0% found this document useful (0 votes)
38 views

NCL Scope - 2022 1

This document summarizes the accreditation of N.C.L. PVT. LTD. laboratory in Ghaziabad, India. The laboratory has been accredited according to ISO/IEC 17025:2017 for calibration in multiple fields including alternating current. The accreditation is valid until January 2022 and covers calibration and measurement capabilities with uncertainties between 0.02-0.23% for measuring and sourcing alternating currents from 10 microamps to 30 amps at frequencies between 10Hz to 30kHz. The scope of accreditation is available on the NABL website.

Uploaded by

Samaksh Sharma
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
38 views

NCL Scope - 2022 1

This document summarizes the accreditation of N.C.L. PVT. LTD. laboratory in Ghaziabad, India. The laboratory has been accredited according to ISO/IEC 17025:2017 for calibration in multiple fields including alternating current. The accreditation is valid until January 2022 and covers calibration and measurement capabilities with uncertainties between 0.02-0.23% for measuring and sourcing alternating currents from 10 microamps to 30 amps at frequencies between 10Hz to 30kHz. The scope of accreditation is available on the NABL website.

Uploaded by

Samaksh Sharma
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 88

N.C.L. PVT. LTD.

has been assessed and accredited in accordance with the standard

ISO/IEC 17025:2017
''General Requirements for the Competence of Testing &
Calibration Laboratories''
for its facilities at
B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH, INDIA

in the field of

CALIBRATION
Certificate Number: CC-2213
Issue Date: 29/01/2020 Valid Until: 28/01/2022

This certificate remains valid for the Scope of Accreditation as specified in the annexure subject to continued
satisfactory compliance to the above standard & the relevant requirements of NABL.
(To see the scope of accreditation of this laboratory, you may also visit NABL website www.nabl-india.org)

Signed for and on behalf of NABL

N. Venkateswaran
Chief Executive Officer
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 1 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Permanent Facility

1 ELECTRO- AC Current@10Hz to Using Fluke 8588A by 1 A to 10 A 0.04% to 0.09%


TECHNICAL- 10kHz Direct Method
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

2 ELECTRO- AC Current@10Hz to Using Fluke 8588A by 1 mA to 10 mA 0.06% to 0.08%


TECHNICAL- 10kHz Direct Method
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

3 ELECTRO- AC Current@10Hz to Using Fluke 8588A By 10 µA to 100 µA 0.2% to 0.03%


TECHNICAL- 10kHz Direct Method
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

4 ELECTRO- AC Current@10Hz to Using Fluke 8588A by 10 A to 30 A 0.09% to 0.07%


TECHNICAL- 10kHz Direct Method
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

5 ELECTRO- AC Current@10Hz to Using Fluke 8588A by 10 mA to 100 mA 0.08%


TECHNICAL- 10kHz Direct Method
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 2 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
6 ELECTRO- AC Current@10Hz to Using Fluke 8588A By 100 µA to 1 mA 0.03% to 0.06%
TECHNICAL- 10kHz Direct Method
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

7 ELECTRO- AC Current@10Hz to Using Fluke 8588A by 100 mA to 1 A 0.08% to 0.041%


TECHNICAL- 10kHz Direct Method
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

8 ELECTRO- AC Current@10kHz to Using Fluke 8588A by 1 mA to 10 mA 0.06% to 0.07%


TECHNICAL- 30kHz Direct Method
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

9 ELECTRO- AC Current@10kHz to Using Fluke 8588A by 10 µA to 100 µA 0.2% to 0.02%


TECHNICAL- 30kHz Direct Method
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

10 ELECTRO- AC Current@10kHz to Using Fluke 8588A by 10 mA to 100 mA 0.07% to 0.08%


TECHNICAL- 30kHz Direct Method
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 3 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
11 ELECTRO- AC Current@10kHz to Using Fluke 8588A by 100 µA to 1 mA 0.03% to 0.06%
TECHNICAL- 30kHz Direct Method
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

12 ELECTRO- AC Current @ 10Hz to Using Fluke Calibrator 1.1 A to 2.99 A 0.17% to 0.23%
TECHNICAL- 45Hz 5502A By Direct
ALTERNATING Method
CURRENT (< 1 GHZ)
(Source)

13 ELECTRO- AC Current @ 10Hz to Using Fluke Calibrator 1.9 mA to 3.29 mA 0.11%


TECHNICAL- 45Hz 5502A By Direct
ALTERNATING Method
CURRENT (< 1 GHZ)
(Source)

14 ELECTRO- AC Current @ 10Hz to Using Fluke Calibrator 3.29 mA to 1.1 A 0.11% to 0.17%
TECHNICAL- 45Hz 5502A By Direct
ALTERNATING Method
CURRENT (< 1 GHZ)
(Source)

15 ELECTRO- AC Current @ 10Hz to Using Fluke Calibrator 33 µA to 330 µA 0.14% to 0.23%


TECHNICAL- 45Hz 5502A By Direct
ALTERNATING Method
CURRENT (< 1 GHZ)
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 4 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
16 ELECTRO- AC Current @ 10Hz to Using Fluke Calibrator 330 µA to 1.9 mA 0.23% to 0.11%
TECHNICAL- 45Hz 5502A By Direct
ALTERNATING Method
CURRENT (< 1 GHZ)
(Source)

17 ELECTRO- AC Current @ 1kHz o Using Fluke Calibrator 33 µA to 190 µA 0.14% to 0.21%


TECHNICAL- 30kHZ 5502A By Direct
ALTERNATING Method
CURRENT (< 1 GHZ)
(Source)

18 ELECTRO- AC Current @ 45Hz to Using Fluke Calibrator 1.1 A to 2.99 A 0.07%


TECHNICAL- 1kHz 5502A By Direct
ALTERNATING Method
CURRENT (< 1 GHZ)
(Source)

19 ELECTRO- AC Current @ 45Hz to Using Fluke Calibrator 1.9 mA to 3.29 mA 0.14% to 0.23%
TECHNICAL- 1kHz 5502A By Direct
ALTERNATING Method
CURRENT (< 1 GHZ)
(Source)

20 ELECTRO- AC Current @ 45Hz to Using Fluke Calibrator 11 A to 20 A 0.11% to 0.09%


TECHNICAL- 1kHz 5502A By Direct
ALTERNATING Method
CURRENT (< 1 GHZ)
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 5 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
21 ELECTRO- AC Current @ 45Hz to Using Fluke Calibrator 2.99 A to 3.3 A 0.07% to 0.09%
TECHNICAL- 1kHz 5502A By Direct
ALTERNATING Method
CURRENT (< 1 GHZ)
(Source)

22 ELECTRO- AC Current @ 45Hz to Using Fluke Calibrator 3.29 mA to 329 mA 0.14% to 0.21%
TECHNICAL- 1kHz 5502A By Direct
ALTERNATING Method
CURRENT (< 1 GHZ)
(Source)

23 ELECTRO- AC Current @ 45Hz to Using Fluke Calibrator 3.3 A to 11 A 0.09% to 0.11%


TECHNICAL- 1kHz 5502A By Direct
ALTERNATING Method
CURRENT (< 1 GHZ)
(Source)

24 ELECTRO- AC Current @ 45Hz to Using Fluke Calibrator 329 mA to 1.1 A 0.21% to 0.07%
TECHNICAL- 1kHz 5502A By Direct
ALTERNATING Method
CURRENT (< 1 GHZ)
(Source)

25 ELECTRO- AC Current @ 45Hz to Using Fluke Calibrator 33 µA to 330 µA 0.1% to 0.2%


TECHNICAL- 1kHz 5502A By Direct
ALTERNATING Method
CURRENT (< 1 GHZ)
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 6 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
26 ELECTRO- AC Current @ 45Hz to Using Fluke Calibrator 330 µA to 1.9 mA 0.14% to 0.23%
TECHNICAL- 1kHz 5502A By Direct
ALTERNATING Method
CURRENT (< 1 GHZ)
(Source)

27 ELECTRO- AC Current@1kHz to Using Fluke 5502 190 µA to 190 mA 0.14% to 0.22%


TECHNICAL- 30kHz Calibrator by Direct
ALTERNATING Method
CURRENT (< 1 GHZ)
(Source)

28 ELECTRO- AC Current@1kHz to Using Fluke 5502 190 mA to 329 mA 0.11% to 0.23%


TECHNICAL- 30kHz Calibrator by Direct
ALTERNATING Method
CURRENT (< 1 GHZ)
(Source)

29 ELECTRO- AC Current@45Hz to Using Fluke 5502 20 A to 400 A 0.16 to 0.14


TECHNICAL- 100Hz Calibrator with Current
ALTERNATING Coil by Direct Method
CURRENT (< 1 GHZ)
(Source)

30 ELECTRO- AC Current@45Hz to Using Fluke 5502 400 A to 1000 A 0.14% to 0.17%


TECHNICAL- 100Hz Calibrator with current
ALTERNATING coil by Direct Method
CURRENT (< 1 GHZ)
(Source)

31 ELECTRO- DC Current Using Fluke 8588A By 1 µA to 10 µA 0.5% to 0.05%


TECHNICAL- DIRECT Direct Method
CURRENT (Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 7 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
32 ELECTRO- DC Current Using Fluke 8588A By 1 A to 10 A 0.013% to 0.02%
TECHNICAL- DIRECT Direct Method
CURRENT (Measure)

33 ELECTRO- DC Current Using Fluke 8588A By 10 µA to 100 µA 0.05%


TECHNICAL- DIRECT Direct method
CURRENT (Measure)

34 ELECTRO- DC Current Using Fluke 8588A By 10 A to 20 A 0.02% to 0.07%


TECHNICAL- DIRECT Direct Method
CURRENT (Measure)

35 ELECTRO- DC Current Using Fluke 8588A By 10 mA to 100 mA 0.005%


TECHNICAL- DIRECT Direct Method
CURRENT (Measure)

36 ELECTRO- DC Current Using Fluke 8588A By 100 µA to 1 mA 0.05%


TECHNICAL- DIRECT Direct Method
CURRENT (Measure)

37 ELECTRO- DC Current Using Fluke 8588A By 100 mA to 200 mA 0.005% to 0.004%


TECHNICAL- DIRECT Direct Method
CURRENT (Measure)

38 ELECTRO- DC Current Using Fluke Calibrator 100 µA to 330 µA 0.01%


TECHNICAL- DIRECT 5502A By Direct
CURRENT (Source) Method

39 ELECTRO- DC Current Using Fluke Calibrator 11 A to 20 A 0.07%


TECHNICAL- DIRECT 5502A By Direct
CURRENT (Source) Method

40 ELECTRO- DC Current Using Fluke Calibrator 2.9 A to 11 A 0.01% to 0.07%


TECHNICAL- DIRECT 5502A By Direct
CURRENT (Source) Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 8 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
41 ELECTRO- DC Current Using Fluke Calibrator 20 A to 400 A 0.07% to 0.11%
TECHNICAL- DIRECT with Current Coil
CURRENT (Source) 5502A By Direct
Method

42 ELECTRO- DC Current Using Fluke Calibrator 3.3 mA to 33 mA 0.01%


TECHNICAL- DIRECT 5502A By Direct
CURRENT (Source) Method

43 ELECTRO- DC Current Using Fluke Calibrator 33 µA to 100 µA 0.02% to 0.01%


TECHNICAL- DIRECT 5502A By Direct
CURRENT (Source) Method

44 ELECTRO- DC current Using Fluke Calibrator 33 mA to 330 mA 0.01%


TECHNICAL- DIRECT 5502A By Direct
CURRENT (Source) Method

45 ELECTRO- DC Current Using Fluke Calibrator 330 µA to 3.3 mA 0.01%


TECHNICAL- DIRECT 5502A By Direct
CURRENT (Source) Method

46 ELECTRO- DC Current Using Fluke Calibrator 330 mA to 2.9 A 0.01%


TECHNICAL- DIRECT 5502A By Direct
CURRENT (Source) Method

47 ELECTRO- DC Current Using Fluke Calibrator 400 A to 1000 A 0.11%


TECHNICAL- DIRECT with Current Coil
CURRENT (Source) 5502A By Direct
Method

48 ELECTRO- AC Current@10kHz to Using Fluke 8588A by 100 mA to 1 A 0.08% to 0.04%


TECHNICAL- 30kHz Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 9 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
49 ELECTRO- AC Power/active Using Power Meter 0.4 wh to 12 KWh 0.24% to 0.09%
TECHNICAL- Energy 50 Hz(-)0.1 PF AC/DC by Direct
ELECTRICAL to upf to0.1 PF to Method
EQUIPMENT UPF(1-Phase and 3-
(Measure) Phase)4V to 600 V(0.1
A to 20 A)

50 ELECTRO- AC Volatge@10kHz to By Using Fluke 8588A 1 mV to 10 mV 1.4% to 0.04%


TECHNICAL- 100kHz By Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

51 ELECTRO- AC Volatge@10kHz to Using Fluke 8588A By 10 mV to 100 mV 0.04% to 0.013%


TECHNICAL- 100kHz Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

52 ELECTRO- AC Volatge@10kHz to By Using Fluke 8588A 100 mV to 1 V 0.013% to 0.02%


TECHNICAL- 100kHz By Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

53 ELECTRO- AC Voltage@ 10kHz to By Using Fluke 8588A 1 V to 100 V 0.02%


TECHNICAL- 100kHz By Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 10 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
54 ELECTRO- AC Voltage@10 Hz to Using Fluke 8588A By 1 mV to 10 mV 1.07% to 0.04%
TECHNICAL- 10kHz Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

55 ELECTRO- AC Voltage@10 Hz to Using Fluke 8588A By 1 V to 100 V 0.008% to 0.009%


TECHNICAL- 10kHz Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

56 ELECTRO- AC Voltage@10 Hz to Using Fluke 8588A By 10 mV to 100 mV 0.04% to 0.009%


TECHNICAL- 10kHz Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

57 ELECTRO- AC Voltage@10 Hz to Using Fluke 8588A By 100 mV to 1 V 0.009% to 0.008%


TECHNICAL- 10kHz Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

58 ELECTRO- AC Voltage@10 Hz to Using Fluke 8588A By 100 V to 1000 V 0.009%


TECHNICAL- 10kHz Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 11 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
59 ELECTRO- AC VoltageAt 10 KHz Using Fluke 8588A by 100 V to 1000 V 0.01%
TECHNICAL- to 100 KHz Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

60 ELECTRO- DC Capacitance Using Fluke 8588A By 1 µF to 10 uF 0.08%


TECHNICAL- Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

61 ELECTRO- DC Capacitance Using Fluke 8588A By 1 mF to 100 mF 0.08% to 0.09%


TECHNICAL- Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

62 ELECTRO- DC Capacitance Using Fluke 8588A By 1 nF to 10 nF 0.2% to 0.05%


TECHNICAL- Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

63 ELECTRO- DC Capacitance Using Fluke 8588A By 10 µF to 1 mF 0.08%


TECHNICAL- Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 12 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
64 ELECTRO- DC Capacitance Using Fluke 8588A By 10 nF to 100 nF 0.05%
TECHNICAL- Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

65 ELECTRO- DC Capacitance Using Fluke 8588A By 100 nF to 1 µF 0.05%


TECHNICAL- Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

66 ELECTRO- DC Current Using Fluke 8588A By 1 mA to 10 mA 0.05% to 0.005%


TECHNICAL- Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

67 ELECTRO- DC Current Using Fluke 8588A By 200 mA to 1 A 0.004% to 0.013%


TECHNICAL- Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

68 ELECTRO- DC Power (1phase)10 Using Power Meter 1 W to 20 KW 0.03% to 0.06%


TECHNICAL- V to 1000 V(0.1 A to 20 AC/DC By Direct
ELECTRICAL A) Method
EQUIPMENT
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 13 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
69 ELECTRO- DC Resistance 2 Wire Using Fluke 8588A By 10 M ohm to 100 M 0.001% to 0.008%
TECHNICAL- Direct/Simulation ohm
ELECTRICAL Method
EQUIPMENT
(Measure)

70 ELECTRO- DC Resistance 2 Wire Using Fluke 8588A By 100 ohm to 1 k ohm 0.0009% to 0.0008%
TECHNICAL- Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

71 ELECTRO- DC Voltage Using Fluke 8588A By 1 mV to 100 mV 0.001% to 0.0002%


TECHNICAL- Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

72 ELECTRO- DC Voltage Using Fluke 8588A By 1 V to 10 V 0.0003% to 0.00009%


TECHNICAL- Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

73 ELECTRO- DC Voltage Using Fluke 8588A By 10 V to 100 V 0.00009% to 0.0005%


TECHNICAL- Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 14 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
74 ELECTRO- DC Voltage Using Fluke 8588A By 100 mV to 1 V 0.0002% to 0.0003%
TECHNICAL- Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

75 ELECTRO- DC Voltage Using Fluke 8588A By 100 V to 1000 V 0.0005 to 0.0001


TECHNICAL- Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

76 ELECTRO- Resistance 2 wireDC Using Fluke 8588A By 1 G ohm to 10 G ohm 0.5%


TECHNICAL- Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

77 ELECTRO- Resistance 2 wireDC Using Fluke 8588A By 1 M ohm to 10 M ohm 0.001%


TECHNICAL- Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

78 ELECTRO- Resistance 2 wireDC Using Fluke 8588A By 1 ohm to 10 ohm 0.001%


TECHNICAL- Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 15 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
79 ELECTRO- Resistance 2 wireDC Using Fluke 8588A By 10 ohm to 100 ohm 0.001% to 0.0009%
TECHNICAL- Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

80 ELECTRO- Resistance 2 wireDC Using Fluke 8588A By 100 k ohm to 1 M ohm 0.0008% to 0.0009%
TECHNICAL- Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

81 ELECTRO- Resistance 2 wireDC Using Fluke 8588A By 100 M ohm to 1 G ohm 0.008% to 0.05%
TECHNICAL- Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

82 ELECTRO- Resistance 2 wireDC Using Fluke 8588A By 100 ohm to 100 k ohm 0.0009% to 0.0008%
TECHNICAL- Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

83 ELECTRO- Resistance 4 wireDC Using Fluke 8588A By 1 ohm to 10 ohm 0.001%


TECHNICAL- DIrect Method
ELECTRICAL
EQUIPMENT
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 16 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
84 ELECTRO- Resistance 4 wireDC Using Fluke 8588A By 10 ohm to 100 ohm 0.001% to 0.0009%
TECHNICAL- Direct MEthod
ELECTRICAL
EQUIPMENT
(Measure)

85 ELECTRO- Resistance 4 wireDC Using Fluke 8588A by 100 ohm to 1 k ohm 0.0009% to 0.0008%
TECHNICAL- Direct Method
ELECTRICAL
EQUIPMENT
(Measure)

86 ELECTRO- AC Using Fluke 5502 10.9 µF to 30 µF 0.2% to 0.4%


TECHNICAL- Capacitance@100Hz Calibrator by Direct
ELECTRICAL Method
EQUIPMENT (Source)

87 ELECTRO- AC Using Fluke 5502 330 nF to 10.9 µF 0.1% to 0.2%


TECHNICAL- Capacitance@100Hz Calibrator by Direct
ELECTRICAL Method
EQUIPMENT (Source)

88 ELECTRO- AC Capacitance@1kHz Using Fluke 5502 0.22 nF to 11 nF 0.9% to 0.2%


TECHNICAL- Calibrator by Direct
ELECTRICAL Method
EQUIPMENT (Source)

89 ELECTRO- AC Capacitance@1kHz Using Fluke 5502 11 nF to 330 nF 0.2%


TECHNICAL- Calibrator by Direct
ELECTRICAL Method
EQUIPMENT (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 17 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
90 ELECTRO- AC Using Fluke 5502 30 µF to 109 µF 0.4% to 0.5%
TECHNICAL- Capacitance@50Hz Calibrator by Direct
ELECTRICAL Method
EQUIPMENT (Source)

91 ELECTRO- AC Voltage@100kHz Using Fluke Calibrator 0.33 V to 3.3 V 0.05% to 0.06%


TECHNICAL- to 500MHz 5502A By Direct
ELECTRICAL Method
EQUIPMENT (Source)

92 ELECTRO- AC Voltage@100kHz Using Fluke Calibrator 1 mV to 30 mV 0.19% to 0.17%


TECHNICAL- to 500MHz 5502A By Direct
ELECTRICAL Method
EQUIPMENT (Source)

93 ELECTRO- AC Voltage@100kHz Using Fluke Calibrator 30 mV to 300 mV 0.05%


TECHNICAL- to 500MHz 5502A By Direct
ELECTRICAL Method
EQUIPMENT (Source)

94 ELECTRO- AC Voltage@10Hz to Using Fluke Calibrator 0.33 V to 3 V 0.060% to 0.059%


TECHNICAL- 45Hz 5502A By Direct
ELECTRICAL Method
EQUIPMENT (Source)

95 ELECTRO- AC Voltage@10Hz to Using Fluke Calibrator 1 mV to 30 mV 0.17% to 0.19%


TECHNICAL- 45Hz 5502A By Direct
ELECTRICAL Method
EQUIPMENT (Source)

96 ELECTRO- AC Voltage@10Hz to Using Fluke Calibrator 3 V to 3.3 V 0.059% to 0.086%


TECHNICAL- 45Hz 5502A By Direct
ELECTRICAL Method
EQUIPMENT (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 18 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
97 ELECTRO- AC Voltage@10Hz to Using Fluke Calibrator 3.3 V to 30 V 0.086% to 0.056%
TECHNICAL- 45Hz 5502A By Direct
ELECTRICAL Method
EQUIPMENT (Source)

98 ELECTRO- AC Voltage@10Hz to Using Fluke Calibrator 30 mV to 33 mV 0.059% to 0.17%


TECHNICAL- 45Hz 5502A By Direct
ELECTRICAL Method
EQUIPMENT (Source)

99 ELECTRO- AC Voltage@10Hz to Using Fluke Calibrator 33 mV to 330 mV 0.059% to 0.058%


TECHNICAL- 45Hz 5502A By Direct
ELECTRICAL Method
EQUIPMENT (Source)

100 ELECTRO- AC Voltage@10Hz to Using Fluke Calibrator 330 mV to 3.3 V 0.058% to 0.057%
TECHNICAL- 45Hz 5502A By Direct
ELECTRICAL Method
EQUIPMENT (Source)

101 ELECTRO- AC Voltage@10kHz to Using Fluke Calibrator 0.33 V to 3.3 V 0.08% to 0.07%
TECHNICAL- 100kHz 5502A By Direct
ELECTRICAL Method
EQUIPMENT (Source)

102 ELECTRO- AC Voltage@10kHz to Using Fluke Calibrator 1 mV to 30 mV 0.1% to 0.4%


TECHNICAL- 100kHz 5502A By Direct
ELECTRICAL Method
EQUIPMENT (Source)

103 ELECTRO- AC Voltage@10kHz to Using Fluke Calibrator 3.3 V to 33 V 0.08%


TECHNICAL- 100kHz 5502A By Direct
ELECTRICAL Method
EQUIPMENT (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 19 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
104 ELECTRO- AC Voltage@10kHz to Using Fluke Calibrator 30 mV to 330 mV 0.1% to 0.8%
TECHNICAL- 100kHz 5502A By Direct
ELECTRICAL Method
EQUIPMENT (Source)

105 ELECTRO- AC Voltage@10kHz to Using Fluke Calibrator 33 V to 330 V 0.08% to 0.06%


TECHNICAL- 100kHz 5502A By Direct
ELECTRICAL Method
EQUIPMENT (Source)

106 ELECTRO- AC Voltage@45Hz to Using Fluke Calibrator 1 mV to 30 mV 0.11% to 0.14%


TECHNICAL- 10kHz 5502A By Direct
ELECTRICAL Method
EQUIPMENT (Source)

107 ELECTRO- AC Voltage@45Hz to Using Fluke Calibrator 3.3 V to 33 V 0.05% to 0.05%


TECHNICAL- 10kHz 5502A By Direct
ELECTRICAL Method
EQUIPMENT (Source)

108 ELECTRO- AC Voltage@45Hz to Using Fluke Calibrator 30 mV to 3.3 V 0.03% to 0.11%


TECHNICAL- 10kHz 5502A By Direct
ELECTRICAL Method
EQUIPMENT (Source)

109 ELECTRO- AC Voltage@45Hz to Using Fluke Calibrator 33 V to 330 V 0.056% to 0.05%


TECHNICAL- 10kHz 5502A By Direct
ELECTRICAL Method
EQUIPMENT (Source)

110 ELECTRO- AC Voltage@45Hz to Using Fluke Calibrator 330 V to 1000 V 0.056% to 0.05%
TECHNICAL- 10kHz 5502A By Direct
ELECTRICAL Method
EQUIPMENT (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 20 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
111 ELECTRO- DC Capacitance Using Fluke 5502A By 1.1 mF to 3.3 mF 0.1% to 0.2%
TECHNICAL- Direct Method
ELECTRICAL
EQUIPMENT (Source)

112 ELECTRO- DC Capacitance Using Fluke 5502 A by 200 µF to 330 µF 0.9% to 0.2%
TECHNICAL- Direct Method
ELECTRICAL
EQUIPMENT (Source)

113 ELECTRO- DC Capacitance Using Fluke 5502A By 3.3 mF to 33 mF 0.2% to 0.4%


TECHNICAL- Direct Method
ELECTRICAL
EQUIPMENT (Source)

114 ELECTRO- DC Capacitance Using Fluke 5502A By 33 mF to 110 mF 0.4% to 0.1%


TECHNICAL- Direct Method
ELECTRICAL
EQUIPMENT (Source)

115 ELECTRO- DC Capacitance Using Fluke 5502A By 330 µF to 1.1 mF 0.2% to 0.1%
TECHNICAL- Direct Method
ELECTRICAL
EQUIPMENT (Source)

116 ELECTRO- DC High Resistance 2 Using HV mega ohm 1 G ohm 4.6%


TECHNICAL- wire@0-5000 VDC Box by Direct Method
ELECTRICAL
EQUIPMENT (Source)

117 ELECTRO- DC High Resistance 2 Using HV mega ohm 1 T ohm 2.3%


TECHNICAL- wire@0-5000 VDC Box by Direct Method
ELECTRICAL
EQUIPMENT (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 21 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
118 ELECTRO- DC High Resistance 2 Using HV mega ohm 10 G ohm 2.3%
TECHNICAL- wire@0-5000 VDC Box by Direct Method
ELECTRICAL
EQUIPMENT (Source)

119 ELECTRO- DC High Resistance 2 Using HV mega ohm 100 G ohm 2.3%
TECHNICAL- wire@0-5000 VDC Box by Direct Method
ELECTRICAL
EQUIPMENT (Source)

120 ELECTRO- DC High Resistance 2 Using HV mega ohm 2 G ohm 4.0%


TECHNICAL- wire@0-5000 VDC Box by Direct Method
ELECTRICAL
EQUIPMENT (Source)

121 ELECTRO- DC High Resistance 2 Using HV mega ohm 20 G Ohm 3.6%


TECHNICAL- wire@0-5000 VDC Box by Direct Method
ELECTRICAL
EQUIPMENT (Source)

122 ELECTRO- DC High Resistance 2 Using HV mega ohm 20 M ohm 3.6%


TECHNICAL- wire@0-5000 VDC Box by Direct Method
ELECTRICAL
EQUIPMENT (Source)

123 ELECTRO- DC High Resistance 2 Using HV mega ohm 200 G ohm 2.3%
TECHNICAL- wire@0-5000 VDC Box by Direct Method
ELECTRICAL
EQUIPMENT (Source)

124 ELECTRO- DC High Resistance 2 Using HV mega ohm 200 M ohm 3.6%
TECHNICAL- wire@0-5000 VDC Box by Direct Method
ELECTRICAL
EQUIPMENT (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 22 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
125 ELECTRO- DC High Resistance 2 Using HV mega ohm 500 G ohm 2.3%
TECHNICAL- wire@0-5000 VDC Box by Direct Method
ELECTRICAL
EQUIPMENT (Source)

126 ELECTRO- DC High Resistance 2 Using HV Mega ohm 2 M ohm 3.6%


TECHNICAL- wire@0-5000VDC Box by Direct Method
ELECTRICAL
EQUIPMENT (Source)

127 ELECTRO- DC Power (1phase)1 V Using Fluke 5502A 0.1 W to 20 KW 0.7% to 0.1%
TECHNICAL- to 1000 V0.1 A to 20 A Calibrator by Direct
ELECTRICAL method
EQUIPMENT (Source)

128 ELECTRO- DC Resistance 2 wire Using Fluke 5502 1 M ohm to 3 M ohm 0.01% to 0.04%
TECHNICAL- Calibrator by Direct
ELECTRICAL Method
EQUIPMENT (Source)

129 ELECTRO- DC Resistance 2 wire Using Fluke 5502 1 ohm to 1 M ohm 0.01% to 0.01%
TECHNICAL- Calibrator by Direct
ELECTRICAL Method
EQUIPMENT (Source)

130 ELECTRO- DC resistance 2 wire Using Fluke 5502 10 M ohm to 30 M ohm 0.07% to 0.09%
TECHNICAL- Calibrator by Direct
ELECTRICAL Method
EQUIPMENT (Source)

131 ELECTRO- DC resistance 2 wire Using Fluke 5502 100 M ohm to 300 M 0.1% to 0.5%
TECHNICAL- Calibrator by Direct ohm
ELECTRICAL Method
EQUIPMENT (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 23 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
132 ELECTRO- DC resistance 2 wire Using Fluke 5502 3 M ohm to 10 M ohm 0.04% to 0.07%
TECHNICAL- Calibrator by Direct
ELECTRICAL Method
EQUIPMENT (Source)

133 ELECTRO- DC resistance 2 wire Using Fluke 5502 30 M ohm to 100 M 0.09% to 0.1%
TECHNICAL- Calibrator by Direct ohm
ELECTRICAL Method
EQUIPMENT (Source)

134 ELECTRO- DC resistance 2 wire Using Fluke 5502 300 M ohm to 1100 M 0.5% to 1.7%
TECHNICAL- Calibrator by Direct ohm
ELECTRICAL Method
EQUIPMENT (Source)

135 ELECTRO- DC Resistance 4 wire Using Standard 1 K ohm 3.5%


TECHNICAL- Resistance Box By
ELECTRICAL Direct Method
EQUIPMENT (Source)

136 ELECTRO- DC Resistance 4 wire Using Standard 1 m ohm 3.5%


TECHNICAL- Resistance Box By
ELECTRICAL Direct Method
EQUIPMENT (Source)

137 ELECTRO- DC Resistance 4 wire Using Standard 1 ohm 3.5%


TECHNICAL- Resistance Box By
ELECTRICAL Direct Method
EQUIPMENT (Source)

138 ELECTRO- DC Resistance 4 wire Using Fluke 5502 1 ohm to 1 K ohm 0.01%
TECHNICAL- Calibrator by Direct
ELECTRICAL Method
EQUIPMENT (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 24 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
139 ELECTRO- DC Resistance 4 wire Using Standard 10 m ohm 3.5%
TECHNICAL- Resistance Box By
ELECTRICAL Direct Method
EQUIPMENT (Source)

140 ELECTRO- DC Resistance 4 wire Using Standard 10 ohm 3.5%


TECHNICAL- Resistance Box By
ELECTRICAL Direct Method
EQUIPMENT (Source)

141 ELECTRO- DC Resistance 4 wire Using Standard 100 µ ohm 0.02%


TECHNICAL- Resistance Box By
ELECTRICAL Direct Method
EQUIPMENT (Source)

142 ELECTRO- DC Resistance 4 Wire Using Standard 100 m ohm 3.5%


TECHNICAL- Resistance Box By
ELECTRICAL Direct Method
EQUIPMENT (Source)

143 ELECTRO- DC Resistance 4 wire Using Standard 100 ohm 3.5%


TECHNICAL- Resistance Box By
ELECTRICAL Direct Method
EQUIPMENT (Source)

144 ELECTRO- DC Resistance 4 wire Using Standard 50 µ ohm 0.03%


TECHNICAL- Resistance Box By
ELECTRICAL Direct Method
EQUIPMENT (Source)

145 ELECTRO- DC Voltage Using Fluke 5502 A 1 mV to 330 mV 0.0071% to 0.0069%


TECHNICAL- calibrator By direct
ELECTRICAL Method
EQUIPMENT (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 25 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
146 ELECTRO- DC Voltage Using Fluke Calibrator 3.3 V to 33 V 0.0060%
TECHNICAL- 5502A By Direct
ELECTRICAL Method
EQUIPMENT (Source)

147 ELECTRO- DC Voltage Using Fluke Calibrator 33 V to 330 V 0.0060% to 0.0066%


TECHNICAL- 5502A By Direct
ELECTRICAL Method
EQUIPMENT (Source)

148 ELECTRO- DC Voltage Using Fluke Calibrator 330 mV to 3.3 V 0.0069% to 0.0060%
TECHNICAL- 5502A By Direct
ELECTRICAL Method
EQUIPMENT (Source)

149 ELECTRO- DC Voltage Using Fluke Calibrator 330 V to 1000 V 0.0066% to 0.0065%
TECHNICAL- 5502A By Direct
ELECTRICAL Method
EQUIPMENT (Source)

150 ELECTRO- Inductance Using Standard 100 µH to 10 H 1.22%


TECHNICAL- Inductance Box by
ELECTRICAL Direct Method
EQUIPMENT (Source)

151 ELECTRO- Oscilloscope Using Fluke 5502A, 0.1 mV to 2.49 mV 5.5% to 0.46%
TECHNICAL- Vertical(Amplitude) SC600 Scope Option
ELECTRICAL by Direct Method
EQUIPMENT (Source)

152 ELECTRO- Oscilloscope Using Fluke 5502A, 11 V to 130 V 0.08% to 0.07%


TECHNICAL- Vertical(Amplitude) SC600 Scope Option
ELECTRICAL by Direct Method
EQUIPMENT (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 26 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
153 ELECTRO- Oscilloscope Using Fluke 5502A, 2.2 V to 11 V 0.06% to 0.08%
TECHNICAL- Vertical(Amplitude) SC600 Scope Option
ELECTRICAL by Direct Method
EQUIPMENT (Source)

154 ELECTRO- Oscilloscope Using Fluke 5502A, 2.49 mV to 500 mV 0.46% to 0.05%
TECHNICAL- Vertical(Amplitude) SC600 Scope Option
ELECTRICAL by Direct Method
EQUIPMENT (Source)

155 ELECTRO- Oscilloscope Using Fluke 5502A, 500 mV to 2.2 V 0.05% to 0.06%
TECHNICAL- Vertical(Amplitude) SC600 Scope Option
ELECTRICAL by Direct Method
EQUIPMENT (Source)

156 ELECTRO- Oscilloscope Using Fluke 5502A, 1.8 mV to 90 mV 1.9% to 0.39%


TECHNICAL- Vertical(Amplitude)Squ SC600 Scope Option
ELECTRICAL are Wave by Direct Method
EQUIPMENT (Source)

157 ELECTRO- Oscilloscope Using Fluke 5502A, 30 V to 55 V 0.06% to 0.03%


TECHNICAL- Vertical(Amplitude)Squ SC600 Scope Option
ELECTRICAL are Wave by Direct Method
EQUIPMENT (Source)

158 ELECTRO- Oscilloscope Using Fluke 5502A, 90 mV to 900 mV 0.39% to 0.05%


TECHNICAL- Vertical(Amplitude)Squ SC600 Scope Option
ELECTRICAL are Wave by Direct Method
EQUIPMENT (Source)

159 ELECTRO- Oscilloscope Using Fluke 5502A, 900 mV to 30 V 0.05% to 0.06%


TECHNICAL- Vertical(Amplitude)Squ SC600 Scope Option
ELECTRICAL are Wave by Direct Method
EQUIPMENT (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 27 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
160 ELECTRO- OscilloscopeBandwidth Using Fluke 5502A, 1 MHz to 600 MHz 3.9% to 4.1%
TECHNICAL- (Leveling) SC600 Scope Option (5mV to 5.5V)
ELECTRICAL By Direct Method
EQUIPMENT (Source)

161 ELECTRO- OscilloscopeHorizontal Using Fluke 5502A, 100 ns to 20 ms 0.03%


TECHNICAL- (Time Based) SC600 Scope Option
ELECTRICAL By Direct Method
EQUIPMENT (Source)

162 ELECTRO- OscilloscopeHorizontal Using Fluke 5502A, 2 ns to 50 ns 0.05% to 0.02%


TECHNICAL- (Time Based) SC600 Scope Option
ELECTRICAL By Direct Method
EQUIPMENT (Source)

163 ELECTRO- OscilloscopeHorizontal Using Fluke 5502A, 20 ms to 5 s 0.03% to 0.02%


TECHNICAL- (Time Based) SC600 Scope Option
ELECTRICAL By Direct Method
EQUIPMENT (Source)

164 ELECTRO- OscilloscopeHorizontal Using Fluke 5502A, 50 ns to 100 ns 0.02% to 0.03%


TECHNICAL- (Time Based) SC600 Scope Option
ELECTRICAL By Direct Method
EQUIPMENT (Source)

165 ELECTRO- Power Active Only45 Using Fluke 5502 0.4 W to 12.0 KW 0.09% to 0.11%
TECHNICAL- Hz to 50Hz(-)0.1 PF to Calibrator By Direct
ELECTRICAL (-)0.9 PF0.1 PF to UPF Method
EQUIPMENT (Source) (1phase & 3phase)40V
to 600 V0.1 A to 20 A

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 28 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
166 ELECTRO- Power Factor/Phase By Fluke 5502A 0.1 pF to 1.0 pF 0.0029pF
TECHNICAL- Angle50Hz , 240 Calibrator by Direct
ELECTRICAL Volt(Lead & Lag) Method
EQUIPMENT (Source)

167 ELECTRO- Resistance 4 W Using Standard 1 m ohm to 1 k ohm 3.71% to 1.75%


TECHNICAL- Resistance Box by
ELECTRICAL Direct Metho
EQUIPMENT (Source)

168 ELECTRO- Temperature Using Fluke 8588A By 600 °C to 1800 °C 0.06°C


TECHNICAL- SimulationTemp Direct/Simulation
TEMPERATURE Indicator/Controller/Dat Method
SIMULATION a Logger/RecordsB -
(Measure) Type

169 ELECTRO- Temperature Using Fluke 8588A By (-)250 °C to 1000 °C 0.06°C


TECHNICAL- SimulationTemp Direct/Simulation
TEMPERATURE Indicator/Controller/Dat Method
SIMULATION a Logger/RecordsE -
(Measure) Type

170 ELECTRO- Temperature Using Fluke 8588A By (-)200 °C to 1200 °C 0.03°C


TECHNICAL- SimulationTemp Direct/Simulation
TEMPERATURE Indicator/Controller/Dat Method
SIMULATION a Logger/RecordsJ-
(Measure) Type

171 ELECTRO- Temperature Using Fluke 8588A By (-)200 °C to 1200 °C 0.03°C


TECHNICAL- SimulationTemp Direct/Simulation
TEMPERATURE Indicator/Controller/Dat Method
SIMULATION a Logger/RecordsK -
(Measure) Type

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 29 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
172 ELECTRO- Temperature Using Fluke 8588A By (-)200 °C to 1300 °C 0.03°C
TECHNICAL- SimulationTemp Direct/Simulation
TEMPERATURE Indicator/Controller/Dat Method
SIMULATION a Logger/RecordsN -
(Measure) Type

173 ELECTRO- Temperature Using Fluke 8588A By 0 °C to 1750 °C 0.03°C


TECHNICAL- SimulationTemp Direct/Simulation
TEMPERATURE Indicator/Controller/Dat Method
SIMULATION a Logger/RecordsR -
(Measure) Type

174 ELECTRO- Temperature Using Fluke 8588A By (-)200 °C to 800 °C 0.061°C


TECHNICAL- SimulationTemp Direct/Simulation
TEMPERATURE Indicator/Controller/Dat Method
SIMULATION a Logger/RecordsRTD
(Measure) (PRT-100)

175 ELECTRO- Temperature Using Fluke 8588A By 0 °C to 1750 °C 0.02°C


TECHNICAL- SimulationTemp Direct/Simulation
TEMPERATURE Indicator/Controller/Dat Method
SIMULATION a Logger/RecordsS -
(Measure) Type

176 ELECTRO- Temperature Using Fluke 8588A By (-)250 °C to 400 °C 0.04°C


TECHNICAL- SimulationTemp Direct/Simulation
TEMPERATURE Indicator/Controller/Dat Method
SIMULATION a Logger/RecordsT -
(Measure) Type

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 30 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
177 ELECTRO- Temperature Using Fluke 5502A 600 °C to 1800 °C 0.5°C
TECHNICAL- SimulationTemp Calibrator by Direct
TEMPERATURE Indicator/Controller/Dat method
SIMULATION (Source) a
logger/Recorder/Calibr
atorB- Type

178 ELECTRO- Temperature Using Fluke 5502 Fluke (-) 250 °C to 1000 °C 0.5°C
TECHNICAL- SimulationTemp Calibrator By direct
TEMPERATURE Indicator/Controller/Dat Method
SIMULATION (Source) a
logger/Recorder/Calibr
atorE- Type

179 ELECTRO- Temperature Using Fluke Calibrator (-) 200 °C to 1200 °C 0.2°C
TECHNICAL- SimulationTemp 5502A by Direct
TEMPERATURE Indicator/Controller/Dat Method
SIMULATION (Source) a
logger/Recorder/Calibr
atorJ-Type

180 ELECTRO- Temperature Using Fluke 5502 (-) 200 °C to 1200 °C 0.4°C
TECHNICAL- SimulationTemp Calibrator By Direct
TEMPERATURE Indicator/Controller/Dat Method
SIMULATION (Source) a
logger/Recorder/Calibr
atorK-Type

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 31 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
181 ELECTRO- Temperature Using Fluke 5502 A (-) 200 °C to 1300 °C 0.4°C
TECHNICAL- SimulationTemp calibrator By direct
TEMPERATURE Indicator/Controller/Dat Method
SIMULATION (Source) a
logger/Recorder/Calibr
atorN- Type

182 ELECTRO- Temperature Using Fluke 5502 0 °C to 1750 °C 0.5°C


TECHNICAL- SimulationTemp Calibrator By Direct
TEMPERATURE Indicator/Controller/Dat method
SIMULATION (Source) a
logger/Recorder/Calibr
atorR Type

183 ELECTRO- Temperature Using Fluke 5502A (-) 200 °C to 800 0.09°C
TECHNICAL- SimulationTemp Calibrator By Direct
TEMPERATURE Indicator/Controller/Dat Method
SIMULATION (Source) a
logger/Recorder/Calibr
atorRTD (PRT-100)

184 ELECTRO- Temperature Using Fluke 5502A 0 °C to 1750 °C 0.5°C


TECHNICAL- SimulationTemp calibrator By Direct
TEMPERATURE Indicator/Controller/Dat method
SIMULATION (Source) a
logger/Recorder/Calibr
atorS Type

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 32 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
185 ELECTRO- Temperature Using Fluke 5502 (-) 250 °C to 400 °C 0.7°C
TECHNICAL- SimulationTemp calibrator By Direct
TEMPERATURE Indicator/Controller/Dat Method
SIMULATION (Source) a
logger/Recorder/Calibr
atorT Type

186 ELECTRO- Frequency Using Fluke 8588A By 1 kHz to 10 kHz 0.005% to 0.0006%
TECHNICAL- TIME & Direct Method
FREQUENCY
(Measure)

187 ELECTRO- Frequency Using Fluke 8588A By 1 MHz to 10 MHz 0.005%


TECHNICAL- TIME & Direct Method
FREQUENCY
(Measure)

188 ELECTRO- Frequency Using Fluke 8588A By 10 Hz to 1 kHz 0.004% to 0.005%


TECHNICAL- TIME & Direct Method
FREQUENCY
(Measure)

189 ELECTRO- Frequency Using Fluke 8588A By 10 kHz to 100 kHz 0.0006%
TECHNICAL- TIME & Direct Method
FREQUENCY
(Measure)

190 ELECTRO- Frequency Using Fluke 8588A By 10 MHz to 100 MHz 0.005% to 0.0006%
TECHNICAL- TIME & Direct Method
FREQUENCY
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 33 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
191 ELECTRO- Frequency Using Fluke 8588A By 100 kHz to 1 MHz 0.0006% to 0.005%
TECHNICAL- TIME & Direct Method
FREQUENCY
(Measure)

192 ELECTRO- Time Using Time Calibrator 0.1 sec. to 60 sec. 0.02sec. to 0.04sec.
TECHNICAL- TIME & by Direct Method
FREQUENCY
(Measure)

193 ELECTRO- Time Using Time Calibrator 18000 Sec to 86400 7.0Sec to 20sec
TECHNICAL- TIME & by comparision method Sec
FREQUENCY
(Measure)

194 ELECTRO- Time Using Time Calibrator 3600 sec to 18000 sec 1.2sec. to 7.0sec.
TECHNICAL- TIME & by Direct Method
FREQUENCY
(Measure)

195 ELECTRO- Time Using Time Calibrator 60 sec. to 3600 Sec 0.04sec. to 1.2sec.
TECHNICAL- TIME & by Direct Method
FREQUENCY
(Measure)

196 ELECTRO- Frequency@3V Using Fluke 5502 1 kHz to 100 kHz 0.005% to 0.009%
TECHNICAL- TIME & Calibrator by Direct
FREQUENCY (Source) Method

197 ELECTRO- Frequency@3V Using Fluke 5502 10 0.002%


TECHNICAL- TIME & Calibrator by Direct Hz
FREQUENCY (Source) Method to 119 Hz

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 34 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
198 ELECTRO- Frequency@3V Using Fluke 5502 100 kHz to 1000 kHz 0.009% to 0.004%
TECHNICAL- TIME & Calibrator by Direct
FREQUENCY (Source) Method

199 ELECTRO- Frequency@3V Using Fluke 5502 1000 kHz to 2 MHz 0.004% to 0.009%
TECHNICAL- TIME & Calibrator by Direct
FREQUENCY (Source) Method

200 ELECTRO- Frequency@3V Using Fluke 5502 119 Hz to 1 kHz 0.002% to 0.005%
TECHNICAL- TIME & Calibrator by Direct
FREQUENCY (Source) Method

201 MECHANICAL- Tachometer Using Digital 10 rpm to 1000 rpm 5.3%


ACCELERATION AND Calibrator/Stroboscope/ Tacometer (Non
SPEED RPM Meter/Centrifuge Contact) by
Etc comparision method

202 MECHANICAL- Tachometer Using Digital 1000 rpm to 99900 rpm 1.2%
ACCELERATION AND Calibrator/Stroboscope/ Tacometer (non
SPEED RPM Meter/Centrifuge Contact) by
Etc comparision method

203 MECHANICAL- TachometerContact Using Tachometer 10 rpm to 5000 rpm 6.0% rdg to 2.5% rdg
ACCELERATION AND type Calibrator contact and
SPEED non contact type by
comparision method

204 MECHANICAL- TachometerNon - Using tachometer 10 rpm to 99900 rpm 5.42% rdg to 2.5% rdg
ACCELERATION AND Contact Type calibrator contact and
SPEED non contact type by
comparision method

205 MECHANICAL- Sound Level Meter Using Sound Level 94 to 114 dB 0.7dB
ACOUSTICS Calibrator by
Comparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 35 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
206 MECHANICAL- Bevel ProtectorAngle Using Angle Gauge 0 to 90 to 0 ° 4min.
DIMENSION (BASIC ProtectorCombination Set. by Comparison
MEASURING Set. Method
INSTRUMENT,
GAUGE ETC.)

207 MECHANICAL- Bore Gauge Travel Using Universal Length up to 1 mm 1.5µm


DIMENSION (BASIC only( 0-1 mm) L.C.:- Measuring Machine by
MEASURING 0.001 mm Comparison Method
INSTRUMENT,
GAUGE ETC.)

208 MECHANICAL- Caliper ( Dial / Digital / Using Slip Gauge Set & up to 1000 mm 12µm
DIMENSION (BASIC Vernier Caliper ) L.C. :- Steel Gauge Block with
MEASURING 0.01 mm Accessories by
INSTRUMENT, Comparison Method
GAUGE ETC.)

209 MECHANICAL- Caliper ( Dial / Digital / Using Slip Gauge Set. up to 300 mm 7µm
DIMENSION (BASIC Vernier Caliper ) L.C. :- with Accessories by
MEASURING 0.01 mm Comparison Method
INSTRUMENT,
GAUGE ETC.)

210 MECHANICAL- Caliper ( Dial / Digital / Using Slip Gauge Set.& up to 600 mm 10.0µm
DIMENSION (BASIC Vernier Caliper ) L.C. :- Steel Gauge Block with
MEASURING 0.01 mm Accessories by
INSTRUMENT, Comparison Method
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 36 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
211 MECHANICAL- Caliper Checker Using Slip Gauge Set , up to 600 mm 8.2µm
DIMENSION (BASIC Long Gauge Block &
MEASURING Dial Test Indicator by
INSTRUMENT, Comparison Method
GAUGE ETC.)

212 MECHANICAL- Coating Thickness Using Standard Foils up to 700 µm 4µm


DIMENSION (BASIC Gauge by Comparison Method
MEASURING
INSTRUMENT,
GAUGE ETC.)

213 MECHANICAL- Comparator Stand # Using Electronic Level up to 300X300 mm 6µm


DIMENSION (BASIC Flatness by Comparison Method
MEASURING
INSTRUMENT,
GAUGE ETC.)

214 MECHANICAL- Cube Mould Using Digimatic Caliper 300 x to 300 mm 90.5µm
DIMENSION (BASIC by comparision method
MEASURING
INSTRUMENT,
GAUGE ETC.)

215 MECHANICAL- Depth Caliper L.C.:- Using Slip Gauge Set. up to 300 mm 11.2µm
DIMENSION (BASIC 0.01 mm by Comparison Method
MEASURING
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 37 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
216 MECHANICAL- Depth Micrometer Using Slip Gauge Set & up to 300 mm 3µm
DIMENSION (BASIC L.C.:-0.001mm Steel Gauge Block by
MEASURING Comparison Method
INSTRUMENT,
GAUGE ETC.)

217 MECHANICAL- Dial Gauge ( Lever Using Universal Length up to 2 mm 1.5µm


DIMENSION (BASIC Type ) L.C.:- 0.001 mm Measuring Machine by
MEASURING Comparison Method
INSTRUMENT,
GAUGE ETC.)

218 MECHANICAL- Dial Gauge ( Plunger Using Universal Length up to 100 mm 1.5µm
DIMENSION (BASIC Type )L.C. 0.001 mm Measuring Machine by
MEASURING Comparison Method
INSTRUMENT,
GAUGE ETC.)

219 MECHANICAL- Dig./Dial Thickness Using Slip Gauge Set up to 50 mm 1.0µm


DIMENSION (BASIC TesterL.C. :- 0.001 mm by Comparison Method
MEASURING
INSTRUMENT,
GAUGE ETC.)

220 MECHANICAL- Ext. Micrometer , L.C. Using Slip Gauge Set & 100 mm to 300 mm 2.5µm
DIMENSION (BASIC :- 0.001mm Steel Gauge Block by
MEASURING Comparison Method
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 38 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
221 MECHANICAL- Ext. Micrometer , L.C. Using Slip Gauge Set & 300 mm to 500 mm 9µm
DIMENSION (BASIC :- 0.001mm Steel Gauge Block by
MEASURING Comparison Method
INSTRUMENT,
GAUGE ETC.)

222 MECHANICAL- Ext. Micrometer , L.C. Using Slip Gauge Set. up to 100 mm 1.6µm
DIMENSION (BASIC :- 0.001mm by Comparison Method
MEASURING
INSTRUMENT,
GAUGE ETC.)

223 MECHANICAL- Ext. Micrometer , L.C. Using Slip Gauge Set & 500 mm to 1000 mm 14µm
DIMENSION (BASIC :- 0.01mm Steel Gauge Block by
MEASURING Comparison Method
INSTRUMENT,
GAUGE ETC.)

224 MECHANICAL- Feeler Gauge Using Universal Length up to 1 mm 1µm


DIMENSION (BASIC Measuring Machine by
MEASURING Comparison Method
INSTRUMENT,
GAUGE ETC.)

225 MECHANICAL- Height GaugeL.C. :- Using Using Slip up to 1000 mm 15µm


DIMENSION (BASIC 0.01 mm Gauge Set , Steel
MEASURING Gauge Block ,Dial Test
INSTRUMENT, Indicator with
GAUGE ETC.) Accessories by
Comparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 39 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
226 MECHANICAL- Height GaugeL.C. :- Using Slip Gauge Set , up to 300 mm 8µm
DIMENSION (BASIC 0.01 mm Long Slip Block & Dial
MEASURING Test Indicator With
INSTRUMENT, Acc. by Comparison
GAUGE ETC.) Method

227 MECHANICAL- Height GaugeL.C. :- Using Using Slip up to 600 mm 12µm


DIMENSION (BASIC 0.01 mm Gauge Set , Steel
MEASURING Gauge Block ,Dial Test
INSTRUMENT, Indicator with
GAUGE ETC.) Accessories by
Comparison Method

228 MECHANICAL- Height GuugeL.C. :- Using Slip Gauge Set , up to 450 mm 10µm
DIMENSION (BASIC 0.01 mm Long Slip Block & Dial
MEASURING Test Indicator With
INSTRUMENT, Acc. by Comparison
GAUGE ETC.) Method

229 MECHANICAL- Inside Dial Caliper L.C. Using Slip Gauge Set. 5 mm to 300 mm 9µm
DIMENSION (BASIC 0.001 mm with Accessories by
MEASURING Comparison Method
INSTRUMENT,
GAUGE ETC.)

230 MECHANICAL- Inside Micrometer Two Using Using Slip 300 mm to 1000 mm 19µm
DIMENSION (BASIC Jaw & Sticks Type Gauge Set , Steel
MEASURING L.C.:-0.001mm Gauge Block , with
INSTRUMENT, Accessories by
GAUGE ETC.) Comparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 40 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
231 MECHANICAL- Inside Micrometer Two Using Slip Gauge Set 5 mm to 50 mm 1.4µm
DIMENSION (BASIC Jaw & Sticks Type with slip Accessories
MEASURING L.C.:-0.001mm by Comparison Method
INSTRUMENT,
GAUGE ETC.)

232 MECHANICAL- Inside Micrometer Two Using Slip Gauge Set 50 mm to 300 mm 3µm
DIMENSION (BASIC Jaw & Sticks Type with slip Accessories
MEASURING L.C.:-0.001mm by Comparison Method
INSTRUMENT,
GAUGE ETC.)

233 MECHANICAL- Length Measuring Using Slip gauges, long up to 1000 mm 8.7µm
DIMENSION (BASIC Machine (LMM)L.C - slip gauges by
MEASURING 0.001 mm comparision method
INSTRUMENT,
GAUGE ETC.)

234 MECHANICAL- Linear Height Gauge Using Using Slip up to 1000 mm 10µm
DIMENSION (BASIC L.C.:- 0.0001 mm Gauge Set , Steel
MEASURING Gauge Block , with
INSTRUMENT, Accessories by
GAUGE ETC.) Comparison Method

235 MECHANICAL- Measuring Tape/Pie Using Length up to 50 meter 578*sqrt(L)µm


DIMENSION (BASIC tape Measuring Machine
MEASURING Machine by
INSTRUMENT, comparision method
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 41 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
236 MECHANICAL- Metric Steel Scales Using Length up to 3000 mm 578 * sqrt(L)µm
DIMENSION (BASIC Measuring(LMM), by
MEASURING compasion method
INSTRUMENT,
GAUGE ETC.)

237 MECHANICAL- Micrometer Setting Using Universal Length 200 mm to 300 mm 3.5µm
DIMENSION (BASIC Rods/Length Measuring Machine &
MEASURING Long Slip Gauge. by
INSTRUMENT, Comparison Method
GAUGE ETC.)

238 MECHANICAL- Micrometer Setting Using Universal Length 100 mm to 200 mm 2.5µm
DIMENSION (BASIC Rods/Length Bar Measuring Machine &
MEASURING Long Slip Gauge. by
INSTRUMENT, Comparison Method
GAUGE ETC.)

239 MECHANICAL- Pin Gauge Using Universal Length 0.5 mm to 10 mm 0.9µm


DIMENSION (BASIC Measuring Machine by
MEASURING Comparison Method
INSTRUMENT,
GAUGE ETC.)

240 MECHANICAL- Pin Gauge Using Universal Length 10 mm to 20 mm 1.0µm


DIMENSION (BASIC Measuring Machine by
MEASURING Comparison Method
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 42 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
241 MECHANICAL- Plain Plug Gauges / Air Using Universal Length 1 mm to 100 mm 3.3µm
DIMENSION (BASIC Plug Gauges ( GO & Measuring Machine by
MEASURING NOGO Comparison Method
INSTRUMENT,
GAUGE ETC.)

242 MECHANICAL- Plain Plug Gauges / Air Using Universal Length 100 mm to 200 mm 3.5µm
DIMENSION (BASIC Plug Gauges ( GO & Measuring Machine by
MEASURING NOGO Comparison Method
INSTRUMENT,
GAUGE ETC.)

243 MECHANICAL- Radius Gauge Using Profile Projector, up to 40 mm 20.7µm


DIMENSION (BASIC comparision method
MEASURING
INSTRUMENT,
GAUGE ETC.)

244 MECHANICAL- Setting Ring Gauge Using Universal Length 100 mm to 200 mm 4µm
DIMENSION (BASIC Measuring Machine by
MEASURING Comparison Method
INSTRUMENT,
GAUGE ETC.)

245 MECHANICAL- Setting Ring Gauge Using Universal Length 3 mm to 100 mm 3.3µm
DIMENSION (BASIC Measuring Machine by
MEASURING Comparison Method
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 43 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
246 MECHANICAL- Snap Gauge ( GO & Using Slip Gauge Set 100 mm to 200 mm 4µm
DIMENSION (BASIC NOGO ) by Comparison Method
MEASURING
INSTRUMENT,
GAUGE ETC.)

247 MECHANICAL- Snap Gauge ( GO & Using Slip Gauge Set 3 mm to 100 mm 2µm
DIMENSION (BASIC NOGO ) by Comparison Method
MEASURING
INSTRUMENT,
GAUGE ETC.)

248 MECHANICAL- Straight Edge ( Using Dial Test up to 500 mm 10.7µm to 11.0µm
DIMENSION (BASIC Parallelism & Indicator & Slip Gauge
MEASURING Straightness ) Set. by Comparison
INSTRUMENT, Method
GAUGE ETC.)

249 MECHANICAL- Surface Plate Using Electronic by up to 2000X2000 mm 1.81v L+W/150 mm


DIMENSION (BASIC Comparison Method
MEASURING (L+W in mm )
INSTRUMENT,
GAUGE ETC.)

250 MECHANICAL- Taper Scale Using Profile Projector 1 mm to 30 mm 65.6µm


DIMENSION (BASIC By comparision method
MEASURING
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 44 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
251 MECHANICAL- Test Foils Using Universal Length up to 2 mm 0.9µm
DIMENSION (BASIC Measuring Machine by
MEASURING Comparison Method
INSTRUMENT,
GAUGE ETC.)

252 MECHANICAL- Test Mandrel ( Run out Using Universal Length up to 150 mm 2.6µm
DIMENSION (BASIC & Diameter ) Measuring Machine &
MEASURING Dial Test Indicator by
INSTRUMENT, Comparison Method
GAUGE ETC.)

253 MECHANICAL- Test Seives Using Digimatic Caliper 4 mm to 100 mm 20µm


DIMENSION (BASIC by comparision method
MEASURING
INSTRUMENT,
GAUGE ETC.)

254 MECHANICAL- Test Sieves Using Profile Projector 32 µm to 4 mm 7µm


DIMENSION (BASIC by Comparison Method
MEASURING
INSTRUMENT,
GAUGE ETC.)

255 MECHANICAL- Thread Pitch Gauge Using Profile Projector 0.25 mm to 20 mm 6.5µm
DIMENSION (BASIC By comparision Method
MEASURING
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 45 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
256 MECHANICAL- Thread Pitch Gauge Using Profile Projector up to 60 ° 73sec of arc'
DIMENSION (BASIC By comparision Method
MEASURING
INSTRUMENT,
GAUGE ETC.)

257 MECHANICAL- Thread Plug Gauge / Using Universal Length 1 mm to 100 mm 2µm
DIMENSION (BASIC Wear Check Plug Measuring Machine by
MEASURING Gauge Comparison Method
INSTRUMENT,
GAUGE ETC.)

258 MECHANICAL- Thread Plug Gauge / Using Universal Length 100 mm to 200 mm 3µm
DIMENSION (BASIC Wear Check Plug Measuring Machine by
MEASURING Gauge Comparison Method
INSTRUMENT,
GAUGE ETC.)

259 MECHANICAL- Thread Ring Gauge / Using Universal Length 3 mm to 100 mm 2.1µm
DIMENSION (BASIC Wear Check Ring Measuring Machine by
MEASURING Gauge Comparison Method
INSTRUMENT,
GAUGE ETC.)

260 MECHANICAL- Three Wire Set. Using Universal Length 0.17 mm to 6.35 mm 1.1µm
DIMENSION (BASIC Measuring Machine by
MEASURING Comparison Method
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 46 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
261 MECHANICAL- Ultrasonic Thickness Using Slip Gauge Set up to 200 mm 111.4µm
DIMENSION (BASIC GaugeL.C - 0.01 mm By comparision method
MEASURING
INSTRUMENT,
GAUGE ETC.)

262 MECHANICAL- Universal Length Using Slip Gauge Set 0 to 100 mm 1µm
DIMENSION (BASIC Measuring Machine / by Comparison Method
MEASURING Single Axis Machine
INSTRUMENT,
GAUGE ETC.)

263 MECHANICAL- V Using Digimatic upto 300 mm 5.6µm


DIMENSION (BASIC BlockFlatnessParallelis Indicator, test Mandrel
MEASURING mSymmetricity and Angle gauge by
INSTRUMENT, comparision method
GAUGE ETC.)

264 MECHANICAL- welding fillet gauge Using Profile Projector 0 mm to 60 mm 0.260µm


DIMENSION (BASIC by comparision method
MEASURING
INSTRUMENT,
GAUGE ETC.)

265 MECHANICAL- Welding Fillet gauge Using Profile Projector 1 ° to 90 ° 3'


DIMENSION (BASIC by comparision method
MEASURING
INSTRUMENT,
GAUGE ETC.)

266 MECHANICAL- Comparator Stand Using Dial gauge with 150 X150 mm 3.0µm
DIMENSION stand
(PRECISION
INSTRUMENTS)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 47 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
267 MECHANICAL- LVDT/Probe with DRO Using Universal Length up to 100 mm 1.5µm
DIMENSION L.C - 0.0001 mmL.C - Measuring Machine
(PRECISION 0.001 mm and slip gauge set by
INSTRUMENTS) comparision method

268 MECHANICAL- LVDT/Probe with DRO Using Universal Length up to 25 mm 1 1µm


DIMENSION L.C - 0.0001 mmL.C - Measuring Machine by
(PRECISION 0.001 mm comparision method
INSTRUMENTS)

269 MECHANICAL- Profile Projector / Using Glass Scale up to 100 X 0.4%


DIMENSION Measuring Microscope Angle Gauge & Dig.
(PRECISION L.C.:- 0.001 mm / 1 Caliper by Comparison
INSTRUMENTS) sec. Linear Angle Method
Magnification

270 MECHANICAL- Profile Projector / Using Glass Scale up to 300 mm 12.7um


DIMENSION Measuring Microscope Angle Gauge & Dig.
(PRECISION L.C.:- 0.001 mm / 1 Caliper by Comparison
INSTRUMENTS) sec. Linear Angle Method
Magnification

271 MECHANICAL- Profile Projector / Using Glass Scale up to 360 deg. 1min.
DIMENSION Measuring Microscope Angle Gauge & Dig.
(PRECISION L.C.:- 0.001 mm / 1 Caliper by Comparison
INSTRUMENTS) sec. Linear Angle Method
Magnification

272 MECHANICAL- Rubber Hardness Using Rubber 0 to 100 Shore ( A&D) 1.3Shore (A&D)
DUROMETER Tester ( Shore A&D ) Hardness Tester
Calibrator Comparison
Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 48 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
273 MECHANICAL- Low Pressure Gauge ( Using Precision Digital 0 to 1 bar 0.0012bar
PRESSURE Dial / Digital / Gauge Comparison
INDICATING DEVICES Transmitter / Switch / Method Based on DKD
Transducers ) -R6-1

274 MECHANICAL- Low Pressure Gauge ( Using Precision Digital 0 bar to 10 bar 0.0083bar
PRESSURE Dial / Digital / gauges by comparison
INDICATING DEVICES Transmitter / Switch / Method
Transducers )

275 MECHANICAL- Low Pressure Gauge ( Using Precision Digital 0 to 2000 Pa 1.72Pa
PRESSURE Dial / Digital / Gauge Comparison
INDICATING DEVICES Transmitter / Switch / Method Based on DKD
Transducers / -R6-
Magnehelic)

276 MECHANICAL- Pressure Gauge ( Dial / Using Precision Digital 0 to 100 bar 0.09bar
PRESSURE Digital / Transmitter / Gauge Comparison
INDICATING DEVICES Switch / Transducers ) Method Based on DKD
-R-6-1

277 MECHANICAL- Pressure Gauge ( Dial / Using Precision Digital 0 to 1000 bar 0.25bar
PRESSURE Digital / Transmitter / Gauge Comparison
INDICATING DEVICES Switch / Transducers )( Method Based on DKD
Hydraulic) -R6-1

278 MECHANICAL- Pressure Gauge ( Dial / Using Precision Digital 0 to 700 bar 0.14bar
PRESSURE Digital / Transmitter / Gauge Comparison
INDICATING DEVICES Switch / Transducers )( Method Based on DKD
Hydraulic) -R6-1

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 49 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
279 MECHANICAL- Vaccumme Gauge ( Using Precision Digital 0 bar to (-) 0.90 bar 0.0012bar
PRESSURE Dial / Digital / guages by Comparision
INDICATING DEVICES Transmitter / Switch / Method
Transducers )(
Hydraulic )

280 MECHANICAL- Torque Screw Driver ( Using Digital Torque 0.5 Nm to 5 Nm 2.1%
TORQUE Type A - Class D ) ( Sensor with Indicator
GENERATING Type B - Class D&E by Comparison Method
DEVICES

281 MECHANICAL- Torque Wrench ( Type Using Digital Torque 0.5 to 100 Nm 1.7%
TORQUE I - Class B & C ) Sensor with Indicator
GENERATING by Comparison Method
DEVICES

282 MECHANICAL- Torque Wrench ( Type Using Digital Torque 50 to 500 Nm 1.2%
TORQUE II - Class A & B ) Sensor with Indicator
GENERATING by Comparison Method
DEVICES

283 MECHANICAL- Glassware Pipettes / Using Precision 0.1 ml to 1 ml 0.3µl


VOLUME Burettes Balance and Distilled
Water of Known
Density Gravimetric
Method

284 MECHANICAL- Glassware Pipettes / Using Precision 1 ml to 10 ml 0.8µl


VOLUME Burettes Balance and Distilled
Water of Known
Density Gravimetric
Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 50 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
285 MECHANICAL- Glassware Pipettes / Using Precision 10 ml to 50 ml 8.2µl
VOLUME Burettes Balance and Distilled
Water of Known
Density Gravimetric
Method

286 MECHANICAL- Measuring Cylinder / Using Precision 0.1 ml to 10 ml 3.4µl


VOLUME volumetric flask / Balance and Distilled
Graduated jar / Can / Water of Known
Beaker etc Density Gravimetric
Method

287 MECHANICAL- Measuring Cylinder / Using Precision 10 ml to 100 ml 8.2µl


VOLUME volumetric flask / Balance and Distilled
Graduated jar / Can / Water of Known
Beaker etc Density Gravimetric
Method

288 MECHANICAL- Measuring Cylinder / Using Precision 100 ml to 250 ml 196µl


VOLUME volumetric flask / Balance and Distilled
Graduated jar / Can / Water of Known
Beaker etc Density Gravimetric
Method

289 MECHANICAL- Measuring Cylinder / Using Precision 1000 ml to 5000 ml 3.4ml


VOLUME volumetric flask / Balance and Distilled
Graduated jar / Can / Water of Known
Beaker etc Density Gravimetric
Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 51 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
290 MECHANICAL- Measuring Cylinder / Using Precision 250 ml to 1000 ml 1.03ml
VOLUME volumetric flask / Balance and Distilled
Graduated jar / Can / Water of Known
Beaker etc Density Gravimetric
Method

291 MECHANICAL- Micropipette Using Precision 10 µl to 100 µl 0.5µl


VOLUME Balance and Distilled
Water of Known
Density Gravimetric
Method

292 MECHANICAL- Micropipette Using Precision 100 µl to 1000 µl 0.6µl


VOLUME Balance and Distilled
Water of Known
Density Gravimetric
Method

293 MECHANICAL- Weighing Balances @ Using Standard 0 to 200 g 0.5mg


WEIGHING SCALE Readability = 0.1 mg Weights of Accuracy
AND BALANCE and Coarser Class F1 Based on
OIML R - 76 - 1 by
Comparison Method

294 MECHANICAL- Weighing Balances @ Using Standard 0 to 100 kg 2.3g


WEIGHING SCALE Readability = 1 g and Weights of Accuracy
AND BALANCE Coarser Class F1 & F2 Based
on OIML R - 76 - 1 by
Comparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 52 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
295 MECHANICAL- Weighing Balances @ Using Standard 0 to 1 kg 3.0mg
WEIGHING SCALE Readability = 1 mg and Weights of Accuracy
AND BALANCE Coarser Class F1 Based on
OIML R - 76 - 1 by
Comparison Method

296 MECHANICAL- Weighing Balances @ Using Standard 0 to 200 kg 10g


WEIGHING SCALE Readability = 10 g and Weights of Accuracy
AND BALANCE Coarser Class F1 , F2 & M1
Based on OIML R - 76
- 1 by Comparison
Method

297 MECHANICAL- Weighing Balances @ Using Standard 0 to 300 kg 22.7g


WEIGHING SCALE Readability = 10 g and Weights of Accuracy
AND BALANCE Coarser Class F1 , F2 & M1
Based on OIML R - 76
- 1 by Comparison
Method

298 MECHANICAL- Weighing Balances @ Using Standard 0 to 10 kg 15mg


WEIGHING SCALE Readability = 10 mg Weights of Accuracy
AND BALANCE and Coarser Class F1 Based on
OIML R - 76 - 1 by
Comparison Method

299 MECHANICAL- Weighing Balances @ Using Standard 0 to 50 kg 1.1g


WEIGHING SCALE Readability = 100 mg Weights of Accuracy
AND BALANCE and Coarser Class F1 Based on
OIML R - 76 - 1 by
Comparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 53 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
300 MECHANICAL- Weights of Accuracy Using Standard 1g 0.1mg
WEIGHTS Class F2 Coarse Weights of Accuracy
Class F1 Substitution
Method of Weighing
and ABBA Weighing
Cycle Based on OIML
R111-1 & Precision
Balance of Readability,
0.01 mg

301 MECHANICAL- Weights of Accuracy Using Standard 1 kg 5mg


WEIGHTS Class F2 Coarse Weights of Accuracy
Class F1 Substitution
Method of Weighing
and ABBA Weighing
Cycle Based on OIML
R111-1 & Precision
Balance of Readability,
1 mg

302 MECHANICAL- Weights of Accuracy Using Standard 10 g 0.2mg


WEIGHTS Class F2 Coarse Weights of Accuracy
Class F1 Substitution
Method of Weighing
and ABBA Weighing
Cycle Based on OIML
R111-1 & Precision
Balance of Readability,
0.01 mg

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 54 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
303 MECHANICAL- Weights of Accuracy Using Standard 10 mg 0.02mg
WEIGHTS Class F2 Coarse Weights of Accuracy
Class F1 Substitution
Method of Weighing
and ABBA Weighing
Cycle Based on OIML
R111-1 & Precision
Balance of Readability,
0.01 mg

304 MECHANICAL- Weights of Accuracy Using Standard 100 g 0.5mg


WEIGHTS Class F2 Coarse Weights of Accuracy
Class F1 Substitution
Method of Weighing
and ABBA Weighing
Cycle Based on OIML
R111-1 & Precision
Balance of Readability,
0.01 mg

305 MECHANICAL- Weights of Accuracy Using Standard 100 mg 0.05mg


WEIGHTS Class F2 Coarse Weights of Accuracy
Class F1 Substitution
Method of Weighing
and ABBA Weighing
Cycle Based on OIML
R111-1 & Precision
Balance of Readability,
0.01 mg

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 55 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
306 MECHANICAL- Weights of Accuracy Using Standard 2g 0.12mg
WEIGHTS Class F2 Coarse Weights of Accuracy
Class F1 Substitution
Method of Weighing
and ABBA Weighing
Cycle Based on OIML
R111-1 & Precision
Balance of Readability,
0.01 mg

307 MECHANICAL- Weights of Accuracy Using Standard 2 kg 10.5mg


WEIGHTS Class F2 Coarse Weights of Accuracy
Class F1 Substitution
Method of Weighing
and ABBA Weighing
Cycle Based on OIML
R111-1 & Precision
Balance of Readability,
10 mg

308 MECHANICAL- Weights of Accuracy Using Standard 2 mg 0.02mg


WEIGHTS Class F2 Coarse Weights of Accuracy
Class F1 Substitution
Method of Weighing
and ABBA Weighing
Cycle Based on OIML
R111-1 & Precision
Balance of Readability,
0.01 mg

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 56 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
309 MECHANICAL- Weights of Accuracy Using Standard 20 g 0.25mg
WEIGHTS Class F2 Coarse Weights of Accuracy
Class F1 Substitution
Method of Weighing
and ABBA Weighing
Cycle Based on OIML
R111-1 & Precision
Balance of Readability,
0.01 mg

310 MECHANICAL- Weights of Accuracy Using Standard 20 mg 0.03mg


WEIGHTS Class F2 Coarse Weights of Accuracy
Class F1 Substitution
Method of Weighing
and ABBA Weighing
Cycle Based on OIML
R111-1 & Precision
Balance of Readability,
0.01 mg

311 MECHANICAL- Weights of Accuracy Using Standard 200 g 0.5mg


WEIGHTS Class F2 Coarse Weights of Accuracy
Class F1 Substitution
Method of Weighing
and ABBA Weighing
Cycle Based on OIML
R111-1 & Precision
Balance of Readability,
0.01 mg

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 57 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
312 MECHANICAL- Weights of Accuracy Using Standard 200 mg 0.06mg
WEIGHTS Class F2 Coarse Weights of Accuracy
Class F1 Substitution
Method of Weighing
and ABBA Weighing
Cycle Based on OIML
R111-1 & Precision
Balance of Readability,
0.01 mg

313 MECHANICAL- Weights of Accuracy Using Standard 5g 0.16mg


WEIGHTS Class F2 Coarse Weights of Accuracy
Class F1 Substitution
Method of Weighing
and ABBA Weighing
Cycle Based on OIML
R111-1 & Precision
Balance of Readability,
0.01 mg

314 MECHANICAL- Weights of Accuracy Using Standard 5 kg 25mg


WEIGHTS Class F2 Coarse Weights of Accuracy
Class F1 Substitution
Method of Weighing
and ABBA Weighing
Cycle Based on OIML
R111-1 & Precision
Balance of Readability,
10 mg

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 58 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
315 MECHANICAL- Weights of Accuracy Using Standard 5 mg 0.02mg
WEIGHTS Class F2 Coarse Weights of Accuracy
Class F1 Substitution
Method of Weighing
and ABBA Weighing
Cycle Based on OIML
R111-1 & Precision
Balance of Readability,
0.01 mg

316 MECHANICAL- Weights of Accuracy Using Standard 50 mg 0.03mg


WEIGHTS Class F2 Coarse Weights of Accuracy
Class F1 Substitution
Method of Weighing
and ABBA Weighing
Cycle Based on OIML
R111-1 & Precision
Balance of Readability,
0.01 mg

317 MECHANICAL- Weights of Accuracy Using Standard 500 g 2.5mg


WEIGHTS Class F2 Coarse Weights of Accuracy
Class F1 Substitution
Method of Weighing
and ABBA Weighing
Cycle Based on OIML
R111-1 & Precision
Balance of Readability,
1 mg

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 59 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
318 MECHANICAL- Weights of Accuracy Using Standard 500 mg 0.08mg
WEIGHTS Class F2 Coarse Weights of Accuracy
Class F1 Substitution
Method of Weighing
and ABBA Weighing
Cycle Based on OIML
R111-1 & Precision
Balance of Readability,
0.01 mg

319 MECHANICAL- Weights of Accuracy Using Standard 1 mg 0.02mg


WEIGHTS Class F2 Coarser Weights of Accuracy
Class F1 Substitution
Method of Weighing
and ABBA Weighing
Cycle Based on OIML
R111-1 & Precision
Balance of Readability,
0.01 mg

320 MECHANICAL- Weights of Accuracy Using Standard 50 g 0.3mg


WEIGHTS Class F2 Coarser Weights of Accuracy
Class F1 Substitution
Method of Weighing
and ABBA Weighing
Cycle Based on OIML
R111-1 & Precision
Balance of Readability,
0.01 mg

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 60 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
321 MECHANICAL- Weights of Accuracy Using Standard 10 kg 1.6g
WEIGHTS Class M3 Coarse Weights of Accuracy
Class F1 Substitution
Method of Weighing
and ABBA Weighing
Cycle Based on OIML
R111-1 & Precision
Balance of Readability,
1g

322 MECHANICAL- Weights of Accuracy Using Standard 20 kg 2.1g


WEIGHTS Class M3 Coarse Weights of Accuracy
Class F1 Substitution
Method of Weighing
and ABBA Weighing
Cycle Based on OIML
R111-1 & Precision
Balance of Readability,
1g

323 MECHANICAL- Weights of Accuracy Using Standard 50 kg 7g


WEIGHTS Class M3 Coarse Weights of Accuracy
Class F1 Substitution
Method of Weighing
and ABBA Weighing
Cycle Based on OIML
R111-1 & Precision
Balance of Readability,
1g

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 61 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
324 THERMAL- SPECIFIC Dial / Digital / Analog Using Humidity 10 % RH to 95 % RH 0.83%
HEAT & HUMIDITY Thermo Hygrometer / Chamber & Digital RH @ 25°C
RH Sensors / with & Temperature
Indicator / Recorder / Indicator with SPRT by
Data Logger Comparison Method

325 THERMAL- SPECIFIC Dial / Digital / Analog Using Humidity 5 °C to 55 @50 % RH 0.4°C
HEAT & HUMIDITY Thermo Hygrometer / Chamber & Digital RH
RH Sensors / with & Temperature
Indicator / Recorder / Indicator with SPRT by
Data Logger Comparison Method

326 THERMAL- Black Body Source ( e IR Thermometer 50 °C to 500 °C 3.69°C


TEMPERATURE = 0.95)

327 THERMAL- IR thermometer/Laser Using Black body 50 °C to 500 °C 3.62°C


TEMPERATURE Gun/Pyrometer/Therma source and IR
l Imaging Camera etc. Thermometer by
comparision method

328 THERMAL- RTD" s , Using SPRT with (-) 25 °C to 140 °C 0.21°C


TEMPERATURE Thermocouples With or Temperature Indicator
Without Indicator / Data & Drywell Furnace by
Logger / Recorder , Comparison Method
Temperature
Transmitter , Digital
Thermometer etc.

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 62 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
329 THERMAL- RTD" s , Using SPRT with 150 °C to 600 °C 0.63°C
TEMPERATURE Thermocouples With or Temperature Indicator
Without Indicator / Data & Drywell Furnace by
Logger / Recorder , Comparison Method
Temperature
Transmitter , Digital
Thermometer etc.

330 THERMAL- RTD" s , Using " S" type 600 °C to 1200 °C 1.67°C
TEMPERATURE Thermocouples With or Thermocouple with
Without Indicator / Data Temperature Indicator
Logger / Recorder , & Drywell Furnace by
Temperature Comparison Method
Transmitter , Digital
Thermometer etc.

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 63 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

1 ELECTRO- AC Current@50 Hz to Using Fluke Dig. 10 µA to 10 A 0.25%


TECHNICAL- 1 kHz Precision Multimeter
ALTERNATING 8846A by Direct
CURRENT (< 1 GHZ) Method
(Measure)

2 ELECTRO- AC Current@50 Hz to Using Fluke Calibrator 10 A to 20 A 0.4% to 0.9%


TECHNICAL- 1 kHz 5080A by Direct
ALTERNATING Method
CURRENT (< 1 GHZ)
(Source)

3 ELECTRO- AC Current@50 Hz to Using Fluke Calibrator 3.3 mA to 3 A 0.3% to 0.6%


TECHNICAL- 1 kHz 5080A by Direct
ALTERNATING Method
CURRENT (< 1 GHZ)
(Source)

4 ELECTRO- AC Current@50 Hz to Using Fluke Calibrator 30 µA to 300 µA 3.2% to 0.6%


TECHNICAL- 1 kHz 5080A by Direct
ALTERNATING Method
CURRENT (< 1 GHZ)
(Source)

5 ELECTRO- AC Current@50 Hz to Using Fluke Calibrator 20 A to 1000 A 0.9% to 1.1%


TECHNICAL- 1kHz 5080A With Current
ALTERNATING Coil ( 50 turn ) by
CURRENT (< 1 GHZ) Direct Method
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 64 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

6 ELECTRO- AC Current@50 Hz to Using Fluke Calibrator 3 A to 10 A 0.3% to 0.4%


TECHNICAL- 1kHz 5080A by Direct Metho
ALTERNATING
CURRENT (< 1 GHZ)
(Source)

7 ELECTRO- AC Current@50 Hz to Using Fluke Calibrator 300 µA to 3.3 mA 0.6% to 0.3%


TECHNICAL- 1kHz 5080A by Direct
ALTERNATING Method
CURRENT (< 1 GHZ)
(Source)

8 ELECTRO- DC Current Using Fluke Dig. 1 µA to 10 µA 0.4% to 0.09%


TECHNICAL- DIRECT Precision Multimeter
CURRENT (Measure) 8846A by Direct
Method

9 ELECTRO- DC Current Using Fluke Dig. 10 µA to 100 µA 0.09%


TECHNICAL- DIRECT Precision Multimeter
CURRENT (Measure) 8846A by Direct
Method

10 ELECTRO- DC Current Using Fluke Dig. 100 µA to 10 A 0.09% to 0.2%


TECHNICAL- DIRECT Precision Multimeter
CURRENT (Measure) 8846A by Direct
Method

11 ELECTRO- DC Current Using Fluke Calibrator 10 µA to 300 µA 1.5% to 0.13%


TECHNICAL- DIRECT 5080A by Direct
CURRENT (Source) Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 65 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

12 ELECTRO- DC Current Using Fluke Calibrator 20 A to 500 A 0.9% to 1.8%


TECHNICAL- DIRECT 5080A With Current
CURRENT (Source) Coil ( 50 turn ) by
Direct Method

13 ELECTRO- DC Current Using Fluke Calibrator 3 A to 20 A 0.2% to 0.9%


TECHNICAL- DIRECT 5080A by Direct
CURRENT (Source) Method

14 ELECTRO- DC Current Using Fluke Calibrator 3 mA to 30 mA 0.09% to 0.07%


TECHNICAL- DIRECT 5080A by Direct
CURRENT (Source) Method

15 ELECTRO- DC Current Using Fluke Calibrator 30 mA to 300 mA 0.07%


TECHNICAL- DIRECT 5080A by Direct
CURRENT (Source) Method

16 ELECTRO- DC Current Using Fluke Calibrator 300 µA to 3 mA 0.13% to 0.09%


TECHNICAL- DIRECT 5080A by Direct
CURRENT (Source) Method

17 ELECTRO- DC Current Using Fluke Calibrator 300 mA to 3 A 0.07% to 0.2%


TECHNICAL- DIRECT 5080A by Direct
CURRENT (Source) Method

18 ELECTRO- DC Current Using Fluke Calibrator 500 to 1000 1.8%


TECHNICAL- DIRECT 5080A With Current
CURRENT (Source) Coil ( 50 turn ) by
Direct Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 66 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

19 ELECTRO- AC High Voltage(1- Using High Voltage 1 kV to 100 kV 1.07% to 1.5%


TECHNICAL- Phase) Divider With DMM by
ELECTRICAL Direct Method
EQUIPMENT
(Measure)

20 ELECTRO- AC Power/active Using Power Meter 0.4 wh to 12 KWh 0.24% to 0.09%


TECHNICAL- Energy 50 Hz(-)0.1 PF AC/DC by Direct
ELECTRICAL to upf to0.1 PF to Method
EQUIPMENT UPF(1-Phase and 3-
(Measure) Phase)4V to 600 V(0.1
A to 20 A)

21 ELECTRO- AC Voltage@ 50 Hz 1 Using Fluke Dig. 1 mV to 100 mV 4.75%


TECHNICAL- kHz Precision Multimeter
ELECTRICAL 8846A by Direct
EQUIPMENT Method
(Measure)

22 ELECTRO- AC Voltage@ 50 Hz 1 Using Fluke Dig. 100 mV to 1000 V 0.12%


TECHNICAL- kHz Precision Multimeter
ELECTRICAL 8846A by Direct
EQUIPMENT Method
(Measure)

23 ELECTRO- Capacitance Using Fluke Dig. 1 nF to 1 mF 1.17% to 0.014%


TECHNICAL- Precision Multimeter
ELECTRICAL 8846A by Direct
EQUIPMENT Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 67 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

24 ELECTRO- DC High Voltage Using High Voltage 1 kV to 100 kV 0.9%


TECHNICAL- Divider By Direct
ELECTRICAL Method
EQUIPMENT
(Measure)

25 ELECTRO- DC Voltage Using Fluke Dig. 1 mV to 100 mV 0.5% to 0.01%


TECHNICAL- Precision Multimeter
ELECTRICAL 8846A by Direct
EQUIPMENT Method
(Measure)

26 ELECTRO- DC Voltage Using Fluke Dig. 10 V to 1000 V 0.007% to 0.008%


TECHNICAL- Precision Multimeter
ELECTRICAL 8846A by Direct
EQUIPMENT Method
(Measure)

27 ELECTRO- DC Voltage Using Fluke Dig. 100 mV to 10 V 0.01% to 0.007%


TECHNICAL- Precision Multimeter
ELECTRICAL 8846A by Direct
EQUIPMENT Method
(Measure)

28 ELECTRO- Resistance Using Fluke Dig. 1 k ohm to 1 M ohm 0.02% to 0.01%


TECHNICAL- Precision Multimeter
ELECTRICAL 8846A by Direct
EQUIPMENT Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 68 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

29 ELECTRO- Resistance Using Fluke Dig. 1 M ohm to 10 M ohm 0.01% to 0.05%


TECHNICAL- Precision Multimeter
ELECTRICAL 8846A by Direct
EQUIPMENT Method
(Measure)

30 ELECTRO- Resistance Using Fluke Dig. 1 ohm to 10 ohm 0.4% to 0.05%


TECHNICAL- Precision Multimeter
ELECTRICAL 8846A by Direct
EQUIPMENT Method
(Measure)

31 ELECTRO- Resistance Using Fluke Dig. 10 M ohm to 100 M 0.05% to 1%


TECHNICAL- Precision Multimeter ohm
ELECTRICAL 8846A by Direct
EQUIPMENT Method
(Measure)

32 ELECTRO- Resistance Using Fluke Dig. 10 ohm to 1 k ohm 0.04% to 0.02%


TECHNICAL- Precision Multimeter
ELECTRICAL 8846A by Direct
EQUIPMENT Method
(Measure)

33 ELECTRO- Resistance Using Fluke Dig. 100 M ohm to 1 G ohm 1% to 2.3%


TECHNICAL- Precision Multimeter
ELECTRICAL 8846A by Direct
EQUIPMENT Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 69 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

34 ELECTRO- AC Voltage @50 Hz to Using Fluke Calibrator 10 mV to 30 mV 1.5% to 0.66%


TECHNICAL- 1 kHz 5080A by Direct
ELECTRICAL Method
EQUIPMENT (Source)

35 ELECTRO- AC Voltage @50 Hz to Using Fluke Calibrator 30 mV to 300 mV 0.66% to 0.02%


TECHNICAL- 1 kHz 5080A by Direct
ELECTRICAL Method
EQUIPMENT (Source)

36 ELECTRO- AC Voltage @50 Hz to Using Fluke Calibrator 300 mV to 1000 V 0.03% to 0.02%
TECHNICAL- 1 kHz 5080A by Direct
ELECTRICAL Method
EQUIPMENT (Source)

37 ELECTRO- Active power UPF , 50 Using Fluke Calibrator 0.1 W to 12 kW 1.76%


TECHNICAL- Hz(-)0.1 PF to (-) 0.9 5080A by Direct Metho
ELECTRICAL PF & 0.1 to UPF( 1 & 3
EQUIPMENT (Source) Phase )10 V to 600
V0.1 A to 20A

38 ELECTRO- Capacitance @1kHz Using Standard 1 nF to 1 mF 2.2%


TECHNICAL- Capacitance Box by
ELECTRICAL Direct Method
EQUIPMENT (Source)

39 ELECTRO- DC Capacitance using Decade 100 pF to 100 µF 0.9% to 1.1%


TECHNICAL- Capacitance box by
ELECTRICAL Direct Method
EQUIPMENT (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 70 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

40 ELECTRO- DC High Resistance 2 Using HV mega ohm 1 G ohm 4.6%


TECHNICAL- wire@0-5000 VDC Box by Direct Method
ELECTRICAL
EQUIPMENT (Source)

41 ELECTRO- DC High Resistance 2 Using HV mega ohm 1 T ohm 2.3%


TECHNICAL- wire@0-5000 VDC Box by Direct Method
ELECTRICAL
EQUIPMENT (Source)

42 ELECTRO- DC High Resistance 2 Using HV mega ohm 10 G ohm 2.3%


TECHNICAL- wire@0-5000 VDC Box by Direct Method
ELECTRICAL
EQUIPMENT (Source)

43 ELECTRO- DC High Resistance 2 Using HV mega ohm 100 G ohm 2.3%


TECHNICAL- wire@0-5000 VDC Box by Direct Method
ELECTRICAL
EQUIPMENT (Source)

44 ELECTRO- DC High Resistance 2 Using HV mega ohm 2 G ohm 4.0%


TECHNICAL- wire@0-5000 VDC Box by Direct Method
ELECTRICAL
EQUIPMENT (Source)

45 ELECTRO- DC High Resistance 2 Using HV mega ohm 20 G Ohm 3.6%


TECHNICAL- wire@0-5000 VDC Box by Direct Method
ELECTRICAL
EQUIPMENT (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 71 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

46 ELECTRO- DC High Resistance 2 Using HV mega ohm 20 M ohm 3.6%


TECHNICAL- wire@0-5000 VDC Box by Direct Method
ELECTRICAL
EQUIPMENT (Source)

47 ELECTRO- DC High Resistance 2 Using HV mega ohm 200 G ohm 2.3%


TECHNICAL- wire@0-5000 VDC Box by Direct Method
ELECTRICAL
EQUIPMENT (Source)

48 ELECTRO- DC High Resistance 2 Using HV mega ohm 200 M ohm 3.6%


TECHNICAL- wire@0-5000 VDC Box by Direct Method
ELECTRICAL
EQUIPMENT (Source)

49 ELECTRO- DC High Resistance 2 Using HV mega ohm 500 G ohm 2.3%


TECHNICAL- wire@0-5000 VDC Box by Direct Method
ELECTRICAL
EQUIPMENT (Source)

50 ELECTRO- DC High Resistance 2 Using HV Mega ohm 2 M ohm 3.6%


TECHNICAL- wire@0-5000VDC Box by Direct Method
ELECTRICAL
EQUIPMENT (Source)

51 ELECTRO- DC Power1 V -600 Using Fluke Calibrator 0.1 W to 12 KW 0.70%


TECHNICAL- V0.1 A - 20 A 5080A by Direct
ELECTRICAL Method
EQUIPMENT (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 72 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

52 ELECTRO- DC Resistance 2 wire Using Fluke 5502 1 M ohm to 3 M ohm 0.01% to 0.04%
TECHNICAL- Calibrator by Direct
ELECTRICAL Method
EQUIPMENT (Source)

53 ELECTRO- DC Resistance 2 Wire Using Decade 1 ohm to 1 M ohm 1.3%


TECHNICAL- Resistance Box by
ELECTRICAL Direct Method
EQUIPMENT (Source)

54 ELECTRO- DC Resistance 4 wire Using Standard 1 K ohm 3.5%


TECHNICAL- Resistance Box By
ELECTRICAL Direct Method
EQUIPMENT (Source)

55 ELECTRO- DC Resistance 4 wire Using Standard 1 m ohm 3.5%


TECHNICAL- Resistance Box By
ELECTRICAL Direct Method
EQUIPMENT (Source)

56 ELECTRO- DC Resistance 4 wire Using Standard 1 ohm 3.5%


TECHNICAL- Resistance Box By
ELECTRICAL Direct Method
EQUIPMENT (Source)

57 ELECTRO- DC Resistance 4 wire Using Standard 10 m ohm 3.5%


TECHNICAL- Resistance Box By
ELECTRICAL Direct Method
EQUIPMENT (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 73 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

58 ELECTRO- DC Resistance 4 wire Using Standard 10 ohm 3.5%


TECHNICAL- Resistance Box By
ELECTRICAL Direct Method
EQUIPMENT (Source)

59 ELECTRO- DC Resistance 4 wire Using Standard 100 µ ohm 0.02%


TECHNICAL- Resistance Box By
ELECTRICAL Direct Method
EQUIPMENT (Source)

60 ELECTRO- DC Resistance 4 Wire Using Standard 100 m ohm 3.5%


TECHNICAL- Resistance Box By
ELECTRICAL Direct Method
EQUIPMENT (Source)

61 ELECTRO- DC Resistance 4 wire Using Standard 100 ohm 3.5%


TECHNICAL- Resistance Box By
ELECTRICAL Direct Method
EQUIPMENT (Source)

62 ELECTRO- DC Resistance 4 wire Using Standard 50 µ ohm 0.03%


TECHNICAL- Resistance Box By
ELECTRICAL Direct Method
EQUIPMENT (Source)

63 ELECTRO- DC Voltage Using Fluke Calibrator 1 mV to 30 mV 0.06% to 1.4%


TECHNICAL- 5080A by Direct
ELECTRICAL Method
EQUIPMENT (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 74 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

64 ELECTRO- DC Voltage Using Fluke Calibrator 30 mV to 300 mV 0.06% to 0.03%


TECHNICAL- 5080A by Direct
ELECTRICAL Method
EQUIPMENT (Source)

65 ELECTRO- DC Voltage Using Fluke Calibrator 300 mV to 1000 V 0.02%


TECHNICAL- 5080A by Direct
ELECTRICAL
EQUIPMENT (Source)

66 ELECTRO- Inductance Using Standard 100 µH to 10 H 1.22%


TECHNICAL- Inductance Box by
ELECTRICAL Direct Method
EQUIPMENT (Source)

67 ELECTRO- Power Factor (Lag & Using Fluke Calibrator (-) 0.1 pF to (+) 0.1 pF 0.05pF
TECHNICAL- Lead ) 5080A by Direct
ELECTRICAL Method
EQUIPMENT (Source)

68 ELECTRO- Resistance Using Standard 1 m ohm to 1 K ohm 1.4%


TECHNICAL- Resistance Box by
ELECTRICAL Direct Method
EQUIPMENT (Source)

69 ELECTRO- Resistance 4 W Using Fluke Calibrator 1 ohm to 190 M ohm 1.4%


TECHNICAL- 5080A&Standrad
ELECTRICAL Resistance box by
EQUIPMENT (Source) Direct Metho

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 75 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

70 ELECTRO- J Type Using Fluke (-) 20 °C to 750 °C 0.82°C


TECHNICAL- (Indicator/Controller/Re Multifunction Process
TEMPERATURE corder) Calibrator 725 by Direct
SIMULATION (Source) Method

71 ELECTRO- K Type Using Fluke (-) 140 °C 1.4°C


TECHNICAL- (Indicator/Controller/Re Multifunction Process to 1300 °C
TEMPERATURE corder) Calibrator 725 by Direct
SIMULATION (Source) Method

72 ELECTRO- N Type Using Fluke 0 °C to 1400 °C 1.1°C


TECHNICAL- (Indicator/Controller/Re Multifunction Process
TEMPERATURE corder) Calibrator 725 by Direct
SIMULATION (Source) Method

73 ELECTRO- R Type Using Fluke 600 °C to 1600 °C 1.8°C


TECHNICAL- (Indicator/Controller/Re Multifunction Process
TEMPERATURE corder) Calibrator 725 by Direct
SIMULATION (Source) Method

74 ELECTRO- RTD (PT-100) Using Fluke (-)100 °C to 650 °C 0.41°C


TECHNICAL- (Indicator/Controller/Re Multifunction Process
TEMPERATURE corder) Calibrator 725 by Direct
SIMULATION (Source) Method

75 ELECTRO- S Type Using Fluke 0 °C to 1600 °C 1.8°C


TECHNICAL- (Indicator/Controller/Re Multifunction Process
TEMPERATURE corder) Calibrator 725 by Direct
SIMULATION (Source) Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 76 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

76 ELECTRO- T Type Using Fluke 0 °C to 1200 °C 1.1°C


TECHNICAL- (Indicator/Controller/Re Multifunction Process
TEMPERATURE corder) Calibrator 725 by Direct
SIMULATION (Source) Method

77 ELECTRO- Frequency Using Fluke Dig. 10 Hz to 1 MHz 0.06% to 0.5%


TECHNICAL- TIME & Precision Multimeter
FREQUENCY 8846A by Direct
(Measure) Method

78 ELECTRO- Time Using Time Calibrator 0.1 sec. to 60 sec. 0.02sec. to 0.04sec.
TECHNICAL- TIME & by Direct Method
FREQUENCY
(Measure)

79 ELECTRO- Time Using Time Calibrator 18000 Sec to 86400 7.0Sec to 20sec
TECHNICAL- TIME & by comparision method Sec
FREQUENCY
(Measure)

80 ELECTRO- Time Using Time Calibrator 3600 sec to 18000 sec 1.2sec. to 7.0sec.
TECHNICAL- TIME & by Direct Method
FREQUENCY
(Measure)

81 ELECTRO- Time Using Time Calibrator 60 sec. to 3600 Sec 0.04sec. to 1.2sec.
TECHNICAL- TIME & by Direct Method
FREQUENCY
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 77 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

82 ELECTRO- Frequency Using Fluke Calibrator 45 Hz to 1000 Hz 0.02%


TECHNICAL- TIME & 5080A by Direct
FREQUENCY (Source) Method

83 FLUID FLOW- FLOW Flow meter, , Flow By Using Ultrosonic 1.0 m³/hr to 718 m³/hr 1.5% rdg
MEASURING Transmitter, Flow handheld flow
DEVICES indicating devices calibrator by
comparision method

84 MECHANICAL- Tachometer Using Digital 10 rpm to 1000 rpm 5.3%


ACCELERATION AND Calibrator/Stroboscope/ Tacometer (Non
SPEED RPM Meter/Centrifuge Contact) by
Etc comparision method

85 MECHANICAL- Tachometer Using Digital 1000 rpm to 99900 rpm 1.2%


ACCELERATION AND Calibrator/Stroboscope/ Tacometer (non
SPEED RPM Meter/Centrifuge Contact) by
Etc comparision method

86 MECHANICAL- Bench Centre Using Dig. Indicator , up to 700 mm 12µm


DIMENSION (BASIC (Parallelism & Taper Mandrel &
MEASURING Coaxiality) Standard Mandrel by
INSTRUMENT, Comparison Method
GAUGE ETC.)

87 MECHANICAL- Comparator Stand # Using Electronic Level up to 300X300 mm 6µm


DIMENSION (BASIC Flatness by Comparison Method
MEASURING
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 78 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

88 MECHANICAL- Cube Mould Using Digimatic Caliper 300 x to 300 mm 90.5µm


DIMENSION (BASIC by comparision method
MEASURING
INSTRUMENT,
GAUGE ETC.)

89 MECHANICAL- Length Measuring Using Slip gauges, long up to 1000 mm 8.7µm


DIMENSION (BASIC Machine (LMM)L.C - slip gauges by
MEASURING 0.001 mm comparision method
INSTRUMENT,
GAUGE ETC.)

90 MECHANICAL- Surface Plate Using Electronic by up to 2000X2000 mm 1.81v L+W/150 mm


DIMENSION (BASIC Comparison Method
MEASURING (L+W in mm )
INSTRUMENT,
GAUGE ETC.)

91 MECHANICAL- Test Mandrel ( Run out Using Universal Length up to 150 mm 2.6µm
DIMENSION (BASIC & Diameter ) Measuring Machine &
MEASURING Dial Test Indicator by
INSTRUMENT, Comparison Method
GAUGE ETC.)

92 MECHANICAL- Universal Length Using Slip Gauge Set 0 to 100 mm 1µm


DIMENSION (BASIC Measuring Machine / by Comparison Method
MEASURING Single Axis Machine
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 79 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

93 MECHANICAL- Comparator Stand Using Dial gauge with 150 X150 mm 3.0µm
DIMENSION stand
(PRECISION
INSTRUMENTS)

94 MECHANICAL- Profile Projector / Using Glass Scale up to 100 X 0.4%


DIMENSION Measuring Microscope Angle Gauge & Dig.
(PRECISION L.C.:- 0.001 mm / 1 Caliper by Comparison
INSTRUMENTS) sec. Linear Angle Method
Magnification

95 MECHANICAL- Profile Projector / Using Glass Scale up to 300 mm 12.7um


DIMENSION Measuring Microscope Angle Gauge & Dig.
(PRECISION L.C.:- 0.001 mm / 1 Caliper by Comparison
INSTRUMENTS) sec. Linear Angle Method
Magnification

96 MECHANICAL- Profile Projector / Using Glass Scale up to 360 deg. 1min.


DIMENSION Measuring Microscope Angle Gauge & Dig.
(PRECISION L.C.:- 0.001 mm / 1 Caliper by Comparison
INSTRUMENTS) sec. Linear Angle Method
Magnification

97 MECHANICAL- Low Pressure Gauge ( Using Precision Digital 0 to 1 bar 0.0012bar


PRESSURE Dial / Digital / Gauge Comparison
INDICATING DEVICES Transmitter / Switch / Method Based on DKD
Transducers ) -R6-1

98 MECHANICAL- Low Pressure Gauge ( Using Precision Digital 0 bar to 10 bar 0.0083bar
PRESSURE Dial / Digital / gauges by comparison
INDICATING DEVICES Transmitter / Switch / Method
Transducers )

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 80 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

99 MECHANICAL- Low Pressure Gauge ( Using Precision Digital 0 to 2000 Pa 1.72Pa


PRESSURE Dial / Digital / Gauge Comparison
INDICATING DEVICES Transmitter / Switch / Method Based on DKD
Transducers / -R6-
Magnehelic)

100 MECHANICAL- Pressure Gauge ( Dial / Using Precision Digital 0 to 100 bar 0.09bar
PRESSURE Digital / Transmitter / Gauge Comparison
INDICATING DEVICES Switch / Transducers ) Method Based on DKD
-R-6-1

101 MECHANICAL- Pressure Gauge ( Dial / Using Precision Digital 0 to 1000 bar 0.25bar
PRESSURE Digital / Transmitter / Gauge Comparison
INDICATING DEVICES Switch / Transducers )( Method Based on DKD
Hydraulic) -R6-1

102 MECHANICAL- Pressure Gauge ( Dial / Using Precision Digital 0 to 700 bar 0.14bar
PRESSURE Digital / Transmitter / Gauge Comparison
INDICATING DEVICES Switch / Transducers )( Method Based on DKD
Hydraulic) -R6-1

103 MECHANICAL- Vaccumme Gauge ( Using Precision Digital 0 bar to (-) 0.90 bar 0.0012bar
PRESSURE Dial / Digital / guages by Comparision
INDICATING DEVICES Transmitter / Switch / Method
Transducers )(
Hydraulic )

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 81 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

104 MECHANICAL- UTM, Compression / Using Load Cell with 1 kN to 1000 kN 0.80%
TENSION CREEP Universal Testing Indicator by
AND TORSION Machine / Load Testing Comparison Method
TESTING MACHINE Machine / Spring
Testing Machine /
Flexural Testing
Machine etc. @
Compression

105 MECHANICAL- UTM, Compression / Using Load Cell with 1 N to 10 N 0.81%


TENSION CREEP Universal Testing Indicator by
AND TORSION Machine / Load Testing Comparison Method
TESTING MACHINE Machine / Spring
Testing Machine /
Flexural Testing
Machine etc. @
Compression

106 MECHANICAL- UTM, Compression / Using Load Cell with 10 N to 1 kN 0.5%


TENSION CREEP Universal Testing Indicator by
AND TORSION Machine / Load Testing Comparison Method
TESTING MACHINE Machine / Spring
Testing Machine /
Flexural Testing
Machine etc. @
Compression

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 82 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

107 MECHANICAL- UTM, Tension / Universal Using Load Cell with 1 kN to 100 kN 0.8%
TENSION CREEP Testing Machine / Load Indicator by
AND TORSION Testing Machine / Comparison Method
TESTING MACHINE Spring Testing Machine
/ Flexural Testing
Machine etc. @
Tension

108 MECHANICAL- UTM, Tension / Universal Using Load Cell with 10 N to 1 kN 0.82%
TENSION CREEP Testing Machine / Load Indicator by
AND TORSION Testing Machine / Comparison Method
TESTING MACHINE Spring Testing Machine
/ Flexural Testing
Machine etc. @
Tension

109 MECHANICAL- Weighing Balances @ Using Standard 0 to 200 g 0.5mg


WEIGHING SCALE Readability = 0.1 mg Weights of Accuracy
AND BALANCE and Coarser Class F1 Based on
OIML R - 76 - 1 by
Comparison Method

110 MECHANICAL- Weighing Balances @ Using Standard 0 to 100 kg 2.3g


WEIGHING SCALE Readability = 1 g and Weights of Accuracy
AND BALANCE Coarser Class F1 & F2 Based
on OIML R - 76 - 1 by
Comparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 83 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

111 MECHANICAL- Weighing Balances @ Using Standard 0 to 1 kg 3.0mg


WEIGHING SCALE Readability = 1 mg and Weights of Accuracy
AND BALANCE Coarser Class F1 Based on
OIML R - 76 - 1 by
Comparison Method

112 MECHANICAL- Weighing Balances @ Using Standard 0 to 200 kg 10g


WEIGHING SCALE Readability = 10 g and Weights of Accuracy
AND BALANCE Coarser Class F1 , F2 & M1
Based on OIML R - 76
- 1 by Comparison
Method

113 MECHANICAL- Weighing Balances @ Using Standard 0 to 300 kg 22.7g


WEIGHING SCALE Readability = 10 g and Weights of Accuracy
AND BALANCE Coarser Class F1 , F2 & M1
Based on OIML R - 76
- 1 by Comparison
Method

114 MECHANICAL- Weighing Balances @ Using Standard 0 to 10 kg 15mg


WEIGHING SCALE Readability = 10 mg Weights of Accuracy
AND BALANCE and Coarser Class F1 Based on
OIML R - 76 - 1 by
Comparison Method

115 MECHANICAL- Weighing Balances @ Using Standard 0 to 50 kg 1.1g


WEIGHING SCALE Readability = 100 mg Weights of Accuracy
AND BALANCE and Coarser Class F1 Based on
OIML R - 76 - 1 by
Comparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 84 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

116 THERMAL- SPECIFIC Humidity Indicator of Using Dig. RH Indicator 10 % RH to 95 % RH 0.83%


HEAT & HUMIDITY Humidity Calibration / with sensor @ 25°C
Generator , Humidity Temperature Indicator
Chamber with SPRT@ Single
Position Calibration at
Measuring Location in
DUC by Comparison
Method

117 THERMAL- SPECIFIC Humidity Indicator of Using Dig. RH Indicator 5 °C to 55 °C @ 50 % 0.33°C


HEAT & HUMIDITY Humidity Calibration / with sensor
Generator , Humidity Temperature Indicator
Chamber with SPRT @ Single
Position Calibration at
Measuring Location in
DUC by Comparison
Method

118 THERMAL- Black Body Source ( e IR Thermometer 50 °C to 500 °C 3.69°C


TEMPERATURE = 0.95)

119 THERMAL- Deep Using Data Logger with (-) 70 °C to 50 °C 1.44°C


TEMPERATURE Freezer,Refrigrator, RTD Sensors Multi
Oven , BOD Incubator , Position by
Environmental Comparison Method
Chamber, Vaccum
Oven

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 85 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

120 THERMAL- Dry Block Furnaces / Using Data Logger with 200 °C to 600 °C 5.5°C
TEMPERATURE Muffle Furnace "K" Type Sensors Multi
Position by
Comparison Method

121 THERMAL- Dry Block Furnaces / Using Data Logger with 600 °C to 1000 °C 8.0°C
TEMPERATURE Muffle Furnace "K" Type Sensors Multi
Position by
Comparison Method

122 THERMAL- Oven , Vacuum Oven , Using Data Logger with 50 °C to 250 °C 2.44°C
TEMPERATURE Aging Oven , BOD RTD Sensors Multi
Incubator , Incubator , Position by
Centrifuge Chamber , Comparison Method
Environment Chamber
, Furnaces

123 THERMAL- RTD" s , Using SPRT with (-) 25 °C to 140 °C 0.21°C


TEMPERATURE Thermocouples With or Temperature Indicator
Without Indicator / Data & Drywell Furnace by
Logger / Recorder , Comparison Method
Temperature
Transmitter , Digital
Thermometer etc.

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 86 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

124 THERMAL- RTD" s , Using SPRT with 150 °C to 600 °C 0.63°C


TEMPERATURE Thermocouples With or Temperature Indicator
Without Indicator / Data & Drywell Furnace by
Logger / Recorder , Comparison Method
Temperature
Transmitter , Digital
Thermometer etc.

125 THERMAL- RTD" s , Using " S" type 600 °C to 1200 °C 1.67°C
TEMPERATURE Thermocouples With or Thermocouple with
Without Indicator / Data Temperature Indicator
Logger / Recorder , & Drywell Furnace by
Temperature Comparison Method
Transmitter , Digital
Thermometer etc.

126 THERMAL- Temperature Indicator Using SPRT with (-) 80 °C to 600 °C 0.3°C
TEMPERATURE with sensor of Liquid Temperature Indicator
bath , Oven , Dry Block @ Single Position
Furnace , Freezers , Calibration at
Auto Clave , BOD Measuring Location in
Incubator , DUC by Comparison
Environmental Method
Chamber , Furnace

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name N.C.L. PVT. LTD., B.D. NAGAR, MEERUT ROAD, GHAZIABAD, UTTAR PRADESH ,
INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2213 Page No. : 87 / 87

Validity 29/01/2020 to 28/01/2022 Last Amended on -

S.No Discipline / Group Measurand or Calibration or Measurement range * Calibration and


Reference Measurement and additional Measurement
Material/Type of Method or parameters where Capability(CMC)(±)
instrument or procedure applicable(Range
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

127 THERMAL- Temperature Indicator Using "S" type 600 °C to 1200 °C 2.03°C
TEMPERATURE with sensor of Oven , Thermocouple with
Dry Block Furnaces / Temperature Indicator
Muffle Furnace @ Single Position
Calibration at
Measuring Location in
DUC by Comparison
Method

* CMCs represent expanded uncertainties expressed at approximately the 95% level of confidence, using a coverage factor of k = 2.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

You might also like