Workstation Guide
Workstation Guide
Workstation Guide
OpenLAB CDS Workstation
Agilent Technologies
Notices
© Agilent Technologies, Inc. 2016-2017 Warranty receive no greater than Restricted Rights as
defined in FAR 52.227-19(c)(1-2) (June
No part of this manual may be reproduced
The material contained in this docu- 1987). U.S. Government users will receive
in any form or by any means (including
ment is provided “as is,” and is sub- no greater than Limited Rights as defined in
electronic storage and retrieval or transla-
tion into a foreign language) without prior
ject to being changed, without notice, FAR 52.227-14 (June 1987) or DFAR
agreement and written consent from Agi- in future editions. Further, to the max- 252.227-7015 (b)(2) (November 1995), as
lent Technologies, Inc. as governed by imum extent permitted by applicable applicable in any technical data.
United States and international copyright law, Agilent disclaims all warranties,
laws. either express or implied, with regard Safety Notices
to this manual and any information
Manual Part Number contained herein, including but not C AU T I O N
limited to the implied warranties of
M8413-90002 merchantability and fitness for a par- A CAUTION notice denotes a
ticular purpose. Agilent shall not be hazard. It calls attention to an
Edition liable for errors or for incidental or
consequential damages in connection
operating procedure, practice, or
06/2017
with the furnishing, use, or perfor- the like that, if not correctly per-
Printed in Germany
mance of this document or of any formed or adhered to, could
Agilent Technologies information contained herein. Should result in damage to the product
Hewlett-Packard-Strasse 8 Agilent and the user have a separate
76337 Waldbronn written agreement with warranty
or loss of important data. Do not
terms covering the material in this proceed beyond a CAUTION
document that conflict with these notice until the indicated condi-
terms, the warranty terms in the sep- tions are fully understood and
Software Revision arate agreement shall control.
met.
This guide is valid for revision 2.2 of Agilent
OpenLAB CDS. Technology Licenses
The hardware and/or software described in WA R N I N G
this document are furnished under a
license and may be used or copied only in A WARNING notice denotes a
accordance with the terms of such license. hazard. It calls attention to an
Restricted Rights Legend operating procedure, practice,
If software is for use in the performance of
or the like that, if not correctly
a U.S. Government prime contract or sub- performed or adhered to, could
contract, Software is delivered and licensed result in personal injury or
as “Commercial computer software” as
defined in DFAR 252.227-7014 (June 1995),
death. Do not proceed beyond a
or as a “commercial item” as defined in WARNING notice until the indi-
FAR 2.101(a) or as “Restricted computer cated conditions are fully
software” as defined in FAR 52.227-19
(June 1987) or any equivalent agency regu- understood and met.
lation or contract clause. Use, duplication
or disclosure of Software is subject to Agi-
lent Technologies’ standard commercial
license terms, and non-DOD Departments
and Agencies of the U.S. Government will
Term Description
3 Licensing
This chapter provides basic information on licensing. It describes how you
generate a license file with SubscribeNet and install the license in the Control
Panel.
5 Optional Procedures
This chapter describes the installation or upgrade of additional instrument
driver software. It also contains information on the installation of OpenLAB
Help and Learning only, and on performance improvement on offline
machines.
Contents
3 Licensing 35
About OpenLAB CDS Licensing 36
Get a License 38
Install Your License 42
5 Optional Procedures 55
Install OpenLAB Help and Learning Only 56
Improve Performance on Offline Machines 57
10 Appendix 103
Hardware and Software Requirements 104
Network Requirements 108
Operating System Configuration 113
Instrument Connections 127
Software Compatibility 147
Privileges in the Control Panel 149
Sales and Support Assistance 157
1
Install OpenLAB CDS Workstation
Installation Workflow Overview 8
Before you Begin 9
Run the OpenLAB Installer 10
Silent Installation 22
Export Properties File 22
Run Installation 23
Parameters and Return Codes 24
Logging and Tracing 24
Install or Upgrade Driver Software 25
Register driver software with OpenLAB CDS 26
Agilent Technologies 7
1 Install OpenLAB CDS Workstation
Installation Workflow Overview
1 Insert the USB media, right-click the setup.exe file, and run it as
administrator. Alternatively, copy the content of the USB media to a
network share, and run the setup.exe file from there.
If User Account Control (UAC) is switched on, this step requires active confirmation to
NOTE continue.
If .NET 3.5 cannot be enabled, for example, because the computer has no internet access,
NOTE install .NET 3.5 from the Windows installation media (see Method 3 under
https://support.microsoft.com/en-us/kb/2734782). If you do not have installation media,
create them as described under
http://windows.microsoft.com/en-US/windows-8/create-reset-refresh-media?woldogcb=0.
4 Click Install/Upgrade.
10 Storage Configuration: Provide a folder for storing the project data. Do not use
the root folder of any drive.
If you want allow access to the projects root path only from within
OpenLAB CDS, select Secure 'Projects Root Path'.
1
To run the site preparation tool separately before installing: Start the OpenLAB Installer, select
the Planning page, and click System Configuration Checker.
12 Review: All components that will be installed are listed with their version
numbers.
• To save a properties file for a future silent installation (see “Silent
Installation” on page 22), click Save to config File.
• To start the installation, click Install.
15 Finish:
• To confirm everything has been installed correctly, click Run Software
Verification1.
• To complete the installation, select the Reboot the computer now check box,
and click Finish.
16 In case of errors during the installation: Check the installation log files
under C:\ProgramData\Agilent\InstallLogs\[date and time]. Note that ProgramData
is a hidden folder.
The installation includes a set of standard instrument drivers. If you need
other instrument driver software, install it in a separate step. See “Install or
Upgrade Driver Software” on page 25.
1
To start the tool separately at a later point in time, select Start > Agilent Technologies > Software
Verification Tool.
Silent Installation
Run Installation
Prerequisites • You have prepared a properties file for silent installation. See “Export Properties
File” on page 22.
• .Net Framework is present on your system.
If it is not present, it will automatically be installed. You must then manually select
Accept to agree with the license agreement.
Return Codes
After installation, uninstallation, upgrade, or repair in the command-line
mode, the system will return a number code which is explained below.
Failure. Verify against the log file to see what any other number
failed.
3 Click Connect to enable the input boxes for the user credentials.
4 Under Step 2 - Authentication, enter your user credentials.
5 Make sure the Register as Instrument Controller check box is cleared.
6 Click Register.
2
Post Installation Tasks
Set Account to Enable Automatic Printing 28
Configure the Antivirus Program 29
Configure Internet Explorer for OpenLAB Help and Learning 32
Local Windows Account for Secure Projects Root Folder 32
Disable Windows 10 Updates 33
This chapter describes tasks that are relevant after finishing the installation.
Agilent Technologies 27
2 Post Installation Tasks
Set Account to Enable Automatic Printing
OpenLAB CDS allows you to process your data already during the acquisition,
without opening Data Analysis. During this automatic processing you can also
generate reports and print them to printer or save them as files. Exporting the
reports to a network share is a typical way how they are sent to an external
system (for example, LIMS).
To access a printer, and also to access a network share, Data Analysis must be
started from an account with the required configuration and privileges.
1 Create a domain user or local user with access to the relevant printer and
network share.
2 On the Workstation, add the new user to the local Administrators group.
3 Log in as the new user, and configure the relevant printer as this user's
default printer.
4 Change the Agilent OpenLAB Instrument Service to run under this user.
a In the Windows command line, run services.msc as a user with
administrative rights.
b In the Services window, right-click the Agilent OpenLAB Instrument Service,
and select Properties.
c On the Log On tab, select This account, and enter the credentials of the new
user.
5 Reboot the PC.
1 Be sure to open the firewall ports listed in the Firewall Settings in the
OpenLAB CDS Requirements guide.
2 For best performance, consider the following folder exclusions. These
folders should only be scanned while the instruments are idle, and no data
acquisition or data analysis takes place.
• [C:\]CDSProjects
• [C:\]Program Files (x86)\Agilent Technologies
• [C:\]ProgramData\Agilent
• [C:\]ProgramData\Agilent IPB Files
• [C:\]ProgramData\Agilent Technologies
• [C:\]ProgramData\ChromatographySystem
• [C:\]ProgramData\Firebird
• [C:\]ProgramData\IsolatedStorage
Refer to your specific antivirus software documentation on how to
configure folder exclusions.
If you installed OpenLAB CDS using the Secure 'Projects Root Path' option, your
projects folder (by default, C:\CDSProjects) and anything within this folder is
locked. Only the local Windows user AgtSfsUser and the local Windows group
AgtSfsGroup are able to browse this folders. They have been created during
installation.
To ensure that the system is working properly:
• Do not delete this group and account, and
• Do not disable Windows logins for local accounts.
Your company's security policy may require that Windows updates not be
automatically applied. Updates may need to be tested first, and then be
distributed internally.
With Windows 10, automatic updates cannot be turned off in the Windows
settings as in earlier Windows versions. Instead, you must disable the
Windows Update service. As this service is required during installation, you
can only disable it after finishing the installation.
1 In the Start menu, search for services.msc and press Enter to open the Services
window.
2 Double-click the Windows Update service.
3 Set the startup type to Disabled.
The computer will not be automatically updated anymore. Make sure you keep the
NOTE computer up to date by other means.
3
Licensing
About OpenLAB CDS Licensing 36
Software subscriptions and Software Maintenance Agreement
(SMA) 36
License Types 37
License File 37
Get a License 38
Obtain a License with SubscribeNet 38
Other Ways to Obtain a License 40
Install Your License 42
Agilent Technologies 35
3 Licensing
About OpenLAB CDS Licensing
License Types
The license file is a collection of Product, Instruments and Add-on licenses (or
activation keys), and is installed to your OpenLAB CDS System. Both the
OpenLAB Workstation PC, or the OpenLAB Server in a Client/Server system
will act as the license server.
The licenses or activation keys in the license file can either be Shared or
Counted:
• Shared licenses — system computers and other components can have
shared, or add-on, licenses — because they share a core license.
• Counted licenses — these licenses are part of the OpenLAB CDS floating
licensing strategy. They are not permanently assigned to any one
component. Instead they are automatically assigned to components, such as
AICs and instruments, while the components are starting up. The licenses
are automatically returned when the component is closed. The license
management program controls license issuance and retrieval.
In this case, the only requirement is that a component is licensed while
running. You only need enough licenses for all components running
concurrently, rather than for each installed component.
A startup license for the system allows you to run OpenLAB CDS for 60 days
after the installation. In order to run the data system software after the 60-day
period, you must install your license file.
License File
A license file will contain your software license. This file is installed on the
workstation. The license file is bound to this computer, and cannot be moved
to another workstation without regenerating the license in SubscribeNet.
Information in the license file defines the number of instruments and other
options that may be used concurrently with your system.
The most efficient way to manage and maintain your licensing is through the
Internet.
Get a License
Prerequisites To generate, download, and install a final license for your product, you will need:
• The authorization code label provided in the lavender envelope containing your
Software Entitlement Certificate.
If you have not received a lavender envelope for your product, contact your vendor or
internal support.
• The URL for SubscribeNet from the Software Entitlement Certificate.
• The host name of the computer where the Control Panel is running.
• The MAC address.
To retrieve your MAC address from a computer where OpenLAB CDS is already
installed, open the Control Panel and browse to the Administration > Licenses section.
Use the Copy MAC Address or Save MAC Address function to obtain the MAC address
for license generation.
During this process you will have to enter the MAC address of your license server. For
workstations, this is the local computer. For client/server systems, this is the server.
If any changes are made to the computer name or domain reference after the license is
NOTE installed, remove the license. A new license will need to be created in SubscribeNet,
downloaded, and installed.
If the network adapter that provides the MAC address used during license creation is
NOTE removed from the machine, your license will no longer be valid. A new license will need to
be generated with a currently available MAC on the license server.
New Users
1 From a computer with Internet access, enter the URL provided in the
Software Entitlement Certificate in an Internet browser
(https://agilent.subscribenet.com/control/agil/AgilRegisterToAccount).
2 On the registration page, enter the authorization code from the label and
complete the profile information (required fields are marked with an
asterisk *).
The email address you enter will become your login ID.
3 Click Submit. The system will generate and display an account name for you.
SubscribeNet will send a welcome email with your login ID and password.
4 Log in to SubscribeNet using your login ID and password.
Once you log in, you can use the online user manual link for help with any
questions you have.
5 Select Generate or View licenses from the left navigation bar.
6 Follow the prompts to generate your new license.
You will be prompted for the HOST NAME of the computer. The host name
you enter must match with the network name of the computer where the
Control Panel is running. Do not include any DNS suffix (domain.com)
references in the entered machine name.
7 When the system generated the license, view its details, then click Download
License File. Save the license file to your computer and to a backup location
(such as a portable storage device).
Use your login ID and password when you revisit the Agilent SubscribeNet
site to regenerate a license file, add new authorization codes, or further
configure the license for your system.
Users registered with SubscribeNet
1 Login to SubscribeNet with your e-mail address and password.
2 Select the SubscribeNet account associated with this authorization code, if
you have more than one account.
3 From the SubscribeNet navigation pane, select Register Authorization Code.
This will allow you to enter your new authorization code and make available
the new license entitlements
4 Follow steps 7 through 9 in the previous procedure, New Users, to generate
or view your new licenses.
Offline Licensing
If an internet connection is not available in your laboratory:
You or your local on-site service engineer will collect the necessary
information from you to allow Agilent to create a license account on your
behalf. For phone support in your region, call the sales and service number for
your region. See the Appendix for a list of numbers for various countries.
Required Customer Information for Agilent License Support:
The following information must be provided to Agilent in order to enable us to
create a licensing account on your behalf.
1 Collect Account Information:
Your account name will be your company name and Lab name separated by
a comma. Employee information provided here will be used to define the
first administrator of your account for future access to the system as
required. Please prepare the following pieces of information prior to
contacting your local Agilent sales and service center in order to expedite
service:
• Company Name
• Lab/Department Name
• First Name
• Last Name
• E-mail address
• Job Title
• Phone #
• Address, City, State/Province, Postal Code, Country
2 Collect Authorization Code(s):
The authorization code is an alpha-numeric code provided on a label which
is enclosed in a lavender envelope. If you have received more than one code
you must provide all codes to ensure that all ordered licenses are granted to
your account.
3 Receiving your license:
Once the above information is provided Agilent will then work on your
behalf to generate a license file through SubscribeNet. The license file will
either be sent to your shipping address (on a CD), or your local FSE will
deliver it in person (usually on USB media). Once your license is received
follow the below section on “Install your License” to finish installing your
license on your CDS system(s).
The license must be added to your system using the Control Panel.
1 Start the Control Panel shortcut on the desktop or go to Start > All Programs >
Agilent Technologies > OpenLAB Shared Services > Control Panel.
2 Navigate to Administration > Licenses.
4
Configure OpenLAB CDS Workstation
Configure Authentication 44
Configure Security Policy 46
Configure users, roles, and privileges 46
Create or import users 47
Groups 48
Roles and Privileges 48
Add users or groups to a role 49
Specific Roles for Individual Instruments or Projects 50
Configure the Storage Location 51
Enable File System Security 52
Configure Initial Project 52
Configure Initial Instrument 53
Other settings in the Control Panel 53
This chapter describes the initial configuration steps after installing the
software. For more details, refer to the Control Panel section in OpenLAB Help
& Learning.
This chapter is not applicable if you are migrating from OpenLAB CDS ChemStation
C AU T I O N
Edition.
➔ If you are upgrading from OpenLAB CDS ChemStation Edition be sure to refer to
the "OpenLAB Migration Tool user guide" prior to configuring OpenLAB CDS.
Agilent Technologies 43
4 Configure OpenLAB CDS Workstation
Configure Authentication
Configure Authentication
With the authentication provider None, any activity logs will display a generic System
NOTE operator with no additional identification. This is not recommended for compliant setups.
• Internal
In this mode, the user's credentials are stored in the OpenLAB CDS system.
You are asked to create an administrator account for OpenLAB CDS before
setting up other users. This is the only mode in which you can create new
users within the system; in all other modes you can only map to users that
exist in a different system.
• Windows Domain
You import existing Windows users into OpenLAB CDS system. The
authentication is done either by Windows Active Directory domain or NT
4.0 Domain within the Enterprise. OpenLAB CDS only uses the identity and
password of the mapped users; roles and privileges for OpenLAB CDS are
still configured in the Control Panel.
1 Launch the Control Panel.
2 Navigate to Administration.
3 In the navigation pane, select System Configuration.
5 Select the required authentication provider from the drop-down list, then
click Next.
If you need to comply with specific standards, adjust the security policy as
required.
With the authentication provider Internal, you can set all parameters in the
Control Panel. With an external authentication provider (Windows Domain),
you can only set the inactivity time in the Control Panel; all other parameters
are defined by the external system.
1 Launch the Control Panel and navigate to Administration.
2 In the navigation pane, select Security Policy.
3 In the ribbon, click Edit Security Policy.
With internal authentication, you create the required users in the Control
Panel. With Windows domain as an external authentication system, you
import the Windows domain users.
To define what users are allowed to view or do, OpenLAB CDS offers
predefined roles and allows you to define your own specific roles. Roles are
equipped with numerous specific privileges.
Each user can be member of multiple groups. You must assign one or more
specific roles to each group. You can also assign roles to single users; however,
for the sake of clarity, it is strongly recommended to assign roles only on the
group level. Every member of a group automatically has all roles of this group.
1 Launch the Control Panel and navigate to Administration.
2 In the navigation pane, select Users, Groups, or Roles.
3 Create new items, or edit the existing ones.
Groups
If you use an external authentication provider, you can either import the
names of groups that exist in the external system, or create new internal
groups. There is no limit on the number of groups that can be mapped or
created.
Assign users to groups either in the external system or in the Control Panel. If
you need additional user assignments that are relevant only for OpenLAB
CDS, create them in the Control Panel. Otherwise it is sufficient to only import
the groups and assign the required roles to the groups.
If you delete or unmap a group, the users who were members in this group
remain unchanged.
You can also create a tree of different projects or project groups in the Control
Panel, and assign different Project roles for different projects (see also
“Specific Roles for Individual Instruments or Projects” on page 50). For
example, a user can have the role Project Administrator in one project, so that he
can manage the settings in the Control Panel. In a second project, he may have
a role that allows him to edit the content of a project, but not to change the
project settings.
For more information on privileges, see the Appendix.
Administrative privileges These privileges are globally assigned to a user or group and cannot
be changed on the instrument/location level. They are the typical
administration privileges such as Backup and restore, Manage
security, Manage printers etc.
Project privileges Privileges for accessing or modifying different levels of data. You can
assign these privileges globally or on project or project group level.
All project data is stored on the local file system. Each project has its own
project folder, but all project folders are located in the Projects Root Path. The
following procedure explains how to change Projects Root Path.
Prerequisites To be able to change the Projects Root Path, the directory must be empty. You did not yet
create any projects.
If you already created projects, move the data away temporarily, then delete the the
NOTE project. After you have set the correct path, create the projects again and move the data
back.
With File System Security, you prevent users from modifying OpenLAB files
outside OpenLAB CDS. File System Security may have been enabled already
during installation, but you can enable it at any time. File System Security is
optional.
1 Launch the Control Panel.
2 Navigate to Administration.
3 In the navigation pane, select System Configuration.
5
Optional Procedures
Install OpenLAB Help and Learning Only 56
Improve Performance on Offline Machines 57
Agilent Technologies 55
5 Optional Procedures
Install OpenLAB Help and Learning Only
Use this option to install OpenLAB Help and Learning content without
installing OpenLAB CDS applications.
Do not use this option on a machine where OpenLAB CDS is, or will be,
installed.
1 Insert the USB media, right-click the setup.exe file, and run it as
administrator.
2 On the start screen, select OpenLAB CDS, and click OK.
3 In the OpenLAB CDS Installer, click Documentation.
4 Click Install OpenLAB Help and Learning Only.
5 Select your language, and click Next.
6 On the welcome screen, click Next.
7 Confirm Agilent terms and conditions, and click Next.
8 Review the installation directory. If desired, click Change... to specify a
different directory.
9 Click Install.
10 When the installation is complete, click Finish.
11 If you plan to use Internet Explorer to view the content, set the Internet
Options as described under “Configure Internet Explorer for OpenLAB
Help and Learning” on page 32 .
Without these settings, you will need to click Allow blocked content when
opening the help.
No settings are required for Google Chrome.
You can uninstall or repair OpenLAB Help and Learning from the same link in
the installer (see “Uninstall OpenLAB Help and Learning Only” on page 97).
Computers running OpenLAB CDS may exhibit slow performance when they
are not connected to the Internet.
The windows operating system has routines built into its operation that
causes it to continuously search for an online connection in order to update to
all the latest Windows security certificates when using secure software.
Use the following system settings on all workstations to remedy this problem.
1 Open Internet Explorer and select Tools > Internet Options. In the Advanced
tab, clear the following check boxes:
• Security > Check for publisher's certificate revocation
• Security > Check for server certificate revocation
2 Change the following registry keys:
• [HKEY_LOCAL_MACHINE\SOFTWARE\Policies\Microsoft\
SystemCertificates\AuthRoot]
"DisableRootAutoUpdate"=dword:00000001
• [HKEY_LOCAL_MACHINE\SOFTWARE\Wow6432Node\Policies\
Microsoft\SystemCertificates\AuthRoot]
"DisableRootAutoUpdate"=dword:00000001
3 Document that you turned off the Root Certificates, as this can prevent
users from installing other applications.
6
About the OpenLAB CDS Software
Software Architecture 60
Local File System 62
Customization 63
Customization via custom calculations 63
Customization via report templates 63
Customize application to start external programs 64
Agilent Technologies 59
6 About the OpenLAB CDS Software
Software Architecture
Software Architecture
OpenLAB CDS is a data system solution for analytical workflows that controls
a wide variety of instruments including the industry-leading GC and
GC/MS-SQ instruments, along with best-in-class LC and LC/MS-SQ. By
combining chromatography and single-quad mass spectrometry into a single
scalable solution with centralized system administration, you can streamline
your laboratory workflows and maximize productivity. A tailored and
simplified user interface with a new state-of-the-art user experience, along
with e-learning tools, to help you to get up to speed and productive as fast as
possible.
The Agilent OpenLAB CDS software is provided on read-only USB media that
contain all required installables and documents. This includes:
• Acquisition
• Data Analysis and Reporting
• Shared Services
• Content Management
• Custom Calculation Editor
• Help and Learning Platform
• User documentation
• Instrument driver software for Agilent LC, GC, LC/MS, GC/MS, or A/D
• Instrument driver software for virtual instruments (Data Player)
• Agilent Parts Finder
• Third party tools
Workstation: All components on a single PC; results are stored in the local file
system; the system supports up to four instrument connections.
,QWUDQHW,QWHUQHW
:RUN6WDWLRQ RSWLRQDO
:RUNVWDWLRQ3OXV
'DWD$QDO\VLV&OLHQW &RQWURO3DQHO
$FTXLVLWLRQ&OLHQW 6KDUHG6HUYLFHV
,QVWUXPHQW6HUYHU &RQWHQW0DQDJHPHQW
'DWDEDVH
WR,QVWUXPHQWV
Customization
7
System Setup and Maintenance
Control Panel 66
Instrument Management / Lab Status at a Glance 66
License Management 67
System Activity Log 68
Diagnostics 69
Administrative Reports 69
Authentication Provider 70
Security Policy 70
User Management 70
Shared Services Maintenance 71
Maintenance Procedures 73
Introduction 73
Software backup procedure 74
Data Backup Procedure 81
Software Restore Procedure 84
Data Recovery Procedure 85
This chapter contains information on the Control Panel and Shared Services
Maintenance. In addition, it describes various maintenance procedures.
Agilent Technologies 65
7 System Setup and Maintenance
Control Panel
Control Panel
Using the Control Panel, you can access Shared Services control features such
as security policy, central configuration, or lab status at a glance.
License Management
This service includes the administration of all licenses that are required for
your system.
Before adding a license file, you must first purchase the license and generate
the license file using SubscribeNet. For more information on generating new
license files, see “Obtain a License with SubscribeNet” on page 38.
License Management in the Control Panel provides the following functions:
• You can add license files to the license server.
• You can navigate to the license monitor and view the properties of all
licenses installed on a given license server.
• You can remove license files from the license server. This may be useful if an
invalid license file has been added.
• You can view or change the license server.
• You can view, copy, or save the MAC Address of the license server.
• You can navigate to the Agilent Electronic Software and License Delivery
web page to obtain a license.
The messages can come from other components, such as the user management,
or from an instrument module. Instrument messages include error messages,
system messages, or event messages. The System Activity Log records these
events irrespective of whether you have been alerted to them or not. To get
more information on an event, expand the line of interest in the activity
logbook viewer.
Diagnostics
The Diagnostics view allows you to access several reports and tools for
diagnostic purposes:
• Ping the server.
• Create a report, either for the local system or for the server, with
information on the operation system, processors, disk drives, processes,
network, and connections.
• Centrally access and download all the log files, trace files, etc. that are
created by the registered modules.
Administrative Reports
In the Administrative Reports view, you can additionally create and export
various XML or PDF reports related to the system configuration:
Instruments Report
Provides detailed information about configuration and access privileges for all
instruments on the system. On client-server systems, this report includes all
instruments on all Instrument Controllers.
Projects Report
Provides detailed information about configuration and access privileges for all
projects on the system.
System Report
This report provides a consolidated view of the system, which includes all
information about instrument controllers, instruments, projects, roles, users,
and groups.
Authentication Provider
The authentication provider is described under Configure an OpenLAB CDS
File System Workstation. For details, see “Configure Authentication” on
page 44.
Security Policy
The security policy is described under Configure OpenLAB CDS. For details,
see “Configure Security Policy” on page 46.
User Management
The user management is described under Configure OpenLAB CDS. For
details, see “Configure users, roles, and privileges” on page 46.
1
In Windows 7 or 8.1: Start > All Programs > Agilent Technologies > OpenLAB Shared Services >
Shared Services Maintenance
To perform a backup:
1 Specify the backup directory and retention time.
The retention time is used to delete any files older than specified.
2 Click Backup.
The backup is placed in the specified backup directory. Backups older than
the retention time are deleted.
The tool automatically generates filenames for the backup files. Never change these
NOTE filenames, as the tool relies on a specific naming convention.
Maintenance Procedures
Introduction
Disaster Recovery Planning
Prepare a recovery plan for the unlikely case of OpenLAB CDS becoming
inoperable due to a hardware or software failure. This plan must include
information and procedures for completely restoring the operating system, the
software, and data. Make sure that the disaster recovery plan has been tested
and confirmed to be working.
The Disaster Recovery Plan must include the following:
• Hardware information: CPU, Memory, and Hard disk configuration
information.
• Computer identity: Name, IP, domain, URI, etc.
• Computer administrator information: username and passwords for
logging in to the server
• Software information: OS version, Patch level.
• Installation parameters:
• Installation folder
• Project folder
• Installation log files
• Shared Services language
• Installed licenses
• Registered applications
• 3rd party software information: applications and their revisions and install
paths.
• Backup procedures (see “Software backup procedure” on page 74, “Data
Backup Procedure” on page 81)
• Backup media location and organization details.
• Restore procedures (see “Software Restore Procedure” on page 84, “Data
Recovery Procedure” on page 85)
The AutoPlay window displays the first time the drive is connected to the PC. Close the
NOTE AutoPlay window.
• In Windows 8.1 or Windows 10, click File History in the Microsoft Control
Panel, then click System Image Backup.
6 Click Next.
8 If you are using Windows 7, create a system repair disc according to the
Windows instructions.
If you are using Windows 8.1 or Windows 10, use the Recovery program
from the Microsoft Control Panel and create a recovery drive according to
the Windows instructions.
9 Close all remaining windows.
10 Eject the newly created Windows Recovery disc.
11 Disconnect the USB drive.
12 Complete the backup Solution Checklist.
13 Keep the USB drive, Windows Recovery disc, completed Backup Solution
checklist, Backup Guide, and Backup Solution Recovery checklist in a safe
place.
Database Backup
1 Launch the Shared Services Maintenance utility and select the Backup and
Restore tab.
It is recommended that you use a safe location as the backup directory (for example, a
NOTE thumb drive, a network location, or an external storage device).
3 Click Backup.
The backup is placed in the specified backup directory. Backups older than
the retention time are deleted.
Content Backup
1 Copy the contents of the project folder specified during installation into a
safe location (for example, a thumb drive, a network location, or an external
storage device).
If you are using Secure Storage, your current Windows user needs to be a member of the
NOTE AgtSfsGroup local group.
Scheduled Backup
To automate the data backup procedure, you can create a batch file
performing all of the steps above and schedule it to run regularly using
Windows Task Scheduler.
For your convenience, there is an example provided in the installation folder
(for example, C:\Program Files (x86)\Agilent Technologies) under \BackupScripts.
Please examine it carefully and adjust according to your system configuration.
The script needs to be run with administrative privileges and as a user who has access to
NOTE the project folder. If the project folder is secure, the user must be a member of the
AgtSfsGroup local group.
• If you are using Windows 8.1, click Troubleshoot > Advanced Options >
System Image Recovery > Choose a target operating system.
6 In the Select a system image backup screen, Use the latest available system image
is selected by default.
• If the auto-filled information is correct, click Next.
• If no image match is found,
a Select Select a system image and click Next
b Select the appropriate image and click Next.
c Select the date and time and click Next.
7 Select Format and repartition disks and click Next.
8 Click Finish.
9 When the re-image process is complete, restart the system to finish the
restoration.
Content Restore
1 Copy the contents of previously preserved project folder to the location
used at the moment of backup.
2 To restore a system with secured project folder:
a Your current Windows user needs to be a member of the AgtSfsGroup local
group.
b Open an Administrative command prompt (type cmd into the Windows
Start Menu search, right click cmd.exe, then select Run as administrator).
c Execute the following commands:
cd C:\Program Files (x86)\Agilent Technologies\OpenLAB
Services\Server
SecureFolder.exe /group C:\CDSProjects
Incorrect usage of the SecureFolder.exe utility (specifying folder other than the Project
NOTE Root) can harm other applications or even your operating system, and potentially prevent it
from starting up.
Database Restore
1 Launch the Shared Services Maintenance utility and select the Backup and
Restore tab.
In case of a successful restore, the Shared Services service will start automatically.
NOTE
8
Upgrade OpenLAB CDS
License Upgrade 92
Get Upgraded License File 92
Add Upgraded License File to the System 93
Upgrade OpenLAB CDS Workstation to Latest Version 94
Agilent Technologies 91
8 Upgrade OpenLAB CDS
License Upgrade
License Upgrade
If the relevant host name does not appear, you may be managing your
licenses in multiple SubscribeNet accounts. You will need to log into those
accounts to upgrade those workstation licenses.
4 If your license(s) are eligible for an upgrade, you will see the Upgrade All
button. Otherwise you will need to contact your Agilent Sales
Representative to renew your Software Maintenance Agreement (see “Sales
and Support Assistance” on page 157). To proceed with generating your
upgrade license, click the button.
5 On the Upgrade All Licenses for License Host page, review the data, and
confirm by clicking Upgrade All.
This upgrades the license file to the most current version. SubscribeNet will
send you an email with a new license file.
6 Put the new license file on your system (see “Add Upgraded License File to
the System” on page 93.
If you have multiple standalone Workstations, repeat this step for each
individual workstation.
Note that each workstation's MAC address is the file name. This helps
identify the correct license file to import into the workstation's Control
Panel.
9
Uninstall OpenLAB CDS With All of its
Components
Uninstall OpenLAB CDS 96
Uninstall OpenLAB Help and Learning Only 97
Agilent Technologies 95
9 Uninstall OpenLAB CDS With All of its Components
Uninstall OpenLAB CDS
1 Log in as an administrator.
2 In the Microsoft Control Panel, open Programs and Features.
3 To uninstall OpenLAB CDS double-click Agilent OpenLAB CDS.
If you installed OpenLAB Help and Learning only, follow this procedure to
uninstall it.
1 Insert the USB media, right-click the setup.exe file, and run it as
administrator. Alternatively, copy the content of the USB media to a
network share, and run the setup.exe file from there.
If User Account Control (UAC) is switched on, this step requires active confirmation to
NOTE continue.
3 On the Documentation page, select Install OpenLAB Help and Learning Only.
5 Click Remove.
The wizard removes OpenLAB Help and Learning from your system.
10
Appendix
Hardware and Software Requirements 104
General Software Requirements 104
Language Compatibility 105
Disk Space 105
Hardware 106
Number of Instruments 106
Software 107
Network Requirements 108
Introduction 108
LAN Connectivity 108
Operating System Configuration 113
Configure Windows 10 113
Configure Windows 8.1 118
Configure Windows 7 122
Instrument Connections 127
RC.NET Drivers compatible with OpenLAB CDS 127
Agilent LC and LC/MS 129
Agilent GC and GC/MS 139
Other supported Agilent Instruments 143
Third Party Instruments 143
Incompatible Instruments and Modules 145
Software Compatibility 147
Agilent Software 147
Libraries and Databases 148
Privileges in the Control Panel 149
Project Privileges 149
Instrument Privileges 155
Administrative Privileges 156
Sales and Support Assistance 157
.NET framework • .NET 3.5.1 must be enabled on systems running on Windows 8.1,
Windows 10, or Windows Server 2012 R2.
and
• .NET 4.6.2 or above (if needed, it will be installed automatically
by the OpenLAB Installer)
1
The listed antivirus software has been tested and is recommended by Agilent. Check with your Agi-
lent service person in case you want to use other software.
If these tools are not installed on your system, you can install them from the OpenLAB
NOTE CDS Installer.
Language Compatibility
The English version of OpenLAB CDS is validated on Windows English and
Western European language operating systems. Regional settings (locales) can
be adjusted as required. In a Client/Server system make sure regional settings
are the same for the server, AIC and clients.
Localized versions of OpenLAB CDS are supported on localized language
versions of Windows, using default system fonts:
• Chinese: Microsoft YaHei
• Japanese: Meiryo UI
• Brazilian Portuguese: Calibri
OpenLAB CDS can run on other language versions of Windows as well, but the
user interface will be English.
Non-localized instrument drivers are supported and will appear in English
even when running localized versions of OpenLAB CDS.
Disk Space
Disk space requirements should be adjusted based on the number and type of
instruments and archival periodicity. Agilent recommends providing enough
disk space for one year of lab operation, in addition to the operating system
and OpenLAB CDS requirements.
Hardware
Table 5 Tested and recommended hardware configuration for an OpenLAB CDS Work-
station
Item Requirement
Processor speed (CPU) Intel® Xeon E3, 3.2 GHz, 8MB Cache, 4 Core,
Hyper Threading, Intel® vPro TM Technology,
Intel® HD Graphics, or equivalent
Physical memory (RAM) 8 GB
Hard disk 1 x 500 GB SATA 7200 RPM
See Table 4 on page 105
Graphic Resolution 1600 x 900
RS-232 port 1 serial port (certain instruments require a
serial port)
USB port 1 USB port required for installation.
LAN card Integrated Intel® I217LM PCIe GbE Controller
Number of Instruments
With the standard configurations you can configure any number of
instruments that sums up to 4 instrument points per OpenLAB CDS
Workstation.
With the VL configurations you can configure only 1 instrument with up to 2
instrument points per OpenLAB CDS VL Workstation.
Instrument points are an indicator for the data stream that needs to be
managed, and thus also for the required hardware and software. Instruments
count as follows:
Software
Table 6 Supported operating systems for a workstation
Network Requirements
This chapter describes the network requirements that must be met in order to
support the environmental computing needs of an OpenLAB Chromatography
Data System (CDS).
Introduction
OpenLAB CDS systems rely on network infrastructure in order to support the
communication between various system nodes. This communication is based
on standard TCP/IP protocols. In order to provide optimum performance and
uptime, the network must meet design criteria for available bandwidth, IP
address assignment, name resolution and appropriate isolation of the lab
subnet from the corporate network.
LAN Connectivity
When using LAN communications to connect workstations or instrument
controllers to an instrument, use one of these methods:
• Connect via an isolated switch using standard CAT-5 network cabling
• LAN communication hardware should be 100/1000 mbps (or higher) speed
capable.
• NIC teaming1: LAN cards should not be teamed on workstations,
instrument controllers, or clients.
• LAN communication must be on the same subnet as instruments, and
preferably on the same segment.
See the separate driver installation guides for further information regarding vendor specific
NOTE instrument connections. GPIB or RS232 might be required.
1
Network Interface Card (NIC) teaming is also known as Load Balancing and Failover (LBFO)
1
View the items by icon to see a list of all items.
Firewall Settings
If you are using a third party firewall or antivirus software on the network
where OpenLAB CDS is installed, open these firewall ports to allow
communication between the system components of OpenLAB CDS. These
apply to workstations as well as to Client/Server systems as component
communications rely on these communication channels:
Ports 8085-8089: Alternative to port 8084 if that port is in use by another page or
process
Port 8090: Hosts the viewing page of current license grants and consumptions
found in the Control Panel administration interface
Ports 10000-10010: Status and acquisition communication with Agilent 7890 GCs
The OpenLAB CDS installer will automatically open these ports on an enabled
Windows firewall during installation.
Domain Requirements
Domains support the flow of information and user access rights across
machines in the network. This means that all machines and instruments
within the networked OpenLAB CDS system must reside within the same
domain or have the appropriate cross domain trusts to allow name based
communications between all components in the system. In the case of a
workstation installation, domains are only relevant if you are using a Windows
domain-based authentication model. In this case the workstation or client
must always be able to communicate with domain components in order to
function as expected.
The domain name server (DNS) must be able to resolve the IPv4 address of all instrument
NOTE controllers and instruments. Any unresolved instrument controller or instrument will
disrupt the functionality of OpenLAB resulting in errors or delays. IPv6 is not supported
and must be deactivated.
OpenLAB CDS components must not be installed on the same machine as the domain
NOTE controller.
The domain components above host a variety of services and settings that
must be configured appropriately to allow communication across machines.
The following services and settings will need to be configured to fit your
domain. Your internal IT group is responsible for proper configuration of any
custom domain solutions. These include settings for:
• Lookup zones and hostnames
• Group and security policies
• Subnet masks and Virtual LANs
• IP reservation (static or DHCP)
Configure Windows 10
[MUST] 1 System (Microsoft Control Panel)1: Register Windows with Microsoft.
2 File Explorer Options (Microsoft Control Panel): In the View tab:
• Select Always show menus.
• Select Display the full path in the title bar.
• Clear Hide extensions for known file types.
• Clear Use Sharing Wizard.
[MUST] 3 Start > Settings > Update and Security:
a Click Check for updates to check for updates and apply all critical security
patches.
Before proceeding, ensure that all updates are downloaded and installed.
Ensure that there is no reboot pending.
b Click Advanced options.
c Select the Defer feature updates check box.
d Click Choose how updates are delivered.
e Turn off Updates from more than one place.
[MUST] 4 Settings for updates: Windows Update service MUST NOT be running during
installation.
[MUST] 5 Indexing Options (Microsoft Control Panel): Disable indexing.
Click the Modify button. Clear all drives and locations.
[MUST] 6 Start > Settings > System > Tablet Mode: For When I sign in, select Use desktop
mode.
1
View the items by icon to see a list of all items.
If this procedure does not work as expected, or the computer has no internet access,
NOTE install .NET 3.5 from the Windows installation media (see details for Windows 10 under
https://support.microsoft.com/en-us/kb/2734782). If you do not have installation media,
create them as described under
https://www.microsoft.com/en-us/software-download/windows10.
c To make sure that all the net.tcp components are properly initialized,
Non-HTTP activation must be enabled. Expand the .NET Framework 3.5
(includes .NET 2.0 and 3.0) node and select the Windows Communication
Foundation Non-HTTP Activation check box.
d Select the .NET Framework 4.6 Advanced Services check box. Use the default
values for sub items.
e Select the Internet Explorer 11 check box.
f Select the Telnet Client check box.
g Select the TFTP Client check box.
h Reboot the PC.
[MUST] 14 Start > Settings > System > Default Apps: Select Internet Explorer as default
Web browser.
Do not use Edge, it is not supported. Unpin Edge from the task bar.
[MUST] 15 Disable Compatibility View in Internet Explorer.
a Open Internet Explorer.
b Click the Tools menu, and then click Compatibility View Settings.
c Clear the Display intranet sites in Compatibility View check box.
[MUST] 16 Enable the navigation pane:
Open Windows Explorer, then select View > Navigation pane from the ribbon
and make sure that Navigation pane is selected.
Do not the change system locale if you are using an English, Portuguese, Japanese or
NOTE Chinese Operating System.
[OPTIONAL] 25 Right-click the taskbar to open the Taskbar and Start Menu Properties dialog. In
the Taskbar tab, under Taskbar buttons select Combine when taskbar is full.
This will simplify switching between open CDS instances.
1
View the items by icon to see a list of all items.
If this procedure does not work as expected, or the computer has no internet access,
NOTE install .NET 3.5 from the Windows installation media (see Method 3 under
https://support.microsoft.com/en-us/kb/2734782). If you do not have installation media,
create them as described under
http://windows.microsoft.com/en-US/windows-8/create-reset-refresh-media?woldogcb
=0.
c To make sure that all the net.tcp components are properly initialized,
Non-HTTP activation must be enabled. Expand the .NET Framework 3.5
(includes .NET 2.0 and 3.0) node and select the Windows Communication
Foundation Non-HTTP Activation check box.
[MUST] 11 Disable Compatibility View in Internet Explorer.
a Open Internet Explorer.
b Click the Tools menu, and then click Compatibility View Settings.
c Clear the Display intranet sites in Compatibility View check box.
[OPTIONAL] 15 Recycle Bin Properties: (right-click on desktop icon Recycle Bin) Select the
following options:
• Custom size: Select a size corresponding to approximately 10% of the
complete disk space for the drive.
• Select Display delete confirmation dialog.
Repeat these steps for all drives of your computer.
16 Region (Microsoft Control Panel): Language for non-Unicode programs:
On the Administrative tab, click Change system locale.... From the drop down
list, select English (United States).
Do not the change system locale if you are using an English, Portuguese, Japanese or
NOTE Chinese Operating System.
Configure Windows 7
[MUST] 1 System (Microsoft Control Panel)1: Register Windows with Microsoft.
[MUST] 2 Folder Options (Microsoft Control Panel): In the View tab,
• Select Always show menus.
• Select Display the full path in the title bar.
• Clear Hide extensions for known file types.
• Clear Use Sharing Wizard.
[MUST] 3 Windows Update (Microsoft Control Panel):
Click Check for updates to check for updates and apply all critical security
patches.
[MUST] 4 Windows Update (Microsoft Control Panel): Change the settings for updates:
Click Change settings. In the Important updates section, select Never check for
updates. Clear the other update options.
1
View the items by icon to see a list of all items.
Do not the change system locale if you are using an English, Portuguese, Japanese or
NOTE Chinese Operating System.
Instrument Connections
The following Agilent instrument driver software revisions are part of the
OpenLAB CDS 2.2 software package, and are installed by default with the
software:
Table 7 Agilent driver packages shipped and installed with OpenLAB CDS 2.2
Agilent LC A.02.16
Agilent GC B.01.02
Headspace B.01.07
Pal 3 A.01.03
Pal-xt B.01.08
Additional RC.NET instrument drivers are supported with OpenLAB CDS. You
will need to install them separately (see section Install or Upgrade Driver
Software in chapter 1 of your Workstation or Client Server guide).
Supported Agilent drivers are listed in the following table with their minimum
software revisions.
Table 8 Other compatible Agilent Instrument Drivers (on the installation medium)
PAL XT B.01.08
ELSD A.01.07
Recommended Firmware
Agilent recommends using current LC firmware sets with your OpenLAB CDS.
The latest LC firmware sets are 7.01, 6.50, and 6.30. Firmware Set 7.01 can be
downloaded from
http://www.agilent.com/en-us/firmwareDownload?whid=99818.
LC and CE firmware revisions are grouped into sets for each module or system. Firmware
NOTE sets include just the latest firmware of each module.
Modules combined into one LC instrument always need to have firmware from the same
set. Firmware of one set is fully compatible with your CDS that supports this firmware set.
However, if one module uses a firmware revision lower that the specified Minimum
Firmware Revision some functionality might not be supported.
Do not mix firmware revisions from one set with older or newer sets. Firmware is not
tested across set borders!
Device Recommended
Firmware
Agilent 1100 Series, 1200 Series and 1200 Infinity A.07.01 or later
Agilent 1200 Series, 1200 Infinity and 1120 Compact LC B.07.20 or later
You need to upgrade all existing modules to the latest version only if
• you add a new LC module to the existing system (recommended).
• one of the existing modules requires an upgrade due to a bug fix solved
with the latest release.
For more information on downloading the current LC firmware, please refer
to the Agilent website atLC Firmware News and Downloads and Firmware
Update Tools & Procedures (, ).
The tables below “Supported LC modules” on page 131 list the supported LC
modules with the minimum required firmware. Please note that using some
driver features like valve-thermostat clusters and new temperature control
modes require firmware updates beyond the minimum required firmware for
related modules. For more information please refer to the respective release
notes. The current Release Note for Agilent LC and CE drivers Revision
A.02.16LC Driver Release Note A0216.pdf is available in the Docs/Enu folder of
the OpenLAB CDS media.
Supported LC modules
Check for the product number at the lower right of each module or system.
HINT
Agilent LC – Pumps
Agilent LC – Detectors
ELSD
Agilent LC Systems
G4286B 1220 Infinity LC System Isocratic, Man. Inj., VWD, 600 bar B.06.50
G4287A 1120 Compact LC, Isocratic with Oven and ALS B.06.50
G4287B 1220 Infinity LC Isocratic, ALS, TCC, VWD, 600 bar B.06.50
G4288B 1220 Infinity LC Gradient, Man. Inj., VWD, 600 bar B.06.50
G4289B 1220 Infinity LC Gradient, ALS, TCC, VWD, 600 bar B.06.50
G4290B 1220 Infinity LC Gradient, ALS, Man. Inj., TCC, VWD, 600 bar B.06.50
G4291B 1220 Infinity LC Isocratic, Man. Inj., TCC, VWD, 600 bar B.06.50
G4294B 1220 Infinity LC Gradient, ALS, TCC, DAD, 600 bar B.06.50
G4286C 1220 Infinity LC System VL, Isocratic, Man. Inj., VWD, 400 bar B.06.50
G4287C 1220 Infinity LC System VL, Isocratic, ALS, TCC, VWD, 400 bar B.06.50
G4288C 1220 Infinity LC System VL, Gradient, Man. Inj. VWD, 400 bar B.06.50
G4289C 1220 Infinity LC System VL, Gradient, Man. Inj. VWD, 400 bar B.06.50
G4290C 1220 Infinity LC System VL, Gradient, ALS, TCC, VWD, 400 bar B.06.50
G4291C 1220 Infinity LC System VL, Isocratic, Man. Inj. TCC, VWD, 400 bar B.06.50
G4292C 1220 Infinity LC System VL, Isocratic, ALS, VWD, 400 bar B.06.50
G4293C 1220 Infinity LC System VL, Gradient, ALS, VWD, 400 bar B.06.50
Agilent LC/MS
Recommended Firmware
Always use the most recent Firmware installation package that comes with
the driver package.
G1948B ESI
Supported GC Systems
G2630A 6850
Serial # >= US10243001 A.06.02
Serial # < US00003200 A.03.07
Supported Autosamplers
Table 10 Agilent GC Autosampler
G1888A A.01.10
G1888 Headspace
G6500-CTC CTC Combi-Pal for Liquid and Headspace Agilent 2.6.8 or 4.3.0
Injection
Agilent GC/MS
Recommended Firmware
Always use the most recent MS Firmware installation package that comes with
the driver package.
PAL-XT CTC
PAL-3
G7697A Headspace
G1888
Download the most recent instrument drivers available from SubscribeNet to control
NOTE non-Agilent Instruments.
Table 15 Drivers for non-Agilent instruments compatible with OpenLAB CDS 2.2
Part Number 3rd Party Vendor Driver Type Instrument Name Driver Version
Table 15 Drivers for non-Agilent instruments compatible with OpenLAB CDS 2.2
Part Number 3rd Party Vendor Driver Type Instrument Name Driver Version
G1390A UIB
Incompatible CE Instruments
OpenLAB CDS 2.2 can not be used to control CE Instruments.
Software Compatibility
Agilent Software
The following Agilent software is compatible with version 2.2 of OpenLAB
CDS:
• LabAdvisor for LC
• Instrument Utilities for GC
• ACE
• Remote Advisor
• Sample Scheduler 2.0 for OpenLAB CDS
• OpenLAB ELN
• Match Compare
The privileges described in the following can be associated with different roles
in the Control Panel. The following roles are available:
Project Administrator
In the Control Panel under Administration > Roles, you can view or change the
associated privileges, or create your own roles.
Project Privileges
Table 16 Project Management
Name Description
Manage project or project group User can create or edit project properties and
can move the project but cannot view or edit
the project access settings.
Manage project or project group access User can view and edit the project access
settings.
View project or project group User can see a project and project details but
cannot edit.
Note: This privilege is required for all users.
Access content using web client User can view the data via the Content
Management web interface.
Edit content of project User can create new versions of documents
(e.g. data, methods, or templates).
Table 17 E-Signature
Name Description
Name Description
Create and modify sample prep View, edit, and save an autosampler sample
prep file
Name Description
Create and modify acquisition method Create, edit and save an acquisition method file
(*.amx)
Name Description
Name Description
Edit chromatogram extraction parameters View and edit the parameters in the Extraction
> Chromatogram section of a method.
Edit spectrum extraction parameters View and edit the parameters in the Extraction
> Spectrum section of a method.
Edit MS library search parameters View and edit the parameters in the MS
Library Search > Properties section of a
method.
Edit system suitability parameters View and edit the parameters in the
Compounds > System Suitability section of
method.
Edit custom calculation parameters View and edit the parameters in the Tools >
Custom Calculation section of a method.
Edit signal parameters View and edit the parameters in the General >
Signals section of a method.
Edit sample purity parameters View and edit the parameters in the MS
Sample Purity section of a method.
Edit reporting parameters View and edit the parameters in the Reports >
Injection Report section of a method.
Edit general parameters View and edit the parameters in the General >
Propertiess section of a method.
Privilege Description
Unlock/lock report template items Lock and unlock report template items (tables,
chromatograms, groups of items, ...) to control
who is allowed to modify those.
Name Description
Create and modify sequence template Create, edit and save sequence creation
templates (*.stx).
Table 23 Sequence
Name Description
Edit any users running sequence Edit any user's running sequence (status
Acquiring in the Run Queue).
Create and modify sequence Create, edit and save sequences (*.sqx)
Edit users own running sequences Edit your own running sequences (status
Acquiring in the Run Queue).
Name Description
Change method audit trail settings Edit and save method audit trail settings
(project properties in the Control Panel).
Add manual audit trail entry Add a manual entry to document your own
actions in the audit trail.
Table 25 Control
Name Description
Abort any running sample Abort any running sequence or single run.
Manual control (in run) Access manual control functions while the
instrument is running.
Manual control (only when instrument idle) Access manual control functions while the
instrument is idle.
Delete any pending run Remove pending runs from the Run Queue.
MS autotune Perform MS autotune and check tune.
Name Description
Update master processing method Save changes from a result set method to the
corresponding master processing method in
the Methods folder.
Name Description
Launch Custom Calculation Editor Start the Custom Calculation Editor from Data
Analysis.
Name Description
Table 28 Data
Name Description
Table 29 Lock
Name Description
Name Description
Access Custom Tools section Start external programs that were added to the
application via the customization tool
Table 31 Snapshot
Name Description
Instrument Privileges
Table 32 Instrument Management
Name Description
View instrument or location User can view and access a location in the
tree, but not edit access security, can view
properties.
Manage Instrument or location User can create and move locations and edit
properties (name, description, etc).
Manage instrument or location access User can view and edit the location access
settings.
Administrative Privileges
Table 33 System Administration
Name Description
Edit activity log properties Can change the Activity log Settings in the
Control Panel (that is, can turn logging on for
the System Activity Log).
Create administrative reports Can create any of the system admin reports.
Please check the following web site for your local sales and support contact:
http://www.agilent.com/en-us/contact-us/page
In This Book
*M8413-90002*
*M8413-90002*
M8413-90002
Agilent Technologies