X10, Unit switch, Over/Under, Counting. z Setup: Calibration, Filter, Baud rate, Print format etc. z Battery operation (Optional) z IP65 protection (IND226) z RS232 communication 1.3 Dimensions IND221: 160mm x 120mm x 40mm IND226: 160mm x 120mm x 60mm (with bracket) 1.4 Order information IND221-XX IND226-XX XX: Load cell input type, e.g. 04 for 4-350ohm load cell. Accessories: Battery pack, RS232 cable,

Uploaded by

Osvaldo Carpio
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
247 views

IND22x English Manual

X10, Unit switch, Over/Under, Counting. z Setup: Calibration, Filter, Baud rate, Print format etc. z Battery operation (Optional) z IP65 protection (IND226) z RS232 communication 1.3 Dimensions IND221: 160mm x 120mm x 40mm IND226: 160mm x 120mm x 60mm (with bracket) 1.4 Order information IND221-XX IND226-XX XX: Load cell input type, e.g. 04 for 4-350ohm load cell. Accessories: Battery pack, RS232 cable,

Uploaded by

Osvaldo Carpio
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 27

IND221

IND226
Industry Terminal
User/Service Manual

METTLER TOLEDO
182837 (2006-5-25)
PRECAUTIONS

READ this manual BEFORE WARNING


operating or servicing this
equipment. DISCONNECT ALL POWER TO THIS UNIT
BEFORE INSTALLING, SERVICING, CLEANING,
OR REMOVING THE FUSE. FAILURE TO DO SO
COULD RESULT IN BODILY HARM AND/OR
PROPERTY DAMAGE.
FOLLOW these instructions
carefully.

SAVE this manual for future CAUTION


reference.
OBSERVE PRECAUTIONS FOR HANDLING
ELECTROSTATIC SENSITIVE DEVICES.

DO NOT allow untrained


personnel to operate, clean,
inspect, maintain, service, or
tamper with this equipment. WARNING
ONLY PERMIT QUALIFIED PERSONNEL TO
SERVICE THIS EQUIPMENT. EXERCISE CARE
ALWAYS DISCONNECT this WHEN MAKING CHECKS, TESTS AND
equipment from the power ADJUSTMENTS THAT MUST BE MADE WITH
POWER ON. FAILING TO OBSERVE THESE
source before cleaning or PRECAUTIONS CAN RESULT IN BODILY
performing maintenance. HARM.

WARNING
FOR CONTINUED PROTECTION AGAINST
SHOCK HAZARD, CONNECT TO PROPERLY
GROUNDED OUTLET ONLY. DO NOT REMOVE
Note: If the unit has been stored or THE GROUND PRONG.
transported in below freezing
temperatures, allow the unit to warm
up to room temperature before turning
on AC power.
CAUTION
BEFORE CONNECTING OR DISCONNECTING ANY INTERNAL ELECTRONIC
COMPONENTS OR INTERCONNECTING WIRING BETWEEN ELECTRONIC
EQUIPMENT, ALWAYS REMOVE POWER AND WAIT AT LEAST THIRTY (30)
SECONDS BEFORE ANY CONNECTIONS OR DISCONNECTION’S ARE MADE.
FAILURE TO OBSERVE THESE PRECAUTIONS COULD RESULT IN DAMAGE
TO OR DESTRUCTION OF THE EQUIPMENT, OR BODILY HARM.

CAUTION
THE IND226 SOCKET-OUTLET SHALL BE INSTALLED NEAR THE
EQUIPMENT AND SHALL BE EASILY ACCESSIBLE.

CAUTION
RISK OF EXPLOSION IF BATTERY IS REPLACED BY EN INCORRECT TYPE.
DISPOSE OF USED BATTERIES ACCORDING TO THE INSTRUCTIONS

CAUTION
REMOVE THE BATTERY OUT IF THE TERMINAL WILL BE STORED IN A LONG
TIME.

CAUTION
REPLACE THE FUSES ONLY BY SERVICE PERSONNEL
In conformance with the European Directive 2002/96/ EC
on Waste Electrical and Electronic Equipment (WEEE) this
device may not be disposed of in domestic waste.
This also applies to countries outside the EU, per their
specific requirements.
Please dispose of this product in accordance with local
regulations at the collecting point specified for electrical and
electronic equipment.
If you have any questions, please contact the responsible
authority or the distributor from which you purchased this
device.
Should this device be passed on to other parties (for private
or professional use), the content of this regulation must also
be related.
Thank you for your contribution to environmental protection.
Contents
1.0 Overview................................................................................................................................ 1
1.1 Specification ....................................................................................................................... 1
1.2 Main functions .................................................................................................................... 1
1.3 Dimensions ......................................................................................................................... 2
1.4 Order information ............................................................................................................... 2
2.0 Install ...................................................................................................................................... 3
2.1 Open the package................................................................................................................ 3
2.2 Electronic Connect.............................................................................................................. 3
2.2.1 Open the terminal................................................................................................. 3
2.2.2 Load Cell Connect................................................................................................ 3
2.2.3 Com1 RS232......................................................................................................... 4
2.3 Lead seal ............................................................................................................................. 4
2.4 Battery Pack Option ............................................................................................................ 5
2.4.1 Battery install......................................................................................................... 5
2.4.2 Recharge Battery.................................................................................................. 5
2.4.3 Use Battery............................................................................................................ 5
3.0 Operation ............................................................................................................................... 1
3.1 Operation HMI.................................................................................................................... 1
3.2 Basic function operation ..................................................................................................... 1
3.2.1 On/Off key.............................................................................................................. 1
3.2.2 Zero ........................................................................................................................ 1
3.2.3 Tare......................................................................................................................... 1
3.2.4 Clear ....................................................................................................................... 2
3.2.5 Print ........................................................................................................................ 2
3.3 Expand functions ................................................................................................................ 2
3.3.1 X10 Function ......................................................................................................... 2
3.3.2 Unit switch ............................................................................................................. 2
3.3.3 Over/Under Function............................................................................................ 2
3.3.4 Counting Function ................................................................................................ 4
4.0 Setup ...................................................................................................................................... 1
4.1 Enter Setup.......................................................................................................................... 1
4.2 Keys in setup....................................................................................................................... 1
4.3 Setup Details ..................................................................................................................... 10
5.0 Terminal Maintenance ....................................................................................................... 17
5.1 Daily maintenance ............................................................................................................ 17
5.2 Error Messages.................................................................................................................. 17
5.3 Software download ........................................................................................................... 18
1.0 Overview

1.1 Specification

z 6 digits 1.2” /30.5mm large green LED display. Robust long life.
z 6 Function keys, simple and easy.
z IND221: Plastic, Protection IP5x.
z IND226: Stainless steel, Protection: IP65.
z Executive voltage:+5VDC.
z Load Cell capability: Maximum 4-350ohm analog load cell.
z Zero signal input ranges: 0-5mV.
z SPAN signal input ranges: 1-10mV.
z Resolution: 1,000,000.
z Increments: 1,000 - 30,000
z A/D Rates: 27 /seconds.
z Working voltage: 100-240VAC, 0.1A
DC: NI-HM rechargeable battery (Optional)
DC: C Size Dry Battery (only for IND221)
z RS232 Serial port
z Working temperature: -10°C - +40°C, Relative Humidity < 90%.
z Storage temperature: -20°C - +60°C, Relative Humidity < 90%.

1.2 Main functions

z Basic weighing: Zero, Tare, Clear, Print.


z Auto print function.
z Units switch: kg, lb.
z x10 function / Simple check weighing / Counting.
z English/Chinese print formats.
z Support micro ticket printer PQ16.
z Power saving technology. Low battery icon.
z Auto power off.
z Software download

1
1.3 Dimensions

129.1mm

1.5
150mm 90°

45°
136.8mm

220mm

IND221
220mm 90mm

150mm

132mm 66mm

IND226

1.4 Order information

Model Name Descriptions P/N


IND221-1000 Plastic Standard(with dry battery case) 72183995
IND221-1001 Plastic With Ni-MH rechargeable battery 72183997
IND226-1000 Harsh, Standard 72183987
IND226-1001 Harsh, With Ni-MH rechargeable battery 72183989

2
2.0 Install
This part will talk about the installation for IND221 and IND226. Please read
this chapter carefully before install.

2.1 Open the package

Open the package, and check all the parts with the checklist. Make sure no
parts were damaged and missing.
Remove the terminal from its protective package.

2.2 Electronic Connect

2.2.1 Open the terminal


The IND221 terminal use 4 screws to lock the front cover.
The front panel of the IND226 terminal is locked in place by four spring clips attached
to the enclosure body. To gain access to the terminal’s PCB for internal wiring and setting
switches, separate the front panel from the enclosure as follows:
Insert the tip of a flat-blade screwdriver into one of the two slots located on the bottom
of the front panel assembly and gently push in toward the enclosure. A “pop” sound is
made when the cover is released.

2.2.2 Load Cell Connect


7 Pins terminal strip.
1 Pin 1 - +EXC Pin 5 - -SIG
Pin 2 - +SEN Pin 6 - -SEN
+EXC

Shield
+SEN

-EXC
-SEN
+SIG

Pin 3 - +SIG
-SIG

Pin 7 - -EXC
Pin 4 - Shield

3
For 4 wires load cell, you should short the W1 two pins, and short the W2
two pins in PCB.
IND221 use PG9 cable bush grip, the allow cable diameter is 4 to 9mm.
IND226 use PG11 cable bush grip, the allow cable diameter is 5 to 10mm.
We recommend use CMX or CM cables for load cell connect.
The cable information:
z Type: CMX, CM
z Parameter: 6x0.2mm2
z Voltage: Max. 380V
z Temperature range: -30~+80°
z Cable diam: 6.1±0.15mm.
z Isolation material: UBEC180 polyethylene.
z Protection jacket: 90°C HZ-90(318#)

2.2.3 Com1 RS232


IND221:
RS232 use D-Sub Male connector
1 2 3 4 5
Pin 2 - RXD
Pin 3 - TXD
Pin 5 - GND

6 7 8 9
IND226:
Serial port use 3 Pins terminal strip.
1 Pin 1 - TXD
Pin 2 - RXD IN D226
C om puter
Pin 3 - GND D B9
GND
RXD
TXD

1 2

2 3

3 5
2.3 Lead seal

After setup and calibration you can seal the terminal.


Lead sealing:

IND221: IND226:

4
2.4 Battery Pack Option

2.4.1 Battery install


Recharge PCB install:

IND221 IND226

NI-HM battery package install:

IND221 IND226

2.4.2 Recharge Battery


When the low battery cursor light on red color, it means the battery voltage is
low, you can still continuous your work for about one hour..
When the low battery cursor flash on red color, it means the battery voltage is
too low, you should charge the battery immediately.
Plug the AC power line will automatic recharge the battery, and the low battery
cursor will light on green color. Normally the charge time is about 12 hours.
For new terminal, please charge the battery for 12 hours before use the battery.

2.4.3 Use Battery


Full charged battery can continuous work 35 hours.

5
The following setup can longer the work hours:
z Set timeout (F3.1.1), When the scale hasn’t operate for certain time, the
terminal will automatic turn off the display. Only kg cursor light. And when
the scale was added weight or press any key on terminal. The terminal will
automatically turn on the display.
z Set brightness to low (F3.1.2). When the terminal use battery, the display
brightness will turn to low, when use AC power, the display brightness will
automatically change to high.
z Set Auto power off (F3.2), When the scale haven’t any operation in certain
times, the terminal will automatically power off.

Note: Remove the battery out if the terminal will be stored for a long time.

6
3.0 Operation

3.1 Operation HMI

METTLER TOLEDO

Under OK Over NET >|1|< >|2|< lb kg Battery


Function
Units
Motion
Ranges
Net

0 T F C

On/Off Zero Tare Function Clear Print

“Under OK Over” is use for check weighing. You can stick ”Count APW
PCS” label for counting function.

3.2 Basic function operation

3.2.1 On/Off key


Hold the key 2 seconds, All display segments will light. The
terminal will show software part number [178037] [L 1.00].
Then the terminal will show the normal weight.
In normal display state, Hold this key 2 seconds, the terminal
will show [ -OFF- ], then power off.

3.2.2 Zero
0
Zero the scale.

3.2.3 Tare
T
Tare the scale, the display will change to Net mode..

1
3.2.4 Clear
Clear the tare, the display will go back to gross mode.
C

3.2.5 Print
Print the current weight.

3.3 Expand functions

3.3.1 X10 Function


F2.1 is set to ÑÖUL10 - x10 function.

F
Under OK Over NET >|1|< >|2|< lb kg Press key, The terminal will show
more accuracy weight for 20 seconds.

Under OK Over NET >|1|< >|2|< lb kg


Print is forbidden in this mode.

3.3.2 Unit switch


F2.1 is set to Unit – Unit switch

F
Under OK Over NET >|1|< >|2|< lb kg
Press key, you can switch the
weight units at kg or lb.

Count APW PCS NET >|1|< >|2|< lb kg

3.3.3 Over/Under Function


F2.1 is set to OVEr - Over/Under function
z F2.1.1 is set to CHECh - Check weighing mode.
Target weight setting:
F
Press key to switch Over/Under
mode.
F
Under OK Over NET >|1|< >|2|< lb kg
Hold key for 2 seconds. Or you can
press Tare key to tare the weight.
The terminal show the old target weight..
Under OK Over NET >|1|< >|2|< lb kg

2
If F2.2.2 is set to WEIGHt- Get target
weight from weighing. Then you can put
the target weight in scale platform. Press
Under OK Over NET >|1|< >|2|< lb kg

F
key to set new target weight.

Under OK Over NET >|1|< >|2|< lb kg


If F2.2.2 is set to MAnUAL- Manually set the
target weight. You can input target weight
directly, then press Enter to confirm.
Operation:

Current weight less than target weight and


over tolerance.
Under OK Over NET >|1|< >|2|< lb kg

Current weight near the target weight and


within tolerance.
Under OK Over NET >|1|< >|2|< lb kg

Current weight larger than target weight


and over tolerance.

Under OK Over NET >|1|< >|2|< lb kg

z F2.1.1 is set to CLASS- Classifying mode.


Target weight setting:

F
Press key to switch Over/Under
Under OK Over NET >|1|< >|2|< lb kg
mode.

F
Hold key for 2 seconds. Or you can
Under OK Over NET >|1|< >|2|< lb kg

press Tare key to tare the weight.

The terminal show the old target weight..

If F2.2.2 is set to WEIGHt- Get target


Under OK Over NET >|1|< >|2|< lb kg
weight from weighing. Then you can put
the target weight in scale platform. Press

F
Under OK Over NET >|1|< >|2|< lb kg
key to set new target weight.
If F2.2.2 is set to MAnUAL- Manually set the
target weight. You can input target weight
directly, then press Enter to confirm.
Operation:

3
Current weight less than target weight and
over tolerance.
Under OK Over NET >|1|< >|2|< lb kg

Current weight near the target weight and


within tolerance.
Under OK Over NET >|1|< >|2|< lb kg

Current weight larger than target weight


and over tolerance.
Under OK Over NET >|1|< >|2|< lb kg

3.3.4 Counting Function


F2.1 is set to Count - Counting function.
Stick “Count APW PCS” label at “Under OK Over” position.
Display Mode switch:

Normal display mode.

Count APW PCS NET >|1|< >|2|< lb kg

F
Press key to switch counting PCS
mode, show current pieces.
Count APW PCS NET >|1|< >|2|< lb kg

F
Press key to switch counting APW
Count APW PCS NET >|1|< >|2|< lb kg mode, show average piece weight.

Sampling:

F
Hold key 2 seconds. Or you can

Count APW PCS NET >|1|< >|2|< lb kg


press Tare key to tare the weight.

F
Press key to select sample pieces:
Count APW PCS NET >|1|< >|2|< lb kg
5,10,20,50. Put the sample pieces on scale
platform, press Enter to confirm.
Display mode will go back to counting PCS
Count APW PCS NET >|1|< >|2|< lb kg mode.

APW Enhancement:
If F2.3 = On, Enable APW enhancement function. Then after the pieces
number increased, the terminal will automatically adjust the average weight to
get more accuracy pieces number.

4
4.0 Setup

4.1 Enter Setup

Hold key 2 seconds, terminal will show [MAStEr] . Ask for password:

0 T 0 T
Supervisor password: (Can setup every
parameters).

0 0 0
Operator password: (Can only setup F2).

Press to confirm. Terminal will show [SEtUp] ..


If F1.1 is set OIML or NTEP. Then F1, F5.1 and F5.4 will be disabled for setup.
The only way you can setup them is turn off the terminal, hold the S1 key in
PCB, then turn on the terminal. The terminal will show “SEtUp” directly, now
you can setup everything.

4.2 Keys in setup

C
0 Go back to previous setup
Previous parameter
F
T Go back up level
Next parameter

Confirm.

F
Digits Input: Press key to edit current number. Change input cursor
to left.
C
Press key to clear current number.

0 T
Press key to decrease. Press key to increase.

1
Press to confirm

4.3 Setup Details

F1 – Scale

F1.1 – Approval
Selection: no(Factory default) - none
OIÑÖL - OIML
ntEP - NTEP
otHEr - Other approval

F1.2 – Scale Capacity & Increments


F1.2.1 – Units
Selection: 1 - kg (Factory default)
2 - lb

F1.2.2 – Ranges
Selection: 1 r - 1 Range (Factory default)
2 r - 2 Ranges

F1.2.3 – Capacity (First Range)


Selection: 3 … 20’000 (Factory default 6)

F1.2.4 – Increments (First Range)


Selection: 0.0001 … 10 (Factory default 0.001)

F1.2.5 – Capacity (Second Range)


Selection: 3 … 20’000 (Factory default 3)
Notes: Capacity (Second Range) should less than Increments (First Range).

F1.2.6 –Increments (Second Range)


Selection: 0.0001 ~ 10 (Factory default 0.001)

F1.3 – Calibration
F1.3.1 GEO Adjustment
Selection: 0…31 (Factory default16)

F1.3.2 – Linearity
10
Selection: On - Enabled
OFF - Disabled (Factory default)

F1.3.3 – Calibration
z If F1.3.2 is set to On
[E SCL] Empty scale.
Remove any load from scale platform, then press Enter to confirm.
Terminal will count down from [10 CAL] to [ 0 CAL].
[FULL Ld] Add full load.
Add load on the scale platform, then press Enter.
[000000] Input weight number, then press Enter to confirm. Terminal will
count down from [10 CAL] to [ 0 CAL].
[ donE ] Complete.

z If F1.3.2 is set to OFF


[E SCL] Empty scale.
Remove any load from scale platform, then press Enter to confirm.
Terminal will count down from [10 CAL] to [ 0 CAL].
[Add Ld]Add middle load
Add load on the scale platform, then press Enter.
[000000] Input weight number, then press Enter to confirm. Terminal will
count down from [10 CAL] to [ 0 CAL].
[FULL Ld] Add full load.
Add load on the scale platform, then press Enter.
[000000] Input weight number, then press Enter to confirm. Terminal will
count down from [10 CAL] to [ 0 CAL].
[ donE ] Complete.

F1.4 – Zero Function


F1.4.1 – AZM
Selection: OFF, 0.5d(Factory default),1d,3d

F1.4.2 – Power up Zero


Selection: OFF, 2%,10%(Factory default), 20%

F1.4.3 – Pushbutton Zero


Selection: OFF,2%(Factory default),10%,20%
If F1.1 is set to OIML, Then F1.4.2 will set to 10%, and F1.4.3 will set to 2%.

F1.5 – Tare Function


F1.5.1 –Auto Tare
Selection: On, OFF(Factory default)

11
F1.5.2 – Auto Clear
Selection: On, OFF(Factory default)

F1.5.3 – Tare Interlock


Selection: On, OFF(Factory default)

F1.5.4 – Auto Tare threshold.


Selection: 0 - FS(Factory default10d)

F1.5.5 – Auto Tare reset threshold


Selection: 0 - FS(Factory default10d)

F1.6 – Digital filter


F1.6.1 – Digital filer
Selection: Lo - Low
ÑÖEd (Factory default) - Middle
HIGH - High

F1.6.2 – Motion range


Selection: 0.5d(Factory default),1d,3d

F1.10 – F1 Factory default


Reset all F1 parameters to Factory default. Not include calibration values.

F2 – Application

F2.1 – F key function


Selection: ÑÖUL10(Factory default) - x10
Unit - Unit switch
OVEr - Over/Under
Count - Counting

F2.2 – Over/Under function


F2.2.1 – Display mode
Selection: CHECh(Factory default) - Check weighing
CLASS - Classifying

F2.2.2 – Target Input


Selection: WEIGHt(Factory default)- By weight
MAnUAL - Manually

12
F2.2.3 – Plus Tolerance
Selection: 0…FS(Factory default 10d)

F2.2.4 – Minus Tolerance


Selection: 0…FS(Factory default10d)

F2.3 – APW Enhancement(F2.1 is set to Counting)


Selection: On,OFF(Factory default)

F2.10 – F2 reset to Factory default


All F2 parameters will be set to Factory default.

F3 – Terminal

F3.1 – Display
F3.1.1 – Screen saver timeout
Selection: 0, 10 - 999 seconds (Factory default 60s) 0 will disabled this
function.

F3.1.2 – Brightness
Selection: Lo(Factory default) - Low bright
HIGH - High bright
Recommend to set Low if you use battery to operation.

F3.2 – Auto power off


Selection: 0, 5~60 minutes (Factory default 5 minutes) 0 will disabled this
function.

F3.3 – Battery type


Selection: drY(Factory default) - Dry battery
ni-ÑÖx - Ni-MH rechargeable battery
LEAd-A - Lead-Acid rechargeable battery

F3.10 – F3 reset to Factory default


All F3 Parameters will be set to Factory default.

F4 – Communication

13
F4.1 – Connections
Selection: Print(Factory default) - Demand print
APrint - Auto print
SICS - SICS
Contin - Toledo continuous mode

F4.2 – Format
F4.2.1 – Line format
Selection: ÑÖULti(Factory default) - Multi line
SinGLE - Single line

F4.2.2 – Print format


Selection: StAndr(Factory default)- Standard
OVEr - Over/Under
Count - Count

F4.2.3 – Print languages


Selection: EnG(Factory default) - English
CHn - Chinese

F4.2.4 – Add Line Feed


Selection: 0~9 (Factory default 3)

F4.2.5 – Auto print threshold


Selection: 0 ~ FS (Factory default10)

F4.2.6 – Auto print reset threshold


Selection: 0 ~ FS (Factory default10)

F4.3 – Com1
F4.3.1 – Baud rate
Selection: 1200, 2400, 4800, 9600(Factory default),19200

F4.3.2 – Data bits/Parity


Selection: 7-odd - 7 bits odd parity
7-EvEn - 7 bits even parity
8-nonE(Factory default)- 8 bits none parity

F4.3.3 – Xon/Xoff
Selection: On - Enabled
OFF (Factory default) - Disabled

14
F4.3.4 – Checksum
Selection: On - Enabled
OFF (Factory default) - Disabled

F4.10 – F4 reset to Factory default


All F4 parameters will reset to Factory default.

F5 – Maintenance

F5.1 – Calibration values


F5.1.1 – Zero Counts
F5.1.2 – Middle load weight (half capacity)
F5.1.3 – Middle load counts
F5.1.4 – Full load weight
F5.1.5 – Full load Counts

F5.2 – Keypad test


Terminal show “PrESS”, You can press Zero, Tare, F, Clear, Print. On/Off key
to exit.

F5.3 – Display test


All display segments will light.

F5.4 – Display internal resolution


Display internal resolution.

F5.5 – COM1 test


Connect COM1 to computer to test COM1.

F5.6 – Print setup


Print all setup parameters.

F5.10 – Reset all parameters to Factory default


Reset all F1 ~ F4 Reset all parameters to Factory default. Not include
calibration values.

F6 – Exit Setup
15
[SAVE] Press Enter to confirm save changes and exit..
Press Tare key, the terminal will show[AbOrt] , will not save changes and exit.

16
5.0 Terminal Maintenance

5.1 Daily maintenance

Main PCB:

Cal switch S1
U8

U7 DC in
J7

J1 Keypad
1 AC out

J8
Software 4 wire 1
download load cell
connectors Fuse
J3
F1
w3 W1 J2 W2 J6
1
1 1

Load cell AC input RS232 COM1

J1 - Keypad connector.
J2 - Load cell connector.
J3 - RS232 connector.
J6 - AC power in, 87~264VAC.
J7 - DC power in, from rechargeable PCB and Dry battery.
J8 - AC power out, to rechargeable PCB.
W1 ,W2 - 4 wires load cell connectors.
W3 - Factory use.
S1 - Calibration switch.
F1 - Fuse, 250V 1.5A.

5.2 Error Messages

Message Possible reasons Solution

17
{||||} Over load, more than 9d Reduce the load
above scale capacity

<____> Under Zero 5d Zero the scale

{|no|} Over the zero range Remove the load

<_no_>

--no-- Key forbidden Check setup

Err 3 EEPROM verify error Reset the terminal

Err 35 Scale is in motion when Check the scale


calibration

Err 4 Samples number too small. Add sample number

Err 6 EEPROM W/R error Replace EEPROM

Err 70 The keys hold too long Replace keypad


The key may be short

Terminal auto Terminal is set auto power off. Press On/off key
power off
Battery voltage to low. Recharge battery

Terminal not light Fuse break Replace fuse


after power on

battery can’t be Batter type set incorrect Set F3.3 the correct
charged battery type.
Battery was damaged
Replace battery.

Battery working Battery charge not full. Charge battery longer


hour decreased
Battery life was reached Replace new battery
rapidly

5.3 Software download

IND221 and IND226 can download software in the field by using Windows
HyperTerminal software. Protocol: 19200, 8, None, Xmodem.

18