Manual Lifegain INGLES
Manual Lifegain INGLES
CU-HD1
This Operation manual is intended to provide information needed to use CU-HD1 that is developed and
manufactured by CU Medical Systems, Inc.
Also this Operation manual is subject to change without prior notice.
1
CU-HD1 Instructions for use
◎ Table of Contents
2
CU-HD1 Instructions for use
◎ Table of Contents
3.1 Unpacking.................................................................................................................... 49
3.1.1 Package of the Main Body ....................................................................................................................... 49
3.1.2 Package of Accessories .............................................................................................................................. 50
3
CU-HD1 Instructions for use
◎ Table of Contents
4
CU-HD1 Instructions for use
◎ Table of Contents
5
CU-HD1 Instructions for use
◎ Table of Contents
6
CU-HD1 Instructions for use
◎ Table of Contents
7
CU-HD1 Instructions for use
◎ Table of Contents
13.5 Delivered Defibrillating Energy according to the Load Impedance ............. 188
8
CU-HD1 Instructions for use
CU-HD1 and CU Medical Systems, Inc. hereinafter referred to as “the Product” and “the Company” respectively.
This device provides the Automated External Defibrillator (AED) function, Manual Defibrillator function, Non-
Invasive Pacing function and Patient Monitoring function.
High-voltage and high-current electric energy is applied to the defibrillator. Therefore, be sure to
read through this Operation manual so that you can fully understand the safety cautions, operation
methods, and general cautions before using this product.
9
CU-HD1 Instructions for use
◎ General Instructions
◎ General Instructions
When using the product, be sure to follow instructions described in this Operation manual.
It is recommended to place the instruction manual near the product and refer to it when unsure about an item
or a defect has possibly occurred.
In no event, shall the company be liable for any product problems arising out of careless operation or
negligent misuse by the user.
All of the repair services for the product can provided only by CU MEDICAL SYSTEMS, INC. or its authorized
agents.
Use only parts and accessories that are recommended by the company.
When you want to use the product in connection with other devices for which no usage instructions are
provided in this Operation manual, be sure to contact us before starting operation.
If this product does not operate normally, stop using the product immediately, contact our company or a
certified agency, notify of the failure details and request a repair.
10
CU-HD1 Instructions for use
※ If an injury has occurred to the user or a patient due to a clear instance of user negligence or
misuse, the company or its authorized agent holds no liability.
A case that could result in dangerous situations, including death or severe injury if instructions are
not observed
Instruction that directly or indirectly addresses the company policy in order to protect people or
property
Explanation of reference terms or additional operating methods for normal product usage
11
CU-HD1 Instructions for use
Avoid installing or storing the product in a location with high humidity where is
exposed to moisture or poor ventilation.
Avoid installing or storing the product in a location where it may receive excessive
impacts from vibrations.
When disconnecting the power cord from the wall outlet, have it gently pulled out
by grabbing the plug and not the cord.
12
CU-HD1 Instructions for use
※ If the product has been submerged under water, contact and have the retailer inspect the product before
using it.
※ The status of the battery charge must be periodically monitored during the storage. Ensure that the
remaining battery is sufficient (for operating the device).
※ If the battery is completely discharged while using the device, you can use the device while charging the
battery by connecting the AC power module. At this time, turn off the device, connect the AC power
module, and then turn on the device again.
※ Do not operate the device in electrically noisy environments near motors, generators, X-ray equipment,
wireless transmitters or mobile phones as these will interfere with the signals being acquired. Such
electrical noise interference may lead to a malfunction of the device.
※ After using the device, thoroughly clean the main body using a soft, dry cloth.
※ If the battery has been stored for a long period of time, charge the battery periodically. This will help
preventing the occurrence of low battery level.
Use the AC power adapter and car cigar jack in order to only recharge the battery.
Do not use the device with the AC power adapter and car cigar jack connected to it.
Remove the battery if the device is not likely to be used for some time.
13
CU-HD1 Instructions for use
This product is a medical device designed to be used by Level 1 and Level 2 emergency medical technicians
and medical professionals.
This product provides the automated external defibrillation function and the manual defibrillation function. This
product also provides the synchronized cardioversion function, transcutaneous pacing function, and patient
monitoring function.
This product also has the Synchronized Cardioversion function, Transcutaneous Pacing function, and Patient
Monitoring function. This product can be used by emergency medical technicians and medical professionals.
You can use the functions of the device with simple button operations, and check various information through
the screen while using the product.
14
CU-HD1 Instructions for use
The CU-HD1 monitor/defibrillator is a complete acute cardiac care response system designed for basic
life support(BLS) and advanced life support (LAS) patient management protocols.
▶Indications
The Defibrillation function delivers an electric shock to a patient showing the symptoms of
sudden cardiac arrest, such as ventricular fibrillation and ventricular tachycardia, in order to
restore the normal ECG rhythm.
In the Automated External Defibrillation (AED) mode, the patient's ECG obtained through the
defibrillation pads is analyzed automatically and the guidance is provided through the voice and
message, informing the user to press the electric shock button.
▶Contraindications
None known.
The product must not be used in AED Mode on patients who show any of the following
symptoms: responsiveness, normal movement, normal breathing and a detectable pulse.
15
CU-HD1 Instructions for use
Manual Mode is divided into two functions such as asynchronous defibrillation and synchronized
cardioversion.
The Synchronized Cardioversion function analyzes the QRS of the patient's ECG to enable
execution of the defibrillation according to the R wave.
In the manual mode, the synchronized cardioversion treatment can be used for a patient with
unstable tachyarrhythmia, such as atrial flutter or atrial fibrillation, and a patient with cardiac
ischemia who has insufficient blood volume.
▶Indications
Manual defibrillation is indicated for the termination of certain potentially fatal arrhythmias, such
as ventricular fibrillation and symptomatic ventricular tachycardia. Delivery of this energy in the
synchronized mode is a method for treating atrial fibrillation, atrial flutter, paroxysmal
supraventricular tachycardia and, in relatively stable patients, ventricular tachycardia.
▶Contraindications
When the product is used for asynchronous defibrillation treatment in Manual mode, do not
use it on patients who show any of the following symptoms:
- responsiveness, normal movement, normal breathing and detectable pulse.
There is a possibility of explosion or fire if the product is used in the presence of flammable
agents or in an OXYGEN enriched atmosphere due to the arc discharged caused by electric
shock.
Do not deliver an electric shock when the patient's ECG signal is in the asystole state. It may
lead to a failure to restore cardiac pacemaker functions in the heart, meaning the cardiac
function will not be restored.
This product must not be applied on patients implanted with the implantable pacemaker. If
patients show all of the symptoms including no response, and abnormal breathing, use the
product in the following ways:
- Attach the pad at least 3cm away from the implantable pacemaker attached to the patient.
- Do not attach the pad right on the area implanted with the implantable pacemaker.
16
CU-HD1 Instructions for use
The CU-HD1 functions to support non-invasive pacing, a way of helping maintain a patient’s pulse
by attaching its electrode to the patient’s skin and delivering artificial electric stimulation to the
heart.
Pacing mode is divided into the ‘Fixed mode’ and the ‘Demand mode’.
Consult a physician and follow the manufacturer's instructions before treating a person with a
permanent pacemaker or an implantable cardiac defibrillator. Pacing therapy should only be
performed by trained medical personnel.
▶Indications
▶Contraindications
Noninvasive pacing is contraindicated for the treatment of ventricular fibrillation and asystole.
17
CU-HD1 Instructions for use
For the ECG monitoring function, you can selectively use the 3-lead, 5-lead, or 10-lead ECG cables.
If the patient's ECG is analyzed while monitoring the patient and if ventricular fibrillation or
ventricular tachycardia occurs or it exceeds or falls below the range of set ECG, the alarm function
will be activated.
▶ Indications
The electrocardiogram is used to identify, diagnose, and treat patients with cardiac disorders and
is useful in the early detection and prompt treatment of patients with acute
STelevationmyocardial infarction (STEMI).
▶Contraindications
None known
SpO2 is a noninvasive method of measuring functional oxygen saturation (SpO 2) in arterial blood.
SpO2 readings indicate the percentage of hemoglobin molecules in arterial blood which are
saturated with oxygen.
▶Indications
Pulse oximetry is indicated for use in any patient who is at risk of developing hypoxemia.
SpO2 monitoring may be used during no motion and motion conditions.
▶Contraindications
None known..
18
CU-HD1 Instructions for use
A noninvasive blood pressure measuring device is divided based on patient status (adult, child
and infant - classified as For adults, children, and infants and is used with an appropriate cuff to
measure a patient’s blood pressure.
▶Indications
▶Contraindications
None known..
End-tidal carbon dioxide (EtCO2) provides end-tidal CO2 level of the patient to monitor breathing
or to determine whether CPR is being performed correctly.
▶Indications
The electrocardiogram is used to identify, diagnose, and treat patients with cardiac disorders and
is useful in the early detection and prompt treatment of patients with acute
STelevationmyocardial infarction (STEMI).
▶Contraindications
None known..
※ For more detailed information about the patient monitoring mode, please refer to
“Chapter 7_Patient Monitoring”.
19
CU-HD1 Instructions for use
To apply the defibrillation or pacing functions on the emergency patient who is currently using
the implantable cardiac defibrillator (ICD) or the cardiac resynchronization therapy defibrillator
(CRT-D), you must contact a medical specialist.
Do not use this Product on more than one patient at the same time.
When using the storage function, initialize equipment usage time so that you can identify the
specific patient information from others. If you set the Rotary switch to OFF for about 10
seconds, the device usage time will be initialized.
20
CU-HD1 Instructions for use
This chapter is intended to provide information about the exterior view of the product, various buttons and
indicators on the main body, Bluetooth communications with linkage to the outside, a real-time printing printer,
screen symbols, texts and voice signals that intend to bring convenience to users.
Also, this chapter includes the guidelines on accessories mounted on the main body which are the power
module, ECG cable, SpO2 sensor, NIBP cuffs, and EtCO2 module, etc.
As for the pads designed and manufactured for the therapeutic purpose, ECG cables, and other
related accessories except for disposable consumables, you must use what is provided by the
CU Medical Systems, Inc.
When there is any damage in disposable consumables or accessories, stop using the device
and contact a customer service center for replacement. Also, if damage occurs to the
equipment cable or reusable paddles, please contact a service center.
If your CU-HD1 does not have some of the optional functionality listed in this chapter,
disregard these controls and the related information described throughout this manual.
21
CU-HD1 Instructions for use
22
CU-HD1 Instructions for use
Button Function
Charge Button
When an electricity shock needs to be done in the manual defibrillator mode, the
defibrillation energy is charged according to the setting on the Rotary switch by
pressing this button.
When the battery has been completely recharged, the orange lamp of the Shock
button will be on.
Shock Button
When a defibrillation electric shock is needed through an ECG signal analysis, the
orange lamp of the Shock button will be on.
By pressing the button at this time, the electric shock will be delivered to the
patient.
Menu Knob
This key is used to navigate around the menu and select each mode. In addition,
the entry of patient information and device setup can be done.
Soft Keys
These buttons facilitate selecting necessary functions in each mode.
23
CU-HD1 Instructions for use
Button Function
SYNC Button
In the manual defibrillator mode and pacer mode, it is used for synchronized
cardioversion.
It analyzes the patient's ECG signal and synchronizes the transfer of defibrillation
energy with the R wave among the QRS of the ECG signal within 60ms.
Print Button
Outputs the real-time ECG information or stops any printing during the printing
process.
NIBP Button
Starts/Stops noninvasive blood pressure measuring.
Home Button
While changing the settings in the menu, pressing this button exits out from the
menu screen.
Event Button
The medicine information given to patients is entered.
2) CU-CM1
Button Function
Power Button
This button is used to select power on/off of CU-CM1
24
CU-HD1 Instructions for use
2.3 Indicators
1) CU-HD1
The indicators are located right over the main body’s LCD display as shown below. They function
to indicate the power state of the device or any problems that may happen in the device.
Indicator Description
This indication shows whether the power is fed from the commercial power source
through an AC power module, or from the car cigar jack.
Once the battery is attached to the device, the LED lights up in green. When the
green LED is blinking, it indicates that the battery is being charged through the AC
power module and the car cigar jack. Once the charging is completed, it lights up
in green.
When there is any problem in the system, the SERVICE LED lights up.
If this happens, the device does not operate normally. Therefore, stop using it
immediately and contact an authorized service center to repair the product.
2) CU-CM1
The four indicators are located on the left side of CU-CM1.
Indicator Description
Power/Connection Indicator
The blue indicator will light up when the product turns on. If the CU-CM1 is in
communication with the CU-HD1 via Bluetooth while measuring CO2, the blue
indicator will blink in 1 second intervals.
Low Battery Indicator
The yellow indicator will light up when the remaining battery of the CU-CM1 falls
below 20%. Recharge the battery when the Low Battery Indicator turns on.
Charging Status Indicator
The green indicator will light up when the battery is charged with AC power.
The indicator will turn off when battery charging is complete
25
CU-HD1 Instructions for use
26
CU-HD1 Instructions for use
1) AED Mode
(A)
(D)
(B)
(E)
(C)
Screen Layout
Device state Shows mode in use, connector state, Duration of device
(A)
indication area usage, current time, date, and power state
(B) Sector 1 Display of ECG measured through the pads
Shows text prompt guides and progress of AED
(C) Sector 2
procedure
Heart rate indication
(D) Shows beat per minute(bpm)
area
(E) SpO2 indication area Shows SpO2(%)
Menu & Soft button Soft buttons to start analyze, change CPR type,
(F)
area start/stop CPR, and menu creation
27
CU-HD1 Instructions for use
(A)
(B) (D)
(E)
(C)
(F)
(G)
Screen Layout
Device state indication Shows mode in use, connector state, Duration of device usage,
(A)
area current time, date, and power state
(B) Sector 1 Vital signs measured through the ECG electrodes, pads, paddles
(C) Sector 2 or sensors
Heart rate is expressed as the vital sign in Sector 1, while the vital sign set for Sector 2 does
not affect the heart rate display.
When the ECG cables are not connected or when in Lead Fault condition, the ECG lead graph
is displayed in dotted lines.
When the pads or paddles are not connected or when the pads are not attached, the pad lead
graph is displayed in dotted lines.
28
CU-HD1 Instructions for use
(A)
(B)
(C)
Screen Layout
Shows mode in use, connector state, Duration of device
(A) Device state indication area
usage, current time, date, and power state
12-lead ECG indication 10-ECG signal measured form the lead ECG cable
(B)
area (I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6)
Patient vital signs & Displays heart rate, SpO2, NIBP, Bluetooth
(C) Bluetooth communication communication connection status, soft button menu, and
area alarm message
29
CU-HD1 Instructions for use
In addition, the icons related to the battery are summarized in the following table.
Symbol Description
AC power module input
It is recommended that battery charging must be done if it’s Step 4, or two LEDs light at the
battery gauge at least.
30
CU-HD1 Instructions for use
Starts CPR.
Stops CPR.
Turns off an alarm that is occurring. Pressing the soft button will turn off an
alarm and the button will change into the [Pause Alarm] button.
Transfers the information of the patient being examined to the computer in real
time. To see this, Bluetooth must be connected.
Once the pad is detached, the menu is activated. (Applied to all modes)
31
CU-HD1 Instructions for use
Symbol Description
Indicates it is printing.
Indicates the voice and ECG storage functions are being executed.
32
CU-HD1 Instructions for use
Conformité Européenne
Complies with the requirements of the Medical Device Directive
93/42/EEC as amended by 2007/47/EC
For EU only:
Electrical and electric equipment shall be collected and recycled in
accordance with Directive 2002/96/EC
Manufacturer
Date of manufacture
Use by date
Serial Number
33
CU-HD1 Instructions for use
Symbol Description
Direct current
Alternating current
Do not re-use
LOT Number
Catalogue number
Do not fold.
34
CU-HD1 Instructions for use
2) CU-CM1
Symbol Description
DC power input
Serial port
Gas Inlet
Warning
Serial Number
Date of manufacture
35
CU-HD1 Instructions for use
Serial Number
Catalog number
Rx only
Caution (U.S.): Federal law restricts this device to sale by or on the order
of a physician.
For EU only:
Electrical and electric equipment shall be collected and recycled in
accordance with Directive 2002/96/EC.
36
CU-HD1 Instructions for use
Serial Number
Catalog number
Rx only
Caution (U.S.): Federal law restricts this device to sale by or on the order
of a physician.
Gas Inlet:
Gas inlet for connecting the Nomoline Family sampling lines
For EU only:
Electrical and electric equipment shall be collected and recycled in
accordance with Directive 2002/96/EC.
Date of manufacture
37
CU-HD1 Instructions for use
※ Attach pads
Attach the disposable defibrillation pads to the patient’s chest skin below the right clavicle and
on the middle glands in the armpit below the left nipple, and connect the connector on the
opposite side of the defibrillation electrode to the defibrillation pad connector correctly.
※ Shock advised
It means that the patient needs to have electric shock treatment.
※ Stand clear
It means that a patient must avoid all contact with other people.
※ Charging
It indicates that sufficient energy is charged for electric shock treatment.
※ Shock delivered
It indicates that electric shock treatment was delivered to the patient.
※ No shock advised
It indicates that electric shock treatment is unnecessary.
38
CU-HD1 Instructions for use
39
CU-HD1 Instructions for use
※ The alarm settings can be changed under alarm list in the MENU. Default alarm settings
are restored when the product is powered off and on. To change the default alarm setting,
refer to the service manual.
• Text
If the set limit of alarm occurrence is exceeded according to the patient’s condition, an alarm of
the patient's condition will be issued and the relevant alarm text will be displayed on the top left
side of LCD. Also, if the cables, paddles or pads are not connected, a message indicating the
relevant condition is displayed on the LCD.
When it is temporarily stopped, the alarm is done so for a predetermined period of time, and the
temporal stop time is lapsed in 10-second interval until its preset time, enabling a checkup for a
temporary alarm stop.
If the alarm condition continues to happen even after discontinuing the alarm, the same alarm
continues to be issued.
If the volume level of the alarm sound is lower than the ambient sound pressure level, it may
interfere with the operator’s awareness of the alarm condition.
40
CU-HD1 Instructions for use
41
CU-HD1 Instructions for use
- Once the alarm is triggered, it will clear automatically when the patient recovers to
normal.
- This includes Medium priority order alarms.
ECG alarms (VT/VF, asystole, heart rate) are generated only when Sector 1 is lead II or pads.
To monitor ECG alarms, Sector 1 must be changed to 'Lead II' or 'Pads'.
2.6.2 Errors
The product may produce errors when there are problems during its operation in addition to
alarms, and each error is represented in the LCD screen as a code.
※ To troubleshoot the errors, refer to “Chapter 12_Troubleshooting”.
42
CU-HD1 Instructions for use
Setting Alarm limits to extreme values can render the Alarm system useless.
If the rotary switch is not in the correct position, the 'Check the rotary switch.' message
appears with a periodic alert sound. If this alarm message appears, check that the rotary switch
is positioned correctly.
2.7 Accessories
The CU-HD1’s accessories are composed of disposable accessories (disposable defibrillation pads,
ECG electrodes, printing papers, etc.) as well as external defibrillation paddles used for electric shock,
ECG cables, SpO2 sensor and extension cable, NIBP cuff and NIBP connection tube, SD card, power
supply devices (AC power module, car cigar jack, and battery module), and bed rack.
External
External defibrillation paddles are pressed against the patient's
Defibrillation
chest to deliver electric shock.
Paddle
43
CU-HD1 Instructions for use
3-Lead ECG This ECG cable is used for measuring 3-lead ECG waveforms
Cable ∙ Leads: I, II, III
5- Lead ECG This ECG cable is used for measuring 7-lead ECG waveforms
Cable ∙ Leads: I, II, III, aVR, aVL, aVF, V
10- Lead ECG This ECG cable is used for measuring 12-lead ECG waveforms
Cable ∙ Leads: I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
Disposable ECG
ECG monitoring electrodes are attached on the patient to
Monitoring
measure ECG signals.
Electrodes
SpO2 Sensor This extension cable is used to connect SpO2 sensor to the CU-
Extenstion Cable HD1.
Connection This connection tube is used to connect the NIBP measuring cuff
Tube for NIBP to the CU-HD1.
Measuring
CO2 Sensor If the EtCO2 option is selected for your CU-HD1, this
Communication communication module connects the CU-HD1 with the IRMA
Module Mainstream or ISA Sidestream over Bluetooth.
44
CU-HD1 Instructions for use
IRMA
Mainstream This mainstream analyzer module measures EtCO2.
Analyzer
ISA Sidestream
This Sidestream analyzer module measures EtCO2.
Analyzer
IRMA Airway
This adapter is connected to the IRMA Mainstream to measure
Adapter
EtCO2 of an adult or child.
(Adult/
Single patient use. Disposable.
Pediatric)
Nomoline
This is connected to the ISA Sidestream gas analyzer to measure
(Adult/
EtCO2.
Pediatric/
Single patient use. Disposable.
Infant)
Nomoline
Adapter This is connected to the ISA Sidestream gas analyzer to measure
(Adult/ EtCO2.
Pediatric/ Multi patient use. Disposable.
Infant)
Nomoline
Airway This is connected to the ISA Sidestream gas analyzer to measure
Adapter Set EtCO2.
(Adult/ Single patient use. Disposable.
Pediatric)
When charged, the battery module can supply power to the CU-
HD1 without the need for an external power source.
Battery Module
※ For details on attaching, detaching and charging the battery,
see "3.2.1 Installing & Charging Battery".
45
CU-HD1 Instructions for use
AC Power The power adapter can be connected to the rear of the CU-HD1
Adapter to charge it.
The car cigar jack can be connected to the rear of the CU-HD1
Car Cigar Jack
to charge it through the cigar jack in a car.
Power Adapter
The power adapter can supply power to the CU-CM1.
for CU-CM1
The bed rack allows the CU-HD1 to be secured onto the patient's
Bed Rack
bed.
SD Card
The SD card is used to save or export data generated by the
(Secure Digital
CU-HD1.
Card)
46
CU-HD1 Instructions for use
Paddles or pads designed specifically for children are recommended when using the
defibrillator on children aged 8 years or under or weighing 25 kg or below. In emergency
situations, pads for adults may be used on children.
Using the defibrillator on an adult with pads for children (without defibrillation energy
reduction module) may cause necrosis of the cardiac muscles.
ECG cables with AHA in the product name are generally ECG cables following the US naming
convention, while products with IEC in the product name are generally ECG cables following
the EU naming convention
47
CU-HD1 Instructions for use
For details regarding the operation of CU-HD1 besides the installation of accessories, refer to Chapters 4~7.
If the product is initially installed in an emergency situation, please check out if the product components are
properly installed after the product has been used or during the periodic checkup session.
If the floor surface and your hands are wet, you may get shocked. Move to a dry location first and install the
product.
Before using, turn on the device using the Rotary switch and check the charging status visually.
48
CU-HD1 Instructions for use
3.1 Unpacking
Take a careful look to see if there is any damage to the package container.
Check out whether there is any obvious damage to the device, which may have been caused during
transportation.
Check if all of the components and accessories have been accurately provided according to the
package item list.
49
CU-HD1 Instructions for use
Consist of accessories may be differed from an order. When unpacking, it is very important to
check to make sure all accessories you have placed an order are included
50
CU-HD1 Instructions for use
-- +
Finger latch
The internal battery pack of the Product is fully charged before leaving the factory. Upon
receiving the product, please charge the battery module.
To separate the battery from the main body, press the finger latches on both sides of the battery
and pull the battery out.
Avoid exposing battery module to hot, humid or wet conditions.
If the Low Battery status is indicated, please charge the battery module. When the battery is
being recharged, the Battery Recharge indicator will blink. When the recharging process has been
completed, the Battery Recharge indicator will shows green light.
To check the battery charging status, refer to the battery status displayed in the LCD screen.
To check the battery status in terms of the charge level, refer to “2.4.2 Battery Indication
Symbols” or use the Battery Level Indication button in the rear side of the battery.
For more information about the safety tips for battery usage, please refer to “11.6
Considerations for Handling Power and Battery”.
For more detail information about the charging battery, please refer to “10.2 Power
Management“.
51
CU-HD1 Instructions for use
-- +
--
-
Finger latch
To separate the AC power module from the main body, press the finger latches on both sides of
the AC power module and pull the AC power module out.
Avoid exposing AC power module to hot, humid or wet conditions.
If the SUPPLY MAINS is interrupted for more than 30 s, the subsequent operation does one of the
following:
52
CU-HD1 Instructions for use
Vendors and users should note that the AC power module has been rated for electromagnetic
compatibility for work use (A Class). The defibrillator is appropriate for use in places other than
homes.
Caution should be taken on the mounting position of the battery module and the AC power
module shown in the instruction manual.
The battery module is mountable in both the A and B slots, but the AC power module is
mountable only in the B slot.
For more information about the safety tips for the AC power module, please refer to “11.6
Considerations for Handling Power and Battery”.
53
CU-HD1 Instructions for use
The protruding part of the car cigar jack connector faces upward when mounting the car cigar
jack.
-- + -- +
The car cigar jack and AC power adapter are designed only to charge the battery. Therefore,
do not use them for running the device.
For more information, please refer to “11.6 Considerations for Handling Power and Battery”.
54
CU-HD1 Instructions for use
① In the left figure, pull the lever forward on the right side of the printer from the CU-HD1.
② The front cover of the printer will be open as shown in the figure on the right.
③ Place the printer paper into the printer, and pull out some paper.
④ Push the printer cover back to the default location until it makes a “click” sound.
⑤ If the Rotary switch is set to the monitor mode, the printer power amp will be light up green.
You can use the [Feed] button to take as much printer paper as you want
⑥ When replacing the printer paper, repeat steps ① and ②, then detach the printer paper
with your hand.
55
CU-HD1 Instructions for use
AC Adapter Port
3.3 Self-test
Whenever the product is turned on, it periodically initiates a self-test. This test is designed to ensure
that the whole system is ready for use in emergencies. The Self-test performs the battery state
checkup, the control system state checkup, and evaluates all functions provided by the product.
This product can also run a manual self-test. It is recommended to perform a manual self-test for
first-time use. For detailed contents, refer to “10.1 Self-test”.
Do not disconnect the battery pack during storage. The battery must be charged fully enough to be
turned ON during emergencies through the self-test.
If there are errors other than the battery error, contact CU Medical Systems, Inc. or an
authorized agent. If a “Low Battery” error occurs, recharge the battery pack or plug in the AC
power according to the instructions described in the Operation manual.
For long-term storage, do not store the product in connection with the disposable
defibrillation pads.
56
CU-HD1 Instructions for use
This defibrillator supports two AED modes: Auto Analysis Mode (Auto Analysis Mode ON) which
automatically analyzes the patient's ECG, and Manual Analysis Mode (Auto Analysis Mode OFF) which
analyzes ECG when the 'Analyze' soft key is pressed. In addition, the defibrillator provides voice
guidance to facilitate performance of CPR. In addition, the product provides a voice guide to facilitate
the CPR procedure.
Other medical devices that may be affected by defibrillation energy (strong electric shock) or
become an obstruction to the protection of defibrillator should be removed from the vicinity of
the patient.
The CU-HD1 is not intended to be used for supplementing the abnormal functionality of the
internal pacemaker. Therefore, if needed to recover the heart function of patients using the
internal pacemaker, consult a medical specialist.
The CU-HD1 does not have any functionality generating the alarm sound or warning message
saying whether an internal pacemaker is used for the patient or not.
To safely shut the CU-HD1 down while in AED (Automated External Defibrillator) mode, rotate
the rotary switch to the OFF position.
57
CU-HD1 Instructions for use
When removing the defibrillation pads and paddles connector from the main body, for
disconnection, turn a cable connector terminal in the unlocking direction, and pull it out from
the main connector.
58
CU-HD1 Instructions for use
Shock
Cable
AED Usage Da Time Ti Power
Connection
Mode Time te m status
Status
Lead e
Informat
Heart
ion
Rate
Text
Messa SpO2
ge
59
CU-HD1 Instructions for use
2) Setting
2.1) Auto Analysis
Press the menu knob from “Auto Analysis” at the “Main Menu” and then the following screen
will be prompted, where you can change the Auto Analyzing setting at the AED mode.
In Auto Analysis ON mode, patient ECG is automatically analyzed once the pads are correctly
connected to the patient.
In Auto Analysis OFF mode, patient ECG is not analyzed even when the pads are correctly
connected to the patient.
To analyze ECG, press the "Start Analysis" soft key.
60
CU-HD1 Instructions for use
2.2) CPR
Press the menu knob from “CPR” at the “Main Menu” and then the following screen will be
prompted, where you can change the CPR setting at the AED mode. From the CPR menu, you
can choose whether to turn on/off the CPR guide, or the method of CPR.
After changing the CPR settings, the changed values are applied by pressing the menu knob.
Select Exit from “CPR” menu and press the menu knob to retrieve the “Main Menu”.
61
CU-HD1 Instructions for use
2.4) Volume
This is a sub-menu to control the speaker volume of the CU-HD1.
2.5) Filter
At the “Filter” menu, you can set a bandwidth with which you can check out the ECG signal
detected by the CU-HD1.
Support for the function of filtering the LCD monitor and printer ECG signal, and individual
optional item is listed as follows.
62
CU-HD1 Instructions for use
Sector 1
As for the ECG signals displayed in the Sector 1, click the menu knob to choose the ECG
signal size from Auto Gain, 5mm/mV, 10mm/mV, and 20mm/mV.
Sector 2
As for the ECG signals displayed in the Sector 2, click the menu knob to choose the ECG
signal size from Auto Gain, 5mm/mV, 10mm/mV, and 20mm/mV.
63
CU-HD1 Instructions for use
※ Attach and connect the defibrillation pads in the order described below.
① Take off all the upper-body clothes including any under-garments.
② If the chest to which the pads are attached is too hairy, shave it using a razor or with
scissors.
③ Tear off the defibrillation pads package along the cutting line. Select the adult or pediatric
pads according to the patient.
64
CU-HD1 Instructions for use
④ Attaching pads – Following the directions indicated in the figure below. Attach the pads to
the upper body of the patient. Individual pads have a drawing showing the attaching
location. It is best to follow the exact locations.
If the pediatric pads are used in the AED mode, make sure they have the defibrillation energy
attenuation module.
Maintain the position where the pads are attached dry. If the patient’s skin surface is moist or
wet, a problem may occur when the device recognizes the patient and defibrillation energy
may leak during defibrillation.
If using the disposable pads, do not use the defibrillation-specific gel. The defibrillation-specific
gel must be used for the external defibrillation paddles.
Check for any damage to the pads and the packages of the pads as well as the expiration date.
If damaged or expired, discard the package without using it.
When the patient is less than 8 years old or weighs less than 25 kg (55 lb), use Pediatric
defibrillation pads. Do not delay therapy to determine the patient’s exact age or weight.
65
CU-HD1 Instructions for use
① Once attaching the pads to the patient correctly, his or her ECG is automatically analyzed.
② If the patient’s ECG analysis results require defibrillation, it delivers the defibrillation
electricity shock and reanalyzes the ECG automatically.
③ During CPR after the defibrillation electricity shock has been delivered, there is no ECG
analysis.
66
CU-HD1 Instructions for use
① Push the soft key called “Analyze” to analyze the patient’s ECG.
② As this time, the voice and text messages that are “If no pulse, press ‘Analyze’.” shall be
displayed.
③ Using the manual analysis mode, therefore, the user must use the device according to the
patient’s ECG on the LCD screen.
④ In addition, if the ECG analysis results in requiring defibrillation, it does not automatically
reanalyze the patient’s ECG after delivering the defibrillation electricity shock. Rather, press
the “Analyze” button again to reanalyze his or her ECG.
Do not analyze the patient ECG during patient movement. A patient must be motionless
during ECG analysis. Do not touch the patient during analysis. Cease all movement via
stretcher or vehicle before analyzing the ECG.
67
CU-HD1 Instructions for use
※ Effective shock level of the CU-HD1 is set at 200J for adults and 50J for children.
Do not allow defibrillation pads of paddles to touch each other or to touch other ECG
monitoring electrodes, lead wires, dressings, etc. Contact with metal objects may cause
electrical arcing and patient skin burns during defibrillation and may divert current away from
the heart.
While charging or carrying out the defibrillation, do not allow the operator or other individuals
to come into contact with the patient or any device connected to the patient.
Do not allow exposed portions of the patient’s body to come in contact with metal objects,
such as a bed frame, as unwanted pathways for defibrillation current may result.
If the Shock button is not pressed within 15 seconds after the charge energy has been charged,
the charged energy is discharged internally.
68
CU-HD1 Instructions for use
3) Performing CPR
As the function to help CPR, CU-HD1 provides the guidance on the CPR procedure with voice
and text messages.
At the AED mode, press the “Start CPR” soft button to perform CPR without defibrillation.
The CPR rate (Compression : Respiration) can be selected through the “CPR Type 30:2” / “CPR
Type 15:2” soft button.
While delivering the defibrillation energy, the patient’s ECG through the pads does not show
up in the screen. When it’s delivered, the Biphasic wave is shown.
If the patient's ECG changes to ECG that does not require defibrillation in the fully
charged state, the fully charged energy of ME equipment discharge to internal resistor
and is reanalyzed after CPR guide.
69
CU-HD1 Instructions for use
70
CU-HD1 Instructions for use
In the manual defibrillation mode, both paddles and pads can be used.
When the synchronized cardioversion function is used, the synchronization with the R wave will be
established and an electric shock will be delivered.
To safely shut the CU-HD1 down while in Manual Defibrillation mode, rotate the rotary switch
to the OFF position.
Defibrillating asystole can inhibit the recovery of natural pacemakers in the heart and
completely eliminate any chance of recovery. Asystole should not be routinely shocked. Begin
CPR.
To avoid stress to the defibrillator or the tester, never attempt to repeatedly charge and
discharge the defibrillator in rapid succession. If a need for repetitive testing arises, allow a
waiting period of at least 2 minutes after every third discharge.
71
CU-HD1 Instructions for use
To remove the paddles and the pad connector from the main body, remove these items by
rotating them in the unlock direction on the lock figure shown in the connector input terminal
of the main body.
72
CU-HD1 Instructions for use
Selected
Status
Charging
Lead Shock
Status Shock Hear
Informatio Energy
Display Times Rate
n
Blood
Pressur
e
Lead
SpO
Informatio
Impedance Status 2
n
Bar
Disarm
Soft Key
73
CU-HD1 Instructions for use
2) Setting
※ Synchronized Cardioversion
The synchronized cardioversion transfers energy in sync with the ‘R’ wave of the ECG signals measured.
Paddles and pads may be used for the synchronized cardioversion. Synchronized Cardioversion is
enabled in the manual defibrillation mode.
When using the supplied pediatric pads which are specified for the CU-HD1, the energy level
must not exceed 50J.
74
CU-HD1 Instructions for use
1) Defibrillation Pads
For how to use the pads, follow the same steps as explained in Section 4.1.3, which describes
pads attachment and connecting pads for AED.
75
CU-HD1 Instructions for use
The impedance connection status bar must be green or above. This may not be possible
according to a patient’s physical conditions. In such cases, maintain the status bar in yellow (5
bars) or above.
Rubbing the paddles together without applying enough conductive gel on the paddle
electrode surface may scratch or damage the surface.
After using the conductive gel, clean off the gel that may be left on the paddles with wet
clothes or gauze. If any leftover gel dries on the pads or paddles, it may cause problems in
future use.
Apply an ample amount of conductive gel on the paddle electrodes. Do not allow the gel to
dry or accumulate in between the chest wall and paddle electrodes. Failure to remove gel
residue from previous use may result in burns to the patient or a reduction of delivered energy.
When the impedance connection status bar is not in the green area or above, incorrect
defibrillation or vital sign measurement may occur. The status bar must be in the green area or
above for correct defibrillation.
76
CU-HD1 Instructions for use
① Push the yellow switches on the paddles in ② While pressing the yellow switches, pull
the direction of the arrow as indicated. them in the direction of the arrow.
77
CU-HD1 Instructions for use
1) Selecting Energy( )
[Default Screen after Pad Connection] [Default Screen after Paddle Connection]
2) Charging Energy
Using Pads: Press the Charge button ( ) on the front of the defibrillator to start charging
energy.
Using Paddles: Press the yellow button next to the paddle handle to start charging.
[Charging Screen when using pads] [Charging Screen when using paddles]
3) Performing Defibrillation
Using disposable pads: Press the Shock button ( ) on the front of the defibrillator to deliver
defibrillation energy to the patient.
Using Paddles: Press the two orange buttons on the front of the paddle handles at the same time
to deliver defibrillation energy to the patient.
[Defibrillation Screen when using pads] [Defibrillation Screen when using paddles]
78
CU-HD1 Instructions for use
If the energy level is changed while charging the defibrillation energy, charging will be
canceled. To charge the device again at the newly selected energy level, press the Charge
button again.
When the paddle is used and you press the Shock button while the impedance is not
recognized after charging is finished, the charged energy will be discharged internally. Charge
the device again by pressing the Charge button on the paddle handle, then carry out the
defibrillation.
When using the synchronized cardiac pacing function, check that the marker indication,
position and heart rate are consistent.
The white inverted triangle mark indicates the position of the R wave signal which is measured
through synchronization.
79
CU-HD1 Instructions for use
Pressing the "Sync" button above the rotary switch lights the lamp in blue and activates the
synchronized cardiac pacing mode. Pressing the "Sync" button again dims the blue light and
deactivates the synchronized cardiac pacing mode.
If markers do not appear above the R waveform, select another ECG lead. If synchronization
markers do not appear, it means that the R wave could not be detected and therefore
synchronized cardiac pacing energy cannot be delivered.
If there is interference generated by patient movement while attached with the paddles or pads, or
external contact, it may sense the R-wave and deliver the defibrillation energy to the patient.
Therefore, avoid any contact with the patient during pacing.
Since defibrillation may bring harm to the performer or surrounding people. Make sure to keep a
safe distance from the patient and any electronic devices and conductive metals connected to the
patient during defibrillation.
Poor adherence and/or air under the defibrillation pads can lead to the possibility of arcing and
skin burns.
After pressing the Shock button, keep hands away from the electrode plates.
Medical electrical equipment which does not incorporate defibrillator protection should be
disconnected during defibrillation.
Do not allow defibrillation pads of paddles to touch each other or to touch other ECG monitoring
electrodes, lead wires, dressings, etc. Contact with metal objects may cause electrical arcing and
patient skin burns during defibrillation and may divert current away from the heart.
80
CU-HD1 Instructions for use
Yellow alarm
Heart rate measurement falls
HR Low Medium message
below the minimum set value.
with alert sound
Yellow alarm
Pulse measurement exceeds the
Pulse High Medium message
maximum set value.
with alert sound
Yellow alarm
Pulse measurement falls below the
Pulse Low Medium message
minimum set value.
with alert sound
SpO2
Yellow alarm
SpO2 measurement exceeds the
SpO2 High Medium message
maximum set value.
with alert sound
Yellow alarm
SpO2 measurement falls below the
SpO2 Low Medium message
minimum set value.
with alert sound
81
CU-HD1 Instructions for use
Alarm
Classification Priority Indication Condition
Message
Yellow alarm
EtCO2 measurement exceeds the
EtCO2 High Medium message
maximum set value.
with alert sound
Yellow alarm
EtCO2 measurement falls below
EtCO2 Low Medium message
EtCO2 the minimum set value.
with alert sound
Yellow alarm
Respiration Respiration rate measurement
Medium message
Rate High exceeds the maximum set value.
with alert sound
Yellow alarm
Respiration Respiration rate measurement falls
Medium message
Rate Low below the minimum set value.
with alert sound
82
CU-HD1 Instructions for use
2) Technical Alarms
Alarm Message Priority Indication Condition
Turquoise alarm
Real-time Bluetooth transmission failed or
Transmission failed Low message
12-ch Bluetooth transmission failed.
with alert sound
Turquoise alarm
Low battery level. Low message Low battery level.
with alert sound
Turquoise alarm
NIBP Measurement
Low message NIBP measurement failure.
Failure
with alert sound
Turquoise alarm
NIBP signal has
Low message Oscillometric signal has noise.
noise artifacts
with alert sound
Turquoise alarm
NIBP Pneumatic NIBP operation is interrupted by bent
Low message
Blockage tubes, etc.
with alert sound
Turquoise alarm
NIBP air leak or
Low message Air is leaking from tubes, etc.
loose cuff
with alert sound
Turquoise alarm
NIBP Cuff
Low message NIBP is over-pressurized.
Overpressure
with alert sound
Turquoise alarm
There is a problem with NIBP module
NIBP Error Low message
operation.
with alert sound
Turquoise alarm
NIBP Equipment
Low message NIBP device is faulty.
Malfunction
with alert sound
Turquoise alarm
SpO2 Error Low message There is a problem with the SpO2 module.
with alert sound
Turquoise alarm
Unidentified
Low message Invalid accessory ID.
accessories
with alert sound
83
CU-HD1 Instructions for use
Turquoise alarm
CO2 : Out of range Measurement value of the CO2 sensor is
Low message
accuracy outside the reference accuracy range.
with alert sound
Turquoise alarm
CO2 : Need zero
Low message The CO2 sensor requires zero calibration.
calibration
with alert sound
Turquoise alarm
Software Error Low message The CO2 sensor has a software error.
with alert sound
Turquoise alarm
Hardware Error Low message The CO2 sensor has a hardware error.
with alert sound
Turquoise alarm
Motor Speed out of The CO2 sensor motor exceeds the speed
Low message
bounds range.
with alert sound
Turquoise alarm
Factory calibration There is an error with the CO2 sensor
Low message
data loss calibration value.
with alert sound
84
CU-HD1 Instructions for use
An electric stimulus is delivered through the pads attached to the patient in a noninvasive manner.
The pacing method consists of two modes: ‘Demand mode’ that the pacing signals are transferred to
when the patient’s heart rate is slower than the preset pacing rate, and the ‘Fixed mode’ where the electric
stimulus is delivered to the patient at a fixed heart rate.
Do not deliver any defibrillation energy to any patients using the pacing function.
If defibrillation is necessary, remove the cables connected with the pacing electrodes before
carrying out the defibrillation.
Do not use the noninvasive pacer to any patients using the internal pacemaker.
85
CU-HD1 Instructions for use
⑥ ⑤
③ ④
① ②
Pacer Mode Shows whether the device is running in the Pacer mode.
①Pacer Mode Shows the Pacer’s mode (Fixed/Demand)
②Pacer Status Shows whether the pacer is running or stopped.
③Current Shows the current to be delivered.
④Pacing Rate Shows the pacing rate (pacing times per minute).
⑤R-Sync Mark Shows the R-Sync mark in sensing the R-wave while analyzing the ECG.
⑥Pacing Display Shows a white indication when delivering the pacing pulse to the patient.
During pacing, heart rate is shown on the screen as '---' due to inaccurate ECG.
Since the diagnosis function on patient conditions is not offered by pacing, continue to watch
the patient during pacing.
86
CU-HD1 Instructions for use
For how to connect and attach the pad, refer to "4.1.1 Connecting to the Device" and "4.1.3
Attaching and Connecting the Defibrillation Pads".
Attach the pads on the positions as shown in the figure below according to the patient’s
condition and circumstance.
87
CU-HD1 Instructions for use
For connecting the ECG cable, refer to “7.1.1.1 Connecting ECG Cable”.
Remove hair or foreign substances on the patient’s body where the electrodes are attached
using a razor or scissors.
Attach the ECG electrodes within an appropriate distance from the pads. If the ECG electrodes
are placed close to the pads, the ECG signals may be distorted due to the pacing current. For
the attachment position, refer to the figure above.
Maintain a proper distance between the pacing pads and the conductive part of the ECG
electrodes, so as not to attach them together.
If the pads are detached from the patient, a confirmation alarm will be issued.
88
CU-HD1 Instructions for use
Disposable defibrillation pads may be used for non-invasive demand mode pacing. And for details
of how to attach disposable defibrillation pads to the patient, refer to Section “4.1.3 Attaching
the Defibrillation Pads”.
Maintain a proper distance between the pacing pads and the conductive part of the ECG
electrodes, so as not to attach them together.
The patient’s skin where the pads and the electrodes are attached should be kept dry. The
patient’s skin should be dry in order not to affect the measurement of the ECG signal. In such
a case, no current will leak and/or the adhesive strength of the pads will not be maintained
during pacing.
Before using the pads and the electrodes, check the expiration date and any damage to the
packaging. If damaged or expired, discard the package without using it.
If pacing is carried out for a long period of time, replace the pads periodically.
If the pads are detached from the patient, a confirmation alarm will be issued.
89
CU-HD1 Instructions for use
③ Using the Rate and Current buttons on the pacer menu, the pacing rate and pacing current
can be adjusted.
④ To start pacing, use the Start/Stop button. To stop it during pacing, stop it using the
Start/Stop button.
⑥ Check the patient’s heart rate with respect to pacing through the ECG, SpO2, pulse, NIBP,
and EtCO2. If the heart rate is not enough after checking his or her symptoms, increase the
⑦ Patient conditions may require a change in the current level as time goes on. During pacing,
Do not touch the patient nor have the patient touch any devices while the pacing electric
stimulus is active. Otherwise, the ECG will be distorted, impacting the pacing.
If touching the patient to check the heart rate of the patient while delivering the pacing energy
to him or her, it may be possible to be exposed to the pacing current.
90
CU-HD1 Instructions for use
③ Using the Rate and Current buttons on the pacer menu, the pacing rate and pacing current
can be adjusted.
⑤ Check the patient’s heart rate. If no pulse is recognized, increase the current level till the
heart rate is sensed, and gradually adjust the current to the minimum level where the pulse
is detected.
It’s recommended to use the Demand Pacing mode. The Fixed mode is normally used when
there is ECG interference or noise that makes it difficult to sense the reliable R-wave in the
Demand mode.
91
CU-HD1 Instructions for use
92
CU-HD1 Instructions for use
Classification Alarm
Priority Indication Condition
Message
While pacing is being carried out, an ECG-related alarm will not be issued.
93
CU-HD1 Instructions for use
2) Technical Alarms
Alarm Message Priority Indication Condition
94
CU-HD1 Instructions for use
Motor Speed out of Turquoise alarm message The CO2 sensor motor exceeds the
Low
bounds with alert sound speed range.
Factory calibration Turquoise alarm message There is an error with the CO2 sensor
Low
data loss with alert sound calibration value.
95
CU-HD1 Instructions for use
In addition, with the alarm setting on the menu, an alarm is issued whenever an abnormality of the
patient’s ECG is detected, leading to appropriate treatment.
The 12-lead ECG measures and patient monitoring information can be checked by connecting into a
computer through the Bluetooth communication. For details of information transfer, refer to
“Chapter 9_Communication and Data Management”.
If attaching the pads to the patient to use the defibrillation function while measuring the ECG
in the patient monitoring mode, make sure that the electrodes used in measuring the ECG and
defibrillation paddles or pads do not make any contact.
If it is necessary to monitor for a long period of time, replace the ECG electrodes or the pads
periodically.
To safely shut the CU-HD1 down while in Monitor mode, rotate the rotary switch to the OFF
position.
96
CU-HD1 Instructions for use
Heart Rate
Alarm Limit
NIBP Alarm
Sign
Systolic/Diastolic
Lead
Information NIBP Alarm
Limit
Mean Blood
Pressure
SpO2
Alarm Sign
SpO2
Alarm
Limit
While printing, pressing the lead change button in Sector 1 stops the printing.
If only Sector 1 is printing, the lead for Sector 2 can be changed while printing.
If both Sector 1 and Sector 2 are printing, the lead for Sector 2 cannot be changed while
printing.
97
CU-HD1 Instructions for use
The equipment analyzes ECG and provides alarms based on heart rate, ventricular fibrillation or
ventricular tachycardia.
For ECG monitoring over long periods or for more precise ECG measurement, make sure to select
an appropriate channel (I, II, III, aVR, aVF, aVL, V1 - V6) based on the patient's condition as
monitored with ECG electrodes.
The patient's ECG obtained through the amplification and operation of differential voltages
between minute electromotive forces occurred due to the cardiac activities measured through the
electrodes is displayed on the monitor as waveforms and values. When this value exceeds the
range of set values for alarm, an alarm will be issued, indicating the abnormal condition of the
patient.
98
CU-HD1 Instructions for use
If you would like to measure ECG by using 3-lead, 5-lead, 10-lead, all of the cables must be
inserted into the ECG terminal.
7.1.1.2 Setting
1) Inputting the information for patient monitoring
In Monitor Mode, use the Menu knob to enter patient information.
For more details on entering information, see "8.1 Patient Information".
2) Filter
At the “Filter” menu, you can set a bandwidth with which you can check out the ECG signal
detected by the CU-HD1.
Support for the function of filtering the LCD monitor and printer ECG signal, and individual
optional item is listed as follows.
99
CU-HD1 Instructions for use
3) ECG Gain
This is the menu item that takes control of the ECG sensitivity degree. If the ECG signal is too
high or too low, you can change the signal level to an extent that you can easily verify it.
Default setting is 10mm/mV.
Sector 1
As for the ECG signals displayed in the Sector 1, click the menu knob to choose the ECG
signal size from Auto Gain, 5mm/mV, 10mm/mV, and 20mm/mV.
Sector 2
As for the ECG signals displayed in the Sector 2, click the menu knob to choose the ECG
signal size from Auto Gain, 5mm/mV, 10mm/mV, and 20mm/mV.
Changes made to the settings in the menu are not saved. To change the default settings, see
the Service Manual.
100
CU-HD1 Instructions for use
Do not allow the conductive unit of the electrodes, including the neutral electrode and relevant
connections, to come into contact with any other conductors, including the ground.
When using ECG electrodes, always check the expiry date. Also, remove the ECG electrodes
from sealed packaging immediately before use.
When using the expired disposable ECG electrode and the disposable ECG electrode whose
envelope is lost, the accurate ECG measurement is not guaranteed.
101
CU-HD1 Instructions for use
2) 5-Lead
RA/R Underneath the right shoulder collarbone,
or the right arm
LA / L Underneath the left shoulder collarbone, or
the left arm
LL / F Left lower abdomen or left leg
RL/N Right lower abdomen
V/C Among 12-lead locations, select the desired
one from V1 ~ V6
3) 12-Lead
RA/R Underneath the right shoulder collarbone, or
the right arm
LA / L Underneath the left shoulder collarbone, or
the left arm
LL / F Left lower abdomen or left leg
RL/N Right lower abdomen
V1/C1 Right 4th sterna intercostal
102
CU-HD1 Instructions for use
Cable Connection
Status
Lead
Information
Lead
Information
1) Selecting Lead
Type of ECG Type of applicable lead
3-Led I, II, III
5-Lead I, II, III, aVR, aVL, aVF, V
12-Lead I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
Setting "Device Management / ECG Size" to “Auto Gain” automatically shows the adequately
sized ECG waveform on the screen.
Should a “Lead Fault” message appear, check the ECG cable or electrode connections. If the
problem persists, replace the cables and electrodes.
Dotted lines on the ECG denote invalid ECG signals in the waveform sector.
In this case, check that an appropriate lead has been selected, and that the pads, ECG
electrode cable and electrodes are attached correctly.
When replacing the lead cables, dotted lines appear momentarily.
Dotted lines appear when the “Lead Fault” error occurs.
103
CU-HD1 Instructions for use
◎ Layout
Push the “Start 12-lead” soft key. And check the ECG signal while the patient keeps his/her
body in the same position.
If needed, press the MENU key to enter patient information.
Do not connect many devices to the patient at once. The leakage limit of the current may be
exceeded.
Adjusting the ECG waveform size on the screen does not affect the ECG data used in
arrhythmia analysis.
104
CU-HD1 Instructions for use
In case of a patient implanted with a pacemaker, the heart rate meter may count the
pacemaker rate even if cardiac arrest or other arrhythmias occurs. To avoid a shock hazard and
interference from nearby electrical equipment, keep electrodes and patient cables away from
grounded metal and other electrical equipment.
ECG monitoring of the patient can be measured accurately when the patient is motionless.
Physiological alarms may be triggered by environmental factors.
2) Technical Alarms
Alarm Message Priority Indication Condition
Turquoise alarm
Transmission Real-time Bluetooth transmission failed
Low message
failed or 12-ch Bluetooth transmission failed.
with alert sound
105
CU-HD1 Instructions for use
Arrhythmias-related alarms only occur when lead II of the pads or the ECG electrodes is
selected for Sector 1 in Monitor mode. To monitor the patient's ECG, set the pads or ECG lead
II for the lead section from Sector 1 (yellow box in the figure below).
106
CU-HD1 Instructions for use
2) Asystole
This is a screen to set it to On/Off to generate the alarm sound when detecting Asystole.
107
CU-HD1 Instructions for use
3) Heart Rate
This is an item to set the alarm according to Heart Rate (HR) measured through the ECG cable
or pad.
Alarm On/Off
You can set it to On/Off to generate an alarm sound when the patient HR goes beyond
the assigned numeric setting.
Upper Limit
This is the maximum value of patient HR that generates an alarm sound, which can be
changed by 5bpm with the Menu selection button. Numeric values you can set range
from 35 to 300bpm.
Lower Limit
This is the minimum value of patient HR that generates an alarm sound, which can be
changed by 5bpm with the Menu selection button. The numeric values that you can set
ranges from 30 to 295bpm.
108
CU-HD1 Instructions for use
The product uses the principle of spectrophotometry. SpO2 is measured percutaneously using a
difference in the optical density occurring when two lights at a natural wavelength pass through
material with a different concentration, while pulse waveform, value of SpO2 concentration, and
heart rate are displayed on the monitor through the operation. When this value exceeds the
range of set values for alarm, an alarm will be issued, indicating the abnormal condition of the
patient.
109
CU-HD1 Instructions for use
110
CU-HD1 Instructions for use
When measuring SpO2 only, the screen shows [SpO2 Measurement Screen] above, along with a
pulse. When pads, paddles, or ECG cables are connected (not in Lead Fault condition) and ECG is
measured, pulse is shown in HR bpm.
SpO2 is one of the supplementary measures to check the patient‘s status; the measurement value is
subject to change according to the patient’s status and ambient condition. Measurement values may
change in the following cases.
- Hypothermic patient or Acidotic patient
- Patients that are receiving a photosensitive drug
- Patients that are receiving vasoconstrictor medications
- Patients that have poor circulation
- Hemoglobin malfunction patient
- Severe anemia patient
- Elevated levels of bilirubin
- Interference by carboxyhemoglobin and methemoglobin
- Injected dyes such as methylene blue
- Exposure to excessive illumination such as surgical lamps, bilirubin lamps, fluorescent lights, infrared
heating lamps or direct sunlight
- Equipment with an inaccurate sensor
111
CU-HD1 Instructions for use
Always use DS100A sensors and extension cables approved by CU Medical Systems, Inc. Check
compatibility before use of sensors or extension cables from other manufacturers, as they may
affect defibrillator performance.
The CU-HD1 is calibrated to display functional oxygen saturation.
Incorrect usage of the sensors under excessive pressure for long periods may result in pressure
damage.
Some models of commercially available bench top functional testers and patient simulators can
be used to verify the proper functionality of pulse oximeter sensors, cables and monitors. See
the individual testing device's operation manual for the procedures specific to the model of
tester that you are using.
SpO2 measurement accuracy can only be evaluated in vivo by comparing pulse oximeter
readings with values traceable to SpO2 measurements obtained from simultaneously sampled
arterial blood using a laboratory CO-oximeter.
Functional test equipment designed for SpO2 testing cannot be used to assess the accuracy of
the SpO2 readings.
See the sensor’s operation manual for the maximum temperature possible at the sensor-skin
interface and other information such as intended patient population, sensor application sites
and use criteria.
Information about wavelength range can be useful to clinicians, especially those performing
photodynamic therapy
112
CU-HD1 Instructions for use
1) Physiological Alarms
Alarm Message Priority Indication Condition
2) Technical Alarms
Alarm Message Priority Indication Condition
Turquoise alarm
There is a problem with the SpO2
SpO2 Error Low message
module.
with alert sound
Alarm On/Off
You can set it to On/Off to generate the alarm sound when the measured SpO2 value goes
beyond the assigned numeric range.
Upper Limit
This is the maximum value of SpO2 that generates the alarm sound, which can be changed
by 1% with the Menu selection button. The numeric values that you can set range from 2
to 100%.
113
CU-HD1 Instructions for use
Lower Limit
This is the minimum value of SpO2 that generates the alarm sound, which can be changed
by 1% with the Menu selection button. The numeric values that you can set ranges from 1
to 99%.
Alarm On/Off
You can set it to On/Off to generate an alarm sound when the patient HR goes beyond
the assigned numeric setting.
Upper Limit
This is the maximum value of patient HR that generates an alarm sound, which can be
changed by 5bpm with the Menu selection button. Numeric values you can set range
from 35 to 300bpm.
Lower Limit
This is the minimum value of patient HR that generates an alarm sound, which can be
changed by 5bpm with the Menu selection button. The numeric values that you can set
ranges from 30 to 295bpm.
114
CU-HD1 Instructions for use
1) Connecting the Cuff and the Connection Tube for NIBP Measuring
115
CU-HD1 Instructions for use
Before measurement, choose the correct Patient Category from the menu. It is important for the
safety of the patient that you choose the correct “Patient Category” (Adult, Pediatric, Neonate) since
the maximum expansion pressure is determined based on your selection.
For details on changing Patient Category, see Menu. The default setting is "Adult".
③ Ensure the patient is lying down or comfortably seated with legs uncrossed, both feet on
the floor, and back supported. The limb to be used for NIBP measurement should be
relaxed, extended, and placed on a smooth surface for support. The operator position is
not restricted during NIBP measurement.
④ Sqeeze as much air from the cuff as possible before placing it on the patient.
⑤ Place the suff 2 to 5cm above the elbow crease.
⑥ Adjust the cuff so that the artery marker on the cuff is over the artery, pointing to the
hand or foot.
⑦ Check that the cuff ends between the range lines marked on the cuff.
⑧ If they do not line up, use a different size cuff.
⑨ Wrap the deflated cuff snugly around the limb without impeding blood flow.
⑩ Ensure that the hose is routed th avoid kinking or compression.
116
CU-HD1 Instructions for use
Avoid using the cuff on parts of the body where damage to the tissue from external air
pressure can be expected.
Never use anything other than the cuff supplied with the defibrillator. CU Medical Systems, Inc.
will not be liable for any problems caused by the use of products other than the cuff supplied.
Neonatal brood pressure measurements must always use a 3 meter patient hose in order to
avoid overpressure errors caused by a lack of air volume within the overall pneumatic system.
Select and use a proper cuff according to the patient for accurate measurement of noninvasive
blood pressure (NIBP) by referring to “2.7 Accessories”. If a cuff which is too small is used, the
measured blood pressure will be higher than the actual blood pressure of the patient, and if a
cuff which is too large is used, the measured blood pressure will be lower than the actual
blood pressure of the patient.
When the blood pressure is measured while the cuff is loose or air remains in the cuff, the
measured blood pressure may be higher than the actual blood pressure. Use the cuff by
adhering it to the patient as closely as possible while it is being deflated.
Accuracy of any blood pressure measurement may be affected by the position of the subject,
his or her physical condition and use outside of the operating instructions detailed in this
manual. Interpretation of blood pressure measurements should be made only by a physician or
trained medical staff.
Intended patient populations include adult, pediatric and neonate patients. Safety and
effectiveness have not been proven on pregnant women.
For neonate populations, the clinical effectiveness of this device has not been established in
the presence of dysrhythmias. Arterial reference sites included femoral, umbilical and radial
arteries.
Mode Description
117
CU-HD1 Instructions for use
② Measuring NIBP
Pressing the “NIBP” button starts the measurement according to the selected mode. During
measurement, the cuff attached to the patient expands and pressure applied is shown on the
screen. When measurement is complete, the systolic and diastolic blood pressures (mean blood
pressure) are displayed on the screen. Pressing the NIBP button once during measurement
stops the measurement.
Use clinical judgment to decide whether or not to perform automatic blood pressure
measurements on patients with severe blood clotting disorders because of the risk of
hematoma in the limb wearing the cuff.
Do not apply the cuff to a limb that has an intravenous infusion or arterial catheter in place.
This could cause tissue damage around the catheter when the infusion is slowed or blocked
during cuff inflation.
Do not attach the cuff to a limb being used for IV infusions or any other intravascular access,
therapy or an arterio-venous (A-V) shunt. The cuff inflation can temporarily block blood flow,
potentially causing harm to the patient.
Prolonged series of NIBP measurements in automatic mode may be associated with purpura,
ischemia and neuropathy in the limb wearing the cuff. When monitoring a patient, examine the
extremities of the limb frequently for normal color, warmth and sensitivity. If any abnormality is
observed, stop the blood pressure measurements immediately
Care should be taken when using an NIBP measurement mode on patients with decreased
consciousness, neuropathy, irregular cardiac rhythm, labile high blood pressure, increased arm
activity, or arterial insufficiency especially if the unit is utilized for a prolonged period. Pay
particular attention to unconscious patients since they cannot alert you if the pain is present.
Accurate measurements are not guaranteed in circumstances such as dramatically fluctuating
barometric pressure (inside airplane or elevator) or hyperbaric chamber.
Do not allow the NIBP tube to become kinked or crushed. This could prevent normal cuff
deflation resulting in patient injury due to prolonged restriction of blood flow.
Do not place cuff over a wound, as this can cause further injury.
118
CU-HD1 Instructions for use
Repeated use of SpO2 measurements on the same patient over a short time interval can affect
blood pressure readings, limit circulation to the limb, and cause injury to the patient.
Observe the patient’s limb periodically to ensure that circulation is not impaired for a
prolonged period of time.
Do not place the NIBP cuff on the same arm or leg as an SpO2 sensor. Inflation of the cuff
causes the SpO2 monitor to read incorrectly.
Do not attach the NIBP cuff to the arm on the side of the body where a mastectomy has been
performed. Attach the cuff on the other arm for blood pressure measurement.
The pulse measured during NIBP measurement is not displayed on the screen. For patient
heart rate information, use the SpO2 sensor or ECG electrodes.
A major air leak can be preventing cuff inflation. Check hose and cuff connections, replace a
defective hose or cuff, as necessary, and reattempt NIBP measurement. Check hose for kinks.
Kinked hose can be preventing the correct measurement.
Do not compress or restrict pressure tubes during an NIBP measurement.
If a spill occurs and liquid appears inside the tubing, contact your service personnel.
If the circumference of the upper arm is less than 8-13 cm, noninvasive blood pressure cannot
be measured for the patient.
Allergic exanthema (symptomatic eruption) in the area of the cuff may result, including the
formation of urticaria (allergic reaction including raised edematous patches of skin or mucous
membranes and intense itching) caused by the fabric material of the cuff.
Petechia (a minute reddish or purplish spot containing blood that appears in the skin)
formation or Rumple-Leede phenomenon (multiple petechia) on the forearm following the
application of the cuff, which may lead to Idiopathic thrombocytopenia (spontaneous persistent
decrease in the number of platelets associated with hemorrhagic conditions) or phlebitis
(inflammation of a vein) may be observed.
119
CU-HD1 Instructions for use
An NIBP alarm is issued due to an occasional measurement. Once the alarm is issued, the EXIT
button will be activated immediately.
1) Physiological Alarms
120
CU-HD1 Instructions for use
2) Technical Alarms
NIBP
Turquoise alarm message
Measurement Low NIBP measurement failure.
with alert sound
Failure
NIBP signal has Turquoise alarm message
Low Oscillometric signal has noise.
noise artifacts with alert sound
121
CU-HD1 Instructions for use
2) Alarm On/Off
Turns the alarm on or off for noninvasive blood pressure. You can set the alarm only for the
selected items out of the results of noninvasive blood pressure measuring.
122
CU-HD1 Instructions for use
Changes made to the alarm settings in the menu are not automatically saved. When the
equipment reboots, the default alarm settings are restored. To change the default alarm
settings, they must be changed and saved in Administrator Mode. For details on entering
Administrator Mode or changing default alarm settings, see the Service Manual.
Calibrate the NIBP measurement device every year by contacting CU Medical Systems, Inc. for
service.
If the device does not work properly, see Troubleshooting for Problems Related to NIBP
Measuring. For other problems, please contact the CU Medical Systems Service Center.
123
CU-HD1 Instructions for use
The CU-CM1 allows you to measure CO2, EtCO2, and RR (Respiration Rate). CO2 can be measured
in Patient Monitoring Mode, Pacer Mode and Manual Mode on the CU-HD1.
The CU-CM1 offers two methods for measuring CO2, namely Mainstream and Sidestream and it is
measured through infrared absorption of CO2.
CO2 continues to be measured through a sampling line or the adapter. EtCO 2 is measured
through the concentration of CO2 at every end of the patient's exhalation. It can be a critical
means of monitoring the respiration of the patient. For example, it allows you to see if chest
compressions were administered properly to a patient during CPR, and is used to confirm
successful intubation. It also allows you to check respiratory ailments in the lungs or bronchi.
CU-CM1 is not compatible with any other products other than CU-HD1.
Either Mainstream or Sidestream measurement method can be selected and used.
When cleaning the IRMA Mainstream gas analyzer, always wipe it with a soft cloth. Remove the
airway adapter before cleaning the analyzer.
Wipe the outside surface using a formula of 70% or less ethanol and a 70% or less isopropyl
alcohol.
This device does not provide an automatic air pressure compensation function.
If the zero calibration of the CO2 sensor is necessary, if correct air pressure is not set of if the
pre-heating time is not enough, the EtCO2 and RR measurements may not be accurate. A
sensor application error or environmental conditions may also affect the measurements.
Accessories of IRMA Mainstream gas analyzer and ISA Sidestream gas analyzer can only be
used once. Do not clean, sterilize, or reuse them.
Check if CO2 waveform (capnogram) is appropriate on the monitor screen. Always check if the
patient's ventilator is connected correctly.
If it is loosely connected or damaged, the patient's breathing may become abnormal or the
measurement of respiratory gases may become inaccurate. Connect it securely and check the
connected part for leaks.
If spray medicine or anesthetic gas is used, the CO2 measurement may be inaccurate.
Carry out zero calibration while the airway adapter is not connected to the patient.
124
CU-HD1 Instructions for use
When selecting an accessory, always check the patient category (adult, pediatric, neonate),
patient airway (ventilation), and ventilation status (humidified ventilation).
Use only accessories provided by CU Medical Systems, Inc. for correct and accurate
measurement of CO2.
125
CU-HD1 Instructions for use
7.4.1.2 Setting
The CU-CM1 measures the amount of CO2 and transmits the measured value to the CU-HD1 via
Bluetooth. The CU-HD1 displays it on the LCD screen. CO2 can be measured in Patient
Monitoring Mode, Pacer Mode and Manual Mode on the CU-HD1.
First, pairing the CU-CM1 with the CU-HD1 is needed for using CU-CM1. For details on pairing
the CU-CM1 is shown below.
1) Pairing
If a new CU-CM1 is purchased or replaced, it can be used only after pairing with CU-HD1.
① Select "CO2 Sensor Initialization" in Bluetooth as below for pairing via Bluetooth. (“MENU”
→ “Device Management” → “Bluetooth” → “CO2 Sensor Initialization”)
CU-HD1 may be obstructed by other devices, even if those other devices meet the KN
emission requirements.
126
CU-HD1 Instructions for use
② When the CU-CM1 is turned on and is paired with the CU-HD1 via Bluetooth, a long
beep will sound three times. The following window will pop up on the CU-HD1 screen to
show that pairing is complete.
Communication is not possible with devices that are not designated by CU Medical Systems,
Inc.
Before pairing the CU-HD1, turn on the power of the device to be connected.
③ Select CO2 by pressing the lead selection (Sector 2) button on the CU-HD1. When Sector
2 is selected as CO2, the CO2 measurement screen appears and the CU-HD1 attempts to
connect to the CU-CM1.
127
CU-HD1 Instructions for use
④ When you turn the CU-CM1 on by pressing the power button, the CU-HD1 is connected
to the CU-CM1. When the CU-HD1 is connected to the CU-CM1, the screen shows
information on the Mainstream and Sidestream connections, as well as the battery status
of the CU-CM1.
,,
,
,
2) CO2 Display
You can change the sweep speed, scale, and unit by selecting “MENU → Device Management
→ Etc. → CO2 → CO2 Display”.
128
CU-HD1 Instructions for use
2.3) Unit
In Menu on the CU-HD1, you can change the display unit in the EtCO2. You can change it
within the following range.
- mmHg (default setting)
- %
129
CU-HD1 Instructions for use
② Snap the IRMA probe on top of a new IRMA airway adapter. It will click into place when
properly seated.
③ If necessary, perform zero calibration for the analyzer on the CU-HD1. (MENU Device
Management Etc. CO2 Zero Calibration)
④ During the calibration a message “Zero calibration in progress” will appear on CU-HD1
and the green LED on the IRMA Mainstream Analyzer will blink.
⑤ When zero calibration is completed, the message will disappear and the green LED will
be steady.
⑥ Connet IRMA airway adapter 15mm male connector to the breathing circuit Y-piece.
130
CU-HD1 Instructions for use
⑦ Connet the IRMA airway adapter 15mm female connector to the patient’s endotracheal
tube.
⑧ Connect the patient’s endotracheal tube to the patient, and measure the CO2.
Perform zero calibration for the analyzer while the Airway adapter is not connected to the
patient.
For more accurate measuring CO2, perform the zero calibration whenever you replace the
Airway adapter.
When zero calibrating the IRMA Mainstream Analyzer, additional calibration gas is not required.
Before measuring the amount of CO2 using the IRMA Mainstream Analyzer, it is required for
the analyzer to warm-up for 10 seconds after you turn it on. For accurate measurement, use
the equipment approximately 30 seconds after the IRMA Mainstream gas analyzer has been
powered on.
131
CU-HD1 Instructions for use
③ After ensuring the nasal cavity is clear, insert the tip of the nasal cvity Nomoline into the
patient’s nostril for measurement.
Before measuring the amount of CO2 using the ISA Sidestream Analyzer, it is required for the
analyzer to warm-up for 1 minutes after you turn it on. For accurate measurement, use the
equipment approximately 2 minutes after the ISA Sidestream gas analyzer has been powered
on.
If the "Sampling line is interrupted" message appears or the measurement value begins to
show error, replace the nomoline connector.
Disposable accessories, such as disposable nomoline adapter set and T adapter set, should be
used for only one patient.
If the patient's nostrils are blocked partially or completely or the patient is breathing using
his/her mouth when the ISA Sidestream gas analyzer is used, the EtCO2 measurement may
become lower.
Reflux of gastric contents, mucus, pulmonary edema fluid or endotracheal epinephrine
introduced into the detector may increase airway resistance and affect breathing. If such a case
occurs, dispose of the accessories.
EtCO2 should be measured after ensuring the nomoline airway is not bent and a physical
blockade is not occurring as a result of the patient lying on top of nomoline airway.
Using a flammable anesthetic mixed with air, oxygen or nitric oxide poses a danger of
explosion. If it is exposed directly to laser, the ESU device, or high heat while oxygen exists, the
nomoline airway may ignite. If a procedure using laser, an electrosurgery device, or high heat
is carried out on the patient's head and neck, special caution should be taken to prevent the
occurrence of fire in the nomoline or the surrounding environment.
The sidestream CO2 sensor port should be ventilated to the outside. If the discharge port of
the sidestream sensor is blocked, the measurement may be significantly delayed while such a
problem is not displayed.
Place the nomoline airway carefully, so as not to choke the patient’s neck or body.
132
CU-HD1 Instructions for use
1) Physiological Alarms
2) Technical Alarms
Motor Speed Turquoise alarm message The CO2 sensor motor exceeds the
Low
out of bounds with alert sound speed range.
Factory
Turquoise alarm message There is an error with the CO2
calibration Low
with alert sound sensor calibration value.
data loss
133
CU-HD1 Instructions for use
2) EtCO2
You can set the alarm range of EtCO2 out of the measured values.
134
CU-HD1 Instructions for use
3) Respiration Rate
You can set the alarm range of respiration rate out of the measured values.
4) Apnea
You can set the alarm range of apnea out of the measured values.
Changes made to the alarm settings in the menu are not automatically saved. When the
equipment reboots, the default alarm settings are restored. To change the default alarm
settings, they must be changed and saved in Administrator Mode. For details on entering
Administrator Mode or changing default alarm settings, see the Service Manual.
135
CU-HD1 Instructions for use
To change the menu composition, use the menu knob or the soft button at the bottom of the LCD
screen.
The MENU key is activated in the condition where there is no connection to a patient with pads or
paddles in other modes than the Patient monitoring mode (Monitor mode).
136
CU-HD1 Instructions for use
To change patient information, turn your Menu knob left or right to select the patient information
item. If you press the Menu knob, a text input window will be prompted at the bottom of the menu.
Also, you can change the information about the number or gender information by switching between
Increase and Decrease, On and Off using the Menu knob.
After changing the sub items of Patient Information, press the Menu knob to apply the changed data.
Select “Exit” from each of the Patient Information input menu to return to the “Main Menu”.
137
CU-HD1 Instructions for use
Make sure to enter a patient category before measuring noninvasive blood pressure.
138
CU-HD1 Instructions for use
8.2 Alarm
Press the Menu knob from the “Alarm” item at the “Main Menu”, and then the following MENUs will
be prompted so that you can change the alarm setting. For the setting of the alarm menu, you can
change the on/off status of alarm, and the maximum/minimum value of the alarming frequency.
After the detailed items have been changed, the changed values are applied by pressing the Menu
knob.
Select “Exit” from each of the alarm menus, and press the Menu knob, the “Main Menu” will be
prompted.
Changes made to the alarm settings in the menu are not automatically saved. When the
equipment reboots, the default alarm settings are restored. To change the default alarm
settings, they must be changed and saved in Administrator Mode. For details on entering
Administrator Mode or changing default alarm settings, see the Service Manual.
Changes made to the filter settings in the menu are not automatically saved. To change the
default settings, see the Service Manual.
139
CU-HD1 Instructions for use
8.3 Printer
Press the Menu knob from “Print” at the “Main Menu” to view the following menu display where you
can change the printer settings.
After the detailed items have been changed, the changed values are applied by pressing the Menu
knob. Select “Exit” from the “Printer” menu, and press the Menu knob, the “Main Menu” will be
prompted.
Auto Mode
This is a function to set it to the automatic printing mode, printing the defibrillation process
after permitting defibrillation shock. The Menu knob may be used to set the automatic mode
setting to On or Off.
Manual Print Duration
A function to set the Manual Print Duration time, for printing after the Printer button has been
pressed. By clicking the Menu knob, the duration time can be changed by 10 seconds. The
amount of time that you can set ranges from 30 seconds to 2 minutes.
Sector
Set the number of vital sign sectors which are printed by pressing the Print button. 'Sector 1' prints Sector 1 only
and 'Sector 2' prints both Sector 1 and Sector 2.
140
CU-HD1 Instructions for use
8.4.2 Volume
This is a sub-menu to control the speaker volume of the CU-HD1.
141
CU-HD1 Instructions for use
8.4.4 Bluetooth
The Product supports wireless communication using Bluetooth. The “Bluetooth” menu is
composed of the following menus that connect Bluetooth to external devices.
From the “12-lead send duration” on the “Bluetooth” menu, you can set the 12-lead ECG time
that is transferred when the 12-lead ECG information is transferred to other devices. The time
bracket you can set ranges from 10 seconds to 2 minutes by the 10 second unit. The following
shows the menu for “12-lead send duration”.
142
CU-HD1 Instructions for use
If you select Initialize Bluetooth from the “Bluetooth” menu, the following window will be
prompted, waiting for the connection to Bluetooth communication.
If the Bluetooth communication device is not detected, press the “Cancel” button. The window will
be prompted, saying that Bluetooth is not connected yet.
Through the following window, you can verify that there is Bluetooth communication established
between the CU-HD1 and the computer. For information on computer and communication, please
refer to “Chapter 9_Communication and Data Management”.
CO2 connection can be reset using a product with the EtCO2 function. For details on
connecting to a product with the EtCO2 function, see “7.4.1.1 Connecting to the Device”.
8.5 Etc.
The following screen displays the “Etc.” sub-menu which is comprised of self-test, data management,
and CO2.
143
CU-HD1 Instructions for use
After the device has been inspected, it is impossible to use the function of device treatment and
diagnosis. So, please turn off the device and turn it on again. The list of items tested in the
manual test is as follows.
Normal System Test Inspects functions necessary for the system operation.
Mode Rotary Key Test Checks out the Rotary switch functions.
Charging Button Test Checks the button status to prevent unintended recharging.
Shock Button Test Checks the Shock button status.
Audio Test Inspects whether the audio function works normally.
Defibrillation Test Inspects the charging/discharging functions.
Pacer Test Inspects the pacer function.
ECG Lead Test Inspects the ECG function.
Battery (Part A/B) Test Inspects the function of battery terminals (A/B)
SpO2 Test Inspects the function of SpO2 module.
Printer Test Inspects the printer working status.
※ For more detailed information on the manual diagnosis process, please refer to Section
10.1.3 – Manual Self-test.
144
CU-HD1 Instructions for use
Other than communicating with an external device, you can transfer digitalized voice and patient
measurement information externally by the use of an SD Card. Also, you can use the real-time printer
to print out ECG information and device usage history.
The CU-HD1 uses a data management function to record/manage patient vital sign data in the
internal memory. Up to 100 records of context information can be stored. When storing a single data
type, data can be stored for up to 192 hours continuously.
Context information is recorded in the following modes.
- AED Mode
- Manual Defibrillation Mode
- Patient Monitoring Mode
- Pacer Mode
※ Context information in data management is displayed as date of the context, start time, duration,
and usage mode.
General functions are unavailable while using the data management function. Entering Data
Management mode disables Normal mode. Exiting Data Management mode turns the
equipment on again.
When 100 context information records have been stored in the internal memory, the
defibrillator will store new context information by overwriting the oldest context information.
145
CU-HD1 Instructions for use
※ In data management, Event Review, ECG Review, Copy Data, or Data Erasing can be used for
context information.
Event Review shows the start date and time of the event and details of the event. ECG Review shows
ECG information of up to 16 seconds per page on the screen. Partial and Total Print are also
available in Event Review and ECG Review.
Copy Data allows you to select and copy context information stored in the internal memory it to
external memory (SD card).
Data Erasing erases all context information stored in the internal memory.
.
Do not insert or remove the external memory (SD card) while the defibrillator is in use. Doing
so may cause the defibrillator to malfunction.
146
CU-HD1 Instructions for use
147
CU-HD1 Instructions for use
If there is not enough printer paper before the printing job is finished, the remaining
information will be printed out by feeding paper into the printer without turning off the device.
For the information on how to feed printing paper into the printer, please refer to “Chapter
3_Product Installation”.
When the noninvasive blood pressure measuring process is complete, the measured values are
printed.
While printing, EtCO2 pauses and the "Will reconnect after printing" message appears.
When using the saving function, install an SD card before turning on the product. When using
a new SD card, connect the SD card to a PC and format it in FAT32 before inserting it in the
defibrillator.
If there are 10 MB or less of free space on the SD card, replace it or delete unnecessary data
before use. Otherwise, further data storage may not be possible or the existing data may be
damaged.
Do not remove the SD card from the product during the usage. It may cause a malfunction.
When verifying the saved information, close the current process and remove the SD card.
148
CU-HD1 Instructions for use
③ When executing “My Bluetooth Places”, the following screen will be prompted. Double
click “Search for devices in range” on the following screen.
If you purchase your PC and the CU-HD1 together, Bluetooth has should already have been
completed. If you need to initialize the Bluetooth connection due to a PC problem, you must
disconnect Bluetooth first, and then initialize the Bluetooth connection.
149
CU-HD1 Instructions for use
④ If the Bluetooth device is detected, the following screen will be prompted, and the CU-
HD1 can be identified with its serial number.
⑤ After clicking the detected Bluetooth device, right-click the device connection. And then
the following window will be prompted. At this time, type the security code.
※ For more information on how to type security code, please contact us.
⑥ Click the “OK” button after you have accurately entered the secure code. If the security
code has been typed correctly, a check mark will be displayed to the Bluetooth device
icon as shown below.
150
CU-HD1 Instructions for use
⑦ Double click the detected Bluetooth device and check the port connected to the device.
151
CU-HD1 Instructions for use
③ The Bluetooth device is searched. At this time, the name of the searched Bluetooth device
should be matched to the product serial number..
152
CU-HD1 Instructions for use
If the connection for transferring patient information is made between the computer and the
Bluetooth device, it should be possible to verify the above 12-lead measurement screen. If
Bluetooth is connected, the Bluetooth icon at the bottom of the screen appears blue, and the
“Send” soft button should be activated. If Bluetooth is disconnected, the icon should appear red,
and the “Send” soft button should disappear.
③ Press the “Send” soft button and transfer patient information via transfer program.
be transferred to the computer via the Bluetooth connection. (10 seconds later after
Bluetooth is connected, the “Real-time transfer” soft key will be created.)
153
CU-HD1 Instructions for use
③ On your smart phone, go to Setting > Bluetooth and scan for available devices.
④ Add the scanned device to your smart phone. (The PIN is the same as the device serial
number.)
⑤ Launch the 'EMS12 Agent ' app, select Settings > Scan Devices, then add the registered
device to the app.
⑥ Tap the ‘Connect CU’ button to connect to the equipment over Bluetooth.
⑦ When the Bluetooth connection is established, the ‘Bluetooth connected’ message appears
on the equipment message window.
154
CU-HD1 Instructions for use
⑥ Press the ‘OK’ button with the Menu knob to exit normal mode.
⑦ Rotate the Menu knob left/right to select a context information record and press the Menu
knob.
⑧ Select ‘Event Review’ and press the Menu knob.
② At the message window confirming total printing, press the ‘OK’ button to exectue total
printing.
155
CU-HD1 Instructions for use
⑦ Rotate the Menu knob left/right to select a context information record and press the Menu
knob.
⑧ Select ‘ECG Review’ and press the Menu knob.
printing.
156
CU-HD1 Instructions for use
② Rotate the rotary switch to Monitor, Pacer, Manual or Automatic Defibrillation Mode. The
pads must be OFF.
③ Press the Menu knob to select ‘MENU’.
④ Use the Menu knob to select ‘Device Management’ and press the Menu knob.
⑤ Select ‘Etc.’ and press the Menu knob.
⑥ Select ‘Data Management’ and press the Menu knob.
⑦ Press the ‘OK’ button with the Menu knob to exit the normal mode.
⑧ Rotate the Menu knob left/right to select a context information record and press the Menu
knob.
⑨ Select ‘Copy Data’ and press the Menu knob.
⑩ At the message confirming data copying, select the ‘OK’ button.
⑪ The selected context information is automatically copied to the external memory.
We advise regular back up of the context information in the internal memory and erasure of
the data for more systematic management.
Before erasing data, copy all context information in the internal memory to the external
memory (SD card).
Data erasing deletes all context information from the internal memory.
157
CU-HD1 Instructions for use
◎ Chapter 10_Maintenance
Familiarize yourself with the product functions and management method introduced in this chapter
in order to maintain the product at the condition for optimal and immediate use.
You can check all of the main systems that are needed to properly run the product through the
automatic or manual self-test.
10.1 Self-test
The product is a medical device. Therefore, it performs a self-test in order to make sure that all of its
functions work normally. If an error lamp is ON, immediately stop using the device and contact CU
Medical Systems, Inc. or an authorized sales agent.
158
CU-HD1 Instructions for use
◎ Chapter 10_Maintenance
1) Daily Self-test
This test is done on a daily basis and the following items are checked.
This test is the same as the battery capacity test done during the “power on” self-test.
Checks the Charge button status to prevent unintended recharging.
Checks the Shock button status of the device.
2) Weekly Self-test
This test is done on the weekly basis and the following items are checked.
Checks the daily self-test items.
Verifies the functionality of the SpO2 Pulse Oximeter.
Verifies the functionality of the internal ECG circuit.
3) Monthly Self-test
This test is done on a monthly basis and the following items are checked.
Weekly self-test items are tested.
Charging and Discharging test: Tests whether the product can replenish itself with 2J energy, and checks the
functionality of charging and discharging through internal discharging.
The manual mode test evaluates all functions tested in all the self-test modes. To initiate a
manual self-test, use the “Self Test” in the “Device Management” menu. During the self-test
process, there is a process for the user to make sure whether the device functions properly.
159
CU-HD1 Instructions for use
◎ Chapter 10_Maintenance
The Product cannot be used during the manual self-test session. The following section will give
you the step-by-step testing guidelines.
① If the mode is switched to the “Self Test” mode, product information (model name, serial
number), software version information, and the date when the self-test had been done at the
last time will be displayed on screen.
② Normal System Test : The test is run on the general system.
③ Mode Rotary Key Test : Checks the Rotary switch. If the instruction window is prompted,
move the Rotary switch to 150J of the manual defibrillator mode.
④ Charging Button Test : Process to verify the functionality of the Charge button. Press the
Charge button accoridng to the instructions.
⑤ Shock Button Test : Process to verify the functionality of the Shock button. Press the Shock
button according to the instructions.
⑥ Audio Test : Checks the speaker functions. After the voice test, there is a process to check
whether to operate the voice function by clicking the Menu knob.
⑦ Defibrillation Test : Items to check the defibrillation functionality, including the test of
automatically charging 200J and delivering the defibrillation energy. To test 200J
defibrillation operation, you must press the Shock button, and use the checkup resistance to
implement the test. If there is no resistance or simulator connection to deliver the
defibrillation energy, the device discharges the internally charged energy, leading to failure.
⑧ Pacer Test : Test about the pacer function.
⑨ ECG Lead Test : Tests 12-lead ECG measurement.
⑩ Battery Part A Test : Inspects the functionality of power terminal ‘A’.
⑪ Battery Part B Test : Inspects the functionality of power terminal ‘B’.
⑫ SpO2 Test : Inspects the functionality of SpO2 measurement module.
⑬ Printer Test : Inspects the functionality of the printer.
If the above manual test function has been finished, a message window will be prompted, saying
that the device must be closed.
In the course of the self-test process, if there is any problem with the test results, or the self-test
is put into pause during the test, or the test result indicates failure because the verification
request had not been properly performed, an additional self-test can be provided according to
the result of the power on the self-test.
160
CU-HD1 Instructions for use
◎ Chapter 10_Maintenance
The CU-HD1 is an emergency medical product that must be usable even in a situation without AC
power. So it is important to check the battery status when preparing for an emergency. For the
charged battery capacity, check the remaining battery gauge or remaining battery status displayed
on the LCD screen.
The AC power module and battery must be mounted at the same time. In case where AC power
is supplied to the AC power module, the Product can be operated by the use of power supplied
by an AC power module, and the mounted battery will proceed into its charging process. It is
recommended to use the car cigar jack only for the purpose of battery charging.
If the battery is completely discharged, it takes less than 5 hours to completely charge the battery.
Battery life depends on the frequency and duration of use. When properly cared for and used in
its intended environment, the battery module has a useful life of approximately 2 years. Use
outside those conditions could significantly reduce battery life.
For details of information on battery module and power, please refer to “Chapter 13_Product
Specifications”.
During storage, regularly ensure the battery is adequately charged. The battery is a
consumable part which requires regular inspection and replacement.
161
CU-HD1 Instructions for use
◎ Chapter 10_Maintenance
10.3 Cleaning
For any device problems resulting from the negligence of the below cleaning instructions, the free
repair service may not be applied even during the warranty period.
Keep the main body and cables away from dust and pollution, and in the normal time, clean it using
a soft cloth.
There is a risk of device damage from forcible pressing or shock.
162
CU-HD1 Instructions for use
◎ Chapter 10_Maintenance
Detailed technical information required for service support and servicing by certified personnel
is provided in the Service Manual.
163
CU-HD1 Instructions for use
◎ Chapter 10_Maintenance
CU-HD1
Date
Schedule
Exterior Condition
Accessory Maintenance
Status
Error Correction or
Measures
Inspected by Signature
of Operator
The used supplies must be discarded according to the related law applicable to the local region.
When discarding the used defibrillation pads, special caution must be paid not to cause
environmental pollution. As for the battery replacement, please consult the manufacturer or the
authorized sales agent. If there is a need for discarding the battery, it must be discarded according to
the related law.
164
CU-HD1 Instructions for use
A case that could result in a dangerous situation, including death or severe injury if instructions
are not observed
Instruction that directly or indirectly addresses the company policy in order to protect people
or property
Explanations for referential terms or additional operation tips helping you properly use the
product
165
CU-HD1 Instructions for use
166
CU-HD1 Instructions for use
Operation of the device below specified amplitude or value of vital sign may cause
inaccurate results.
While performing maintenance, calibration, etc., without tools, once the cover,
connectors, etc. are removed, do not touch the patient at the same time as
touching any of the non-electric parts of the medical device which can be touched
in the patient environment.
For the patient’s safety, do not place the device in a location where a risk factor
may occur, such as a location where the device may fall down on top of the
patient.
If the floor surface and your hands are wet, you may get shocked. Move to a dry
location first and install the product.
If you discover any abnormality while using the product, turn off the power and
refer to the user manual.
Do not use the ECG cables or connectors accompanied with the product with
other vender’s ECG monitoring devices.
167
CU-HD1 Instructions for use
With the ECG Monitoring mode, the patient’s heart rate is measured and displayed
on the LCD screen. It neither analyzes the ECG nor delivers defibrillating shocks.
This product can be used in a place with high-frequency surgical equipment
without an additional protective device. The accuracy of this product may be
degraded temporarily during electric surgery or defibrillation but this does not
affect the safety of the product or a patient. See the electrosurgery device’s
Instructions for Use for information on reducing hazards of burns in the event of a
defect in its equipment. This product must not be exposed to X-ray or strong
magnetic fields (MRI).
168
CU-HD1 Instructions for use
If you do CPR during the analysis of a patient’s ECG signal analysis, you may
perform an incorrect analysis due to the interference of the patient’s ECG signal
analysis.
When attaching the defibrillation pads to the patient’s skin, following the
instructions described in the rear side of the pad. Do not use the damaged pad.
The defibrillation pad is disposable, do not reuse.
After using the conductive gel, remove the remaining gel on the paddles
completely using a wet towel or gauze.
In the course of CPR, the device provides the beep sound based on the 5 time
cycle according to the 2005 CPR Guideline Instructions (30:2, 15:2).
169
CU-HD1 Instructions for use
170
CU-HD1 Instructions for use
Make sure that the connection tube of the cuff is not twisted or folded.
171
CU-HD1 Instructions for use
172
CU-HD1 Instructions for use
◎ Chapter 12_Troubleshooting
When the device malfunctions, the corresponding text message(s) shall be issued to inform the
device state. If the device still malfunctions after the associated troubleshooting described in this
chapter has been carried out, contact an authorized representative.
Any repairs on the product must be carried out by trained service personnel. When a problem
that may not be resolved takes place, do not disassemble the product arbitrarily. Otherwise, it
could result in an injury.
173
CU-HD1 Instructions for use
◎ Chapter 12_Troubleshooting
the charged Auto internal button has not been pressed within 15 seconds.
energy at the discharging - Recharge the energy by pressing the Charge button if
manual the charged energy discharged.
defibrillator - Use the conductive gel if the contact between the
mode paddles and the patient skin is not good.
Bad paddle contact
- Check if there is any impurity on the conductive plates
of the paddles that has contact with the patient.
"Check the
rotary switch" Rotary switch is not in
- Adjust the rotary switch to the correct position.
message with the correct position.
alert sound
174
CU-HD1 Instructions for use
◎ Chapter 12_Troubleshooting
175
CU-HD1 Instructions for use
◎ Chapter 12_Troubleshooting
176
CU-HD1 Instructions for use
◎ Chapter 12_Troubleshooting
177
CU-HD1 Instructions for use
◎ Chapter 12_Troubleshooting
178
CU-HD1 Instructions for use
◎ Chapter 12_Troubleshooting
179
CU-HD1 Instructions for use
◎ Chapter 12_Troubleshooting
180
CU-HD1 Instructions for use
◎ Chapter 12_Troubleshooting
181
CU-HD1 Instructions for use
◎ Chapter 12_Troubleshooting
182
CU-HD1 Instructions for use
CU-HD1
Dimentions (Paddle 326mm (W) x 253mm (L) x 358mm (H) (Width×Length×Height)
included)
Body: 4.7kg or below / 8.2kg or below if paddles, cables (ECG cable,
SpO2 sensor), print paper, and storage device included.
Weight
Paddle (with cables): 1.2kg or below.
Battery, AC power module: 0.5kg, 0.7kg respectively.
CU-CM1
Dimentions 128.6mm (W) x 78.7mm (L) x 32mm (H) (Width×Length×Height)
210g or below if battery included.
Weight IRMA Mainstream, ISA Sidestream analyzer: 25g, 130g or below,
respectively.
183
CU-HD1 Instructions for use
CU-HD1
Condition where both the equipment and pads must be stored together, which are
Operation immediately usable in an emergency case
Environment Temperature: 0 ~ 40 ℃
Humidity : 5% ~ 95%, Non-condensing
Condition where the equipment and pads are not stored together, with only the
Storage equipment stored or transported for a long time
Environment Temperature: -20 ~ 60 ℃
Humidity : 5%~95%, Non-condensing
Shock/ Fall/
Abuse Satisfying the condition of IEC 60601-1 Section 21
tolerance
Vibration MIL-STD-810E Method 514.4 Category 10
Package Satisfying the condition of IEC 60601-1 Section 44
ESD Satisfying the condition of IEC 61000-4-2:2001
Satisfying the condition of IEC 60601-1-2
EMI (emission)
EN55011:1998+ A1:1999 +A2:2002, Group 1, Class B
IEC 60601-1-2 limits, method EN 61000-4-3: 2001 Level 3 (10V/m 80MHz to
EMI (tolerance)
2500MHz)
Dustproof/
Waterproof IP43 according to the condition of IEC 60529
Classification
Standby Temperature: 0 ~ 43 ℃
Pads Usage
Usage Temperature: 0 ~ 40 ℃
Environment
Humidity : 5%~95%, Non-condensing
Battery Usage Condition to store or transport only the battery for a long time
and Storage Temperature: -20 ~ 45 ℃
Environment Humidity : 5%~95%, Non-condensing
184
CU-HD1 Instructions for use
CU-CM1
Operation Temperature: 0℃ ~ 40℃
Environment Humidity : 10% ~ 95%, Non-condensing
Storage Temperature: -20℃ ~ 60℃
Environment Humidity : 5% ~ 95%, Non-condensing
Altitude 0~4,572m
ESD Satisfying the condition of IEC 61000-4-2:1995+A1:: 1998+A2: 2001
Satisfying the condition of IEC 60601-1-2
EMI (emission)
EN 55011:2007+A2:2007, Group 1, Class B
Satisfying the condition of IEC 60601-1-2
EMI (tolerance)
EN 61000-4-3: 2006+A1:2008 Level 3 (10V/m 80MHz to 2500MHz)
Defibrillation-proof type BF
Operation Temperature: 0℃ ~ 40℃
Environment Humidity : 10% ~ 95%, Non-condensing
Storage Temperature: -40℃ ~ 75℃
Environment Humidity : 5% ~ 100%, Non-condensing
Altitude 0~4,572m
Dustproof/
Waterproof IP44
Classification
Defibrillation-proof type BF
Operation Temperature: 0℃ ~ 50℃
Environment Humidity : 10% ~ 95%, Non-condensing
Storage Temperature: -40℃ ~ 70℃
Environment Humidity : 5% ~ 100%, Non-condensing
Altitude 0~4,572m
Dustproof/
Waterproof IPX4
Classification
185
CU-HD1 Instructions for use
a. A statement for health Professionals From the AHA(American Heart Association) Task Force on AED,
Subcommittee on AED Safety and Efficacy. Automatic External defibrillators for Public Access
Defibrillation: Recommendations for Specifying and Reporting Arrhythmia Analysis Algorithm
Performance, Incorporating New Waveforms, and Enhancing Safety. 발행 1977;95:1677-1682
b. According to AHA Recommendations (a) and AAMI-based DF80, SVT is clearly included in the non
shock able rhythm grade.
Output Waveform
(Manual / Automatic)
186
CU-HD1 Instructions for use
187
CU-HD1 Instructions for use
1 1 1 1 1 1 1 1 ±1 J
2 2 2 2 2 2 2 2 ±1 J
3 3 3 3 3 3 3 3 ±1 J
4 4 4 4 4 4 4 4 ±1 J
5 5 5 5 5 5 5 5 ±2 J
6 6 6 6 6 6 6 6 ±2 J
7 7 7 7 7 7 7 7 ±2 J
8 8 8 8 8 8 8 8 ±2 J
9 9 9 9 9 9 9 9 ±2 J
10 10 10 10 10 10 10 10 ±2 J
15 15 15 15 15 15 15 15 ±3 J
20 20 20 20 20 20 20 20 ±3 J
30 30 30 30 30 30 30 30 ±15 %
50 50 50 50 50 50 50 50 ±15 %
70 70 70 70 70 70 70 70 ±15 %
188
CU-HD1 Instructions for use
189
CU-HD1 Instructions for use
190
CU-HD1 Instructions for use
191
CU-HD1 Instructions for use
ECG Monitoring
Recovery time of
defibrillation-proof type 5 seconds (Recovery time after delivering the defibrillation voltage)
applied unit
Heart Rate Averaging Determined by calculation algorithm of heart rate every 2 seconds
∙ A step increase from 80bpm to 120bpm: approximately 6.52 seconds
Response Time
∙ A step decrease from 80bpm to 40bpm: approximately 18.46 seconds
Time to alarm for
Approximately 10.5 seconds, regardless of amplitude and heart rate
Tachycardia
SpO2 Monitoring
Pulse rate 20 ~ 250 bpm (± 3 bpm)
SpO2 Measuring Rage 1 ~ 100%
SpO2 Accuracy 80 ~ 100%(± 3 digit)
Perfusion 0.2%
Resolution 1%
NIBP Monitoring
Patient Category Adult, Pediatric, Neonate
Measuring Method Oscillometric
Operation Mode Manual/ Auto mode
Time Interval for Auto 1, 3, 5, 10, 15, 30, 60, 120 munutes
Mode
Display Systolic / Diastolic / Mean blood pressure, Alarm setting
Error Range for Pressure ± 3 mmHg
192
CU-HD1 Instructions for use
NIBP Monitoring
Systolic
- Adult: 40 ~ 260 mmHg
- Pediatric: 40 ~ 160 mmHg
- Neonate 20 ~ 130 mmHg
Measuring Range
Diastolic
- Adult: 20 ~ 200mmHg
- Pediatric: 20 ~ 120 mmHg
- Neonate: 20 ~ 100 mmHg
Adult: 300mmHg
Overpressure Limit Pediatric: 300mmHg
Neonate: 150 mmHg
Adult: 23~33cm
Cuff Type Pediatric: 12~19cm
Neonate: 8~13cm
Material: Polyurethane
Connection Tube
Length: approximately 3m
EtCO2 Monitoring
The result of measuring CO2 acquired from the IRMA Mainstream analyzer
Capnography Input
or ISA Sidestream analyzer is displayed on the CH-HD1 LCD.
Display Range 0 ~ 99 mmHg (0~14 %)
0~99 mmHg: ± (1.5 mmHg + 2% of reading)
EtCO2Accuracy
0~14 vol%: ± (0.2 vol% + 2% of reading)
6.25 mm/sec (Default setting)
Sweep Speed 12.5 mm/sec
25 mm/sec
0~100 mmHg or 0~14 % (Default setting)
Scale 0~50 mmHg or 0~7 %
0~20 mmHg or 0~4 %
Display Unit mmHg or %
Minimum: 10 ~ 94 mmHg
EtCO2 Alarm Setting
Maximum: 11 ~ 95 mmHg
Respiration Rate
± 1 bpm
Accuracy
Respiration Rate Alarm Minimum: 1 ~ 149 bpm
Setting Maximum: 2 ~ 150 bpm
193
CU-HD1 Instructions for use
Accuracy (22 ±℃, 1013 ± 40hPa) 0~14 vol%: ± (0.2 vol% + 2% of reading)
Resolution 1 mmHg
Isolation of Patient
Defibrillation-proof type BF
(Check Defibrillation)
Recovery time of defibrillation-proof 5 seconds (Recovery time after delivering the defibrillation
type applied unit: voltage)
Note 1. The accuracy specification is valid for the operating temperature and humidity conditions
specified, except for interference specified in the table “EtCO2 Measuring – Interfering gas
and vapor effects”.
Isolation of Patient
Defibrillation-proof type BF
(Check Defibrillation)
Recovery time of defibrillation-proof 5 seconds (Recovery time after delivering the defibrillation
type applied unit: voltage)
Note 1. The accuracy specification is valid for the operating temperature and humidity conditions
specified, except for interference specified in the table “EtCO2 Measuring – Interfering gas
and vapor effects”.
194
CU-HD1 Instructions for use
13.10 Display
Display
Type TFT LCD (including backlights)
Screen Size 152.4(W) X 91.44(H) mm
Resolution 800 X 480 X 3(RGB) pixels
Dot Pitch 0.0635(W) X 0.1905(H) mm
Backlight LED Life Time 20,000hours (time when brightness is reduced to 50%)
ECG Viewing Time 6 seconds
13.11 Event Storage
Event Storage
External Storage Store up to 100 events and ECG data.
SD Card (if 1GB) Store more than 192 hours of single event or ECG data.
Or, store more than 8 hours of events, ECG data, and voice data
ECG Data Print Output ECG directly from the CU-HD1.
Printer Paper
Type Roll type
Width: 58mm
Size
Roll size: Maximum diameter 40mm
195
CU-HD1 Instructions for use
13.13 Bluetooth
Bluetooth
Parani-ESD210
Applied Module
(Bluetooth – Serial Module)
Version Bluetooth v 1.2
Frequency Range 2.402 GHz ~ 2.480GHz
Send Output Max. +4 dBm
Receive Sensitivity -80 dBm(0.1%BER)
Antenna Standard antenna and dipole antenna
Communication Distance Within 30m (based on open space)
Operation Temperature -10°C ~ 55°C (Humidity: 90%, Non-condensing)
Storage Temperature -20°C ~ 70°C (Humidity: 90%, Non-condensing)
Transmission Method :
FHSS(Frequency Hopping Spread Spectrum)
Miscellaneous
Modulation Method :
GFSK(Gaussian-filtered Frequency Shift Keying)
196
CU-HD1 Instructions for use
197
CU-HD1 Instructions for use
198
CU-HD1 Instructions for use
■ About Service
The products of CU MEDICAL SYSTEMS, INC. are designed and manufactured in compliance with the
Medical Devices Act and the relevant notifications of the Ministry of Food and Drug Safety, including
the Standards and Specifications of Medical Devices, General Requirements for Basic Electric and
Mechanical Safety of Medical Devices, and Medical Devices Manufacturing and Quality Control
Standards.
※ You are entitled to free servicing of the product for any problem that occurs during the warranty
period (2 years for the product and 1 year for the battery) under normal usage conditions.
(Warranty period shall be reduced by half if the device is used for commercial purposes.)
199
CU-HD1 Instructions for use
※ Even within the warranty period, the following defects are not covered by warranty.
Failures caused by operations performed against instructions in the Operation manual or other
incorrect operation.
Failures caused by repair or modification in service centers other than those designated by CU
Medical Systems, Inc.
Failure or damage caused by a fall or external shock after purchase.
Damage by natural disasters such as fire, earthquake, flood and/or lightning.
Damage caused by use or storage of the device in environments subjected to high temperature,
high humidity, chemical compounds, microorganisms, etc. which are detrimental to use of the
device.
Failure due to depletion of consumables.
Failure caused by sand and/or soil getting inside the device.
The purchase date, customer name, distributor name, batch number and other listed information
being arbitrarily changed.
No proof of purchase provided along with the device warranty.
Usage of accessories (such as adapter, battery, etc. and parts not recommended by the
manufacturer.
Other failure or damage caused by inappropriate operation.
■ Product Registration
After purchasing, please register the product on our website (www.cu911.com) to receive continued
customer support and product information.
Website
http://www.cu911.com
Online Support
Go to our website (www.cu911.com) -> Customer Service -> Contact Us
Europe
Authorized EU Representative / Medical Device Safety Service, GmbH
Schiffgraben 41, 30175 Hannover, Germany
200
CU-HD1 Instructions for use
Do not use the product near other electronic devices. If you use the product in such an
environment, check if it is working properly.
If cables or accessories that are not designated by CU Medical Systems, Inc. are used, the
product may be affected by EMC
201
CU-HD1 Instructions for use
<5 % UT <5 % UT
(>95% dip in UT): (>95% dip in UT):
for 5s for 5s
Power frequency 3 A/m 3 A/m Power frequency magnetic fields
(50/60 Hz) should be at levels characteristic of
Magnetic field a typical location in a typical
IEC 61000-4-8 commercial or hospital
environment.
NOTE. UT is the a.c. mains voltage prior to application of the test level.
202
CU-HD1 Instructions for use
The CU-HD1 is intended for use in the electromagnetic environment specified below. The customer
or the user of the CU-HD1 should assure that it is used in such an environment.
10 V/m
10 V/m where P is the maximum output power
80MHz ~ 2.5GHz rating of the transmitter in watts (W)
Radiated RF
according to the transmitter manufacturer
IEC 61000-4-3
20 V/m and d is the recommended separation
20 V/m
80MHz ~ 2.5GHz distance in meters (m).b
Field strengthsc from fixed RF transmitters,
as deter-mined by an electromagnetic site
survey, should be less than the compliance
level in each frequency ranged.d
Interference may occur in the vicinity of
equipment marked with the following
symbol:
203
CU-HD1 Instructions for use
NOTE 1. At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2. These guidelines may not apply in all situations. Electromagnetic propagation is
The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are
a 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz;
and 40.66 MHz to 40.70 MHz.
The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and
in the frequency range 80 MHz to 2.5 GHz are intended to decrease the likelihood
that mobile/portable communications equipment could cause interference if it is
b
inadvertently brought into patient areas. For this reason, an additional factor of 10/3
is used in calculating the recommended separation distance for transmitters in these
frequency ranges.
Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an
the location in which the CU-HD1 is used exceeds the applicable RF compliance level
above, the CU-HD1 should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting
or relocating the CU-HD1.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1]
d
V / m.
204
CU-HD1 Instructions for use
output bands
power of 35 12 12 23
𝑑 𝑃 𝑑 𝑃 𝑑 𝑃 𝑑 𝑃
transmitter
𝑉 1 𝑉 2 𝐸 1 𝐸 2
[W] E1 = E1 = E1 = E1 =
V1 = 3Vrms V2 = 10Vrms
10V/m 20V/m 10V/m 20V/m
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be determined using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1. At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2. The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6.765
MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz
to 40.70 MHz.
NOTE 3. An additional factor of 10/3 is used in calculating the recommended separation distance for
transmitters in the ISM frequency bands between 150 kHz and 80 MHz and in the
frequency range 80 MHz to 2.5 GHz to decrease the likelihood that mobile/portable
communications equipment could cause interference if it is inadvertently brought into
patient areas.
NOTE 4. These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and refection from structures, objects and people
205
CU-HD1 Instructions for use
Halothane 4 ㅡ 1
Enflurane
Isoflurane 5 +8% of reading 4
Sevoflurane
Desflurane 15 +12% of reading 4
206
CU-HD1 Instructions for use
Operation Manual
[Sales]
5F Cheonggye Plaza, 221, Anyangpangyo-ro, Uiwang-si,
Gyeonggi-do, Republic of Korea
Tel: +82 31 421 9700
Fax: +82 31 421 9911
Website: www.cu911.com
2460
The CU-HD1 complies with the requirements of the Medical Device Directive
93/42/EEC as amended by 2007/47/EC.
207