200 VOLTS PEAK AC INPUT

70 KV PEAK PER COIL

140 KV PEAK TOTAL +70KV

R1
100 MEG

HVAC1

T1 R2
5 FS1 TUBE CURRENT PCB ASSY. DB1 100 MEG
140-0060
SP1 1 SHOWN FOR REF. ONLY AC1 +
D2
6
0
1N4001 SOLDER TERM. SG2 R3
TB1 D1 75V 20K/1%
FP1 2 SS1 SEC1
1 0
2 R1 SG1
6 SOLDER TERM. 1N4001
3 10K 75v
7 D3
4 SS2 SEC2
5 FP2 0 SG3 R4
6 3
SOLDER TERM. 1N4001 75V 20K/1% V1
7
D4
8 3 X-RAY TUBE
9 0
10
1N4001 SOLDER TERM.
11 SP2
12 4
AC2 - R5
HEADER 12 8 FS2 100 MEG

HV TRANSFORMER

HVAC2
R6
100 MEG
GROUND TO
TANK COVER
-70KV

T2
FIL. TRANS

CHASSIS GROUND
120 VOLTS AC RMS AT
CURRENT SENSE, 10 VOLTS = 1MA. WITHOUT EXTERNAL RESISTOR WHT BLK PULSE WIDTH MODULATED
100% DUTY CYCLE.

NEG SENSE OUTPUT, -1 VOLT = -10 KV

POS SENSE OUTPUT, +1 VOLT = +10 KV

CHASSIS GROUND

GROUND STUD

Hologic, Inc.,
HOLOGIC 35 Crosby Drive,
Bedford, MA 01730
PROPOSED

REV AUTHORED BY DATE


PROPRIETARY: This document contains proprietary data of Hologic, Inc.
T. WONG 11/30/05 No disclosure, reproduction or use of any part thereof may be made except
by written permission from Hologic.
REV DRAFTED BY DATE TITLE DOCUMENT NUMBER REV
T. WONG 11/30/05 SD, X-RAY TANK 10 DEG TUBE 010-0575-SD 001

REV RELEASE DATE: 12/13/05 SCHEMATIC DIAGRAM SIZE C SHEET 1 OF 1


SCHFILE: A0575S1C.DSN