Soil Sample Analysis UV-VIS Spec
Soil Sample Analysis UV-VIS Spec
1. SCOPE
1.1 This method determines the Soil Organic Content (SOC) based on the amount of humic acid present in the
soil sample by using Ultraviolet-Visible Spectroscopy (UV-Vis).
1.2 UV-Vis uses a beam of light with a wavelength of 300 nanometers to pass through a sample in a cuvette.
The amount of light absorbance of the sample is measured and used to determine the SOC.
2. DEFINITIONS
2.1 Blank–Solution that consists of reagents without soil sample.
2.2 Constant Weight-Soils are oven-dried at a temperature of 104 ± 5°F, such that they will not lose more than
0.1% moisture after 4 hr. of drying. Verify constant weight by measuring the moisture content from weighing
a sample before and after consecutive 4-hr. drying periods.
2.3 Cuvette–A small, transparent tube of square cross section sealed at one end, made of fused quartz and
designed to hold sample solution for spectroscopic analysis.
2.4 Filtrate Solution–Soil and reagents that has passed through a filter.
2.6 Ultraviolet-Visible Spectroscopy (UV-Vis)—The process of measuring the soil organic content by the use of
UV-Vis light absorption properties from soil extracts.
3. APPARATUS
3.1 Balance, Class B in accordance with Tex-901-K.
3.2 Bottles, 1 L, glass or polyethylene for storage of 1N hydrochloric acid and sodium pyrophosphate solutions
(minimum of two required).
3.4 Crusher.
3.5 Cuvettes, 10 mm path length, capable of measuring at a wavelength of 300 nm (minimum of seven required).
3.7 Funnel.
3.15 Stopwatch.
5. MATERIALS
5.1 Hydrochloric Acid solution – 1.0N (Certified).
5.2 Humic acid powder – laboratory grade, CAS number 1415-93-6, used to prepare standards with organic
contents of 0.5, 1.0, and 1.5%.
6. PROCEDURES
6.1 This test procedure does not claim to address the safety concerns associated with its use. It is the
responsibility of the user of this test procedure to establish the appropriate safety, health, and environmental
practices and determine the applicability of regulatory limitations before use.
6.2.1 Obtain a representative soil sample in accordance with Tex-100-E. Sample a minimum of 5 lb. of soil at
approximately in the middle of the proposed treatment layer.
6.2.2 Dry the sample in an oven at 104 ± 9°F to constant weight, as defined in Section 2.2. Allow the sample to
cool to room temperature. Alternatively, air-dry the sample to constant weight.
6.2.3 Crush and grind the entire dried sample to pass the No. 4 sieve.
6.2.6 Use a spatula and stir the pulverized sample. Weigh three individual soil test samples at 0.10 ± 0.01 g per
sample, place each sample into individual centrifuge tubes and label them appropriately.
6.3 Ensure all lab equipment used when preparing reagents, standards, blank and test samples are clean and
free of any residue before use.
6.4.1.1 Pour approximately 500 mL of Hydrochloric Acid solution from Section 5.1 into an empty wash bottle using a
funnel and label it appropriately.
6.4.2.1 Pour 500 mL of distilled or deionized water into an empty 1,000 mL volumetric flask using a funnel.
6.4.2.2 Weigh 44.6 ± 0.1 g of Sodium Pyrophosphate Decahydrate from Section 5.3 and place it into the flask.
6.4.2.3 Weigh 10.0 ± 0.1 g of Sodium Hydroxide from Section 5.4 and place it into the flask.
6.4.2.4 Add additional distilled or deionized water into the flask until it reaches the 1,000 mL mark. Stir the solution
until the chemicals have dissolved.
6.4.2.5 Pour approximately 500 mL of Sodium Pyrophosphate solution into an empty wash bottle using a funnel and
label it appropriately.
6.4.2.6 Pour remaining solutions into individual polyethylene bottles and label them appropriately. Store the bottles
for future testing.
Note 2-Chemicals should not be stored near heat sources such as laboratory ovens or in direct sunlight .
6.5.1 Label three individual 500 mL volumetric flasks as 0.5%, 1.0%, and 1.5% respectively.
6.5.2 Pour 100 mL of Hydrochloric Acid solution from Section 6.4.1 into each flask.
6.5.4 Prepare the 0.5% standard sample. Weigh 0.01 ± 0.001 g of humic acid powder from Section 5.2 and
carefully place it into the appropriate flask not to lose any powder.
6.5.5 Prepare the 1.0% standard sample. Weigh 0.02 ± 0.001 g of humic acid powder and carefully place into the
appropriate flask not to lose any powder.
6.5.6 Prepare the 1.5% standard sample. Weigh 0.03 ± 0.001 g of humic acid powder and carefully place into the
appropriate flask not to lose any powder.
6.5.7 Stir the solution in each flask until the chemicals have dissolved.
6.5.8 Label three syringes from Section 3.16 as 0.5%, 1.0%, and 1.5% respectively.
6.5.9 Use the appropriate syringe and draw a minimum of 5 mL from each standard sample.
6.5.10 Attach a new syringe filter from Section 3.17 to each syringe. Use a new syringe filter for each syringe.
6.5.12 Pour remaining solutions into individual polyethylene bottles and label them appropriately. Store the bottles
for future testing.
Note 3-Chemicals should not be stored near heat sources such as laboratory ovens or in direct sunlight.
6.6.1 Pour 5 mL of hydrochloric acid solution from Section 6.4.1 into a graduated cylinder.
6.6.3 Repeat Section 6.6.1 and pour 5 mL into each test sample prepared in Section 6.2.
6.6.4 Vigorously shake each centrifuge tube of soil and Hydrochloric Acid solution by hand for 10 sec. at 1-min.
intervals, five times.
6.6.5 Temporarily store each centrifuge tube in a vertical position after shaking.
6.6.6 Add 20 mL of Sodium Pyrophosphate solution prepared from Section 6.4.2 into a graduated cylinder.
6.6.8 Repeat Section 6.6.6 and pour 20 mL into each centrifuge tube containing the test samples prepared in
Section 6.2.
6.6.9 Vigorously shake each centrifuge tube of soil and Sodium Pyrophosphate solution by hand for 10 sec. at 1-
min. intervals, five times.
6.6.10 Store each centrifuge tube in a vertical position after shaking for a minimum of 15 min. to allow the soil
samples to settle and the air bubbles to dissipate.
6.6.11 Label four additional clean syringes from Section 3.16 as ‘Blank’ and for each test sample prepared in
Section 6.2.
6.6.12 Use the appropriate syringe and draw a minimum of 5 mL from each centrifuge tube in Section 6.6.10.
6.6.13 Attach a syringe filter from Section 3.17 to each syringe. Use a new syringe filter for each syringe.
6.6.14 Fill one cuvette for the blank and each test sample.
6.7.1 Prepare the UV-Vis spectrometer for testing at a wavelength of 300 nanometers according to the
manufacturer’s guidelines and recommendations.
6.7.2 Use latex gloves and wipe the sides of each cuvette from Section 7.5.14 using a lint-free wipe to remove
fingerprints and any other residue.
Note 4–Fingerprints and residue will obstruct the light beam from the spectrometer that passes through the
cuvette and will produce inaccurate test results.
6.7.3 Visually inspect the cuvettes for bubbles and particulates. When bubbles and particulates are present, do not
test the cuvette. Prepare a new cuvette sample.
6.7.4 Place a cuvette into the UV-Vis spectrometer in the order of blank; 0.5%, 1.0%, and 1.5% standard samples;
and test samples.
6.7.5 Determine the absorbance value of each cuvette and enter the results into the SiteManager template in
Section 4.1. The template will graph the absorbance values from each cuvette, determine an R-square value,
and calculate the soil organic content of each test sample.
6.7.6 When the R-square value of the graph is less than 0.9500, prepare and test additional blank and standard
soil samples to improve the best-fit line to achieve the minimum R-square value.
Note 5- Refer to the manufacturer’s guidelines and recommendations for the UV-Vis spectrometer when an
R-square value of 0.9500 or greater cannot be achieved.
7. CALCULATIONS
7.1 Calculate the soil organic content of each test sample:
𝑦−𝑏
𝑆𝑂𝐶 =
𝑚
Where:
𝑆𝑂𝐶 = 𝑆𝑜𝑖𝑙 𝑂𝑟𝑔𝑎𝑛𝑖𝑐 𝐶𝑜𝑛𝑡𝑒𝑛𝑡, %
𝑦 = 𝑎𝑏𝑠𝑜𝑟𝑏𝑎𝑛𝑐𝑒
𝑏 = 𝑦 𝑖𝑛𝑡𝑒𝑟𝑐𝑒𝑝𝑡
7.2