Antibbiotics Testing by Kit Method
Antibbiotics Testing by Kit Method
Kit Information
Introduction
The Charm SL Beta-lactam Test is a receptor assay utilizing ROSA® (Rapid One Step Assay) lateral flow
technology. The SL Beta-lactam Test detects beta-lactam drugs at or below the U.S. Safe Level/Tolerance,
see Table 1.
Cross Reactivity
There are no known interferences from the following non-beta-lactam animal drugs at 100 ppb:
sulfadiazine, sulfanilamide, sulfathiazole, sulfamethazine, sulfapyridine, sulfadimethoxine, tetracycline,
oxytetracycline, chlortetracycline, doxycycline, gentamicin, neomycin, streptomycin, ivermectin,
erythromycin, pirlimycin, tilmicosin, novobiocin, furosemide, trichlormethiazide, chlorothiazide,
thiabendazole, oxytocin, phenylbutazone, dexamethasone, PABA and dipyrone.
Table 2 lists some other cross reactive beta-lactam (intra-family) drugs.
Table 2: Intra-family Cross Reactivity Levels Detected in Milk with High Confidence
Cefacetrile 30 ppb Cefoperazone 15 ppb Dicloxacillin 50 ppb
Cefadroxil 50 ppb Cefquinome 30 ppb Nafcillin 100 ppb
Cefalexin 50 ppb Cefuroxime 15 ppb Oxacillin 50 ppb
Cefalonium 4 ppb Cloxacillin 50 ppb Ticarcillin 100 ppb
Cefazolin 40 ppb
Training
Equipment setup and use is simple and can be self-taught from the manual. In addition, training
materials and/or video is available for review of procedure, equipment use, or maintenance.
Proficiency samples (blind coded positive and negative samples) are available for retraining and
procedural validation (see Additional Test Information).
For questions contact your local representative or Charm Sciences at +1.978.687.9200 or
[email protected].
Precautions:
High fat samples (greater than 6.5%) may cause invalid results.
Keep equipment clean.
Keep dust and milk off exterior of test strips.
ROSA® Incubator must be clean, and the temperature indicator green (indicating 56 1°C). Keep
ROSA Incubator lid lowered, but not latched, unless performing procedure. A daily temperature check
is required.
Invalid Inspect C (Control) line for even development before placing strip
in ROSA Pearl Reader.
C C C The test is INVALID if any of the following are observed:
T T T C (Control) line is missing
C line is smeared or uneven
Sample is obscuring either the C or T (Test) lines
DO NOT PUT INVALID STRIPS IN READER. IF INVALID THE
SAMPLE MUST BE RETESTED.
Reader Interpretation
Within 5 minutes of completing the test, insert the test strip, with valid
C line, into ROSA Pearl Reader. Slide completely into slot.
Read results on SLBL channel of ROSA Reader. SLBL channel displayed in
upper left of reader screen should blink slowly. Use the down arrow to
enter other optional information such as SAMPLE or OPERATOR ID. If strip
is not fully inserted, the reader will display an error message; re-insert
strip and read again. Press ENTER. A numerical READING and a
RESULT are displayed on screen in 8 seconds.
Reader results including lot number and time/date are stored in memory
for later review, download, or printing, refer to ROSA Reader Manual.
NCIMS policy requires a hard copy of data be printed daily.
Technical Assistance:
Call 1-800-343-2170 hours 8:30 AM - 5:30 PM EST.
For emergency 24-hour 365-day per year service call 1-978-687-9200 and follow prompts.
E-mail: [email protected] (monitored during 8:30 AM – 5:30PM EST).
1000 disposable pipet tips, 300-1000 LF-ROSA-PIPET- 5000 disposable ROSA-Pipets, 300
1-MLT-X4
µl 5000 µl
Thermal printer for data output to 4 Blind Proficiency Samples: coded,
PRN-THERMAL-SYS reader with RS-232 adapter and UNKSTD-LF freeze-dried negative and positive
cable samples, 8-10 ml rehydration
Positive
Positive - T line is clearly lighter/less than C line, or T line is absent, or
C C C partially or unevenly colored. Sample should be retested. See Retest of
T T T
Initial Positive.
Warranty Information:
Charm Sciences, Inc. (“Charm”) warrants each reagent product, including but not limited to test kits, to be free from defects in materials and workmanship and to be free from
deviations from the specifications and descriptions of Charm’s reagent products appearing in Charm’s product literature, when stored under appropriate conditions and given
normal, proper and intended usage, until the expiration of such reagent product’s stated shelf life, or, if none is stated, for one year from the date of delivery of such reagent
product to the end-user purchaser. THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, WHETHER STATUTORY, EXPRESS, IMPLIED (INCLUDING
WARRANTIES OF TITLE, NON-INFRINGEMENT, MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE AND ALL WARRANTIES ARISING FROM COURSE
OF DEALING OR USAGE OF TRADE). The warranty provided herein may not be altered except by express written agreement signed by an officer of Charm. Representations,
oral or written, which are inconsistent with this warranty are not authorized and if given, should not be relied upon. In the event of a breach of the foregoing warranty, Charm’s
sole obligation shall be to replace any reagent product or part thereof that proves defective in materials or workmanship within the warranty period, provided the customer
notifies Charm promptly of any such defect prior to the expiration of said warranty period. The exclusive remedy provided herein shall not be deemed to have failed of its
essential purpose so long as Charm is willing to replace any nonconforming reagent product or part. Charm shall not be liable for consequential, incidental, special or any
other indirect damages resulting from economic loss or property damages sustained by any customer from the use of its reagent products. Except for Charm’s
obligation set forth above to replace any reagent product that proves defective within the warranty period, Charm shall not be liable for any damages of any kind arising out of or
caused by any incorrect or erroneous test results obtained while using any such reagent product, whether or not caused by a defect in such reagent product.