Iso 3104 2020 en
Iso 3104 2020 en
se/std-80024544
INTERNATIONAL ISO
STANDARD 3104
Third edition
2020-09
Reference number
ISO 3104:2020(E)
© ISO 2020
This preview is downloaded from www.sis.se. Buy the entire standard via https://www.sis.se/std-80024544
ISO 3104:2020(E)
ISO 3104:2020(E)
Contents Page
Foreword......................................................................................................................................................................................................................................... iv
Introduction...................................................................................................................................................................................................................................v
1 Scope.................................................................................................................................................................................................................................. 1
2 Normative references....................................................................................................................................................................................... 1
3 Terms and definitions...................................................................................................................................................................................... 1
4 Principle......................................................................................................................................................................................................................... 2
5 Reagents and materials.................................................................................................................................................................................. 2
6 Apparatus design and requirements................................................................................................................................................ 3
7 erification.................................................................................................................................................................................................................. 6
V
7.1 Viscometer................................................................................................................................................................................................... 6
7.2 Liquid-in-glass thermometer...................................................................................................................................................... 6
7.3 Digital contact thermometer........................................................................................................................................................ 7
7.4 Timer................................................................................................................................................................................................................ 7
8 Re-calibration........................................................................................................................................................................................................... 7
9 Quality control......................................................................................................................................................................................................... 7
10 Sample preparation............................................................................................................................................................................................ 7
10.1 Pre-analysis sample conditioning............................................................................................................................................ 7
10.2 Visual inspection and filtering.................................................................................................................................................... 8
11 Procedure A — Manual equipment.................................................................................................................................................... 8
12 Procedure B — Automated equipment.......................................................................................................................................10
13 Cleaning of the viscometer tube.........................................................................................................................................................11
14 Calculation................................................................................................................................................................................................................ 11
14.1 Procedure A — Manual viscometers.................................................................................................................................. 11
14.2 Procedure B — Automated viscometers......................................................................................................................... 12
15 Expression of results......................................................................................................................................................................................13
16 Precision of procedure A............................................................................................................................................................................13
16.1 Determinability, d .............................................................................................................................................................................. 13
16.2 Repeatability, r ..................................................................................................................................................................................... 14
16.3 Reproducibility, R .............................................................................................................................................................................. 14
17 Precision of Procedure B............................................................................................................................................................................15
17.1 Determinability, d .............................................................................................................................................................................. 15
17.2 Repeatability, r ..................................................................................................................................................................................... 15
17.3 Reproducibility, R .............................................................................................................................................................................. 15
18 Test report................................................................................................................................................................................................................. 15
Annex A (normative) Viscometer types, calibration and verification.............................................................................17
Annex B (normative) Thermometers for kinematic viscosity test......................................................................................18
Annex C (normative) Conditioning of samples prior to manual or automated analysis..............................22
Annex D (normative) Calculation of acceptable tolerance zone (band) to determine
conformance with a certified reference material............................................................................................................23
Bibliography.............................................................................................................................................................................................................................. 24
ISO 3104:2020(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the
different types of ISO documents should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of
any patent rights identified during the development of the document will be in the Introduction and/or
on the ISO list of patent declarations received (see www.iso.org/patents).
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and
expressions related to conformity assessment, as well as information about ISO's adherence to the
World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT), see www.iso.org/
iso/foreword.html.
This document was prepared by Technical Committee ISO/TC 28, Petroleum and related products, fuels
and lubricants from natural or synthetic sources.
This third edition cancels and replaces the second edition (ISO 3104:1994), which has been technically
revised.
The main changes compared to the previous edition are as follows:
— precision data have been updated to all actual fuels on the market;
— biodiesel (FAME) blends and paraffinic diesel have been included in the scope;
— the procedure description and allowance of automated techniques have been included.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www.iso.org/members.html.
ISO 3104:2020(E)
Introduction
Many petroleum products, and some non-petroleum materials, are used as lubricants, and the correct
operation of equipment depends upon the appropriate viscosity of the liquid being used. In addition,
the viscosity of many petroleum fuels is important for the estimation of optimum storage, handling
and operational conditions. Thus, the accurate measurement of viscosity is essential to many product
specifications.
1 Scope
This document specifies Procedure A, using manual glass viscometers, and Procedure B, using glass
capillary viscometers in an automated assembly, for the determination of the kinematic viscosity, ν,
of liquid petroleum products, both transparent and opaque, by measuring the time for a volume of
liquid to flow under gravity through a calibrated glass capillary viscometer. The dynamic viscosity, η,
is obtained by multiplying the measured kinematic viscosity by the density, ρ, of the liquid. The range
of kinematic viscosities covered in this test method is from 0,2 mm2/s to 300 000 mm2/s over the
temperature range –20 °C to +150 °C.
NOTE The result obtained from this document is dependent upon the behaviour of the sample and is intended
for application to liquids for which primarily the shear stress and shear rates are proportional (Newtonian flow
behaviour). If, however, the viscosity varies significantly with the rate of shear, different results can be obtained
from viscometers of different capillary diameters. The procedure and precision values for residual fuel oils,
which under some conditions exhibit non-Newtonian behaviour, have been included.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments) applies.
ISO 3105, Glass capillary kinematic viscometers — Specifications and operating instructions
ISO 3696:1987, Water for analytical laboratory use — Specification and test methods
ASTM E1137, Standard Specification for Industrial Platinum Resistance Thermometers
ASTM E2877, Standard Guide for Digital Contact Thermometers
ISO 3104:2020(E)
3.1
kinematic viscosity
ν
resistance to flow of a fluid under gravity
Note 1 to entry: For gravity flow under a given hydrostatic head, the pressure head of a liquid is proportional to
its density, ρ. For any particular viscometer, the time of flow of a fixed volume of fluid is directly proportional to
its kinematic viscosity, ν:
ν = η/ρ
3.2
dynamic viscosity
coefficient of dynamic viscosity
viscosity
η
ratio between the applied shear stress and rate of shear of a liquid
Note 1 to entry: It is a measure of the resistance to flow or deformation of a liquid.
Note 2 to entry: The term dynamic viscosity is also used in a different context to denote a frequency-dependent
quantity in which shear stress and shear rate have a sinusoidal time dependence.
3.3
density
ρ
mass per unit volume of a substance at a given temperature
4 Principle
The time is measured for a fixed volume of liquid to flow under gravity through the glass capillary
of a calibrated viscometer under a reproducible driving head and at a known and closely controlled
temperature. The kinematic viscosity is the product of the measured flow time and the calibration
constant of the viscometer.
5.1 Cleaning solution, strongly-oxidizing cleaning solution or alkaline cleaning solutions can be used.
Alkaline cleaning solutions with a pH of greater than 10 are not recommended as they have been shown
to change the viscometer calibration. If these are used, then the viscometer calibration should be
verified to ensure there is no change.
5.2 Sample solvent, completely miscible a prewash with an aromatic solvent such as toluene or
heptane might be necessary to remove asphaltenic material. When cleaning capillaries inside the bath,
the boiling point of the cleaning solution shall be higher than the bath temperature.
5.3 Drying solvent, suitable and volatile at the used temperature. Filter before use. If moisture
remains, use a drying solvent miscible with water (5.4).
NOTE When cleaning capillaries inside the bath and if the bath temperature is higher than 50 °C, acetone is
not suitable.
5.4 Water, deionized or distilled, conforming to Grade 3 of ISO 3696:1987. Filter before use.
ISO 3104:2020(E)
5.5 Certified viscosity reference standards (CRM), with data provided by an accredited calibration
laboratory — traceable to the international agreed value of distilled water (1,003 4 mm2/s at 20 °C as
specified in ISO/TR 3666) and calibrated in accordance with a standard practice for the basic calibration
of master viscometers and viscosity oils, such as in ASTM D2162[8]. ISO 17034 specifies the requirements
for CRM producers.
6.1 Drying tubes, consisting of a desiccant drying system, consisting of either externally mounted
drying tubes or an integrated desiccant drying system designed to remove ambient moisture from the
capillary tube. Ensure that they are packed loosely and that the desiccant is not saturated with water.
6.2 Sample filter, micron screen or fretted (sintered) glass filter, no more than 75 µm.
6.3 Reagent filter, micron screen or fretted (sintered) glass filter, no more than 11 µm.
6.4 Ultrasonic bath, unheated — with an operating frequency between 25 kHz to 60 kHz and a typical
power output of ≤100 W, of suitable dimensions to hold container(s) placed inside of bath, for use in
effectively dissipating and removing air or gas bubbles that can be entrained in viscous sample types
prior to analysis. It is permissible to use ultra-sonic baths with operating frequencies and power outputs
outside this range; however, it is the responsibility of the laboratory to conduct a data comparison study
to confirm that results determined with and without the use of such ultrasonic baths does not materially
impact results.
The viscometer shall have a certificate of calibration provided by a laboratory that meets ISO/IEC 17025.
The calibration constant should be checked before first use of the capillary and only changed if
necessary.
The calibration constant, C, is dependent upon the gravitational acceleration at the place of calibration.
The variation in the value of g across the earth’s surface is about 0,5 % due to latitude plus approximately
0,003 % per 100 m altitude. Apply a gravity correction to the viscometer calibration constant as in
Formula (1), if the acceleration of gravity of the testing laboratory differs by more than 0,1 % of the
calibration laboratory.
g
C2 = 2 C (1)
g1
where the subscripts 1 and 2 indicate, respectively, the calibration laboratory and the testing laboratory.
NOTE Calculation of acceleration of gravity values can be found at www.NPL.co.uk.
IMPORTANT — Viscometers used for silicone fluids, fluorocarbons and other liquids, which are
difficult to remove using a cleaning agent, shall be reserved for the exclusive use of those fluids,
except during their calibration. Subject such viscometers to calibration checks at frequent
intervals. The solvent washings from these viscometers shall not be used for the cleaning of
other viscometers. If the viscometer is cleaned using the material in 5.1 then the user shall
verify the calibration before further use.
ISO 3104:2020(E)
6.5.2 Viscometer holder or mounting device within the temperature-controlled bath, enabling
the glass viscometer to be suspended so that the upper meniscus is directly above the lower meniscus
vertically within 1° in all directions.
Those viscometers whose upper meniscus is offset from directly above the lower meniscus shall be
suspended vertically within 0,3° in all directions (see ISO 3105).
The proper alignment of vertical parts may be confirmed by using a plumb line, but for rectangular
baths with opaque ends, this might not be possible.
6.5.3 Temperature-controlled bath, containing a transparent liquid of sufficient depth such that at
no time during the measurement is any portion of the sample in the viscometer less than 20 mm below
the surface of the bath liquid or less than 20 mm above the bottom of the bath.
Temperature control of the bath liquid shall be such that, for each series of flow-time measurements,
within the range of 15 °C to 100 °C the temperature of the bath medium does not vary by more
than ±0,02 °C from the selected temperature over the length of the viscometer, and/or between the
position of each viscometer, and/or at the location of the thermometer i.e. the temperature needs to
be constant at the capillary and at the position of the thermometer within a maximum difference of
0,04 °C. For temperatures outside this range, the deviation from the desired temperature shall not
exceed ±0,05 °C.
Adjust and maintain the viscometer bath at the required test temperature within the limits given
in 6.5.3, taking account of the conditions given in Annex B and of the corrections supplied on the
certificates of calibration for the thermometers. Maintain the bath temperature at the test temperature
using the readings of the temperature measuring device with the corrections supplied by the certificate
of calibration.
Thermometers shall be held in an upright position under the same conditions of immersion as when
calibrated. In order to obtain the most reliable temperature measurement, it is recommended that two
thermometers with valid calibration certificates be used. They should be viewed with a lens assembly
giving approximately 5× magnification and be arranged to eliminate parallax errors.
6.5.4 Temperature-measuring device, for the range 0 °C to 100 °C, being either:
6.5.4.1 The calibration data should be traceable to a calibration or metrology standards body and
meet the uncertainty of measurement required. The calibration certificate shall include data covering the
series of temperature test points which are appropriate for its intended use. When two thermometers are
used in the same bath in this range, they shall agree within 0,04 °C. See Annex B for the list of complying
thermometers.
If calibrated liquid-in-glass thermometers are used, the use of two thermometers is recommended.
Outside the range 0 °C to 100 °C, a calibrated liquid in-glass thermometer (see Annex B) or a DTC as
described in 6.5.4.2 with a CMC of ±0,05 °C or better shall be used, and when two thermometers are
used in the same bath they shall agree within ±0,1 °C.
When using liquid-in-glass thermometers, use a magnifying device to read the thermometer to
the nearest 1/5 division (e.g. 0,01 °C or 0,02 °C) to ensure that the required test temperature and
temperature control capabilities are met. It is recommended that thermometer readings (and any
corrections supplied on the certificates of calibrations for the thermometers) be recorded on a periodic
basis to demonstrate compliance with the test method requirements.
ISO 3104:2020(E)
NOTE The resulting uncertainty of calibration can be dependent upon the immersion depth.
6.5.4.3 The DCT probe is to be immersed no less than the immersion depth stated on the calibration
certificate.
NOTE With respect to DCT probe immersion depth, a procedure is available in ASTM E644-11:2011, Section
7[10], for determining the minimum depth. With respect to an ice bath, ASTM E563[9] provides guidance on
the preparation of an ice bath however variance from the specific steps is permitted provided preparation is
consistent as it is being used to track change in calibration.
6.6.1 General
Automated viscometers, which use the technical principles of this document, are acceptable provided
they meet the accuracy and precision of all the equipment listed in 6.5. In addition, if they are used to
measure kinematic viscosity in samples subject to conditioning using the steps in Annex C, a heated
sample tray shall be used if the sample is not analysed immediately after conditioning. This sample
tray (6.6.2) shall be heated to a temperature which will not allow the sample to fall below the testing
temperature.
Flow times of less than 200 s are acceptable; however, the kinetic energy correction shall be calculated
and should not exceed 3 % of the measured viscosity. Where a value of greater than 3 % is achieved the
analysis should be repeated using a smaller diameter viscometer tube.
Some automated equipment contain sample loading trays for analysis of multiple samples. When a
sample has been subjected to conditioning using the steps in Annex C, the sample shall not be allowed
to cool below the testing temperature on the loading tray as this will result in an increase in measured
viscosity as compared to the manual procedure. As the temperature of the conditioned sample can
drop very quickly, for analysis of these samples, the sample loading trays shall be heated above the test
temperature to ensure the temperature of the sample does not fall below the test temperature at the
time of analysis, and the nature of the sample is not changed. For analysis of these samples at 50 °C a
sample tray heated at 54 °C has been shown to be sufficient to maintain the sample temperature above
50 °C for at least 40 min. For analysis of these samples at other temperatures the lab should establish
the correct sample tray temperature and time before analysis.