MEG Manual - Baby CPAP
MEG Manual - Baby CPAP
Page 1 of 13
© Copyright Diamedica (UK) Ltd 2017
Read this page first
INTENDED USE
This device is suitable for use in hospital settings with limited resources or in any field or outreach
locations (where a suitable power source is available) and is suitable for neonate and paediatric
patients.
The Diamedica CPAP provides effective and efficient non-invasive airway pressure support to
neonate and paediatric patients in difficult environments.
FOREWORD
This manual is intended to provide guidance on the function, performance and user maintenance
of the Diamedica Baby CPAP. The information given in this manual is correct at the date of
publication.
A Separate IFU manual for the oxygen concentrator is also included and must be reviewed for
specific safety and maintenance requirements before use.
The policy of Diamedica (UK) Ltd is to continuously improve its products. Changes may be made to
this manual without notice being given.
Users of the Diamedica Baby CPAP must read, understand and follow the guidance given in this
manual before using the system.
WARNING
The Diamedica Baby CPAP delivers mixtures of oxygen-enriched air to the patient and the device
should be monitored at all times.
It is essential that the patient’s oxygen saturation and other vital functions are also monitored.
The ultimate responsibility for patient safety remains with the operator.
Observations of the patient must take precedence over machine settings in judging the condition
of the patient.
Page 2 of 13
© Copyright Diamedica (UK) Ltd 2017
Only persons who have read and understood this entire manual and
therefore deemed competent, are authorized to operate this equipment
Humidifier bottle
Expiratory limb
Inspiratory limb
Surge protector
Oxygen / Air mixing
chart
Expiratory limb
Page 4 of 13
© Copyright Diamedica (UK) Ltd 2017
Connect the nasal prongs to the child as below:
8. Dial the flows of air and oxygen required to see a constant stream of bubbles in the water.
Start with 5 L/min of oxygen and 5 L/min of air. Check and record the flows dialled and the
oxygen concentration this provides. A total flow of 10 L/min is sufficient in most cases to
deliver CPAP, which you can see by continuous bubbling in the bottle.
9. Check the child for signs of respiratory distress, check the SpO2, and check if there are bubbles
in the bottle.
a. If the SpO2 is below 90% or the child has severe respiratory distress, first increase
the CPAP level to 8 or 10cmH20 and then, if no response the oxygen flow meter to
as much as 8 L/min.
b. If there are not continuous bubbles, check the nasal prongs are attached properly,
and reposition them so they fit snugly inside the nostrils.
c. If there are still not continuous bubbles, check for leaks along the circuit, and adjust
the oxygen or air flows according to the chart on the CPAP.
Page 5 of 13
© Copyright Diamedica (UK) Ltd 2017
SpO2 >90% and SpO2 >90% but SpO2 <90%
mild or no moderate to severe
respiratory distress respiratory distress
At least once each day, children in the ward who are clinically stable (have no emergency signs and
SpO2 >90%) should be disconnected from CPAP for 10–15 minutes, and carefully examined for
changes in clinical signs and SpO2, to assess whether supplemental oxygen and CPAP is still
required. Trials off CPAP are best done first thing in the morning, when there is likely to be
adequate staff to observe the child throughout the day. If trials off CPAP are started in the late
afternoon, low staff numbers overnight and the oxygen desaturation that sometimes occurs
during sleep mean that there is a risk of hypoxaemia developing unrecognized overnight.
Children who have an SpO2 <90% while still on CPAP or who are unstable or very unwell should
not be given trials on room air.
Before a trial off CPAP, the SpO2 should be checked to determine if the trial is safe
(i.e. SpO2 >90%). The child should then be disconnected from the oxygen source and observed
carefully to avoid any adverse complications of hypoxaemia. If severe hypoxaemia (SpO2 <80%),
apnoea or severe respiratory distress occurs, children should be immediately restarted on oxygen.
Some children will become hypoxaemic very rapidly when they are taken off oxygen, and this is a
marker of very severe disease and a high risk of death. Parents and nursing staff should be advised
to watch the child to see if he/she develops cyanosis or severe respiratory distress.
Page 6 of 13
© Copyright Diamedica (UK) Ltd 2017
If the SpO2 is 90% or more after a trial on room air, they should remain off oxygen and the SpO2
should be rechecked after one hour, as late desaturation can sometimes occur.
If the SpO2 on room air is <90% or there is severe respiratory distress they should go back on
CPAP.
Any child who appears to deteriorate clinically should have their SpO2 checked to determine
whether they need oxygen. Children with acute respiratory disease should not be discharged until
their SpO2 has been stable at 90% or more while breathing room air for at least 24 hours, until
danger signs have resolved, and until appropriate home treatment can be organized.
This is essential to reduce the risk of cross-infection. This should be done after every patient has
used the CPAP, and weekly if the same child is on CPAP for over a week. A spare circuit is essential
to replace the one that is being cleaned without interruption to the child’s treatment.
The CPAP circuit (inspiratory and expiratory limb, bottle and lid and connections) must be
thoroughly cleaned as follows:
1. Wipe off any gross soiling. Clean first with detergent (soap) to remove gross contamination
(e.g. blood, sputum);
2. Rinse with water that has been boiled and allowed to cool to tepid. Let it dry
3. Disinfect before re-using.
• Staff cleaning the equipment must wear protective clothing to avoid splash exposure or
contact with dirty equipment: wear apron, gloves and glasses.
• Good ventilation of the area is needed where you are cleaning the equipment
Page 7 of 13
© Copyright Diamedica (UK) Ltd 2017
How to Wash and Disinfect
1. Wash hands
2. Prepare detergent (soapy water) and disinfectant solution wearing personal protective
equipment (PPE): gown or impermeable apron as a minimum, eye protection and
surgical mask for bleach solution
3. Wash first in soapy water, to remove blood and respiratory secretions. In clean sink or
bucket brush the equipment under water to prevent splash and ensure all visible soiling
is removed, rinse with water that has been boiled but allowed to cool to tepid
4. Let dry
5. Wash next in diluted bleach or disinfectant. Bleach needs one hour of soaking. Soak all
items together, do not keep putting one in and taking one out. Once used bleach
should not be re-used or kept in storage, discard after use.
6. Rinse with water that has been boiled and cooled to tepid (rinse also inside, for
example, using a sterile syringe), let it drip dry over the sink, do not leave it coiled on
the sink.
7. Remove all protective gear (apron, gloves, mask) and wash hands thoroughly
8. Check that there is no pooled water in the circuit. Store the circuit and bottle in a clean
plastic bag (labelled and dated). Store in the dry and clean area (separate from a soiled
equipment area)
Routine Cleaning and Maintenance of the CPAP-Concentrator
Each week the oxygen concentrator will require approximately 30 minutes of attention.
Concentrators have a large particulate filter over the air inlet opening (usually at the back of
freestanding or portable models). This filter stops dust and other airborne particles from entering
the unit. The filter should be removed and cleaned in warm soapy water, completely dried with an
absorbent towel and replaced. Have a spare dry filter to replace with so there are minimal
interruptions to the concentrator function.
The exterior of the oxygen concentrator should be cleaned with a mild disinfecting cleaning agent
or a diluted solution of bleach (usually 5.25% sodium hypochlorite). A solution in the range of
1:100 to 1:10 of bleach to water can be used effectively. Allow the solution to remain on the
surface for 10 minutes and then rinse off and dry.
Page 8 of 13
© Copyright Diamedica (UK) Ltd 2017
An additional water trap may be added to
remove condensed water from the system
To empty the water trap, unscrew bottom section empty and replace. The CPAP can remain running
whist removed without effecting CPAP flow.
If the product is returned to the manufacturer at the end of its life the company will ensure
disposal in line with the Waste Electrical and Electronic Equipment Directive (WEEE) 2012/19/EU
Page 9 of 13
© Copyright Diamedica (UK) Ltd 2017
Instructions For Use – Additional information
This Baby CPAP unit has been fitted with an AVS30 Voltage Stabilizer to automatically protect the
on-board circuitry from damage due to unstable voltage.
Input good
GREEN LIGHT
Power OK
Input unstable
RED LIGHT
No Power
OR
OR
Page 10 of 13
© Copyright Diamedica (UK) Ltd 2017
DIAMEDICA (UK) LTD
Grange Hill Industrial Estate, Bratton Fleming
Barnstaple, Devon, EX31 4UH, UK
Tel : +44(0)1598 710066
Email : [email protected]
www.diamedica.co.uk
0088
SYMBOLS GLOSSARY
Some or all the following symbols may be used within this manual or found on the product or
packaging labels. Please familiarize yourself with them:
Page 11 of 13
© Copyright Diamedica (UK) Ltd 2017
Serial number Indicates the manufacturer's serial number
so that a specific medical device can be
identified.
Class II equipment
Type BF applied
part
Page 12 of 13
© Copyright Diamedica (UK) Ltd 2017
The battery Chemical symbol for battery type included
recycling symbol beneath
Indicates that an
object is capable of
being recycled
Page 13 of 13
© Copyright Diamedica (UK) Ltd 2017