23-18013-00 BD Accuri C6 Plus System User's Guide
23-18013-00 BD Accuri C6 Plus System User's Guide
User’s Guide
23-18013-00
1/2016
bdbiosciences.com
[email protected]
Copyrights
© 2016, Becton, Dickinson and Company. All rights reserved. No part of this publication may be reproduced,
transmitted, transcribed, stored in retrieval systems, or translated into any language or computer language, in
any form or by any means: electronic, mechanical, magnetic, optical, chemical, manual, or otherwise, without
prior written permission from BD Biosciences.
The information in this guide is subject to change without notice. BD Biosciences reserves the right to change
its products and services at any time to incorporate the latest technological developments. Although this guide
has been prepared with every precaution to ensure accuracy, BD Biosciences assumes no liability for any errors
or omissions, nor for any damages resulting from the application or use of this information. BD Biosciences
welcomes customer input on corrections and suggestions for improvement.
Trademarks
Alexa Fluor® is a registered trademark of Life Technologies Corporation.
Cy™ is a trademark of GE Healthcare. Cy™ dyes are subject to proprietary rights of GE Healthcare and
Carnegie Mellon University, and are made and sold under license from GE Healthcare only for research and in
vitro diagnostic use. Any other use requires a commercial sublicense from GE Healthcare, 800 Centennial
Avenue, Piscataway, NJ 08855-1327, USA.
© 2016 BD. BD, the BD Logo, and all other trademarks are property of Becton, Dickinson and Company.
Regulatory information
For Research Use Only. Not for use in diagnostic or therapeutic procedures.
FCC information
WARNING: Changes or modifications to this unit not expressly approved by the party responsible for
compliance could void the user’s authority to operate the equipment.
NOTICE: This equipment has been tested and found to comply with the limits for a Class A digital device,
pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against
harmful interference when the equipment is operated in a commercial environment. This equipment generates,
uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction
manual, may cause harmful interference to radio communications. Operation of this equipment in a
residential area is likely to cause harmful interference in which case the user will be required to correct the
interference at his or her own expense. Shielded cables must be used with this unit to ensure compliance with
the Class A FCC limits. This Class A digital apparatus meets all requirements of the Canadian Interference-
Causing Equipment Regulations. Cet appareil numérique de la classe A respecte toutes les exigences du
Réglement sur le matériel brouilleur du Canada.
History
Chapter 1: Introduction 7
About this guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Additional documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Safety symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Technical support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Index 279
1
Introduction
This chapter covers the following topics:
In this guide This guide provides information for setting up and running the
BD Accuri™ C6 Plus system during a typical workflow, whether
you are acquiring samples manually or using a BD CSampler™
Plus accessory.
Additional documentation
Introduction This topic describes additional documentation available for the
BD Accuri C6 Plus system.
Documents The following table lists the available documents for the BD Accuri
C6 Plus system.
Document Description
BD Accuri™ C6 Plus Provides the site requirements. Read this
System Site Preparation guide before the system is installed.
Guide
Safety symbols
Introduction This topic describes the safety symbols used in this guide. For a
complete description of all safety hazards, see the BD Accuri C6
Plus System Safety and Limitations Guide.
Safety symbols The following table describes the symbols used in this guide.
Symbol Meaning
Caution. Indicates the need for the user to consult the
system user’s guide for important cautionary information
such as warnings and precautions that cannot, for a
variety of reasons, be presented on the device itself.
Biological hazard
Electrical hazard
Laser hazard
Technical support
Introduction This section describes how to get technical support.
C6 Plus system The C6 Plus system includes the BD Accuri C6 Plus flow
cytometer, an all-in-one desktop workstation running BD Accuri
C6 Plus software (or BD CSampler Plus software) for acquisition
and analysis, and the optional BD CSampler Plus. The system also
includes BD™ CS&T RUO beads for instrument QC. An optional
barcode reader can be attached to a USB port on the computer.
Reagents BD CS&T RUO beads are used to check and monitor the
cytometer performance. We recommend running instrument QC
daily, ensuring that the test passes before running samples.
CSampler Plus The BD CSampler Plus is an optional accessory that saves you time
option and effort by automating the sample loading step. Place a 24-tube
rack of standard 12 x 75-mm tubes or a 48- or 96-well microplate
on the CSampler Plus and select RUN. The CSampler Plus
automatically agitates the tubes to keep the cells in suspension,
then loads each tube for acquisition. Use the software to program
sample injection probe (SIP) cleaning and rinsing during the run.
Outside of The following figures show the front and back of the cytometer
cytometer and locations of the indicator lights, power button, sample
injection probe (SIP), and connectors.
Front
1 5
2
3
Chapter 2: About the system 17
Back
4
20 BD Accuri C6 Plus System User’s Guide
During operation, the software will notify you if the fluid level for
a fluid bottle is getting low or the waste bottle is getting full.
Always attend to the fluid bottles when instructed.
5 6
3
8
4
Adjusting the Note: This procedure applies to manual acquisition only. If you
height of the are using a CSampler Plus for acquisition, the sample stage was
sample stage removed to accommodate for the CSampler Plus.
The bottom of the sample stage contains a metal platform that can
be adjusted to two different positions. It can be raised or lowered
to accommodate both standard 12 x 75-mm tubes and
BD Trucount™ tubes, respectively.
Three fixed tube locations at the back of the CSampler Plus tray
allow you to load tubes for cleaning, rinsing, and backflushing the
SIP.
Chapter 2: About the system 25
Loading a rack/ Always use the Eject Plate/Rack and Load Plate/Rack buttons in
plate the software to move the CSampler Plus. Never move the
CSampler Plus by hand.
The rack or plate fits one way on the tray, ensuring that
position A1 is in the upper-left corner.
The CSampler Plus positions the selected tube under the SIP
for acquisition. A plate map on the screen allows you to see the
tube that is being acquired. The location flashes blue and is
outlined in red.
Cleaning tube The CSampler Plus is equipped with three fixed tube locations to
locations support the system cleaning functions. Check and replace these
tubes after system startup, after every run, and as needed.
Chapter 2: About the system 27
Location Description
Triangle ( ) Place a tube containing 2 mL of BD FACSClean in
this location. Used for the first tube in a SIP Clean.
About the software BD Accuri C6 Plus software allows you to control the BD Accuri
C6 Plus flow cytometer system to acquire data, generate statistics,
and analyze results.
About BD CSampler BD CSampler Plus software allows you to control the BD Accuri
Plus software C6 Plus flow cytometer system and the BD CSampler Plus to
acquire data, generate statistics, and analyze results.
Software The main C6 Plus software window is called the workspace. The
workspace workspace contains controls and displays that provide access to all
the functions for acquiring and analyzing data.
CSampler Plus The main BD CSampler Plus software window is called the
workspace workspace. The workspace contains controls and displays that
provide access to all functions required for data acquisition and
analysis.
Software menus This topic contains tables that list and describe the menu options
and options for BD Accuri C6 Plus software and BD CSampler Plus software.
File menu
New workspace Opens a new, blank workspace. Replaces any previously open
workspace.
Save Saves the open workspace under the current name. If the file has
not already been named, you will be prompted to name the file.
Save template as Creates a template from the currently open workspace. Markers,
regions, gates, parameter names, sample names, and user-defined
compensation settings are saved without any data points.
Auto-save Data Settings Allows you to enable or disable the auto-save feature.
Import FCS File Imports an FCS file previously exported from another file to the
currently open workspace. Only FCS files created using
BD Accuri C6 Plus software and BD Accuri™ C6 software can
be imported into the software.
32 BD Accuri C6 Plus System User’s Guide
Export ALL Samples as Saves all of the data wells as individual FCS 3.1 files in the folder
FCS FCS Exports on the computer desktop. Exported files are
compatible with off-line analysis programs such as FCS Express,
FlowJo, and WinList.
Export Plot Data as CSV Saves an individual file in CSV format for further analysis in
spreadsheet programs. All data for every event in the selected
plot is exported.
Export Sample Settings Exports sample settings as a CSV file for viewing data in a
spreadsheet. Sample settings include acquisition criteria, sample
names, parameter names, and compensation values. Available in
BD CSampler Plus software only.
Set Plot Drag and Drop Allows plot format to be toggled between .png (low resolution)
Format or .eps (high resolution).
Edit menu
Rename Parameters/Color Allows you to rename individual parameters for either the
Compensation current sample or all samples at the same time. Also allows you
to view and/or change the color compensation settings.
Display menu
Auto-Select Next Well Opens a dialog to configure whether the software automatically
selects wells vertically or horizontally, or not at all at the
completion of each sample acquisition.
Remove All VirtualGain™ Removes all VirtualGain settings from the entire workspace.
VirtualGain settings are removed from all wells of data.
Instrument menu
Run Backflush cycle Runs the backflush cycle to clean the SIP and remove clogs at the
base of the SIP.
Extended Flow Cell Clean Cleans the flow cell for an extended time.
34 BD Accuri C6 Plus System User’s Guide
Auto-adjust volume Adjusts the fluidics to ensure that the cytometer accurately
settings reports the correct volume of sample.
Update firmware Used to update the firmware. Use only when directed by a BD
representative.
Remote Interfacing Optional feature available with automation key. Allows control
Module of the cytometer from a remote location.
About menu
Technical Support Opens a dialog that displays information about the current
Information version of C6 Plus software (or CSampler Plus software) and the
cytometer. Each time an activation key is used to install a new C6
Plus software component, the dialog is updated to reflect the
change.
Create Cytometer Log Creates and opens a Microsoft Notepad document with
cytometer- and software-specific information for use by a
technical support representative. The file is saved to
C:\Cytometer Support Files.
3
Startup and shutdown
This chapter covers the following topics:
Startup workflow
Introduction This topic describes the workflow for daily system startup.
Workflow The following table lists the tasks that should be performed each
day to start up the system and prepare it for running samples.
Step See
1 Filling the fluid bottles (page 36)
About the fluid Visually check all the bottles at the start of each day and fill the
bottles sheath, BD FACSClean, and detergent solution bottles, as needed.
Required materials The following fluids are used with the system:
• Sheath fluid
Add one bottle (5 mL) of BD Sheath Additive to 1 liter of 0.2-
µm filtered DI water.
38 BD Accuri C6 Plus System User’s Guide
About the waste Empty the waste bottle daily or when prompted by the software to
bottle prevent spillover and possible biological safety risk.
About startup During startup, the power indicator flashes blue and the traffic
light turns yellow while the laser warms up and the cytometer
flushes the fluid lines with fresh sheath fluid. This process takes
approximately 15 minutes.
Note: Do not open the lid of the cytometer during the startup
process. Opening the lid interrupts the laser warm-up and extends
the time before samples can be acquired.
Once the fluid lines are flushed, the traffic light turns green
and the software displays the message Cytometer is connected
and ready. The power indicator on the front of the cytometer
turns solid blue.
Starting the To start up the cytometer when using the CSampler Plus:
cytometer when 1. Press the power button on the front of the cytometer to turn it
using the CSampler on.
Plus
Once the lasers are warmed up and the fluid lines are flushed,
the traffic light turns green and the software displays the
message Cytometer is connected and ready. The power
indicator on the front of the cytometer turns solid blue.
Chapter 3: Startup and shutdown 43
2. Click Eject Plate and load the following fresh cleaning tubes in
the designated fixed locations on the CSampler Plus tray:
About shutdown When you turn off the power to the cytometer, the Clean Fluidics
cycle is run automatically. The cycle takes about 13 minutes to
complete. See Cleaning the fluidics (page 200) for details. At the
end of the cycle, leave the SIP in the tube of water to keep it from
drying out.
Clean the sample stage at shutdown and whenever you see stains
or spills on it.
44 BD Accuri C6 Plus System User’s Guide
When this occurs, the cytometer takes additional time to run the
Clean Fluidics cycle at startup. The startup time can take
approximately 25 minutes.
Shutting down the To shut down the cytometer when running manually:
cytometer when 1. Use a disposable towel or wipe moistened with BD FACSClean
running manually or a 10% bleach solution to wipe down the sample stage.
Follow with a wipe moistened with water.
Shutting down the To shut down the cytometer when using the CSampler Plus:
cytometer when 1. Ensure the following cleaning tubes are loaded in the
using the CSampler designated fixed locations on the CSampler Plus tray:
Plus
• A tube with 2 mL of BD FACSClean in the triangle ( )
• A tube with 2 mL of DI water in the circle ( )
• A tube with 2 mL of DI water in the square ( )
Chapter 3: Startup and shutdown 45
QC overview
Introduction This topic describes the quality control module and recommended
workflow.
About quality The instrument QC module allows you to perform quality control
control on the system. Run quality control daily using the BD CS&T RUO
beads to check and monitor the instrument’s performance. The
CS&T RUO beads have a known median fluorescence intensity
(MFI) and distribution (rCV), and allow you to characterize, track,
and report measurements made by the cytometer.
During instrument QC, the software sets regions around the dim
beads and the mid + bright beads. The locations of the regions are
based on target values, not on the actual locations of the bead
populations. The system measures the brightness and distribution
of the bright beads and compares the results to expected values.
Instrument sensitivity is also calculated. In addition, the
compensation values are updated based on the CS&T RUO bead
results. When the QC test is complete, a Passed or Failed result is
displayed.
Daily workflow The following table shows the recommended daily workflow for
running QC and samples.
Step Description
1 Prepare BD CS&T RUO beads according to the BD CS&T
RUO Beads data sheet.
Step Description
3 Prepare samples. See the appropriate reagent kit package
insert or data sheet.
Running instrument QC
Introduction This topic describes how to run the BD CS&T RUO beads to
check the instrument performance.
Before you begin Prepare BD CS&T RUO beads according to the instructions in the
BD CS&T RUO Beads data sheet.
If you are using the Selectable Lasers Module, the laser will
automatically switch to the standard 3 blue 1 red configuration for
instrument QC. Ensure that the standard optical filters are
installed when running instrument QC. When QC is complete and
you close the QC module, the system will switch back to the
configuration that was previously selected, if different from 3 blue
1 red.
50 BD Accuri C6 Plus System User’s Guide
2. Select the bead lot from the BD CS&T Bead Lot menu.
If you need to install a bead lot file for a new lot of beads, see
Install a new bead lot (page 51).
Install a new bead If this is the first time you are running a bead lot, you will need to
lot install the BD CS&T RUO bead lot file. The bead lot file contains
information specific to the given lot of beads, such as expiration
date, rCV, target MFIs, and sensitivity specifications. Once
installed, the bead lot number will be available to select from the
BD CS&T Bead Lot menu for subsequent QC runs.
2. You can either locate a bead lot file on your computer or scan
the bead lot barcode.
More information • Adjusting the bead regions and markers (page 53)
• Viewing QC results (page 58)
• Removing bead lots from the CS&T bead lot menu (page 185)
• QC troubleshooting (page 231)
Chapter 4: Quality Control 53
The software sets the region and markers for the bright beads
based on target values, not on the actual locations of the
populations. Therefore, you will need to check and optimize the
region and markers, as necessary.
About adjusting Once the bead acquisition is complete and while the QC results are
the regions and displayed, you can adjust the regions and markers to optimize
markers them.
Adjusting the Adjust both the dim and mid + bright regions.
regions in the dot
plot To adjust the bead regions:
1. Use the zoom tool to zoom in on the bead data before
adjusting the regions. See Zooming plots (page 56).
The region appears with a bold red outline and eight handles.
Unzoomed Zoomed
3. Adjust the region for the dim beads and the region for the
mid + bright beads. Exclude the doublet population to the
right of the mid + bright beads.
Adjusting the To adjust the bright-bead peak markers in all four histograms:
markers in the 1. Use the zoom tool to zoom in on the bead populations before
histogram plots adjusting the markers. See Zooming plots (page 56).
The marker appears with a bold red outline and two handles.
Unzoomed Zoomed
Zooming plots You can zoom in on data to make it easier to see where to set the
markers. If the plots are zoomed when you close the QC report,
the report will be saved with the zoomed plots.
zoom expand
tool tool
Chapter 4: Quality Control 57
2. Click and drag over the area that you want to zoom in on.
Repeat as necessary.
3. Click the expand tool to zoom out once. Or, click the Default
Zoom button to go back to the original data range.
Next step • If the result passed, you are ready to run test samples. See
Manual data acquisition (page 67) or CSampler Plus data
acquisition (page 87).
• If the result failed, even after you adjusted the bead regions
and markers, check the QC messages for additional help. See
QC troubleshooting (page 231) for possible causes and
solutions.
• To rerun the beads, mix the tube, click New QC, then click
RUN.
58 BD Accuri C6 Plus System User’s Guide
Viewing QC results
Introduction This topic describes the information that appears on the QC results
screen.
About instrument The instrument QC results are displayed as Passed or Failed. If the
QC results result fails, adjust the mid + bright bead region and the
fluorescence markers. The QC messages that appear below the QC
Report (results) table provide information on the conditions that
the system encountered.
QC results screen While the QC results window is open, you can make adjustments
to the regions and markers to optimize them. See Adjusting the
bead regions and markers (page 53). Once you close the window
or click on the History: Levey-Jennings tab, the QC results are
locked and no longer editable; however you can add comments.
You can also open and view previous reports and enter comments.
Chapter 4: Quality Control 59
Instrument noise The system measures the background instrument noise and
measurement plots displays this in the Instrument Noise Measurement plots. Markers
are automatically set in the lower channels around the target
locations for noise peaks. Do not adjust the noise peak markers.
BD CS&T Beads The system acquires 25,000 total events and displays the data in
plots the BD CS&T Beads plots. Regions are set around the dim bead
population and the mid + bright bead population in the dot plot
based on target values. Markers are set around the bright bead
target location in the histogram for each fluorescence parameter.
You can adjust the regions and/or the histogram markers. See
Adjusting the bead regions and markers (page 53) for information.
QC Report table The QC Report table displays the values and Pass/Fail result for
each parameter. If any of the results for an individual parameter
60 BD Accuri C6 Plus System User’s Guide
fails, the failed parameter result appears in red and the overall
instrument QC result fails.
Measurement Description
Bright Bead Median Measured median value for the bright bead
population
Comments The Comments section allows you to add comments about the run.
You can also add comments before you run the beads or after the
QC window is closed, by viewing previous QC results.
Viewing previous You can view the QC report for any previous QC run.
QC results
To view QC results from a previous run:
1. Click View Previous QC in the upper-right corner of the
Instrument QC tab.
2. Select a run, then click OK to view the results for that run.
Printing QC reports You can print QC reports after the run is complete. You can also
print the reports from previous QC runs.
Next step • If instrument QC passed, you are ready to run controls and
test samples. See Manual data acquisition (page 67) or
CSampler Plus data acquisition (page 87). Prepare the control
along with the test samples. See the appropriate reagent
package insert or data sheet for use for sample preparation
information.
• If instrument QC still fails after adjusting regions and markers,
check the QC messages at the bottom of the screen for more
information. Refer to QC messages (page 232) for possible
causes and solutions.
About Levey- Levey-Jennings reports track QC data over time, allowing you to
Jennings reports view the system’s performance and ensure that the system is
reproducing consistent results. The graphs in the report show you
random errors or shifts and trends in the data for each parameter,
and help you diagnose possible problems with the system.
The graphs show the bright bead median values, %rCV, and
sensitivity with standard deviations for FSC, SSC, FL1, FL2, FL3,
and FL4.
2. Select the number of days from the View menu in the upper-
left corner of the tab.
You can view data for the last 30, 60, or 90 days. Use the
scroll arrows to the right of the View menu to change the start
date.
3. Scroll down to view all the graphs for FL1–FL4, FSC, and SSC
median, FL1–FL4 rCVs, and sensitivity for each parameter.
Chapter 4: Quality Control 65
The graphs show CS&T RUO bead data plotted over time. If
multiple bead lots were used, a different symbol and different
color line appear for each lot.
Collect tab
Introduction This topic describes the Collect tab in BD Accuri C6 Plus software.
About the Collect The Collect tab allows you to set data collection criteria, start and
tab stop data acquisition, and view data on collected samples. You can
use pre-defined templates for acquisition or create your own
workspace with custom acquisition criteria. The tab contains
buttons and controls for performing the following functions:
• Acquiring samples
• Creating plots (histogram, dot, or density) for viewing data
• Setting stop criteria (run limits) and thresholds
• Controlling the fluidics
• Using regions and markers to create gates and obtain statistics
• Saving and print plots and data
• Accessing the following analytic functions:
– Creating nested gates
– Setting and/or adjusting fluorescence compensation
– Generating statistics
Chapter 5: Manual data acquisition 69
Viewing the Collect The Collect tab is displayed when the software opens. You can also
tab view the tab by clicking on Collect from any of the other tabs
(Analyze, Statistics, or Batch Analysis).
• Data display. Area on the right that shows sample data in plots
and in a statistics table.
70 BD Accuri C6 Plus System User’s Guide
Collect tab controls The following table describes each of the controls and indicators in
the Collect tab:
Control Description
Sample naming field Text box for naming the sample.
Sample grid Matrix laid out in the configuration of a 96-well plate to help organize
experiments and collect data from sample tubes. Each sample is
acquired into its own well in the sample grid. The color-coded wells can
be filled with data in any order.
White. Does not contain data.
Blue. Contains data.
Red outline. Currently selected for viewing or collecting data.
Traffic light and Indicator that displays software readiness and system messages. Before
message data collection can begin, the software must display a green traffic light
with the message Cytometer is connected and ready.
The traffic light status is color-coded:
Green. Software is ready to collect data or is collecting data.
Yellow. Cytometer is preparing to perform an action or a non-critical
error has occurred.
Red. Critical error has occurred.
Run Settings Contains a set of controls that allow you to define criteria for
automatically stopping data collection. See Run settings (page 73) for
more information.
Fluidics Contains a set of controls for defining flow rate and core size. See
Fluidics rate (page 75) for more information.
Threshold Allows you to set the event threshold to eliminate debris and noise from
sample data. The default value is 80,000 on FSC-H. See Threshold
(page 79) for more information.
Chapter 5: Manual data acquisition 71
Control Description
RUN/PAUSE/ADD Button that performs the following functions:
TO
RUN. Starts sample acquisition.
PAUSE. Pauses acquisition. Click ADD TO to resume acquisition.
ADD TO. Allows you to collect more sample data in a well that already
contains data.
Set Color Opens the Color Compensation dialog for correcting fluorescence
Compensation spillover. See Correcting fluorescence spillover (page 149) for more
information.
Acquisition Displays the following information about the most recent acquisition
counters for the selected well (Last Run) and all acquisitions for the selected well
(Cumulative) in real-time:
Events. Number of events acquired.
Time. Elapsed acquisition time.
Microliters. Volume of sample acquired.
Events/Sec. Events acquired per second. When the run is completed,
this is the average value.
Events/µL. Events acquired per microliter. When the run is completed,
this is the average value.
Delete Events Permanently deletes all events from the current sample. Select the show
warning checkbox to display a warning message before deleting sample
data. Also contains a Data Capacity Used meter that displays the
amount of data storage capacity currently used in the software.
Plots pane Area displaying two rows of plot corrals for graphically viewing data
for the selected sample. Scroll up or down to view additional plots. You
can create multiple plots for each sample. Each plot corral contains
buttons for creating histogram, dot, and density plots. See Plots and
gates (page 156) for more information on plots.
Statistics table Table below the plots that displays statistical information on all plots.
72 BD Accuri C6 Plus System User’s Guide
Data collection Complete the following workflow steps to collect sample data.
workflow
Step Description
1 (Optional) Open a template. See Creating a workspace
template (page 189).
3 Open plots, create and apply gates. See Creating a new plot
(page 157) and Creating and applying gates (page 160).
4 Run the sample. See Running samples using the Collect tab
(page 81).
Run settings
Introduction This topic explains run settings and provides instructions for
enabling and disabling run limits for data collection.
About run settings Set a run limit to indicate when to stop collecting data. You can set
a run limit based on any of the following criteria:
Setting an event To stop the run after a specified number of events have been
run limit collected:
1. Select Run with Limits.
Setting a time run To stop the run after a time has expired:
limit 1. Select Run with Limits.
Setting a volume To stop the run after a specified volume has been collected:
run limit 1. Select Run with Limits.
Fluidics rate
Introduction This topic describes how to set the fluidics rate for data collection.
About fluidics rates The system can accommodate an upper limit of 10,000 events per
second, but we recommend acquiring samples at a rate of 2,500
events per second or less to ensure the best data resolution.
Three preset rates are provided—Slow, Medium, and Fast. You can
also customize the fluidics rate by selecting a rate outside the preset
values.
• Slow is 14 µL/min.
• Medium is 35 µL/min.
• Fast is 66 µL/min.
Note: We recommend starting data collection on Slow and
observing the event rate. You can then switch the setting to
Medium or Fast, if necessary.
Adjusting the core Advanced users can adjust the core size of the sample stream for a
size of the sample range of cell sizes.
stream
To customize the sample core size:
1. Click Set Core Size under Fluidics in the Instrument Control
Panel.
Note: Certain core sizes are not compatible with certain flow
rates, and the software does not allow these combinations to
be set. Use the following table to determine allowable
combinations.
Minimum Maximum
Core size flow rate flow rate
5 10 11
6 10 16
7 10 22
8 10 29
9 10 36
10 10 45
11 10 54
12 10 65
13 10 76
14 12 88
15 14 100
16 15 100
17 17 100
18 19 100
19 22 100
20 24 100
21 26 100
22 29 100
23 32 100
24 35 100
78 BD Accuri C6 Plus System User’s Guide
Minimum Maximum
Core size flow rate flow rate
25 38 100
26 41 100
27 44 100
28 47 100
29 50 100
30 54 100
31 58 100
32 61 100
33 65 100
34 69 100
35 74 100
36 78 100
37 82 100
38 87 100
39 91 100
40 96 100
Chapter 5: Manual data acquisition 79
Threshold
Introduction This topic explains the threshold and provides instructions for
setting a collection threshold.
About thresholds Thresholds eliminate debris and noise from cell samples so that
sample data are not compromised by non-cellular events. By
default, events lower than channel 80,000, for the selected
parameter, are filtered out.
You can change the threshold settings at any time before, during,
or after data acquisition. The primary threshold is the parameter
that triggers data collection. FSC-H is the default primary
threshold. You can optionally set a secondary threshold to filter
out additional data.
All thresholds are set on the Height signal for any given parameter.
For best results when setting or changing thresholds, create a plot
that displays the Height signal for the threshold channel and
observe the effect on the data as the threshold is raised or lowered.
3. Type 80000 in the less than text box to set the threshold
minimum to channel 80,000.
About collecting • Before data collection can begin, the software must display a
sample data green traffic light with the message Cytometer is connected
and ready.
• Each data well holds a maximum of 1 million events. You can
add more events (up to a total of 1 million) to a well already
containing data.
• When a well already contains data, the RUN button becomes
ADD TO.
Note: Run limits may need to be adjusted to accommodate
additional data.
Before you begin Set the acquisition parameters. See Setting acquisition parameters
(page 72).
A1 is selected by default.
2. Enter a sample name in the text box above the 96-well grid.
82 BD Accuri C6 Plus System User’s Guide
3. Resuspend the cells in the sample tube and place the tube on
the SIP.
If you have not yet saved the workspace, you will be prompted
to do so.
The well flashes blue during data collection. After the run limit
is reached, the well stops flashing and remains blue, indicating
that the well contains data.
6. To run the next sample, click the next well in the sample grid.
If necessary, adjust the acquisition parameters, then click
RUN.
Adding more You can collect new samples and add the data to a workspace file
sample data to a that already contains sample data, by adding data to a well that
file already contains data or to an empty well. If the well already
contains data, you may need to increase the number of events
defined for the run limit to allow for more events to be acquired.
If you select an empty well, any plots and gates you created
earlier are still displayed, but they do not contain data.
The cytometer stops sampling from the tube when the run
limit is reached.
Pausing data You can interrupt sample acquisition any time during a run.
collection
To stop a run:
1. Click PAUSE.
If you want to delete the data that was already collected before
you paused, click Delete Events at the bottom of the
Instrument Control Panel, then click RUN to acquire new
data.
Auto-save data If you have the Auto-save Data Settings feature enabled, the
feature software will automatically save the workspace file at the end of
the run. If you make any changes to the workspace after the run is
finished, the changes will not be automatically saved. You will
need to manually save the file. See Saving the workspace file
(page 85).
Saving the To save the workspace file at the end of the run:
workspace file 1. Select File > Save.
Performing a SIP When you finish collecting samples, clean the SIP to ensure cells or
clean other particles are not left in the SIP.
About the Manual The Manual Collect tab allows you to set the data collection
Collect tab criteria, start and stop data acquisition, and view data on collected
samples. You can use pre-defined templates for acquisition or
create your own workspace with custom acquisition criteria. The
tab contains buttons and controls for performing the following
functions:
Viewing the The Manual Collect tab is organized into two major sections:
Manual Collect tab
• Instrument Control Panel. Panel on the left side of the window
that contains controls for collecting data.
• Data display. Large area on the right side of the window that
shows the sample data in plots and in a Statistics table.
90 BD Accuri C6 Plus System User’s Guide
Manual Collect tab The following table describes each of the controls and indicators in
controls the Manual Collect tab.
Control Description
Plate Type List for selecting the plate type.
Load Plate/Eject Moves the sample plate into position to be loaded onto or ejected from
Plate the flow cytometer.
Sample grid Matrix laid out in the configuration of a 96-well plate or a 24-tube
rack to correspond to the sample vessel. Each sample is acquired into
its own well in the sample grid. The color-coded wells can be filled with
data in any order.
White. Does not contain data.
Blue. Contains data.
Red outline. Currently selected for viewing or collecting data.
Traffic light and Indicator that displays software readiness and system messages. Before
message data collection can begin, the software must display a green traffic light
with the message Cytometer is connected and ready.
The traffic light status is color-coded:
Green. Software is ready to collect data or is collecting data.
Yellow. Cytometer is preparing to perform an action, performing an
action other than data collection (such as clean or agitate), or a non-
critical error has occurred.
Red. Critical error has occurred or the CSampler Plus has collided.
Run Settings Contains a set of controls that allow you to define criteria for
automatically stopping data collection. See Run settings (page 73) for
more information.
Chapter 6: CSampler Plus data acquisition 91
Control Description
SIP Rinse Rinses the outside of the SIP between samples.
Agitate Agitates the plate/rack to keep cells in suspension.
Backflush Forces sample out of the SIP.
SIP Clean Performs a SIP Clean to clean the inside of the SIP.
Fluidics Contains a set of controls for defining flow rate and core size. See
Fluidics rate (page 75).
Threshold Allows you to set the event threshold to eliminate debris and noise
from sample data. The default value is 80,000 on FSC-H. See Setting
the threshold (page 107).
Set Color Opens the Color Compensation dialog for correcting fluorescence
Compensation spillover. See Correcting fluorescence spillover (page 149).
Acquisition counters Displays the following information about the most recent acquisition
for the selected well (Last Run) and all acquisitions for the selected well
(Cumulative) in real-time:
Events. Number of events acquired.
Time. Elapsed acquisition time.
Microliters. Volume of acquired sample.
Events/sec. Events acquired per second. When the run is completed,
this is the average value.
Events/µL. Events acquired per microliter. When the run is completed,
this is the average value.
92 BD Accuri C6 Plus System User’s Guide
Control Description
Delete Events Permanently deletes all events from the current sample. Select the show
warning checkbox to display a warning message before deleting sample
data. Also contains a Data Capacity Used meter that displays the
amount of data storage capacity currently used in software.
Plots pane Area displaying two rows of plot corrals for graphically viewing data
for the selected sample. Scroll up or down to view additional plots. You
can create multiple plots for each sample. Each plot corral contains
buttons for creating histogram, dot, and density plots. See Plots and
gates (page 156).
Statistics table Table below the plots that displays statistical information on all plots.
Manual data Complete the following workflow stages to collect sample data
collection with the BD CSampler Plus software.
workflow
Stage Description
1 In the Manual Collect tab, do one of the following:
Select a plate type. Name the plate (optional).
Open a template. See Creating a workspace template
(page 189).
3 Open plots, create and apply gates. See Creating a new plot
(page 157) and Creating and applying gates (page 160).
Before you begin Set the acquisition parameters. See Setting acquisition parameters
(page 72).
About collecting • Before data collection can begin, the software must display a
sample data green traffic light with the message Cytometer is connected
and ready.
• Each data well holds a maximum of 1 million events. You can
add more events (up to a total of 1 million) to a well already
containing data.
• When a well already contains data, the RUN button displays
ADD TO.
Note: Run limits may need to be adjusted to accommodate
additional data.
Procedure To manually run the sample with the BD CSampler Plus software:
1. Select a plate type from the Plate Type list.
94 BD Accuri C6 Plus System User’s Guide
3. Type the sample name into the text box above the 96-well grid.
Chapter 6: CSampler Plus data acquisition 95
4. Resuspend the cells in the sample tube or plate, and load the
tube rack or plate.
The well flashes blue during data collection. After the run limit
is reached, the well stops flashing and remains blue, indicating
that the well contains data.
8. To run the next sample, click the next well in the sample grid.
If necessary, adjust the acquisition parameters, then click
RUN.
Rinsing the SIP A rinse cycle can be used to rinse the SIP between samples. The SIP
between samples goes to the water tube in the square location of the CSampler Plus
tray and aspirates the contents of the SIP up then out of the SIP,
then rinses the SIP with clean fluid.
Adding more You can collect new samples and add the data to a CSampler Plus
sample data to a workspace file that already contains data, by adding data to a well
CSampler Plus file that already contains data or to an empty well.
If you select an empty well, any plots and gates you created
earlier are still displayed, but they do not contain data.
The flow cytometer stops sampling from the tube or well when
the run limit is reached.
Pausing data You can interrupt sample acquisition any time during a run.
acquisition
To pause a run:
1. Click PAUSE.
If you want to delete the data that was already collected before
you paused, click Delete Events, then click RUN to acquire
new data.
Ending a data When you finish collecting samples, clean the SIP to ensure cells or
acquisition session other particles are not left in the SIP.
About the Auto The Auto Collect tab allows automated sampling from a well plate
Collect tab or sample tube rack. The tab contains buttons and controls for
performing the following functions:
Viewing the Auto The Auto Collect tab is organized into two major sections:
Collect tab
• Instrument Control Panel. Panel on the left side of the window
that contains controls for collecting data.
Auto Collect tab The following table describes each of the controls and indicators in
controls the Auto Collect tab.
Control Description
Plate Type List for selecting the plate type.
Load Plate/Eject Moves the sample plate into position to be loaded onto or ejected from
Plate the flow cytometer.
Control Description
Sample grid Matrix laid out in the configuration of a multi-well plate to help
organize experiments and collect data from sample tubes or wells. Each
sample is acquired into its own well in the sample grid. The wells are
color-coded:
White. Does not contain data.
Colored fill. Acquisition settings are applied. Each unique acquisition
setting is indicated by a color change.
Small blue box. Contains data.
Red outline. Currently selected for viewing or collecting data.
Traffic light and Indicator that displays software readiness and system messages. Before
message data collection can begin, the software must display a green traffic light
with the message Cytometer is connected and ready.
The Traffic Light status is color-coded:
Green. Software is ready to collect data or is collecting data.
Yellow. Cytometer is preparing to perform an action, performing an
action other than data collection (such as cleaning or agitating), or a
non-critical error has occurred.
Red. Critical error has occurred or the CSampler Plus has collided.
Run Limits Contains a set of controls that allow you to define criteria for
automatically stopping data collection.
SIP Rinse Settings Allows you to select a SIP Rinse cycle to perform between each sample.
See Rinsing the SIP between samples (page 96).
Fluidics Allows you to set the flow rate. See Setting the fluidics rate (page 107).
Custom flow rates are not available in the Auto Collect tab.
Set Threshold Sets the event threshold to eliminate debris and noise from samples.
The default value is 80,000 on FSC-H. You can also set a secondary
threshold. See Threshold (page 79).
Apply Settings/ To apply or remove sample settings, such as run limits, fluidics,
Remove Settings threshold, and SIP rinse settings.
Agitate Plate Performs an agitate cycle to keep particles in suspension. See Agitating
samples (page 108).
102 BD Accuri C6 Plus System User’s Guide
Control Description
SIP Clean Performs a SIP Clean at the end of the run.
Run Horizontally/ Allows you to select how you want to acquire samples—horizontally or
Run Vertically vertically. See Setting the run direction (page 109).
Open Run Display Opens the run display for starting and stopping a run, and viewing
data acquisition counters and two plots of data.
Sample Annotation Contains fields for naming samples, renaming parameters, and adding
table a notation to each sample. Information can be entered manually for
each sample or copied and pasted from a spreadsheet program. See
Creating the Sample Annotation table (page 110).
Auto Collect Use the following workflow to run samples using Auto Collect.
workflow
Stage Description
1 Creating plot displays and gates in the Manual Collect tab.
2 Display the Auto Collect tab and select the plate type.
About sample sets A sample set is a group of samples that use the same data
acquisition criteria (such as number of events or threshold setting).
Acquisition criteria include:
If you are not creating sample sets and all samples will have the
same collection criteria, proceed to Setting parameters for Auto
Collect experiments (page 106).
• Run limits
• (Optional) SIP Rinse Settings
• Fluidics
• Threshold – default is FSC-H 80,000
See Setting parameters for Auto Collect experiments
(page 106) for more information.
All sample settings are saved each time you click Apply
Settings. Acquisition settings displayed in the Control Panel do
not change. To see the actual settings for a specific set, press
Ctrl and click a well of interest within the set.
Setting run limits The run limit defines when data acquisition will stop. The
following parameters can be used individually or in combination to
set a run limit.
See Run settings (page 73) for more information about setting
run limits.
Chapter 6: CSampler Plus data acquisition 107
Setting the fluidics The cytometer can accommodate an upper limit of 10,000 events
rate per second, but it is recommended to acquire samples at a rate of
2,500 events per second or less to ensure the best data resolution.
• Slow is 14 µL/min.
• Medium is 35 µL/min.
• Fast is 66 µL/min.
You cannot customize the flow rate or the sample stream core
size in Auto Collect experiments.
Setting the Use thresholds to eliminate light scatter and/or fluorescence signals
threshold caused by debris in cell samples and electronic noise inherent in the
system.
Rinsing the SIP Multiple rinse cycles of the SIP can be specified in the Auto Collect
between samples tab. Up to three rinse cycles can be specified at a time.
The specified SIP rinse cycles are performed after each sample
acquisition, and can be applied to individual wells.
Setting the run Data can be collected from plates either horizontally (A1, A2, A3,
direction etc) or vertically (A1, B1, C1, etc).
Creating the Use the Sample Annotation table to name samples, rename
Sample Annotation parameters, and add a notation to each sample. Information can be
table entered manually for each sample or copied and pasted from a
spreadsheet program.
Opening the Run The Run Display is used to begin, view, interrupt, and abort data
Display collections in Auto Collect.
Before you begin • Set up plots and regions in the Manual Collect tab. You can
display two plots in the Auto Collect tab. See Creating a new
plot (page 157) and Creating and applying gates (page 160).
About collecting The BD CSampler Plus software begins data collection in the first
sample data well that has settings applied (most commonly A1) and progresses
horizontally or vertically as defined in the settings.
The Run Display also updates the acquisition counters and the
sample data in the plots.
114 BD Accuri C6 Plus System User’s Guide
To restart a run:
• To restart during an Interrupt plate state, click AUTORUN.
The software continues acquiring data where it left off.
• To restart a run during an Abort well state, do one of the
following:
– To begin collecting the sample in the next well, click
AUTORUN.
– To continue collecting from the same well, close the run
display on the Auto Collect tab, select the Manual Collect
tab and run the sample manually.
Viewing sample All plots created in the Manual Collect tab are listed in the Select
plots Plot list beneath the plot window. They can be selected at any time
during data collection. Two plots can be viewed at a time.
To view plots:
1. Select plots to view from the Select Plot list under each plot
corral.
Saving the To save the workspace file at the end of the run:
workspace file 1. Select File > Save.
Performing a SIP When you finish collecting samples, clean the SIP to ensure cells or
clean other particles are not left in the SIP. If you select Run SIP Clean
after samples at the bottom of the Control Panel, a SIP Clean will
be performed automatically at the end of the run.
About the Analyze The Analyze tab allows you to simultaneously view data from
tab multiple samples using the same plots and gating.
Viewing the The Analyze tab is organized into two major sections:
Analyze tab
• Setup panel. Panel on the left side of the window that contains
controls for selecting samples and plots.
• Data display. Large area on the right side of the window that
shows sample data in plots and in a statistics table.
When you open the Analyze tab for the first time, the
workspace is empty. Plots can be copied from the Collect/
Manual Collect tab or created from scratch. Gating strategies
that were set up in the Collect (or Manual Collect) tab can be
applied in Analyze, or new gates can be drawn and new gating
strategies set up.
Chapter 7: Data analysis 121
Analyze tab The following table describes each of the controls and indicators in
controls the Analyze tab:
Control Description
Plate Type Disabled in the Analyze tab. For the BD CSampler
Plus software only.
Plate Name Text box for naming the plate. For the
BD CSampler Plus software only.
Control Description
Sample grid Matrix laid out in the configuration of a 96-well
plate to help organize sample data. Each sample
has its own well in the sample grid.
The wells are color-coded:
White. Does not contain data.
Blue. Contains data.
Black check mark. Currently selected for viewing
data.
Copy Plots from Copies specified plots from the Collect (or Manual
Collect Collect) tab.
See Copying plots from the Collect or Manual
Collect tab (page 124) for more information.
Plot List Lists the plots that are available in the Analyze
tab. Available plots include plots copied from the
Collect tab and plots created in the Analyze tab.
Example data file An example data file of a four-color analysis of human peripheral
for analysis blood (HPB 4 Color Tutorial.c6) can be found on the BD Accuri
C6 Plus software installation disc. The data file can be used to
explore the various software analysis tools.
The example file consists of data from four sample tubes used to
assess the CD3+CD4+ and CD3+CD8+ cell populations. Samples
were prepared by staining peripheral blood with directly
conjugated antibodies, followed by red cell lysis, according to
standard preparation methods. The following table describes the
experimental design:
Copying plots from In the Analyze workspace, you can copy plots from the Collect (or
the Collect or Manual Collect) tab or create new plots. Plots that are copied from
Manual Collect tab the Collect (or Manual Collect) tab are appended with a “C” (for
example, Plot 1C, Plot 2C).
If the plots you are copying contain regions and markers, the
regions and markers will copy over. The gating strategy does not
copy over.
To copy plots:
1. In the Analyze tab, click Copy Plots from Collect.
3. Click OK. The selected plots are added to the Plot List.
5. Click the sample well that contains the data you want to view.
For more information on viewing the data, see Viewing data in
plots from the Plot List (page 125).
Any gates copied from the Collect tab are renamed (for
example, P1 in Collect is P2 in Analyze). These gates can be
adjusted in the Analyze tab without changing the position of
the gates in the Collect tab.
• Histogram
• Dot
• Density
• Overlay Histogram
An FSC-A histogram or FSC-A vs SSC-A dot or density plot is
displayed.
3. Click the sample well that contains the data you want to view.
Statistics tab
Introduction This topic describes the Statistics tab and how to create custom
statistics.
About the Statistics The Statistics tab provides a way of tabulating data from multiple
tab samples in one master table. It also allows you to do the following:
Viewing the The Statistics tab is organized into two major sections:
Statistics tab
• Setup controls (Statistics Column Selector, Sample Selector,
and Display Plot Preview) for the Master Statistics table
• Master Statistics table
When the Statistics tab is first opened, the workspace is empty.
Statistics tab The following table describes each of the controls and indicators in
controls the Statistics tab:
Control Description
Plot Previews the selected plot.
Creating the The Master Statistics table allows you to immediately view selected
Master Statistics data across multiple samples. You can customize which data you
table want to view for each plot, and you can modify the table at any
time.
Stats are added in the order they are selected after they are
grouped by plot and region.
The plot preview is available for viewing only. You can modify
the zoom level and other plot settings in the Analyze or Collect
tab.
134 BD Accuri C6 Plus System User’s Guide
About VirtualGain In certain instances, a particular peak should have the same
position across different samples or be located at a specific channel
number, regardless of the staining. Instruments that have voltage
and amp gain controls allow you to adjust peak position from
sample to sample. BD CSampler Plus software uses VirtualGain
instead of these controls.
VirtualGain is only applied to the displayed data and does not alter
FCS data. The adjustment is recorded only in the data file.
6. If needed, use the Zoom tools in the Analyze tab to change the
zoom level in the VirtualGain dialog.
7. Click the small sample grid icon in the center of the Sample to
Align plot.
plots for individual samples without affecting the plots for other
samples.
Viewing the Batch The Batch Analysis tab is organized into two major sections:
Analysis tab
• Setup panel. Panel at the top that contains the sample grid and
available plots for selecting samples and plots.
• Data display. Large area at the bottom that shows sample data
in plots and in a Statistics table.
When the Batch Analysis tab is opened for the first time, the
workspace is empty. To set up the tab, plots can be created or
copied from the Collect (or Manual Collect) or Analyze tab.
144 BD Accuri C6 Plus System User’s Guide
Control Description
Sample grid Matrix laid out in the configuration of a 96-well
plate for selecting the samples to analyze.
The wells are color-coded:
White. Does not contain data.
Blue. Contains data.
Black check mark. Currently selected for batch
analysis.
Exporting data Plots and statistics can be exported in two ways, either as a
Microsoft PowerPoint® file or as a Microsoft Excel® spreadsheet.
To export data:
1. Do one of the following in the lower right-hand corner of the
Batch Analysis tab:
Compensation
Introduction This topic explains the concept of fluorescence spillover, how to
recognize when spillover occurs, and how to correct it using
compensation.
Recognizing The following figure shows data collected for a PE-Cy™7 single-
fluorescence stained control. Most of the fluorescence signal from PE-Cy7–
spillover positive cells was detected in FL3 (670 LP), as expected.
Chapter 7: Data analysis 147
However, there is also PE-Cy7 signal detected in FL1 (533 BP) and
FL2 (585 BP), so that data plots for those detectors appear to have
positively fluorescing cells. No signal from PE-Cy7 appears in
detector FL4.
Compensation preferences
The Preferences button in the compensation window allows you to
select between using your own custom user-defined compensation
settings or the compensation settings set by the software during
Instrument QC using BD CS&T RUO beads. Instrument QC sets
compensation for FITC, PE, PerCP, PerCP-Cy™5.5, and APC. If
you choose to use the settings from Instrument QC, you can select
to use FL3 compensation values for either PerCP or PerCP-Cy5.5.
If you use fluorochromes other than those set by instrument QC,
compensation will not be set for those fluorochromes. You will
need to manually adjust compensation for other fluorochromes.
Chapter 7: Data analysis 149
Correcting When color compensation has been properly applied to a data set,
fluorescence the median fluorescence channel value in non-primary detectors
spillover for any given single-stained control sample should be the same as
that of an unstained control sample.
About plots Plots allow you to view sample data in histogram, density, dot, and
overlay histogram plots. You can view multiple plots for each
sample you collect.
Plot tools Each plot contains a set of gating and marking tools and a set of
viewing tools.
• Expand tool. Zooms out one level with each click of the tool.
• Histogram Plot
• Dot Plot
• Density Plot
158 BD Accuri C6 Plus System User’s Guide
Changing plot The Plot Spec tool allows you to change the way data are displayed
specifications in a plot. You can change the axis parameter, specify channel
ranges, and toggle between linear and logarithmic scales. The Plot
Spec tool is available in the Collect and Analyze tabs.
2. In the Set Plot Specs dialog, do the following for each axis:
Renaming plots Plots can be renamed. By default, plots are named after the order
in which the plot was created and the associate plate map well, for
example Plot 1: A01.
To rename a plot:
1. Double-click the plot name.
3. Press Enter.
The well number will remain in the new plot name, for
example, LWB: A01. The new name will apply to the selected
plot for all samples in the run, for example, LWB: A02.
• Polygon
• Rectangle
• Ellipse
• Quadrants
• Hinged quadrants
2. Use the mouse to draw a region (labeled P1 for polygon, R1
for rectangle, E1 for ellipse, Q1 for quadrants, and H1 for
hinged quadrants).
Creating a vertical You can create a single vertical marker per histogram to obtain
marker in a statistics on the events to the left and right of the marker.
histogram plot
To create a vertical marker in a histogram plot:
1. Click the Vertical Marker tool.
2. Click the cursor at the point along the x-axis where you want
to place the marker.
Chapter 8: Plots and gates 163
2. Drag the cursor horizontally across the area you want to gate.
164 BD Accuri C6 Plus System User’s Guide
If the plot to which you are applying a gate already has a gate,
it will not appear in the list. Only gates in other plots appear.
3. Select one of the following gating icons to the left of the region
in the list.
4. Click Apply.
Creating a live gate Live gating is used to exclude events outside a designated region
from being displayed and stored as part of the workspace or FCS
file. Any region may be used as a live gate. This selection can be
remembered as part of the template if desired.
About nested gates You can create a series of nested gates in which each gate is a
subset of the previous one. This allows you to fine tune your gating
strategy for viewing and analyzing specific subsets of data.
168 BD Accuri C6 Plus System User’s Guide
2. Click GATE in a second plot and apply the gate. See Applying
a gate to a plot (page 164). This is the parent gate.
Chapter 8: Plots and gates 169
3. Close the dialog. The plot displays only the populations within
the parent gate.
Modifying regions
Introduction This topic describes how to modify regions by moving and resizing
them and coloring the events in a region.
Moving and You can move or resize a region. You can also move a region label
resizing regions by clicking and dragging the label.
To move a region:
1. Click the border of the region. The outline appears in bold.
172 BD Accuri C6 Plus System User’s Guide
To resize a region:
1. Click the border of the region. The outline appears in bold.
Applying color All events within one or more regions can be designated to appear
gating as a specified color within other histogram or dot plots. The event
coloring will remain and automatically update during data
acquisition.
2. Double-click the region name and click the small white square
to display a color palette.
3. Select a color.
174 BD Accuri C6 Plus System User’s Guide
Events that fall within the gate are now colored and displayed
in the same color in other dot and histogram plots.
To rename a region:
1. Double-click the region name.
Modifying plots
Introduction This topic describes how you can change the number of events in a
plot and change the plot parameters and axes names.
Changing the You can change the number of events displayed in all plots across
number of events all samples to make it easier to view the data. This option allows
in a plot you to visually remove a number of events from the plot without
deleting data.
The software displays the plot and shows a message in the plot
that some events are not being displayed.
Changing plot You can change the parameters that a plot displays for the x- and
parameters y-axes. By default, the software displays the area parameter
(signified with a suffix of -A), but you can also choose a height,
width, or time parameter.
Chapter 8: Plots and gates 177
To change a parameter:
1. Click the x- or y-axis label and select the option you want
from the menu.
Renaming plot axes You can rename the axis labels for a plot to identify the antibody
staining or fluorochrome used in the sample.
Zooming a plot
Introduction This topic explains the zoom function you can use to zoom in on
specific areas of a plot.
2. Repeat as needed.
180 BD Accuri C6 Plus System User’s Guide
Copying and To copy and paste plots from the Collect or Analyze tab:
pasting plots 1. Click anywhere on a plot and drag it to an open Microsoft®
Office compatible application.
You cannot use Ctrl+C and Ctrl+V to copy and paste plots
from the software into other applications.
For information on copying and pasting plots from the Collect tab
to the Analyze tab, see Copying plots from the Collect or Manual
Collect tab (page 124).
Selecting a drag You can select one of two file formats for plots during drag and
and drop format drop actions.
Printing plots and You can print selected plots and their associated statistics from the
statistics Collect or Analyze tab.
More information • Copying plots from the Collect or Manual Collect tab
(page 124)
• Copying analysis data into other applications (page 142)
9
Managing data
This chapter covers the following topics:
Managing files
Introduction This topic describes the files generated by BD Accuri C6 Plus
software.
About C6 Plus files The BD Accuri C6 Plus system generates the following files.
File Description
FCS FCS 3.1 files are not saved by default, but can be
exported at any time following acquisition. See
Exporting and importing data (page 190) for
information.
Results CSV CSV files are saved only if you select to save them. A
single CSV file is saved for each sample. Files are
saved to the location you choose. See Exporting and
importing data (page 190) for more information.
Removing QC You can remove old QC runs that appear in the history file. See
report files from Viewing previous QC results (page 61).
the previous QC list
1. Navigate to C:\Cytometer Support
Files\InstrumentPerformance.
The QC results for the dates you deleted will no longer appear
in the history file list. However, the results will still appear in
the Levey-Jennings plots.
Removing bead lots You can remove old CS&T RUO bead lots so that they no longer
from the CS&T appear in the bead lot menu on the QC workspace.
bead lot menu
1. Navigate to C:\Cytometer Support
Files\InstrumentPerformance\BeadLots.
Workspace files
Introduction This topic describes how to save the data to a workspace file.
About saving Always save data as a workspace file (.ci). A workspace file is a
workspace files comprehensive (and often large) file that contains instrument
settings, FCS files, and plot layouts.
• Sample data
• Plot layouts
• Parameter names
• Zoom levels
• Gating
• Color compensation
• Threshold settings
• Collect (or Manual Collect) tab settings
• Changes made in the Analysis or Statistics tab
When a file is saved, the software displays the file name in the
upper-left corner of the workspace.
Auto-saving data By default, the software automatically saves the event data any
time the cytometer reaches a run limit or if you click Pause during
a run. Auto-save does not save changes to acquisition settings,
plots, or gating strategies that occur after the initial save when
naming the file.
Chapter 9: Managing data 187
3. Click OK.
Manually saving You can manually save a workspace file at any time.
files
Note: For BD CSampler Plus software, you will be prompted to
save the workspace file after you select the plate type.
If this is the first time the file is being saved, a save dialog
appears. Enter a file name and click Save.
2. Navigate to the location where you want to save the file, enter
the file name, and click Save. The file is saved with the
extension .ci.
Opening a new A new workspace displays a single FSC-A versus SSC-A density
workspace plot that is pre-zoomed to channel values of 1,600,000 and
800,000, respectively. Run settings are set to Run Unlimited, the
primary threshold is set to channel 80,000 on the FSC-H signal.
Compensation is set to the values defined by the last Instrument
QC run performed, unless user defined compensation is selected.
3. Navigate to the location to save the file, enter the file name,
and click Save. The file is saved with the extension .cit.
190 BD Accuri C6 Plus System User’s Guide
Exporting data You can export sample data, including compensation settings,
from all or individual sample wells as FCS 3.1 files. No gates or
analysis settings are exported with the data. You can also export
plot data and sample settings.
To export data:
1. Do one of the following:
• Select File > Export FCS File to export and save the data
from the currently selected well as an FCS 3.1 file. Enter a
file name and click Save.
• Select File > Export ALL Samples as FCS to export and
save the data for all wells as individual FCS 3.1 files. A
default folder is created on the desktop. Click Ok.
Importing data Only FCS files created using BD Accuri C6 Plus software or
BD Accuri C6 software can be imported into a workspace file or
template.
Maintenance schedule
Introduction This topic provides a list of daily, scheduled, and unscheduled
maintenance procedures.
Daily maintenance Daily maintenance is part of the startup and shutdown procedures.
See Starting up the system (page 41) and Shutting down the system
(page 43). Additionally, the following procedures can be
performed as needed.
2. Wipe down the outside panels and the sample stage. If you
have a CSampler Plus, wipe down the CSampler Plus tray and
mat.
3. Moisten a clean paper towel with DI water and wipe the areas
you wiped with the cleaning solution.
About the The backflush cycle forces fluid out of the flow cell and out the SIP
backflush cycle to remove bubbles in the flow cell and/or clogs in the SIP. Perform
a backflush if you notice the event rate slowing or stopping and
you suspect a clog.
You can also click Backflush from the control panel of the
Collect or Manual Collect tab.
You can also click Backflush from the control panel of the
Collect or Manual Collect tab.
198 BD Accuri C6 Plus System User’s Guide
About SIP cleans The SIP clean runs BD FACSClean through the SIP for
approximately 5 minutes followed by water for approximately
5 minutes. Regular SIP cleans are vital for keeping your fluidics
system clear and free of clogs.
When to run a SIP We recommend always running a SIP clean at the end of each run.
clean You can also run a SIP clean as often as you like to prevent SIP
clogs.
About cleaning the The fluidics cleaning cycle is automatically run when you shut
fluidics down the instrument. You can also run it at any time, if necessary.
During the fluidics cleaning cycle, BD FACSClean is run through
the fluid lines, followed by sheath. A second cycle runs detergent
solution through the fluid lines, followed by sheath.
About an extended An extended flow cell clean cycle provides a way to thoroughly
flow cell clean clean the flow cell. Run an extended flow cell clean cycle if you
notice excessive debris or high CVs during instrument QC. During
Chapter 10: Maintenance 201
extended flow cell cleaning, the flow cell fills completely with
cleaning solution from a sample tube on the SIP. The cycle
automatically shuts down the cytometer, allowing the flow cell to
soak with the cleaning solution.
3. Load the tube of Extended Flow Cell Clean Solution and click
Proceed.
About cleaning the Clean the bottles every month to ensure that residue and
bottles contaminants do not build up inside the bottles. Do not allow the
bottle filters to dry out while you clean the bottles.
5. Refill each bottle with the appropriate fluid. See Filling the
fluid bottles (page 36).
About fluid bottle The sheath, BD FACSClean, and detergent bottles each contain a
filters disk filter. We recommend replacing these filters every 2 months.
2
204 BD Accuri C6 Plus System User’s Guide
When to replace We recommend replacing the in-line sheath filter every 2 months.
the in-line sheath
filter If you notice a yellow discoloration, fluid leaking, or if the filter is
less than 50% full, change the filter immediately.
Before you begin Write today’s date on the new in-line sheath filter to help you keep
track of unscheduled filter replacements.
2. Lift the cytometer lid and remove the plastic storage bin.
Chapter 10: Maintenance 205
3. Each time you change the in-line sheath filter, visually inspect
all tubing and connectors for fluid leaks.
This filter has a male and female end to ensure that it can only
be installed in the correct orientation.
206 BD Accuri C6 Plus System User’s Guide
7. Replace the plastic storage bin and close the cytometer lid.
Note: Upon startup you may see an error message indicating that
a fluidics system error was detected. This error is normal after
replacing the in-line sheath filter. See Hardware fault errors
(page 225) for information.
6. Remove the pump tubing from the pump head and discard it
as you would a biological sample according to standard
laboratory protocols and regulations.
Note: Upon startup you may see an error message indicating that
a fluidics system error was detected. This error is normal after
replacing the peristaltic pump tubing. See Hardware fault errors
(page 225) for information.
About this Perform this procedure if the traffic light message indicates that the
procedure sheath is empty or the waste is full, when in fact all fluid levels are
as they should be. Fluid in the sensing lines can lead to erroneous
tank level messages.
Before you begin Start with the instrument turned off. Ensure the sheath bottle is full
and the waste bottle is empty, except for bleach. Place the sheath
bottle on the benchtop where you can monitor the fluid level.
Procedure 1. Lift the cytometer lid and remove the plastic storage bin.
2. Locate the round, finger-size access hole near the in-line sheath
filter.
210 BD Accuri C6 Plus System User’s Guide
3. Place your finger over the pin hole inside the access hole.
4. With your finger sealing the pin hole, turn on the cytometer.
5. Keep your finger over the hole for 30 seconds. You should see
continuous bubbling in the sheath bottle.
7. Replace the storage bin and close the lid to allow the
cytometer to complete the normal fluidics startup.
Chapter 10: Maintenance 211
About this Perform the syringe SIP unclog if the SIP is clogged and performing
procedure backflushes and SIP cleans have not fixed the problem.
Before you begin Before using the syringe to unclog the SIP, try unclogging the SIP
by performing several backflushes followed by several SIP cleans.
3. Disconnect the three flow cell fluid lines (blue sheath, red
waste, clear purge) from the chassis by turning each connector
approximately 1/8 turn counter-clockwise.
9. Slowly pull the plunger to pull the cleaning solution back into
the syringe.
10. Repeat the push/pull cycle a few more times to ensure the clog
is cleared. Finish by pushing the BD FACSClean into the tube.
4. Slowly pull the plunger to pull the water back into the syringe.
214 BD Accuri C6 Plus System User’s Guide
5. Repeat the push/pull cycle a few more times to ensure the SIP
is thoroughly rinsed. Finish by pushing the water into the tube.
7. Disconnect the blue (sheath) and red (waste) lines from each
other.
About precise Each BD Accuri C6 Plus flow cytometer is calibrated at the factory
volume for accurate volume measurement. This enables the system to
measurements provide absolute cell counts per sample volume. The system
displays the volume in µL and automatically calculates counts per
µL for any gated population. Follow this procedure when accurate
volume measurement is required during sample acquisition. For
more detailed information, refer to the Technical Bulletin, A Guide
to Absolute Cell Counting on the BD Accuri C6 Flow Cytometer,
which can be found at bdbiosciences.com.
Before you begin Ensure that the peristaltic pump tubing and in-line sheath filter
have been replaced within the last 60 days.
About alignment The CSampler Plus performs an alignment to verify that the tube
rack is aligned to the SIP every time the flow cytometer is powered
up or if the CSampler Plus arm collides into an object. A manual
alignment can be performed using the software at any time.
Samples left in the tube rack can be recovered by turning off the
cytometer and gently pushing down on the white cylindrical motor
housing.
Chapter 10: Maintenance 219
3. Fill the fluid bottles with fresh fluid and ensure the waste
bottle is empty except for 200 mL of bleach. See Filling the
fluid bottles (page 36).
6. Turn on and off the cytometer two more times, allowing the
fluid cycles to complete before cycling the power.
220 BD Accuri C6 Plus System User’s Guide
11
Troubleshooting
This chapter covers the following topics:
Troubleshooting overview
Introduction This topic describes troubleshooting for the BD Accuri C6 Plus
system.
Additional help If, after reading through the possible problems and recommended
solutions, you still have questions or are experiencing problems,
contact BD Technical Support. See Technical support (page 11) for
information.
Hardware troubleshooting
Introduction This topic describes how to troubleshoot hardware problems with
the cytometer or CSampler Plus.
Cytometer and/or
computer does not Possible causes Recommended solutions
power on
Power supply not Make sure the power supplies and cords are
plugged in plugged into an appropriate outlet.
Power and event Power and event indicator lights flash simultaneously
indicator lights
flash on startup Possible causes Recommended solutions
Waste tank full Turn off the power to the cytometer. Empty the
waste bottle, then restart the cytometer.
Message “Extra
startup time Possible causes Recommended solutions
needed due to
Extended flow cell Allow the extended startup to run. It will take
cleaning or clean cycle was run. approximately 25 minutes.
improper
shutdown” Fluidics error
occurred during
shutdown.
Instrument forcibly
shut down, for
example power
outage.
CSampler Plus
collision Possible causes Recommended solutions
Collision message Time-out occurred because the CSampler Plus
appears but no did not reach its designation in the time
collision occurred permitted. Select Instrument > Align CSampler
Plus. See Aligning the CSampler Plus after a
collision (page 217)
Traffic light
displays message Possible causes Recommended solutions
that waste is full or
Fluid bottle sensing Purge the fluid bottle sensing lines. See Purging
sheath is empty lines have fluid in the fluid sensor lines (page 209).
when they are not them
Message “Your
print job failed” Possible causes Recommended solutions
Printer not Ensure a printer is connected, either locally or
connected or not through a network, and turned on.
turned on
Hardware fault If any of the following errors occur, click Close this window, then
errors follow the steps listed.
Software troubleshooting
Introduction This topic describes problems related to the software.
Software hangs at
startup and will not Possible causes Recommended solutions
launch
Software launching Software requires more time to launch. Give
slowly the application time to launch.
USB flash drive Do not use a USB port on the same hub as the
plugged into same cytometer. For example, use USB ports on the
hub as cytometer, front of the computer for external flash drives.
disrupting
Disconnect flash drive.
communication
Disconnect the USB cable between the
cytometer and computer, then reconnect it.
Software not
responding Possible causes Recommended solutions
USB flash drive Do not use a USB port on the same hub as the
plugged into same cytometer. For example, use USB ports on the
hub as cytometer, front of the computer for external flash drives.
disrupting
Disconnect the flash drive.
communication
Disconnect either end of the USB cable
between the cytometer and computer, then
reconnect it.
Software appears
to shut down but is Possible causes Recommended solutions
still running in the
Javaw.exe process Use the task manager to end the javaw.exe
background interfering process.
1. Press Ctrl + Alt + Delete.
2. Click Start Task Manager.
3. Click the Processes tab.
4. Select javaw.exe and click End Process.
If the process still won’t quit, unplug the USB
cable from the cytometer, then plug it back in.
Chapter 11: Troubleshooting 229
Acquisition troubleshooting
Introduction This topic describes how to troubleshoot problems you encounter
during acquisition, including the data displayed in the plots.
Auto-adjust volume
settings procedure Possible causes Recommended solutions
failed
System fluidics not Perform a Clean Fluidics cycle, then repeat the
clean procedure.
QC troubleshooting
Introduction This topic describes how to troubleshoot problems with
instrument QC.
No beads were
detected Possible causes Recommended solutions
No beads in sample Make sure you are running the correct sample
tube.
QC result failed
Possible causes Recommended solutions
Regions and/or Adjust regions and markers, as necessary. See
markers not set Adjusting the bead regions and markers
correctly (page 53).
QC messages
Message Recommended solutions
Event count for Mid Beads settled in bottom of tube. Resuspend
+ Bright too low the beads and rerun the sample.
Ensure that the region is encompassing the
mid + bright bead population in the FSC vs
SSC plot. Adjust the region, if necessary.
Event count for FL1/ Make sure the bead peak in the corresponding
FL2/FL3/FL4 Bright plot is located within the markers. Adjust the
too low marker, if necessary. If the message still
appears, prepare a new bead suspension.
Selectable lasers
Introduction This topic describes the Selectable Lasers Module option and how
to install, validate, and configure the lasers and optical filters.
About selectable The detectors and lasers of the BD Accuri C6 Plus flow cytometer
lasers operate in a predefined standard configuration: FL1, FL2, and FL3
detect blue laser-excited fluorescence emissions and FL4 detects red
laser-excited emissions. This configuration is referred to as 3 blue 1
red.
Components The standard optical filters supplied with the instrument are 533/
supplied 30 nm (FL1), 585/40 nm (FL2), 670 LP nm (FL3), and 675/25 nm
(FL4).
Chapter 12: Selectable Lasers Module 237
2. Copy the Selectable Lasers Activation Key from the USB drive
to the desktop.
6. Click Install.
About the selected The selectable laser buttons show the most recently selected
laser configuration configuration only. They do not change if you select another data
well with a different configuration. Therefore, we recommend
annotating your data wells in the sample naming field to indicate
the laser configuration used during data collection of each well,
especially when not using the default 3 blue 1 red option. (For
Chapter 12: Selectable Lasers Module 239
Laser configurations are not saved in the C6 Plus data file or with a
C6 Plus template. Any previously saved C6 Plus file will default
back to the 3 blue 1 red option when it is opened.
About filter Due to the unique optical layout of the cytometer, it is critical that
placement any optional filters used with the Selectable Lasers Module are
placed in the proper positions for optimal performance.
Configurations Use the tables below as a guide to optical filter placement for
various fluorochrome combinations.
4 blue: Configuration 1
4 blue: Configuration 2
FL2 585/40 PE
FL4 610/20 PI
4 blue: Configuration 3
FL2 585/40 PE
FL4 610/20 PI
FL2 585/40 PE
FL2 585/40 PE
Example 2: 4 blue
FITC, PE, PE-Cy5, PE-Texas Red: Human Peripheral Blood
Subsets
FL2 585/40 PE
About user tracking User tracking is an optional upgrade that allows laboratory
administrators to track the activities of cytometer operators by
assigning a user name and password to each individual. The user
name and password are required to log into the software. The
passwords are unrelated to any Windows passwords used on the
host computer or network. You must be the administrator to use
the user tracking features.
4. Click Install.
Signing in to the Once the tracking software is installed, users must now sign in
software each time they want to use the C6 Plus software. To use the
tracking feature, sign in to the software as the administrator.
2. When the Sign In dialog is displayed, enter the user name and
password provided to you during installation.
3. Click OK.
Signing out of the Be sure you sign out of the software when you are finished using it.
software If you do not sign out, software will continue to log your time.
Adding user The Users menu option is only visible to the administrator, and
accounts only after the User Tracking software has been installed.
4. Type a user name and password for the user in the blank fields.
6. Click Save.
3. Click Save.
Chapter 13: User Tracking Module 255
Changing a The administrator can change their password or any user password
password at any time.
2. Delete the text in the Password field and type a new password.
3. Click Save.
Restoring an The password file is encrypted and contains all user name and
administrator password information created in the User Logging feature. If you
password forget your administrator password, you can delete the password
file and recreate your administrator account.
About the Each time a user signs into or out of the software, an entry is made
userUsage log in the userUsage Log. The userUsage log is a CSV file that you can
open using Microsoft Excel or similar spreadsheet program.
Remote Interfacing
Introduction This topic describes the Remote Interfacing Module option and
how to install and use it.
About the Remote The Remote Interfacing Module is an optional module that enables
Interfacing Module the BD Accuri C6 Plus software and BD CSampler Plus software to
receive commands from other software applications via a TCP/IP
connection. Remote Interfacing is designed for advanced users
only.
4. Click Install.
<Connected/>.
Implemented commands
Introduction This section describes the available remote interfacing commands
to C6 Plus or CSampler Plus software and the return values from
the software.
Commands <AutomaticWellAdvance/>
Return <OK/>
<ChangeWell wellCode=“A1”/>
Return <OK/>
<CollectSample/>
Return <OK/>
<NewWorkspace/>
Return <OK/>
<ExportAllAsFCS path=“C:\NewDirectory”/>
Return <OK/>
<OpenWorkspace path=“C:\NewDirectory\Filename”/>
Return <OK/>
<Pause/>
Return <OK/>
<ReportStatus/>
Return <Status><Time>14:28:22.640</Time><Hardware
Status>Running</HardwareStatus><TotalEvent
Count>445785</TotalEventCount><UserMessage>
Events are being recorded.</UserMessage></Status>
<SaveFile path=“C:\NewDirectory\Filename”/>
Description Saves the file to the specified location and name. This
command requires both a file location and file name.
The extension must be included in the file name. To
save the file as a template, the extension is .cit, to
save as a C6 Plus software or CSampler Plus software
workspace file, the extension is .ci.
Return <OK/>
Command Response
<Connected/>
<OpenWorkspace <OK/>
path=“C:\ExampleFolder\Examplefile.cit”/>
<AutomaticWellAdvance/> <OK/>
<CollectSample/> <OK/>
<SaveFile <OK/>
path=“C:\NewDirectory\NewFilename.ci”/>
266 BD Accuri C6 Plus System User’s Guide
15
FCS keywords
This chapter covers the following topics:
FCS keywords
Introduction The following tables list the standard and custom keywords used
in FCS files generated by BD Accuri C6 Plus software and
BD CSampler Plus software.
Standard keywords The following table describes the FCS keywords found in FCS files
that are exported by BD Accuri C6 Plus software and
BD CSampler Plus software.
Keyword Description
$FIL Filename including FCS extension.
Keyword Description
$PnN The name of parameter n. Parameters are the
default values from the software.
Keyword Description
$BEGINSTEXT Default FCS 3.1 tag to mark the beginning of
the text section.
Custom keywords The following table describes custom keywords found in FCS files
that are exported by C6 Plus software and CSampler Plus
software. Custom keywords are not required as part of the FCS 3.1
standard.
Keyword Description
#BDACCURIDECADESn The number of decades for
parameter n. It is always
7.224719870049579.
Keyword Description
#ACQUISITIONTIMEMILLI The cumulative run data’s
acquisition time in milliseconds.
System specifications
Introduction This topic describes the system specification.
Cytometer
specifications Item Specification
Dimensions Height: 27.9 cm (11 in.)
Width: 37.5 cm (14.8 in.)
Depth: 41.9 cm (16.5 in.)
With fluid bottles:
Height: 27.9 cm (11 in.)
Width: 54.6 cm (21.5 in.)
Depth: 41.9 cm (16.5 in.)
Item Specification
Emission detection 4 colors, standard optical filters
FL1 533/30 nm
FL2 585/40 nm
FL3 >670 nm
FL4 675/25 nm
Item Specification
Signal processing 24-bit datapath
Computer
specifications Item Specification
General PC desktop workstation with removable
SD card
RAM 8 GB
Optional CSampler
Plus Item Specification
Dimensions Length: 50.8 cm (20 in.)
Width: 35.6 cm (14 in.)
Height: 20.3 cm (8 in.)
Connection Serial
Item Specification
Standards Code 39
Codabar
Code 128
Interleaved 2 or 5
Universal Product Code (UPC)
Maxicode
DataMatrix
PDF417
Index
A barcode reader 14, 278
acquisition batch analysis 142–144
auto collect 112–115 BD Accuri C6 Plus system
collecting precise volumes 214–217 See also cytometer
CSampler Plus 93–96 overview 14, 16–22
manual 81–83 BD FACSClean
parameters, setting 72 bottle 21
troubleshooting 229 filling 38
workflow, auto collect 102 beads
workflow, CSampler Plus 92 See CS&T RUO beads
workflow, manual 72 bottle
adjusting BD FACSClean 21
height of sample stage 23 detergent solution 21
peak position with VirtualGain 134 filling 38
QC markers 54–56 filters, replacing 203–204
sample stream core 76 sheath 21
volume settings 215 waste 21
agitating samples 96, 108
analysis C
batch 142–144 C6 Plus system
sample data file 123 See also cytometer
setting up plots 124–125 overview 14, 16–22
statistics 130 Clean Fluidics cycle 200
using VirtualGain 135–140 cleaning
viewing data 125 backflushing SIP 197
auto-adjust volume settings 215 Clean Fluidics cycle 200
auto-saving data 186 extended flow cell clean 200–201
outside of instrument 196
B sample stage 43
backflushing SIP 197 SIP clean 198–199
tubes, in CSampler Plus rack 26
280 BD Accuri C6 Plus System User’s Guide
importing 191 H
managing 184–185 histogram overlay 127–129
sample analysis file 123
saving 187
workspace 186–188 I
filter importing data 191
fluid bottle, replacing 203–204 indicator
in-line sheath 22, 204–206 event 17
optical 20 power 17
fluid bottles in-line sheath filter
emptying waste 39 overview 22
filling 38 replacing 204–206
fluid sensing line 204 installing
overview 21 CS&T RUO bead lot 51
replacing filters 203–204 sample tube 22
fluidics selectable lasers module 237
cleaning 200 user tracking module 250
components, overview 20–22
harness 18 K
tubing, replacing 206–209 keywords
fluidics rate custom 270
customizing 76 standard 268
setting 75
fluids required 37
fluorescence spillover L
See also compensation lasers 20
about 146–147 Levey-Jennings report
correcting 149–153 printing 65
viewing 63
live gates 166
G
loading
gates 24-tube rack 25
See also regions sample tube 22
applying to plot 164
color gating 172–174
creating 161 M
deleting 164 maintenance
live 166 See also cleaning
nested 168–170 aligning CSampler Plus 217–218
types 160–164 Clean Fluidics cycle 200
daily 194
282 BD Accuri C6 Plus System User’s Guide
T emptying 39
template, workspace 189 pump 22
threshold workflow
about 79 auto collect 102
setting 80 CSampler Plus acquisition 92
troubleshooting manual acquisition 72
acquisition 229 QC 48
hardware 222 startup 36
QC 231 workspace
software 227 C6 Plus 29–30
tube rack creating templates 189
loading 25 CSampler Plus 30–31
map 26 files 186–188
specifications 24 opening new 188
tubes workstation
loading 22 overview 15
specifications 15 specifications 276
tubing, replacing 206–209
Z
U zooming plots 56, 179
USB port 18
user tracking module
installing 250
managing user accounts 253
monitoring user activity 256
signing in 252
V
VirtualGain
applying 135–140
changing views 141
purpose 134
removing 142
viewing 140
W
waste
bottle 21