Quick Guide Zeiss Callisto Eye 3.5 Matching The Reference Image
Quick Guide Zeiss Callisto Eye 3.5 Matching The Reference Image
5 matching the
reference image
vs.
• ZEISS CALLISTO eye® aligns the microscope image (colored picture) according to the reference picture (black & white,
captured by the ZEISS IOLMaster ®) regarding the position of the scleral blood vessels. Fine adjustment has to be done by the
user after successful matching to ensure reliable assistant functions.
• The good quality of the reference picture is the starting point of a successful toric IOL alignment. If the reference
picture is not in good quality, repeat the ZEISS IOLMaster measurement!
Carl Zeiss Meditec AG, Sylvia Andrea Schwarz, Learning Management MED EN_32_012_0009I 2019-09-03 2
Quick Guide - ZEISS CALLISTO eye 3.5 matching the reference image
Capture the IOLMaster reference picture
• Advise the patient to blink once before starting the measurement • Lids too close
to ensure a proper tearfilm keratometric values of anterior • Defocussed
• Nasally disturbing
corneal surface is measured again while capturing the reference light – scleral
picture. Use artificial drops if patient has tearfilm of bad quality blood vessels not
visible at all!
• Eyes opened widely – if patient cannot open the eyes widely by • Lids too close
his own (e.g. due to Ptosis), lift the upper lid by using a Q-tip. Be
carefull to not squeeze on the cornea while lifting the upper lid,
otherwise keratometric values will be falsified
• Too dark
Carl Zeiss Meditec AG, Sylvia Andrea Schwarz, Learning Management MED EN_32_012_0009I 2019-09-03 3
Quick Guide - ZEISS CALLISTO eye 3.5 matching the reference image
Capture the IOLMaster reference picture
• The green arrows, the traffic light and the color code help you to find the right
measurement position for focus & centration
• Check the image quality directly after the ZEISS IOLMaster measurement: focus,
centration, sharpness & the amount of scleral blood vessels. Repeat the
measurement if needed! defocussed
• The good quality of the reference picture ensures your success of the
matching on ZEISS CALLISTO eye!
decentered
Carl Zeiss Meditec AG, Sylvia Andrea Schwarz, Learning Management MED EN_32_012_0009I 2019-09-03 4
Quick Guide - ZEISS CALLISTO eye 3.5 matching the reference image
Microscope requirements
a. White balance test was performed successfully on a piece of white paper *b)
b. Microscope head is not tilted above the patients eye – 2 small black lines
are coincident, if not adjust over the big black knob*
SCI 100%
FF 30-50%
Carl Zeiss Meditec AG, Sylvia Andrea Schwarz, Learning Management MED EN_32_012_0009I 2019-09-03 5
Quick Guide - ZEISS CALLISTO eye 3.5 matching the reference image
Check availability and quality of the reference image on CALLISTO eye
Carl Zeiss Meditec AG, Sylvia Andrea Schwarz, Learning Management MED EN_32_012_0009I 2019-09-03 6
Quick Guide - ZEISS CALLISTO eye 3.5 matching the reference image
Preparation Phase
1
1. Select the patient
• Treatment position
• Apply ZEISS IOLMaster data
• Incisions & LRIs: size & position
• Rhexis size
• Z ALIGN: axis for toric IOL
alignment
Carl Zeiss Meditec AG, Sylvia Andrea Schwarz, Learning Management MED EN_32_012_0009I 2019-09-03 7
Quick Guide - ZEISS CALLISTO eye 3.5 matching the reference image
Centering of the eye - Parfocality
3. Maximum magnification
Carl Zeiss Meditec AG, Sylvia Andrea Schwarz, Learning Management MED EN_32_012_0009I 2019-09-03 8
Quick Guide - ZEISS CALLISTO eye 3.5 matching the reference image
Centering of the eye - Parfocality
Carl Zeiss Meditec AG, Sylvia Andrea Schwarz, Learning Management MED EN_32_012_0009I 2019-09-03 9
Quick Guide - ZEISS CALLISTO eye 3.5 matching the reference image
Requirements before the matching
*The video image section can be checked by means of the video frame of the integrated
data injection system. The video frame can be set under “Lumera 700 configuration –
Displays – additional settings”. The frame is then visible in the eye piece of the microscope
oculars.
Attention: The frame will not be visible on the video screen! This picture is just an example!
Carl Zeiss Meditec AG, Sylvia Andrea Schwarz, Learning Management MED EN_32_012_0009I 2019-09-03 10
Quick Guide - ZEISS CALLISTO eye 3.5 matching the reference image
Matching the reference image - markerless
Markerless option:
Attention -
matching can fail in case of
chemosis or bleeding!
Carl Zeiss Meditec AG, Sylvia Andrea Schwarz, Learning Management MED EN_32_012_0009I 2019-09-03 11
Quick Guide - ZEISS CALLISTO eye 3.5 matching the reference image
Alignment of the microscope image
Carl Zeiss Meditec AG, Sylvia Andrea Schwarz, Learning Management MED EN_32_012_0009I 2019-09-03 12
Quick Guide - ZEISS CALLISTO eye 3.5 matching the reference image
Alignment of the microscope image
Check position of
the blood vessels
• After pressing the button for the matching, the software is The two image
prealigning the microscope image to the reference image by modes
rotation until the scleral blood vessels are on the same position “side_by_side” and
“overlay” help you
• The fine adjustment has to be done by the user by using the for the alignment
arrow buttons
Carl Zeiss Meditec AG, Sylvia Andrea Schwarz, Learning Management MED EN_32_012_0009I 2019-09-03 13
Quick Guide - ZEISS CALLISTO eye 3.5 matching the reference image
Alignment of the microscope image
Carl Zeiss Meditec AG, Sylvia Andrea Schwarz, Learning Management MED EN_32_012_0009I 2019-09-03 14
Quick Guide - ZEISS CALLISTO eye 3.5 matching the reference image
Matching the reference image - markerbased
Markerbased option:
Attention –
Manual marking of the reference axis
works on CALLISTO eye only if the
eye of the patient was previously
marked with blue ink!
Carl Zeiss Meditec AG, Sylvia Andrea Schwarz, Learning Management MED EN_32_012_0009I 2019-09-03 15
Quick Guide - ZEISS CALLISTO eye 3.5 matching the reference image
Matching the reference image - markerbased
Markerbased option
Carl Zeiss Meditec AG, Sylvia Andrea Schwarz, Learning Management MED EN_32_012_0009I 2019-09-03 16
Quick Guide - ZEISS CALLISTO eye 3.5 matching the reference image
Treatment Phase
Carl Zeiss Meditec AG, Sylvia Andrea Schwarz, Learning Management MED EN_32_012_0009I 2019-09-03 17
Quick Guide - ZEISS CALLISTO eye 3.5 matching the reference image
Do’s & Don’ts
Do’s Don’ts
Reference picture:
Reference picture: • Eye lids too narrow
• Eye wide open • Scleral blood vessels defocused
• Homogeneous • Nasal disturbing light reflections
illumination
• Focus on scleral
blood vessels
Live image:
Carl Zeiss Meditec AG, Sylvia Andrea Schwarz, Learning Management MED EN_32_012_0009I 2019-09-03 18
Quick Guide - ZEISS CALLISTO eye 3.5 matching the reference image
FAQs
Advise the patient to blink once before starting the measurement and to open the eye widely. Check the image quality directly
…..after the ZEISS IOLMaster measurement and repeat the measurement if needed.
The IOLMaster did not export the reference picture on the USB stick
• Always use the empty USB stick that was delivered with the device for the export
• ZEISS IOLMaster 700: Click on to open the settings. Go to “Advanced settings” “Export” and choose “File” for DICOM
export. Clicking on “Export” after performing the measurement safes the printout and the reference picture now to the USB stick.
• ZEISS IOLMaster 500: Click on button after measuring the reference picture to export the image & the ZEISS
IOLMaster report to the stick. Do not use the USB stick symbol from the tool bar
Carl Zeiss Meditec AG, Sylvia Andrea Schwarz, Learning Management MED EN_32_012_0009I 2019-09-03 19
Quick Guide - ZEISS CALLISTO eye 3.5 matching the reference image
FAQs
Which button do I have to press to import the patient from USB stick or from Forum?
• Click on “Import patient” in the patient administration screen. A menu opens:
Overlays are not visible on the screen and in the oculars after successful matching of the reference picture
• Confirm the alignment of the microscope image always by pressing the “OK” button. Make sure that the data injection system is
activated:
Export of patient data from FORUM to ZEISS CALLISTO eye is not possible
• Make sure that you only marked the ZEISS IOLMaster report you want to export to ZEISS CALLISTO eye
Carl Zeiss Meditec AG, Sylvia Andrea Schwarz, Learning Management MED EN_32_012_0009I 2019-09-03 20
Quick Guide - ZEISS CALLISTO eye 3.5 matching the reference image
Documents
2. Quick guide ZEISS IOLMaster 500 – option reference image: QUG IOLMaster 500 option reference image.pdf
(000000-1865-713-KurzGA01-GB-040713)
Carl Zeiss Meditec AG, Sylvia Andrea Schwarz, Learning Management MED EN_32_012_0009I 2019-09-03 21