIGURE 2-2
0.5
mfd
1000V
Coil
220
ohms
115V 1/4W
230V 1W
FIGURE 2-3
MOV
R C
Inductive
Load
11
2.5.2 SENSOR PLACEMENT (THERMOCOUPLE OR RTD)
If the temperature probe is to be subjected to corrosive or abrasive conditions, it should be
protected by the appropriate thermowell. The probe should be positioned to reflect true
process temperature:
To determine the temperature error resulting from the lead length resistance, use the following
equation:
TABLE 1
TABLE 2
Example
A recorder is to be located in a control rrom 660 feet away from the process. Using 16 AWG,
Type J thermocouple, how much error is induced?
Terr = 3.3 °F
12
RTD LEAD RESISTANCE
RTD lead length can affect instrument accuracy. Size (gauge) and length of the wire used
affects lead length resistance.
To determine the temperature error resulting from the lead length resistance, use the following
equation:
Terr = TLe * L where; TLe = value from Table 3 if 3 wire or Table 4 is 2 wire.
L = length of leadwire in thousands of feet.
Example
An application uses 2000 feet of 18 AWG copper lead wire for a 3-wire RTD sensor. What is
the worst-case error due to this leadwire length?
Terr = TLe * L
TLE = +/- .46 (°F/1000 ft) from Table 1
13
Wiring Connections 2.6
All wiring connections are typically made to the instrument at the time of installation. Connec-
tions are made at the terminal boards provided, two 12 gauge wires maximum. Terminal
boards are designated TB1 through TB13. See Figure 2-4 for the terminal board locations.
The number of terminal boards present on the instrument depend upon the model number/
hardware configuration.
FIGURE 2-4
1 2 1 2 3 4 1 2 3 4 1 2 3 4
RELAY/SSR Driver
Board
TB9
Processor Board
1 2 3 4
1 2
TB3 TB4 TB5 TB10 TB11 TB12 TB13
12 1 2 3 4 5 1 2 3 4 5 1 2 1 2 1 2 1 2
14
FIGURE 2-5
AC Instrument Power Input
Connect the 115 VAC hot and neutral to terminals 1 and 2 respectively of TB1. See Figure 2-4
(page 14) for Terminal Board locations on the instrument. Connect the 230 VAC one leg to
each terminal, be sure to check the position of the Voltage Selector switch provided with 230
VAC instruments. The switch position must match the voltage input to the instrument.
FIGURE 2-6
Thermocouple Inputs
Use TB4 for the Pen 1 input, and TB5 for the Pen 2 input. Connect the positive leg of the
thermocouple to terminal 1, and the negative to terminal 2. Be sure that the input conditioning
jumpers are properly positioned for a thermocouple input. See Appendix A-1 (page 62).
TB4 or TB5
+ -
1 2 3 4 5
Grounded or
Ungrounded
Thermocouples
may be used
FIGURE 2-7
RTD Inputs
Use TB4 for the Pen 1, and TB5 for the Pen 2 input. Connections are shown for 3 wire and 2
wire RTD inputs. If a three wire device is used, install the common legs to terminals 2 and 3. If
a two wire device is used, install a jumper between terminals 2 and 3. Be sure that the input
conditioning jumpers are properly positioned for an RTD input. See Appendix A-1 (page 62).
15
FIGURE 2-8
Volt, Millivolt and milliamp Input
Make the volt, millivolt and milliamp connections as shown below. Use TB4 for thePen 1 input,
and TB5 for the Pen 2 input. Terminal 1 is positive and terminal 2 is negative. The milliamp
input requires the installation of an appropriate shunt resistor (ordered separately) between
terminals 1 and 2. Be sure that input conditioning jumpers are in the correct positions for the
input being connected. See Appendix A-1 (page 62).
TB4 or TB5
+ -
1 2 3 4 5
SHIELDED
TWISTED
PAIR
+ - MAY BE
GROUNDED
SOURCE OR
UNGROUNDED
FIGURE 2-9
Remote Setpoint Input VDC , mADC
If Remote Setpoint option has been specified, make connections as shown. The remote
setpoint input may be selected as either 0 to 5 VDC or 1 to 5 VDC input in the Program mode
section. Make sure the configuration properly matches the input used. Use TB4 for Pen 1, and
TB5 for Pen 2 if Pen 2 is provided and specified with the Remote Setpoint option. Connect
the positive lead to terminal 4, and the negative lead to terminal 3 (Terminal 3 is the ground,
terminal 4 is the input, terminal 5 is 5 VDC.) If a 4 to 20 mADC remote setpoint is to be used,
the instrument remote setpoint input should be configured for 1 to 5VDC in the Program
mode, and a 250 ohm resistor should be installed across terminals 4 and 3. A 250 ohm
resistor is provided with the instrument, one per pen. The resistor(s) are shipped in the plastic
bag that is clipped to the inside of the instrument cover. This is the same bag that contains
the spare pen cartridge(s).
SHIELDED
TWISTED
PAIR SHIELDED
MULTI-CONDUCTOR
CABLE
- +
SOURCE 150 OHM
TO 10K OHM
POTENTIOMETER
16
FIGURE 2-10
Digital Communications Options
Connections are made as shown using TB2. Refer to the Protocol Manual, Form #2878 for
more details regarding the connections and how to use this option. This document is provided
only when this option has been specified. If the communications network continues on to other
instruments, connect the cable shields together, but not to the instrument. A terminating
resistor should be installed at the terminals of the last unit in the communications loop. If the
communications network ends at the instrument, the shield is not connected.
TB2
Serial A Serial B
1 2
FIGURE 2-11A
SPST Relay Output
Connections are made to relays A through F as shown. Terminal connections are made using
TB6 (Relay/SSR Driver A, B), TB7 (Relay/SSR Driver C, D) and TB8 (Relay/SSR Driver E, F).
HOT
NEU POWER
LOAD
5 AMPERES
1 2 3 4 MAXIMUM
N.O. C N.O. C AT 115 VAC
TB6 Relay A & B, Relay A Terminals 1 & 2
TB7 Relay C & D, Relay C Terminals 1 & 2
TB8 Relay E & F, Relay D Terminals 1 & 2
17
FIGURE 2-11B
SPDT Relay Output
HOT
NEU POWER
LOAD
5 AMPERES
1 2 3 MAXIMUM
N.O. C N.C. AT 115 VAC
TB6 Relay A
TB7 Relay B
FIGURE 2-12
SSR Driver Output
Connections are made to relays H through A as shown. Terminal connections are made using
TB9, TB8, etc. depending on the number of SSR Driver outputs specified.
1 2 3 4 + -
SSR
TB6 THRU TB9
FIGURE 2-13
Current Output
Connections are made to current outputs A thruough D as shown. Each current output is
programmable as either 4 to 20 mADC or 0 to 20 mADC. Each output must be assigned to the
desired function in the Program mode. Terminal connections are made using TB10 through
TB13 for current output A through D respectively. Connect positive lead (+) to terminal 1 and
the negative lead (-) to terminal 2. Each current output will operate up to a 650 ohms
maximum load.
1 2
SHIELDED
TWISTED
PAIR
+ -
650 OHMS
LOAD MAXIMUM
18
FIGURE 2-14
Transmitter Power Supply Input
If the isolated 24 VDC regulated transmitter power supply has been specified, the connections
should be made as shown. Connections are made using TB3, terminal 1 is positive and
terminal 2 is negative. The power supply is capable of providing the power needed by up to 2
transducers (40 mADC maximum).
SHIELDED
TWISTED
PAIRS
+ - + - + -
FIGURE 2-15
Position Proportioning Control Output
Position Proportioning control requires that two relays (or SSR Drivers) and the Position
Proportioning Auxiliary input be specified. On a dual pen instrument, either pen may be
configured with Position Proportioning control provided the outputs and auxiliary inputs have
been properly specified.
L2
OPEN
L
CLOSE C
TB4
4 or 5
TB6, TB7
or TB8 TB5
3 4
2 3
L1
19
Configuration 3.1
After completing installation and wiring of the instrument the configuration (set up) procedures
must be performed to prepare the instrument for operation on the intended application. The
procedures include selecting specific parameters, entering data and possible jumper position-
ing. Once properly configured the instrument will retain the user selections in memory so this
procedure need not be repeated unless required by changes in the application.
Parameter selections and data entry are made via the front keypad. To ease configuration
and operation, user entered data has been divided up into several sections referred to as
modes. Each mode contains a different type of data or may be used for specific operating
functions. For two pen instruments, some modes are common to both pens. These modes are
as follows:
Associated with each mode is a series of unique displays that are accessed via the front
keypad.
Prior to first time operation of the instrument, the configuration procedures for the Program
and Tune modes must be performed as applicable. Calibration and Test modes are not used
as part of the instrument configuration or operation. These are used for service and mainte-
nance functions and are discussed in Section 5.4 & 5.5 of this manual (page 46 - 54).
20
Shipped Configuration/Jumper Positioning 3.2
Each instrument is factory shipped ready to accept a thermocouple input on TB 4 and TB 5.
All parameters in each mode are set to default values. These defaults are shown in tabular
form after the description for each mode. Instrument AC power input is as specified in the
instrument model number and is shown on the ratings label. The 230 VAC option includes a
switch in the instrument for selecting either 230 VAC or115 VAC input power. If this feature is
provided, verify AC input and switch position before applying power to the instrument.
The sensor input jumpers JU4, JU5, JU6 and JU7 condition the sensor input signals
and must be used in conjunction with input type selections made in the Program mode.
(page 28).
Entry into any mode except the Control, Tune, Manual, Off and Enable modes, will cause the
output(s) to turn off and any process re-transmission value output(s) to be 0 %. Entry into the
Off mode will cause process re-transmission to remain active.
For a single pen instrument, parameters for each mode are displayed in the upper right
display. If the instrument being configured is a two pen model, a sequence of applicable
parameters will be displayed in the Pen 2 display after the Pen 1 parameters have been
reviewed and configured. After the Pen 2 parameters have been completed, parameters
common to both pens will be configured and displayed in the Pen 1 display.
The instrument is provided with a time out feature. If the instrument is in any mode (except
while executing a calibration or test procedure) and no keypad activity takes place for 30
seconds, the instrument will time out and exit the mode automatically. The display will become
the code for the respective mode. If a mode code is displayed for 5 seconds with no keypad
activity, then the time out will cause the instrument to proceed to either the Control or Off
mode, depending upon whether the mode entered was an operational mode (Tune, Manual,
Enable) or non operational mode (Test, Program, Cal).
21
3.4.1 POWER UP PROCEDURE
Step 1
Verify that all electrical connections have been properly made before applying power to the
instrument.
Step 3
If any error messages are displayed, refer to Section 5.6 (page 55) for a definition of the
error message and the required action.
22
3.5.2 KEYPAD CONTROLS
The keys on the keypad functions include:
AUTO/MAN: Used to: 1. In the Control mode to enter the Manual mode if proportional
The upper key is for output(s) selected
Pen 1, the lower key 2. In the Manual mode to enter the Control mode if proportional
AUTO
is for Pen 2 (if provided) output(s) selected
MAN
AUTO
MAN
23
FIGURE 3-1
Pen 1
AUTO/MAN SCROLL
Key Key
UP
Key
Pen 2
DOWN
AUTO/MAN
Key
Key
LAMP TEST
From the Off or Control modes, all display and status LEDs can be illuminated simultaneously
by depressing the UP and DOWN keys at the same time. Any defective LEDs will not light.
PEN RESYNCHRONIZATION
To resynchronize the pen(s), depress the UP and DOWN keys for 4 seconds, the display will
show P dn momentarily, the pen(s) are driven to the "Pen Home" position below the bottom
graduation on the chart while the unit resynchronizes their position. After about 14 seconds,
the pen(s) will return to their correct position. All other functions of the unit's present mode
continue to operate normally.
24
PROGRAM MODE FLOW CHART
Prog
inPS AL2
iCor diSP
out1 dPoS
o1uL Euu
o1LL EuL
out2 HyCo
o2uL HyAo
A B Scroll Key
Numeric Display -
Use arrow keys
to change value
Up Arrow Key
Down Arrow
25
B C
rSPL PoL
SPuL Cru
SPLL CrL
AtFr P1EC
Prnd P2EC
dFF PAEC
PFF rLyA
KEY
Actual Display Pout rLyb
ON On/Off Display -
OFF Use arrow keys
to turn on or off Pou rLyC
Scroll Key
C D
Numeric Display -
Use arrow keys
to change value
Up Arrow Key
Down Arrow
26
D E
rLyd Cobr
rLyE CoCr
rLyF Codr
rLyg Crt
rLyH PAPu
CurA C00
Com (Optional)
Curb CCon
CurC Cbs
Curd CAd1
CoAr CAd2
27
TABLE 3-1 PROGRAM MODE CONFIGURATION PROCEDURE
Press the SCROLL key until Prog is displayed. Press the DOWN key to enter the Program
mode. Pen 1 will be displayed in the upper display. To enter the Pen 1 parameter, press the
DOWN key. To enter the Pen 2 parameter, if provided, press the SCROLL key, then the
DOWN key. To enter the unit parameter, press the SCROLL key with either Pen 1 or Pen 2
displayed until unit is displayed, then press the DOWN key. Press the SCROLL key to
advance the display through the parameter codes and their values. Use the UP and DOWN
keys to adjust the values. After adjusting a parameter, press the SCROLL key to proceed to
the next parameter. After all selections have been made, press the UP key with a parameter
in the display (not a setting) to exit the mode. For two pen instruments the parameters and
values which are applicable to Pen 1 will appear in the upper display, then the parameters
and values which are applicable to Pen 2 will appear in the lower display. Then the unit
parameters and values will appear in the upper display.
For illustration purposes all available Program mode parameters have been listed. The
parameters that will appear on the specific instrument will depend upon the model number
(hardware configuration) of the instrument and on the parameter selections previously made.
For future reference, record the parameter selections for the application in the “Your Setting”
column and on the Software Reference Sheet in Appendix E (Page 72). Use the Software
Reference Sheet to program the instrument.
To prevent unauthorized changes to the Program mode, the mode can be disabled (turned
off) in the Enable mode.
28
DISPLAY AVAILABLE FACTORY YOUR
STEP DESCRIPTION CODES SETTINGS SETTING SETTING
9 Alarm 1 AL1 0 to 6 0
the Alarm Point 0 = None
is selected in the 1=Process Alarm-Direct
Tune mode 2=Process Alarm-Reverse
3=Deviation Alarm-Direct
4=Deviation Alarm-Reverse
5=Deviation Band Alarm-Open within band
6=Deviation Band Alarm-Closed within band
29
DISPLAY AVAILABLE FACTORY YOUR
STEP DESCRIPTION CODES SETTINGS SETTING SETTING
30
DISPLAY AVAILABLE FACTORY YOUR
STEP DESCRIPTION CODES SETTINGS SETTING SETTING
Pressing the SCROLL key with the PAEC parameter value displayed in the Pen 1 window will
advance the display of a single pen instrument to the unit parameters. Pressing the SCROLL
key with the PAEC parameter displayed in the Pen 1 window of a two pen instrument will
advance the display to be PEns in the Pen 2 window. The Pen 2 Program mode parameter
selections can be made now. Pressing the SCROLL key with the PAEC parameter value
displayed in the Pen 2 window will cause the display to advance to the unit parameters
displayed in the Pen 1 window.
UNIT PARAMETERS
31
DISPLAY AVAILABLE FACTORY YOUR
STEP DESCRIPTION CODES SETTINGS SETTING SETTING
53 Communication CCon 0 to 4 3
Configuration 0=Off
1=Monitor Mode (Read Only)
2=Normal Mode (Read and Write)
3=Total Access with Limit Checking
4=Total Access without Limit Checking
55 Communications CAd1 0 to 99 1
Address-Pen 1
Will not be seen if
Con=0
56 Communications CAd2 0 to 99 2
Address-Pen 2
(If provided)
Will not be seen if
Con=0
32
TUNE MODE FLOW CHART
tunE
SoP rSEt
PAL1 ArSt
dAL1 rAtE
bAL1 Ct1
PAL2 Ct2
dAL2 SEnS
bAL2 FoP
Key
Pb1 Actual Display
ON On/Off Display -
OFF Use arrow keys
Pb2 to turn on or off
Scroll Key
A
Numeric Display -
Use arrow keys
to change value
Up Arrow Key
Down Arrow
33
TABLE 3-2 TUNE MODE CONFIGURATION PROCEDURE
The Tune mode allows the entry, review or altering of the process control Tune adjustments
and alarm setting(s).
To enter the Tune mode, press and release the SCROLL key until tunE is displayed, then
press the DOWN key. Press the SCROLL key to advance the display through the parameters
and their values. Use the UP and DOWN keys to select (adjust) the values. Each time the
DOWN key is pressed while a parameter code is being displayed, such as SoP, the next
parameter code in the sequence will be displayed (PAL1).
After selecting a parameter, press the SCROLL key to proceed to the next parameter. Pen 1
selections will appear in the Pen 1 window and the Pen 2 ( if provided) selections will appear
in the Pen 2 window after the Pen 1 parameters have been displayed. After all selections
have been made, press the UP key with a parameter in the display (not a setting) to exit the
mode.
For illustration purposes, all available Tune mode parameters have been listed. The
parameters that will appear on the specfic instrument will depend upon the parameter
selections previously made in the Program mode.
For future reference, record the parameter selections for the application in the "Your Setting"
column and on the Software Reference Sheet in Appendix E (page 72). Use the Software
Reference Sheet to program the instrument.
To prevent unauthorized changes to the Tune mode, the mode can be disabled (turned off) in
the Enable mode.
The Tune mode is adjusted on-line. The instrument will react to changes as they are
made.
34
DISPLAY AVAILABLE FACTORY YOUR
STEP DESCRIPTION CODES SETTINGS SETTING SETTING
11 Automatic Reset ArSt 0.0 to 100.0 repeats 0.0
Integration per minute
(Will be seen if Pb1 or
Pb2 was shown)
35
TABLE 3-3 ENABLE MODE CONFIGURATION PROCEDURE
To enter the Enable mode, press the UP and DOWN keys while in CtrL or oFF modes. All the
display lamps will light. After 2 seconds, the display will show Cchg and the pen(s) will move
to and remain at a point above the top graduation on the chart. Continue to press the UP and
DOWN keys, after 2 additional seconds, the display will show P dn and the Pen(s) are driven
below the bottom graduation on the chart. After 6 more seconds, the display will show EnAb.
Release the keys and the display will change to EtSt. Press and release the DOWN key and
each mode to be enabled/disabled will be displayed. With the enable mode prompt for the
desired mode displayed press the SCROLL key to verify that the displayed mode is either on
(enabled) or oFF (disabled). Press the DOWN key to turn off the mode, press the UP key to
turn on the mode or press the SCROLL key to advance the display to the next enable mode
prompt. Use the "Your Setting" column in the table to record your settings.
A hardware jumper located on the Processor Board (See Appendix A-1, Page 62) can be
used to lock/unlock the Enable mode. When the jumper is moved to the locked position, entry
into the Enable mode is not possible until the jumper is moved to the unlocked position.
EnAb
EtSt ON
OFF
ECAL ON
Key OFF
Actual Display
EPro ON
OFF
ON On/Off Display -
OFF Use arrow keys
to turn on or off Etun ON
OFF
Scroll Key
ESbY ON
Numeric Display - OFF
Use arrow keys
to change value
ESPS ON
Up Arrow Key
OFF
Down Arrow
ESPC ON
36 OFF
Operation 4.1
4.1.1 OFF MODE
In the Off mode, the instrument control and alarm function (s) are turned off. Process
Retransmission signal(s) remain active. The chart rotation can be selected in the Program
mode to stop or continue to rotate when the instrument is in the Off mode. The pen(s) will
remain active. The Off mode can be entered by pressing and releasing the SCROLL key until
the display reads oFF, then pressing the DOWN key. The display will read oFF and then
current process variable at two second intervals. Entering the Off mode of a dual pen
instrument will cause both pens to enter the Off mode. The second pen display will be blank
as the upper display reads oFF and displays the process value for the second pen at the
same time as the upper display.
Dual Pen Instruments: The setpoint of each instrument can be reviewed separately. Pen 1:
To review and change the local setpoint for Pen 1, press the UP key. The green LED in the
upper display will light indicating that the setpoint value is displayed. The setpoint can now be
raised by pressing and holding the UP key or decreasing by pressing the DOWN key. Pen 2:
The setpoint for Pen 2 is viewed and changed in the same manner, except the DOWN key
must be pressed to access the Pen 2 setpoint. The setpoint will change slowly at first, then
faster as the key is held. If the setpoint will not increase, check the Program mode to see that
you are not trying increase the setpoint above the setpoint upper limit SPuL. If the setpoint
will not decrease, check the Program mode to see that you are not trying to decrease the
setpoint below the setpoint lower limit SPLL. If the setpoint will not increase or decrease,
check that the Setpoint change mode is On in the Enable mode.
37
Dual Pen Instruments: If the instrument is specified and provided with the Remote Setpoint
capability for either or both pens and the Remote Setpoint has been properly configured in the
Program mode for either or both pens, the Setpoint Select mode will be accessible, if enabled.
Press and release the SCROLL keys until SPS appears in the display, then press the DOWN
key. The upper display will light and the lower display will blank if pen 1 is selected for a
remote setpoint. If only pen 2 is selected as a remote setpoint, then the upper display will
blank and the lower display will light. If both pens are selected for remote setpoint, the lower
display is accessed by pressing the DOWN key while the upper display shows the setpoint
mode selected for that pen. The display(s) will show Loc if the local (keypad entered) setpoint
is being used or rSP if the remote setpoint is in use. Press the SCROLL key to select the
setpoint mode that will be used by the instrument. Press the UP key to exit the Setpoint Select
mode. If a problem develops with the remote setpoint signal for either pen, the upper display
will show rSEr, the remote setpoint error code.
Digital Communications: The setpoint can be adjusted from a supervisory computer system.
Dual pen instruments are capable of having independent setpoint adjustments for each pen.
SPS
KEY
Actual Display
Scroll Key
rSP
Numeric Display -
Use arrow keys
to change value
Up Arrow Key
Down Arrow
38
4.1.2.4 TIME PROPORTIONING CONTROL
Time Proportioning Control can be implemented on controllers provided with SPST relay or
SSR driver output(s). Time proportioning can be programed for output 1 and/or 2 for each
pen. Time proportioning control is accomplished by cycling the output On and Off when the
process value is within the proportional bandwidth selected at a prescribed time period. The
time period is selected in the Tune mode by adjusting Ct1 and/or Ct2. The On time is a
percentage of the Cycle Time.
Position proportioning control permits the use of PID control where the final control element is
a modulating device such as a motorized valve. In this form, each of the two required relays
or SSR Drivers will be used to control the valve. One output will be used to open the valve,
the other is used to close the valve. The slidewire feedback is used to provide a signal
relative to the valve armature position to the instrument.
As with the other proportioning control forms, the process input, tuning parameters and the
setpoint are used by the control algorithm to calculate the output % required to correct for the
deviation between setpoint and process.
Proportional Output 1 Action P1EC on error condition does not apply to position proportioning
control. In an error condition, the unit goes only to 0% output (fully closed).
With Position Proportioning control, it may be necessary to adjust the Sensitivity Tune SEnS
mode parameter to reduce or eliminate oscillations of the motor around setpoint. If oscillation
occurs, increase the SEnS value until the motor stops oscillating. If the differential between
the Open and Closed rotation is too large, then decrease the SEnS value. Also, for proper
Position Proportioning operation, it is necessary to specify the actuation time of the valve or
damper from full open to full closed. If the motor has a stroke duration of 60 seconds, change
the value in the Cycle Time parameter Ct1 to 60. This ensures that controller will move the
motor for the proper amount of time when making adjustments.
39
4.1.2.7 PROPORTIONAL OUTPUT PERCENTAGE DISPLAY
While in the Control mode, pressing the UP and the SCROLL keys at the same time will
cause the display to sequence through a series of display codes and values:
Each code and output value will be displayed only if the corresponding proportional output is
present. Each code or value will be displayed for 1 second. This sequence of displays will
continue until the SCROLL key is pressed, which will then return the display to the normal
mode. Displays will sequenced for both Pen 1 and 2 as applicable on dual pen instruments.
To enter the Manual mode, pres and release the AUTO/MAN key for the specific pen. If the
Standby mode is On in the Enable mode and a control output is selected for proportional
control, the instrument will enter the Manual mode. The Manual mode status LED will light for
that pen to indicate that the Manual mode is in use. Shifting from the Control to the Manual
mode is bumpless. The proportional output(s) will stay at the last value(s) calculated by the
control algorithm. Po1 will appear on the display if output 1 is a proportional output or Po2 if
output 1 is not a proportional control.
In order to vary a proportional output value, press and release the SCROLL key until the
display code for the output is displayed, Po1 or Po2. Press the SCROLL key again to see the
percentage of output value. Press the UP key to increase the output percentage value. Press
the DOWN key to decrease the output percentage value.
If no keys are pressed, the display will sequence through the following displays:
Proc will be displayed then current process value for that pen
In order to change a proportional value once the cyclic display begins, press and release the
UP or DOWN key then press the SCROLL key until the display code for the output displayed
is Po1 or Po2. Press the SCROLL key again to see the percentage of output value. Press
the UP key to increase the output percentage value. Press the DOWN key to decrease the
output percentage value.
To return to the Control mode of operation from the Manual mode, press the specified AUTO/
Man key. If the Automatic Transfer feature is selected in the Program mode, the instrument
will switch from Manual mode to the Control mode when process value reaches the setpoint
value.
40
STANDBY MODE FLOW CHART
Stby
KEY
Actual Display
Po1
ON On/Off Display -
OFF Use arrow keys
to turn on or off
Scroll Key
Po2
Numeric Display -
Use arrow keys
to change value
Up Arrow Key
Down Arrow
1. Process Alarm Direct - the alarm will be On if the process value is greater than the process
value selected.
2. Process Alarm Reverse - the alarm will be On if the process value is less than the process
value selected.
3. Deviation Alarm Direct - the alarm will be On if the process value is greater than the
setpoint plus the deviation value selected.
4. Deviation Alarm Reverse - the alarm will be On if the process value is less than the
setpoint plus the deviation value selected.
5. Deviation Band Alarm Open Within - the alarm will be On if the process value is greater
than one half the deviation band alarm values selected above or below the setpoint.
6. Deviation Band Alarm Close Within - the alarm will be On if the process value is less than
one half the deviation band value selected above or below the setpoint.
The alarms will be active while the instrument is in the Control mode. Relay and solid state
relay drivers can be assigned to provide output capability for the alarm functions.
The alarm value (Process deviation or bandwidth) is selected in the Tune mode.
Alarm outputs chatter can be reduced by using the hysteresis for the alarm output adjustable
in the Program mode to create a deadband around the alarm point.
41
Tune Mode Operation 4.3
Proportional output controllers may require the adjustment (tuning) of the PID and other
related parameters. This provides a means for the instrument's control algorithm to be
adjusted to meet specific application requirements.
2. After making any changes in tuning parameters, a disturbance should be introduced into
the process so that the process reaction may be observed. This process reaction, or
recovery, will tell whether the tuning parameters provide the desired control. It is usually
easiest to make a step change in setpoint to introduce this disturbance.
3. The change in setpoint or disturbance, referenced above should be large enough to cause
an observable deviation of process from setpoint. However, this change should not be so
large that it will cause the controller output to proceed to either extreme limit.
4. Controller tuning for optimal control is not hard and fast, BE PATIENT: The process will
take a certain amount of time to react to the setpoint changes during tuning. The amount
of time depends upon the specific process, however, a period of 8 to 12 minutes should
be allowed between changes. The important point to remember is to allow the process to
react completely, do not rush through tuning of the controller. If the complete process
reaction is not observed, optimum control may never be achieved.
5. Time Proportioning control output(s) require the cycle time to be adjusted for the
application. Short cycle times typically result in the most accurate process control, but will
cause the quickest "wear out" of any mechanical components.
6. Leave all other tuning parameters (except for the alarm settings) at the factory default
settings. Obtain the best possible process reaction by adjusting the Proportional Band
width parameter. The setting that achieves the best response for the process should be
left in the controller programming, and should be noted on the Software Reference Sheet
in Appendix E (page 72).
7. If there are to be no setpoint or load changes in the process, the Proportional Band
adjustment may be all that is necessary for proper control. If an offset still exists (the
process does not settle out at setpoint for the best possible proportional band adjustment),
Manual Reset may be added to eliminate this offset.
8. Auto Reset may be added to eliminate process offsets with respect to setpoint. Increase
Auto Reset from 0 to 0.2 increments. Start with a small amount. Increase this increment if
there is no apparent reaction. Remember to allow the process 8 to 12 minutes to react to
any changes.
9. If necessary, Rate may be added. Rate is a dynamic tuning parameter. Rate may be
required to compensate for load changes or to inhibit process oscillations that occur when
a large amount of auto reset (4 or 5 repeats/minute) is being used.
10. Controller tuning is not hard and fast. It may be necessary to adjust the tuning
parameters over a period of time to obtain optimal control of the process.
42
4.3.2 ZIEGLER NICHOLS TUNING METHOD
This procedure has been determined empirically to yield ideal 1/4 amptitude decay tuning
parameters that are determined by watching the system in a sustained oscillation (curve C,
page 44, the ultimate proportional band and ultimate time period) and then using these values
from this sustained oscillation to calculate ideal parameters.
To aid in determining the process oscillation, the instrument configuration parameters can be
adjusted. By reducing the chart upper value Cru and increasing the chart lower value CrL,
the resolution of the oscillation can e better observed. Also the chart rotation time Crt can be
reduced to improve the observation of the oscillation cycle time.
1. Set Manual Reset rSet to 0.0, set ArSt to 0.0 and set rAtE to 0.0.
3. Set the Proportional Band (PB) at 100 and upset the process and observe the response.
One easy method for imposing the upset is to move the setpoint for a few seconds and
then return it to its original value.
4. Achieve a response curve similar to the sustained oscillation (curve C), this is the Ultimate
Proportional Band (UPB) and Ultimate Time Period (UTP).
a) If the response curve from step 3 does not damp out, as in curve A from the
drawing, the PB is too low. The PB should be increased and step 3 repeated.
b) If the response in step 3 damps out, the PB is too high. The PB should be
decreased and step 3 repeated.
These values obtained for Ultimate Proportional Band (UPB) and Ultimate Time Period (UTP)
are used to calculate ideal P, PI, PD, PID tuning parameters using the following Ziegler-
Nichols equations:
Proportional plus automatic reset (PI) P (Pb) = 2.2 x UPB (degrees or units)
I (ArSt) = 1.2 / UTP (repeats per minute)
43
Period
Curve A : unstable
Curve B : stable
Curve C : continuously cycling, ultimate PB and period
44
Service 5.1
This section contains information regarding calibration and test procedures that can be
performed in the field as well as items concerning the normal maintenance of the instrument.
CAUTION: The chart flange assembly pin is sharp to perforate the chart. Use caution
while installing the chart to avoid coming into contact with the pin.
1. Depress and hold the UP and DOWN keys for between 2 and 3 seconds. Immediately
after depressing the keys, the unit will do a Lamp Test with all LED segments and indicators
lighted on the upper (and lower) display.
2. After 2-3 seconds, the display will show Cchg, the pen(s) will move to and remain at a
point above the top graduation on the chart and the chart will stop rotating. All other functions
remain active (Control, Display, etc).
3. Open the instrument door, snap up the chart hold down lever on the chart flange assembly,
gently lift the pen(s) and remove old chart.
4. With the pen(s) held up, install new chart. Be sure to line up the chart time line so that the
current time is aligned with the time setting mark on the chart platen.
5. Snap down chart hold down lever and close the instrument door.
6. Depress and hold the UP and DOWN keys for between 2 and 3 seconds. Immediately
after depressing the keys, the unit will do a Lamp Test with all LED segments and indicators
lighted on the upper (and lower) display.
7. After 2-3 seconds, the display will show Cchg momentarily, then the pen(s) position and
chart rotation will be restored.
FIGURE 5-1 2
Changing Pens
To install pen, slide pen into
holder (1) and push down (2)
as shown by arrows
1
2 1
45
Calibration 5.4
CAUTION: Do not attempt any calibrations without the proper test equipment that meets or
exceeds the specifications listed.
Press and release the SCROLL key until CAL appears on the display , then press the DOWN
key to enter the mode. The display will change to CAL1. Press the SCROLL key to advance
the display to the other calibration modes available. For two pen units, CAL5 will only need
to be required on TB4 to calibrate Pen 1 and Pen 2 inputs. Both TB 4 and TB 5 inputs must
be calibrated for thermocouple inputs. Table 5-1, page 47, provides a listing of field calibra-
tion routines. All instruments are calibrated prior to shipment from factory.
CAL Prog
CAL1
CAL2
CAL3
CAL4
KEY CAL5
Actual Display
ON On/Off Display -
Use arrow keys
CAL6
OFF
to turn on or off
Scroll Key
CAL7
Numeric Display -
Use arrow keys
to change value
Down Arrow
46
TABLE 5-1 CALIBRATION PROCEDURES
Calibration
Procedure Description
CAL 1 Reinitialization of program and tuning values.
CAL 4 Cold Junction Utility, displays temperature the cold junction compensator
is sensing. No adjustment is made with this procedure.
With CAL 1 displayed, while pressing the DOWN key, press the SCROLL key. The display
will momentarily blank while the initialization is in progress. Upon completion of the routine, a
Pen Home function is automatically performed; P.dn is displayed during this period, with the
decimal point flashing to indicate this mode. Once the pen(s) reach Pen Home, CAL1 will be
displayed.
With CAL 2 displayed, press and hold the DOWN key, then press the SCROLL key. Release
the keys when the instrument displays hLd1. Short the input terminals or apply 0.00, ±.01
mVDC to TB 4 terminals 1 and 2. Press the DOWN key and dELy will appear for 10 sec-
onds, then SCAn will appear for 10 seconds. A calibration reference number will then appear,
this should be 0 ± 50. If this number falls outside these limits, press the SCROLL key and
CAL 2 will be displayed. Perform the calibration again. Repeat the calibration until the number
falls within the tolerance limits. If the number remains outside the limits, check the connec-
tions and try the calibration again. If the number does not approach the tolerance limits
contact the nearest representative or the factory for assistance. With an acceptable reference
number on the display, remove the short and connect a 50.00, ±.01 mVDC source to the TB
4 terminals 1 and 2. Be sure to observe the proper polarity when connecting the source.
Terminal 1 is positive and terminal 2 is negative. Press the DOWN key, deLy will be dis-
played for 10 seconds and then SCAn for 10 seconds. When the calibration is complete CAL
2 will reappear.
(Continued on next page)
47
Error recovery:
See section 5.6 (page 55) for details. Ensure that the millivolt source is connected correctly
and functioning properly.
The calibration can be exited when hLd1 or the calibration reference number is displayed by
pressing the SCROLL key.
With CAL2 displayed, press and hold the DOWN ARROW key, then press the SCROLL key.
Release both keys and the instrument will display hLd1. Press and hold the UP ARROW key,
then press the SCROLL key. The display will momentarily blank and then CAL1 will be
displayed. Release both keys and depress the UP ARROW key. CAL will be displayed.
Operation:
With CAL 3 displayed press and hold the DOWN key, then press the SCROLL key . The
instrument will display hoLd. Release both keys. Press the DOWN key , deLy will be dis-
played for up to 10 seconds, then SCAn for 10 seconds. If SCAn remains in the display for
much longer than ten seconds, refer to the note below. The instrument will compute and
display the cold junction temperature to the nearest tenth of a degree C. Compare the display
reading with thermometer and use the UP and DOWN keys to correct the reading. To exit,
press the SCROLL key and CAL 3 will be displayed again.
NOTE: If the display stays in SCAn for more than 10 seconds press the SCROLL key.
CAL3 should be displayed. With CAL3 displayed, while pressing the DOWN key, press the
SCROLL key. The instrument will display hoLd. Press the UP key. The instrument will
begin the calibration procedure with a default value and proceed to dELy as described
above.
Error recovery:
See section 5.6 (page 55) for details on specific errors. The calibration can be exited if hoLd
or the CJC temperature is displayed by pressing the SCROLL Key.
With CAL 4 displayed, press and hold the DOWN key, then press the SCROLL key . Release
both keys and SCAn will be displayed for 10 seconds while the instrument senses the CJC
temperature. The result will then be displayed to a tenth of a degree C. The input terminals
must be shorted. CAL 3 must first be performed. The displayed temperature is not the
ambient temperature. It is the temperature of the CJC . To exit, press the SCROLL key
and CAL 4 will be displayed.
48
5.4.5 CAL 5 RTD INPUT
This procedure determines and saves calibration values relating to RTD inputs. This calibra-
tion must be preceded by CAL 2 to properly calibrate the instrument. Both RTD inputs must
be calibrated and both inputs must have valid inputs during the calibration. Decade boxes
with .01% resolution or equivalent are required. Make sure that the Processor board
jumpers JU4, JU6 and JU5, JU7 are in the proper positions. See Appendix A-1 (page
62).
With CAL 5 displayed, press and hold the DOWN key and then press the SCROLL key. The
display will now be PEn1 to indicate that the instrument is set to calibrate the RTD input on
TB 4 . Press the DOWN key to change the display to PEn2 to calibrate the RTD input at TB
5. Choose the input TB to be calibrated and press the SCROLL key, hLd1 will then be
displayed. Connect the decade box at 100 ohm setting across the input terminals 1 and 2 and
a jumper wire from terminal 2 to 3. Press the DOWN key and dELy will be displayed for 10
seconds, then SCAn for ten seconds. When hLd2 is displayed, change the decade box
setting to 277 ohms to the input terminals (do not disturb the wiring) and press the DOWN
key. The display will change to dELy for 10 seconds, followed by SCAn for ten more
seconds. CAL 5 will be displayed after the calibration is completed.
Error recovery:
See section 5.6 (page 55) for details on specific errors.
The calibration can be exited when the instrument displays hLd1 by pressing the SCROLL
key.
With CAL 6 displayed, press and hold the DOWN key, then press the SCROLL key. The
instrument will display C6 and the number of the mode in effect. Press the UP or DOWN key
to change the mode selection, indicated by the number to the right of the C6. Pressing the
SCROLL key will exit the calibration with the last mode number displayed in effect.
The selected mode will remain in effect if power is interrupted. To return the instrument to
normal operation, CAL 6 must be exited, with mode zero selected, or CAL 1 must be exe-
cuted to initialize all parameters.
49
5.4.7 CAL 9 PEN CALIBRATION
This procedure is used to calibrate the pen(s). No special test equipment required.
Valid inputs must be connected to TB 4 and TB 5 before performing this calibration. With CAL
9 displayed, push and hold the DOWN key, then press the SCROLL key . Release both keys
and the display will indicate PEn1. For 2 Pen instruments, press the DOWN key to toggle the
display between pen 1 and pen 2. With desired pen displayed, press the SCROLL key.
Pen Arc - To adjust the pen arc, the UP and DOWN keys are pressed simultaneously to draw
an arc for visual reference. The display will show P uP while the selected pen moved upscale
to the top stop position. The pen then moves back downscale to nominally place the pen at
the outer ring (100%) of the chart. The display will now display ArC. The pen arm may be
loosened and the length adjusted to a position and angle to place the pen tip near the outer
ring then re-tighten the pen arm. Press any key and the pen will drive to "HOME", then back
to the inner ring (0%) and PEnL will be displayed.
Zero and Span - To begin a pen zero, with PEnL in the display, the UP or DOWN key may be
pressed to move the pen to the low end of the chart span (0%). When adjusted, press the
SCROLL key and P uP is displayed while the pen is moved to the outer edge (100%) of the
chart, then PEnh is displayed. To adjust pen span, use the UP and DOWN keys to adjust the
pen for the high end (100%) of the chart. When adjusted, press the SCROLL key and P dn
will be displayed for about 5 seconds while the pen is moved to the 50% point on the chart,
then PEn1 is displayed. Calibration may now be performed again to verify the accuracy, or
press the UP key twice to exit back to the CAL mode.
50
Test Mode 5.5
To enter the Test mode, press and release the SCROLL key until tESt appears on the
display then press the DOWN key. tSt1 will be displayed, press and release the SCROLL key
to advance the display to the desired test. Tests 1, 2 and 3 are performed as a unit so the
display will advance directly to tSt4 from tSt1. Listed in Table 5-2, page 52, are the test
procedures available. Test 1, 2 and 3 are performed on start up, periodically during operation,
and on entry into the Test mode. Test 4 is executed on entry into and periodically during the
Operation mode. These tests can be used as a trouble shooting aid.
tESt
tSt1
tSt4
tSt5
tSt6
tSt7
KEY
Actual Display
tSt8
ON On/Off Display -
OFF Use arrow keys
to turn on or off
tSt9
Scroll Key
Numeric Display -
tStA Use arrow keys
to change value
Up Arrow Key
Down Arrow
51
TABLE 5-2 TEST PROCEDURES AND DESCRIPTION
Test Description
Test 1 Microprocessor internal RAM test. Used to check the processor RAM to
make sure it is functioning correctly.
Test 2 External RAM test, used to test the RAM chip for proper function.
Test 3 EPROM checksum test, used to check that the EPROM program is correct.
Test 4 External RAM checksum test; instrument test and identifies how many
times Errors16 or 17 have occurred.
Test 5 Verifies that all keys are functional and all LEDs are working.
Test 6 Used to verify that all relays and solid state relay driver outputs are working
correctly.
Test 7 This procedure will allow operator to adjust the current output value to
check the output and to test the operation of the external device.
With tSt1 displayed, press and hold the DOWN key, then press the SCROLL key. tSt1 will be
displayed momentarily while the test is in progress. Upon successful completion, the instru-
ment will proceed to Test 2.
After completion of Test 1, tSt2 will be displayed momentarily while the test is in progress.
Upon successful completion, the instrument will proceed to Test 3.
After completion of Test 2, tSt3 will be displayed momentarily while the test is in progress.
Upon successful completion of Test 3 tst1 will be displayed.
With tst4 displayed, press and hold the DOWN key then press the SCROLL key. The display
will blank momentarily, then momentarily display two numbers, and then tst4 will be dis-
played. These numbers indicate the number of times Error 16 and 17 have occurred respec-
tively. Test 4 can be executed again, or another test may be selected.
52
5.5.5 TEST 5 KEYPAD/DISPLAY TEST
This test allows the operator to verify that the keys work and that all display elements can be
lighted. No special test equipment is required.
With tSt5 displayed, press and hold the DOWN key, then press the SCROLL key and then
release both keys. The display will go blank. The following code will be displayed while the
corresponding key is pressed:
Key Display
SCROLL SCrL
UP uAro
DOWN dAro
AUTO/MAN (Pen 1) Autu
AUTO/MAN (Pen 2) AutL
UP and DOWN (All LED’s and segments lit)
SCROLL and UP (exit)
To exit, press the SCROLL and UP keys simultaneously, tSt5 will be displayed.
With tSt6 displayed, press and hold the DOWN key, then press the SCROLL key. oFF will be
displayed. For SPST Relay outputs, connect the meter across the relay output in the ohm
scale. For SPDT Relay outputs, connect the meter across the N.O. and COM output in the
ohm scale. The meter should read continuity with the relay On and infinity when the relay is
Off. For SSR Driver outputs, connect the meter across the output in the Volt DC scale. The
meter should read 5 VDC when the SSR driver is On and 0 VDC when the driver is Off. Press
and release the DOWN key to advance through the following sequence:
RELAY ON RELAY ON
DISPLAY SPST SPDT
rLYA A A
rLYb B B
rLYC C C
rLYd D D
rLYE E None
rLYF F None
rLYg G G
rLYH H H
oFF None None
To exit, press the SCROLL key and tSt6 will be displayed. The existence of Relay and
Solid State Relay Driver output(s) is dependent upon the hardware configuration.
With tSt7 displayed press and hold the DOWN key, then press the SCROLL key. The display
will indicate CurA, for the first current output. By pressing the DOWN key, the instrument will
advance through the selection sequence of Curb, CurC, Curd, and then back to CurA.
Choose the desired output to be tested and connect the milliamp meter across the output
terminals being tested. Be sure to observe the proper polarity when connecting the meter.
Terminal 1 is positive and terminal 2 is negative. Press the SCROLL key and the display will
indicate 4 milliamps. Use the UP and/or DOWN key to increase or decrease the current
output in 1 mADC steps from 0 to 20 mADC. (Continued on next page)
53
The current output reading should be ± 0.1 mADC at any output value. A ± 5 % of span
adjustment for the current output(s) is provided by using the potentiometer adjacent to the
current output on the Current Output board. See Appendix A-4 (page 65). To exit the test,
press the SCROLL key and tSt7 will be displayed. The existence of a mADC current
output is dependent upon the hardware configuration.
With tSt8 displayed press and hold the DOWN key, then press the SCROLL key. The display
will show ALL while the test is in progress. This test functions automatically. The chart speed
will be at maximum, 72 seconds per revolution. Starting at the current pen position, pen 1 will
move (sweep) upscale for about 5 seconds, then downscale for about 5 seconds. Pen 2 (if
provided) will sweep in the opposite direction. The pens will continue to cycle until the test is
exited. To exit, press the SCROLL key. The display will show tSt8.
Auxiliary inputs are the remote setpoint and position proportioning options. With tSt9 dis-
played, press and hold the DOWN key, then press the SCROLL key. The display will show
Ai1, Auxiliary Input 1. By pressing the DOWN key, the instrument will sequence through the
selections for testing: PF1, Ai2, PF2, then back to Ai1 (PF=Pen Feedback). With the desired
input selected, press the SCROLL key. The corresponding value will be displayed. For
software revision R2.99 and below only, the pen position feedback input voltage will be
displayed to the nearest tenth of a millivolt. Manually move the pen up and down scale and
verify that the displayed reading changes from about 0 to 50 millivolts. For software revision
R3.00 and above, the pen position feedback input voltages are disabled and will display a
fixed value of zero. To terminate a particular test, press the SCROLL key. The display will
show tSt9.
With tStA displayed, press and hold the DOWN key then press the SCROLL key. The display
will indicate SEnd. Press the DOWN key to toggle the display between SEnd and rEC . With
the desired function selected, press the SCROLL key. In the SEnd (send or transmit) mode,
the instrument will repeat the following sequence. First the transmitter will output a logic 1 on
the line for one second. Next the transmitter will change the logic level to 0 for one second.
Then the transmitter will disable for one second. In the rEC (receive) mode, the transmitter is
continuously disabled. In either mode, the instrument will monitor the line and display rEC1
when a logic 1 is on the line or rEC0 when a logic 0 is on the line . In the SEnd mode, the
instrument will display rEC when the transmitter is disabled. To perform an internal test and
verify the basic operation of the hardware, place the instrument in the Send mode and verify
that the display cycles through rEC1, rEC0, and rEC. To verify that the transmitter functions
properly, two LED’s, each with a current limiting resistor, can be connected with opposite
polarity observed to the TB 2 terminals 1 and 2. The following three states should be ob-
served: one LED On, then the other LED On, then both Off. Alternately, a load resistor can be
placed on the terminals, and observe that the voltage generated across the load resistor is as
follows: >+3 VDC, then >-3 VDC, and then 0 VDC.
Another method, to test the communications hardware, would be to connect one or more
instruments in the Receive mode to an instrument in the Send mode. All of the instruments in
the Receive mode should have their displays alternating in sync with the instrument in the
Send mode. When the sending instrument displays rEC, the receiving instruments should
54 display rEC1. To terminate the test, press the SCROLL key for one second. Upon exit, tStA
will be displayed. The existence of the communications hardware is dependent on the
hardware configuration.
Trouble-shooting and Diagnostics 5.6
The Trouble-shooting Guidelines Section consists of two columns. The first column is a list of
some possible instrument conditions. The second column is a list of steps that should improve
the condition. The steps should be performed in order until the condition improves or all the
steps have been completed. If the instrument condition has not improved, contact the nearest
representative or the factory for assistance.
Trouble-shooting should be performed by qualified personnel using the proper equipment and
following all safety precautions. Whenever possible the trouble-shooting should be accom-
plished with the electrical power disconnected. The instrument contains static sensitive
components so care should be taken to observe anti-static procedures.
55
Model Number Displayed 1. Turn off the instrument power, wait 5 seconds then
during power up is re-apply the power. Verify that the number displayed
incorrect during the power up sequence is the same as
indicated on the label affixed to the platen. If the
number displayed is incorrect, perform the following
steps:
a. Turn off the power to the instrument. Press
and hold the UP and DOWN keys. Turn on
the power and keep the keys pressed until
the model number displayed resets to 72X0-
0000-0X21 where the digits X are determined
by the firmware installed. Release the keys
and turn off the power.
b. To enter the correct model number, press
and hold the SCROLL and DOWN keys and
turn on the instrument power. 72X0 should
be displayed. Wait about 5 seconds and
release the keys. The display should stay
72X0. Use the UP/DOWN keys as necessary
to change the displayed number to match
the first 4 digits of the model number as
indicated on the label on the platen. After
adjusting the first 4 digits to the proper
values, press the SCROLL key and the
display will change to 0000. Use the UP/
DOWN keys to set the next 4 digits of the
model number to the correct values. Press
the SCROLL key and the display will be
0X21. Use the UP/DOWN keys as necessary
to adjust the last 4 digits of the number.
Press the SCROLL key and the power up
sequence will complete. The Program and
Tune mode parameters will need to be
re-entered (pages 28 thru 35 or Software Ref.
Sheet, page 72, if already filled out).
56
5. If the output appears not to turn off remove the power
to the instrument. Open the cover and loosen the
platen hold down screw. Swing the platen open. Clip
the resistor located on the Relay Board adjacent to
the output(s) that seem to stay on (See Appendix A-2,
page 63). A .01 microfarrad, 1 KV capacitor should
be connected from the terminal listed below, for the
output where the resistor indicated was removed, to
the AC ground.
Relay A R1 TB 6 Terminal 1
Relay B R2 TB 6 Terminal 3
Relay C R3 TB 7 Terminal 1
Relay D R4 TB 7 Terminal 3
Relay E R5 TB 8 Terminal 1
Relay F R6 TB 8 Terminal 3
Close the platen and tighten the hold down screw. Close
the cover and turn the power on to the instrument.
Check the operation of the output(s).
Pen(s) do not operate 1. Verify that the Program mode parameters are properly
set (page 28 or Software Ref. Sheet, page 72, if
already filled out)
2. If the process value in the display exceeds the chart
upper or the chart lower values selected in the
Program mode, the pen will appear stuck at the chart
upper or lower edge respectively. Re-configure values
as needed by the application (May require a new
chart).
3. Perform Test 8 as described in the Test section of the
manual (page 54). If the pen(s) operate in Test 8,
perform the pen calibration, Cal 9, as described in the
Calibration section of the manual (page 50). Return
the instrument to the oPEr and check the operation of
the pen(s).
(Continued on next page)
57
4. For software revision R2.99 and below, perform Test
9 as described in the Test Section of the manual
(page 54). If the pen feedback voltage does not vary,
check the pen Potentiometer Segment board for
proper ribbon cable connection to the Processor
board (Appendix A-1, page 62) and that the pen
position fingers are making contact with Potentiometer
Segment board.
5. For software revision R2.99 and below, inspect to see
that the Potentiometer Segment of the pen feed back
is clean. Use a non-residue type cleaner that does
not affect the plastic to clean the segment if
necessary.
6. Verify the instrument is not in Change Chart function.
SnSr - Sensor Break or 1. Inspect the sensor for proper operation and
Over Range connection to the instrument. Acceptable sensor
ranges for the instrument are listed in the
Specifications section of Appendix D (page 69).
2. Verify that the Program mode input selection matches
the sensor input connected (page 28 or Software Ref.
Sheet, page 72, if already filled out).
3. Check that the input conditioning jumpers on the
Processor board (Appendix A-1, page 62) are in the
proper position for the sensor input. Check that the
jumpers are in the proper position for the Terminal
Board where the sensor is connected.
4. Perform the calibration procedure(s), as described in
the Calibration section (page 46), for the sensor input
type.
Hi - Input more than 10% 1. Perform the steps listed for the SnSr Error Condition.
Over Span (above)
Lo - Input more than 10% 1. Perform the steps listed for the SnSr Error Condition.
Under Span (above)
Er 1 - Microprocessor RAM 1. Turn off the power to the instrument. Wait 5 seconds,
Failure and turn the power on.
2. Turn off the power to the instrument. Open the
cover, and loosen the platen hold down screw. Swing
open the platen and inspect that the microprocessor
chip is properly seated in the socket located on
the Processor board (Appendix A-1, page 62). Close
the platen and tighten the screw. Close the cover and
turn on the power.
Er 2 - External RAM 1. Turn off the power to the instrument. Wait 5 seconds,
Failure and turn the power on.
Er 4 - RTD Mismatch Error 1. Check the connections to the instrument for the RTD
Input Calibration (CAL5) as described in the
Calibration section (page 49). Repeat the RTD Input
Calibration.
Er 5 - No Zero Crossings 1. Turn off the power to the instrument. Wait 5 seconds,
Detected and turn the power on.
2. Connect the instrument to another AC power source.
Er 6 - AC line below 43 HZ 1. Turn off the power to the instrument. Wait 5 seconds,
and turn the power on.
2. Connect the instrument to another AC power source.
Er 7 - AC line over 64 HZ 1. Turn off the power to the instrument. Wait 5 seconds,
and turn the power on.
2. Connect the instrument to another AC power source.
59
Er 9 - ADC Reference 1. Perform the CAL2 procedure as described in the
Number Error Calibration section (page 47).
Er12 - CAL2 Voltage Error 1. Check that 50 mVDC is properly connected to the
instrument and is within the tolerance limits as
indicated in the CAL2 procedure of the Calibration
section (page 47).
2. Turn off the power to the instrument and open the
cover. Loosen the platen hold down screw and swing
the platen open. Inspect the Processor board
(Appendix A-1, page 62) to insure that the input
conditioning jumper JU4 is in the non-volt position.
3. Perform the CAL2 procedure as described in the
Calibration section (page 47).
Er13 - RTD CAL5 Input 1. Check that resistance device is of the correct value
Error and properly connected to the instrument and is
within the tolerance limits as indicated in the CAL5
procedure of the Calibration section (page 49).
2. Turn off the power to the instrument and open the
cover. Loosen the platen hold down screw and swing
the platen open. Inspect the Processor board
(Appendix A-1, page 62) to insure that the input
conditioning jumpers are in the correct position for the
RTD input for the Terminal Board being calibrated.
For TB 4, JU4 should be in the non-volt position and
JU6 in the RTD position. For TB 5, JU5 should be in
the non-volt position and JU7 in the RTD position.
3. Perform the CAL5 procedure as described in the
Calibration section (page 49).
Er16 - Program/ASEt Mode 1. Record all Program and Tune mode parameters.
Checksum Error Perform CAL 1 procedure as described in the
Calibration section (page 47). Re-enter the
Program and ASEt mode parameters.
Er21 - Pen Feedback 1. Perform the steps listed for pens not moving.
Error
60
Er36 - Incorrect Crystal 1. Turn off the power to the instrument, wait 5 seconds,
For Digital Comm. then turn the power on.
Er37 - Incorrect Micro. 1. Turn off the power to the instrument wait 5 seconds,
For Digital Comm. then turn the power on.
Er38 - Incorrect RAM 1. Turn off the power to the instrument, wait 5 seconds,
for profiler then turn the power on.
2. Re-configure for non-profiler.
61
Appendix A
Board Layouts
Top
SWI for Rev. Y and above SWI for Rev. X and below
JU1
ENABLE MODE
230
TB1
UNLOCKED LOCKED
230
JU3
EPROM SIZE
230/115 VAC
32K 64K SWITCH 230 VAC
MODEL ONLY
U5 JU3
JU4 PEN 1 INPUT U7
T/C,mV, RTD
VOLT/mA
JU1
T/C,mV, RTD
Battery JU6
VOLT/mA JU5 JU7
JU4
TB2
JU6 /JU7
RTD
INPUT
NON- RTD
RTD
62
FIGURE A-2 - SPST RELAY/SSR DRIVER OUTPUT BOARD
C1 C2 C3 C4 C5 C6
K1 K2 K3 K4 K5 K6
P7
TB9
(rlyG) (rlyH)
63
FIGURE A-3 - SPDT RELAY/SSR DRIVER OUTPUT BOARD
C1 C2 C3 C4 C5 C6
K1 K2 K4
K3
(rlyA) (rlyB)
(rlyC) (rlyD)
P7
TB9
(rlyG) (rlyH)
64
FIGURE A-4 - CURRENT OUTPUT BOARD
U1
U3 U5 U7 U9
If this option board was ordered, you will find it located in the lower right hand corner of the
instrument.
65
Appendix B
Glossary
Automatic Reset (Integral)
This parameter is used so that the instrument will compensate for process variable deviations
from setpoint that occur when the process load characteristics change. Instructions for
determining the automatic reset settings are given in Table 3-2 (Page 34). Factory default
is.0.0. Display code ArSt.
Automatic Transfer
This feature, if configured, allows manual control of the process until setpoint is reached, at
which point the controller automatically transfers from manual to automatic control. Factory
default value is 0 = no auto transfer. Display code AtFr.
Bumpless Transfer
This feature prevents step changes in proportional outputs when changing from automatic to
manual control only.
Cycle Time
This Tune mode parameter is used to select the on/off cycle time for time proportioning
outputs (Ct1 for Output 1 and/or Ct2 for Output 2).
When usng the Position Proportioning option, Ct1 must be selected for the stroke time of the
motor.
Hysteresis
This parameter is adjustable from 0 to 300 units representing the width of the band (half
above and half below setpoint). Used with ON/OFF or Alarm outputs to reduce cycling. For
instance, with a value of 4 and a setpoint of 70 the output will turn ON when the process
variable drops to 68 and stay ON until 72 is reached, then turn OFF the output. Factory
default = 3. Display code HyAo for Alarm outputs. Display code HyCo for ON/OFF Control
outputs.
Input Correction
This parameter is adjustable from -300 to 300 units and is used as a method to compensate
for a linear sensor error. Factory default is 0 = no correction. Display code iCi1, iCi2.
Manual Reset
This parameter is adjustable from -1500 to 1500 units representing a manual shift of propor-
tional band(s) relative to the normal position which is an even strattle of the setpoint. Manual
reset is intended to be used when automatic reset is not used to allow compensation for
deviations from setpoint which remain after the process has stabilized. Factory default is 0.
Increasing the value increases the process variable, i.e. if the process variable stabilizes too
low, increase the manual set. Integral action, and conversely reset-windup inhibit apply over
66 the same process value range regardless of the manual reset value. Display code rSEt.
Pen Action on Power Up
This parameter specifies whether the pen, on a power-up will drive to the "Home Position"
(center of chart), then return to its correct postion. This is done as a cal check. Settings are
0=go to "home" and 1=remain in last position prior to power down. Default is 0.
Platen
The flat surface in the instrument upon which the chart rotates.
Process Rounding
This parameter is adjustable from 0 to 100 units and is used to round the process value to the
nearest value specified. This is for display only and does not affect the recorded value or
control action. Intended for use where the engineering units span is large, to reduce display
fluctuation. Factory default is 0 = no rounding. Display code Prnd. (e.g. Prnd = 3, Process
Value = -6, -3, 0, 3, 6, 9....)
Rate (Derivative)
This parameter is adjustable from 0.0 to 10.0 minutes and specifies how the control action
responds to the rate of change in the process variable. For example, if the process variable is
rising rapidly to setpoint, power is turned off sooner than it would be if the rise were slow. In
effect, derivative action anticipates lags within the system and shifts the proportioning band by
an amount determined by the rate of change of the input sensor.
Magnitude of the shift is determined by a derivative time constant. If the time constant is, say,
.1 minute (6 seconds), for every unit per second rate of change of the process variable at the
sensor, the proportiong band is moved 7 units in the direction that helps control. Likewise, if
the time constant is 1 minute (60 seconds), for every unit per second rate of change of the
process variable at the sensor, the proportioning band is moved 60 units in the direction that
helps control. Factory default is 0.0 Display code rAtE.
CSA Approved
*Relay (SPST) Outputs
4 115VAC Input
0 None
5 115/230VAC Input
1 One
2 Two
4 Four
Enclosure Options
6 Six 2 Standard Cover
7 One SPDT (Plastic Windows)
8 Two SPDT 4 Door Lock**
9 Two SPDT and Two SPST 6 Sealed Conduit Conn.
7 Combination of 4& 6**
*SSR Driver Outputs
Communications
0 None
1 One 0 None
2 Two 2 RS-485 Total Access
4 Four
6 Six Pen 2 Auxiliary Input
8 Eight 0 None
1 Position Proportioning
4 to 20mA Outputs 2 Remote Setpoint
0 None
1 One Pen 1 Auxiliary Input
2 Two 0 None
3 Three 1 Position Proportioning
4 Four 2 Remote Setpoint
*Total quantity of SPST Relays and SSR Drivers must be less than or equal to (8) eight.
**This option comes with a structural foam cover.
68
Appendix D Product Specifications
Measurement Error Limit • Type J,K,T,E,N,C T/C’s and RTD +/-0.25% of reading
plus 1 degree @ 25 degrees C
• Type R,S, B T/C’s +/-0.25% of span @ 25 degrees C
• mA, mV and VDC +/-0.25% of scaled span plus 1 Whole
Digit @ 25 degrees C
Ambient Temperature Error 0.01% of span per degree C deviation from 25 degrees C
Display Decimal Positions One, two or three decimal places (0.1 or 1 degrees for
T/C or RTD)
Sensor Fault Detection Displays SnSr for sensor or transmitter break. Outputs go
off. PV Out to 100%. Fault detection is not functional for
0-5V or 0-20mA inputs
69
Input Specifications
THERMOCOUPLE
TYPE RANGE TYPE RANGE
J 0 to 760C E 0 to 750C
0 to 1400F 0 to 1400F
S 200 to 1650C
400 to 3000F
CONTROL ADJUSTMENTS
Proportional Band 1 to 3000 units
Position Proportioning
Sensitivity 0.0 to 50.0%
CONTROL OUTPUTS
Relay SPST 115 VAC: 5.0A Resistive, 1/8HP or 250VA
230 VAC: 2.5A Resistive, 1/8HP or 250VA
SSR Driver Open collector output
Short circuit protected @ 100mA maximum
Provides 4VDC at 20mA or 3VDC at 40mA
mADC Current 4-20 or 0-20 mADC into 650 ohms maximum load
DISPLAY
Digital Display Two possible; one per installed pen. Red LED’s 0.56" high.
Status Indicators LED indicators for Output 1, Output 2, Manual Operation, Alarm,
Setpoint, minus sign and process value units (°C, °F, or E).
RECORD
Chart 10 in circular chart; 100 charts furnished with each instrument if
standard range
Chart Range -9999 to 9999 °/ units
Chart Drive DC stepper motor
Chart Rotation User configurable from 0.1 to 999.9 hours per revolution
Pen Type Disposable Fiber-tip
Pen Color Pen 1-Red, Pen 2-Green
Pen Response Time < 9 seconds over chart span
Accuracy ±1.0% of chart span max. error from displayed value
Chart Rotation Accuracy ± 0.5% of Chart Rotation Time
DIGITAL COMMUNICATIONS
Type RS-422/485 serial communications port. Half-duplex bi-directional
communications.
Character Format ASCII
Protocol Per ANSI X3.28 subcategories 2.5 & A4
Configuration User configurable to Monitor (read only) or Normal (read and write)
Bit Rate User configurable to 300, 600, 1200, 2400, 4800, or 9600 bits per
second
Address User configurable for each pen; 0 to 99
71
Appendix E
Software Reference/Record Sheet
PROGRAM MODE TUNE MODE
Pen 1 Pen 2 unit Pen 1 Pen 2
inPS rLyA SoP
iCor rLyb PAL1
out1 rLyC dAL1
o1uL rLyd bAL1
o1LL rLyE PAL2
out2 rLyF dAL2
o2uL rLyg bAL2
o2LL rLyh Pb1
AL1 CurA Pb2
AL2 Curb rSEt
diSP CurC ArSt
dPoS Curd rAtE
Euu CoAr Ct1
EuL Cobr Ct2
HyCo CoCr SEnS
HyAo Codr FoP
rSP Crt
rSPu PAPu
rSPL Coo
SPuL CCon
SPLL CbS
AtFr CAd1
Prnd CAd2
dFF ENABLE MODE
PFF EtSt
Pout ECAL
Pou EPro
PoL Etun
Cru ESby
CrL ESPS
P1EC ESPC
P2EC
PAEC
72
Warranty and Return Statement
These products are sold by The Partlow-West Company under the warranties set forth in the
following paragraphs. Such warranties are extended only with respect to a purchase of these
products, as new merchandise, directly from The Partlow-West Company or from a Partlow-
West Company distributor, representative or reseller, and are extended only to the first buyer
thereof who purchases them other than for the purpose of resale.
Warranty
These products are warranted to be free from functional defects in materials and workman-
ship at the time the products leave The Partlow-West Company factory and to conform at that
time to the specifications set forth in the relevant Partlow-West instruction manual or manuals,
sheet or sheets, for such products for a period of three years.
Limitations
The Partlow-West Company shall not be liable for any incidental damages, consequential
damages, special damages, or any other damages, costs or expenses excepting only the cost
or expense of repair or replacement as described above.
Products must be installed and maintained in accordance with The Partlow-West Company
instructions. Users are responsible for the suitability of the products to their application.
There is no warranty against damage resulting from corrosion, misapplication, improper
specifications or other operating condition beyond our control. Claims against carriers for
damage in transit must be filed by the buyer.
This warranty is void if the purchaser uses non-factory approved replacement parts and
supplies or if the purchaser attempts to repair the product themselves or through a third party
without factory authorization.
Returns
Partlow-West's sole and exclusive obligation and buyer’s sole and exclusive remedy under the
above warranty is limited to repairing or replacing (at Partlow-West's option), free of charge,
the products which are reported in writing to The Partlow-West Company at its main office
indicated below.
The Partlow-West Company is to be advised of return requests during normal business hours
and such returns are to include a statement of the observed deficiency. The buyer shall pre-
pay shipping charges for products returned and The Partlow-West Company or its representa-
tive shall pay for the return of the products to the buyer.
73
THE PARTLOW-WEST COMPANY
2 CAMPION ROAD • NEW HARTFORD, NY 13413 USA
1-800-866-6659 • 315-797-2222 • FAX 315-797-0403
74