0% found this document useful (0 votes)
649 views15 pages

Oiml R 149-1

This document defines terminology and metrological requirements for non-invasive automated sphygmomanometers, which are used to measure blood pressure. It establishes maximum permissible errors for cuff pressure indication and blood pressure measurements under different conditions. It also outlines technical requirements for components like the cuff and bladder as well as requirements for electromagnetic compatibility, durability, safety, and metrological controls.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
649 views15 pages

Oiml R 149-1

This document defines terminology and metrological requirements for non-invasive automated sphygmomanometers, which are used to measure blood pressure. It establishes maximum permissible errors for cuff pressure indication and blood pressure measurements under different conditions. It also outlines technical requirements for components like the cuff and bladder as well as requirements for electromagnetic compatibility, durability, safety, and metrological controls.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 15

OIML R 149-1:2020 (E)

Contents
Foreword................................................................................................................................................ 4
1 Scope ............................................................................................................................................ 5
2 Terminology ................................................................................................................................ 5
2.1 auscultatory method ........................................................................................................... 5
2.2 bladder ................................................................................................................................ 5
2.3 cuff ..................................................................................................................................... 5
2.4 deflation valve .................................................................................................................... 5
2.5 diastolic blood pressure (value) .......................................................................................... 5
2.6 manometer .......................................................................................................................... 5
2.7 mean arterial blood pressure (value) .................................................................................. 6
2.8 non-invasive automated sphygmomanometer .................................................................... 6
2.9 non-invasive blood pressure measurement ......................................................................... 6
2.10 oscillometric method .......................................................................................................... 6
2.11 patient simulator ................................................................................................................. 6
2.12 pneumatic system ............................................................................................................... 6
2.13 rapid exhaust valve ............................................................................................................. 6
2.14 sleeve .................................................................................................................................. 6
2.15 systolic blood pressure (value) ........................................................................................... 6
2.16 tamper proofing .................................................................................................................. 7
2.17 Zero adjustment of a measuring system (VIM 3.11) .......................................................... 7
3 Description of the category of instrument ................................................................................ 7
4 Units of measurement ................................................................................................................. 7
5 Metrological requirements ......................................................................................................... 7
5.1 Maximum permissible errors of the cuff pressure indication under ambient conditions ... 7
5.2 Maximum permissible errors of the blood pressure measurement as determined by
clinical investigation........................................................................................................... 7
5.3 Maximum permissible errors of the cuff pressure indication under storage conditions ..... 8
5.4 Blood pressure measurement range .................................................................................... 8
5.5 Repeatability of blood pressure indication ......................................................................... 8
6 Technical requirements .............................................................................................................. 8
6.1 General ............................................................................................................................... 8
6.2 Technical requirements for the cuff and bladder ................................................................ 8
6.3 Effect of voltage variations of the power source ................................................................ 8
6.4 Pneumatic system ............................................................................................................... 9
6.5 Electromagnetic compatibility.......................................................................................... 10
6.6 Durability.......................................................................................................................... 10
6.7 Technical requirements for the pressure indicating device .............................................. 10
6.8 Signal input and output ports ............................................................................................ 11
6.9 Safety requirements .......................................................................................................... 11
6.10 Resistance to vibration and shock .................................................................................... 12
6.11 Durability of markings ..................................................................................................... 12
7 Metrological controls ................................................................................................................ 12
7.1 Type approval ................................................................................................................... 12
7.2 Verification....................................................................................................................... 12
7.3 Sealing .............................................................................................................................. 13
7.4 Marking of the device....................................................................................................... 13
7.5 .............................................................................................. 14

3
OIML R 149-1:2020 (E)

Foreword
The International Organisation of Legal Metrology (OIML) is a worldwide, intergovernmental
organisation whose primary aim is to harmonise the regulations and metrological controls applied by
the national metrological services, or related organisations, of its Member States.
The main categories of OIML publications are:
International Recommendations (OIML R), which are model regulations that establish the
metrological characteristics required of certain measuring instruments and which specify
methods and equipment for checking their conformity. OIML Member States shall implement
these Recommendations to the greatest possible extent;
International Documents (OIML D), which are informative in nature and which are intended
to harmonise and improve work in the field of legal metrology;
International Guides (OIML G), which are also informative in nature and which are intended
to give guidelines for the application of certain requirements to legal metrology; and
International Basic Publications (OIML B), which define the operating rules of the various
OIML structures and systems.
OIML Draft Recommendations, Documents and Guides are developed by Project Groups linked to
Technical Committees or Subcommittees which comprise representatives from the Member States.
Certain international and regional institutions also participate on a consultation basis. Cooperative
agreements have been established between the OIML and certain institutions, such as ISO and the IEC,
with the objective of avoiding contradictory requirements. Consequently, manufacturers and users of
measuring instruments, test laboratories, etc. may simultaneously apply OIML publications and those
of other institutions.
International Recommendations, Documents, Guides and Basic Publications are published in English
(E) and translated into French (F) and are subject to periodic revision.
Additionally, the OIML participates in Joint Committees with other Institutions for the development of
Vocabularies (OIML V) and Joint Guides (G) and periodically commissions legal metrology experts
to write Expert Reports (OIML E). Expert Reports are intended to provide information and advice,
and are written solely from the viewpoint of their author, without the involvement of a Technical
Committee or Subcommittee, nor that of the CIML. Thus, they do not necessarily represent the views
of the OIML.
This publication - reference OIML R 149-1, edition 2020 (E) - was developed by the OIML Technical
Subcommittee TC 18/SC 1 Blood pressure instruments. It was approved for final publication by the
International Committee of Legal Metrology in 2020 and supersedes OIML R 16-2:2002 (E).
OIML Publications may be downloaded from the OIML website in the form of PDF files. Additional

11, rue Turgot - 75009 Paris France


Telephone: 33 (0)1 48 78 12 82
Fax: 33 (0)1 42 82 17 27
E-mail: [email protected]
Internet: www.oiml.org

4
OIML R 149-1:2020 (E)

Non-invasive automated sphygmomanometers


Part 1: Metrological and technical requirements

1 Scope
This Recommendation specifies general, performance, efficiency and mechanical safety requirements
for non-invasive automated sphygmomanometers and their accessories which, by means of an inflatable
cuff, are used for the non-invasive measurement of arterial blood pressure.
This Recommendation only applies to devices measuring at the arm, the wrist or the thigh.

2 Terminology

2.1 auscultatory method


method whereby sounds (known as Korotkoff sounds) are heard or detected (e.g. by a microphone) over
an occluded artery as the occluding pressure is slowly released, the appearance of sounds coinciding
with the systolic blood pressure and the disappearance of sounds with the diastolic blood pressure

2.2 bladder
inflatable component of the cuff

2.3 cuff
component of the non-invasive automated sphygmomanometer, comprising a bladder and a sleeve,
which is wrapped around the limb of the patient
Note: A cuff might comprise a bladder and an inelastic part that encloses the bladder, or have an
integral bladder (i.e. the cuff including the bladder are fixed together or are one piece).

2.4 deflation valve


valve for controlled exhaust of the pneumatic system during measurement

2.5 diastolic blood pressure (value)


minimum value of the arterial blood pressure as a result of relaxation of the systemic ventricle
Note: Because of hydrostatic effects, this value should be measured with the cuff at the heart level.

2.6 manometer
instrument used to measure and display pressure

5
OIML R 149-1:2020 (E)

2.7 mean arterial blood pressure (value)


value of the integral of one cycle of the blood pressure curve divided by the time of one heart beat
period
Note: Because of hydrostatic effects, this value should be measured with the cuff at the heart level.
The calculation of the mean arterial blood pressure using only the systolic and diastolic blood
pressure values is not recommended.

2.8 non-invasive automated sphygmomanometer


medical measuring instrument used for the intermittent non-invasive estimation of the blood pressure
by utilising an inflatable cuff, a pressure transducer, a valve for deflation, and/or displays used in
conjunction with automated methods for estimating blood pressure. Hereafter referred to as

2.9 non-invasive blood pressure measurement


indirect measurement of the arterial blood pressure without arterial puncture

2.10 oscillometric method


method that estimates systolic, diastolic and mean arterial pressures during the slow inflation or
deflation of an occluding cuff at the brachial artery
Note: During the inflation and deflation of the cuff, small pressure changes (oscillations) occur in
the cuff as a result of the arterial blood pressure pulses. These oscillations are detected and
stored together with the corresponding cuff pressure values in the measurement system. With
these stored values the systolic, diastolic and mean arterial blood pressure values can be
mathematically derived using an appropriate algorithm.

2.11 patient simulator


device for simulating the oscillometric cuff pulses and/or auscultatory sounds during inflation and
deflation
Note: This device is not used for testing measurement accuracy but is required in assessing stability
of performance.

2.12 pneumatic system


system that includes all pressurised and pressure-controlling parts such as cuff, tubing, connectors,
valves, transducer and pump

2.13 rapid exhaust valve


valve for rapidly exhausting the pneumatic system

2.14 sleeve
essentially inelastic part of the cuff that encloses the bladder

2.15 systolic blood pressure (value)


maximum value of the arterial blood pressure as a result of the contraction of the systemic ventricle
Note: Because of hydrostatic effects, this value should be measured with the cuff at the heart level.

6
OIML R 149-1:2020 (E)

2.16 tamper proofing


means of preventing the user from gaining easy access to the measuring mechanism of the device

2.17 zero adjustment of a measuring system (VIM 3.11)


procedure that corrects a deviation of the pressure reading to 0.0 kPa (0 mmHg) at atmospheric pressure
(gauge pressure: 0 kPa (0 mmHg))

3 Description of the category of instrument

a system for applying and releasing pressure to the bladder in the cuff, and a means of measuring and
displaying blood pressure values automatically.
Note 1: Specific device types included in this category are: sphygmomanometers for self-
measurement, blood pressure monitors and multi-parameter patient monitors used for home
healthcare, or public use.
Note 2: Components of a sphygmomanometer include: manometer, cuff, valve for deflation (often in
combination with the valve for rapidly exhausting the pneumatic system), pump for inflation
of the bladder, and connection tubing.

4 Units of measurement
The blood pressure shall be indicated either in kilopascals (kPa) or in millimetres of mercury (mmHg).

5 Metrological requirements

5.1 Maximum permissible errors of the cuff pressure indication under ambient
conditions
For any set of conditions within the ambient temperature range from 10 C to 40 C and the relative
humidity range from 15 % to 85 %, both for increasing and for decreasing pressure, the maximum
permissible error for the measurement of the cuff pressure at any point of the measurement range shall
be 0.4 kPa ( 3 mmHg) or % of the reading, whichever is greater.

5.2 Maximum permissible errors of the blood pressure measurement as determined by


clinical investigation 1
The following maximum permissible errors shall apply for the sphygmomanometer:
mmHg);
maximum experimental standard deviation: 1.1 kPa (8 mmHg).
For further recommended test methods see Annex A.

1
carried out by the manufacturer

7
OIML R 149-1:2020 (E)

5.3 Maximum permissible errors of the cuff pressure indication under storage
conditions
The sphygmomanometer shall maintain the requirements specified in this Recommendation after
storage for 24 h at a low temperature of -5 h at a high
temperature of 50 f 85 % (non-condensing).

5.4 Blood pressure measurement range


The sphygmomanometer shall be capable of indicating diastolic blood pressure over at least the range
from 2.7 kPa (20 mmHg) to 8.0 kPa (60 mmHg) in neonatal mode, and 5.3 kPa (40 mmHg) to 17.3 kPa
(130 mmHg) otherwise.
The sphygmomanometer shall be capable of indicating systolic blood pressure over at least the range
from 5.3 kPa (40 mmHg) to 14.7 kPa (110 mmHg) in neonatal mode, and 8.0 kPa (60 mmHg) to
30.7 kPa (230 mmHg) otherwise.

5.5 Repeatability of blood pressure indication


For any set of conditions within the ambient temperature range from 10 C to 40 C and the relative
humidity in the range from 15 % to 85 %, the experimental standard deviation of the blood pressure
indication of the sphygmomanometer shall not exceed 0.4 kPa (3 mmHg).

6 Technical requirements

6.1 General
Equipment, or parts thereof, using materials or designs different from those detailed in this
Recommendation shall be accepted if it can be demonstrated that an equivalent degree of safety and
performance is obtained.

6.2 Technical requirements for the cuff and bladder


The cuff shall contain or incorporate a bladder. The cuff shall be designed and marked (i.e. using
permitted circumference indicators) to ensure and restrict the use of the appropriate cuff size
corresponding to a given limb circumference.
The bladder length should be approximately 0

limb at the midpoint of the intended range of the cuff.


For reusable cuffs the manufacturer shall indicate the method for cleaning in the accompanying
documents.

6.3 Effect of voltage variations of the power source

6.3.1 Internal electrical power source


Changes in the voltage within the working range specified by the manufacturer (see 7.5) shall not
influence the cuff pressure indication.
Outside this working range no cuff pressure indication and no result of the blood pressure measurement
shall be displayed.

8
OIML R 149-1:2020 (E)

6.3.2 External electrical power source


Changes in the voltage within the working range specified by the manufacturer (see 7.5) shall not
influence the cuff pressure indication.
Outside the working range specified by the manufacturer, no cuff pressure indication and no result of
the blood pressure measurement shall be displayed.

6.4 Pneumatic system

6.4.1 Air leakage


Air leakage shall not exceed a pressure drop of 0.8 kPa/min (6 mmHg/min).

6.4.2 Pressure reduction rate of devices using the auscultatory method


The pressure reducing system for manually operated and automated deflation valves shall be capable
of maintaining a deflation rate of 0.3 kPa/s to 0.4 kPa/s (2 mmHg/s to 3 mmHg/s) within the target
range of systolic and diastolic blood pressure. For devices which control the pressure reduction as a
function of the pulse rate, a deflation rate of 0.3 kPa/pulse to 0.4 kPa/pulse (2 mmHg/pulse to
3 mmHg/pulse) shall be maintained.
Note: Manually operated deflation valves should be easily adjustable to these values.

6.4.3 Rapid exhaust


During the rapid exhaust of the pneumatic system, with the valve fully opened, the time for the pressure
reduction from 34.7 kPa to 2.0 kPa (260 mmHg to 15 mmHg) shall not exceed 10 s.
For the sphygmomanometer having the capability to measure in a neonatal/infant mode, the time for
the pressure reduction from 20.0 kPa to 0.7 kPa (150 mmHg to 5 mmHg) during the rapid exhaust of
the pneumatic system with the valve fully opened shall not exceed 5 s.

6.4.4 Zero adjustment of a measuring system


The sphygmomanometer shall be capable of automatic zero adjustment. The zero adjustment shall be
carried out at appropriate intervals, at least when the device is powered on. After a zero adjustment, the
device shall keep the indication of a gauge pressure of 0.0 kPa (0 mmHg).
The sphygmomanometer shall repeat a zero adjustment or shall be switched off automatically when the
output of the pressure transducer drifts one scale interval (0.1 kPa or 1 mmHg) or more.

6.4.5 Manometer test mode


The sphygmomanometer shall have a manometer test mode that permits static pressure measurement
over at least the nominal blood pressure indication range. This mode shall not be available in normal
use, but restricted to service / test personnel.
When the sphygmomanometer is put into the test mode, all air outlets shall be closed.
The manufacturer shall confirm that the test results obtained in 5.1 and 6.4.4 are identical to the results
in the normal use mode.

9
OIML R 149-1:2020 (E)

6.4.6 Maximum time for which the cuff is inflated


The total time for which the pressure exceeds 2.0 kPa (15 mmHg) shall be no longer than 180 s in the
case of adult patients. The total time for which the pressure exceeds 0.7 kPa (5 mmHg) shall be no
longer than 90 s in the case of neonatal/infant patients.

6.5 Electromagnetic compatibility

6.5.1 Immunity
The following requirements apply:
electrical and/or electromagnetic interferences shall not lead to degradations in the cuff pressure
indication, i.e. the maximum permissible error for the measurement of the cuff pressure shall
be 0.4 kPa ( 3 % of the reading, whichever is greater; or
if electrical and/or electromagnetic interferences lead to an abnormality, the abnormality shall
be clearly indicated and it shall be possible to restore normal operation within 30 s after
cessation of the electromagnetic disturbance.
Testing shall be carried out in accordance with 202 of IEC 80601-2-30:2018 Medical electrical
equipment - Part 2-30: Particular requirements for the basic safety and essential performance of
automated non-invasive sphygmomanometers.

6.5.2 Electrosurgery interference recovery


If a sphygmomanometer is intended to be used together with HF surgical equipment, it shall return to
the previous operating mode within 10 s after exposure to the field produced by the HF surgical
equipment, without loss of any stored data.
Testing shall be carried out in accordance with 202.8.101 of IEC 80601-2-30:2018 Medical electrical
equipment - Part 2-30: Particular requirements for the basic safety and essential performance of
automated non-invasive sphygmomanometers.

6.6 Durability
The change in the cuff pressure indication shall not be greater than 0.4 kPa (3 mmHg) throughout the
pressure range after 10 000 simulated measurement cycles.

6.7 Technical requirements for the pressure indicating device

6.7.1 Nominal range and measurement range of the cuff pressure measurement
The nominal range for the cuff pressure measurement shall be specified by the manufacturer. The
measurement range of the cuff pressure shall be equal to the nominal range. Values of blood pressure
measurement results outside the nominal range of cuff pressure shall be clearly indicated as out of range.
Testing shall be carried out by visual inspection.

10
OIML R 149-1:2020 (E)

6.7.2 Digital indication


The digital scale interval shall be 0.1 kPa (1 mmHg).
If the measured value of a parameter is to be indicated on more than one display, all the displays shall
indicate the same numerical value.
Measured numerical values on the display(s), and the symbols defining the units of measurement shall
be arranged in such a way so as to avoid misinterpretation.
Numbers and characters should be clearly legible.
Testing shall be carried out by visual inspection.

6.7.3 Technical requirements for the display


The display shall be designed and arranged so that all information can be read and easily recognised.
If abbreviations are used on the display they shall be as follows:

Single letter abbreviations shall be positioned in such a way to avoid confusion with SI units.
Testing shall be carried out by visual inspection.

6.8 Signal input and output ports


The construction of the signal input and output ports (excluding internal interfaces, e.g. microphone
signal input) relevant to the non-invasive blood pressure measurement shall ensure that incorrectly fitted
or defective accessories shall not result in erroneous indication of cuff pressure or erroneous indication
of blood pressure.
Note: An error message or a blank display is sufficient.

6.9 Safety requirements

6.9.1 Aborting a measurement


It shall be possible to abort any blood pressure measurement at any time by single key operation and
this shall lead to a rapid exhaust (see 6.4.3).

6.9.2 Unauthorised access and tamper proofing


All controls which affect accuracy shall be sealed against unauthorised access.
Tamper proofing of the instrument shall be achieved by requiring the use of a special tool or breaking
a seal.
Testing shall be carried out by visual inspection.
It shall be clear to an operator if tampering or unauthorised access has occurred.

11
OIML R 149-1:2020 (E)

6.9.3 Tubing connectors


Luer lock and Luer slip connectors shall not be used on sphygmomanometers so as to avoid any risk of
connecting the output of the sphygmomanometer to intravascular fluid systems as air might
inadvertently be pumped into a blood vessel.

6.9.4 Electrical safety


Regional or national regulations may specify electrical safety requirements.

6.10 Resistance to vibration and shock


The sphygmomanometer or its parts not intended for use during patient transport outside a healthcare
facility shall have adequate mechanical strength when subjected to mechanical stress caused by normal
use, pushing, impact, dropping, and rough handling. A fixed (e.g. wall mounted) sphygmomanometer
is exempt from the requirements of this subclause.
After the test for the resistance to vibration and shock, the sphygmomanometer shall comply with the
requirements of 5.1 but only at a temperature of 20 C 5 C and at ambient humidity.

6.11 Durability of markings


The markings shall be removable only with a tool or by appreciable force and shall be sufficiently
durable to remain clearly legible during the expected service life of the sphygmomanometer. In
considering the durability of the markings, the effect of normal use shall be taken into account.

7 Metrological controls
Regional or national regulations may prescribe type approval, initial and/or subsequent verification for
sphygmomanometers. These controls shall meet the conditions in 7.1 7.5.

7.1 Type approval


At least three samples of a new type of sphygmomanometer shall be tested.
The tests to verify conformity to metrological and technical requirements shall be carried out according
to R 149-2. A test report shall be prepared according to R 149-3.

7.2 Verification
After type approval has been granted, verification shall be carried out before the sphygmomanometer
is put into use and during its lifetime. At verification, testing can be conducted at any set of climatic
conditions within the temperature range from 10 %
to 85 %. A climatic chamber is not required.
Each instrument of an approved type of sphygmomanometer shall be verified periodically in accordance
with applicable metrological laws and regulations of a member state, or after repair. At least the
requirements of 5.1, 5.5 and 6.4.1 shall be fulfilled.

12
OIML R 149-1:2020 (E)

7.3 Sealing
Metrological control marks shall be put on seals. These seals shall prevent, without destruction of the
control marks:
in the case of patient-monitors in which the sphygmomanometer is one part of a system: the
manipulation of the metrologically relevant parts for measuring blood pressure;
in the case of all other sphygmomanometers: the opening of the casing.
If the construction of the instrument guarantees security against any interference, the metrological
control marks or the security marks may be attached in the form of labels.
All seals shall be accessible without using a tool.

7.4 Marking of the device

7.4.1 Markings required on the indicating device


The indicating device of the sphygmomanometer shall be marked with the following information:
name and/or trademark of the manufacturer;
type of sphygmomanometer;
units of measurement (kPa/mmHg), positioned close to the displayed values;
measurement range;
type approval number (if applicable);
serial number;
year of fabrication;
country of origin.
Testing shall be carried out by visual inspection

7.4.2 Markings required on the cuff


The cuff of the sphygmomanometer shall be marked with the following information:
limb circumference for which it is appropriate;
marking of the limb circumference indication range;
centre of the bladder, indicating the correct position for the cuff over the artery.
Testing shall be carried out by visual inspection.
For sphygmomanometers applied to the wrist, the marks required in 7.4.1 and 7.4.2 can be positioned
on the indicating device or on the cuff.
For sphygmomanometers used for home healthcare environment, the sales packaging shall display
information needed by the end user including, as a minimum:
the operating and storage temperature and humidity ranges;
any special requirements for a battery-powered sphygmomanometer.

13
OIML R 149-1:2020 (E)

Sphygmomanometers for public use which are intended for self-use in public areas, shall be marked
with the following:
precautions for use, including a statement concerning the need to consult a physician for
interpretation of blood pressure measurements;
adequate operating instructions

7.5 Manufactu
Information supplied by the manufacturer shall comply with the specifications and requirements given
in this Recommendation.

reference to OIML R 149 including the complete title;


explanation of the operating procedures which are important for correct application (such as
the selection of the appropriate cuff size, positioning of the cuff at the heart level and adjustment
of the pressure reduction rate);
methods for cleaning reusable cuffs;
if the bladder is removable, the method for ensuring the correct repositioning of the bladder in
the cuff;
nature and frequency of the maintenance which is required to ensure that the device operates
correctly and safely at all times;
disclosure that applicable national or regional metrological laws and regulations have to be
considered;
list of all components belonging to the pressure measuring system, including accessories;
description of the operating principles of the blood pressure measuring device;
remarks on the environmental or operational factors which may affect the performance (e.g.
electromagnetic fields, arrhythmia);
specification of the signal input/output port(s);
specification of the rated voltage, if applicable;
specification of the intended power source, if applicable;
measurement range for the systolic and diastolic blood pressure measurements;
measurement range of the pulse rate;
operating and storage temperature and humidity ranges;
any special requirements for a battery-powered automatic sphygmomanometer, e.g. safety
warnings;
warm-up time, if applicable;
0 and 0, if applicable);
description of all symbols, abbreviations and error codes used on the instrument; and
name and address of manufacturer.
Testing shall be carried out by visual inspection.

14
OIML R 149-1:2020 (E)

Annex A
Rationale for the maximum permissible errors
of the overall system
(Informative)

Note: This Annex provides a rationale for the values of maximum permissible errors presented in
5.2.

Overall system accuracy


A clinical investigation is strongly recommended to demonstrate compliance with the requirements
specified in 5.2.
A new clinical investigation is necessary only for changes affecting the overall system accuracy.
Recommended protocols for the clinical investigations are given in:
O'Brien E., Petrie J., Littler W., de Swiet M, Padfield P.L., Altman D.G., Coats A. and Aikins
N. The British Hypertension Society protocol for the evaluation of blood measuring devices.
Journal of Hypertension 1993, 11 (Suppl 2): S 43 62
ISO 81060-2.

15

You might also like