TN-310 Module 4 Display Operations
TN-310 Module 4 Display Operations
TN - 310
Reference modules
Contents
-1-
Module 4 Display operation Screen flow
メニュー(各種設定・操作一覧画面)
メニュー(各種設定・操作一覧画面)
DEVICE Device settings … 4・3・3 ※
USB
2
Screen selection is as follows.
DEVICE
INFO.
3
Main screen (status display)
11 58
Present step number is displayed
100.0
While operating in measurement mode, the
54.62 measurement start time, measured value, and
measured voltage value are displayed as the
previous measurement result.
【 OPERATION DETAILS】
4
Main screen (status display)
1 10
11 58
【 警 報 詳 細 】
【Alarm details】
5
Main screen (operation settings)
≪Virtual keypad≫
Displayed when the background color is yellow
00 60
stop stop
FORWARD stop
Note1: Only when the start mode is selected as internal start, it operates according to the
settings.
6
Main screen (schedule settings)
After touching the "Schedule" key on the operation screen, the schedule setting screen is displayed.
Press the "SET" key to save the settings.
0 0 0 0 0 0 0 0
0 0 0 0 0 0 0 0
0 0 0 0 0 0 0 0
【 Setting details】
Process Contents
SET (Days of Set days of SON, MON TUE, WED, THU, FRI, SAT
the week) the week
SET(TIME) Set time Enter the series number 0~1 ※
[SET]key Save the set data
7
Menu screen (Setting operation list)
This is a list screen for setting and operating each device setting, process / load operation, input / output
signal operation, etc. You can select and transfer by directly touching each key on the screen.
【 Operation Permissions 】
Process Contents
BASIC Measurement conditions settings Only when measurement is stopped
DEVICE Device settings 〃
CLOCK Calendar / clock correction 〃
ARBITRARY Arbitrary step operation Operation permitted at all times
FLOW Operation of specific flow Only when measurement is stopped
LOAD Individual load operation 〃
Operation permission for all (with some
PRINT Batch print operation
restrictions)
RECORD Display of each record Operation permitted at all times
PRESENT
Display of present value 〃
VALUE
IN/OUT IN/OUT operation 〃
Display of troubleshooting
MALFUNCTION 〃
measures
INFORMATION LCD monitor display adjustment Operation permitted at all times
USB Batch print operation Only when measurement is stopped… Option
8
Menu screen (general settings)
≪Virtual keypad≫
Displayed when the background color is yellow
1000
1000
1
54.62
1.004
50.00
【 Setting details】
Process Contents
Measurement Measurement operation settings START、STOP Note1
control
MEAS. RANGE Measurement range setting 0.00~9999.99 mg/L Note2
CONC. ALARM Upper concentration limit 0.00~9999.99 mg/L Note2
alarm「C.ALARM」settings
MEAS. COUNT Setting the number of 0~9 times
measurements
ZERO Zero calibration voltage value 0.000~999.99 V
setting
SPAN Span calibration voltage value Zeros setting value~999.99 V
setting
HALF Half calibration voltage value 0.000~99.99 %
setting
9
Menu screen (Device settings)
【 Operation Permissions 】
Process Contents
MEAS. Measurement process setting Only when measurement is stopped
ALARM Setting alarm conditions 〃
OPTION Option settings 〃
DISPLAY LCD monitor display adjustment Operation permitted at all times
Installation location information
LOCATION Only when measurement is stopped
registration
COMM. Communication device settings 〃
Setting the temperature of the
TEMP. 〃
relevant parts
10
Menu screen (Measurement settings)
This screen allows you to make settings related to the sequence during measurement.
≪Virtual keypad≫
Displayed when the background color is yellow
50
15
99
1
24
【 Setting details】
Process Contents
SAMPLE SUCTION Setting the sample suction time 1~999 seconds
SAMPLE WEIGHING Setting of sample weighing 1~99 seconds
(backwash) time
PRELIMINARY WASH Setting the cleaning time before 1~99 hours
measurement
WASH COUNT Setting the count for washing 0~9 times
WASH CYCLE Setting the number for wash 0~999 times
cycles
11
Menu screen (Alarm settings)
≪Virtual keypad≫
0 Displayed when the background color is yellow
0
3
3
3
3
【 Setting details】
Process Contents
SHUT SAMPLE Sample cut 1~9 Times (operation
continues with 0 times
setting)
CONC. ALARM Upper concentration limit 〃
SUCTION FAIL The maximum number of times of suction failure allowed until
the measurement of the analyzer comes to complete stop.
SAMPLE Sample measurement failure 1~9 Times (operation
continues with 0 times
setting)
DIL. WATER Dilution measurement failure 〃
REAGENT Hydrochloric acid solution 〃
measurement failure
BUFFER Buffer measurement failure 〃
12
Menu screen (Option settings)
≪Virtual keypad≫
Displayed when the background color is yellow
1
YES
NONE
NONE
0 000 00
0 000 00
13
Menu screen (Option settings)
【 Setting details】
Process Contents
MEAS. LINES Measurement lines setting 1~6 …
Option
DIL. MEAS. Setting with or without dilution YES, NONE …
mechanism Option
AUTO CAL. Automatic calibration mechanism 〃 …
settings Option
RY OUTPUT1 Arbitrary external output signal STEP(0~999)-(0~99seconds) …
(contact output) setting Option
RY OUTPUT1 Arbitrary external output signal STEP(0~999)-(0~99seconds) …
(contact output) setting Option
AUTO CAL. Set the details of the automatic calibration mechanism
SETTING
CAL. CYCLE Automatic calibration cycle 0~99 day …Display only for
setting automatic calibration
START TIME Setting the start time for 0~23 hour …Display only for
automatic calibration automatic calibration
EFFECTIVE Effective range setting when 0.000~9.999 V…Display only for
RANGE updating automatic calibration automatic calibration
CAL. TIMES Setting the number of automatic 0~99 times …Display only for
calibration cycles automatic calibration
VALID TIMES Setting the number of valid 0~99 times …Display only for
automatic calibrations automatic calibration
CAL. VALUE Calibration value update NONE, YES …Display only for
UPDATE settings for automatic automatic calibration
calibration
Note: The optional setting items will not work even if they are set unless the mechanism and function are
added.
14
Menu screen (Display settings)
31
00
Process Contents
BRIGHTNESS Adjust the brightness of the LCD monitor
[DECREASE]key Set the LCD monitor to dark
[INCREASE]key Set the LCD monitor to bright
BACKLIGHT OFF Time setting to turn off the backlight when there is no screen
TIME (min) operation ※
[-1]key Set the backlight off time (1 minute increments)
[+1]key Set the backlight off time (1 minute increments)
[SET]key Save the set data
15
Menu screen (station name settings)
≪Virtual keypad ≫
TN-310 Displayed when the background color is
yellow
ANATEC YANACO
≪INPUT example≫
INPUT NAME・・・ANATEC YANACO
INPUT ORDER
『ANATEC YANACO』
[A] [N] [A] [T] [E] [C] [SPACE] [Y] [A] [N] [A] [K] [O] ・・
[ENT]
[SET]
【 DISPLAY CONTENTS】
Process Contents
MODEL NAME Displays the model name of the TN-310 STD
device
LOCATION Set and register the Up to 10 full-width
installation location characters (20 half-width
characters)
[SET]KEY Save the set data
16
Menu screen (Communication settings)
≪Virtual keypad≫
Displayed when the background color is yellow
【 Settings 】
Process Contents
PRINTER Set the communication speed of the printer
SPEED SET the communication speed 4800,9600 bps
EXTERNAL COMM. SET the external communication speed
SPEED Speed settings 4800,9600 bps
DATA LENGTH Data length settings 7,8 bits
PARITY BIT Error detection settings EVEN, ODD, NONE
STOP BIT Set the number of end bits 1,2 stop
[SET]key Save the set data
17
Menu screen (Temperature settings)
≪Virtual keypad≫
Displayed when the background color is yellow
2.00
2.00
2.00
2.00
【 Settings】
Process Contents
DEVICE NAME Displays the name of the part for Dehumidifier, Thermo. Oven ,
which the temperature is set. ozonolysis
CURRENT Displays the current temperature in real time
TEMP.
PRESET TEMP. Constant temperature setting 0.00~160.00 ℃
PROPORTIONAL Proportional control range setting
RANGE
(H)SET+ Higher limit setting of 0.00~50.00 ℃
proportional temperature
(L)SET- Lower limit setting of 0.00~50.00 ℃
proportional temperature
[SET]key Save the set data
18
Menu screen (Time settings)
≪Virtual keypad≫
20 11 16 Displayed when the background color is yellow
11 58 00
Process Contents
Y Year settings 00~99Y
M Month settings 01~12M
D Day settings 01~31D
H Hour settings 00~23H
M Month settings 00~59M
S Seconds settings 00~59S
[SET]key Save the set data
19
Menu screen (Arbitrary settings)
≪Virtual keypad≫
Displayed when the background color is yellow
1 10
10
58
Process Contents
PRESENT(STEP) Displays present step
TARGET(STEP) Set the step number arbitrarily STEP 1~999
(used to jump the steps)
[RUN]key After setting the target step, perform the operation
20
Menu screen (Flow operations)
Unselected key
【 Settings 】
Process Contents
[WASH1]key Introduce wash water into the mixing pipe and drain it
[BFWASH] key Introduce wash water into the buffer pipe and drain it
[BF DRAIN] key Drain the buffer pipe
[DROP] key Introduce wash water into the mixing tube and drop it from the slide valve
through the buffer tube
[ME. TUBE] key Drain the liquid from each sample / diluted water measuring tube
[MI.TUBE] key Drain the liquid from the mixing tube
[DEHU.] key Drain the liquid from the dehumidifier tube
[SV DRIVE] key Drives the slide valve back and forth
[SAMP.1] key After sucking and weighing the sample in the sample measuring tube, it is
introduced into the mixing tube. *
[DIL.1] key After sucking and measuring the diluted water in the diluted water
Measuring pipe, it is introduced into the mixing pipe……option
[CAL.] key The calibration liquid is sucked and discharged into the buffer tube.
……option
[RUN] key After selecting the item you want to operate, execute the operation.
* The display of samples 2 to 5 changes depending on the number of smple lines. (option)
21
Menu screen (Load operations)
22
【 Operation details 】
23
Menu screen (Print)
4.3.8 Print
This screen allows you to print internal setting information, daily reports, and curve history all at once.
If you touch each numerical part directly, the virtual keyboard will be displayed.
≪Virtual keypad≫
Displayed when the background color is yellow
20 11 15 9 00
20 11 16 9 00
【 Operation details 】
Process Contents
BASIC All operation settings and basic settings are printed at once.
DEVICE Print all device settings at once
SCHEDULE Prints schedule settings all at once
DAILY REPORT
[TEST PRINT]key Print the test pattern
[PAPER WINDER]key Winding the printing paper
DAILY Displays the period of daily report data saved in the internal
memory data
CURVE Displays the period of curve data stored in internal memory
data
ENTER THE PERIOD Select the print range of internal memory
START Start date and time for specifying the period of internal memory
data
END Specify the end date and time of the period specification of the
internal memory data
[DAILY]key Prints the daily report for the period selected by specifying the
period.
[CURVE]key Prints the curve data for the period selected by specifying the
period.
24
Menu screen (History)
4.3.9 History
This screen allows you to check the history of internal data all at once.
The initial screen displays the measured value data history screen.
* The history retention period is when the number of measurements is set to once every hour.
If the measurement is performed at a cycle earlier than this number, the storage period will be shortened
accordingly.
1 100.0
1 100.0
1 100.0
1 100.0
1 100.0
1 100.0
1 100.0
1 100.0
1 200
25
Menu screen (History)
54.62
Displays the measured voltage, base
1.096 5.048
voltage, and peak voltage bar graph at
the time of measurement.
26
Menu screen (History)
27
Menu screen (History)
28
Menu screen (History)
29
Menu screen (History)
30
Menu screen (Present value)
P.C.B
【 Settings 】
Process Contents
AMP. IN (Filter) The preamplifier input voltage value is displayed in real time
AMP. IN The preamplifier input voltage value is displayed in real time
DEHUMIDIFIER Temperature of the dehumidifier is displayed in real time
OZONOLYSIS CATALYST Temperature of the temperature bath of the ozone decomposition
catalyst is displayed in real time
THERMOSTAT OVEN Temperature of the constant temperature bath is displayed in real
time
P.C.B TEMP Temperature of the PCB board is displayed in real time
AGND
+5V Power supply voltage value is displayed in real time
31
Menu screen (Input/output)
32
Menu screen (Input/output)
33
Menu screen (Input/output)
【 Operation settings】
Process Contents
[ 4.0mA( 0mV)] key( 0%output) Output Current( 4.0mA)and voltage( OmV)key
[ 8.0mA( 250mV)] key ( 25% output) Output Current( 8.0mA)and voltage( 250mV)key
[12.0mA( 500mV)] key ( 50% output) Output Current( 12.0mA)and voltage( 500mV)key
[16.0mA( 750mV)] key ( 75% output) Output Current( 16.0mA)and voltage( 750mV)key
[20.0mA(1000mV)] key (100% output) Output Current( 20.0mA)and voltage( 1000mV)key
[ 7.2mA( 200mV)] key ( 20% output) Output Current( 7.2mA)and voltage( 200mV)key
[10.4mA( 400mV)] key ( 40% output) Output Current( 10.4mA)and voltage( 400mV)key
[13.6mA( 600mV)] key ( 60% output) Output Current( 13.6mA)and voltage( 600mV)key
[16.8mA( 800mV)] key ( 80% output) Output Current( 16.8mA)and voltage( 800mV)key
34
Menu screen (Malfunction)
4.3.12 Malfunction
≪Device malfunction selection screen≫
This screen allows you to check the main causes when an error is occurred in the device.
You can move to the screen of each selection by touching the key.
If you need other abnormalities or more detailed information, see Chapter 7, Failures and Countermeasures.
* The displayed contents are limited to the main causes.
【 Operation settings】
Process Contents
[Inquiries] key Moves to the confirmation screen when making inquiries
[SHUT SAMPLE] key Moves to the cause and confirmation item screen when the
sample shut alarm is issued
[SHUT WATER] key Moves to the cause and confirmation items screen when the
washing water alarm is issued
[FAULT SUCK] key Moves to the cause and confirmation items screen when the
measuring tube suction failure alarm is issued
[MEAS. VALUE(LOW)] key Moves to the cause and confirmation items screen when the
measured value drops
[MEAS. VALUE(RISE)] key Moves to the cause and confirmation items screen when the
measured value rises
[MEAS. VALUE(FAULT)]key Moves to the cause and confirmation items screen when the
measured value is abnormal
35
≪Inquiries≫
This screen allows you to check what kind of information you will need when making inquiries.
Please check the contents of the confirmation items and contact the person in charge.
You can communicate smoothly by confirming the items to be confirmed in advance.
36
≪Each abnormality confirmation screen≫
This screen displays the main possible causes and confirmation items for the event that occurred.
Checking when an alarm occurs will lead to early detection of problems.
37
Menu screen (Information)
4.3.13 Information
This screen allows you to check the device program and initialize the data.
Touch the key to move to next screen.
【 Operation permissions 】
Process Contents
Information on analyzer
PROGRAM Operation permission at all times
program version etc.
INITIALIZE Format internal memory Only when measurement is stopped
38
Menu screen (Program)
4.3.13.1 Program
Device name, firmware information, and installation location are displayed on this screen.
【 Display contents 】
Process Contents
MODEL Displays the model of the device TN-310 STD
PROG.ID Displays the version of the main Version name-Created date
program software
DISP.ID LCD screen ID STD001A 200731
DISP. Ver Displays the screen software
version of the LCD display
LOCATION Station name Displays the set Up to 10 full-width
installation location characters (20 half-width
characters)
OTHER Shows the product version of the
LCD display
39
Menu screen (Initialize)
4.3.13.2 Initialize
This screen initializes the data in the internal memory.
There are three types of clear: data clear that initializes the measurement data, set value clear that initializes
the set value, and ALL clear that initializes all the contents.
After directly touching and selecting the item you want to operate, press the [RUN] key to pop up a
confirmation process, and press the "Yes" key to execute the process.
When clearing the set value and ALL, print or write down the current set value.
≪Indication≫
Formats all the data
【 Operation settings 】
Process Contents
ALL DATA clear Both data clear and set value clear are performed at once.
SET DATA clear Returns various device settings to the default settings
MEAS. DATA clear Clears the measured and recorded data
[Select]key Select the item you want to operate
[Run]key After selecting the item, you want to operate, execute the operation
40
Initial settings list
41
Initial settings list
Measurement settings
Alarm settings
Upper conc. alarm CONC. ALARM [0] times Same as on the left
Option settings
Measurement line MEAS. LINE [1] line As per the delivery spec.
Output signal1 RY OUTPUT1 step[ 0]-[ 0] sec As per the delivery spec.
Output signal2 RY OUTPUT2 step[ 0]-[ 0] sec As per the delivery spec.
Auto cal. start time START TIME [00] 時 Same as on the left
effective range
Device settings Calibration times CAL. TIMES [3] times Same as on the left
Valid calibration range VALI TIMES [3] times Same as on the left
Calibration value update CAL. VALUE UPDATE [update] Same as on the left
Screen settings
Locations settings
Communication settings
Temperature settings
Dehumidifier
Ozonolysis
Combustion chamber
42
Operation keys list
ARBITRARY START LOAD PAUSE STEP OPE. RETURN MENU MAIN RUN
FLOW LOAD PAUSE STEP OPE. RETURN MENU MAIN RUN
OPERATE LOAD OPE. RETURN MENU MAIN OUT
43