TDE10170 11 ReleaseNote

This document provides release notes for QAWeb Enterprise software updates, including: - Portal and Agent releases from July 2022 to May 2021, with new features and bug fixes - The July 2022 release added optional wake settings and remote task triggering for Portal, and new test patterns for Agent - The April 2022 Portal release added single sign-on and multi-factor authentication, and the Agent fixed an issue with missing results - The February 2022 releases added the ability to be a member of multiple organizations in Portal, and improved calibration matching in Agent
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
103 views

TDE10170 11 ReleaseNote

This document provides release notes for QAWeb Enterprise software updates, including: - Portal and Agent releases from July 2022 to May 2021, with new features and bug fixes - The July 2022 release added optional wake settings and remote task triggering for Portal, and new test patterns for Agent - The April 2022 Portal release added single sign-on and multi-factor authentication, and the Agent fixed an issue with missing results - The February 2022 releases added the ability to be a member of multiple organizations in Portal, and improved calibration matching in Agent
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 13

QAWeb Enterprise release notes

Document ID: TDE10170


Download latest release notes: https://www.barco.com/en/support/docs/TDE10170

July 11th, 2022 – Portal 2.12, Agent 2.9


Portal 2.12
New functionality
▪ Optional setting to wake the displays if tasks become due.
▪ Remote trigger execution of all tests of a workstation
▪ Delete users from an organization
▪ Fleet reports + Insights become part of the essential feature set.

Agent 2.9
New functionality
▪ Optionally wake standby displays when tasks become due.
▪ Support tool to optimize I-Guard area
▪ Addition of TG-270 SQC and TG-270 PQC test patterns
▪ Support for MDNC-6121 SNEF CN
▪ Support for MDCC-6530 NC MKI

Apr 26th, 2022 – Portal 2.11


Portal 2.11
New functionality
• Support for configuration Single-Sign-On. As part of this, users are automatically migrated to
the Barco-wide user authentication system. The look and feel of the login process has
changed.
More information on SSO setup: https://www.barco.com/en/support/knowledge-base/kb12591
• Support for MFA. A user can setup MFA by visiting https://www.barco.com/mybarco and
signing in with the same QAWeb Enterprise user and password. Next, MFA setup can be done
under ‘My Account’ (involves scanning a QR code with commonly-used authenticator apps).
• Added column ‘communication type’ in assets > Displays
• The address input (using the MapBox) component is improved to allow setting the address
manually for addresses missing a ZIP code; or when a network error causes the MapBox
component to fail.

Apr 1st, 2022 – Agent 2.8.2


Agent 2.8.2
Solved issues:

QAWeb Enterprise Release Notes 1


• We identified and fixed an issue where the automatic resending of local Agent task results
performed during server unavailability was not properly functioning. After the unexpected
downtime of the online service on March 27th-28th, several workstations in the Portal were
missing calibration or test results (even though in reality, these tasks were executed), causing
them to be incorrectly shown as 'Due' or 'Outdated'.
Updating the Agent to version 2.8.2 will correct this problem and cause these agents to
automatically resend the missing results to the Portal. More info:
https://www.barco.com/en/support/knowledge-base/KB12620
• SteadyColor calibration and SteadyColor response test were disabled on systems that
implement SteadyColor using the MXRT graphics board.

Feb 16th, 2022 – Agent 2.8.1


Agent 2.8.1
Solved issues
• When updating existing agents to version 2.8.0, the custom HTTP proxy settings were lost,
causing a loss of connectivity after the update until the proxy configuration was manually
applied again. (Note: Agents using 'no proxy' or 'use system proxy' were not affected by this
issue. New 2.8.0 installations were also not affected.)

Feb 3rd, 2022 – Portal 2.10.0, Agent 2.8.0


Portal 2.10.0
New functionality
• QAWeb Enterprise Portal users (identified with a single email address) can now be a member
in multiple QAWeb Enterprise organizations.

o This is mostly useful for external physicists or service providers that are active in
multiple healthcare organizations that use QAWeb Enterprise.
o Users that are member of multiple organizations, will be able to select the organization
to login to, following successful user+password entry. After that, the Portal
remembers the last organization as default.
o A menu entry 'Switch organization' has been added in the menu, accessible by clicking
your initials in the upper-right corner of the Portal application.
o Users have separate permissions and scopes in different organizations, but a single
password for login is used.
o We have some remaining work to enable user SSO/ADFS support, this will be added
soon.

• When deleting a workstation in the portal, the user is asked whether to unlink or keep the
associated displays.

Agent 2.8.0
• Added the ability to match the white luminance measurements by the I-Guard sensor of the
display with the luminance measured using an external photometer. For more information, see
the chapter Hardware Maintenance in the agent user manual (https://documentation-qaweb-
agent.healthcare.barco.com).
• The installation process has been made more reliable. The local storage of the agent moved to
C:\ProgramData\Barco\QAWeb (earlier versions:
C:\Windows\System32\config\systemprofile\AppData\Roaming\Barco)
• Uninstalling the agent will now remove all local data and settings. This is done to ensure a
consistently clean system state in case of troubleshooting. Note: when updating the agent

QAWeb Enterprise Release Notes 2


software, it is not required to first uninstall the old version. The installer will automatically
replace the older version, making sure any local settings are saved.
• Improved reliability of USB communication with the display (in some hardware combinations,
users reported an occasional loss of USB communication only solved by power-cycling the
display).
• Various smaller bugfixes.

Oct 26th, 2021 – Agent 2.7.1


Agent 2.7.1
Solved issues
• Fixed an issue introduced in the 2.7.0 release 11 that caused a loss of connection to the online
service when upgrading from an older version.
• Fixed issue where after initial installation of 2.7.0, test patterns could not be shown until a
restart of barcocore.exe

Oct 11th, 2021 – Portal 2.9.0, Agent 2.7.0


Portal 2.9.0
New functionality
• Introduced a refresh button on the workstation detail which triggers the Agent at the
workstation to immediately sync its settings with the latest server policies. Using this, you can
update a task policy and immediately apply it to an agent.
• There is a dedicated detail per display bundling all related information: display characteristics,
calibration settings, task history, warranty info, ...
• Next to that, we improved the address input of a hospital. It requires less steps to complete
the form and provides more accurate information on a map.
• The IP of the workstations is now displayed on the workstation page next to the hostname.
Solved issues
• Solved an issue in larger organizations where the list of workstations was trimmed.
• Sorting on several Insights columns was not possible.
• Could not delete a duplicate policy.

Agent 2.7.0
New functionality
• Agents are now able to run in stand-alone mode. This allows individual workstations to
calibrate and monitor attached displays.
• During the installation of the Agent, the user can postpone the configuration to a later
moment. This allows you to take an image of the workstation and deploy it across multiple
machines.
• We introduce a display communication check which monitors the communication between
workstation and display (video link or USB)
• Details on the task results can be consulted from the Agent systray. They provide the same
detail as what was already available in the Portal. They are available in PDF for portability and
storage.
Solved issues
• Some configurations were suffering from frequent dialog popups to confirm the use of the
displays.
• Fixed use of , and . as decimal separator in the ONR test.
• Solved issue where the visual test results were discarded when they ran on workstations
where the displays didn't go to sleep mode.

QAWeb Enterprise Release Notes 3


• Interaction between IWT and Agent got solved to prevent SpotView/DimView during
calibration and QA tasks.
• Manuals are now included correctly in the Agent installer and are deployed in the installation
folder.

June 28th, 2021 – Portal 2.8.1


Portal 2.8.1:
New features / improvements:
• Technical hostname is now also visualized
• ‘Allow local override’ is removed from QA policies
Bugfixes
• Workstation history contained no task results for unlinked or removed displays
• The last LRT result in the JCI report was not always reflecting the latest result when

June 7th, 2021 – Portal 2.8.0


Portal 2.8.0:
New features / improvements:
• Dashboard widgets are stacked vertically (vertical scrollbars appear)
• Several list table layouts were improved
• Org. Scope filtering was added on all lists
• Hostname column has been added in Assets – Workstations
• Dashboard widgets can be renamed
Bugfixes
• Search functionality in the online manual was very slow
• Could not delete rooms with assigned deleted workstations
• A user that had no scope, could see all workstations, while he should see none

May 25th, 2021 - Agent 2.6.1


Agent 2.6.1
Bugfixes:
• Confirm use dialog pops up too often
• Changing calibration settings does not immediately start a calibration
• Mammography acceptance test: TG18-OIQ letter L in the black area is rendered at wrong
level.
• Mammography acceptance test: one question is positioned initially on top of the relevant
pattern, slowing down the user.
• Blocked internal event loop leading to an unresponsive user interface, IOT-mqtt connectivity
loss, a desynchronized QA status
• In multi-display setups, the OSD unlock feature unlocks only one display
• Failure to start when 3rd party displays with a space character in the serial number are
attached.
• The current display configuration was sometimes inaccurate when connecting an eGFX solution

April 1st, 2021 - Portal 2.7.0


Portal 2.7.0:
New features / improvements:

QAWeb Enterprise Release Notes 4


• Budget report
• Moved excel export, user permissions and user scope restrictions to base tier
• Backlight runtime hours visible in workstation component detail page
• Warnings in QA policy editor when policy is incomplete
Bugfixes
• Rendering issue of workstation components tab when resolution is at 0x0
• Organization names cannot could not contain accents

March 8th, 2021 - Portal 2.6.0, Agent 2.6.0


Portal 2.6.0
New features / improvements:
• Task history report: filterable list of all calibrations and QA task history of the organization
• Recently viewed workstations in top navigation bar
• Search on serial number in Asset: Displays list
• Organizational structure setup: tooltips and search functionality
• Workstation: components tab shows date of last connectivity
• Remote update agent (from 2.6 versions onwards)
• Mammography compliance PDF fleet report (PRO feature)
• Calibration policy editor: show warning when extended calibration is enabled
• Calibration policy editor: show warning when a use is missing.
• Performance improvements for test results lists
• Improved Image quality of charts in PDF reports
Bugfixes
• Email digests were not sent in some configurations.
• Time filter of mammography compliance test also filtered out acceptance test
PDF report of an individual task showed the task of PDF generation instead of the task of
execution

Agent 2.6.0
New features / improvements:
• Agent software mechanism is now included: agents will detect future updates and allow
updating from the UI.
• Visual test questions are now translated
• Unlock the monitor’s Display Menu (OSD menu) for 1 hour
• Status panel is simplified and only shows busy tasks and due manual tasks
• Introduced a workstation details panel, accessible from status panel. Currently, shows all
applicable tasks and components.
• Confirm the use of displays when adding new displays or modifying the use
• MDNC-2122 Ambient light sensor support
• User guide for agent is separated out from the portal manual
Bugfixes
• Could not cancel a running luminance response test with an external sensor
• Reduced check interval of barcocoremonitor from 5 to 30 seconds to avoid spawning barcocore
when windows is shutting down
• Runtime and generalconfig tasks should not log journal event log entries for sleeping displays

QAWeb Enterprise Release Notes 5


December 1st, 2020 - Portal 2.5.1, Agent 2.5.1
Bugfix release with small improvements

Agent 2.5.1
Bugfixes:
• Some installed agents were not correctly executing tasks due to corrupt local task results
cache files. File corruption was caused by earlier versions (<2.4) that occasionally crashed on
USB events. This resulted in ‘Agent has no valid policy’ message in the webclient when trying
to remotely run a calibration or QA task.
• ICC compliance check gave unclear message when no ICC profile is associated with a display
• Mammography compliance test could not run on workstation with displays
Improvements:
• Agents can upload logfiles at Barco remote request for investigation purposes

Portal 2.5.1
Improvements/bugfixes:
• Agent remains out of sync with current policy when only non-clinical displays are used
• Workstations asset list and workstation detail now indicate that a new agent update is
available
• Workstation detail - event log: Repeating N/A task results in the event log are now aggregated
instead of repeating entries
• Display asset list: firmware version column added
• Workstation QA list: online/offline column added
• Show accurate error message when adding user with email that already exists
• Improved rendering of tables on Microsoft Edge browser

November 9th, 2020 - Portal 2.5.0, Agent 2.5.0


Important changes:
• From agent 2.5 onwards, all calibration and QA tests require that the display is on for at least
10 minutes (backlight powered on). If calibration or QA tests are started (either locally, or
through the webclient), they will wait until the 10 minutes is reached.
• Ambient light settings on room-level have been replaced with ambient light policies. Please
refer to the user manual for further reading. An automatic migration from the old settings to
the new system is implemented.
• Questions in the “NYState Visual Test” were rephrased so that ‘Yes’ always is the expected
answer.

Portal 2.5.0
New features / improvements:
• Ambient light policies + optional notifications for ambient light limits (see also: agent,
important changes)
• PDF reports are now automatically removed from the Portal after 7 days
• New PDF summary workstation report:
o DICOM GSDF Compliance report: Contains a chronological overview of all calibration
and luminance response tests, in a table format.
• Fleet reports (PRO feature): Fleet reports combine data from multiple workstations into one
single report. This way, a single PDF with information of all workstations, reducing time spent
on gathering reports. Currently, only one report type is available - more will be added in later
versions
o DICOM GSDF Compliance fleet report: Same content as the DICOM GSDF compliance
report, but for all workstations.
• Downloadable PDF reports for all individual test results - including chart/graphs

QAWeb Enterprise Release Notes 6


o Mammography Acceptance test result
o Mammography Constancy test result
o ICC Compliance test result
o Maximum luminance test result
o Luminance ratio test result
o sRGB Color Space test result
o sRGB Luminance response test result
o White point chromaticity test result
o ONR Acceptance test result
o ONR Constancy test result
o Multi-display luminance test result
o Luminance Uniformity test result
o Greyscale Chromaticity test result
o DICOM GSDF Luminance test result
o Visual test result
o SteadyColor response test result
• Translation of user interface and manual into 9 extra languages: Japanese, Korean, Chinese
(Traditional), Chinese (Simplified), Russian, Serbian, Arabic, Turkish, Ukrainian
Bugfixes:
• Time zone in PDF reports was not correctly set
• Nightly email notifications were not sent for some organizations
• Mammography compliance report: in some cases, with dual head setups, data could be
swapped between display 1 and 2
• QA policies: edit/delete buttons were still clickable for Mammography compliance test in read-
only mode
• NY State and NY City Visual Test: "Bright pixel" question was missing in the setting presets
• Mammography compliance test: The white point luminance test now compares the current
measured white point to effectively the measured calibration white point (L’max) instead of
the requested target luminance at the last calibration time.
• Changing settings of the visual test did not refresh the policy of the agent
• JC report did not show results of luminance response tests executed as part of Mammography
Compliance tests.
• From/to filters in test results were not correctly cut off on the date limits.
• Countries in signup webpage were not in alphabetical order

Agent 2.5.0
New features / improvements:
• Optional ambient light warnings/notifications (requires displays with an integrated ambient
light sensor. See also: ambient light policies in user manual
Bugfixes:
• German translations were missing
• Questions in ONR wizard were not translated
• Mammography compliance test could not run on workstation with displays of different use
• Executing Mammography compliance constancy test temporarily interrupted the IOT
connection.
• Fixed communication issue with MNDC-2221 with older firmware versions (V1.XX)

QAWeb Enterprise Release Notes 7


Aug 31st, 2020 - Portal 2.4.1, Agent 2.4.1
Portal 2.4.1
Bugfixes
• Workstation hostname is now correctly shown again on the workstation detail page
• Mammography compliance test report:
o displayed a green checkmark icon (instead of a red failure icon) when the white point
luminance test exceeded the threshold in a positive direction
o in some cases, other displays than mammography use displays were included
o Incorrect amount of reflected ambient light was added to an L’ value

Agent 2.4.1
Bugfixes
• Improvements to waking displays that are in standby before starting a QA test:
o Abort test results with sensor measurements when the backlight is powered off during
a test (in standby) - this caused storage of false/unrealistic test results
o Additional logic added to avoid displays going into sleep, when the OS turns off the
video output during a QA test
o Multiple repeated executions of remotely waking a display no longer fails
• On MDNC displays, draw large sensor patch to measure for maximum luminance test - in line
with other QA tests
• Not all translations of QA tasks were included

July 14th, 2020 - Portal 2.4.0, Agent 2.4.0


Portal 2.4.0
New features & improvements
• New QA tests:
o Mammography compliance test (acceptance test and constancy test)
o sRGB luminance response test
o sRGB color space test
o Grayscale chromaticity test
o White point chromaticity test
o ICC compliance test
• Generation PDF summary reports (currently, only mammography compliance report is
implemented)
• Calibration: sRGB luminance response option, sRGB gamut option
• Added ‘pathology’ use for calibration policies and QA policies
• Workstations can be deactivated (=logically deleted)
• Unlinked displays can be deactivated (=logically deleted)
• Schedule QA tests on non-Barco displays: allow selecting all use types for any display.
Important: calibration of non-Barco displays is not enabled. QA tests require a supported
external optical sensor to be executed.
• License page: activate [Pro] features by purchasing DisplayCare / ManagedCare subscription
• View Barco service providers / partners to which login access has been granted by hospitals
through Partner Portal
• Roadmap / Ideas portal
• Signup page: sort countries by name
Important changes / modifications
• Remotely triggering a QA test or calibration in the web-client will now wake sleeping displays
• User activation emails are valid for 7 days (previous versions: 24 hours)
• Changes to default Barco QA policy

QAWeb Enterprise Release Notes 8


o ‘Mammography’ and ‘Digital Breast Tomography’ use:
▪ Added mammography compliance test (replacing existing DICOM GSDF and
inter-display uniformity test)
▪ Removed maximum luminance test & luminance ratio test
o ‘Reviewing’ use:
▪ Removed ‘DICOM GSDF Luminance response test’ and ‘Luminance ratio test’

• ONR constancy test no longer contains a daily visual test


• Renaming:

o ‘Color Response Test’ is renamed to ‘SteadyColor Response test’


o ‘Luminance Response test’ is renamed to ‘DICOM GSDF Luminance response test’
o ‘Inter-display uniformity test’ is renamed to ‘multi-display uniformity test’

Bugfixes
• Issue with timeformat caused agents to not always automatically execute a due task
• New workstation would not immediately (but after 15 minutes or a restart) receive a correct
policy
• fixed excel export of user list
• Sorting on connectivity date column in workstation asset list
• Workstations with an empty display name are not correctly shown

Agent 2.4.0
New features & improvements
• New QA tests and calibration settings, as listed in [Portal 2.4.0] notes
• Remotely triggered QA tasks will wake displays in standby
• ONR 195240-20:2017 (ONorm, Austria) support: multi-select ONR use classifications during
acceptance test
• Installer: Option ‘deactivate qaweb1’ added. This will set the Qaweb1 windows service as
‘disabled’.
• Detection of external optical sensors is now controlled using a user interface button, instead of
continuous background polling.
Bugfixes
• Improved communication stability of IBA LXCan optical sensor (using LXCan driver 2.9.5)
• Fixed an issue with Barco IWT interaction (also requires update to MXRT driver version
10.184.1.0)
• Systems without support for 'virtual accounts' could not install the agent
• Agent that cannot readout MAC causes 4xx errors in logs.
• ONR acceptance test: unit cd/m² is not correctly displayed.
• Luminance Ratio test occasionally runs on display with disabled backlight.

Feb 06, 2020 - Portal 2.3.0


Portal 2.3.0
New features & improvements
• ONR 195240-20:2017 (ONorm, Austria) support
• QA Status widget has been improved to better include schedule status
• Introduction of Essential/Pro features.

QAWeb Enterprise Release Notes 9


• Export lists to Excel [Pro]
• Insights [Pro]
• Ambient light presets also list their lux values
• Non-clinical displays are not listed in JC Report
Bugfixes
• User account email addresses are no longer case sensitive
• Search filters are case insensitive
• Org tree is sorted alphabetically.
• Remote actions show untranslated error messages.

Agent 2.3.0
New features & improvements
• ONR 195240-20:2017 (ONorm, Austria) support
• Prevent displays to go in sleep during QA tests
Bugfixes
• Installer rolls back on unpatched Windows 7
• QA Status out of sync with Portal
• Better handling of displays in standby mode
• QA Tasks that errored once didn’t always retry automatically

Dec 2, 2019 - Agent 2.2.2


Agent 2.2.2
Bugfixes
• BarcoCore.exe high CPU usage when connection to Portal cannot be made (due to e.g. no
network connection or incorrect time setting

Nov 28, 2019 - Portal 2.2.1, Agent 2.2.1


Portal 2.2.1
New Features / improvements
• Adding asset notes: create/edit/remove/list notes related to a display
Bugfixes
• Added missing translations in policy assignment.
• Redirect agent to dedicated IoT endpoint. Do not use data.iot.

Agent 2.2.1
New features & improvements
• Log Service (windows service) is integrated with BarcoCore. Logging is not longer sent to
Portal.
• Barcoconfig.exe has been improved: set the log-level, mark log file, run connectivity test.
Bugfixes
• Do not try to use external sensors for MDNG-6121
• Core Monitor is made more robust. BarcoCore process is monitored throughout it's life cycle.
• BarcoCore will exit if another instance is running.
• Agent IoT connection is established using local system time (requirement: Workstation time
needs

QAWeb Enterprise Release Notes 10


Oct 14, 2019 - Portal 2.2.0, Agent 2.2.0
Portal 2.2.0
New Features / improvements
• Handling QA Status and scheduling has been improved. QA tests and workstations now
maintain a 'QA status' and a 'Schedule status'
• Image Quality policies have been renamed to Calibration policies
• Policy engine has been simplified: There now is a single QA policy and Calibration policy; and
only one
• Rework and improvements of user interface to adapt to improved QA status & single policy
• Permission system and user org. scope restriction are not available in the Essential version
• Translation of user interface and manual to all languages of European Union (22 languages)
• It is now possible to remotely trigger all QA tests that do not depend on an external sensor or
manual user interaction
Bugfixes
• Filtering on organization level in workstations list returned more results than expected
• Logging improvements
• Database query performance improvements
• Resend activation mail: link in new activation mail ends up in "Forbidden" after user
confirmation
• Many small glitches

Agent 2.2.0
New features & improvements
• Support Raysafe X2 sensor
• update LxCan driver to latest version (2.9)
• Support new Barco displays (MDDC-6530 & 4330 ; MDRC-2321 & 2324)
• Translation of user interface and manual to all languages of European Union (22 languages)
• Support hotplugging of thunderbolt eGFX
Bugfixes
• Improve handling of displays that are physically attached to the workstation but not "active" in
the Windows desktop
• BarcoCore crashes when display is not attached to Windows desktop
• Improve handling of running a QA task on a system with displays that are not "ON"
• Agent does not attempt to reconnect when IAM was unresponsive in the beginning.
• When agent encounters a DDC error it only recovers after a reboot
• Timestamp format of agent incorrectly states it is in Zulu time
• Systray sometimes crashes when switching users
• CloneView does not work on a Fusion Display in dual view mode
• Many other fixes encountered during internal testing procedures

June 24, 2019 - Portal 2.1.2


Portal 2.1.2
New Features / improvements

QAWeb Enterprise Release Notes 11


• The language of the webclient can now be explicitely specified for each individual user: On the
user profile, you can choose between 'Use browser setting' or a specific language (Before this
update, the language of the browser was always used).
• The documentation is now more directly accessible by clicking in the top-right corner and
selecting the Help menu item. Both the online and PDF version are available in this way. The
help is shown in the current user language.
• Firmware and part number are now displayed in the details (requires Agent 2.1.1)
• The drag-and-drop experience and performance is improved on all places where an
organization tree is displayed. (internal component update)
Bugfixes
• Fixed: first time click on instruction manual or agent download lands on Amazon S3 page
instead of launching QAWE manual.
• As system is added to clear to ensure that the browser correctly clears previously cached
translations (which resulted in missing text files until a forced browser cache clear)
• Fixed some missing translation texts (e.g. when unassigning a policy, taskresults.nativewhite)
• Fixed: When a Luminance Response Test was launched remotely, it reported incorrectly to the
Portal that is was triggered by a local user (requires Agent 2.1.1)
• When using Internet Explorer, the user is notified that this browser is not supported
• (Note: This release contains several features for the back-office functionality for Barco
Employees - not visible to end-users)

May 14, 2019 - Portal 2.1.1, Agent 2.1.1


Portal 2.1.1
New Features / improvements
• Email notifications: It is now possible to receive a nightly summary email with a list non-
compliant workstations. It will only be sent when there is at least one non-compliant
workstation within the scope of the user. To enable this feature, two steps must be taken:
o Enable the functionality on organization level and configure on which weekdays the
nightly mail should be sent out (mails are sent out at 01:00 am)
o Each user should individually subscribe to the emails . This can be done on the user
profile.
• Command-line instructions: On the My Organization page, command-line installation and un-
installation instructions were added below the Download section.
Bugfixes
• HTML online help was not reachable
• Fixed a page redirect issue on session expiry
• When creating a room, the 'Custom lux value' field is now only shown when selecting 'custom'
• Newly created users did not have all permissions, even though it appeared so visually
• Fixed some Dutch translation typo's
• Added missing text strings for ambient light presets, org structure buttons and permissions
(technical codes were shown instead)
• Fixed a Bad request error when confirming password change in the 'forgot password' flow
• Fixed sending out a redundant 'forgot password' e-mail again after resetting the password

Agent 2.1.1
Serviceability improvements
• The firmware version and part number (K93....) of Barco displays is now read and sent to the
Portal. This is mainly used for improved serviceability.
• The agent now writes a technical logfile
in C:\Windows\System32\config\systemprofile\AppData\Roaming\Barco\LogSend.

QAWeb Enterprise Release Notes 12


• CE Label change from CE 0120 to CE 1639
Bugfixes
• Fixed the problem where calibration or QA tests where indefinitely repeating at an hourly
interval
• Fixed a meaningless error being shown when trying to perform a measurement with a Barco
LCD sensor which is in Ambient Light measurement mode.
• Fixed some DPI scaling / small font sizes problems in certain configurations
• Could not cancel an ongoing color response test. Fixed.
• When a Luminance Response Test was launched remotely, it reported incorrectly to the Portal
that is was triggered by a local user.
• Increased timeout for registration timeouts during unattended installation (and improved
logging to investigate possible future occurrences)