0% found this document useful (0 votes)
439 views60 pages

MC-55-097-001 r2 DRE ASG 300 UG-E

Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
439 views60 pages

MC-55-097-001 r2 DRE ASG 300 UG-E

Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 60

USER’S GUIDE

User ’s Guide • ASG-300 i


ii 800-477-2006 • www.dremed.com
This manual and the equipment it describes are for use only by qualified medical professionals trained in the particular technique
and surgical procedure to be performed. It is intended as a guide for using the ASG-300 only.

Equipment Covered in this Manual


ASG-300:
Reference No.: ASG-300

For Information Contact


DRE, Inc. • 1800 Williamson Court • Louisville, KY 40223 USA
U.S. Phone 1-800-477-2006 Fax 1-502-244-0369 • International Phone +1-502-244-4444 Fax +1-502-244-0369
www.dremed.com • [email protected]

Distributed by:
DRE, Inc.
1800 Williamson Court
Louisville, KY 40223

Part Number - MC-55-097-001 Rev. 2

©2013 DRE, Inc. All rights reserved. Contents of this publication may not be reproduced without the written permission of
DRE, Inc.

CONVENTIONS USED IN THIS GUIDE


WARNING:
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious
injury.

CAUTION:
Indicates a hazardous situation which, if not avoided, may result in minor or moderate injury.

NOTICE:
Indicates an operating tip, a maintenance suggestion, or a hazard that may result in product
damage.

User ’s Guide • ASG-300 iii


TABLE OF CONTENTS
Equipment Covered in this Manual............................................................................iii
For Information Contact .............................................................................................iii
Conventions Used in this Guide ........................................................................................iii
Introducing the ASG-300 ......................................................................................................1-1
Key Features...................................................................................................................1-2
Components and Accessories.........................................................................................1-3
Safety ..............................................................................................................................1-4
Controls, Indicators, and Receptacles................................................................................2-1
Front Panel......................................................................................................................2-2
Symbols on the Front Panel....................................................................................2-3
Preset Controls................................................................................................................2-4
Cut and Blend Controls...................................................................................................2-5
Coag Controls .................................................................................................................2-6
Bipolar Controls...............................................................................................................2-7
Indicators.........................................................................................................................2-8
Power Switch and Receptacles.......................................................................................2-9
Rear Panel ....................................................................................................................2-10
Symbols on the Rear Panel ..................................................................................2-10
Getting Started ......................................................................................................................3-1
Initial Inspection ..............................................................................................................3-2
Installation .......................................................................................................................3-2
Function Checks..............................................................................................................3-2
Setting Up the Unit ..................................................................................................3-2
Checking the Return Electrode Alarm.....................................................................3-2
Confirming Modes ...................................................................................................3-3
Checking Bipolar Mode (with bipolar footswitch) ....................................................3-3
Checking Monopolar Mode (with monopolar footswitch) ........................................3-3
Checking Monopolar Mode (with handswitch) ........................................................3-3
Performance Checks.......................................................................................................3-3
Using the ASG-300................................................................................................................4-1
Inspecting the Generator and Accessories .....................................................................4-2
Setup Safety....................................................................................................................4-2
Setting Up .......................................................................................................................4-3
Preparing for Monopolar Surgery....................................................................................4-4
Applying the Return Electrode ................................................................................4-4
Connecting Accessories ..........................................................................................4-4
Preparing for Bipolar Surgery..........................................................................................4-5
Setting and Recalling Memory Presets...........................................................................4-5
Memory....................................................................................................................4-5
Memory Function Overview.....................................................................................4-5
Setting Your Presets................................................................................................4-6
Memory Feature (Last Selected Preset) .................................................................4-6
Activating the Unit ...........................................................................................................4-7
Activation Safety..............................................................................................................4-8
Maintaining the ASG-300......................................................................................................5-1
Cleaning ..........................................................................................................................5-2
Periodic Inspection..........................................................................................................5-2
Fuse Replacement ..........................................................................................................5-2

iv 800-477-2006 • www.dremed.com
Troubleshooting ....................................................................................................................6-1
Repair Policy and Procedures .............................................................................................7-1
Responsibility of the Manufacturer..................................................................................7-2
Returning the Generator for Service ...............................................................................7-2
Step 1 – Obtain a Return Merchandise Authorization Number...............................7-2
Step 2 – Clean the Generator .................................................................................7-2
Step 3 – Ship the Generator ...................................................................................7-2
Technical Specifications......................................................................................................A-1
Performance Characteristics...........................................................................................A-2
Input Power.............................................................................................................A-2
Duty Cycle...............................................................................................................A-2
Dimensions and Weight ..........................................................................................A-2
Operating Parameters.............................................................................................A-2
Transport and Storage ............................................................................................A-2
Audio Volume..........................................................................................................A-3
Return Electrode Sensing.......................................................................................A-3
Low Frequency (50-60 Hz) Leakage Current .........................................................A-3
High Frequency (RF) Leakage Current ..................................................................A-4
Standards and IEC Classifications .................................................................................A-4
Class I Equipment (IEC 60601-1)...........................................................................A-4
Type CF Equipment (IEC 60601-1) / Defibrillator Proof .........................................A-4
Drip Proof (IEC 60601-2-2).....................................................................................A-4
Electromagnetic Interference ..................................................................................A-4
Electromagnetic Compatibility (IEC 60601-1-2 and IEC 60601-2-2) ......................A-4
Voltage Transients (Emergency Generator Mains Transfer) ..................................A-4
Output Characteristics ....................................................................................................A-5
Maximum Output for Monopolar and Bipolar Modes..............................................A-5
EMC Compliance............................................................................................................A-5
Output Power Curves .....................................................................................................A-9

Warranty ................................................................................................................................B-1

User ’s Guide • ASG-300 v


LIST OF FIGURES
Figure 2 – 1 Layout of controls, indicators, and receptacles on the front panel..................2-2
Figure 2 – 2 Controls for setting and recalling presets........................................................2-4
Figure 2 – 3 Controls for the Cut and Blend modes............................................................2-5
Figure 2 – 4 Controls for the Coag mode ............................................................................2-6
Figure 2 – 5 Controls for the Bipolar mode .........................................................................2-7
Figure 2 – 6 Indicators for power, return electrodes, and footswitch control.......................2-8
Figure 2 – 7 Location of the unit power switch and front panel receptacles .......................2-9
Figure 2 – 8 Layout of connectors and controls on the rear panel....................................2-10
Figure 5 – 1 Fuse holder......................................................................................................5-2
Figure A – 1 Display setting vs. output power in watts at rated load ..................................A-9
Figure A – 2 Output voltage (Vpeak) versus power setting at rated load ...........................A-9
Figure A – 3 Output power vs impedance for Cut I mode.................................................A-10
Figure A – 4 Output power vs impedance for Cut II mode ...............................................A-10
Figure A – 5 Output power versus impedance for Blend I (minimum) mode.....................A-11
Figure A – 6 Output power versus impedance for Blend II (maximum) mode...................A-11
Figure A – 7 Output power vs impedance for Pinpoint mode............................................A-12
Figure A – 8 Output power vs impedance for Spray mode ...............................................A-12
Figure A – 9 Output power vs impedance for Bipolar mode .............................................A-13

vi 800-477-2006 • www.dremed.com
INTRODUCING THE ASG-300
This section includes the following information:

● Key Features

● Components and Accessories

● Safety

Cautions:
Read all warnings, cautions, and instructions provided with this generator before using.
Read the instructions, warnings, and cautions provided with electrosurgical accessories before
using. Specific instructions are not included in this manual.

User ’s Guide • ASG-300 1-1


Key Features
The ASG-300 includes the latest technology. This unit offers unsurpassed performance, flexibility, reliability, and convenience.

It includes the following features:

• Two Cut Modes, Cut I & Cut II


Two cut modes give the surgeon flexibility to cut all types of tissue without losing performance.
Cut I generates constant output power over a wide range of impedances. Refer to Figure A-1 in the Technical Specifications section
of this guide.
Cut II is a softer cut that generates constant output power over a small range of impedances. Refer to Figure A-2 in the Technical
Specifications section of this guide.

• Blend with 10 settings


The Blend mode is a combination of cutting and hemostasis. The ASG-300 gives the surgeon freedom to adjust the desired level
of hemostasis. A setting of 1 is minimal blend with maximum cutting effect. A setting of 10 is maximum hemostasis (blend) with
minimal cutting effect. This adjustment is easily achieved by a incremental adjustment. Refer to Section 2, Controls, Indicators,
and Receptacles, Cut and Blend Controls. The Blend mode improves the rate of targeted tissue desiccation without increasing the
power delivered by the generator.

• Presets
The surgeon can store 10 user-defined presets for easy recall of frequently used settings.

• Two levels of coagulation: Pinpoint and Spray


Pinpoint provides precise control of bleeding in localized areas.
Spray provides greater control of bleeding in highly vascular tissue over broad surface areas.

• Return electrode sensing and contact quality monitoring


The ASG-300 incorporates a return electrode contact quality monitoring system (Bovie NEM™). This system detects the type of
return electrode: solid or split. The system also continually monitors the contact quality between the patient and the split return
electrode. This feature is designed to minimize patient burns at the return electrode site.

• FDFSTM (Fast Digital Feedback System)


The FDFSTM (Fast Digital Feedback System) measures voltage and current at 5,000 times a second and immediately adjusts the
power to varying impedance during the electrosurgical procedure. The unit’s digital technology senses and responds to changes
in tissue and density. Unlike analog, this feature reduces the need to adjust power settings manually.

Notices:
The Bovie NEM™ system recommends that you use a split return electrode.
Before activation, pad placement and visual verification of the split return electrode (split pad) indi-
cator on the front panel is recommended. After connecting the split pad to the generator and plac-
ing the split pad securely to the patient, give the unit 5 to 10 seconds to recognize the split pad.
The split pad indicator will illuminate green. If the split pad and cord are attached to the generator
without secure contact to the patient, the alarm indicator will illuminate red.

• Memory
The unit automatically powers up to the last selected preset settings.

• Isolated RF output
This minimizes the potential of alternate site burns.

• Standard connectors
These connectors accept the latest monopolar and bipolar instruments. Refer to Section 2, Controls, Indicators, and Receptacles to
learn more.

1-2 800-477-2006 • www.dremed.com


• Self diagnostics
These diagnostics continually monitor the unit to ensure proper performance.

compoNeNts aNd accessories


You should receive the following components with your generator:

• ASG-300

• Hospital-grade power cord (110 VAC and 220 VAC)

• User’s Guide

User ’s Guide • ASG-300 1-3


saFety
The safe and effective use of electrosurgery depends to a large degree on factors solely under the control of the operator. There is no
substitute for a properly trained and vigilant medical staff. It is important that they read, understand, and follow the operating
instructions supplied with this electrosurgical equipment.

Physicians have used electrosurgical equipment safely in numerous procedures. Before starting any surgical procedure, the surgeon
should be familiar with the medical literature, complications, and hazards of using electrosurgery in that procedure.

To promote the safe use of the ASG-300, this section presents the warnings and cautions that appear throughout this user’s guide. It
is important that you read, understand, and follow the instructions in these warnings and cautions so that you can operate this
equipment with maximum safety. It is also important that you read, understand, and follow the instructions for use in this
user’s guide.

warnings:
HazardousElectricaloutput - This equipment is for use only by trained, licensed physicians.
Danger:Fire/ExplosionHazard - Do not use the ASG-300 in the presence of flammable
materials.
Fire/ExplosionHazard - The following substances will contribute to increased fire and explosion
hazards in the operating room:
• Flammable substances (such as alcohol based skin prepping agents and tinctures)
• Naturally occurring flammable gases which may accumulate in body cavities such as
the bowel
• Oxygen enriched atmospheres
• Oxidizing agents (such as nitrous oxide [N20] atmospheres).
The sparking and heating associated with electrosurgery can provide an ignition source.
Observe fire precautions at all times. When using electrosurgery in the same room with any
of these substances or gases, prevent their accumulation or pooling under surgical drapes, or
within the area where electrosurgery is performed.
Connect the power cord to a properly polarized and grounded power source with the frequency
and voltage characteristics that match those listed on the back of the unit.
ElectricshockHazard - Connect the generator power cord to a properly grounded receptacle.
Do not use power plug adapters.
ElectricshockHazard - Always turn off and unplug the generator before cleaning.
FireHazard - Do not use extension cords.
Patientsafety - Use the generator only if the self-test has been completed as described.
Otherwise, inaccurate power outputs may result.
Failure of the high frequency electrosurgical equipment could result in an unintended increase of
output power.
The instrument receptacles on this generator are designed to accept only one instrument at a
time. Do not attempt to connect more than one instrument at a time into a given receptacle. Doing
so will cause simultaneous activation of the instruments.
Use the lowest output setting necessary to achieve the desired surgical effect. Use the active
electrode only for the minimum time necessary in order to lessen the possibility of unintended
burn injury. Pediatric applications and/or procedures performed on small anatomic structures may
require reduced power settings. The higher the current flow, and the longer the current is applied,
the greater the possibility of unintended thermal damage to tissue, especially during use on small
structures.

1-4 800-477-2006 • www.dremed.com


warnings:
Use electrosurgery with caution in the presence of internal or external devices such as
pacemakers or pulse generators. Interference produced by the use of electrosurgical devices can
cause devices such as pacemakers to enter an asynchronous mode or can block the
pacemaker effect entirely. Consult the device manufacturer or hospital Cardiology Department for
further information when use of electrosurgical appliances is planned for patients with
cardiac pacemakers or other implantable devices.
If the patient has an Implantable Cardioverter Defibrillator (ICD), contact the ICD manufacturer for
instructions before performing an electrosurgical procedure. Electrosurgery may cause multiple
activation of ICDs.
Do not use electrosurgical equipment unless properly trained to use it in the specific procedure
being undertaken. Use by physicians without such training has resulted in serious, unintended
patient injury, including bowel perforation and unintended, irreversible tissue necrosis.
For surgical procedures where the high frequency current could flow through parts of the body
having a relatively small cross-sectional area, the use of bipolar techniques may be desirable to
avoid unwanted coagulation.
In some circumstances, potential exists for alternate site burns at points of skin contact (e.g.,
between the arm and the side of the body). This occurs when electrosurgical current seeks a path
to the return electrode that includes the skin-to-skin contact point. Current passing through small
skin-to-skin contact points is concentrated and may cause a burn. This is true for grounded,
ground referenced, and isolated output generators.
To reduce the potential for alternate site burns, do one or more of the following:
• Avoid skin-to-skin contact points, such as fingers touching leg, when positioning the patient.
• Place 5 to 8 cm (2 to 3 in.) of dry gauze between contact points to ensure that contact does
not occur.
• Position the return electrode to provide a direct current route between the surgical site and
the return electrode which avoids skin-to-skin contact areas.
• In addition, place patient return electrodes according to the manufacturer’s instructions.
Potential for alternate site burns increases if the return electrode is compromised. DRE, Inc.
recommends the use of split return electrodes and DRE generators with a contact quality
monitoring system.
Do not wrap the accessory cords or return electrode cords around metal objects. This may induce
currents that could lead to shocks, fires, or injury to the patient or surgical team.

User ’s Guide • ASG-300 1-5


Cautions:
At no time should you touch the active electrode or bipolar forceps. A burn could result.
Do not stack equipment on top of the generator or place the generator on top of electrical
equipment. These configurations are unstable and/or do not allow adequate cooling.
Provide as much distance as possible between the electrosurgical generator and other electronic
equipment (such as monitors). An activated electrosurgical generator may cause interference
with them.
Non-function of the generator may cause interruption of surgery. A backup generator should be
available for use.
Do not turn the activation tone down to an inaudible level. The activation tone alerts the surgical
team when an accessory is active.
When using a smoke evacuator in conjunction with the electrosurgical generator, place the smoke
evacuator a distance from the generator and set the generator volume control at a level that
ensures that the activation tones can be heard.
The use of high frequency current can interfere with the function of other electromagnetic
equipment.
When high frequency surgical equipment and physiological monitoring equipment are used
simultaneously on the same patient, place any monitoring electrodes as far as possible from
the surgical electrodes. Monitoring systems incorporating high frequency current-limiting devices
are recommended.
Do not use needles as monitoring electrodes during electrosurgical procedures. Inadvertent
electrosurgical burns may result.
To avoid the possibility of an electrosurgical burn to either the patient or the physicians, do
not allow the patient to come in contact with a grounded metal object during activation. When
activating the unit, do not allow direct skin contact between the patient and the physician.
To avoid the possibility of a burn to the patient, when using a split pad do not activate the unit if
the solid pad indicator is illuminated green or the red alarm indicator remains illuminated red. This
could indicate improper pad placement or a faulty NEM circuit.
Remove any loose fitting jewelry from the patient before activation.
Examine all accessories and connections to the electrosurgical generator before use. Ensure that
the accessories function as intended. Improper connection may result in arcs, sparks, accessory
malfunction, or unintended surgical effects.
When not using active accessories, place them in a holster or in a clean, dry, non-conductive, and
highly visible area not in contact with the patient. Inadvertent contact with the patient may result
in burns.
Studies have shown that smoke generated during electrosurgical procedures can be potentially
harmful to patients and the surgical team. These studies recommend adequately ventilating the
smoke by using a surgical smoke evacuator or other means.1

1. U.S. Department of Health and Human Services. National Institute for Occupational Safety
and Health (NIOSH). Control of Smoke from Laser / Electric Surgical Procedures. HAZARD
CONTROLS, Publication No. 96-128, September, 1996).

Notices:
If required by local codes, connect the generator to the hospital equalization connector with an
equipotential cable.
Do not clean the generator with abrasive cleaning or disinfectant compounds, solvents, or other
materials that could scratch the panels or damage the generator.

1-6 800-477-2006 • www.dremed.com


CONTROLS, INDICATORS, AND RECEPTACLES
This section describes:

● The Front and Rear Panels

● Controls, Indicators, Receptacles, and Ports

User ’s Guide • ASG-300 2-1


FRONT PANEL
Figure 2 – 1 Layout of controls, indicators, and receptacles on the front panel

2-2 800-477-2006 • www.dremed.com


Symbols on the Front Panel
Refer to the following table for descriptions of symbols found on the front panel of the ASG-300.

SYMBOLS DESCRIPTION

Cut Controls

Cut Mode

Blend Mode

Coag Controls

Pinpoint Mode

Spray Mode

Bipolar Controls

Bipolar Mode

Indicators

Split Return Electrode

Solid Return Electrode

Regulatory Symbology

Read instructions before use.

Defibrillator Proof Type CF Equipment

RF Isolated – patient connections are isolated from earth at high frequency.

Power Switch and Handpiece Connectors

Return Electrode Receptacle

Caution High Voltage

Cut Mode

Coag Mode

Monopolar Handpiece Receptacle

Bipolar Mode

Bipolar Handpiece Receptacle

User ’s Guide • ASG-300 2-3


PRESET CONTROLS
Figure 2 – 2 Controls for setting and recalling presets

Preset Number Display


Indicates the current selection of
one of the 10 presets (0-9).

Recall Button
Toggles through the 10 presets. Stop
at the desired number (0-9) illumi-
nated in the Preset Number Display
to recall a Preset.

Set Button
Sets the desired preset into one of
the 10 user-defined presets. Press Preset Indicator LED
and hold the Set button for three Blinks in the lower right corner of
seconds to save the settings. the Preset display to indicate that
the current setting is not one of the
user-defined presets.

NOTICES:
The ASG-300 incorporates 10 factory-set presets that are all set to zero and can be reset to your
preferred settings.
Set and Recall are disabled while the unit is activated.

2-4 800-477-2006 • www.dremed.com


CUT AND BLEND CONTROLS
Figure 2 – 3 Controls for the Cut and Blend modes

Cut Power Display (watts) Cut Power Control Buttons


Blend Amount Control Buttons Indicates the power set for Increases or decreases the Cut or
Increases or decreases the the Cut / Blend mode. Blend power output in increments
amount of blend (Level 1-10) of 1 to 10 watts.
added in the Blend mode.

Cut I Mode Indicator


Indicates when the
Cut I mode is selected.

Cut II Mode Indicator


Indicates when the
Cut II mode is selected.

Cut and Blend


Cut Activation Indicator Mode Selector
Illuminates when Cut or Toggles between
Blend Amount Indicator Blend mode is activated. Cut I, Cut II, and
Indicates the amount of Blend modes.
blend added in the Blend
mode. More bars illuminated
indicates more blend, divided
into 10 steps.
Blend Mode Indicator
Indicates when the Blend
mode is selected.

NOTICE:
When selecting the Blend mode, the unit defaults to a setting of minimum blend (only the first bar
is illuminated).

User ’s Guide • ASG-300 2-5


COAG CONTROLS
Figure 2 – 4 Controls for the Coag mode

Coag Power Display (watts)


Indicates the power set for Coag Power Control Buttons
the Coag mode. Increases or decreases the Pinpoint
or Spray Coag power output in
increments of 1 to 10 watts.

Coag Activation Indicator


Illuminates when Coag
mode is activated.
Pinpoint Mode Indicator
Indicates when the Pinpoint
mode is selected.

Pinpoint and Spray


Mode Selector
Toggles between
Pinpoint mode and
Spray mode.

Spray Mode Indicator


Indicates when the Spray
mode is selected.

2-6 800-477-2006 • www.dremed.com


BIPOLAR CONTROLS
Figure 2 – 5 Controls for the Bipolar mode

Bipolar Power Display (watts)


Indicates the power set for the
Bipolar mode.
Displays error code in the event Bipolar Power Control Buttons
of an error. Increases or decreases the Bipolar
power output in increments of 1 to
5 watts.

Bipolar Activation Indicator


Illuminates when Bipolar mode
is activated.

User ’s Guide • ASG-300 2-7


INDICATORS
Figure 2 – 6 Indicators for power, return electrodes, and footswitch control

Monopolar Footswitch Indicator Bipolar Footswitch Indicator


Illuminates when monopolar Illuminates when bipolar
footswitch control is plugged footswitch control is plugged
in and available. in and available.

Power Indicator
Illuminates when the
Split Return Solid Return Alarm Indicator
main power is on.
Electrode Indicator Electrode Indicator Illuminates when the system
Illuminates green Illuminates green detects a return electrode
when the system when the system alarm condition.
detects a split plate is detects a single plate.
properly placed on
the patient. Solid pad indicator
only detects that a pad
is connected to the
unit. The unit does not
monitor pad placement
on the patient.

2-8 800-477-2006 • www.dremed.com


POWER SWITCH AND RECEPTACLES
Figure 2 – 7 Location of the unit power switch and front panel receptacles

Return Electrode Monopolar Handswitching Receptacle


Receptacle Accepts standard 3-pin handpieces.
Accepts a standard Connect handswitching accessories.
return electrode plug.

Power On/Off Switch Monopolar Footswitching Receptacle Bipolar Receptacle


Turns the unit on or off. Accepts cables or adapters equipped Accepts standard cables for
with standard (Bovie #12) active plugs. bipolar handpieces. Connect
Connect footswitching accessories. bipolar accessories.

User ’s Guide • ASG-300 2-9


REAR PANEL
Figure 2 – 8 Layout of connectors and controls on the rear panel

Symbols on the Rear Panel


Refer to the following table for descriptions of symbols found on the rear panel of the ASG-300.

SYMBOLS DESCRIPTION

Equipotential Ground Stud

Non-ionizing Radiation

Volume Control

Danger - Explosion Risk If Used With Flammable Anesthetics.

Fuse Enclosed

Relay Connector

Monopolar Footswitch Input Jack

Bipolar Footswitch Input Jack

Do Not Dispose of Unit in Municipal Waste Stream.

Manufacturer

Read Instructions Before Use

2-10 800-477-2006 • www.dremed.com


GETTING STARTED
This section includes the following information:

● Initial Inspection

● Installation

● Function Checks

● Performance Checks

User ’s Guide • ASG-300 3-1


InItIal InspectIon
When you first unpack your ASG-300, inspect it visually:

• Look for any signs of damage.

• Verify that the shipping package contains all items listed on the packing list.

If the unit or any accessories are damaged, notify DRE, Inc.’s Customer Service immediately.
Do not use any damaged equipment.

InstallatIon
Place the ASG-300 on any flat surface with a tilt angle not more than 10˚. The unit relies on natural convection cooling. Do not block
its bottom or rear vents. Ensure that air flows freely on all sides of the unit.

Warning:
Connect the power cord to a properly polarized and grounded power source with the frequency
and voltage characteristics that match those listed on the back of the unit.

FunctIon checks
Upon initial installation of the unit, perform the tests listed below. Refer to the figures in the previous chapter for the location of
connectors and controls.

Warning:
At no time should you touch the active electrode or bipolar forceps. A burn could result.

setting up the unit


1. Verify that the Power Switch is in the Off (O) position and that no accessories are connected to the unit.

2. Connect a hospital grade power cable to the AC power cable receptacle on the back of the unit, then to a properly grounded
wall outlet.

3. Connect a two-button monopolar pencil to the appropriate receptacle. The use of Bovie pencils is recommended.

4. Do not connect a patient return electrode at this time.

5. Turn the unit on by switching the power switch to the On (|)position.

checking the Return electrode alarm


1. Adjust the power settings for each mode (Cut, Coag, Bipolar) to one watt.

2. Press the Coag button of the pencil. Verify that an alarm sounds for three seconds and the patient return electrode sensing
alarm indicator light illuminates, indicating that no return electrode is connected to the unit.

3. Verify that adjusting the volume control on the back of the unit while the alarm is sounding does not change the alarm volume.

3-2 800-477-2006 • www.dremed.com


confirming Modes
Confirm that you can select each mode and adjust the power up and down.

checking Bipolar Mode (with bipolar footswitch)


1. Plug in the Bipolar footswitch. Verify that the Bipolar footswitch indicator illuminates.

2. Press the pedal on the Bipolar footswitch. Verify that the Bipolar mode activation indicator illuminates and that the system
generates the Bipolar activation tone.

3. While activating the Bipolar mode, rotate the volume control over the full range to verify that the sound is audible throughout
the range.

4. Confirm that releasing the pedal returns the unit to an idle state.

checking Monopolar Mode (with monopolar footswitch)


1. Plug in the Monopolar footswitch. Verify that the monopolar footswitch indicator illuminates.

2. Connect a solid return electrode to the return electrode receptacle. Verify that the green solid return electrode indicator
illuminates.

3. Press the Cut pedal (yellow) on the footswitch. Verify that the Cut mode activation indicator illuminates and that the system
generates the Cut activation tone.

4. While activating the Cut mode, rotate the volume control over the full range to verify that the sound is audible throughout
the range.

5. Press the Coag pedal (blue) on the footswitch. Verify that the Coag mode activation indicator illuminates and that the system
generates the Coag activation tone.

6. While activating the Coag mode, rotate the volume control over the full range to verify that the sound is audible throughout
the range.

checking Monopolar Mode (with handswitch)


1. Connect a handswitching handpiece to the Monopolar handpiece receptacle.

2. Connect a solid return electrode to the return electrode receptacle. Verify that the green solid return electrode indicator
illuminates.

3. Activate, one at a time, the Cut and Coag handswitching controls. Verify that each control causes the correct indicator and tone to
sound.

peRFoRMance checks
After the unit has passed the preliminary functional test, it is ready for performance testing. A qualified biomedical engineer who
is thoroughly familiar with electrosurgical devices should conduct this testing. The testing should include checking all modes of
operation for proper function and power output.

User ’s Guide • ASG-300 3-3


3-4 800-477-2006 • www.dremed.com
USING THE ASG-300
This section contains the following procedures:

● Inspecting the Generator and Accessories

● Setup Safety

● Setting Up

● Preparing for Monopolar Surgery

● Preparing for Bipolar Surgery

● Setting and Recalling Memory Presets

● Activating the Unit

● Activation Safety

Cautions:
Read all warnings, cautions, and instructions provided with this generator before use.
Read the instructions, warnings, and cautions provided with electrosurgical accessories before
use. Specific instructions are not included in this manual.

User ’s Guide • ASG-300 4-1


InspectIng the generator and accessorIes
Before each use of the ASG-300, verify that the unit and all accessories are in good working order:

• Inspect for damage to the Electrosurgical Generator and all its connections.

• Verify that the appropriate accessories and adapters are present.

• Inspect all cords and connectors for signs of wear, damage, and abrasion.

• Verify that no errors occur when you turn on the unit.

setup safety
Warnings:
HazardousElectricaloutput - This equipment is for use only by trained, licensed physicians.
ElectricshockHazard - Connect the generator power cord to a properly grounded receptacle.
Do not use power plug adapters.
Connect the power cord to a properly polarized and grounded power source with the frequency
and voltage characteristics that match those listed on the back of the unit.
FireHazard - Do not use extension cords.
Patientsafety - Use the generator only if the self-test has been completed as described.
Otherwise, inaccurate power outputs may result.
The instrument receptacles on this generator are designed to accept only one instrument at a
time. Do not attempt to connect more than one instrument at a time into a given receptacle. Doing
so will cause simultaneous activation of the instruments.
Failure of the high frequency electrosurgical equipment could result in an unintended increase of
output power.
Do not use electrosurgical equipment unless properly trained to use it in the specific procedure
being undertaken. Use by physicians without such training has resulted in serious, unintended
patient injury, including bowel perforation and unintended, irreversible tissue necrosis.
For surgical procedures where the high frequency current could flow through parts of the body
having a relatively small cross-sectional area, the use of bipolar techniques may be desirable to
avoid unwanted coagulation.
If the patient has an Implantable Cardioverter Defibrillator (ICD), contact the ICD manufacturer for
instructions before performing an electrosurgical procedure. Electrosurgery may cause multiple
activation of ICDs.
In some circumstances, potential exists for alternate site burns at points of skin contact (e.g.,
between the arm and the side of the body). This occurs when electrosurgical current seeks a path
to the patient return electrode that includes the skin-to-skin contact point. Current passing through
small skin-to-skin contact points is concentrated and may cause a burn. This is true for grounded,
ground referenced, and isolated output generators.
To reduce the potential for alternate site burns, do one or more of the following:
• Avoid skin-to-skin contact points, such as fingers touching leg, when positioning the patient.
• Place 5 to 8 cm (2 to 3 in.) of dry gauze between contact points to ensure that contact does
not occur.
• Position the return electrode to provide a direct current route between the surgical site and
the return electrode which avoids skin-to-skin contact areas.
• In addition, place return electrodes according to the manufacturer’s instructions.
Potential for alternate site burns increases if the return electrode is compromised. DRE, Inc.
recommends the use of split return electrodes and generators with a contact quality
monitoring system.

4-2 800-477-2006 • www.dremed.com


Cautions:
Do not stack equipment on top of the generator or place the generator on top of electrical
equipment. These configurations are unstable and/or do not allow adequate cooling.
Provide as much distance as possible between the electrosurgical generator and other electronic
equipment (such as monitors). An activated electrosurgical generator may cause interference with
them.
Non-function of the generator may cause interruption of surgery. A backup generator should be
available for use.
Do not turn the activation tone down to an inaudible level. The activation tone alerts the surgical
team when an accessory is active.
When using a smoke evacuator in conjunction with the electrosurgical generator, place the smoke
evacuator a distance from the generator and set the generator volume control at a level that
ensures that the activation tones can be heard.

notIce:
If required by local codes, connect the generator to the hospital equalization connector with an
equipotential cable.

settIng up
1. Verify that the generator is Off by pressing the power switch Off (O).

2. Place the generator on a stable flat surface, such as a table, platform, or medical cart. Carts with conductive wheels are
recommended. For details, refer to the procedures for your institution or to local codes. Provide at least 10 to 15 cm (4 to 6 in.)
of space from the sides and top of the generator for cooling. Normally, the top, sides, and rear panel are warm when you use the
generator continuously for extended periods of time.

3. Plug the generator power cord into the AC Power Cable Receptacle on the rear panel.

4. Plug the generator power cord into a grounded receptacle.

5. Turn on the generator by pressing the power switch On (|). Verify the following:

• All visual indicators and displays on the front panel illuminate.

• Activation tones sound to verify that the speaker is working properly.

6. If the self-test is successful, a tone sounds. Verify the following:

• A Cut mode is selected; a Coag mode is selected.

• Each display shows a power setting. The unit automatically powers up to the last selected preset settings.

• The Patient Return Electrode Alarm Indicator illuminates red.

If the self-test is not successful, an alarm tone sounds. An error code will appear in the Bipolar display, in most cases, the generator
is disabled. Note the error code and refer to Section 6, Troubleshooting.

Once the self-test is successful, connect the accessories and set the generator controls. Refer to Preparing for Monopolar Surgery or
Preparing for Bipolar Surgery later in this section.

User ’s Guide • ASG-300 4-3


preparIng for Monopolar surgery
Monopolar surgery requires a return electrode.

applying the return electrode


To maximize patient safety, DRE, Inc. recommends using a split return electrode and a generator with a contact quality monitoring
system (Bovie NEM™).

notIces:
The Bovie NEM™ system recommends that you use a split return electrode.
Before activation, pad placement and visual verification of the split return electrode (split pad) indi-
cator on the front panel is recommended. After connecting the split pad to the generator and plac-
ing the split pad securely to the patient, give the unit 5 to 10 seconds to recognize the split pad.
The split pad indicator will illuminate green. If the split pad and cord are attached to the generator
without secure contact to the patient, the alarm indicator will illuminate red.

Refer to the manufacturer’s instructions for application site and placement procedures. When using metal plate return electrodes,
use a conductive gel specifically designed for electrosurgery. Select a return electrode site with good blood flow. While a properly
applied electrode results in minimal tissue heating beneath the electrode, a good blood flow helps carry heat away from the site.

1. Connect the cable to the Return Electrode receptacle on the front of the unit.
The unit will automatically sense the presence of a split or solid return electrode and, if a split return electrode is used,
will constantly monitor the resistance at the contact between the electrode and the patient.

2. Adjust the Blend setting to the desired amount of hemostasis (Level 1 - 10). Adjustment is preformed by pressing the up or down
buttons next to the Blend setting indicator.

Select the desired power settings for Cutting. Adjustment is preformed by pressing the up or down buttons next to the Cut display.

Select the mode of operation for Coagulation, either Pinpoint or Spray.

Select the desired power setting for Coagulation. Adjustment is preformed by pressing the up or down buttons next to the
Coag display.

ifyouareusing… Connectitto…

Standard 3-pin handswitching pencil Monopolar handswitching receptacle


Footswitching pencil Monopolar footswitching receptacle

connecting accessories
1. Connect a 3-pin monopolar device into the monopolar receptacle on the front of the unit.

If footswitching control capabilities are preferred, connect the Bovie monopolar footswitch to the appropriate footswitch connecting
socket on the rear of the unit.

To activate the Monopolar mode, depress the cut or coag button on the monopolar handpiece or the cut or coag pedal on the
monopolar footswitch.

4-4 800-477-2006 • www.dremed.com


preparIng for BIpolar surgery
1. Connect a Bipolar cable to the Bipolar receptacle on the front of the unit.

2. Connect a forceps instrument to the bipolar cable.

3. Connect the bipolar footswitch to the bipolar footswitch connecting socket located on the rear of the unit.

To activate the Bipolar mode, depress the pedal on the bipolar footswitch.

settIng and recallIng MeMory presets


The ASG-300 incorporates 10 user-defined memory preset settings for easy recall of frequently used settings in all
three modes.

Memory
The Memory feature allows the ASG-300 (unit) to display the last selected Preset when the generator is turned on. When activated
by the handpiece or footswitch, the unit will operate in that particular mode and power setting.

The small red blinking dot in the lower right hand corner of the Preset display lets the user know that the Preset values have been
adjusted.

All new settings must be saved as a Preset to be available at startup or as a Preset selection (0 through 9) when using the unit.

Memory function overview


• The unit powers up with the last selected preset (0-9).

• Mode (Cut and Coag) membrane switches are disabled during activation.

• Blend amount control buttons are disabled during activation.

• Recall and Set membrane switches are disabled during activation.

• During activation, the activated mode can be adjusted up and or down a maximum of four steps. Refer to the following table
for power increments.

POWER SETTINGS INCREMENTS Example 1...


1-50 Watts 1 Watts The unit is activated using the same preset values as described in
Example 1 of this section.
50-100 Watts 2 Watts
While activated, the Cut 1 power output of 30 watts can be
100-200 Watts 5 Watts adjusted 4 steps down to 26 watts or 4 steps up to 34 watts. The
Pinpoint and Bipolar can be adjusted to display a different setting
200-300 Watts 10 Watts but can not be saved during activation.

• While operating the unit outside of a user-defined preset (small red dot will be blinking in lower right corner of the Preset
display as an indicator), the unit temporarily stores the power setting for the activated mode (Cut, Coag, or Bipolar).
This temporary power setting is available until either the unit is reset, a preset is selected , or the power setting for the mode in
use is adjusted and the unit is again activated.

• Presets only store one Cut mode (Cut I or Cut II, or Blend) and power setting, one Blend level (if applicable), one Coag mode
(Pinpoint or Spray) and power setting, and Bipolar power setting. When storing, only the information displayed in the display
windows will be saved to the unit’s memory.

User ’s Guide • ASG-300 4-5


setting your presets
Select the desired preset (0-9) by pressing the recall button.

Select the desired modes to be stored by pressing the mode membrane switches (Cut and Coag).

If presetting the Blend mode, select the desired level of hemostasis (Blend Bar 1-10) by pressing the
Blend amount control button.

Select the desired power (Cut, Coag, and Bipolar) to be stored by using the power output up and down membrane switches.

Once all of the settings are selected, depress and hold the Set button for three seconds. To indicate the settings have been
stored, the Preset Memory Number (0-9) will blink.

To recall a Preset, repeatedly press the Recall button to toggle through all of the presets.

notIces:
The ASG-300 incorporates 10 factory-set presets that are all set to zero and can be reset to your
preferred settings.
A small red dot blinking in the lower right corner of the Preset indicator display indicates that the
unit is not presently set to a user-defined preset.
Set and Recall buttons are disabled while the unit is activated.
Presets only store one Cut mode (Cut I or Cut II, or Blend) and power setting, one Blend level (if
applicable), one Coag mode (Pinpoint or Spray) and power setting, and Bipolar power setting.
When storing, only the information displayed in the display windows will be saved to the unit’s
memory.

Memory feature (last selected preset)


The Memory feature allows the unit to display the last selected power preset when the generator is turned on.

notIce:
To have a setting selection available at startup or to be one of the 10 user-defined presets, the
adjustment to the mode and/or power settings must be saved by pressing the Set button on the
Preset display panel.
Examples
Examples 1 through 4 explain how the Memory and temporary memory features work and what happens when the power and/or
mode is adjusted but not saved as one of the 10 Preset selections. Example 5 explains what happens when the power and/or mode is
adjusted and saved as a new Preset setting:

#2. The physician performs a surgical procedure using Preset 2. The Preset has been stored with the following mode and power:

• The mode is set to Cut I


• The power setting for Cut I is 30 watts
• The power setting for Pinpoint is 15 watts
• The power setting for Bipolar is 20 watts.

The procedure is completed and the unit is switched off.


The next time the unit is switched on, the number 2 Preset will be displayed and available when the unit is activated. The
number 2 Preset will be the same as the modes and settings indicated above.

#3. The physician performs a surgical procedure using Preset 2 (same as Example #1 Preset values).
He adjusts the power settings for each mode but does not store the new settings into the Preset.
The next time the unit is switched on, the number 2 Preset will be displayed and available when the unit is activated. The
number 2 Preset will be the same as the modes and settings indicated in Example #1.

4-6 800-477-2006 • www.dremed.com


#4. The physician performs a surgical procedure using Preset 2 (same as Example #1 Preset values).
He changes the settings by selecting the Cut II mode. The displayed power will remain at 30 watts. The physician then adjusts the
power to 100 watts. He resumes the procedure now using Cut II at 100 watts. He then switches the mode back to Cut I. The power
output returns to 30 watts as stored in the # 2 Preset. The physician switches again to the Cut II mode and the output power
returns to the temporary memory of 100 watts as previously selected. The procedure is completed without saving any modes or
power settings. The next time the unit is switched on, the number 2 Preset will be displayed and available when the unit is
activated. The number 2 Preset will be the same as the modes and settings indicated in Example #1.

#5. The physician performs a surgical procedure using Preset 2 (same as Example #1 Preset values.).
As required in the procedure, he selects the Blend mode (the Blend Amount Indicator illuminates to one bar indicating the Blend
mode can be increased to the preferred amount of blend). He adjusts the hemostasis level up to a 30% blend but does not store
the new settings into the Preset. The next time the unit is switched on, the number 2 Preset will be displayed and available when
the unit is activated. The number 2 Preset will be the same as the modes and settings indicated in Example #1.

#6. The physician performs a surgical procedure using Preset 2 (same as Example #1 Preset values). He adjusts the power settings
for a Cut mode, a Coag mode, and a Bipolar mode and presses the Store button for three seconds to save the new settings as
Preset number 2. The next time the unit is switched on, the number 2 Preset will be displayed and available when the unit is
activated. The number 2 Preset will now be the last saved Preset settings for Preset 2.

actIvatIng the unIt


notIce:
Review Activation Safety on page 6 of this section before activating the unit. When you turn on
your unit remember the following feature:
The ASG-300 will power up to the modes and settings displayed when the unit was last
activated. For example, if you set Cut I mode at 50 watts and activate the unit, then turn the unit
off, it will automatically return to Cut I mode at 50 watts when you turn it on again. Similarly, if you
set Pinpoint mode at 40 watts and activate the unit before you turn it off, it will return to Pinpoint
mode at 40 watts when you turn it on again.

1. Monopolar Cut - select the mode of operation for Cut: Cut I, Cut II, or Blend then select the desired Cut power settings by pressing
the up and down buttons next to the Cut power output display.

2. If using Blend, vary the Blend setting by pressing the up and down buttons next to the blend amount indicator graph.

3. Monopolar Coag - select the mode of operation for coagulation: Pinpoint or Spray, then select the coagulation power settings by
pressing the up and down buttons next to the Coag power output display.

4. Bipolar - adjust the Bipolar power settings by pressing the up and down buttons next to the Bipolar power output display.

5. Activate the generator by pressing the appropriate button on the handpiece or pedal on the footswitch.

notIce:
Monopolar and bipolar footswitching operations are controlled by independent foot controls.

User ’s Guide • ASG-300 4-7


actIvatIon safety
Warnings:
Do not wrap the accessory cords or patient return electrode cords around metal objects. This may
induce currents that could lead to shocks, fires, or injury to the patient or surgical team.
Danger:Fire/ExplosionHazard - Do not use the ASG-300 in the presence of flammable anes-
thetics.
Fire/ExplosionHazard - The following substances will contribute to increased fire and explosion
hazards in the operating room:
• Flammable substances (such as alcohol based skin prepping agents and tinctures)
• Naturally occurring flammable gases that may accumulate in body cavities such as the bowel
• Oxygen enriched atmospheres
• Oxidizing agents (such as nitrous oxide [N 2O] atmospheres).
The sparking and heating associated with electrosurgery can provide an ignition source. Observe
fire precautions at all times. When using electrosurgery in the same room with any of these
substances or gases, prevent their accumulation or pooling under surgical drapes, or within the
area where electrosurgery is performed.
Use the lowest output setting necessary to achieve the desired surgical effect. Use the active
electrode only for the minimum time necessary in order to lessen the possibility of unintended
burn injury. Pediatric applications and/or procedures performed on small anatomic structures may
require reduced power settings. The higher the current flow, and the longer the current is applied,
the greater the possibility of unintended thermal damage to tissue, especially during use on small
structures.
Use electrosurgery with caution in the presence of internal or external devices such as
pacemakers or pulse generators. Interference produced by the use of electrosurgical devices can
cause devices such as pacemakers to enter an asynchronous mode or can block the
pacemaker effect entirely. Consult the device manufacturer or hospital Cardiology Department for
further information when use of electrosurgical appliances is planned for patients with
cardiac pacemakers or other implantable devices.

Cautions:
The use of high frequency current can interfere with the function of other electromagnetic
equipment.
When high frequency surgical equipment and physiological monitoring equipment are used
simultaneously on the same patient, place any monitoring electrodes as far as possible from
the surgical electrodes.
Do not use needles as monitoring electrodes during electrosurgical procedures. Inadvertent
electrosurgical burns may result.
To avoid the possibility of an electrosurgical burn to either the patient or the physicians, do not
allow the patient to come in contact with a grounded metal object during activation. When
activating the unit, do not allow direct skin contact between the patient and the physician.
To avoid the possibility of a burn to the patient, when using a split pad do not activate the unit if
the solid pad indicator is illuminated green or the red alarm indicator remains illuminated red. This
could indicate improper pad placement or a faulty NEM circuit.
Remove any jewelry from the patient before activation.
Studies have shown that smoke generated during electrosurgical procedures can be potentially
harmful to patients and the surgical team. These studies recommend adequately ventilating the
smoke by using a surgical smoke evacuator or other means.1
Examine all accessories and connections to the electrosurgical generator before use. Ensure that
the accessories function as intended. Improper connection may result in arcs, sparks, accessory
malfunction, or unintended surgical effects.

4-8 800-477-2006 • www.dremed.com


When not using active accessories, place them in a holster or in a clean, dry, non-conductive, and
highly visible area not in contact with the patient. Inadvertent contact with the patient may result
in burns.
1. U.S. Department of Health and Human Services. National Institute for Occupational Safety
and Health (NIOSH). Control of Smoke from Laser / Electric Surgical Procedures. HAZARD
CONTROLS, Publication No. 96-128, September, 1996.)

User ’s Guide • ASG-300 4-9


4-10 800-477-2006 • www.dremed.com
MAINTAINING THE ASG-300
This section covers the following topics:

● Cleaning

● Periodic Inspection

● Fuse Replacement

User ’s Guide • ASG-300 5-1


DRE, Inc. recommends that you complete periodic inspection and performance testing. Perform inspections and
performance testing every six months. A qualified biomedical technician should conduct this testing to ensure that the unit
is operating effectively and safely.

CLEANING
After each use, clean the unit.

WARNING:
Electric Shock Hazard - Always turn off and unplug the generator before cleaning.

NOTICE:
Do not clean the generator with abrasive cleaning or disinfectant compounds, solvents, or other
materials that could scratch the panels or damage the generator.

1. Turn off the generator, and unplug the power cord from the wall outlet.

2. Thoroughly wipe all surfaces of the generator and power cord with a mild cleaning solution or disinfectant and a damp cloth.
Follow the procedures approved by your institution or use a validated infection control procedure. Do not allow fluids to enter the
chassis. Do not sterilize the generator.

PERIODIC INSPECTION
Every six months, visually inspect the ASG-300 for signs of wear or damage.
In particular, look for any of the following problems:

• Damage to the power cord


• Damage to the power cable receptacle
• Obvious damage to the unit
• Damage to any receptacle
• Accumulation of lint or debris in or around the unit

FUSE REPLACEMENT
Fuses for the unit reside directly below the Power Cable Receptacle on the rear of the unit.

To replace the fuses, follow this procedure:


1. Unplug the power cord from the wall outlet.
2. Remove the power cord from the Power Cable Receptacle on the rear panel.
3. To release the fuse drawer, insert a small flathead screwdriver into the slot
on the drawer below the power cord receptacle. Then, slide the drawer out.
4. Remove the two fuses (T6.3AL250V) and replace them with new fuses with the
same values.
5. Insert the fuse holder into the Power Cable Receptacle.

NOTICE: Figure 5 – 1 Fuse holder


If the unit does not display an error and does not power on, check fuses.

5-2 800-477-2006 • www.dremed.com


TROUBLESHOOTING
This section includes Error Code Descriptions and actions to take to resolve them.

User ’s Guide • ASG-300 6-1


The ASG-300 includes automatic self-diagnostics. If the diagnostics detect an error, the system displays an error code, sounds an
audible tone, and deactivates the unit output power.

Most error codes result from faults in accessories attached to the unit. The following table lists the error codes, describes the errors,
and recommends actions to take to resolve the errors.

All error codes are displayed in the Bipolar display.

If the unit displays any other error code, it requires service.


Error Code Description Recommended Action
F1 Cut handpiece button may be stuck
1. Turn off, then turn on the generator. Do not press buttons or activate
Coag handpiece button
F2 accessory devices during the self-test.
may be stuck
2. If the error code reappears, disconnect all accessories.
F3 Cut footswitch pedal may be stuck Turn off, then turn on the generator again.
3. If the problem persists, replace the handpiece or footswitch
F4 Coag footswitch pedal
and repeat the restart.
may be stuck
4. If the error code reappears, record the number and call
F5 Bipolar footswitch pedal customer service.
may be stuck
The unit does not allow simultaneous activation of the cut and
coagulation modes. The activation mode is “first come, first serve.”
This means that whichever mode is selected first will be the function
the unit is activated to dispense. An example of this functionality
includes, when the handpiece Cut button is pressed, the unit is activated
for Cut. If a footswitch is simultaneously pressed for Coag, the unit will
F6 Simultaneous activation error continue in the Cut mode as long as the handpiece Cut button is
pressed. If the Cut button is released, the unit will sense an error
and both functions will be disabled.
1. Release either the cut or coag button on the handpiece,
or the cut or coag pedal on the footswitch.
2. If the error code reappears, record the number and contact
customer service.

E1 Output current out of specification


1. Turn the unit off.
E2 Output current sensors delta error 2. Turn the unit on.
3. If the error code reappears, record the number and contact
E3 Output voltage sensors delta error customer service.
E4 System power supply voltages error
E5 1. Turn the unit off.
2. Allow the unit to cool for 20 minutes.
Internal temperature of a section of
E6 3. Turn the unit on.
the unit exceeded the limit.
4. If the error code reappears, record the number and contact
E7 customer service.
1. Turn the unit off.
E8 2. Turn the unit on.
NEM circuit error
3. If the error code reappears, record the number and contact
customer service.

Notice:
If the unit does not power on to display an error, check fuses as described in Section 5 of
this guide.

6-2 800-477-2006 • www.dremed.com


REPAIR POLICY AND PROCEDURES
Refer to this section for information on:

● Responsibility of the Manufacturer

● Returning the Generator for Service

User ’s Guide • ASG-300 7-1


RESPONSIBILITY OF THE MANUFACTURER
DRE, Inc. is responsible for the safety, reliability, and performance of the generator only under the following circumstances:

• The user has followed the Installation and Setup Procedures in this User’s manual.
• Persons authorized by DRE, Inc. performed assembly operation, readjustments, modifications,
or repairs.
• The electrical installation of the relevant room complies with local codes and regulatory requirements, such as
IEC and BSI.
• Equipment use is in accordance with the DRE, Inc. instructions for use.

Please note that infected medical devices must be disposed of as medical/biohazard waste and cannot be included in used
electronic equipment disposal/recycling programs. In addition, certain electronic products must be returned directly to distributor.
Contact your DRE, Inc. representative for return instructions.

For warranty information, refer to APPENDIX B - WARRANTY.

RETURNING THE GENERATOR FOR SERVICE


Before you return the generator, call your DRE, Inc. representative for assistance. If instructed to send the generator to DRE, first
obtain a Return Merchandise Authorization Number. Then clean the Generator and ship it as instructed for service.

Step 1 – Obtain a Return Merchandise Authorization Number


Call your DRE representative (1800-477-2006) to obtain a Return Merchandise Authorization Number. Have the following
information ready when you call:
• Serial number
• Hospital / clinic name / customer number
• Description of the problem
• Telephone number
• Type of repair to be done
• Department / address, city, state, and zip code
• P.O. number
• Model number

Step 2 – Clean the Generator


WARNING:
Electric Shock Hazard - Always turn off and unplug the generator before cleaning.

NOTICE:
Do not clean the generator with abrasive cleaning or disinfectant compounds, solvents, or other
materials that could scratch the panels or damage the generator.

A. Turn off the generator, and unplug the power cord from the wall outlet.

B. Thoroughly wipe all surfaces of the generator and power cord with a mild cleaning solution or disinfectant and a damp cloth.
Follow the procedures approved by your institution or use a validated infection control procedure. Do not allow fluids to enter the
chassis. You cannot sterilize the generator.

Step 3 – Ship the Generator


A. Attach a tag to the generator that includes the Return Merchandise Authorization Number and the information
(hospital, phone number, etc.) listed in Step 1 – Obtain a Return Merchandise Authorization (RMA) Number.

B. Be sure the generator is completely dry before you pack it for shipment. Package it in its original shipping container, if available.
Please be sure to include a reference to the RMA Number on the outside of the box/container.

C. Ship the generator, prepaid, to the address given to you by the DRE representative.

7-2 800-477-2006 • www.dremed.com


TECHNICAL SPECIFICATIONS
All specifications are nominal and subject to change without notice. A specification referred to as “typical” is within ± 20% of a
stated value at room temperature (25° C / 77° F) and a nominal input power voltage.

User ’s Guide • ASG-300 A-1


PERFORMANCE CHARACTERISTICS
Input Power
Input Voltage 100-240 ~ VAC ± 10%
Mains line frequency range (nominal): 50 – 60 Hz
Power consumption: 560 VA
Fuses (two): 6.3 A (slow blow)

Duty Cycle
Under maximum power settings and rated load conditions (Pure Cut, 300 watt @ 300 ohm load), the generator is suitable for
activation times of 10 seconds ON followed by 30 seconds OFF for one hour.

The internal temperature of the unit is continuously monitored. If the temperature rises above 850 C, the alarm will sound and
output power will be deactivated.

Dimensions and Weight


Width 31.1 cm (12.25 in.) Depth 41.3 cm (16.25 in.)
Height 15.3 cm (6.00 in.) Weight < 8.75 kg (< 19 lbs)

Operating Parameters
Ambient temperature range 10° to 40° C
Relative humidity 30% to 75%, non-condensing
Atmospheric pressure 70kPa to 106kPa
If transported or stored at temperatures outside the operating temperature range,
Warm-up time
allow one hour for the generator to reach room temperature before use.

Transport and Storage


Generator should fit on all standard Carts for monopolar generators. The device should be stored and used in a room temperature of
approximately 770 F/250 C.
Ambient temperature range -40˚ to +70˚ C
Relative humidity 10% to 100%, including condensation
Atmospheric pressure 50kPa to 106kPa

A-2 800-477-2006 • www.dremed.com


Audio Volume
The audio levels stated below are for activation tones (cut, coag, and bipolar) and alarm tones (return electrode and system alarms)
at a distance of one meter. Alarm tones meet the requirements for IEC 60601-2-2.

Activation Tone
Volume (adjustable) 40 to 65 dB
Frequency Cut I: 610 Hz
Cut II: 610 Hz
Blend: 610 Hz
Pinpoint: 910 Hz
Spray: 910 Hz
Bipolar: 910 Hz
Duration Continuous while the generator is activated

Alarm Tone
Volume (not adjustable) 70 dB ± 5 dB
Frequency 2 kHz 1⁄2 seconds / 1 kHz 1⁄2 seconds
Duration 2 seconds

Return Electrode Sensing


The system presents audible and visible alarms when it senses no return electrode.
Trip resistance: 0 Ω to 5 Ω ± 3 Ω
Continuous measurement:
Solid Once the system establishes the solid return electrode resistance, an
increase to 20 Ω ± 5 Ω in resistance will cause an alarm. When the
alarm condition exists, the system deactivates output power.

Trip resistance: 10 Ω ± 5 Ω to 135 Ω ± 10 Ω


Continuous measurement:
Split Once the system establishes the split return electrode resistance, an
increase of 40% in resistance will cause an alarm. When the alarm
condition exists, the system deactivates output power.

Low Frequency (50-60 Hz) Leakage Current


Enclosure source current, ground open < 500 µA
Normal polarity, intact ground: < 10 µA
Source current, patient leads, all outputs Normal polarity, ground open: < 10 µA
Reverse polarity, ground open: < 10 µA
Sink current at high line, all inputs < 10 µA

User ’s Guide • ASG-300 A-3


High Frequency (RF) Leakage Current
Bipolar RF leakage current < 63 mA rms at 80 watts
Monopolar RF leakage current (additional tolerance) < 150 mA rms

STANDARDS AND IEC CLASSIFICATIONS


Class I Equipment (IEC 60601-1)
Accessible conductive parts cannot become live in the event of a basic insulation failure because of the way in which they are
connected to the protective earth conductor.

Type CF Equipment (IEC 60601-1) / Defibrillator Proof


The ASG-300 provides a high degree of protection against electric shock, particularly regarding allowable
leakage currents. It is type CF equipment. Patient connections are isolated from earth and resist the effects of
defibrillator discharge.

Drip Proof (IEC 60601-2-2)


The generator enclosure is constructed so that liquid spillage in normal use does not wet electrical insulation or other components
which, when wet, are likely to affect adversely the safety of the generator.

Electromagnetic Interference
When other equipment is placed on or beneath a ASG-300, the unit can be activated without interference.
The generator minimizes electromagnetic interference to video equipment used in the operating room.

Electromagnetic Compatibility (IEC 60601-1-2 and IEC 60601-2-2)


The ASG-300 complies with the appropriate IEC 60601-1-2 and IEC 60601-2-2 specifications regarding
electromagnetic compatibility.

Voltage Transients (Emergency Generator Mains Transfer)


The ASG-300 operates in a safe manner when the transfer is made between line AC and an emergency generator
voltage source.

A-4 800-477-2006 • www.dremed.com


OUTPUT CHARACTERISTICS
Maximum Output for Monopolar and Bipolar Modes
Power readouts agree with actual power into rated load to within 20% or 5 watts, whichever is greater.
Crest Factor*
Mode Output Power Output Frequency Repetition Rate Vpeak max (Rated Load)
Cut I 300 W @ 300 Ω 490 kHz ± 4.9 kHz N/A 1000V 1.6 ± 20%
Cut II 300 W @ 300 Ω 490 kHz ± 4.9 kHz N/A 1000V 1.6 ± 20%
Blend (Max) 200 W @ 300 Ω 490 kHz ± 4.9 kHz 30 kHz ± 5 kHz 2000V 3.5 ± 20%
Pinpoint 120 W @ 500 Ω 490 kHz ± 4.9 kHz 30 kHz ± 5 kHz 2400V 4.5 ± 20%
Spray 80 W @ 500 Ω 490 kHz ± 4.9 kHz 30 kHz ± 5 kHz 3600V 6.5 ± 20%
Bipolar 80 W @ 150 Ω 490 kHz ± 4.9 kHz N/A 450V 1.6 ± 20%

• an indication of a waveform's ability to coagulate bleeders without a cutting effect.

EMC COMPLIANCE
Special precautions should be taken regarding the ASG-300. Medical Electrical Equipment needs special precautions regarding EMC
and needs to be installed and put into service according to the EMC information provided in this manual.

Understand that only the Accessories supplied with or ordered from DRE Medical should be used with your device. The use of
accessories, transducers, and cables other than those specified, may result in increased Emissions or decreased Immunity of the
ASG-300. The ASG-300 and its accessories are not suitable for interconnection with other equipment.

Portable and mobile RF communications equipment can affect Medical Electrical Equipment. The ASG-300 should not be used
adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the ASG-300 should be observed to
verify normal operation in the configuration in which it will be used.

User ’s Guide • ASG-300 A-5


Recommended separation distances between portable and mobile RF
communications equipment and the ASG-300
The ASG-300 is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the ASG-300 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the ASG-300 as recommended below, according to
the maximum output power of the communications equipment.

Rated maximum output separation distance according to frequency of transmitter


power of transmitter
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
d = [ 3.5]√ P d = [ 3.5]√ P d = [ 7 ]√ P
W V1 E1 E1

0.01 0.12 0.12 0.23


0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23

For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.

Guidance and manufacturer’ s declaration – electromagnetic emissions

The ASG-300 is intended for use in the electromagnetic environment listed below. The customer or
the user of the ASG-300 should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance

The ASG-300 must emit


electromagnetic energy in order
RF Emissions CISPR 11 Group 2 to perform its intended function.
Nearby electronic equipment
may be affected.

RF Emissions CISPR 11 Class A The ASG-300 is suitable for use


in all establishments other than
Harmonic emissions domestic and those directly
Class A connected to the public
IEC 61000-3-2
low-voltage power supply net-
work that supplies buildings used
Voltage fluctuations/flicker
Complies in domestic purposes.
emissions IEC 61000-3-3

A-6 800-477-2006 • www.dremed.com


Guidance and manufacturer’ s declaration – electromagnetic immunity
The ASG-300 is intended for use in the electromagnetic environment listed below. The customer or
the user of the ASG-300 should assure that it is used in such an environment.
IEC 60601
Immunity test Compliance level Electromagnetic environment - guidance
test level

Floors should be wood, concrete


or ceramic tile. If floors are
Electrostatic discharge ±6 kV contact ±6 kV contact
covered with synthetic material,
(ESD) IEC 61000-4-2 ±8 kV air ±8 kV air
the relative humidity should be at
least 30%.
±2 kV for power ±2 kV for power
Electrical fast Mains power quality should be
supply lines supply lines
transient/burst that of a typical commercial or
±1 kV for ±1 kV for
IEC 61000-4-4 hospital environment.
input/output lines input/output lines
±1 kV differential
±1 kV differential Mains power quality should be
mode
Surge IEC 61000-4-5 mode that of a typical commercial or
±2 kV common
±2 kV common hospital environment.
mode

<5 % Ut <5 % Ut
(<95 % dip in Ut) (<95 % dip in Ut)
for 0.5 cycle for 0.5 cycle Mains power quality should be
that of a typical commercial or
<40 % Ut <40 % Ut
Voltage dips, short (<60 % dip in Ut) (<60 % dip in Ut) hospital environment. If the user
interruptions and voltage for 5 cycles of the ASG-300 requires contin-
for 5 cycles
variations on power sup- ued operation during power mains
ply input lines 70 % Ut 70 % Ut interruptions, it is recommended
(<30 % dip in Ut) (<30 % dip in Ut)
IEC 61000-4-11 that the ASG-300 be powered
for 25 cycles for 25 cycles
from an uninterruptible power
<5 % Ut <5 % Ut supply or a battery.
(>95 % dip in Ut) (>95 % dip in Ut)
for 5 sec for 5 sec
Power frequency magnetic fields
Power frequency (50/60 should be at levels characteristic
Hz) magnetic field 3 A/m 3 A/m of a typical location in a typical
IEC 61000-4-8 commercial or hospital
environment.
NOTE Ut is the a.c. mains voltage prior to application of the test level.

User ’s Guide • ASG-300 A-7


Guidance and manufacturer’ s declaration – electromagnetic immunity continued...
IEC 60601
Immunity test Compliance level Electromagnetic environment - guidance
test level
Portable and mobile RF communications
equipment should be used no closer to any part
of the ASG-300, including cables, than the
Conducted RF 3 Vrms recommended separation distance calculated
IEC 61000-4-6 150 kHz to 80 3 Vrms
from the equation applicable to the frequency
MHz
of the transmitter.

Recommended separation distance


d = [ 3.5]√ P
V1
d = [ 3.5]√ P
E1
80 MHz to 800 MHz

d = [7 ]√ P
E1
800 MHz to 2.5 GHz
where P is the maximum output power rating
of the transmitter in watts (W) according to the
3 V/m transmitter manufacturer and d is the
Radiated RF
80 MHz to 3 V/m recommended separation distance in metres
IEC 61000-4-3 (m)
2.5 GHz
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,a
should be less than the compliance level in
each frequency range.b
Interference may occur in the vicinity of
equipment marked with the following
symbol.

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.

a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicated theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location which the ASG-300 is used exceeds the applicable RF compliance level above, the ASG-300
should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the ASG-300.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.

A-8 800-477-2006 • www.dremed.com


OUTPUT POWER CURVES
Figure A–1 illustrates output power delivered to rated load for all available modes. Figure A–2 illustrates the maximum peak volt-
age available at a given power setting and output mode. Figures A–3 through A–9 illustrate specific output power delivered to a
range of load resistances for each mode.

Figure A – 1 Display setting vs. output power in watts at rated load

Figure A – 2 Output voltage (Vpeak) versus power setting at rated load

User ’s Guide • ASG-300 A-9


Figure A – 3 Output power versus impedance for Cut I mode

Figure A – 4 Output power versus impedance for Cut II mode

A-10 800-477-2006 • www.dremed.com


Figure A – 5 Output power versus impedance for Blend Min mode

Figure A – 6 Output power vs impedance for Blend Max mode

User ’s Guide • ASG-300 A-11


Figure A– 7 Output power vs impedance for Pinpoint mode

Figure A – 8 Output power vs impedance for Spray mode

A-12 800-477-2006 • www.dremed.com


Figure A– 9 Output power vs impedance for Bipolar mode

User ’s Guide • ASG-300 A-13


A-14 800-477-2006 • www.dremed.com
WARRANTY
DRE, Inc. warrants each product manufactured by it to be free from defects in material and workmanship under nor-
mal use and service for the period(s) set forth below.

DRE, Inc. obligation under this warranty is limited to the repair or replacement, at its sole option, of any
product, or part thereof, which has been returned to it or its Distributor within the applicable time period shown
below after delivery of the product to the original purchaser, and which examination discloses, to DRE, Inc.’s
satisfaction, that the product is indeed, defective.

This warranty does not apply to any product, or part thereof, which has been repaired or altered outside DRE, Inc.’s
facility in a way so as, in DRE, Inc.’s judgment, to affect its stability or reliability, or which has been
subjected to misuse, neglect, or accident.

The warranty periods for DRE, Inc. products are as follows:

• Electrosurgical Generators: Two years after delivery of the product to the original purchaser

• Mounting Fixtures (all models): Two years after delivery of the product to the original purchaser

• Footswitches (all models): Ninety days after delivery of the product to the original purchaser

• Patient Return Electrodes: Shelf life only as stated on packaging

• Sterile Single Use Accessories: Only as stated on packaging

• Handpiece: Only as stated on packaging

User ’s Guide • ASG-300 B-1


This warranty is in lieu of all other warranties, express or implied, including without limitation, the warranties of
merchantability and fitness for a particular purpose, and of all other obligations or liabilities on the part of DRE, Inc.

DRE, Inc. neither assumes nor authorizes any other person to assume for it any other liability in connection with the
sale or use of any of DRE, Inc.’s products.

Notwithstanding any other provision herein or in any other document or communication, DRE, Inc.’s liability with
respect to this agreement and products sold hereunder shall be limited to the aggregate purchase price for the goods
sold by DRE, Inc. to the customer.

DRE, Inc. disclaims any liability hereunder or elsewhere in connection with the sale of this product, for indirect or
consequential damages.

This warranty and the rights and obligations hereunder shall be construed under and governed by the laws of the
State of Florida, USA.

The sole forum for resolving disputes arising under or relating in any way to this warranty is the District Court of the
County of Pinellas, State of Florida, USA.

DRE, Inc., its dealers, and representatives reserve the right to make changes in equipment built and/or sold by them
at any time without incurring any obligation to make the same or similar changes on equipment previously built
and/or sold by them.

B-2 800-477-2006 • www.dremed.com

You might also like