RAPIDLab® 348EX Specification

Nama Dagang/ Merk : RAPIDLAB 348EX Blood Gas System and Accessories
Tipe/Ukuran/ Kode Produk : 11065004
Ijin Edar : AKL 20101310763
Nama Pendaftar : PT. SIEMENS INDONESIA
Masa Berlaku Ijin Edar : 1 November 2022

RAPIDLab® 348EX adalah Instrumen untuk pemeriksaan gas darah dan elektrolit pada pelayanan diagnostik.

System Specifications
Dimensions : 38.2 (h) x 38.5 (w) x 35.3 (d) cm
Weight (with bar-code scanner) : 9.8 Kg
Display Interface : Intuitive, icon-based color, Touch Screen
Storage Capacity / Memory : Up to 250 patient test records & Up to 90 results for each level of QC
Data Export : Via integrated USB port/flash drive to PC, or direct to LIS/HIS or RAPIDComm Data Management System
Connectivity Options/Interfaces
# Serial : RS232, LIS1, LIS2, LIS3 protocol
# Bi-Directional : Can be set up for monitoring/control and connectivity with the RAPIDComm Data Management System
# USB Port : Standard USB 2.0

Sample Volume :
# Syringe : 95 uL
# Capillary : 95 uL
# Microsample: : 50 uL

Sample Type : Whole blood & Dialysate fluid


Analysis Time : Approximately 60 seconds to result
Calibration : Automatic or on demand

Parameters : Measured Parameters Calculated Parameters


Parameter Measurement Range Unit of Measure Parameter Measurement Range Unit of Measure
pH : 6.001 – 8.000 pH units O2 Sat : 0 -100 %
pCO2 : 5.0 – 250.0 mmHg O2 CT : 0.0 - 40.0 mL/dL
pO2 : 0.0 – 749.0 mmHg HCO3 act : 0.0 - 60.0 mmol/L
Na+ : 80 - 200 mmol/L HCO3 std : 0.0 - 60.0 mmol/L
K+ : 0.50 - 9.99 mmol/L ctCO2 : 0.0 - 60.0 mmol/L
Ca++ (*) : 0.20 - 5.00 mmol/L BE b : - 29.9 - 29.9 mmol/L
Cl- (*) : 40 –160 mmol/L BE ecf : - 29.9 - 29.9 mmol/L
Hct : 12 – 75 % pO2 (A-a) : 0.0 -749.0 mmHg
Note : (*) Choose either one pO2 (A/a) : 0.00 - 1.00 mmHg
pO2 FiO2 : 0.00 - 5.00 mmHg
Power Requirements Ca++ (7.4) : 0.20 - 5.00 mmol/L
# Voltage : 100V (85–110V), 120V (102–132V) Anion Gap : - 60 - 60 mmol/L
: 220V (187–242V), 240V (204–264V) pH (T) : 6.001 – 8.000 pH units
# Power : 80 VA H+ (T) : 10.0 - 997.0 mmol/L
# Frequency : 50/60 Hz pCO2 (T) : 5.0 – 250.0 mmHg
pO2 (T) : 0.0 – 749.0 mmHg
Environmental pO2 (A-a) (T) : 0.0 -749.0 mmHg
# Temperature : 15°C - 32°C pO2 (A/a) (T) : 0.00 - 1.00 mmHg
# Humidity : 5%–85% non-condensing ct(Hb) est : 2.0 - 25.0 g/dL
# Barometric pressure : 400–825 mmHg

Approvals : UL, CSA, IEC, EN-61010


: CE Marked with full CB Scheme and all National Deviations.
: Complies with the IVD directive

Accessories :
UPS : APC Brand with Capacity 1 KVA or equivalent

Note: (**) Instrument can be managed using a middleware solution either RAPIDComm (closed system middleware) or POCcelerator (open connectivity middleware).

Run critical care tests reliably and economically in low-volume settings