Calibration Guideline For The Infusion Pump Analyzer
Calibration Guideline For The Infusion Pump Analyzer
Abstract
The flow laboratory of the National Institute of Metrology, Thailand (NIMT), in cooperation with the department of
medical sciences, and the department of health service support, ministry of public health developed the periodic testing
and maintenance procedures of the infusion pumps according to the international standards, IEC 60601-2-24 particular
requirements for the safety of infusion pumps and controllers, and requirements from the authorities in Thailand. The
testing procedures are now widely used in hospitals across country. Since the infusion analyzer is used as a tester in one
of the methods applied in the infusion pump testing procedures. Then, the calibration guideline for the infusion analyzer
has been developed to support the secondary laboratories for calibrating the devices. The guidelines provide
characteristic of the three main elements used for calibration which are the supply and flow control, the weighing system
and the data acquisition system. The infusion analyzer has to be calibrated at least three points; 50, 100 and 300 ml/h that
are required by the authorized person and normal conditions for patients in the hospitals. The uncertainties of calibrations
are range from 0.31% to 0.57%.
Shield The flow rates are calibrated at 50, 100 and 300 ml/h.
Variation of the
environment
Those flow rates are normally selected to infuse the
Fluid level
drugs or solutions to patients. At each flow rates, mass
and time are successively collected every 30 seconds
from the weighing system. The calibration period is one
Balance hour and repeated for three loops. The data from the
UUC are recorded depending on its software. However,
the UUC data have to show the measured flow rates.
Figure 2: Shield of weighing balance Other parameters such as the water temperature,
ambient temperature and ambient humidity are also
0.20 recorded.
0.18 0.17208
The measured mass and time are used for calculating
0.16 Largemouth w/o cover
the flow rates. The mass of each increment are
Evaporation rate (ml/h)
Largemouth w cover
0.14 converted to volume by using Equation (1) which comes
Smallmouth w/o cover
0.12 Smallmouth w cover from ISO/TR 20461 [5].
0.10
0.08 V 20=mw ×Z×Y
(1)
0.06
Where mw the collected mass of water
0.04 Z the combined factor for buoyancy correction
0.02088
0.02 0.00864
0.00436 and conversion from mass to volume
0.00 Y the thermal expansion correction factor of the
delivering device
Figure 3: Evaporation rate with four types of vessels
Since the measured mass is collected at every 30
seconds, then, the discrete flow rates are executed every
UUC flowrates Standard flowrates 4.2 Instability of the flowrate, ufluctuation_WB and
ml/h ml/h
110 110 uinstability_UUC,
The instability of the discrete flowrate of the standard
105 105 calibration system, ufluctuation_WB, can be considered from
the standard deviation of N data divided by the total
100 100 collected numbers of data.
SD Qstd
95 95 uinstability =
discrete
WB
√N (3)
90 90
The instability of the discrete flowrate of the UUC,
85 85 uinstability_UUC, can be considered from the standard
0 1000 2000 3000 0 1000 2000 3000 deviation of N data divided by the total collected
Time (s) Time (s) numbers of data.
SD QUUC
uinstability =
discrete
Figure 5: Discrete flow rates of UUC and standard system at 100 ml/h
UUC
√N (4)
UUC flowrates Standard Flowrates
ml/h ml/h 4.3 Repeatability, urepeat_UUC
295 295
urepeat_UUC, can be
290 290 The repeatability of the calibration,
considered from the maximum deviation of the error
285 285 caused by the calibration result.
error max−error min
280 280 urepeat =UUC
2 √3 (5)
275 275
4.4 Resolution, ures_UUC
270 270
ures_UUC, can be
0 1000 2000 3000 0 1000 2000 3000
Time (s) Time (s) The resolution of the calibration,
considered from the finest display of the unit under
Figure 6: Discrete flow rates of UUC and standard system at 300 ml/h calibration that user can read.
The results of calibration of each set point is shown in Table 2 : The trumpet curve
Table 1. All channels of the infusion pump analyzer are Flow rate (ml/h)
Stability (window/min)
calibrated at flow rates; 50, 100 and 300 ml/h. The CH1 CH2 CH3 CH4
50 31 31 31 31
uncertainty of calibration are ranged from (0.31 – 100 11 11 11 11
0.57)% of reading. The error span of all channels (with 300 5 5 5 2
uncertainty) are within ±1% which meets the
specification of the analyzer. This information is considerably useful for the operator
in order to get the correct testing result when they are
Moreover, the calibration system was verified by using using the infusion pump analyzer to test the infusion
the standard flow meter (Bronkhorst, M13) which has pumps. Similarly, the stability time for 100 and 300
accuracy ±0.25%. The standard flow meter was set ml/h are 11 and 5 minutes that the infusion pumps have
instead of the UUC and tested at the same flow rates. to infuse the solution through the analyzer prior to start
The results of the calibration of the meter are compared the test.
with the result from its certificates and the international
comparison [2]. From those results of the standard flow 10.0
0.5
2 -0.03 -0.06 0.49 0.0
50 -0.5 0
3 -0.05 -0.10 0.47 5 10 15 20 25 30
-1.0
4 -0.20 -0.39 0.57 -1.5
1 0.16 0.17 0.31 -2.0
Window (mins)
2 0.28 0.29 0.33
100 1.5
3 0.37 0.39 0.47 1.0
4 0.14 0.15 0.42
Error (%)
0.5
1 0.47 0.17 0.31 0.0
2 1.52 0.54 0.31 -0.5 0 5 10 15 20 25 30
300 -1.0
3 1.47 0.51 0.33
-1.5
4 1.85 0.65 0.31 Window (mins)
0.5