Testing System For Testing Electrical Safety: GM-610 User Manual

This user manual provides instructions on operating and using the GM-610, a testing system for testing the electrical safety of electrical devices. The manual details the technical specifications, functions, standard tests, operation instructions, and service/calibration of the GM-610.

Uploaded by

U CDP
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
292 views

Testing System For Testing Electrical Safety: GM-610 User Manual

This user manual provides instructions on operating and using the GM-610, a testing system for testing the electrical safety of electrical devices. The manual details the technical specifications, functions, standard tests, operation instructions, and service/calibration of the GM-610.

Uploaded by

U CDP
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 35

S.P.L.

GM-610 User manual

User manual English

GM-610
FW Version 5

Testing system for testing electrical safety


Instructions for the user: This manual is supplied in the form of a loose-leaf
publication. Any necessary updates and additional pages can thus be included
without difficulty later so that you always have the latest version of the manual. We
provide you with supplements and/or changes to your manual along with technical
modifications that are made to your testing system.

The DIN/VDE symbols and drawings shown in this manual have been reproduced
with the permission of the Deutsches Institut für Normung e. V. (DIN: German
Institute for Standardization) and the Verband Deutscher Elektrotechniker e. V.
(VDE: German Association of Electrical Engineers). In applying the relevant
standards, it is important to use the versions with the most recent date of issue.

We reserve the right to make any changes to the information in this manual without
prior notice. S.P.L. Elektronik accepts no responsibility in this regard. The software
that comes with this product is supplied within the framework of a License
Agreement. The software may be used and copied only in accordance with the
terms and conditions of the License Agreement.

All rights pertaining to reprinting and reproducing the manual, or parts thereof, are
reserved. No part of the manual may be reproduced in any form whatsoever
(photocopy, microfilm or any other procedure) or processed, reproduced or
distributed using electronic systems without prior written approval from S.P.L.
Elektronik.

Other products mentioned by name are trademarks or registered trademarks of


their respective companies.

S.P.L. Elektronik
Forstweg 1
D - 14656 Brieselang
GERMANY

Tel. +49 33232 39915


Fax +49 33232 39911

10th Edition, 2020

1
S.P.L. GM-610 User manual

European Union Declaration of Conformity


Nr. 20150505102
Hersteller: S.P.L. Elektronik, Forstweg 1, 14656 Brieselang
Bundesrepublik Deutschland
Produktbeschreibung: Mess- und Prüfgerät zur Überprüfung der elektrischen
Sicherheit von elektrischen Geräten
Modell: GM-610
Konform mit den Bestimmungen folgender EU-Richtlinien:
Richtlinie 2014/35/EU des Europäischen Parlaments und des Rates
vom 26. Februar 2014 (Niederspannungsrichtlinie)
Richtlinie 2011/65/EU des Europäischen Parlaments und des Rates
vom 8. Juni 2011 zur Beschränkung der Verwendung bestimmter gefährlicher Stoffe
in Elektro- und Elektronikgeräten (RoHS Richtlinie)
Richtlinie 2014/30/EU des Europäischen Parlaments und des Rates
vom 26. Februar 2014 (Elektromagnetische Verträglichkeit)

Angewendete Normen und technische Spezifikationen:


 EN 61010-1:2011-07,
Sicherheitsanforderungen an elektrische Mess-, Steuer-, Regel-, und Laborgeräte,
Teil1 Allgemeine Anforderungen (IEC 61010-1:2010 + Cor. :2011)
 VDE 0404-1:2002-05,
Prüf- und Messeinrichtungen zum Prüfen der elektrischen Sicherheit von
elektrischen Geräten, Teil 1: Allgemeine Anforderungen
 VDE 0404-2:2002-05,
Prüf- und Messeinrichtungen zum Prüfen der elektrischen Sicherheit von
elektrischen Geräten, Teil 2: Prüfeinrichtungen für Prüfungen nach Instandsetzung,
Änderung oder für Wiederholungsprüfungen
 VDE 0404-3:2005-04,
Prüf- und Messeinrichtungen zum Prüfen der elektrischen Sicherheit von
elektrischen Geräten, Teil 3: Prüfeinrichtungen für Wiederholungsprüfungen und
Prüfungen vor der Inbetriebnahme von medizinischen elektrischen Geräten oder
Systemen
 EN 61326-1:2013-07
Elektrische Mess-, Steuer-, Regel- und Laborgeräte - EMV-Anforderungen –Teil 1:
Allgemeine Anforderungen (IEC 61326-1:2012); Deutsche Fassung EN 61326-
1:2013
 EN 61000-4-2:2009-12
 EN 61000-4-3:2011-04
 EN 61000-4-4:2013-04
 EN 61000-4-5:2015-03
 EN 61000-4-6:2014-08
 EN 61000-4-7:2009-12
 EN 61000-4-8:2010-11
 EN 61000-4-11:2005-02
Brieselang, den 05.05.2015

Bernd Schulz
S.P.L. Elektronik

2
S.P.L. GM-610 User manual

1 Introduction ...................................................................................................... 5

2 Technical Data .................................................................................................. 7


2.1 Intrinsic uncertainty ................................................................................... 8
2.2 Connections / Accessories ....................................................................... 9
3 Warning signs and symbols .......................................................................... 10

............................................................................................................................ 10

4 Functional description ................................................................................... 11

4.1 Specific features of standard ..................................................................... 12

4.2 Additional functions.................................................................................... 12

4.3 Standardization ........................................................................................... 12

5 Manual ............................................................................................................ 13
5.1 Operation .................................................................................................. 13
5.2 Main menu ................................................................................................ 13
5.3 MULTI ........................................................................................................ 14
5.4 NORM ........................................................................................................ 20
5.4.1 IEC 60601............................................................................................ 20
5.4.2 IEC 62353............................................................................................ 21
5.4.3 IEC 61010............................................................................................ 22
5.4.4 VDE 0701-0702 ................................................................................... 22
5.5 TEST PROTOCOLS .................................................................................. 23
5.6 USER ......................................................................................................... 25
5.7 SETUP ....................................................................................................... 28
5.7.1 Results................................................................................................. 28
5.7.2 Test protocols ...................................................................................... 28
5.7.3 Test lead resistance ............................................................................. 28
5.7.4 GM-610 Diagnose ................................................................................ 29
5.7.5 Tester .................................................................................................. 29
5.7.6 Identification number and comment ..................................................... 29
5.7.7 Automatic / Partautomatic .................................................................... 29
5.7.8 Measure type ....................................................................................... 29
5.7.9 Multi – display result time ..................................................................... 30
5.7.10 Time & Date ....................................................................................... 30
5.7.11 Select your language ......................................................................... 30
5.7.12 GM-610 Info ....................................................................................... 31
5.7.13 Header ............................................................................................... 32
5.7.14 Display brightness.............................................................................. 32
5.7.15 Touchtone .......................................................................................... 32
5.7.16 DEF – restore default values .............................................................. 32
5.8 General Hints............................................................................................ 33
5.9 Short cuts ................................................................................................. 34
5.10 PC-controlled ......................................................................................... 35
6 Service and Calibration ................................................................................. 36

3
S.P.L. GM-610 User manual

1 Introduction

The test appliance GM-610 is a measurement system for testing the electrical
safety of medical technical and other technical appliances. The measurements and
tests correspond to the IEC 60601, IEC 62353 and VDE 0701-0702. The standards
EN 61010 are also supported.

To control the GM-610 in PC operation a 100% compatible IBM computer with


industrial standards is needed. The communication between the PC and the
GM-610 follows through the serial interface and/or USB interface.

With the GM-610 the following tests are realized:

Line voltage measurement


Current measurement
Power measurement
Leakage measurements
Alternative leakage current measurement
Protective earth resistance measurement
Insulation resistance measurement
Difference current measurement

The measuring values of the leakage current measurements become automatic,


suitable of the respective norm calculated to the normative demanded value of the
mains voltage.

With GM-610 the separate ac/dc measurement (RMS), of the patient leakage and
patient auxiliary current can be carried out accordingly to the demands of the
standards.

GM-610 can be used as stand-alone unit or PC operated. With this appliance, you
can do single measurements or you can run automatic test instructions.

The operation of the Safety Tester is touch-controlled and with a few decisions the
different functions of the equipment become activated.

In a non-volatile memory of GM-610 can be stored up to 200 test protocols.


User can create tests with up to 50 of his own test instructions that are saved in a
non-volatile memory.

4
S.P.L. GM-610 User manual

Warning – Safety notice!


When taking measurements to test electrical safety, the device to be tested
is operated with mains voltage. This creates an increased risk of injury from
connected devices waiting to be tested.

With certain measurements, there may be an internally generated mains


voltage or insulation test voltage at the component sockets.

Attention!

The measurement of the leakage current can be measured


only in a TN System.

Attention – Hints for the PC operation

After measurements with main voltage on the socket, the main voltage
will be not switched off. The device to be tested not has
to be disconnected.

You can start functional tests without any delay. The main voltage
can be switched off by using the test step SPL50
at the end of the test sequence.

Attention - Hint!

Note the safety tester annex.

5
S.P.L. GM-610 User manual

2 Technical Data


Line voltage: 230 V / 115 V ac, +/- 10 %, 50 / 60 Hz
Output power: maximal 3,5 kW
Protection class: 1
Over voltage category: II

Environmental 5 - +40 °C
temperature:
Storage temperature: -10 - +50 °C

Measurement range
Voltages measurement: 0 – 300 V ac
(Entrance resistance: 10 MOhm)
Discrimination: 0,3 V

Earth conductor 0,00 – 40 Ohm


resistance:
(Test voltage 6 V ac, max. 25 A / max. 5 A)
Discrimination: 10 mOhm
Insulation resistance: 0,2 – 100 MOhm
(Test voltage 500 V dc, max. 3,5 mA)
Discrimination: 0,1 – 2 MOhm

Leakage current: 0 – 10 mA
or 0 – 20 mA
Resistance: 1000 Ohm +/- 1 %
or 2000 Ohm +/- 1 %
Discrimination: 1 µA or 0,2 µA

Difference current: 10 A – 20000 A


Discrimination: 1 µA or 0,2 µA

Output power: 1 – 3,5 kW


Discrimination: 1W

Current: 0 – 16 A
Discrimination: 10 mA

6
S.P.L. GM-610 User manual

2.1 Intrinsic uncertainty

Measurement range error (faults)


Voltage: 0 – 300 V ac +/- 0,3 V or +/- 1 %
of measurement value

Earth conductor resistance: 0,00 – 4,9 Ohm +/- 0,03 Ohm or+/- 5 %
5 – 40 Ohm of measurement value

Insulation resistance: 0,2 – 4,9 MOhm or +/- 0,2 MOhm or +/- 5 %


5 – 100 MOhm of measurement value

Leakage current: 0 – 99 µA +/- 2 µA or +/- 1 %


100 – 20000 µA of measurement value

Difference current: 10 – 20000 A +/- 2 µA or +/- 1 %


of measurement value

Output power: 1 – 3,5 kW +/- 2 W or +/- 5 %


of measurement value

Current: 0 – 16 A +/- 50 mA or +/- 2,5 %


of measurement value
The specified intrinsic uncertainties relate to the respective measuring circuit. The operating uncertainty at the test object
connections is ± 5%. The displayed value is normalized according to the documentation / standard requirement, if required.
12/2020

7
S.P.L. GM-610 User manual

2.2 Connections / Accessories

Interface: 1 x Centronics for printer


1 x USB for PC-connection
1 x RS-232 for PC-connection
1 x PS2
for external keyboard or bar code scanner

Test object connection: 1 ea. VDE test socket


5 x safety sockets (4mm) for applied part,
selectable into 3 groups
1 x safety socket (4mm) for test probe
1 x safety socket (4mm) for PE

Display: 4,3’’ TFT-Display

Operation: Touch panel

Accessories: 1 x measurement line with test probe, 1 m length


1 x test adapter PA-X for self-diagnosis test
1 x USB cable
1 x power cord 16 A

Mechanical data: light weight metal case

Dimensions: 235 x 130 x 310 mm (B x H x T)

Weight: approx. 5,4 kg

8
S.P.L. GM-610 User manual

3 Warning signs and symbols

Please read this documentation before using the device.

Within this manual potential hazards are advised and the actions necessary to
avoid risks described.

HINT: Labels and texts on the device are part of the documentation!

All warning labels and symbols on the GM-610 are explained in the following:

 general warning signs (IEC 61010-1:2011)

Disregard documentation, read the manual!

 terminal for earth (IEC 61010-1:2011)

functional earth terminal

replaced:

OLD symbol

HINT: Detachable power cords may not be replaced by poorly dimensioned


power cords.

9
S.P.L. GM-610 User manual

4 Functional description

The GM-610 is a microprocessor-controlled device for measuring insulation resistance,


protective earth resistance and leakage current of medical technical and other electrical
equipment.

The CPU board controls all the necessary paths, switching for the various
measurements. The measuring paths are switched by relays installed on relay boards.

The test power board generates test voltage for the insulation resistance measurement
(500 V dc), as well as an internal test voltage for the leakage current – alternative
measurement. This board also contains the RS-232 and USB Interface for the
communication to an external PC. A transformer generates the test voltage for the
protective earth resistance measurement. This transformer can supply test current of
25 A.

All measurements (leakage current, insulation resistance, and protective earth


resistance) generate a voltage drop across a resistor (1 kOhm +/- 1 % or 2 kOhm +/-
1 %), which amplified using a chopper amplifier. The amplified voltage signal will be
transformed by a RMS converter, which generates RMS value from the AC voltage.
This DC-voltage is digitized using an A/D converter.

The mains voltages are connected, via a separate, high-resistance input (10 MOhm),
to the chopper amplifier. Then they follow the same way as the other measured values.

The digital value is transmitted serially via digital isolator to the CPU board. The CPU
board converts the value and transmits it to the PC.

It is also possible to use the GM-610 to perform measurements in the IT system, and
for measurements on permanently connected devices. After the power is switched on,
the GM-610 measures the line voltage from the phase conductor to the PE-conductor
contact, and from the directly grounded (N) conductor to the PE-conductor contact.

In this manner, the GM-610 checks whether a phase/zero-sequence net or an


IT system is present. In the event that a phase/zero sequence network is present, the
system then determines whether the line voltage of net carries 115 V or 230 V.

10
S.P.L. GM-610 User manual

To set the GM-610 into operation, first connect it to the power supply.

For a PC-controlled measurement connect it to the PC via the USB interface cable
supplied.

Insert the test-probe lead (supplied with the GM-610) into the jack labelled. This
jack is used for measurement of protective earth resistance and for measurement
of enclosure leakage current. To perform tests involving the patient applied part,
insert the required test leads into the test jacks, which have been labelled “Patient
applied part”.

The GM-610 is now ready to perform tests for the electrical safety of medical
technical appliances.

4.1 Specific features of standard

The measurements of the leakage current in accordance to IEC 60601 require a


TN system as a voltage supply for the test voltage.

With GM-610, measurements can be carried out at devices with fixed protective
earth conductor.

4.2 Additional functions

With GM-610 the separate ac/dc measurement (RMS), of the patient leakage and
patient auxiliary current can be carried out accordingly to IEC 60601. Furthermore,
the difference current can be measured.

4.3 Standardization

The measuring values of the leakage current measurements become automatic,


suitable of the respective norm calculates to the normative demanded value of the
mains voltage.

11
S.P.L. GM-610 User manual

5 Manual

5.1 Operation

In stand-alone mode GM-610 will be touch panel operated.

The explanation of the main touch panel buttons:

ESC / Back OK / Choice Backward Forward / Next

Setup Quick Entry / Edit

5.2 Main menu

After power on, the voltage will be measured for the protective earth conductor and
the two mains conductors. This will determine whether it is a TN system or an IT
system. Furthermore, it is checked whether it is a 115 V or 230 V mains. The
identified results are viewed by a blinking code on the PA-LEDs.

Switch on procedure blinking code

PA1  115 V mains


PA2  230 V mains
PA3  Error in adding the two switch on voltages L1 and L2
PA4  TN-system
PA5  TN-system
PA5 + PA1  IT- system
PA5 + PA2  Solid state relay error
PA5 + PA3  internal temperature above 40 degree

GM-610 can be operated either with 115 V or 230 V mains voltage.

The yellow LED located next to the test socket "mains" indicates that the voltage is
applied. This is also valid to the error condition S.F.C. mains open.

12
S.P.L. GM-610 User manual

After properly initialization the following main menu can be seen.

Start-Menu

Norm Multi
Test protocols User

Setup Info

5.3 MULTI

After choosing menu <Multi>, the GM-610 can be used in a multimeter mode.

You can now choose function steps sorted by the norm, or with <Quick entry> start
measurement entering SPL function number.

MULTI

IEC 60601
IEC 62353
IEC 61010
VDE 0701/02

ESC Quick entry

Norm-Function

Function 1
Function 2
Function 3
Function 4
Results

ESC Setup Backward Forward

13
S.P.L. GM-610 User manual

In the <Setup> menu you can use <SETUP>, <RESULTS> or <QUICK ENTRY>:

Setup
Results
Quick entry

ESC

After choosing <SETUP> you can change different parameters of GM-610. In the
<RESULTS> menu will be saved as the last measurement from the multimeter
mode.

Choosing <Quick entry> is possible to start the measurement entering SPL function
number. In this menu you can use all functions available in GM-610.

Quick entry

SPL function
Numpad

ESC Setup Start

The SPL function number is double-digit, after this number an applied group or
applied parts number can follow.

If the number after SPL function will not be typed, all applied groups and applied
parts will be chosen.

By the applied group you can choose between 1…4 groups (4 means group 1 plus
3).

Applied parts can be chosen by entering 01…05 (it is necessary to put zero before
the number of applied part).

After the function number is entered, the user can add: P, A, AP, D, DP.
P means turn phase position, A – AC measurement, D – DC measurement.

14
S.P.L. GM-610 User manual

If the function number is correct, you will be able to see the function description and
then you can start the measurement.

Function number, function description, actual value and the unit of measurement
will be displayed during the measurement.

With the button <ESC> you can always cancel the measurement.

Function-measurement

Actual value

ESC

In the menu <Results> you can get last measurement results. There will be maximal
400 results saved. If the memory is full, first 150 results will be automatically
deleted.

Results

Result 1
Result 2
Result 3
Result 4

ESC Setup Backward Forward

15
S.P.L. GM-610 User manual

All results can be printed at any time or deleted by entering menu <Setup>.

Delete results

Print results

ESC

The following test steps are available:

IEC 60601:

SPL01 Mains voltage


SPL02 Load current
SPL19 Power
SPL05 Protective earth resistance
SPL06 Earth leakage current NC
SPL07 Earth leakage current SFC line open
SPL08 Touch current NC
SPL09 Touch current SFC PE open
SPL10 Touch current SFC line open
SPL11 Patient leakage current NC
SPL12 Patient leakage current SFC PE open
SPL13 Patient leakage current SFC line open
SPL14 Patient auxiliary current NC
SPL15 Patient auxiliary current SFC PE open
SPL16 Patient auxiliary current SFC line open
SPL17 Patient leakage current line-AP
SPL50 Line off
SPL51 Read detailed information’s of testing equipment

16
S.P.L. GM-610 User manual

IEC 62353

SPL01 Mains voltage


SPL02 Load current
SPL19 Power
SPL20 Protective earth resistance
SPL21 Equivalent leakage current, alternative
SPL22 Equivalent leakage current, direct
SPL23 Equivalent leakage current, difference
SPL24 Leakage current of applied part, alternative
SPL25 Leakage current of applied part, line-AP
SPL26 Leakage current of applied part, internal power source
SPL28 Insulation resistance, AP Typ B
SPL46 Insulation resistance, Equipment
SPL47 Insulation resistance, AP Typ F
SPL50 Line off
SPL51 Read detailed information’s of testing equipment

IEC 61010

SPL01 Mains voltage


SPL02 Load current
SPL19 Power
SPL29 Protective earth resistance
SPL34 Earth leakage current, direct
SPL37 Touch current, direct
SPL50 Line off
SPL51 Read detailed information’s of testing equipment

17
S.P.L. GM-610 User manual

VDE 0701-0702

SPL01 Mains voltage


SPL02 Load current
SPL19 Power
SPL38 Protective earth resistance
SPL39 Insulation resistance
SPL40 Earth leakage current, alternative
SPL41 Earth leakage current, direct
SPL42 Earth leakage current, difference
SPL43 Touch current, difference
SPL44 Touch current, direct
SPL48 Touch current, alternative
SPL50 Line off
SPL51 Read detailed information’s of testing equipment

All measuring values of the leakage current measurements are standardized on the
normative demanded standard values.

230 V +/- 10 % 115 V +/- 10 %


IEC 60601 253 V 127 V
IEC 62353 230 V 115 V
IEC 61010 230 V 115 V
VDE 0701-0702 230 V 115 V

18
S.P.L. GM-610 User manual

5.4 NORM

Automatic tests, sorted by the Norm, can be chosen from the main menu by
pressing the <NORM>.

Norm

IEC 60601
IEC 62353
IEC 61010
VDE 0701/2

ESC

5.4.1 IEC 60601

Choosing the field IEC 60601 the automatic measurement, in accordance to


IEC 60601, can be done.

Depending on the class and device type, proper test instruction can be chosen.

Choice: Norm-Test instruction

Test instruction 1
Test instruction 2
Test instruction 3
Test instruction 4
Test protocol

ESC Setup Backward Forward

All measured values during the test will be compared to the limit value and
automatically evaluated – positive with „+“ and negative with „-“.

The test stops if the result is negative evaluated. With the button <FORWARD> the
step will be repeated and with <ESC> button test instruction will be aborted.

19
S.P.L. GM-610 User manual

The following test instructions are selectable:

60601 CI Type B
60601 CI Type BF
60601 CI Type CF
60601 CII Type B
60601 CII Type BF
60601 CII Type CF

5.4.2 IEC 62353

The same description like IEC 60601. The test steps are different.

The following test instructions are selectable:

62353 CI B difference
62353 CI BF difference
62353 CI CF difference
62353 CI B direct
62353 CI BF direct
62353 CI CF direct
62353 CI B alternative
62353 CI BF alternative
62353 CI CF alternative
62353 CII B difference
62353 CII BF difference
62353 CII CF difference
62353 CII B direct
62353 CII BF direct
62353 CII CF direct
62353 CII B alternative
62353 CII BF alternative
62353 CII CF alternative

CI can be selected for system or for equipment.


CII can be selected for system and equipment.

In the standard IEC 62353, the visual inspection and functionality are required.
Before the electrical safety measurement starts, the menu with visual inspection
will be called. After electrical safety, the functionality can be saved.
You can answer with yes/no and with <ESC> button abort the test instruction.

20
S.P.L. GM-610 User manual

5.4.3 IEC 61010

The same description like IEC 60601. The test steps are different.

The following test instructions are selectable:

61010 CIB with Touch current measurement


61010 CI without Touch current measurement
61010 CII with Touch current measurement

5.4.4 VDE 0701-0702

The same description like IEC 60601. The test steps are different.

The following test instructions are selectable:

70X CI Direct measurement general limit values


70X CI Difference current general limit values
70X CI Alternative measurement general limit values
70X CII Direct measurement general limit values
70X CII Difference current general limit values
70X CII Alternative measurement general limit values

In the standard VDE 0701-0702 the visual inspection will be required.


Before the electrical safety measurement starts, the menu with visual inspection
will be called. You can answer with yes/no and with <ESC> button abort the test
instruction.

21
S.P.L. GM-610 User manual

5.5 TEST PROTOCOLS

In the menu <TEST PROTOCOLS>, up to 200 test protocols can be saved.

Test protocols

Test protocol 1
Test protocol 2
Test protocol 3
Test protocol 4

ESC Edit Backward Forward

All test protocols can be printed at any time or deleted by entering menu <Edit>.

Edit: Test protocols

( All ) Test protocols print

( All ) Test protocols delete

ESC Number of test protocols:

22
S.P.L. GM-610 User manual

If the button <Test protocol> will be selected, the following menu with test protocol
description will be shown:

Test protocol description


Result:
Tester:
Time / Date:
Comment:
Id. number:

ESC Choice Results

The button <RESULTS> call menu with all test protocol results.

Test protocol
description
Single result

Step number SPL number


Function description
Limit value
Measured value
Test step result

ESC Next step

23
S.P.L. GM-610 User manual

One test protocol can be printed at any time or deleted by entering menu
<CHOICE> from the test protocol description menu.

Choice ( Test protocol description )

Id. number

Print test protocol

Delete test protocol

ESC

5.6 USER

USER menu call test instructions created by the user in ConActX Test module PC-
software.

The maximum amount the user can export from Test module software are 50 Test
instructions to the file called GM610imp.pvs placed in ConActX folder. One test
instruction may have a maximum of 200 functions.

The file GM610imp.pvs with user test instructions can be imported to GM-610 with
GMExport-Script called from ConActX Test module and the function number 21 –
SPL PVS-Imp. 1.
All user test instructions saved in GM-610 will be deleted!

Notice: all test protocols and all user test instructions saved in GM-610 will
be deleted during the calibration.

24
S.P.L. GM-610 User manual

User test instruction can be called from menu <USER>.

User

User – test instruction 1


User – test instruction 2
User – test instruction 3
User – test instruction 4
Test protocol

ESC Bearbeiten Backward Forward

All measured values during the test will be compared to the limit value and
automatically evaluated – positive with „+“ and negative with „-“.

The test stops if the result is evaluated negative. With the button <FORWARD> the
step will be repeated and with <ESC> button test instruction will be aborted.

At the end of the test, user can type identification number and comment.

Id. Number:

KEYBOARD

ESC Forward

Comment:

KEYBOARD

ESC Forward

25
S.P.L. GM-610 User manual

In Test instruction menu it is still possible to change the identification number,


comment, tester name and time / date. If the button <ESC> is selected, test
instruction will be aborted and all values will be deleted.

The button <Forward> allows you to print the test protocol.

Save – Test instruction

Tester:
Time & date:
Comment:
Id. Number:

ESC Forward

Print

Yes No

26
S.P.L. GM-610 User manual

5.7 SETUP

In the SETUP menu, you can choose different device and tests parameters and
change them.

Side 1
1. Results
2. Test protocols
3. Test lead resistance
4. GM-610 Diagnose

ESC Backward Forwards

5.7.1 Results

Last measurements from the multimeter mode can be shown in the menu
RESULTS, printed and deleted.

5.7.2 Test protocols

All test protocols and a single test protocol can be shown, printed and deleted in
this menu.

5.7.3 Test lead resistance

The value of the lead resistance will be measured with SPL function 18. This value
will be saved in a non-volatile memory and used for all protective earth resistance
measurements.

27
S.P.L. GM-610 User manual

5.7.4 GM-610 Diagnose

In this menu, the internal self-check of the GM-610 starts. This takes place with the
test adapter PA-X, that is connected into the test socket in addition with TP1, and
one of applied part.

The test units runs in both phase positions.


The insulation resistances and the mains voltage are supervised with +/- 10 %, all
other measurements are supervised with +/- 5 %. If the tolerance is over the rated
value and the self-check stops.

Side 2
5. Tester
6. Identification number & comment
7. Automatic / Partautomatic
8. Measure type

ESC Backward Forward

5.7.5 Tester

Tester name with maximal 20 signs can be typed

5.7.6 Identification number and comment

At the end of the test instruction, the identification number and comment from the
last test instruction will be automatically displayed in the input field.

5.7.7 Automatic / Partautomatic

The measure mode can be selected for the test instruction measurement. If part
automatic is set, every function must be confirmed with the button <FORWARD>.

5.7.8 Measure type

User can choose single or cyclic measure type.

28
S.P.L. GM-610 User manual

Side 3
9. Multi – display result time
10. Time & Date
11. Select your language
12. GM-610 Info

ESC Backward Forward

5.7.9 Multi – display result time

In the multimeter mode you can set the display result time after the measurement
for:
- 0 seconds,
- 3 seconds,
- 7 seconds,
- ESC (wait until the <ESC> button is pressed).

5.7.10 Time & Date

Time and date in a real-time clock can be changed.

5.7.11 Select your language

Languages can be selected:

- 0: Deutsch
- 1: English

29
S.P.L. GM-610 User manual

5.7.12 GM-610 Info

You can get in the Info menu GM-610 device and setup information’s.

Side 1

GM-610 Info

1. Contact

ESC Forward

Side 2

GM-610 Info

2. Firmeware 3. Hardware
4. TFT-Firmeware 5. Serial Number
6. Test steps 7. next calibration

ESC Forward

Side 3

GM-610 Info

8. Measure type 9. Prüfer


10. Display brightness 11. Switch on voltage
12. Auto- / Partautomatic

ESC Rückwärts

30
S.P.L. GM-610 User manual

Side 4
13. Header
14. Display brightness
15. Touchtone
16. DEF – restore default values

ESC Backward Forward

5.7.13 Header

For the printout of test protocol you can select a start line between 0..15.

5.7.14 Display brightness

The display brightness can be selected like 100%, 85%, 70% or 55%.

5.7.15 Touchtone

User can turn on or turn off the device tone. The tone will not be switched off for the
measurement and device warnings.

5.7.16 DEF – restore default values

Default values will be set:


- automatic mode,
- single measure type,
- 15 headers,
- 3 seconds for multimeter result display time,
- touchtone on,
Tester name: Otto Normalprüfer

31
S.P.L. GM-610 User manual

5.8 General Hints

All test instructions will run twice, so that the measurement of the leakage current
occurs in both phase positions. The higher value in each case is documented in the
test protocol.

At the end of every test instruction the measurement of the line voltage, the current
consumption and the power consumption is carried out. In addition, the test object
is switched on. This also happens after equivalent measurements.

By measurements, which demand an isolated installation of the test object, this has
absolutely to be kept. Isolated installation means that no other connections to PE
exist and the tester cannot touch the test object during the test.

32
S.P.L. GM-610 User manual

5.9 Short cuts

IEC 60601 / IEC 62353

CI Class I
C II Class II
C III Class III
Type B Type B Applied Part
Type BF Type BF Applied Part
Type CF Type CF Applied Part
direct Direct measurement

VDE 0701-0702

CI Class I
C II Class II
C III Class III
gen general
direct Direct measurement

33
S.P.L. GM-610 User manual

5.10 PC-controlled

GM-610 is controlled via the RS-232 interface or via the USB interface.

For the operation of the device’s USB interface an operating system specific driver
is required. This driver is provided on the S.P.L. product CD.

It is recommended to set the COM port provided by the system control to COM1 –
COM4 and if present please disable the energy management.

The USB device driver usually must be installed separately for every USB port on
a PC.

34
S.P.L. GM-610 User manual

6 Service and Calibration

The measuring - and test devices quality, manufactured from S.P.L. Elektronik
Brieselang, is guaranteed by a multi-level quality control - and quality assurances
in accordance to DIN EN ISO 9001. After final inspection and calibration each
device gets an examine plaque and a calibration certificate. The calibration is
enforced with standards, which are restorable to the standards of the Physical-
Technical Federal Institution.

In agreement with TÜV CERT, business-group TÜV Rheinland / Berlin-


Brandenburg, is recommended for normal, accordingly to the corresponding use of
the measuring and test devices form S.P.L. Elektronik Brieselang an annually
calibration interval.

The calibration includes a maintenance, update of the firmware, complete control


of the accuracy of measurements for of all test steps by control with one or several
reference-values and calibration. By an update of the firmware for example,
changes in standards, expansions of the function of the devices etc. are
automatically integrated. By the implemented inspection and calibration, the validity
is fixed by the implemented measurement list in the associated user guide. The
warranted characteristics for the PC-operation of the measuring - and test devices
of S.P.L. Elektronik are guaranteed through the offered or authorized software-
programs.

For cleaning, you should wipe off your device with a moist cloth, which is
impregnated with a mild detergent. It is important after that to rub your device dry.

Attention – Hint!

Before cleaning always pulls of the power plugs, otherwise danger for life exists

Please take note in the use with the test system that it is despite of protected
measuring channel inputs a sensitive measuring instrument.

Waste disposal:
Through the used parts of the measuring - and test devices of the S.P.L. Elektronik,
etc. accumulators, Lithium batteries, and similar things, a professional waste
disposal is necessary. This can be done through S.P.L. Elektronik, ear 37506249,
if demand is enforced.

35