Calibration Eraser-C-En
Calibration Eraser-C-En
1
Picture 2: Screen display for enter the password for Calibration mode
o The next key sequence is for the first password character, please follow
bellow steps:
Press and hold Power Down key.
Still keep and hold the Power Down, then press and Power Up key.
Still keep and hold the Power Down and Power Up key, then press
and hold the CW key.
Still keep and hold the Power Down and Power Up, CW key, then
press and hold the SP key.
2
o After the correct sequence, the screen will display * character, this means the
first of the password character is correct, see the below:
o Continue the key sequence is for the second password character, please
follow bellow steps:
Press and hold Power Down key.
Still keep and hold the Power Down, then press and Power Up key.
Still keep and hold the Power Down and Power Up key, then press
and hold the CW key.
Still keep and hold the Power Down and Power Up, CW key, then
press and hold the SP key.
3
o After the correct sequence, the screen will display ** character, this means
the second of the password character is correct, see the below:
4
o After the correct sequence, the screen will display *** character, this means
the third of the password character is correct, see the below:
5
o After the correct sequence, the screen will display the Calibration mode,
this means ready for service engineer to user the Calibration mode, see the
below:
6
o In the Calibration mode, if press the UP key for calibration the energy of the
UP mode from 1w to 7w.
o In the Calibration mode, if press the AIM key for level calibration of the
aiming beam from level 1 to level 9.
- With every parameters we have two columns, one for system display on the LCD
screen, one is for EEPROM data storage, which used for control the energy of the
laser. For example, with the CW mode will be had two columns, one is selectable
from 0.5w to 30w for the energy display, two is for data of every display levels.
- In Calibration mode, move the flashing cursor by Power key, then increase or
decrease the current value by Power Up or Power Down key.
7
3. Example 1 for calibration the energy in UP mode:
a. First, calibration the energy for UP mode:
Form the first screen of Calibration mode in picture 6, follow below steps for
energy calibration of the UP mode:
o Press UP key to select the Calibration for the UP mode, see the picture 7:
Picture 7: the flashing cursor is appointed at the display column which used
to display on the LCD of the system.
o In the UP calibration screen, there are three columns for UP calibration,
including:
Column 1, for power display on the LCD screen from 1w to 7w of the
laser.
8
Column 2, for data to control the laser which store in the EEPROM of
the system for the display energy, see picture 8:
9
Column 3, for data to control the pulse width of the UP pulse which
also store in the EEPROM to control the time of the pulse, see picture
9:
10
b. Example 2 for the Aiming level calibration:
Same to the UP energy calibration, the aiming levels calibration will be the
same way. There are two columns for this aiming calibration, one is for the
aiming level display from 0 to 10, one is the data stored in the EEPROM for
each level.
o In the Calibration mode, press the AIM key to enter the aiming calibration:
see the picture
11
Picture 11
Reference table for the aiming levels and it data:
Column 1: Level Columns 2: Data
0 1200
1 1205
2 1210
3 1215
4 1220
5 1225
6 1230
7 1235
8 1240
9 1235
12
4. Same to the CW mode calibration: please check the reference table for the CW
calibration mode below:
Column 1: Power Columns 2: Data
0.5 55
1.0 62
1.5 69
2.0 76
2.5 83
3.0 90
3.5 97
4.0 104
4.5 111
5.0 118
5.5 125
6.0 132
6.5 139
7.0 146
7.5 153
8.0 160
8.5 167
9.0 174
9.5 181
10 188
11 188
… …
15 244
… …
20 314
… …
25 383
… …
30 454
5. Notes:
- Very careful when using this Calibration mode, only the service engineer who
finished the MEDITECH training for the CO2 laser model, who had a clear and
13
experience about laser service. Otherwise, every mistake in this calibration will
make the system work not correct and failure.
- With any question, please send the e-mail to the MEDITECH in case you not clear
about the calibration.
- The reference table for in this document is not exact and not same from system to
system, it will have some differences for every system be case of the CO2 glass
tube, the laser power specification …
(End of document)