Gossen Metrawatt Metraclip 87 User Manual
Gossen Metrawatt Metraclip 87 User Manual
METRACLIP 87
CONTENTS
1 PRESENTATION............................................................................................................ 8
1.1 THE SWITCH ........................................................................................................................................ 9
1.2 THE KEYS OF THE KEYPAD..........................................................................................................10
1.3 THE DISPLAY UNIT..........................................................................................................................11
1.3.1 The symbols of the display unit....................................................................... 11
1.3.2 Measurement capacity exceeded (O.L) .......................................................... 13
1.4 THE TERMINALS ..............................................................................................................................13
2 THE KEYS ..................................................................................................................... 14
8 ADDRESSES .................................................................................................................. 73
4
English Clamp Multimeter METRACLIP 87
You have just acquired an METRACLIP 87 clamp multimeter and we thank you.
1.5 V battery.
AC – Alternating current.
Earth.
5
Clamp Multimeter METRACLIP 87 English
This device complies with safety standards IEC-61010-1 and 61010-2-032 for
voltages of 1000V in category IV at an altitude OF less than 2000m, indoors, with a
degree of pollution not exceeding 2.
These safety instructions are intended to ensure the safety of persons and proper
operation of the device. If the tester is used other than as specified in this data
sheet, the protection provided by the device may be impaired.
The operator and/or the responsible authority must carefully read and clearly
understand the various precautions to be taken in use.
If you use this instrument other than as specified, the protection it provides
may be compromised, thereby endangering you.
Do not use the instrument in an explosive atmosphere or in the presence of
flammable gases or fumes.
Do not use the instrument on networks of which the voltage or category
exceeds those mentioned.
Do not exceed the rated maximum voltages and currents between terminals or
with respect to earth.
Do not use the instrument if it appears to be damaged, incomplete, or not
properly closed.
Before each use, check the condition of the insulation on the leads, housing,
and accessories. Any element of which the insulation is deteriorated (even
partially) must be set aside for repair or scrapped.
Use leads and accessories rated for voltages and categories at least equal to
those of the instrument. If not, an accessory of a lower category lowers the
category of the combined Clamp + accessory to that of the accessory.
Observe the environmental conditions of use.
Do not modify the instrument and do not replace components with
"equivalents". Repairs and adjustments must be done by approved qualified
personnel.
Replace the batteries as soon as the symbol appears on the display unit.
Disconnect all cords before opening the battery compartment cover.
Use personal protective equipment when conditions require.
Keep your hands away from the unused terminals of the instrument.
6
English Clamp Multimeter METRACLIP 87
When handling the test probes, crocodile clips, and clamp ammeters, keep
your fingers behind the physical guard.
As a safety measure, and to avoid repeated overloads on the inputs of the
device, we recommend performing configuration operations only when the
device is disconnected from all dangerous voltages.
MEASUREMENT CATEGORIES
7
Clamp Multimeter METRACLIP 87 English
1 PRESENTATION
The METRACLIP 87 is a professional electrical measuring instrument that
combines the following functions:
Current measurement;
Measurement of inrush current / overcurrent (True-Inrush);
Voltage measurement;
Frequency measurement;
Measurement of harmonic distortion, total (THD) and order by order;
Continuity test with buzzer;
Resistance measurement;
Power (W, VA, var and PF) and Energy measurements;
Measurement of the Crest Factor (CF), the Displacement Power Factor
(DPF), and RIPPLE;
Recording of data in memory, Wireless data transfer to a PC (by Bluetooth);
1
Item Designation See §
3 3 Switch 1.1
7
4 Function keys 2
8
English Clamp Multimeter METRACLIP 87
9
Clamp Multimeter METRACLIP 87 English
4 5 6
Figure 3 : the keys of the keypad
10
English Clamp Multimeter METRACLIP 87
2
3
Symbol Designation
DC Direct voltage
11
Clamp Multimeter METRACLIP 87 English
V Volt
Hz Hertz
W Active power
A Ampere
% Percentage
Ω Ohm
m Milli- prefix
k Kilo- prefix
VA Apparent power
PF Power factor
CF Crest factor
12
English Clamp Multimeter METRACLIP 87
Continuity test
1 2
Figure 5 : the terminals
Item Function
1 Cold terminal (COM)
2 Hot terminal (+)
13
Clamp Multimeter METRACLIP 87 English
2 THE KEYS
The keys of the keypad respond differently to short, long, and sustained presses.
In this section, the icon represents the possible positions of the switch for which
the key concerned has some action.
2.1 KEY
This key is used to:
store and look up the last values acquired specific to each function (V, A, Ω,
W) according to the specific modes previously activated
(MAX/MIN/PEAK,Hz,THD); the present display is then maintained while the
detection and acquisition of new values continues;
perform an automatic zero correction in ADC/AC+DC et WDC/AC+DC (see also §
3.9.2)
Successive
presses on … serve
See also § 2.5.3 and § 2.6.3 for the action key with the action of the key
and with the action of the key.
14
English Clamp Multimeter METRACLIP 87
Successive
… serve
presses on
short -to select AC, DC or AC+DC. Depending on
your choice, the screen displays AC, DC or
AC+DC
2.3 KEY
This key is used to :
Scroll orders of harmonics or successive pages up;
Activate the back-lighting;
Activate the Bluetooth function.
Successive
… serve
presses on
short to scroll through the various pages of
measurement results, depending on the
function and possibly the active mode
(MAX/MIN/PEAK or THD/Harmonics)
15
Clamp Multimeter METRACLIP 87 English
2.4 KEY
This key is used to :
Scroll down through the orders of harmonics or successive pages;
Activate the recording of the data;
Activate the Bluetooth function.
Successive
… serve
presses on
short to scroll through the various pages of
measurement results, depending on the
function and possibly the active mode
(MAX/MIN/PEAK or THD/Harmonics)
16
English Clamp Multimeter METRACLIP 87
2.5 KEY
Successive presses on
… serve
17
Clamp Multimeter METRACLIP 87 English
Successive
…serves
presses on
long (>2 sec) to enter the True-INRUSH mode
-"Inrh" is displayed for 3s (the backlighting blinks)
-the triggering threshold is displayed for 5s (the
backlighting is steady);
-"------" is displayed and the "A" symbol flashes
-after detection and acquisition, the inrush
current measurement is displayed, after the
calculations stage "------" (backlighting off)
Note: the HOLD function does not interrupt the acquisition of new MAX, MIN, PEAK
values
18
English Clamp Multimeter METRACLIP 87
2.6 KEY
This key is used to display measurements of the frequency of a signal, of power, of
the levels and orders of harmonics.
Successive
…serves
presses on
short to display:
1.the frequency of the signal, the RMS
measurement, and the DC component
2.the crest factor CF, the RMS measurement,
and the DC component
Long (> 2 sec) 1. to enter or exit from the THD calculation and
display mode
2. to display the THDf, the THDr, and the RMS
value.
19
Clamp Multimeter METRACLIP 87 English
Successive
…serve
presses on
short to display the frequency of the order of harmonic
previously selected using the or keys,
instead of order hxx.
A 2nd short press restores display of order (hxx)
or hdC
Successive
…serve
presses on
short To store and display the frequency with the RMS
value and the DC component, then, on a 2nd
consecutive page, the crest factor CF.
Note: the values displayed are those measured
before the HOLD key is pressed
20
English Clamp Multimeter METRACLIP 87
3 USE
3.1 COMMISSIONING
Insert the batteries supplied with the device as follows:
1. Using a screwdriver, unscrew the screw of the battery compartment cover
(item 1) on the back of the housing and open it.
2. Place the 4 batteries in the compartment (item 2), taking care to get the
polarities right.
3. Close the battery compartment cover and screw it to the housing.
21
Clamp Multimeter METRACLIP 87 English
3.4 CONFIGURATION
As a safety measure, and to avoid repeated overloads on the inputs of the device,
we recommend performing configuration operations only when the device is
disconnected from all dangerous voltages.
1. in the OFF position, hold the key down while turning the switch to
, until the "full screen" display ends and a beep is emitted, to enter the
configuration mode. The display unit indicates the percentage overshoot to
apply to the measured current to determine the measurement triggering
threshold.
Remark : The value stored by default is 10%, representing 110% of the established
current measured. The possible values are 5%, 10%, 20%, 50%, 70%, 100%,
150%, and 200%.
2. To change the threshold, press the key. The value flashes: each
press on the key displays the next value. To record the chosen
threshold, apply a long press (>2s) on the key. A confirmation beep is
emitted.
22
English Clamp Multimeter METRACLIP 87
To exit from the programming mode, turn the switch to another setting. The chosen
threshold is stored (emission of a double beep).
1. In the OFF position, hold the key down while turning the switch to
, until the end of the "full screen" display and the emission of a beep,
to enter the configuration mode. The display unit then indicates the
interval of recording of the data in memory.
Remark: the default value is 60 seconds. Possible values range from 1 second to
600 seconds (10 minutes).
2. To change the recording interval, press the key. The right-hand digit
blinks: each press on the key increments its value. To go to the next
digit, apply a long press (>2s) to the key.
When the desired unit is displayed, turn the switch to another setting. The unit
chosen is stored (emission of a double beep).
In the OFF position, hold the key down while turning the switch to
.
The device emits a beep after erasing the records in memory. The "rSt"
and "rEC" symbols are displayed. The device then switches to normal
continuity measurement.
We recommend not having any voltage on the input terminals while doing this.
After 2 s, the clamp emits a double beep, then all of the symbols of the
screen are displayed until the key is released. The default parameters
are then restored:
Recording interval =60 seconds
True Inrush triggering threshold =10%
23
Clamp Multimeter METRACLIP 87 English
24
English Clamp Multimeter METRACLIP 87
- in AC and AC+DC
Display Quantity
Remark: the measurement results are optimal when the conductor is centred in the
jaws (aligned with the centring marks).
3.8.1 AC measurement
For an AC current measurement, proceed as follows:
1. Set the switch to and select AC by pressing the key. The AC
symbol is displayed.
2. Encircle only the conductor concerned with the clamp ;
26
English Clamp Multimeter METRACLIP 87
Display Quantity
st
1 row RMS current A RMS
27
Clamp Multimeter METRACLIP 87 English
Display Quantity
st
1 row Current A RMS
28
English Clamp Multimeter METRACLIP 87
- in AC and AC+DC :
Display Quantity
1. Set the switch to then encircle only the conductor concerned with
the clamp.
2. Effect a long press on the key. The InRh symbol is displayed,
then the triggering threshold. The clamp then awaits detection of the
True-Inrush current.
"------" is displayed and the "A" symbol flashes (central row of the
display).
3. After detection and acquisition for 100 ms, the RMS value of the True-
Inrush current is displayed, along with the PEAK+/PEAK- values
subsequently.
4. A long press on the key or a change of function leads to exiting
from the True-Inrush mode.
Remark : the triggering threshold in A is 10A if the initial current is zero (starting of
installation); it is that set in the configuration (see §3.4.2) for an established current
(overload in a installation)..
29
Clamp Multimeter METRACLIP 87 English
Display Quantity
st
1 row “Inrh”
- PEAK display :
Display Quantity
30
English Clamp Multimeter METRACLIP 87
For the power factor (PF), the displacement power factor (DPF) and the powers VA
and var measurement is possible only in AC or AC+DC.
Display Quantity
st
1 row Active power W (DC, AC or AC+DC)
31
Clamp Multimeter METRACLIP 87 English
Reminder : the arrow on the jaws of the clamp (see the diagram below) must point
in the presumed direction of flow of the current from the source (producer) to the
load (consumer)
32
English Clamp Multimeter METRACLIP 87
Remark : You can also measure the three-phase power on a balanced 4-wire
network by proceeding in the same way, or by proceeding as for the measurement
on a single-phase network, then multiplying the value found by three.
33
Clamp Multimeter METRACLIP 87 English
3. Connecting the black lead to the COM terminal and the red lead to
«+»;
4. Place the test probes or the crocodile clips of the black lead on the
neutral (N), then those of the red lead on the L phase;
5. Place the clamp around the single conductor concerned, respecting
the direction (see §3.10);
6. To access the metering, press the key :
The sequence of use is as follows :
I- ---> ---> --->I
I<-------------------------------------|
1 : hours (h)
2 : minutes (n)
3 : seconds (s)
34
English Clamp Multimeter METRACLIP 87
The duration of the metering uses the following format: XXXh (for hours) XXm (for
minutes) XXs (for seconds)
N.B. Beyond 999h 59m 59s "---h--m--s" is displayed, but the internal metering
duration keeps running correctly.
Conventions :
Load designates the energy received by the load or consumed (W+)
Load C designates the capacitive reactive energy (W+ and var-)
Load L designates the inductive reactive energy (W+ and var+)
Supp designates the energy generated by the load (W-)
Supp designates the capacitive reactive energy (W- and var-)
Supp L designates the inductive reactive energy (W- and var+)
35
Clamp Multimeter METRACLIP 87 English
36
English Clamp Multimeter METRACLIP 87
37
Clamp Multimeter METRACLIP 87 English
1. Set the switch to and press and hold (>2s) the key. The
THDf, THDr and V RMS symbols are displayed.
2. Connect the black lead to the COM terminal and the red lead to «+»;
3. Place the test probes or the crocodile clips on the terminals of the
circuit to be measured;
38
English Clamp Multimeter METRACLIP 87
1. Set the switch to and press and hold (>2s) the key. The
THDf, THDr and A RMS symbols are displayed.
2. Apply the clamp to only the conductor concerned.
39
Clamp Multimeter METRACLIP 87 English
40
English Clamp Multimeter METRACLIP 87
The device allows recording of the data/measurements acquired, using the REC
function. The default recording interval is 60 seconds. It can be set to from 1
second to 600 seconds (10 minutes) in set-up (see §3.4.3).
1. In the function being measured, apply a long press (> 2s) to the
key. The REC symbol is displayed. Recording of the measurements
starts. The data recorded are in the format: "MAX value – AVG Value
– MIN Value – Unit – Mode" (AC, DC, or AC+DC)
2. To stop recording, apply a long press (>2s) to the key. The REC
symbol disappears.
Type of data Max. number of Max. recording time Max. recording time at
records at 1s intervals 600s intervals (10 mn)
V, A, Ω 3000 16 minutes 160 hours
W 3000 3.5 minutes 35 hours
11 minutes
THD 3000 55 hours
(interval 2 s)
Harmonics 3000 8 minutes 80 hours
41
Clamp Multimeter METRACLIP 87 English
42
English Clamp Multimeter METRACLIP 87
43
Clamp Multimeter METRACLIP 87 English
1.3 The device is connected with the PC, after entering password
« 0000 » :
44
English Clamp Multimeter METRACLIP 87
45
Clamp Multimeter METRACLIP 87 English
46
English Clamp Multimeter METRACLIP 87
47
Clamp Multimeter METRACLIP 87 English
1) In the « Matérial » folder, the port COM number dedicated to the device is
displayed (COM18 in the example).
48
English Clamp Multimeter METRACLIP 87
When choose this message, a windows is displayed to ask device PIN code. You
must enter « 0000 ». Then choose « Next » to validate the connection.
49
Clamp Multimeter METRACLIP 87 English
50
English Clamp Multimeter METRACLIP 87
In PAT software, the connection is established. All the informations of the device
are displayed in the following windows.
Remark : this process must be made only at the first connection. Parameter are
stored in the PC for next connections.
3.1 The device is connected. Display the records stored in the device.
Select the record to be transferred.
51
Clamp Multimeter METRACLIP 87 English
3.2 Transfert of the selected record from the device to PAT software.
3.3 The data are recovered in PAT software. Display of the data in
Text mode, in the format « date – time – MIN – AVG – MAX ».
Nota : MAX,AVG and MIN values are calculated with values measured between 2
records spaced with record interval value.
52
English Clamp Multimeter METRACLIP 87
53
Clamp Multimeter METRACLIP 87 English
54
English Clamp Multimeter METRACLIP 87
Note (1) Above 1000V, a repetitive beep indicates that the voltage being
measured is greater than the safety voltage for which the device is guaranteed. The
display indicates "OL"
Note (1) -Above 1000V, a repetitive beep indicates that the voltage being measured
is greater than the safety voltage for which the device is guaranteed. The display
indicates "OL".
- Bandwidth in AC = 3 kHz
Note (2) Any value between zero and the min. threshold of the measurement range
(0.15V) is forced to "----" on the display
56
English Clamp Multimeter METRACLIP 87
Note (1) - The display indicates "OL" above 1400V (in PEAK mode).
- Above 1000V (DC or RMS), a repetitive beep indicates that the voltage
being measured is greater than the safety voltage for which the device is
guaranteed.
- Bandwidth in AC = 3 kHz
Note (2) - Any value between zero and the min. threshold of the measurement
range (0.15V) is forced to "----" on the display.
57
Clamp Multimeter METRACLIP 87 English
Note (2) - The residual current at zero depends on the remanence; It can be
corrected by the “DC zero” function of the HOLD key.
Note (1) - The display indicates "OL" above 1500A (in PEAK mode). The "- "
and "+" signs are not managed.
- Bandwidth in AC = 2 kHz
Note (2) - Any value between zero and the min. threshold of the measurement
range (0.15V) is forced to “----“ on the display.
- Residual current at zero <150mA..
58
English Clamp Multimeter METRACLIP 87
Uncertainties
(2) ± (1% R+10 pt) ± (1% R +3pt)
(zero corrected)
Resolution 0.01A 0.1A 1A
Note (1) - The display indicates "+OL" above 1500A (in PEAK mode). The "-" and
"+" signs are not managed.
- Bandwidth in AC = 2 kHz
Note (2) - In AC, any value between zero and the min. threshold of the
measurement range (0.15A) is forced to "----" on the display.
- Residual current at zero:
• In DC: depends on the remanence. Can be corrected by the "DC
zero" function of the HOLD key
• In AC: <150mA
59
Clamp Multimeter METRACLIP 87 English
Specified
measurement range 0 to 100% of the measurement range
(from 5V or 5A)
Uncertainties (zero
± (2% R + 2 pt) ± (5% R +2 pt) ± (10% R + 2 pt)
corrected in A DC)
Resolution 1 pt
Remark : If one of the terms for the calculation of the RIPPLE is displayed as "OL",
or forced to zero, the RIPPLE displayed is an indeterminate value, "----".
Note (1) - Above the maximum display value, the display unit indicates "OL".
60
English Clamp Multimeter METRACLIP 87
10,00 kW
Measurement 5 W to 100,0 kW to
to 1000 kW (1)
range (2) (4) 9999 W 999,9 kW
99,99 kW
Specified 1 to 100% of the
measurement measurement 0 to 100% of the measurement range
range range
Uncertainties
± (2% R +10 pt) ± (2% R +3 pt)
(3) (7)
Resolution 1W 10W 100W 1000W
Note (1) - Display of O.L or ± O.L above 1000 kW in single-phase (1000V x 1000A).
- Bandwidth in AC in voltage = 3 kHz, in current = 2 kHz
61
Clamp Multimeter METRACLIP 87 English
Note (7) - In balanced three-phases, with deformed signals (THD and harmonics),
uncertainties are guaranted since Ф > 30°. Additionals errors are following ,
depending of THD :
Add +1% for 10% < THD < 20%
Add +3% for 20% < THD < 30%
Add +5% for 30% < THD < 40%
10,00 kW
Measurement 5 W to 100,0 kW to 1000 kW to
to
range (2) (4) 9999 W 999,9 kW 1500 kW (1)
99,99 kW
62
English Clamp Multimeter METRACLIP 87
Specified
1 to 100% of the
measurement 0 to 100% of the measurement range
measurement range
range
Uncertainties until 1000A until 1000A
(3) (7) ± (2% R +10 pt) ± (2% R +3 pt)
from 1000A to 1500A from 1000A to 1500A
± (2.5% R +10 pt) ± (2.5% R +3 pt)
Resolution 1W 10W 100W 1000W
Note (1) - Display of O.L above 1000 kVA in single-phase (1000 V x 1000 A).
- Bandwidth in AC in voltage = 3 kHz, in current = 2 kHz
Notes (2), (3) and (4) of the previous § apply.
63
Clamp Multimeter METRACLIP 87 English
Note (1) - Display of O.L above 1500 kVA in single-phase (1000 V x 1500 A).
- Bandwidth in AC in voltage = 3 kHz, in current = 2 kHz
Notes (2), (3) and (4) of the previous § apply.
10,00 kvar
Measurement 5 var to 100,0 kvar to
to 1000 kvar (1)
range (2) (4) 9999 var 999,9 kvar
99,99 kvar
Specified
1 to 100% of the
measurement 0 to 100% of the measurement range
measurement range
range
Uncertainties
± (2% R +10 pt) ± (2% R +3 pt)
(3) (8)
Resolution 1 var 10 var 100 var 1 kvar
Note (1) - Display of O.L above 1000 kvar in single-phase (1000 V x 1000 A).
- Bandwidth in AC in voltage = 3 kHz, in current = 2 kHz
Notes (2), (3) and (4) of the previous § apply.
Note 5 - In single-phase, the sign of the reactive power is determined by the
phase lead or lag between the U and I signs, while in balanced three-
phase, it is determined by the calculation on the samples.
Note 6 - Signs of reactive powers according to the four-quadrant rule (§4.2.12):
Quadrant 1 : Reactive power Q sign +
Quadrant 2 : Reactive power Q sign +
Quadrant 3 : Reactive power Q sign -
Quadrant 4 : Reactive powerQ sign -
Note (1) - Display of O.L above 1500 kvar in single-phase (1000 V x 1500 A).
- Bandwidth in AC in voltage = 3 kHz, in current = 2 kHz
Notes (2), (3) , (4), 5, 6 and (8) of the previous § apply.
Note (1) - If one of the terms in the calculation of the power factor is displayed as
"OL", or forced to zero, the display of the power factor is an indeterminate
value "----".
Note (1) - If one of the terms in the calculation of the DPF is displayed as
"OL", or forced to zero, the display of the DPF is an indeterminate
value "----".
Note : - The display is "----" if the input signal is too low (U<8V or I<9A) or if the
frequency is less than 5Hz.
67
Clamp Multimeter METRACLIP 87 English
Note : - The display is "----" if the input signal is too low (U<8V or I<9A) or if the
frequency is less than 5Hz.
68
English Clamp Multimeter METRACLIP 87
Mechanical
Free fall: 2m (in accordance with standard IEC-68-2-32)
strength:
A - 0.1%R + 2 pts
0.2%R 0.3%R + 2 pts
Ω 0.25%R 0.5%R + 2 pts
W
10 Hz...1 kHz 1%R + 1 pt 1%R + 1 pt
V
1 kHz...3 kHz 8%R + 1 pt 9%R + 1 pt
Frequency
10 Hz…400 Hz 1%R + 1 pt 1%R + 1 pt
A
400 Hz…2 kHz 4%R + 1 pt 5%R + 1 pt
Position of the Any position on
conductor in the the internal 4%R + 1 pt
A-W 2%R
jaws perimeter of the Full-scale
(f≤400 Hz) jaws
Adjacent Conductor
conductor touching the
carrying a external A-W 40 dB 30 dB
current of 150 A perimeter of the
DC or RMS jaws
Conductor
0-500 A DC or
enclosed by the V < 1 pt 1 pt
RMS
clamp
Application of a
0-1000 V DC or
voltage of the A-W < 1 pt 1 pt
RMS
clamp
1.4 to 3.5
limited A (AC-AC+DC) 1%R 3%R + 1 pt
Peak factor (1)
to 1500 A peak V (AC-AC+DC) 1%R 3%R + 1 pt
1400 V peak
0.7 and I ≥ 5A 1%R + 1 pt
PF (inductive
0.5 and I ≥ 10A W 0.5%R 3%R + 1 pt
and capacitive)
0.2 and I ≥ 20A 8%R + 1 pt
70
English Clamp Multimeter METRACLIP 87
4 MAINTENANCE
The instrument has no parts that can be replaced by personnel who are not trained
and approved. Any non-approved repair or other work, or replacement of a part by
an "equivalent", may severely compromise safety.
4.1 CLEANING
• Disconnect everything connected to the device and set the switch to OFF.
• Use a soft cloth moistened with soapy water. Rinse with a damp cloth and
dry quickly using a dry cloth or forced air.
• Dry perfectly before putting back into use.
The symbol indicates that the batteries are spent. When this symbol appears
on the display unit, the batteries must be replaced. The measurements and
specifications are no longer guaranteed.
4.4 REPAIR
For all repairs before or after expiry of warranty, please return the device to your
distributor.
71
Clamp Multimeter METRACLIP 87 English
5 WARRANTY
Except as otherwise stipulated, our warranty is valid for three years starting from
the date on which the equipment was sold. Extract from our General Conditions of
Sale provided on request.
The warranty does not apply in the following cases:
• Inappropriate use of the equipment or use with incompatible equipment;
• Modifications made to the equipment without the explicit permission of the
manufacturer’s technical staff;
• Work done on the device by a person not approved by the manufacturer;
• Adaptation to a particular application not anticipated in the definition of the
equipment or not indicated in the user’s manual;
• Damage caused by shocks, falls, or floods.
6 DELIVERY CONDITION
72
7 ADDRESSES
Product Support
If required, please contact: