Pipetting Handout 09
Pipetting Handout 09
Introduction:
The ability to accurately pipette is a critical skill required of all clinical laboratory scientists.
Accurate and precise pipetting skills are required to prepare accurate and precise dilutions of
various reagents and patient specimens when performing quantitative measurements in the
clinical laboratory. This may include preparing
1. Dilutions of patient samples when the concentration of a substance is too high to be
accurately measured
2. Stock standards to yield various target concentrations for working standards
3. Controls used to assess instrument performance and validity of test results
4. Reagents
Most clinical laboratory procedures require accurate volumetric measurements to ensure accurate
results. Such measurements may be made by manual pipettes (serologic, volumetric and micro),
dispensing devices, automated pipettes, burets, volumetric flasks and graduated cylinders. The
reliability of these devices can vary considerably, and not all of these devices are suitable for the
most accurate work. The maximum allowable error (or tolerance limits) for these devices are set
by the National Institute of Standards and Technology (NIST). For exact work, only Class A
glassware should be used, as this class of glassware meets NIST specifications.
Calibration of pipettes
Pipettes are checked by the laboratory to ensure accurate delivery of a specified volume. When
the laboratory does this ‘check’, the CLS refers to it as ‘checking or verifying pipette
calibration’. In other words, we are checking to make sure the pipette is still ‘calibrated’ to
deliver a specified volume.
Macro-volume pipettes that are Class A do not need periodic calibration checks by the
laboratory. These pipettes are calibrated to deliver a specified volume by the manufacturer. As
long as the pipette is properly used this is true. Most laboratories will use only class A macro-
pipettes and glassware so that in-house calibration checks are not needed. This saves time and
money.
Most laboratories use volumetric and serologic pipettes. Mohr pipettes are seldom used in the
clinical laboratory due to potential misuse of this pipette which causes invalid test results.
3. MOHR: (figure B)
a. These pipettes are similar to serologic pipettes in that they have
multiple volume graduations, but they do not have graduated volume
markings to the bottom of the pipette tip, nor do they have etched
rings at the top of the pipette.
b. These pipettes require a controlled delivery of the solution between the
upper fill mark and the intended lower mark
c. The tip of the pipette SHOULD NEVER come in contact with the accumulating
fluid in the receiving vessel during drainage or the inside walls of the receiving
vessel when dispensing of sample
d. Most laboratories discourage use of the Mohr pipette because it is often mistaken
as a serologic pipette and is used improperly:
1) Error: fluid remaining in the tip after drainage is ‘blown’ into the
receiving vessel, resulting in extra volume of liquid being delivered
2) Error: when delivering sample to receiving vessel, the Mohr pipette tip is
allowed to touch the inside of the receiving vessel, resulting in an
inaccurate fluid volume delivery
2. Inspect the pipette carefully. The pipette should be clean and the tip not chipped. Dirty
pipettes will drain unevenly and inaccurately; chipped tips alter the volume delivered.
3. NEVER MOUTH PIPET: a rubber bulb or other device is used to fill the pipette with
fluid above the mark. Careful aspiration of the fluid is required so that fluid is not
aspirated beyond the pipette into the bulb. If this should happen, alert your instructor
5. Pipette solutions that are at room temperature to avoid errors caused by contraction
and/or expansion of fluids
6. Any air bubble, regardless of its size, will invalidate the volume of liquid you are trying
to pipette. Should an air bubble appear in the pipette, re-sampling is required.
If the meniscus is adjusted before the outside of the pipette is wiped off, fluid will be
pulled out of the pipette onto the wiping cloth by capillary action, thus decreasing the
desired volume to be delivered.
9. The pipette tip SHOULD NEVER come in contact with the fluid in the receiving vessel
when delivering the sample (refer to step 7g)
1. Generally, when pipetting volumes </= 1.0 ml, a micro-volume pipette is used
2. Several types are available for use; choice depends upon laboratory and intended use
a. Plunger with double stop d. Multiple volume delivery
b. Plunger with single stop e. Automatic or manual tip rejection
c. Single fixed-volume delivery f. Automatic aspiration and delivery
Student Laboratory Rotation (UNMC site): you will learn proper pipetting technique
using the ‘oxford’ pipette. This is a double stop pipette.
Clinical Laboratory Rotation: you will have the opportunity to use other types of
micropipettes such as:
a. ‘MLA’ and ‘Eppendorf’ pipettes: single stop pipettes
b. ‘Focus’ pipette: double stop pipette
3. Initially before use, depress the plunger knob several times. This acts to redistribute the
lubricant within the pipette. It should feel smooth when the plunger knob is depressed
and released. Alert your instructor if sticking is encountered
4. The plastic pipette tip must be properly fitted onto the pipette: it should be straight and
held firmly in place.
5. When using a plastic tip, pre-rinsing is always required (unless otherwise indicated in the
procedure)
6. Pipette solutions that are at room temperature to avoid errors caused by contraction
and/or expansion of fluids
8. Depress and release the plunger with a controlled and smooth action. Try to maintain the
same speed of intake and delivery for all samples (consistency is the key)
9. Never allow the plunger knob to snap back as this will cause aspiration of the liquid past
the pipette tip into the pipette contaminating its interior assembly. If this should happen,
alert your instructor as cleaning of the pipette is required
10. Depress the plunger knob BEFORE inserting the pipette tip into the fluid. Depression of
the plunger knob while it is in the solution will cause bubbles to form in the tip, resulting
in an inaccurate measurement
11. Insert the tip of the pipette into the sample at approximately the same depth each time.
Try not to ‘drown’ the tip into the fluid (never deeper than ¼”)
12. Any air bubble, regardless of its size, will invalidate the volume of liquid you are trying
to pipette. Should an air bubble appear in the pipette, re-sampling is required.
13. When delivering sample to receiving vessel, place plastic pipette tip inside the receiving
vessel without touching the inside wall of the receiving vessel. Touching the wall with
the tip of the pipette while delivering sample alters the ‘flow’ of sample delivery and also
the volume being delivered.
14. Position the sample tip so that the sample does not ‘dribble’ down the entire length of the
inside wall of receiving vessel. This may cause a loss of sample and invalid test results.
15. The plastic pipette tip SHOULD NEVER come in contact with the fluid in the receiving
vessel while delivering the sample (unless otherwise instructed by the procedure)
This is the 3. Immerse the tip approx 1/8" into the solution
pre-wetting
step that is 4. Return the plunger knob to the release position, aspirating the sample into the
required plastic tip
when using
plastic 5. Slightly remove the plastic tip from solution and return the sample to the original
pipette tips container: depress the plunger knob to the first (upper-most) stop, pause slightly, and
then depress the plunger knob to the second (lowest) position
9. Return the plunger knob to the release position, aspirating the sample into the plastic tip
10. Remove the plastic tip from solution, and while removing the plastic tip from the
solution, gently ‘swipe’ the tip against the side of the tube to avoid delivering any fluid
adhering to the outside of the plastic tip into the receiving vessel
11. Place plastic tip inside the receiving vessel without touching the inside wall of receiving
vessel. Position sample tip so that delivery of sample does not result in the sample
dribbling down the entire length of the inside wall of receiving vessel. Do not allow the
pipette tip to touch any fluid in the receiving vessel.
12. Deliver the sample to the receiving vessel: depress the plunger knob to the first/upper
stop, pause slightly, and then depress the plunger knob to the second/lowest position
(this secondary movement is used to BLOW OUT the contents of the pipette tip and is
similar to the blow out step used for serologic pipettes)
13. With the plunger knob still held in its lowest position, withdraw the pipette from vessel
(this is done to eliminate aspirating some of the sample from the receiving vessel back
into the plastic tip resulting in delivering a falsely low volume)
14. Return plunger knob to the release position; remove and properly discard plastic tip
CLS 414 Clinical Chemistry: Student Lab Rotation 9
Pipetting Handout
‘MLA’ pipette:
1. This is a TD (to deliver) pipette, which has only 1 plunger stop position. The same
plunger stroke is used for aspiration and delivery of sample
2. This type of pipette is most often used in clinical laboratories due to its ease of use
This is the 3. Immerse the tip approximately 1/8" into the solution
pre-wetting
step that is 4. Return the plunger knob to the release position, aspirating the sample into the
required pipette tip
when using
plastic 5. Slightly remove the plastic tip from solution and return the sample to the original
pipette tips container: fully depress the plunger knob
9. Return the plunger knob to the release position, aspirating the sample into the pipette tip
10. Remove the plastic tip from solution: and while removing the plastic tip from the
solution, gently ‘swipe’ the tip against the side of the tube to avoid delivering any fluid
adhering to the outside of the plastic tip into the receiving vessel
11. Place plastic tip inside the receiving vessel without touching the inside wall of receiving
vessel. Position sample tip so that delivery of sample does not result in the sample
dribbling down the entire length of the inside wall of receiving vessel; try to deliver the
sample near the bottom of the vessel while not touching the fluid that is already in the
receiving vessel) Do not allow the pipette tip to touch any fluid in the receiving vessel.
12. Deliver the sample to the receiving vessel: fully depress the plunger knob
13. With the plunger knob fully depressed, withdraw the pipette from vessel
(this is done to eliminate aspirating some of the sample from the receiving vessel back
into the plastic tip resulting in delivering a falsely low volume)
14. Return plunger knob to the release position; remove and properly discard plastic tip