CC Quality Control Transes
CC Quality Control Transes
Spread:
o Range
o Standard deviation (SD)
o Coefficient of variation (CV)
1. Range
largest value in the data minus the smallest
value, which represents the extremes of data
one might encounter
Example:
For the sample 5, 4, 6, 5, 3, 7, 5,
ANSWER: 7-3 = 4
2. Standard deviation
Is a measure of the dispersion of values ANSWER
from the mean. SD = 1.29
Helps describe the normal curve
A measure of the distribution range 3. Coefficient of Variation (CV)
Most frequently used in measure of is a percentile expression of the mean, an index
variation of precision
used to determine where errors happened
SD
CV = X 100
Mean
1st: Determine the variance
Variance: SHAPE
Represents the difference between each value Gaussian distribution
and the average of the data. Occurs when the data set can be accurately
A measure of variability described by the SD and the mean.
Standard deviation squared Population probability distribution is symmetric
about the mean.
QUALITY CONTROL
“bell curve”. Allows identification of analytical errors within
Total area under the curve : 1.0 or 100% one week cycle.
2. Interlab QC (External QC)
Involves proficiency testing programs that
periodically provide samples of unknown
concentrations to participating clinical
laboratories
68-95-99 Rule Important in maintaining long term accuracy of
the analytical methods
Used to determine state of the art
interlaboratory performance
The College of American Pathologist (CAP)
68% of all measurements should fall within area proficiency program is the gold standard of
clinical laboratory external QC program.
Conduct of the External QC
A series of unknown samples are sent to the
laboratory from the reference laboratory or
authorized program provider
95% of all measurements should fall within area
Unknown samples must be tested by the
laboratories who regularly perform analysis of the
patient specimen using the same reagents &
equipment for actual patient specimens, and the
results are submitted to the program provider, as
soon as every analysis is done.
99% of all measurements should fall within area Analysis of the unknown samples should be treated
The “68–95–99 Rule” summarizes the above relationships like atient specimen to determine the true essence
between the area under a Gaussian distribution and the SD. of accuracy
INTERPRETATION of Proficiency testing
Bi-modal Distribution Each analyte has a define performance criteria
The “bi” in bimodal distribution refers to “two” (example, +/- 3SD for peer mean), where
and modal refers to the peaks. laboratories using the same method are evaluated
The two peaks in a bimodal distribution also by comparing within the group
represent two local maximums; these are points In external QC, difference of greater than 2SD in the
where the data points stop increasing and start results indicates that a laboratory is not in
decreasing. agreement with the rest of the laboratories
Quality Control included in the program
Look at methods so that we produce a If incase the clinical laboratory failed to identify or
confident inducing result. resolve an error or discrepancy in the test process,
1. Intralab QC (Internal QC) the facility is at risk of continuous operation and
Involves the analytes of control samples may be recommended for closure.
together with the patient specimen Rationale of the External QC/ Proficiency testing
Important for daily monitoring of accuracy & Ultimate goal:
precision of analytical mtds. To ensure our clinicians that patient results are
Detects both random & systematic errors in accurate
daily basis
QUALITY CONTROL
Allows each laboratory to compare and These are expected values represented by intervals
evaluate test results/outcomes with those of acceptable values with upper and lower limits
laboratories that use the same method. Expected control values are within the desired
control limits , the clinicians are assured that the test
results are accurate & precise
Objectives of QC: Control limits: calculated from the mean & SD
1. To check the stability of the machine Linear Regression Analysis
2. To check the quality of reagents a technique that attempts to link 2 correlated
3. To check technical (operator) errors variables x & y in the form of a mathematical
Gives us confidence of our results. equation, such that given value of one variable the
Control Solutions (QC Materials) other may be predicted.
General Chemistry assays used 2 levels of control correlation coefficient (r) : is a measure of the
solutions, while immunoassays uses 3 levels strength of the relationship between the two
Establish statistical quality control on a new methods.
instrument or a new lot numbers of control
materials, the different levels of control material Comparison of Method
must be analyzed for 20days. involves measuring patient specimens by both
For highly précised assays, ( with the CV less than an existing (reference) method and a new (test)
1%) such as blood gases, analysis for 5 days is method
adequate. The data obtained from these comparisons
consist of two measurements for each patient
specimen.
It is easiest to visualize and summarize the
paired-method comparison data graphically