Troubleshooting Problems With Peak Shape
Troubleshooting Problems With Peak Shape
Peak Tailing
Possible
Cause Prevention/Solution
Interaction
with active silanols Use
ultra-high purity silica-based stationary
phase
Add
basic mobile phase additive (eg. TEA)
–
not needed with ultra-high purity phases
Chelation
with metal ions in stationary Use
ultra-high purity silica-based stationary
phase phase
Add
basic mobile phase additive (eg. TEA)
–
not needed with ultra-high purity phases
Wrong
mobile phase pH Decrease
mobile phase pH to suppress
silanol
ionization
Increase
buffer concentration.
https://www.pharmaceuticalsky.com/2022/08/troubleshooting-problems-with-peak-shape.html 1/3
8/13/22, 9:34 AM Troubleshooting Problems with Peak Shape
Blocked
frit Reverse
flush the column
Use
in-line filter
Column
void Reverse
flush the column
Replace
the column
Unswept
dead volume Minimize the number of connections
Use
shorter connection tubing
Check
all fittings are tight
Split Peaks
Possible
Cause Prevention/Solution
Contamination
on guard or Remove
guard cartridge and carry out analysis –
analytical column inlet replace guard if necessary
For
strongly retained contaminants, try the
regeneration procedure
Replace
column
Blocked
frit Reverse
flush the column
Use
in-line filter
Sample
solvent incompatible with Inject
sample in the mobile phase
the mobile phase
Simultaneous
elution of the Use
sample clean-up prior to injection
second component
Change
selectivity by changing mobile phase or
column phase
https://www.pharmaceuticalsky.com/2022/08/troubleshooting-problems-with-peak-shape.html 2/3
8/13/22, 9:34 AM Troubleshooting Problems with Peak Shape
Column
overloaded Use
higher capacity stationary phase
Increase
column diameter
Decrease
sample amount
Peak Fronting
Possible
Cause Prevention/Solution
Formation
of channels in column Replace
the column
Operate
within recommended pH limits of
the column
Column
overloaded Inject a smaller volume or more dilute
sample solution
Use
higher capacity stationary phase
Sample
solvent incompatible with the Inject
sample in the mobile phase
mobile phase
Low
temperature Increase
column temperature
https://www.pharmaceuticalsky.com/2022/08/troubleshooting-problems-with-peak-shape.html 3/3