This document provides instructions for operating the display of an automatic total nitrogen analyzer. It describes the main screens, menu screens, and key functions. The main screens show measurement status, results, and allow controlling the device. The menu screens allow changing settings like measurements, alarms, printing, and maintenance options.

Uploaded by

Mẫn Phan
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
40 views

TN-310 Module 4 Display Operations

This document provides instructions for operating the display of an automatic total nitrogen analyzer. It describes the main screens, menu screens, and key functions. The main screens show measurement status, results, and allow controlling the device. The menu screens allow changing settings like measurements, alarms, printing, and maintenance options.

Uploaded by

Mẫn Phan
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 45

AUTOMATIC TOTAL NITROGEN ANALYZER

TN - 310
Reference modules
Contents

Module 4 Display operation


4.1 Screen Flow…………………………………………………………..2-3
4.2 Main Screen…………………………………………………………..4-7
4.3 Menu Screen………………………………………………………….8-40
4.4 Initial Settings List……………………………………………………41-42
4.5 Operation keys List...…………………………………………………43

-1-
Module 4 Display operation Screen flow

4.1 Screen flow

MAIN STATUS Status display screen … 4・2・1


DISPLAY

Operation screen … 4・2・2


OPERATE

Menu Menu screen … 4・3

メニュー(各種設定・操作一覧画面)

SET BASIC General settings … 4・3・2

メニュー(各種設定・操作一覧画面)
DEVICE Device settings … 4・3・3 ※

CLOCK Time settings … 4・3・4

OPE. ARBITRARY Arbitrary settings … 4・3・5

FLOW Flow operations … 4・3・6

LOAD Load operations … 4・3・7

PRINT Print … 4・3・8

RECORD History … 4・3・9

MAINT PRESENT VALUE Present values … 4・3・10

IN/OUT Common input/output … 4・3・11

MALFUNCTION Malfunction … 4・3・12 ※

OTHER INFORMATION INFROMATION … 4・3・13 ※

USB

※「DEVICE」 「INFORMATION」keys lead to item selection screen.


「MALFUNCTION」

2
Screen selection is as follows.

DEVICE

DEVICE MEAS. Measurement settings … 4・3・3・1


SETTING
ALARM Alarm settings … 4・3・3・2

OPTION Option settings … 4・3・3・3

DISPLAY Display settings … 4・3・3・4

LOCATION Station name settings … 4・3・3・5

COMM. Communication settings … 4・3・3・6

TEMP. Temperature settings … 4・3・3・7

INFO.

INFORMAT PROGRAM Program … 4・3・13・1


ION
INITIALIZE Initialize … 4・3・13・2

3
Main screen (status display)

4.2 Main screen


The measurement operation control key and each screen transition key are present on the main screen.
Operating status and measurement data can be checked. Device settings can also be changed.
The keys related to the operation control of the device can be used only on a specific screen.

4.2.1 Status display screen

≪Normal display screen≫


The current process operation status and the previous measurement result are displayed.
Present time
Time remaining for the
completion of the process
Next step is displayed
"Measurement /Stop" "Measurement / Calibration / Automatic
Calibration / Cleaning" "Maintenance" are displayed
1 10
Present step name is displayed

11 58
Present step number is displayed
100.0
While operating in measurement mode, the
54.62 measurement start time, measured value, and
measured voltage value are displayed as the
previous measurement result.

By directly touching the operation keys, start / stop each


operation can be started, stopped.

【 OPERATION DETAILS】

Key name Process


[START] key Starts and stops measurement operation
[LOAD] key Turns on / off all loads (solenoid valves, pumps, etc.)
[PAUSE] key Pause at the current step
[STEP] key Fast forward steps
[OPE.] key Moves to the operation mode setting screen
[MENU] key Move to the menu screen
[ALARM RESET] key Stop alarm alerts
[PAPER FEED] key Printer paper feed

4
Main screen (status display)

≪Calibration result display screen ≫


The current process operation status and the previous calibration result are
displayed.

1 10

11 58

54.62 While operating in calibration mode,


Calibration start time, calibration value, are
100.0 displayed as calibration result.

Furthermore, measured value (concentration


value) and temperature based on the current
set value are also displayed as calibration
result.

≪Abnormal content display screen≫


When an abnormality occurs, the time of occurrence and the content of the
abnormality will pop up in red.

【 警 報 詳 細 】

【Alarm details】

Alarm display Alarm name


Shut/Sample SHUT SAMPLE
Shut/Washing water SHUT WATER
Suction fail/Sample Improper sample suction
Suction fail/Water Improper washing water suction
Suction fail/Buffer Improper suction (buffer tube)

・For alarm details, refer to "ALARM list" in the appendix.

5
Main screen (operation settings)

4.2.2 Operation screen


Displays the operation mode, measurement cycle, and print settings.
By touching the schedule key, the schedule settings will pop up.

≪Virtual keypad≫
Displayed when the background color is yellow

00 60

stop stop
FORWARD stop

【 Settings and operation details】


Process Contents
OPE. MODE Operation mode settings MEAS., CAL., WASH
START MODE Operation start settings INTERNAL, EXTERNAL, CONTINUE,
SCHEDULE
M. START 'Minute' setting to start 0~59 MINUTES NOTE1
operation
M. Cycle 'Minute' setting to start 5~999 MINUTES NOTE1
measurement cycle
PRINTER Printer operation settings RUN, STOP
BAR GRAPH Print bar graph along with NONE, YES
measured value print
PRINT DIR Print direction setting FORWARD, REVERSE
CURVE GRAPH Print the detection curve along NONE, YES
with the measurement value
print

Note1: Only when the start mode is selected as internal start, it operates according to the
settings.

6
Main screen (schedule settings)

4.2.2.1 Schedule screen

After touching the "Schedule" key on the operation screen, the schedule setting screen is displayed.
Press the "SET" key to save the settings.

0 0 0 0 0 0 0 0

0 0 0 0 0 0 0 0

0 0 0 0 0 0 0 0

【 Setting details】

Process Contents
SET (Days of Set days of SON, MON TUE, WED, THU, FRI, SAT
the week) the week
SET(TIME) Set time Enter the series number 0~1 ※
[SET]key Save the set data

※ If set to "0", measurement will be stopped.

7
Menu screen (Setting operation list)

4.3 Menu screen

The menu screen consists of the following 13 screens.


Please move to the screen of each required setting / operation from the screen of "setting operation list" as
explained below.
However, please note that conditions are set for each screen transition.
4.3.1 Menu (setting operation list)

This is a list screen for setting and operating each device setting, process / load operation, input / output
signal operation, etc. You can select and transfer by directly touching each key on the screen.

【 Operation Permissions 】
Process Contents
BASIC Measurement conditions settings Only when measurement is stopped
DEVICE Device settings 〃
CLOCK Calendar / clock correction 〃
ARBITRARY Arbitrary step operation Operation permitted at all times
FLOW Operation of specific flow Only when measurement is stopped
LOAD Individual load operation 〃
Operation permission for all (with some
PRINT Batch print operation
restrictions)
RECORD Display of each record Operation permitted at all times
PRESENT
Display of present value 〃
VALUE
IN/OUT IN/OUT operation 〃
Display of troubleshooting
MALFUNCTION 〃
measures
INFORMATION LCD monitor display adjustment Operation permitted at all times
USB Batch print operation Only when measurement is stopped… Option

8
Menu screen (general settings)

4.3.2 General settings

This screen is used to make basic settings for measurement.


For multi-series specifications, press the [>>] key to migrate. For single series specifications, the [>>] key
is not displayed.

≪Virtual keypad≫
Displayed when the background color is yellow

1000
1000
1

54.62
1.004
50.00

【 Setting details】

Process Contents
Measurement Measurement operation settings START、STOP Note1
control
MEAS. RANGE Measurement range setting 0.00~9999.99 mg/L Note2
CONC. ALARM Upper concentration limit 0.00~9999.99 mg/L Note2
alarm「C.ALARM」settings
MEAS. COUNT Setting the number of 0~9 times
measurements
ZERO Zero calibration voltage value 0.000~999.99 V
setting
SPAN Span calibration voltage value Zeros setting value~999.99 V
setting
HALF Half calibration voltage value 0.000~99.99 %
setting

Note1: Displayed only for multi-sample line specifications.


Note2: Up to 4 significant digits and 3 digits after the decimal point (5 digits including
The decimal point) can be set.

9
Menu screen (Device settings)

4.3.3 Device settings

Displays each device setting.


Touch the item you want to change to move to the setting change screen.

【 Operation Permissions 】

Process Contents
MEAS. Measurement process setting Only when measurement is stopped
ALARM Setting alarm conditions 〃
OPTION Option settings 〃
DISPLAY LCD monitor display adjustment Operation permitted at all times
Installation location information
LOCATION Only when measurement is stopped
registration
COMM. Communication device settings 〃
Setting the temperature of the
TEMP. 〃
relevant parts

10
Menu screen (Measurement settings)

4.3.3.1 Measurement settings

This screen allows you to make settings related to the sequence during measurement.

≪Virtual keypad≫
Displayed when the background color is yellow
50
15
99
1
24

【 Setting details】

Process Contents
SAMPLE SUCTION Setting the sample suction time 1~999 seconds
SAMPLE WEIGHING Setting of sample weighing 1~99 seconds
(backwash) time
PRELIMINARY WASH Setting the cleaning time before 1~99 hours
measurement
WASH COUNT Setting the count for washing 0~9 times
WASH CYCLE Setting the number for wash 0~999 times
cycles

11
Menu screen (Alarm settings)

4.3.3.2 Alarm settings


This screen is for setting the process operation when each alarm is generated.

≪Virtual keypad≫
0 Displayed when the background color is yellow
0

3
3
3
3

【 Setting details】

Process Contents
SHUT SAMPLE Sample cut 1~9 Times (operation
continues with 0 times
setting)
CONC. ALARM Upper concentration limit 〃
SUCTION FAIL The maximum number of times of suction failure allowed until
the measurement of the analyzer comes to complete stop.
SAMPLE Sample measurement failure 1~9 Times (operation
continues with 0 times
setting)
DIL. WATER Dilution measurement failure 〃
REAGENT Hydrochloric acid solution 〃
measurement failure
BUFFER Buffer measurement failure 〃

12
Menu screen (Option settings)

4.3.3.3 Option settings


This screen is for setting the accessory equipment of the device.

≪Virtual keypad≫
Displayed when the background color is yellow

1
YES
NONE
NONE

0 000 00
0 000 00

13
Menu screen (Option settings)

【 Setting details】

Process Contents
MEAS. LINES Measurement lines setting 1~6 …
Option
DIL. MEAS. Setting with or without dilution YES, NONE …
mechanism Option
AUTO CAL. Automatic calibration mechanism 〃 …
settings Option
RY OUTPUT1 Arbitrary external output signal STEP(0~999)-(0~99seconds) …
(contact output) setting Option
RY OUTPUT1 Arbitrary external output signal STEP(0~999)-(0~99seconds) …
(contact output) setting Option
AUTO CAL. Set the details of the automatic calibration mechanism
SETTING
CAL. CYCLE Automatic calibration cycle 0~99 day …Display only for
setting automatic calibration
START TIME Setting the start time for 0~23 hour …Display only for
automatic calibration automatic calibration
EFFECTIVE Effective range setting when 0.000~9.999 V…Display only for
RANGE updating automatic calibration automatic calibration
CAL. TIMES Setting the number of automatic 0~99 times …Display only for
calibration cycles automatic calibration
VALID TIMES Setting the number of valid 0~99 times …Display only for
automatic calibrations automatic calibration
CAL. VALUE Calibration value update NONE, YES …Display only for
UPDATE settings for automatic automatic calibration
calibration

Note: The optional setting items will not work even if they are set unless the mechanism and function are
added.

14
Menu screen (Display settings)

4.3.3.4 Display settings


This screen is for setting the brightness and backlight of the device.

31

00

【 Settings ・ Operation details 】

Process Contents
BRIGHTNESS Adjust the brightness of the LCD monitor
[DECREASE]key Set the LCD monitor to dark
[INCREASE]key Set the LCD monitor to bright
BACKLIGHT OFF Time setting to turn off the backlight when there is no screen
TIME (min) operation ※
[-1]key Set the backlight off time (1 minute increments)
[+1]key Set the backlight off time (1 minute increments)
[SET]key Save the set data

※ When set to '00', the backlight is always on.

15
Menu screen (station name settings)

4.3.3.5 Station name settings


This screen is for setting the model name and location of the device.
Alphanumeric (full-width) and (half-width) input settings are possible.

≪Virtual keypad ≫
TN-310 Displayed when the background color is
yellow

ANATEC YANACO

≪INPUT example≫
INPUT NAME・・・ANATEC YANACO
INPUT ORDER
『ANATEC YANACO』
[A] [N] [A] [T] [E] [C] [SPACE] [Y] [A] [N] [A] [K] [O] ・・
[ENT]
[SET]

【 DISPLAY CONTENTS】
Process Contents
MODEL NAME Displays the model name of the TN-310 STD
device
LOCATION Set and register the Up to 10 full-width
installation location characters (20 half-width
characters)
[SET]KEY Save the set data

16
Menu screen (Communication settings)

4.3.3.6 Communication settings


This screen is for setting communication to the printer and for other external communication.

≪Virtual keypad≫
Displayed when the background color is yellow

【 Settings 】

Process Contents
PRINTER Set the communication speed of the printer
SPEED SET the communication speed 4800,9600 bps
EXTERNAL COMM. SET the external communication speed
SPEED Speed settings 4800,9600 bps
DATA LENGTH Data length settings 7,8 bits
PARITY BIT Error detection settings EVEN, ODD, NONE
STOP BIT Set the number of end bits 1,2 stop
[SET]key Save the set data

17
Menu screen (Temperature settings)

4.3.3.7 Temperature settings


This screen is for setting the temperature of the relevant parts.

≪Virtual keypad≫
Displayed when the background color is yellow
2.00
2.00

2.00
2.00

【 Settings】

Process Contents
DEVICE NAME Displays the name of the part for Dehumidifier, Thermo. Oven ,
which the temperature is set. ozonolysis
CURRENT Displays the current temperature in real time
TEMP.
PRESET TEMP. Constant temperature setting 0.00~160.00 ℃
PROPORTIONAL Proportional control range setting
RANGE
(H)SET+ Higher limit setting of 0.00~50.00 ℃
proportional temperature
(L)SET- Lower limit setting of 0.00~50.00 ℃
proportional temperature
[SET]key Save the set data

18
Menu screen (Time settings)

4.3.4 Time settings


This screen is for setting the calendar and clock (year, month, day, hour, minute, second) of the device.
If you touch each numerical part directly, the virtual keyboard will be displayed.
Enter the number you want to set and press the [SET] key to start the clock with the changed date.

≪Virtual keypad≫
20 11 16 Displayed when the background color is yellow

11 58 00

【 Settings ・ Operation details】

Process Contents
Y Year settings 00~99Y
M Month settings 01~12M
D Day settings 01~31D
H Hour settings 00~23H
M Month settings 00~59M
S Seconds settings 00~59S
[SET]key Save the set data

19
Menu screen (Arbitrary settings)

4.3.5 Arbitrary settings


This screen is to display the current process operation and to operate the process (step) arbitrarily.
It is possible to turn the load on and off, pause the process, and jump to any process.

≪Virtual keypad≫
Displayed when the background color is yellow

1 10
10
58

【 Settings ・ Operation details】

Process Contents
PRESENT(STEP) Displays present step
TARGET(STEP) Set the step number arbitrarily STEP 1~999
(used to jump the steps)
[RUN]key After setting the target step, perform the operation

20
Menu screen (Flow operations)

4.3.6 Flow operations


This screen selects and executes a specific operation flow process.
After touch-selecting the key you want to operate, press the [RUN] key to start the operation.
The [RUN] key blinks while the flow operation is being executed.

The key indication is as follows

Unselected key

Selected key (bright)

【 Settings 】

Process Contents
[WASH1]key Introduce wash water into the mixing pipe and drain it
[BFWASH] key Introduce wash water into the buffer pipe and drain it
[BF DRAIN] key Drain the buffer pipe
[DROP] key Introduce wash water into the mixing tube and drop it from the slide valve
through the buffer tube
[ME. TUBE] key Drain the liquid from each sample / diluted water measuring tube
[MI.TUBE] key Drain the liquid from the mixing tube
[DEHU.] key Drain the liquid from the dehumidifier tube
[SV DRIVE] key Drives the slide valve back and forth
[SAMP.1] key After sucking and weighing the sample in the sample measuring tube, it is
introduced into the mixing tube. *
[DIL.1] key After sucking and measuring the diluted water in the diluted water
Measuring pipe, it is introduced into the mixing pipe……option
[CAL.] key The calibration liquid is sucked and discharged into the buffer tube.
……option
[RUN] key After selecting the item you want to operate, execute the operation.
* The display of samples 2 to 5 changes depending on the number of smple lines. (option)

21
Menu screen (Load operations)

4.3.7 Load operations


Each load can be operated arbitrarily through this screen.
After directly touch-selecting the key of the load to operate, turn ON / OFF with the [Output] key.
Multiple loads can be selected at the same time. (24V DC load and 100V AC load can be
selected at the same time)

22
【 Operation details 】

DC24V Load DC24V Load


Key Load name Key Load name Key Load name
[L1]key Unused [L18] key Washing water [L35] key Unused
→ mixing tank
[L2] key Negative [L19] key Diluting water [L36] key 〃
pressure tank side
[L3] key Positive [L20] key Mixing tube [L37] key 〃
pressure drain
[L4] key Manifold drain [L21] key Positive [L38] key 〃
pressure pump
[L5] key Sample 1 [L22] key Cleaning water [L39] key Ozone power
suction unit entrance
control
[L6] key 〃 [L23] key Unused [L40] key Temperature
introduction controller
control
[L7] key Sample 2 [L24] key 〃
suction
[L8] key 〃 [L25] key Slider valve AC100V Load
introduction drive
[L9] key Sample 3 [L26] key Slider valve Key Load name
suction return
[L10] key 〃 [L27] key Dehumidifier [L81] key Negative
introduction (SP) drain
pressure pump
[L11] key Diluted water [L28] key Dehumidifier [L82] key Unused
suction (OZ) drain
[L12] key 〃 [L29] key Unused [L83] key 〃
introduction
[L13] key Unused [L30] key 〃 [L84] key 〃
[L14] key 〃 [L31] key 〃 [L85] key 〃
[L15] key Sample / [L32] key 〃 [L86] key 〃
calibration
sol. switching
[L16] key Buffer tube [L33] key 〃 [L87] key 〃
suction
[L17] key Unused [L34] key 〃 [L88] key 〃

23
Menu screen (Print)

4.3.8 Print
This screen allows you to print internal setting information, daily reports, and curve history all at once.
If you touch each numerical part directly, the virtual keyboard will be displayed.

≪Virtual keypad≫
Displayed when the background color is yellow

20 11 15 9 00

20 11 16 9 00

【 Operation details 】
Process Contents
BASIC All operation settings and basic settings are printed at once.
DEVICE Print all device settings at once
SCHEDULE Prints schedule settings all at once
DAILY REPORT
[TEST PRINT]key Print the test pattern
[PAPER WINDER]key Winding the printing paper
DAILY Displays the period of daily report data saved in the internal
memory data
CURVE Displays the period of curve data stored in internal memory
data
ENTER THE PERIOD Select the print range of internal memory
START Start date and time for specifying the period of internal memory
data
END Specify the end date and time of the period specification of the
internal memory data
[DAILY]key Prints the daily report for the period selected by specifying the
period.
[CURVE]key Prints the curve data for the period selected by specifying the
period.

24
Menu screen (History)

4.3.9 History
This screen allows you to check the history of internal data all at once.
The initial screen displays the measured value data history screen.
* The history retention period is when the number of measurements is set to once every hour.
If the measurement is performed at a cycle earlier than this number, the storage period will be shortened
accordingly.

≪Measurement value history batch display screen≫


Displays the measured value history.
The measured value data is saved for 12 months, and after that, it is deleted from the past data at any time.

1 100.0 Displays the measurement data stored in


the internal memory
1 100.0

1 100.0
1 100.0
1 100.0
1 100.0

1 100.0
1 100.0
1 100.0
1 100.0
1 200

25
Menu screen (History)

≪Measurement value history batch display screen≫


Displays the measured value history.
The detection curve data is saved for 2 months, and after that, the past data is deleted at any time.

Displays date, measurement line.

100.0 Displays the measured value.

54.62
Displays the measured voltage, base
1.096 5.048
voltage, and peak voltage bar graph at
the time of measurement.

The detection curve at the time of


measurement is displayed.

Moves to the next history screen.

Use the 「 MEAS. 」 「 CAL. 」 「 ALARM 」


「DAILY」to realize the history data.
The key of the current screen is not
displayed.

26
Menu screen (History)

≪Calibration value history batch display screen≫


Displays the calibration value history.
The calibration value data is saved for 12 months, and after that, past data is deleted at any time.

1 100.0 Displays the measurement data stored in


the internal memory
1 100.0
1 100.0
1 100.0
1 100.0
1 100.0
1 100.0
1 100.0
1 100.0
1 100.0
1 200

27
Menu screen (History)

≪Calibrated value history batch display screen≫


Displays the calibrated value history.
The detection curve data is saved for 2 months, and after that, the past data is deleted at any time.

Displays date, calibration details,


measurement line details.
50.0 Displays the calibrated value.

1.096 5.048 Displays base voltage, and peak


voltage measurement.

The detection curve at the time of


measurement is displayed.

Moves to the next history screen.

28
Menu screen (History)

≪Alarm history display screen≫


The alarm history is displayed.
The latest 10 alarm histories are saved, and after that, the past data is deleted at any time.

Displays the time, date and alarm


details.

29
Menu screen (History)

≪Daily report history display screen≫


Displays the daily report history.
The daily report history saves the latest 16 cases, and after that, it is deleted from the past data at any time.

Displays date, measurement line.

100.0 The maximum measured concentration and


54.62 measured voltage are displayed.

25.00 The minimum measured concentration and


12.58 measured voltage are displayed.

50.00 The average measured concentration and


27.68 measured voltage are displayed.

Moves to the next history screen.

30
Menu screen (Present value)

4.3.10 Present values


This screen allows you to check the current value at once.
Each current value is displayed in real time.

P.C.B

【 Settings 】

Process Contents
AMP. IN (Filter) The preamplifier input voltage value is displayed in real time
AMP. IN The preamplifier input voltage value is displayed in real time
DEHUMIDIFIER Temperature of the dehumidifier is displayed in real time
OZONOLYSIS CATALYST Temperature of the temperature bath of the ozone decomposition
catalyst is displayed in real time
THERMOSTAT OVEN Temperature of the constant temperature bath is displayed in real
time
P.C.B TEMP Temperature of the PCB board is displayed in real time
AGND
+5V Power supply voltage value is displayed in real time

31
Menu screen (Input/output)

4.3.11 Common input/output

≪Common input / output display screen≫


This screen displays input (liquid level detection, start signal, etc.) and output signals (alarm / abnormality
signals, etc.).

Output signal that needs to be realized can be


turned ON / OFF by directly touching the keys.
(Multiple keys can be selected at the same time)

Moves to the analog input / output screen.

32
Menu screen (Input/output)

【 Display contents and operations】


Input signal Output signal
Key Load name Key Load name Key Load name
[01] [MT] Under [01] key SAMPLE
maintenance
switch
[02] Heater malfunction [ST] External [02] key WATER
start signal
[03] [BD] [03] key TROUBLE
[04] Cleaning water unit [R1] [04] key C.ALARM1
[05] [R2] [05] key C.ALARM2
[06] Buffer tube [R3] [06] key C.ALARM3 ※option
[07] Sample/dilution [R4] [07] key Optional contact output
water 1 ※option
[08] [R5] [08] key Optional contact output
2 ※option
[09] Sample water [09] key
collection unit1
※option
[10] Sample water [10] key
collection unit2
※option
[11] Sample water [11] key
collection unit3
※option
[12] [12] key
[13] [13] key
[14] [14] key
[15] [15] key
[16]

33
Menu screen (Input/output)

≪Analog input / output display screen≫


This screen allows you to simulate the analog output of the device to the measured value output terminal.
By pressing the key of the voltage and current value, the intended output value can be realized through
output terminal.

【 Operation settings】

Process Contents
[ 4.0mA( 0mV)] key( 0%output) Output Current( 4.0mA)and voltage( OmV)key
[ 8.0mA( 250mV)] key ( 25% output) Output Current( 8.0mA)and voltage( 250mV)key
[12.0mA( 500mV)] key ( 50% output) Output Current( 12.0mA)and voltage( 500mV)key
[16.0mA( 750mV)] key ( 75% output) Output Current( 16.0mA)and voltage( 750mV)key
[20.0mA(1000mV)] key (100% output) Output Current( 20.0mA)and voltage( 1000mV)key
[ 7.2mA( 200mV)] key ( 20% output) Output Current( 7.2mA)and voltage( 200mV)key
[10.4mA( 400mV)] key ( 40% output) Output Current( 10.4mA)and voltage( 400mV)key
[13.6mA( 600mV)] key ( 60% output) Output Current( 13.6mA)and voltage( 600mV)key
[16.8mA( 800mV)] key ( 80% output) Output Current( 16.8mA)and voltage( 800mV)key

34
Menu screen (Malfunction)

4.3.12 Malfunction
≪Device malfunction selection screen≫
This screen allows you to check the main causes when an error is occurred in the device.
You can move to the screen of each selection by touching the key.
If you need other abnormalities or more detailed information, see Chapter 7, Failures and Countermeasures.
* The displayed contents are limited to the main causes.

【 Operation settings】
Process Contents
[Inquiries] key Moves to the confirmation screen when making inquiries
[SHUT SAMPLE] key Moves to the cause and confirmation item screen when the
sample shut alarm is issued
[SHUT WATER] key Moves to the cause and confirmation items screen when the
washing water alarm is issued
[FAULT SUCK] key Moves to the cause and confirmation items screen when the
measuring tube suction failure alarm is issued
[MEAS. VALUE(LOW)] key Moves to the cause and confirmation items screen when the
measured value drops
[MEAS. VALUE(RISE)] key Moves to the cause and confirmation items screen when the
measured value rises
[MEAS. VALUE(FAULT)]key Moves to the cause and confirmation items screen when the
measured value is abnormal

35
≪Inquiries≫
This screen allows you to check what kind of information you will need when making inquiries.
Please check the contents of the confirmation items and contact the person in charge.
You can communicate smoothly by confirming the items to be confirmed in advance.

36
≪Each abnormality confirmation screen≫
This screen displays the main possible causes and confirmation items for the event that occurred.
Checking when an alarm occurs will lead to early detection of problems.

Trouble names are displayed

Trouble names are displayed

The items to be confirmed for each


abnormality are displayed.

37
Menu screen (Information)

4.3.13 Information
This screen allows you to check the device program and initialize the data.
Touch the key to move to next screen.

【 Operation permissions 】

Process Contents

Information on analyzer
PROGRAM Operation permission at all times
program version etc.
INITIALIZE Format internal memory Only when measurement is stopped

38
Menu screen (Program)

4.3.13.1 Program
Device name, firmware information, and installation location are displayed on this screen.

【 Display contents 】

Process Contents
MODEL Displays the model of the device TN-310 STD
PROG.ID Displays the version of the main Version name-Created date
program software
DISP.ID LCD screen ID STD001A 200731
DISP. Ver Displays the screen software
version of the LCD display
LOCATION Station name Displays the set Up to 10 full-width
installation location characters (20 half-width
characters)
OTHER Shows the product version of the
LCD display

39
Menu screen (Initialize)

4.3.13.2 Initialize
This screen initializes the data in the internal memory.
There are three types of clear: data clear that initializes the measurement data, set value clear that initializes
the set value, and ALL clear that initializes all the contents.
After directly touching and selecting the item you want to operate, press the [RUN] key to pop up a
confirmation process, and press the "Yes" key to execute the process.
When clearing the set value and ALL, print or write down the current set value.

≪Indication≫
Formats all the data

【 Operation settings 】

Process Contents
ALL DATA clear Both data clear and set value clear are performed at once.
SET DATA clear Returns various device settings to the default settings
MEAS. DATA clear Clears the measured and recorded data
[Select]key Select the item you want to operate
[Run]key After selecting the item, you want to operate, execute the operation

40
Initial settings list

4.4 Initial settings list


The table below shows the initial and factory default settings.
If you execute "SET DATA clear" or " ALL DATA CLEAR" on the memory clear screen, the default setting
value will be restored.

[Setting list] (operation settings, schedule settings, basic settings)

Screen Process Print Default values Factory default


settings
Operation mode OPE. MODE [MEAS.] As per internal
inspection
Meas. start START MODE [INTERNAL] Same as on left
Meas. start MEAS. START [ 0] min Same as on the
left
Meas. cycle MEAS. CYCLE [60] min Same as on the
Operation left
Settings Print output PRINT OUTPUT [OPE.] Same as on the
List left
Print direction PRINT DIR. [FORWARD] Same as on the
left
Bar graph PRINT BAR [STOP] Same as on the
left
Curve graph PRINT CURVE [STOP] Same as on the
left
Blank Same as on the
Schedule Time setting all [1]
left
Settings
Blank Same as on the
List Date setting all [O]
left
Meas. control Blank [MEAS.] Same as on the
left
Meas. range MEAS. RANGE [100.0] mg/L As per the
delivery spec.
Upper CONC. ALARM [100.0] mg/L As per the
concentration delivery spec.
Basic limit
Settings Meas. times MEAS. COUNT [1] times Same as on the
List left
Zero value ZERO VOLT [0.000] V As per the
delivery spec.
Span value SPAN VOLT [40.00] V As per the
delivery spec.
Half value HALF [50.00] % As per the
delivery spec.

41
Initial settings list

[Setting list] (Device settings)


Screen Process Print Default values Factory default settings

Measurement settings

Sample suction SAPMLE SUCK [50] sec Same as on the left

Sample measurement SAMPLE BLOW [15] sec Same as on the left

Preliminary wash BEFORE WASH [99] hour Same as on the left

Wash count WASH COUNT [0] times Same as on the left

Wash cycle WASH CYCLE [0] times Same as on the left

Alarm settings

Shut sample NO SAMPLE [0] times Same as on the left

Upper conc. alarm CONC. ALARM [0] times Same as on the left

Suction fail/sample SAMPLE [3] times Same as on the left

Suction fail/water WATER [3] times Same as on the left

Suction fail/reagent REAGENT [3] times Same as on the left

Suction fail/buffer BUFFER [3] times Same as on the left

Option settings

Measurement line MEAS. LINE [1] line As per the delivery spec.

Dilution meas. DILUTION [NONE] As per the delivery spec.

Auto cal. AUTO CAL. [NONE] As per the delivery spec.

Output signal1 RY OUTPUT1 step[ 0]-[ 0] sec As per the delivery spec.

Output signal2 RY OUTPUT2 step[ 0]-[ 0] sec As per the delivery spec.

Auto calibration cycle CAL. CYCLE [7] 日毎 Same as on the left

Auto cal. start time START TIME [00] 時 Same as on the left

Auto calibration EFFECTIVE RANGE [10.00] V Same as on the left

effective range

Device settings Calibration times CAL. TIMES [3] times Same as on the left

Valid calibration range VALI TIMES [3] times Same as on the left

Calibration value update CAL. VALUE UPDATE [update] Same as on the left

Screen settings

Back light intensity No display [31] Same as on the left

Automatic turn-off time No display [00] Same as on the left

Locations settings

Model name No display [TN-310 STD] Same as on the left

Location Loc.: [Blank (no entry)] Same as on the left

Communication settings

Printer comm. speed [9600] bps Same as on the left

Temperature settings

Dehumidifier

Set temperature SET TEMP. [ 3.00] ℃ Same as on the left

Proportional range(H) RATIO RANGE(+) [ 4.00] ℃ Same as on the left

Proportional range(L) RATIO RANGE(-) [ 1.00] ℃ Same as on the left

Ozonolysis

Set temperature SET TEMP. [100.00] ℃ Same as on the left

Proportional range(H) RATIO RANGE(+) [ 0.50] ℃ Same as on the left

Proportional range(L) RATIO RANGE(-) [ 20.00] ℃ Same as on the left

Combustion chamber

Set temperature SET TEMP. [ 50.00] ℃ Same as on the left

Proportional range(H) RATIO RANGE(+) [ 1.00] ℃ Same as on the left

Proportional range(L) RATIO RANGE(-) [ 1.00] ℃ Same as on the left

42
Operation keys list

4.5 Operation keys list


The table below shows the fixed keys that can be operated for each screen.

Classif- Screen1 Screen2 Usable keys


ication

STATUS START LOAD PAUSE STEP OPE. MENU ALARM. PAPER.


DISPLAY
MAIN
OPE. START LOAD PAUSE STEP RETURN MENU ALARM. PAPER.
MENU START LOAD PAUSE STEP OPE. RETURN MENU MAIN

BASIC OPE. RETURN MENU MAIN <<,>>


DEIVE START LOAD PAUSE STEP OPE. RETURN MENU MAIN
MEAS. OPE. RETURN MENU MAIN
ALARM OPE. RETURN MENU MAIN
OPTION OPE. RETURN MENU MAIN
SET DEVICE
DISPLAY OPE. RETURN MENU MAIN
SETTING
LOCATION OPE. RETURN MENU MAIN
COMM. OPE. RETURN MENU MAIN
TEMP. OPE. RETURN MENU MAIN SET <<,>>
CLOCK OPE. RETURN MENU MAIN SET

ARBITRARY START LOAD PAUSE STEP OPE. RETURN MENU MAIN RUN
FLOW LOAD PAUSE STEP OPE. RETURN MENU MAIN RUN
OPERATE LOAD OPE. RETURN MENU MAIN OUT

PRINT OPE. RETURN MENU MAIN


RECORD MEAS. CAL. ALARM REPORT OPE. RETURN MENU MAIN <<,>>

NOW VALUE OPE. RETURN MENU MAIN


MAINT. IN/OUT OPE. RETURN MENU MAIN <<,>>
GUIDE START LOAD PAUSE STEP OPE. RETURN MENU MAIN

INFORMATION START LOAD PAUSE STEP OPE. RETURN MENU MAIN


PROGRAM OPE. RETURN MENU MAIN
INFORMATION
OTHER INITIALIZ OPE. RETURN MENU MAIN RUN
E
USB START LOAD PAUSE STEP OPE. RETURN MENU MAIN RUN

43