Fundamentals of Thermometry - Part 8
Fundamentals of Thermometry - Part 8
where LA (Tse) and LA [T&X)] are the spectral concentrations of the radiance of a
blackbody at wavelength d in vacuum at Tso and Tse(X) respectively, cz = 0.014388
m.K, and T&X) may be the freezing point of silver, gold or copper.
(Plan&s Law derives from the earlier Wien Law for the distribution of en-
ergy in the emission spectrum of a blackbody. Plan&s modification was simply to
add the -1. However in attempting to explain why a much better agreement with
experimental data was thus obtained, it was necessary for him to invent the quan-
tum theory (that all electromagnetic waves can exist only as discrete packages or
quanta) for which discovery he received the Nobel Prize in 1918).
In the design of the Scale, the silver point was chosen as the transition point
between the HTSPRT and the radiation thermometer because 961 o was the practi-
cal upper limit for the platinum thermometer, not because it was the lower limit for
the radiation thermometer. Indeed, all surfaces at temperatures above zero abso-
lute (0 K) emit and absorb some radiant energy. That this energy is not visible to us
until high temperatures are reached is due to the limitation of the human eye to the
electromagnetic spectrum between 0.4 pm and 0.7 pm (longer wavelengths go into
the infra-red; shorter into the ultra-violet). As temperature increases, peak radiance
shifts toward shorter wavelengths. At lower temperatures, the power emitted by the
radiating blackbody is low; at room temperature is 470 W per square meter, at
5OO’C 20 kW, at 1000°C 150 kW, and at 2500°C 3.4 MW (for reference, the full
sun radiates 77 MW). Low power at low temperature limits the accuracy and re-
peatabilty of radiation thermometers; at high temperatures, other errors and un-
certainties are important.
Cells for realizing the freezing points for the calibration of SPRTs between
the triple point of water and the silver point are well understood, well character-
ized, and in common use. With appropriate equipment and techniques, the ITS-90
temperatures can be realized within fractions of a millikelvin (O.OOl°C) at the
lower temperatures and several millikelvins at the higher temperatures.
All real sources reflect, transmit and absorb radiant energy. For an ideal
source, the emissivity E = 1. For a source which also transmits (r) and absorbs (a)
energy:
The emissivity of real surfaces is usually far less than 1, and estimating the emissiv-
ity of the source is one of the real problems of radiation thermometry. Most pre-
sent-day pyrometers have a dial or index which can be set to the estimated emis-
sivity, to correct for its failure to be unity. The emissivity of some common surfaces
is shown in the Table on the next page.
FIG 3: THE NPL PHOTOELECTRIC PYROMflER: (A) off-axis ellipsoidal mirror, (8) plane
mirror, (C) swinging mirror, (D) 0.75 mm diameter target aperture, (E) Sighting graticule, (F)
sector disc, CC) Sighting telescope CH) Sector traverse mechanism (1) EMI Type 9558 pho-
tomultiplier and amplifier (I) Interference filter 662 nm, (K) Neutral density filters. (After
Quinn, in Proceedings of the International School of Physics <Enrico Fermi>, Course
LXVIII, Metrology andFundamenfa/ Constaots, North Holland Publ. Co. 1980).
MATERIAL TEMPERATURE, K
300 500 1200 1300 1600
One novel (and patented) type of blackbody, which does not require either
a sensor, [41 a controller or an equilibrium of a pure substance, is controlled by an
integral gas-buffered heat pipe, which may be factory-set to a pressure to produce
one specific temperature [Fig. 5).
ISOTHERMAL
PART PkINCIPLEOFTHEHEATPIPESOURCE
h
KCEPTABLE j
=ilIu
“:RVARlATlON cl r
AMBIENT
TEMPERATURE
UPILLIARI
i
1 GAS
BUFFER
HEATIN OUT
FIG. 6: OPTICAL PATH OF A COMMERCIAL PYROMETER. (A) source, (f: objective lens,
(C) objective aperture, (D) absorption filter for temperatures above 13CWC, (E) pyrometer
lamp, (F) red filter, (G) microscope objective lens, (H) microscope aperture stop, (I) micro-
scope ocular, (I) eye, (K) instrument for measuring lamp current. (After NBS Monograph
41 ).
Isothermal Technology Ltd. has had many years of experience in the pro-
duction of metal freeze point cells for the calibration of immersion thermometers,
and has recently applied this experience to equivalent cells for calibrating radiation
thermometers. These cells, intended for use in lsotech dedicated furnaces, were de-
veloped with the cooperation of England’s National Physical Laboratory (NPL),
which established the validity of their equilibria. Metal phase-equilibrium fixed
point cells for use with radiation thermometers have a quite different geometry
from cells intended for the immersion of contact thermometers, and so cannot be
verified by SPRTs. However pure metals melt and freeze at temperatures which are
physical constants and independent of the geometry of the containment. Exhaus-
tive measurements with fine-wire thermocouples have found that the lsotech cells
designed for radiation thermometry have virtually unmeasurable equilibrium tem-
perature differences from cells intended for contact thermometers, and reproduce
The technique for verifying the present purity of the metal in a blackbody
cell is similar to that for an immersion thermometer in a cell intended for
immersion instruments.
Set the furnace controller a few degrees above the anticipated melting tem-
perature of the metal (as it was determined in the last paragraph), monitoring the
temperature, as it rises, with the optical thermometer. At the beginning of the phase
change - when the metal begins to melt - the melt arrest is observed. When the
metal is completely molten, the temperature will resume its rise to the controller
setpoint. The rise to the melt arrest, the melt arrest plateau and the rise from the
plateau should all be recorded on a strip-chart recorder if the pyrometer has con-
nections for one, or, failing that, by hand, and then conveyed to a graphical plot.
Freezing analysis can also provide useful information. To freeze the metal,
first assure that it is completely molten, by ascertaining that the metal temperature
has shown a rise above the melt plateau. Then reset the furnace controller to about
1°C below the melt plateau. Allow the furnace temperature to drift downward,
monitoring it all the while, until the downward drift stops. Some metals exhibit su-
percool; that is, the metal remains liquid at a temperature below the temperature at
which it would melt if it were in the solid state. Other metals exhibit very little su-
percool. If the metal does not supercool the temperature will stabilize on the freeze
plateau; if it does supercool, the temperature will rise to the freeze plateau. Con-
tinue to record the temperature until the freeze plateau terminates and the temper-
ature drops to he controller setting and plot this temperature profile as well.
Figure 8(a) shows the characteristic graphs of the melting behavior of metals
6N+ (better than 99.9999%) pure, 5N+ (better than 99.999%) pure and 5N
(99.999%) pure. If the metal is much less than 5N pure, there will be very little
recognizable plateau. Fig. 8(b) shows typical freezing behavior for metals of several
purities. Note in particular the slopes of the temperature drops after the freeze
plateau has been finished.
A - MELTING B - FREEZING
FIG 8: (A) typical melting curves of metals of three purities (8) typical freezing curves
[l I The first optical pyrometer. Objects which radiate within the visible portion of the spectrum
glow whiter and brighter as temperature increases. Experienced workers in high temperature indus-
tries can often estimate temperature by eye to *50°. This cartoon is from Heat and Temperature
Measurement, Robert L. W&w, @ 1950, Prentice-Hall, Inc
T COLOR
13) e.g., lsotech Model 976 with lsotech “Hockey Puck” cells containing either Gallium (29.76”C),
lndium (156,6O”C), Tin (231.93”0, Zinc (419.53’0, or lsotech Model 970 with lsotech Model CC
cells containing either Indium, Tin, Zinc, Aluminum (662.32”C) or Silver (961.8”C)
141Isotech-Phillips Fixed Point Radiation Source, single temperature gas-buffered heat pipe; one
factory-set temperature between 500° and 1000°C.
151Closer control can be obtained if the furnace controllers are occasionally calibrated. This can
be done with little effort by using the metal fixed points themselves as reference.
(a) Set the furnace controller a few degrees below where your data indicates the melt temperature
should lie.
(b) Allow the system to come to thermal equilibrium at that setting. Monitor the temperature to
look for any thermal cycling. Note the controller setting and the mean indicated temperature.
(c) Reset the controller a few degrees higher than the expected melt temperature. Watch for the
melt arrest. Observe the melt plateau temperature as indicated by the instrument.
(d) At the end of the melt arrest, allow the temperature to rise until it stabilizes. When it is in equi-
librium, look for any thermal cycling. Note the controller setting and the mean indicated tempera-
ture.
C1, tl, the controller sening and temperature indication below the melt plateau
Cz, t2, the controller setting and temperature indication above the melt plateau
t), the temperature indicated by your instrument for the melt plateau.
The controller setting corresponding to the furnace temperature at the melt plateau
is then calculated:
For subsequent meItS or freezes, set the controller slightly above Cs(,,), (to melt) or below CS(mp)
(to freeze) or at a setting which assures that the peak furnace temperature when cycling is more
than t3 (to melt) or less than t3 (to freeze).
REFERENCES
(The following books in our library have been found particularly useful)
MEASUREMENT OF RADIANT ENERGY, ed. W. E. Forsythe, McGraw Hill Co, New York and
London, 1937. lDespite its age, this remains one of the classical works.]
TRACEABLE TEMPERATURES, j. V. Nicholas, D. R. White, John Wiley & Sons, New York, 1994
(reviewed in the lsofech journal of Thermometry, Vol. 6 No. 1, 1995
and for a view ol the curring edge and the developmenl of ITS-90
TEMPERATURE MEASUREMENT 1975, Conference Series No. 26, The Institute of Physics (Great
Britain), London, 1975