Calibration of Eddycon CL - ENG
Calibration of Eddycon CL - ENG
of the Eddycon СL
eddy current flaw detectors
for the compliance with DSTU EN ISO 15548-1:2017
(EN ISO 15548-1:2013, IDT; ISO 15548-1:2013, IDT) standard requirements
1 GENERAL PROVISIONS
These methods regulate the procedure of verification of the Eddycon CL
eddy current flaw detectors (hereinafter – the flaw detectors) for the compliance
with DSTU EN ISO 15548-1:2017 (EN ISO 15548-1:2013, IDT; ISO 15548-
1:2013, IDT) standard requirements – Non-destructive testing. Equipment for
eddy current examination. Part I. Instrument characteristics and verification.
2 SAFETY REQUIREMENTS
Note:
Technical characteristics of the used equipment and its stability shall
correspond to the test goals.
Before connecting the oscilloscope and/or generator of the sine signals to the
flaw detector“s receiver, as required in separate test types, it is necessary to
make sure that the devices will not be damaged by a high-voltage pulse.
226
4 CARRYING OUT THE TESTING
Note:
Before connecting the oscilloscope and/or generator of the sine signals to the
flaw detector“s receiver, as required in separate test types, it is necessary to
make sure that the devices will not be damaged by a high-voltage pulse.
326
ATTENTION!
(1)
426
Note:
In case if the signal have a distorted form, it is recommended to increase the
signal amplitude by adjusting the generator of the flaw detector thus
bringing the signal to the regular (smooth) sine form.
Acceptance criteria
For the generator of sine form signals, it is recommended to use the contents of
the harmonic components as a measure of deviation from a pure sine.
ATTENTION!
Methods of testing
5.2.2.1 Assemble the circuit of the oscilloscope connection to the generator
output of the flaw detector via an active (noninductive) load (50 2.5) Ohm in
accordance with Figure 2.
synchronization CН1
generator СН2
R1=50 Оhm
5.2.2.2 Set the sine signal frequency value of 170 kHz at the flaw detector
and an initial voltage of the flaw detector’s generator that equals to 0.5 V (as
per table 2). In the «Settings» menu set the Maximum measurement frequency
value of 1 kHz, аll gain equals to 0 dB.
526
5.2.2.3 Set the scan parameters of the oscilloscope so that one sine signal
cluster of the flaw detector’s generator is displayed on the screen by the second
channel (СН 2) as shown in Figure 3 а) and b).
а)
b)
626
5.2.2.4 When performing measurements by the probe at the oscilloscope
for the second channel (СН 2) it is necessary to enter the «Probe» menu and by
successive pressing of the «Attenuation» button set the «10Х» value.
СН
2
726
activating the cursors of the frequency measurement (in the «Source» item
select the «МАТН» value) ensure that the peak of the main harmonic equals to
the frequency ΔHz=170 kHz as shown in Figure 5.
Cursor 2
Peak of the
main
harmonic
Cursor 1
826
5.2.2.9 After activation of the cursors of the magnitude measurement set
any of the cursors at the top of the peak and record the measured value as
shown in Figure 6.
5.2.2.10 Calculate the root mean square value of the voltage of the main
harmonic using the formula (2)
(2)
(3)
Acceptance criterion
Nonlinear distortion coefficient shall not exceed %.
Methods of testing
5.2.3.1 Assemble the circuit of the oscilloscope connection to the flaw
detector’s generator output via an active (noninductive) load (502.5) Ohm in
accordance with Figure 1.
ATTENTION!
It is recommended to carry out the measurements with double-frequency mode
switched off (in the flaw detector settings).
5.2.3.2 Set the following parameters in the working menu of the flaw
detector:
– Filter – off
– frequency – 100 kHz
– all gain – 0 dB.
926
In the «Settings» menu of the flaw detector set the Maximum measurement
frequency value of kHz.
5.2.3.3 Perform the measurements of the amplitude for all the values of the
excitation signal voltage as per the voltage range given in Table 3. Record the
measured voltage values V1 to the Table 3.
5.2.3.5 Calculate the input impedance value ZS for each value of the rated
voltage from the Table 3 using the formula (4) and fill in the Table 3.
(4)
Acceptance criterion
Maximum output voltage – is the deviation in voltage at the output of the flaw
detector’s generator without load and when the generator is set to the maximum
output signal.
Method of testing
5.2.4.1 Connect the oscilloscope to the output of the flaw detector’s
generator without load and perform the amplitude measurements for all values
1026
of the excitation signal voltage from the range of values indicated in Table 4. Fill
in Table 4 with the measured values of the voltage.
ATTENTION!
5.2.4.2 Calculate the value of the deviation (error) of each rated output
voltage using the received data by the formula (5) and fill in Table 4 with the
data.
(5)
Acceptance criterion
ATTENTION!
1126
This characteristic is the peak one at the minimum gain and corresponds to
the safe input voltage, i.e. voltage at which the flaw detector does not get
damaged.
Methods of testing
5.3.1.1 Assemble the circuit of measurement as per figure 7.
5.3.1.2 Set the following parameters in the working menu of the flaw
detector:
– filter – off
– ECP type – “Refl”
– frequency 100 kHz
– all gain – 0 dB
– voltage – 1 V.
Oscilloscope
СН1
R1=50
Ohm
5.3.1.4 Set the «HScale» and «Vscale» parameters values – ×1/2. Then set
the «Expert» testing mode in the “Settings” menu and go back to the working
menu.
1226
5.3.1.5 Set the following parameters at the sine signal generator:
– frequency 100 kHz;
– signal amplitude – 100 mV.
5.3.1.6 Difference between the generator’s signal frequency and the selected
band frequency of the flaw detector shall not exceed 10% from the bandwidth
of the flaw detector.
5.3.1.7 Balance the flaw detector when the receiver cable is disconnected.
Connect the cable to the flaw detector’s receiver and activate the output voltage
of the signal generator. The input signal is measured by the oscilloscope.
5.3.1.8 Increase the input voltage from 100 mV to 4 V and observe the input
signal change at the flaw detector.
First value out of four that are changing (which corresponds to the lowest input
value) and which stops increasing represents the limitation level of the output
signal VS (measured with the oscilloscope). The received input value
decreasing until the parameter under checking achieves the input value that
corresponds to 90 % of the VS level.
Acceptance criterion
Maximum permissible input voltage shall lay within the range of 2.0 V to
3.0 V.
1326
ATTENTION!
Methods of testing
Flaw detector
Impedance
meter receiver
ATTENTION!
5.3.2.2 Set the following parameters in the working menu of the flaw
detector:
– filter – off
– frequency – 100 kHz
– all gain – 0 dB.
1426
Acceptance criterion
The resistance value R shall be not higher than 1 kOhm, capacity value C —
not higher than 150 pF.
ATTENTION!
The resistor connected to the input shall be shielded. Set the gain of the flaw
detector to the minimum value. The flaw detector shall be balanced and have zero
input voltage.
All the filters shall be set to the lowest impact.
Difference between the generator“s signal frequency and the selected
bandpass frequency of the flaw detector shall not exceed 10% of the flaw
detector“s bandwidth stated by the manufacturer.
5.4.1 Balancing
5.4.1.1 Value of the residual output signal in case of balancing
The value of absolute output signal that is received after the balancing is
expressed in percentage of the specific range, for instance of the maximum
output voltage.
Methods of testing
5.4.1.1.1 Assemble the circuit in accordance with Figure 9.
synchronisation Trig In
Generator
Flaw detector
receiver Output
R1=50 Ohm
1526
Figure 9 – Circuit for the residual output signal measurement in case of
balancing
5.4.1.1.2 Set the following parameters in the working menu of the flaw
detector:
– frequency 100 kHz
– ECP type – “Refl”
– all gain – 0 dB
– filter – off.
5.4.1.1.3 Set the following parameters in the “Settings” menu of the flaw
detector:
– the «Test» mode
– maximum measurement frequency 10 kHz.
5.4.1.1.7 By pressing the «Output» of the generator supply the voltage to the
flaw detector’s receiver.
5.4.1.1.9 Balance the flaw detector and record the value of the residual
signal amplitude to the flaw detector after the balancing.
1626
5.4.1.1.10 Repeat the steps 5.4.1.1.6 – 5.5.1.1.9 and record the received
values into Table 7.
Number of Values of the flaw Measured values of the Maximum value of the
performed input detector’s voltage residual output signal residual output signal in
signal balancing before balancing, V after the balancing, V case of balancing, V
1
2
3
4
5
Acceptance criterion
The value of the residual output signal shall be not more than 0.03 V.
Methods of testing
5.4.1.2.1 Assemble the circuit in accordance with Figure 10.
1726
UZ
T
generator
receiver
R1=50
Ohm
Figure 10 – Circuit for the maximum input voltage measurement that can
be compensated
5.4.1.2.2 Set the «Test» mode in the flaw detector’s «Settings» menu and go
back to the working menu.
5.4.1.2.3 Set the following parameters in the working menu of the flaw
detector:
– filter – off
– frequency – 100 kHz
– ECP type – “Refl”
– generator – 6 V
– all gain – 0 dB
– scale ×1/16
– «HScale» and «VScale» parameters values – 1.
Note:
The voltage is not supplied to the UZT-1.
It is recommended to switch off the dual-frequency mode on the flaw
detector.
1826
5.4.1.2.7 Decrease of the signal attenuation at the UZT-1 is stopped when
the residual voltage achieves the value that is twice as more as the measured one
in para 5.4.1.1.
21
18
15
12
9
8
7
Acceptance criterion
The value of the maximum input voltage that can be compensated shall be
not less than 70 % of the maximum input voltage that meets the linearity
requirements.
ATTENTION!
1926
Flaw detector
Generator
receiver
R1=50
Ohm
Set the following parameters in the working menu of the flaw detector:
– filter – off
– frequency – 100 kHz
– all gain – 0 dB
– scale ×1/16.
When the parameters are set in the «Settings» menu set the «Expert»
testing mode and go back to the working menu.
5.4.2.5 Connect the cable to the flaw detector’s receiver and activate the
output voltage of the signal generator.
5.4.2.6 Record the measured value of the input signal amplitude to the flaw
detector.
2026
100 200 300 400 500 600
Measured value of the voltage
of harmonic, V
Extinction of harmonic, dB
(6)
Acceptance criterion
The value of the extinction of harmonics shall be not less than 15 dB.
Frequency range at which the demodulated signal have the gain not less
than «m» dB and with attenuation not less than «n» dB.
The «m» and «n» values shall be stated in the operating documentation. The
frequency characteristic that are set at the level of ±3 dB.
Methods of testing
5.4.3.1 Assemble the circuit as per Figure 10.
5.4.3.2 Set the «Expert» testing mode in the «Settings» menu of the flaw
detector and set the following parameters:
– scale – 1/16
– filter – off
– frequency – 200 Hz
– all gain – 0 dB
- activate the «Persistence» mode with 0.1s value (for more convenient
perception).
5.4.3.4 Balance the flaw detector with the disconnected cable after all the
settings entered, then connect the cable and activate the output voltage of the
2126
signal generator. As a result, “the ring” will appear on the flaw detector
screen.
5.4.3.5 Activate the “Circle” type ALARM frame in the flaw detector.
5.4.3.6 Set the ALARM frame radius so that the ALARM triggers at the signal
exceedance on the flaw detector screen with a step of +1000, +100, +10, +1.
2226
Measurem Lower limit
ent units of the
Signal Upper limit of the
of Flaw detector Difference in gain frequency-
generator frequency-amplitude
frequency frequency amplitude, dB amplitude
frequency characteristic, dB
characteristi
c, dB
1.0 1.0
5.0 5.0
MHz
10.0 10.0
13,0 13,0
16,0 16,0
Acceptance criterion
Frequency characteristic at the stage of the signal processing in the range of 100
Hz to 16 MHz at the level of not more than ±3 dB.
5.4.4 Bandwidth
The bandwidth is determined on the basis of the values received in para 5.4.3.1
when determining the frequency characteristic at the stage of signal processing
with the extinction of 3 dB.
Methods of testing
5.4.4.1 Assemble the circuit as per Figure 12.
Oscilloscope
Flaw detector
Generator
receiver
R1=50
Ohm
2326
5.4.4.2 Set the «Test» mode in the «Settings» menu of the flaw detector and
go back to the working menu.
5.4.4.3 Set the following parameters in the working menu of the flaw
detector:
– scale – 1/8
– «Hscale» parameter value – 4
– «Vscale» parameter value –
– filter – off
– frequency – 100 kHz
– all gain – 0 dB.
5.4.4.5 Balance the flaw detector with the receiver cable switched off.
5.4.4.6 Connect the cable to the flaw detector’s receiver. Activate the output
voltage of the signal generator.
Acceptance criterion
2426
The bandwidth lays within the range of 40 Hz to 22 MHz.
Methods of testing
5.4.5.1 Assemble the circuit as per Figure 13.
synchronizatio Trig
n In Generator
Flaw detector
receiver Output
R1=50
Ohm
Switch to the «Settings» menu and set the «Expert» mode, then go back to
the working menu.
Switch off the «Circle» type ALARM frame at the flaw detector.
2526
- frequency – 100 kHz.
5.4.5.7 Switch on the output voltage at the signal generator, as a result the
input signal will be displayed in the form of a point on the flaw detector screen.
5.4.5.8 Record the phase value measured with the flaw detector to Table
12 (readings of the flaw detector’s measurer of phase) and the «Phase»
parameter value.
Acceptance criterion
Deviation in the «Phase» parameter and readings of the phase meter of the
flaw detector at each iteration shall not exceed ±0.5°.
2626
5.4.6 Gain adjustment accuracy
Methods of testing
Generator UZT-
1Attenuator Flaw detector
Output Attenuator output receiver
input
R=50
Ohm
Figure 14 – Circuit assembly for determining the gain adjustment accuracy
ATTENTION!
The voltage is not supplied to the UZT-1!
5.4.6.2 Select the «Test» mode in the «Settings» menu of the flaw detector
and go back to the working menu.
Set the following parameters in the working menu of the flaw detector:
– scale – 1/16
– ECP type – «Refl»
– frequency – 100 kHz
– parameters values «HScale» and «Vscale» – 1;
Switch to the «Settings» menu and set the «Expert» testing mode, go back to
the working menu.
5.4.6.4 Balance the flaw detector with the input voltage switched off.
2726
Note:
It is required to balance the flaw detector after each change of the gain
coefficient!
It is necessary to perform balancing at every measurement point!
5.4.6.5 Connect the cable to the flaw detector’s receiver, as a result the input
signal will be displayed in the form of a ring on the flaw detector screen.
5.4.6.6 Activate the «Circle» type ALARM frame in the flaw detector and set
the radius in such a way that the ALARM triggers at the signal exceedance on the
flaw detector screen.
5.4.6.10 Repeat the steps described in paras 5.4.6.7-5.4.6.9 for each value as
per Table 13.
2826
Table 14 — Measurement data when determining the gain adjustment
accuracy
Preamplification coefficient of Signal amplitude ratio
External attenuation, dB
the flaw detector, dB (difference), dB
0
5
10
15
20
25
30
35
Acceptance criterion
The gain adjustment accuracy shall be not more than ±1.0 dB within a range
of 0 to 50 dB with a step of 5 dB.
Phase adjustment accuracy is the difference between the expected and real
change in the output signal vector, when the phase shift takes place at the phase
adjustment. It is necessary to record the amplitude deviation at the phase
adjustment.
R1=50
Ohm
2926
5.4.7.2 Activate the dual-frequency mode in the “Settings” by putting the
check mark in the relevant item and set «1/16» scale. Activate the «Persistence»
mode and set 0.1s value. Set maximum measurement frequency of 9 kHz.
5.4.7.3 Set the generator frequency of 100 kHz and the generator voltage of
1 V at the flaw detector (as per page one).
5.4.7.4 Switch the flaw detector to the page two by pressing the «SWITCH
FREQ» button and set the generator frequency of 200 kHz, the generator voltage
– 1 V.
5.4.7.7 By changing the frequency value of the signal generator achieve the
signal display in the form of a static point on the flaw detector screen.
ATTENTION!
The quality of frequencies matching (para 5.4.7.7) of the flaw detector and
external signal generator affects further measurements.
5.4.7.10 Achieve the signal display fixation by both channels by pressing the
«RUN/STOP» button of the oscilloscope, as shown in Figure 16.
3026
Flaw detector’s Flaw detector’s
generator signal, with generator signal, with
200 kHz frequency 200 kHz frequency
Zero point
5.4.7.11 When the fixed image on the oscilloscope screen coincides with
one represented in Figure 15, it is requested to record the measured value of the
phase «Р1» to the flaw detector.
5.4.7.12 The point of transition from one flaw detector frequency to another
is the zero point.
5.4.7.13 Activate the cursors at the oscilloscope and select the «Time» type.
5.4.7.14 Set Cursor 1 to the first signal of 100 kHz crossing with the
longitudinal axis that comes after the zero point (for convenience, the first point
of signal transition by the dropping in Figure 16).
5.4.7.15 Set Cursor 2 to the first signal crossing of the external generator
signal with the longitudinal axis that comes after the zero point (for convenience,
the first point of signal transition by the dropping in Figure 16).
ATTENTION!
As shown in figure 15, a dropping signal part of the external signal generator
crosses the longitudinal axis earlier relative to zero point than the signal of the
flaw detector generator crosses its longitudinal axis, that is why the measured
value of the time shift of phases «t1» is recorded with the negative value. In case if
the signal of the flaw detector generator crosses the longitudinal axis earlier than
the signal of the external signal generator, «t1» is to be recorded with the positive
value.
5.4.7.16 Record to the protocol the time shift of phases «t1» of the flaw
detector generator signals and external generator signals.
3126
5.4.7.17 Measure the signal period with 100 kHz frequency «Т» by adjusting
the cursors positions and record to the protocol.
5.4.7.18 Calculate the difference in the signal phases 1 of the flaw
detector’s generator and external signal generator by the formula (7):
(7)
5.4.7.21 Calculate the difference in the signal phases 2 of the flaw
detector’s generator and external signal generator by the formula (8):
(8)
(9)
5.4.7.23 Calculate the difference in the measured values Δφ1 and Δφ2 by
the formula (10):
(10)
5.4.7.24 Record the data that have been calculated in paras 5.4.7.16-5.4.7.23
to Table 14.
5.4.7.25 Calculate the value of the absolute error of the phase adjustment
by the formula (11) and record the data to Table 15
(11)
Readings of the Measured value Measured Phases shift of Deviation of Measured Phase adjustment
flaw detector of the values of the the flaw phases as per value of absolute error,
measurer of generator’s phases time detector and of the readings of the degrees,
3226
shift of the
the flaw
main frequency flaw detector the signal deviation
detector’s
phase, degrees, period of the and of the generator, of phases,
measurer of
Р1, Р2 flaw detector, signal degrees, degrees,
phase, degrees,
µs, Т generator, µs, 1, 2
Р
t1, t2
Acceptance criterion:
Phase adjustment absolute error shall lay within the range of ±10.
5.4.8.1 Assemble the measurement circuit in accordance with Figure 16. For
the operating flaw detector 2.1 (16 pin) the adapter with LEMO 16 — LEMO 12
(Bridge) connectors is used.
Matching load
Pos. 1
Signal Flaw detector
generator
Pos. 2 receiver
Attenuator input
R1=50
Ohm
5.4.8.2 Set the following parameters in the «Settings» menu of the flaw
detector:
– testing mode – «Expert»
– dual-frequency mode – off
5.4.8.3 Set the following parameters in the working menu of the flaw
detector:
– scale – 1/16
– all gain — 0 dB
3326
– frequency – 100 kHz
– filter – none
– generator voltage – 1 V.
5.4.8.5 Balance the flaw detector with the signal generator switched off
(zero input voltage).
5.4.8.8 Record to Table 15 the input signal amplitude value measured with
the measurer of the flaw detector V1.
5.4.8.11 Calculate the input signal amplitude of the flaw detector V2 by the
formula (12) and record the value to Table 15.
(12)
5.4.8.12 Calculate the attenuation quality coefficient using the formula (13)
and record the value to Table 16.
(13)
(14)
3426
Input signal
Measured Measured
amplitude
voltage value voltage value Attenuation
reading of the Attenuation quality
in the toggle Coefficient, dB in the toggle quality
flaw detector, coefficient, dB
position 1 position 2 coefficient
Vx , V
V1, V V2, V
40
Acceptance criterion:
Maximum flaw detector noise correspond to the input voltage equivalent to the
maximum residual output signal when the input signal is supplied with
maximum bandwidth to the flaw detector.
Methods of testing
R1=50
Ohm
5.4.9.2 Balance the flaw detector with “zero” input voltage Uin.
5.4.9.3 Set the following parameters in the working menu of the flaw
detector:
- frequency – 100 kHz
- filter – off
- scale – 1/16
- ECP type – “Refl”.
3526
- voltage (in accordance with one measured in para 5.3.1 (maximum
permissible input voltage).
5.4.9.5 Supply the maximum permissible input voltage from the signal
generator Vmaxin to the receiver input of the flaw detector (gain equals to Gmin =1).
5.4.9.6 Fill in Table 16 with the received value of the input signal amplitude
Vmax out, measured by the flaw detector.
5.4.9.7 Disconnect the signal generator from the flaw detector’s receiver.
5.4.9.8 Set the gain of 30 dB at the flaw detector (maximum gain Gmax).
Flaw detector
Oscilloscope
receiver
Figure 19 – Circuit for determining the RMS value of the voltage measured
with the oscilloscope
5.4.9.10 Record the maximum RMS value of the voltage Vnoiseout measured
with the oscilloscope to Table 17.
5.4.9.11 Calculate the equivalent noise value at the input Veq using the
formula (15) and record the data to Table 16.
(15)
Acceptance criterion
3626